Philips 200 User Manual 2

REF 1054728
1072919
JH 6/7/10
CLINICAL MANUAL
CLINICAL MANUAL
FOR CLINICIAN’S USE ONLY
Accessing Prescription Setting Screens
WARNING
The information on this page is ONLY for health care professionals. Remove this page from the manual before giving the manual to the patient!
Full Menu Access Mode
The ventilator has two levels of menu access, Full and Limited. Full Menu Access allows you to alter all available settings. Limited Menu access permits the user to alter only those prescription parameters that aect patient comfort, such as Rise Time, Flex, and Ramp Start Pressure, if those parameters are available as part of the prescription. The ventilator defaults to Full Menu Access mode.
After accessing the Main Menu, if the device is in Limited Menu Access mode, you can use the following Setup key sequence to enter Full Menu Access mode and be able to change prescription settings:
• Press the Down button and the Alarm Indicator/Audio
Pause button simultaneously for several seconds. This will temporarily place the device in Full Menu Access mode.
When you perform this key sequence from the Monitor screen, the Main Menu screen appears and an audible indicator sounds indicating you are now in Full Menu Access mode.
When you perform this key sequence when the airow is o, the Setup screen appears and an audible indicator sounds.
You can go into the Options menu and permanently change the Menu Access setting to Full Menu Access. Otherwise, the device will return to the Menu Access mode stored in the setting once you exit the menu screens or if one minute passes without pressing any device buttons. If you are in the Setup mode, and an SD card is in the device, “Write Event Log to SD Card” will appear in the menu.
Note: The Full Menu Access key sequence can be performed either from the Power O screen or from the Monitor screen.
Note: When the airow is o and AC power is removed from the device for more than 5 minutes, the device will enter a low power mode to save battery life. When the device is in low power mode, the Setup key sequence will be ignored. Press the Start/Stop button, or connect AC power, or insert an SD card to exit the low power mode.
Note: Philips Respironics recommends that after you are nished changing prescription settings, and you give the device to the patient, you set
the device back to Limited Menu Access mode so patients cannot change their prescription settings.
REF 1054728
1072919
JH 6/7/10
© 2010 Koninklijke Philips Electronics N.V. All rights reserved.
Trilogy200
clinical manual
Table of Contents
FOR CLINICIAN’S USE ONLY Accessing Prescription Setting Screens ................................................... i
Chapter 1. Introduction....................................................................................................................................... 1
Package Contents ................................................................................................................. 1
Intended Use .......................................................................................................................... 2
Warnings and Cautions ......................................................................................................3
Warnings ......................................................................................................................... 3
Cautions ..........................................................................................................................8
Notes ..............................................................................................................................10
i
Contraindications ...............................................................................................................11
System Overview ................................................................................................................11
Symbols ..................................................................................................................................12
Front Panel ...................................................................................................................12
Rear and Side Panels .................................................................................................12
How to Contact Philips Respironics .............................................................................13
Chapter 2. System Description ........................................................................................................................15
Front Panel Features ..........................................................................................................15
Buttons ..........................................................................................................................15
Visual Indicators .........................................................................................................16
Display Screen .............................................................................................................16
Side and Rear Panel Features .........................................................................................17
Table of Contents
ii
Chapter 3. Modes, Features, and Alarms......................................................................................................19
Therapy Modes ....................................................................................................................19
Breath Types.................................................................................................................20
Therapy Mode Table .................................................................................................21
Pressure Control Ventilation Therapy Modes ...................................................22
Volume Control Ventilation Therapy Modes ....................................................27
Therapy Mode Features ...................................................................................................30
Flex Comfort Feature ................................................................................................30
Ramp ..............................................................................................................................31
Rise Time .......................................................................................................................32
AVAPS Feature .............................................................................................................32
Flow Pattern Types ....................................................................................................33
Sigh Feature .................................................................................................................35
Dual Prescription Feature .......................................................................................35
Triggering ....................................................................................................................36
BTPS Compensation..................................................................................................41
Ventilator Alarms ................................................................................................................41
Loss of Power Alarm ..................................................................................................41
Ventilator Inoperative Alarm .................................................................................41
Ventilator Service Required Alarm .......................................................................41
Check Circuit Alarm ..................................................................................................42
Low Circuit Leak Alarm ...........................................................................................42
High Expiratory Pressure Alarm ...........................................................................42
Low Expiratory Pressure Alarm ............................................................................42
High Internal Oxygen Alarm .................................................................................42
Circuit Disconnect Alarm ........................................................................................43
Apnea Alarm ...............................................................................................................43
High Vte Alarm ............................................................................................................43
Low Vte Alarm ............................................................................................................43
Trilogy200 clinical manual
High Vti Alarm .............................................................................................................44
Low Vti Alarm .............................................................................................................44
High Respiratory Rate Alarm .................................................................................44
Low Respiratory Rate Alarm ..................................................................................44
High Inspiratory Pressure Alarm ..........................................................................44
Low Inspiratory Pressure Alarm ...........................................................................45
High Minute Ventilation Alarm ............................................................................45
Low Minute Ventilation Alarm ..............................................................................45
Low Battery Alarm .....................................................................................................46
High Temperature Alarm .......................................................................................46
Replace Detachable Battery Alarm ......................................................................46
Ventilator Service Recommended Alarm ..........................................................46
AC Power Disconnected Alarm .............................................................................47
Keypad Stuck Alarm .................................................................................................47
Battery Discharging Stopped due to Temperature Info Message ............ 47
iii
Battery Not Charging due to Temperature Info Message ..........................47
Battery Not Charging Info Message ....................................................................47
Check External Battery Info Message .................................................................48
Battery Depleted Info Message ...........................................................................48
External Battery Disconnected Info Message ..................................................48
Detachable Battery Disconnected Info Message ...........................................48
Start On Battery Info Message ..............................................................................48
Card Error Info Message .........................................................................................48
Chapter 4. Ventilator Setup ...............................................................................................................................49
Position the Device ............................................................................................................50
Install the Air Filter .............................................................................................................50
Supply Power to the Device............................................................................................50
Using AC Power ..........................................................................................................51
Using DC Power ..........................................................................................................52
Device Power Source Indicators ...........................................................................54
Table of Contents
iv
Battery Disposal..........................................................................................................56
First Time Use ..............................................................................................................56
Connect the Breathing Circuit to the Ventilator ......................................................57
Connect a Water Trap .......................................................................................................60
Connect Supplemental Oxygen (Optional) ..............................................................61
Connect the Remote Alarm (Optional) .......................................................................62
Chapter 5. Viewing and Changing Settings ................................................................................................63
Keypad Lock Feature .........................................................................................................63
Accessing the Startup and Monitor Screens ............................................................64
Monitor Screen Indicators ......................................................................................66
On-Screen Button Panel ..........................................................................................71
Navigating the Menu Screens ........................................................................................71
Changing and Viewing Settings in Full Menu Access Mode ...............................72
Changing the Device Settings and Alarms .......................................................73
Device Settings Common to All Therapy Modes ...................................73
Additional Settings Specic to Therapy Modes .....................................78
Viewing and Changing Options Menu Items ..................................................86
Viewing the Alarm Log ............................................................................................89
Trilogy200 clinical manual
Continuous Positive Airway Pressure (CPAP) Mode ..............................78
Spontaneous (S) Mode ....................................................................................80
Spontaneous/Timed (S/T) Mode .................................................................82
Timed (T) Mode .................................................................................................82
Pressure Control (PC) Mode...........................................................................83
Pressure Control Synchronized Intermittent Mandatory
Ventilation (PC-SIMV) Mode ..........................................................................83
Control Ventilation (CV) Mode .....................................................................84
Assist Control (AC) Mode ................................................................................85
Synchronized Intermittent Mandatory Ventilation
(SIMV) Mode........................................................................................................85
Viewing the Event Log .............................................................................................90
Viewing Device Information ..................................................................................90
Updating Prescriptions Using the SD Card ...............................................................91
Changing and Viewing Settings in Limited Menu Access Mode .......................94
Activating Your Primary or Secondary Prescription ......................................95
Viewing and Changing My Settings Menu Items ...........................................96
Connecting the Ventilator to the Patient ...................................................................98
Chapter 6. Ventilator Alarms ............................................................................................................................99
Audible and Visual Alarm Indicators .........................................................................100
Audio Pause and Alarm Reset Features ................................................................... 105
What to Do When An Alarm Occurs ..........................................................................106
Alarm Summary Table .................................................................................................... 107
Chapter 7. Cleaning and Maintenance .......................................................................................................121
Cleaning the Ventilator .................................................................................................. 121
Cleaning and Replacing the Air Inlet Filter .............................................................122
Replacing the Air Inlet Path Foam .............................................................................123
v
Cleaning the Patient Circuit .........................................................................................124
Cleaning Instructions (Reusable Circuits) ...................................................... 124
Preventive Maintenance ...............................................................................................126
Chapter 8. Troubleshooting ............................................................................................................................129
Chapter 9. Accessories ......................................................................................................................................133
Adding a Humidier ....................................................................................................... 133
Adding Supplemental Oxygen to the Device ....................................................... 133
Using a Remote Alarm Unit..........................................................................................134
Using a Nurse Call System ............................................................................................ 135
Using a Secure Digital (SD) Card ................................................................................135
Using the Philips Respironics DirectView Software ............................................136
Using the Optional In-Use Bag ...................................................................................136
Traveling with the System ............................................................................................ 137
Airline Travel ...................................................................................................................... 137
Table of Contents
vi
Chapter 10. System Checkout Procedures ................................................................................................139
Tools Required .................................................................................................................. 139
Visual Inspection ..............................................................................................................139
Initial Setup ........................................................................................................................140
Settings and Alarms Tests ............................................................................................. 140
Battery Function Verication ....................................................................................... 148
Alarm and Event Log Clean-Up ..................................................................................150
Results .................................................................................................................................150
Chapter 11. Technical Specications ...........................................................................................................151
Chapter 12. Glossary .........................................................................................................................................155
Chapter 13. EMC Information ........................................................................................................................161
Index ........................................................................................................................................................................165
Limited Warranty .................................................................................................................................................169
Trilogy200 clinical manual
ap
Trilogy200
clinical manual
1. Introduction
This chapter provides an overview of the Trilogy200 device.
Package Contents
The Trilogy200 system may include the following components. Some components are optional accessories that may not be packaged with the device.
1
In-Use Bag
Trilogy200
Detachable
Battery
Universal Porting Block
(pre-installed)
O2 Inlet Quick Connects
AC Power Cord
Whisper Swivel II
Secure Digital
(SD) Card
(pre-installed)
Reusable Gray Foam Filters
Clinical Manual
Exhalation Valve Line (pre-assembled to AED)
Proximal Pressure Line (pre-assembled to AED)
Flow Sensor
Package Contents
Disposable Active Circuit, No Water Trap
Flexible Tubing
Active Exhalation Device (AED)
Disposable Passive Circuit, No Water Tr
Flexible Trach Adapter (Not Shown)
Chapter 1 Introduction
2
Intended Use
The Philips Respironics Trilogy200 system provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Trilogy200 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.).
The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. It is not intended to be used as a transport ventilator.
The system is recommended to be used only with various combinations of Philips Respironics-approved patient circuit accessories, such as patient interface devices, humidiers, water traps, and circuit tubing.
Trilogy200 clinical manual
Warnings and Cautions
Caution: U.S. federal law restricts this device to sale by or on the order of a physician.
Warnings
A warning indicates the possibility of injury to the user or operator.
Patient Monitoring Prior to placing a patient on the ventilator, a clinical assessment should
be performed to determine:
• The device alarm settings
• Needed alternative ventilation equipment
• If an alternative monitor (i.e., an alarming Pulse Oximeter or
Respiratory Monitor) should be used
Alternative Ventilation
Patient Disconnect Protection
For ventilator dependent patients, always have alternate ventilation equipment, such as a back-up ventilator, manual resuscitator, or similar device, available.
Ventilator dependent patients should be continuously monitored by qualied personnel. These personnel should be prepared to provide alternate therapy in the event of ventilator failure or inoperative equipment.
For ventilator dependent patients, do not rely on any single alarm to detect a circuit disconnect condition. The Low Tidal Volume, Low Minute Ventilation, Low Respiratory Rate, and Apnea alarms should be used in conjunction with the Circuit Disconnect and Low Peak Inspiratory Pressure alarms.
Test the operation of the circuit disconnect function daily and whenever a change is made to the patient circuit. An increase in circuit resistance can prevent proper operation of some alarms.
Speaking valves, Heat Moisture Exchangers (HMEs), and lters create additional circuit resistance and may aect the performance of alarms chosen for circuit disconnect protection.
Do not set the Low Peak Inspiratory Pressure alarm too low, or the system may not detect large circuit leaks or a patient disconnect.
3
Chapter 1 Introduction
4
Personnel Qualications
Modes of Ventilation
SD Card Prescription Changes
Electrical Interference
Battery Back-up Power
Trilogy200 is a restricted medical device designed for use by
Respiratory Therapists or other trained and qualied caregivers under the supervision of a physician.
The prescription and other device settings should only be changed on the order of the supervising physician.
The operator of the ventilator is responsible to read and understand this manual before use.
The device can provide therapies typically associated with both ventilator dependent and non-dependent patients. The mode of ventilation, circuit type, and alarm strategies should be chosen after a clinical evaluation of each patient’s needs.
C-Flex, Bi-Flex, and AVAPS are intended for use by adult patients.
When you change the device prescription, alarms, and other settings using the SD card, Trilogy200 requires that the caregiver review and verify the changes prior to the changes being used by the device. The caregiver or health care professional is responsible to ensure that the prescription settings are correct and compatible with the patient after using this feature. Installing the wrong prescription for a particular patient may result in improper therapy, lack of appropriate safety monitoring, and risk of death or injury to the patient.
This device is intended for use in the electromagnetic environment specied in Chapter 13 of this manual. The user of this device should make sure it is used in a compatible environment.
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated using the information provided in Chapter 13 of this manual.
The internal battery is NOT intended to serve as a primary power source. It should only be used when other sources are not available or briey when necessary; for example, when changing power sources.
The ventilator has a two-stage low battery alarm. The medium priority alarm indicates that approximately 20 minutes of operation remain, and the high priority alarm indicates that less than 10 minutes of operation remain. Actual run time may be more or less than this and varies with battery age, environmental conditions, and therapy.
Immediately seek an alternate power source when the “Low Battery” alarm appears. Complete power failure and loss of therapy is imminent.
Trilogy200 clinical manual
5
Operating and Storage Temperatures
Do not use this device if the ambient temperature is warmer than 40˚ C (104˚ F). If the device is used at room temperatures warmer than 40˚ C, the temperature of the airow may exceed 43˚ C. This could cause system alarms, thermal irritation, or injury to the patient’s airway.
Bacteria Filter Philips Respironics recommends that a main line outlet bacteria lter
(Part Number 342077) be used whenever the device is used for invasive therapy or if the ventilator may be used on multiple patients.
Patient Circuits (General)
The ventilator should only be used with patient interfaces (e.g., masks, circuits and exhalation ports) recommended by Philips Respironics. Proper operation of the device, including alarms, with other circuits has not been veried by Philips Respironics and is the responsibility of the health care professional or respiratory therapist.
When adding any components to the breathing system, the ow resistance and dead space of the added components such as humidiers, speaking valves, Heat Moisture Exchangers (HMEs) and lters should be carefully considered in relation to the potential for adverse eects on the patient’s ventilatory management and device alarms.
Passive Circuits An exhalation port is required when using a passive circuit.
For the passive circuit, at low expiratory pressures, the ow through the exhalation port may be inadequate to clear all exhaled gas from the tubing – some rebreathing may occur. Rebreathing of exhaled air for longer than several minutes can in some circumstances lead to suocation.
Active Circuits Only use the active exhalation devices designed for Trilogy200. Philips
Respironics has not veried proper operation of other active exhalation devices, and their use may result in improper or unsafe device operation.
With active exhalation circuits, the exhalation device must be operating properly for the ventilator to deliver therapy. The exhalation device should be inspected on a daily basis and replaced whenever necessary.
System Checkout Do not use the ventilator on a patient until a system checkout has been
performed. See Chapter 10 of this manual. To make sure the device is operating properly at start-up, always verify
that the audible tone sounds and the alarm LEDs light red and then yellow momentarily. Contact Philips Respironics or an authorized service center for service if these indications do not occur at start-up.
Chapter 1 Introduction
6
Remote Alarms When using a remote alarm, make sure you fully test the remote alarm
connector and cable by verifying that:
– Annunciated alarms on the ventilator are also
annunciated on the remote alarm.
– Disconnecting the remote alarm cable from the
ventilator or from the remote alarm results in an alarm notication at the remote alarm.
The remote alarm should be tested daily.
Oxygen When administering xed-ow supplemental oxygen, the oxygen
concentration may not be constant. The inspired oxygen concentration will vary, depending on the pressures, patient ows and circuit leak. Substantial leaks may reduce the inspired oxygen concentration to less than the expected value. Appropriate patient monitoring should be used, as medically indicated, such as an alarming pulse oximeter.
This device DOES NOT alarm for loss of the low ow oxygen supply. This device is intended to be connected to a low ow (0-15 l/min)
oxygen source such as an oxygen concentrator or other oxygen source equipped with a pressure regulator (set to 50 PSI or less) and a ow regulator/meter.
Do not connect the device to an unregulated or high pressure oxygen source.
The device may result in incorrect ow and tidal volume measurements and improper operation of related alarms if you add low ow oxygen directly into the patient circuit or mask instead of using the oxygen inlet on the back of the ventilator.
Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of an open ame.
If oxygen is used with the device, the oxygen ow must be turned o when the device is not in use. Explanation of the Warning:
When the device is not in operation and the oxygen ow is left on, oxygen delivered into the tubing may accumulate within the device’s enclosure.
Fire or Explosion The ventilator should not be operated in the presence of ammable
gasses. This could cause a re or explosion.
Trilogy200 clinical manual
Alarms Respond immediately to any alarm. It may indicate a potentially
life-threatening condition. Refer to the Alarms and Troubleshooting chapters for more information.
Visually monitor the patient and ventilator at all times during an Alarm Silence period. Allowing alarms to continue without intervention may result in harm to the patient.
If the high priority “Low Internal Battery” message appears, immediately connect the ventilator to an alternate power source. If no alternate power source is available, immediately place the patient on an alternate source of ventilation.
If the “Ventilator Inoperable” alarm occurs, immediately place the patient on an alternate source of ventilation.
You should not rely on any single alarm to detect a circuit disconnect condition. The Low Tidal Volume, Low Minute Ventilation, Low Respiratory Rate, and Apnea alarms should be used in conjunction with the Circuit Disconnect alarm.
Make sure the alarm volume is set loud enough to be heard by the caregiver. Consider the use of a remote alarm.
Trilogy200 oers the following circuit type selections:
• Passive
• Active Flow
• Active PAP (Proximal Airway Pressure )
7
Improperly Functioning Ventilator
The Passive circuit type provides an ESTIMATE of Vte.
Only the Active Flow circuit type directly measures exhaled tidal volume (Vte).
The Active PAP circuit type DOES NOT measure Vte and only provides for an indication of the delivered tidal volume (Vti).
If you notice any unexplained changes in the performance of the device, if it is making unusual sounds, if the device or detachable battery are dropped, if water is spilled into the enclosure, or if the enclosure is cracked or broken, discontinue use and contact Philips Respironics or an authorized service center for service.
Chapter 1 Introduction
8
Maintenance Follow the service recommendations provided in Chapter 7 of this
manual. Periodically inspect electrical cords, cables, and the detachable battery
pack for damage or signs of wear. Discontinue use and replace if damaged.
Repairs and adjustments must be performed by Philips Respironics­authorized service personnel only. Unauthorized service could cause death or injury, invalidate the warranty, or result in costly device damage.
Cleaning
(Refer to Chapter 7 for detailed cleaning instructions.)
To avoid electrical shock, always unplug the power cord from the wall outlet before cleaning the ventilator.
Do not immerse the device in any uids or spray the device with water or cleaners. Clean the device with a cloth dampened with an approved cleaner.
If the device has been exposed to rain or dampness, dry the device including the area around the power cord connection with the power cord disconnected from the device before applying AC power.
Cautions
A caution indicates the possibility of damage to the device.
Storage The internal and detachable batteries will self-discharge in storage. If it
is desired to keep the batteries fully charged (for example, as a back­up ventilator), plug the device into AC power for about eight hours every 16 days. Alternatively, the ventilator may be left continuously connected to AC power without battery degradation.
Allowing the batteries to fully discharge will not harm the batteries or lose device settings, but may require a longer battery charge time prior to use.
Trilogy200 clinical manual
9
Operating and Storage Temperatures
The device may only be operated at temperatures between 5˚ C and 40˚ C (41˚ F and 104˚ F).
Do not operate the device in direct sunlight or near a heating appliance because these conditions can increase the temperature of the airow delivered to the patient.
Prolonged operation or storage at elevated temperatures may reduce the service life of the battery and other internal components of the ventilator.
The ventilator has an internal and detachable Lithium-Ion Battery. Do not expose the device or detachable battery to temperatures above 40˚ C (104˚ F) during use, or above 60˚ C (140˚ F) during storage. This will reduce battery life and may increase the risk of re or damage the battery.
Condensation Condensation may aect operation or accuracy of the device. If the
device has been exposed to either very hot or very cold temperatures during storage, allow it to adjust to ambient temperature before starting therapy.
Air Filter The reusable foam inlet lter is required to protect the ventilator from
dirt and dust. Wash periodically and replace when damaged for proper operation.
Cooling Air Vents Do not block the cooling air vents located on the base and the rear
of the device. This may cause the device to overheat in high ambient temperatures or at high therapy settings.
Battery Life The internal and detachable batteries wear out based on the amount
of use (hours or full charge-discharge cycles). The battery capacity and life are also reduced by operation at higher temperatures.
Detachable Battery Only use the Philips Respironics Trilogy Detachable Battery with the
ventilator.
Cleaning Do not steam autoclave the ventilator. Doing so will destroy the
ventilator. Do not immerse the device in liquid or allow any liquid to enter the
enclosure or inlet lter. Do not spray water or any other solutions directly onto the ventilator. Do not use harsh detergents, abrasive cleaners, or brushes to clean the
ventilator system. Use only cleaning agents and methods listed in this manual.
Chapter 1 Introduction
10
Patient Circuit Exhalation valves, patient circuits, and water traps are shipped clean,
not sterile. Cleaning and disinfection of these parts should follow individual institution processes and conform to guidelines provided by Philips Respironics with each accessory.
External DC Power Do not use the same external battery to operate both the ventilator
and any other equipment such as power chairs. An external battery should only be connected to the ventilator using
the Philips Respironics Trilogy External Battery Cable. This cable is fused, pre-wired, and properly terminated to ensure safe connection to a standard deep-cycle, lead acid battery. Use of any other adapter or cable may cause improper operation of the ventilator.
The ventilator should only be connected to an automotive electrical system using the Philips Respironics Trilogy Automotive Adapter (when available). This adapter is fused, ltered, and designed for safe connection to a standard automotive electrical system. Use of any other adapter or cable may cause improper operation of the ventilator.
Do not operate the ventilator from a car electrical system when starting the vehicle or jump-starting the vehicle. Electrical transients during starting may cause improper operation of the ventilator.
Electrostatic Discharge (ESD)
Do not use antistatic or conductive hoses or conductive patient tubing with the device.
Notes
• This product does not contain natural latex rubber or dry
natural rubber in patient or operator accessible areas or in the air path or breathing circuit.
Trilogy200 clinical manual
Contraindications
If the patient has any of the following conditions, consult their health care professional before using the device in a non-invasive mode:
• Inability to maintain a patent airway or adequately clear
secretions
• At risk for aspiration of gastric contents
• Diagnosed with acute sinusitis or otitis media
• Epistaxis, causing pulmonary aspiration of blood
• Hypotension
System Overview
This ventilator provides both pressure control and volume modes of therapy. The device can provide non-invasive or invasive ventilation. It can be used to provide total therapy to patients as they progress from non-invasive to invasive ventilation.
11
When prescribed, the device provides numerous special features to help make patient therapy more comfortable. For example, the ramp function allows you to lower the pressure when trying to fall asleep. The air pressure will gradually increase until the prescription pressure is reached. Additionally, the Flex comfort feature provides increased pressure relief during the expiratory phase of breathing.
The ventilator can be operated using several dierent power sources, including an internal Lithium-Ion battery. This battery is automatically used when the detachable Lithium-Ion battery pack, external Lead Acid battery, or AC power are not available.
Chapter 1 Introduction
12
O
2
Symbols
The following symbols appear on the device.
Front Panel
Symbol Description
Therapy Start/Stop
Alarm Indicator/Audio Pause
~
AC Power Indicator
Rear and Side Panels
Symbol Description
~
AC Power Connector
Secure Digital (SD) Card Slot
Serial Port Connector
Remote Alarm Connector
Ethernet Connector
DC Power Connector
Oxygen Inlet
Consult accompanying instructions for use.
Type BF Applied Part
Class II (Double Insulated)
Trilogy200 clinical manual
Drip Proof Equipment
For Airline Use. Complies with RTCA-D0160F section 21, category M.
How to Contact Philips Respironics
To have your device serviced, contact Philips Respironics Customer Service department at 1-724-387-4000 or 1-800-345-6443.
13
Chapter 1 Introduction
14
Trilogy200 clinical manual
Trilogy200
clinical manual
2. System Description
This chapter describes the front and rear panel device controls and features.
15
2
Front Panel Features
The front panel contains the control buttons, visual indicators, and display screen.
Buttons
The following buttons are included on the front panel of the device.
1. Start/Stop Button
This button turns the airow on or o, starting or stopping
therapy.
2. Alarm Indicator and Audio Pause Button
This button serves two purposes: it temporarily silences
the audible portion of an alarm, and it also acts as an alarm indicator. When silencing an alarm, if the cause of the alarm is not corrected, the alarm sounds again after one minute. Each time the button is pressed, the alarm silence period resets to one minute. See Chapter 6 for more information.
3. Up/Down Button
This button allows you to navigate the display menu and edit
device settings.
1
3
4
Front Panel Controls and Display Screen
Note: When you start therapy, the display backlight and the backlights on the buttons turn on, the red and yellow alarm LEDs turn on momentarily, and an audible indicator sounds to indicate that therapy has started. The Startup screen appears on the display.
WARNING
To make sure the device is operating properly at start-up, always verify that the audible tone sounds and the alarm LEDs light red and then yellow momentarily. Contact Philips Respironics or an authorized service center for service if these indications do not occur at start-up.
Chapter 2 System Description
5
16
4. Left and Right Buttons
These buttons allow you to select display options or perform
certain actions specied on-screen.
Visual Indicators
Several power and alarm indicators appear on the front panel.
5. AC Power LED In the lower right corner of the front panel, a green LED (~)
indicates that AC power is applied to the device. This light remains on as long as adequate AC power is available.
6. Keypad Backlight LEDs
The Start/Stop, Up/Down, and Left/Right buttons all have a white
LED that lights up if the keypad backlight is turned on in the device Options menu. See Chapter 5 for more information.
7. Red Alarm LED
On the Alarm Indicator/Audio Pause button, a red light ashes to
indicate a high priority alarm.
8. Yellow Alarm LED
On the Alarm Indicator/Audio Pause button, a yellow light
ashes to indicate a medium priority alarm. A solid yellow light indicates a low priority alarm.
Display Screen
The display screen allows you to view settings, system status information, real-time patient data, alarms, and logs. You can also modify certain settings on the display screen.
See Chapter 5 for more information on viewing and modifying device settings.
Trilogy200 clinical manual
Note: See Chapter 6 for more information about high, medium, and low priority alarms.
Side and Rear Panel Features
The ventilator’s side and rear panels contain the following connectors and features, shown at right.
1. AC Power Inlet
You can plug the AC power cord into this connector, located on
the right side of the ventilator.
17
2
3
2. Breathing Circuit Connection
The breathing circuit connector is located on the right side of
the device. You can connect your circuit tubing system here. See Chapter 4 for details.
3. Exhalation Porting Block
The porting block used here depends on the type of exhalation
device you are using. The Universal Exhalation Porting Block is shown here. See Chapter 4 for more information.
4. Secure Digital (SD) Card Slot
On the left side of the device is a slot for the optional SD Card.
You can have the patient record usage and therapy information from the device on the SD card.
1
Right Side Panel
4
Left Side Panel
Chapter 2 System Description
18
5. Serial Connector
You can use this connector to connect the device to a computer
running PC Direct or Sleepware software or to other Philips Respironics devices such as Alice 5 and AOM. Use the Trilogy RS232 Serial Cable to connect the ventilator to the external device or computer.
6. Remote Alarm/Nurse Call Connector
If you are using an optional remote alarm or nurse call system
with the ventilator, you can connect the Philips Respironics remote alarm adapter cable or nurse call adapter cable to this connector.
7. Ethernet Connector (when available)
You can connect a PC or router to this connector to upload
therapy information to a secure web site so you can review therapy information remotely or remotely troubleshoot and service the device.
8. External Battery Connector (DC Power Inlet)
You can connect an external, stand-alone lead acid battery here,
using the Trilogy External Battery cable.
9. Oxygen (O2) Inlet Connector
11
12
Rear Panel
10
8
5
6
9
7
If using low ow supplemental oxygen, connect the oxygen
source to this connector using one of the O2 Inlet Quick Connects provided with the device.
10. Air Inlet and Filter
Insert the lter supplied with the device into the air inlet.
11. Detachable Battery Pack Slot
If you are using the Philips Respironics Lithium-Ion detachable
battery pack to power the device, attach it here.
12. Cord Retainer
Secure the power cord using the cord retainer to prevent
someone from accidentally disconnecting the power cord. See Chapter 4 for more information.
Trilogy200 clinical manual
Trilogy200
clinical manual
3. Modes, Features, and Alarms
Therapy Modes
The device provides Pressure Control Ventilation (PCV) and Volume Control Ventilation (VCV) for non-invasive and invasive patients.
Pressure Control ventilation delivers a prescribed pressure to the patient according to set breath rate and set inspiration time parameters. This means that each breath is controlled so that a prescribed amount of pressure is delivered to the patient. The device oers six dierent Pressure Control modes of operation:
19
• CPAP – Continuous Positive Airway Pressure
• S – Spontaneous Ventilation
• S/T – Spontaneous/Timed Ventilation
• T – Timed Ventilation
• PC – Pressure Control Ventilation
• PC-SIMV – Pressure Controlled Synchronized Intermittent
Mandatory Ventilation
Volume Control ventilation delivers a prescribed inspired tidal volume to the patient according to set breath rate and set inspiratory time parameters. This means that each breath is controlled so that a prescribed tidal volume is delivered to the patient. The device oers three dierent Volume Control modes of operation:
• AC – Assist Control Ventilation
• CV – Control Ventilation
• SIMV – Synchronized Intermittent Mandatory Ventilation
Chapter 3 Modes, Features, and Alarms
20
Breath Types
There are four breath types that apply to the Volume Control and Pressure Control ventilation therapy modes:
• Spontaneous
• Mandatory
• Assisted
• Sigh
Spontaneous Breath
A Spontaneous breath is triggered by the patient. Breaths are initiated by the patient’s inhalation eort, and air delivery is controlled based on the current pressure or volume setting. Breaths are terminated by either the ventilator settings or by the patient’s exhalation eort, depending on the mode selected.
Mandatory Breath
A Mandatory breath (or machine breath) is completely controlled by the ventilator. The ventilator controls both the beginning (triggering) and end (cycling) of the inspiratory phase.
Assisted Breath
An Assisted breath is controlled by both the patient and the ventilator. Breaths are initiated by the patient’s eort and air delivery is controlled by the current pressure or volume settings. Volume Assisted breaths will deliver the prescribed Tidal Volume within the prescribed Inspiratory Time. Pressure Assisted breaths will deliver the prescribed Inspiratory Pressure for the prescribed Inspiratory Time. Breaths are terminated when the Inspiratory Time setting has been reached.
Sigh
A Sigh breath is a breath where 150% of the prescribed volume is delivered. The device will deliver this breath once every 100 Mandatory or Assist breaths when the Sigh setting is enabled. Sigh breaths are only available in volume modes of ventilation.
Trilogy200 clinical manual
Therapy Mode Table
The following table summarizes all of the therapy modes and the settings available in each mode. Some settings in the table are dependent upon other settings. For instance, if the circuit type is set to Active with Flow, then the Flow Trigger Sensitivity, Leak Compensation, and Flow Cycle settings will display.
Note: Pressure Support, referred to in the table below and later in this manual, is dened as IPAP - EPAP or Pressure ­PEEP (PC-SIMV).
Therapy Modes
CPAP S S/T T PC PC-SIMV CV AC SIMV
Dual Prescription
Circuit Type √ CPAP √ IPAP √ AVAPS (On, O)* √ IPAP Max Pressure √ IPAP Min Pressure √ EPAP √ Pressure √ Pressure Support (PS) √ PEEP √ Tidal Volume √ Breath Rate √ Inspiratory Time √ Trigger Type* √ Flow Trigger Sensitivity √ Leak Compensation √ Flow Cycle Sensitivity √ Ramp Length √ Ramp Start Pressure √ Flex * √ Rise Time √ Flow Pattern
Therapy Parameters
Sigh √ Circuit Disconnect √ Apnea √ Apnea Rate √ High Vte* √ Low Vte* √ High Vti* √ Low Vti* √ High Minute Ventilation √ Low Minute Ventilation √ High Respiratory Rate √ Low Respiratory Rate √ High Inspiratory Pressure √ Low Inspiratory Pressure
* Trigger Type, AVAPS, and Flex settings are only available with the Passive circuit type. High Vte and Low Vte settings are only available for Passive and Active Flow circuit types. High Vti and Low Vti settings are only available with the Active PAP circuit type. Leak compensation is only available with the Active Flow circuit type.
21
Chapter 3 Modes, Features, and Alarms
22
Pressure Control Ventilation Therapy Modes
Pressure Control ventilation modes deliver a prescribed pressure to the patient.
Continuous Positive Airway Pressure (CPAP) Mode
In the Continuous Positive Airway Pressure (CPAP) mode, the device delivers a continuous pressure to the patient at all times. All breaths in this mode are Spontaneous breaths.
Spontaneous (S) Mode
In Spontaneous (S) mode, the device delivers bi-level pressure support. This mode provides only spontaneous breaths. In this mode, an Inspiratory Positive Airway Pressure (IPAP) is delivered during inhalation and a lower Expiratory Positive Airway Pressure (EPAP) is delivered during exhalation. The following illustration describes these concepts.
PRESSURE
26
20
10
6
Enter S mode
5 10 15 20
Trilogy200 clinical manual
IPAP = 26 cm H2O
EPAP = 6 cm H
PS = IPAP-EPAP = 20 cm H
5 seconds 4 seconds 8 seconds
1 1 1
O
2
O
2
= Pressure Supported
1
Spontaneous Breath
1
S Mode
TIME
Spontaneous/Timed (S/T) Mode
In Spontaneous/Timed (S/T) mode, the device delivers bi-level pressure support. This mode provides Spontaneous and Mandatory breaths. A Mandatory breath is delivered if the patient does not spontaneously breathe within the prescribed Breath Rate (BPM) setting. This ensures that the patient receives a minimum number of breaths per minute. In this mode, an IPAP is delivered during inhalation and a lower EPAP is delivered during exhalation. The duration of a Spontaneous breath is determined by the patient eort. The duration of a Mandatory breath is determined by the inspiratory time setting. The following illustration describes these concepts.
23
PRESSURE
26
20
10
6
Enter S/T mode
IPAP = 26 cm H
EPAP = 6 cm H
Rate = 10 BPM
PS = IPAP-EPAP = 20 cm H
O
2
O
2
6 seconds
1 1
O
2
6 seconds 6 seconds
= Pressure Supported
1
Spontaneous Breath
= Mandatory Breath
2
(note longer inspiratory
time)
2
5 10 15 20
S/T Mode
1
TIME
Chapter 3 Modes, Features, and Alarms
24
Timed (T) Mode
In Timed (T) mode, the device delivers bi-level pressure support. This mode delivers Mandatory breaths only. A Mandatory breath is delivered according to the prescribed BPM and inspiratory time settings. This also means that the ventilator will not respond to patient eort. In this mode, an IPAP is delivered during inhalation and a lower EPAP is delivered during exhalation. The following illustration describes these concepts.
PRESSURE
26
20
10
6
IPAP = 26 cm H2O
EPAP = 6 cm H
Rate = 10 BPM
PS = IPAP-EPAP = 20 cm H
O
2
O
2
6 seconds 6 seconds 6 seconds 6 seconds
2 2 22
Enter T mode
5 10 15 20 25
= Mandatory Breath
2
TIME
T Mode
Trilogy200 clinical manual
Pressure Control (PC) Mode
In Pressure Control (PC) mode, the device delivers bi-level pressure support. This mode delivers Assist and Mandatory breaths. This mode is identical to S/T mode, except that all breaths have a xed inspiratory time.
25
PRESSURE
26
20
10
6
IPAP = 26 cm H
EPAP = 6 cm H
Rate = 10 BPM
PS = IPAP-EPAP = 20 cm H
6 seconds
3 3
Enter PC mode
5 10 15 20
O
2
O
2
O
2
6 seconds 6 seconds
3
2
= Mandatory Breath
2
= Assist Breath
3
TIME
Pressure Control - Synchronized Intermittent Mandatory
Ventilation (PC-SIMV) Mode
The Pressure Control – Synchronized Intermittent Mandatory Ventilation (PC-SIMV) mode provides Spontaneous, Assist, and Mandatory breaths. This mode uses a time window to decide what type of breaths should be delivered. This time window is the maximum time between breaths according to the set Breath Rate.
When you enter PC-SIMV mode, the time window is started. If the patient does not trigger a breath, the ventilator will provide a Mandatory breath when the time expires and then begin the process again. Mandatory breaths deliver the Pressure setting during inhalation and a lower Peak End Expiratory Pressure (PEEP) during exhalation.
PC Mode
Chapter 3 Modes, Features, and Alarms
26
During the time window, if patient eort is detected, either a Spontaneous or an Assist breath is delivered for the rst eort. If the last breath delivered was Mandatory, then a Spontaneous breath is delivered for the rst patient eort of the window. If the last breath delivered was an Assist or Spontaneous breath, then an Assist breath is delivered for the rst patient eort of the window. Assist breaths deliver the Pressure setting during inhalation and the PEEP setting during exhalation.
During the time window, if patient eort continues to be detected (after the rst patient triggered breath), Spontaneous breaths are delivered for the remainder of the window. The gure below provides example breath patterns in PC-SIMV mode.
PRESSURE
26
20
10
6
Pressure = 26 cm H2O
Pressure Support = 10 cm H
PEEP = 6 cm H
Rate = 5 BPM
12 seconds
S S
2
Enter PC-SIMV mode
O
2
12 seconds 12 seconds 12 seconds
3
O
2
111
= Spontaneous
1
Breath
= Mandatory Breath
2
= Assist Breath
3
M MM M
= Mandatory Window
M
= Spontaneous Window
S
5 10 15 20 25 30 35 40 45
Sample Breath Patterns in
2
TIME
PC-SIMV Mode
Trilogy200 clinical manual
Volume Control Ventilation Therapy Modes
Volume Control ventilation modes deliver a prescribed inspired tidal volume according to a set Breath Rate and a set Inspiratory Time.
Control Ventilation (CV) Mode
In Control Ventilation (CV) mode, the device delivers volume control therapy. This mode provides only Mandatory breaths. A Mandatory breath is delivered according to the prescribed BPM setting for the prescribed Inspiratory Time. This also means that the ventilator will not respond to patient eort. In this mode, the Tidal Volume is delivered during inhalation and PEEP is delivered during exhalation. The following illustration describes these concepts.
27
50
-50
FLOW
0
Enter CV mode
Tidal Volume = 500 mL
Rate = 10 BPM
Inspiratory Time = 1.5 seconds
Flow Pattern = Square
2 2 2
5
= Mandatory Breath
2
6 seconds 6 seconds6 seconds 6 seconds
15 10
20
Control Mode
TIME
Chapter 3 Modes, Features, and Alarms
28
Assist Control (AC) Mode
In Assist Control (AC) mode, the device delivers volume control therapy. This mode provides Assist and Mandatory breaths. An Assist breath is started when there is patient eort, but it is ended when the Inspiratory Time setting has been met. A Mandatory breath is delivered if the patient does not spontaneously breathe within the prescribed BPM setting. This ensures that the patient receives a minimum number of breaths per minute. In this mode, the Tidal Volume is delivered during inhalation and PEEP is delivered during exhalation. The following illustration describes these concepts.
= Mandatory Breath
6 seconds6 seconds 5 seconds
15 10
2
= Assist Breath
3
20
TIME
50
-50
FLOW
0
Tidal Volume = 500 mL
Rate = 10 BPM
Inspiratory Time = 2.0 seconds
Flow Pattern = Square
Enter A/C mode
2 3 2
5
Assist Control Mode
Synchronized Intermittent Mandatory Ventilation (SIMV) Mode
In Synchronized Intermittent Mandatory Ventilation (SIMV) mode, the device delivers both volume control and pressure control therapy. This mode provides Spontaneous, Assist, and Mandatory breaths. This mode uses a time window to decide what type of breaths should be delivered. This time window is the maximum time between breaths according to the set Breath Rate.
When you enter SIMV mode, the time window is started. If the patient does not provide any eort, the ventilator will provide a Mandatory Volume breath when the time expires and then begin the process again. Mandatory breaths deliver the Tidal Volume setting during inhalation and provide a lower Peak End Expiratory Pressure (PEEP) during exhalation.
Trilogy200 clinical manual
During the time window, if patient eort is detected, either a Spontaneous or an Assist breath is delivered for the rst eort. If the last breath delivered was Mandatory, then a Spontaneous breath is delivered for the rst patient eort of the window. If the last breath delivered was an Assist or Spontaneous breath, then an Assist breath is delivered for the rst patient eort of the window. Assist breaths deliver the prescribed Tidal Volume during inhalation and the PEEP setting during exhalation.
During the time window, if patient eort continues to be detected (after the rst patient triggered breath), Spontaneous breaths are delivered for the remainder of the window. Spontaneous breaths deliver the prescribed Pressure Support setting above PEEP during inhalation and PEEP during exhalation. The gure below provides example breath patterns in SIMV mode.
29
Tidal Volume = 500 mL
Breath Rate = 5 BPM
Inspiratory Time = 3.0 seconds
Pressure Support = 10 cm H
PEEP= 6 cm H
PRESSURE
26
20
10
2
6
Enter SIMV mode
5 10 15 20 25 30 35 40 45
O
O
2
2
S S
3
= Pressure Supported
1
Breath
= Mandatory Breath
2
= Assist Breath
3
111
= Mandatory Window
M
= Spontaneous Window
S
M MM M
12 seconds12 seconds12 seconds12 seconds
2
SIMV Mode
TIME
Chapter 3 Modes, Features, and Alarms
30
Therapy Mode Features
The device has several additional features that enhance patient comfort.
Flex Comfort Feature
The device consists of a special comfort feature called Flex. The device provides the Flex feature in CPAP mode and S mode. This feature is only available when Auto-Trak is enabled.
C-Flex
When in CPAP mode, if C-Flex is enabled, it enhances patient comfort by providing pressure relief during the expiratory phase of breathing. In the following diagram, the dashed lines represent normal CPAP therapy in comparison to the bold line representing C-Flex. C-Flex levels of 1, 2, or 3 progressively reect increased pressure relief.
PRESSURE
CPAP
Inhalation Exhalation
1
2
3
TIME
Note: Flex is not available if AVAPS is enabled or if an active circuit is used.
Note: The patient has access to this setting, if Flex is enabled.
Note: C-Flex is not available if CPAP is set to 4 cm H2O.
Note: C-Flex, Bi-Flex and AVAPS are intended for use by adult patients.
C-Flex in Comparison to Traditional CPAP Therapy
C-Flex pressure relief is determined by the C-Flex setting and the amount of patient ow. C-Flex returns to the set pressure by the end of exhalation, when the airway is most vulnerable to closure.
Trilogy200 clinical manual
Bi-Flex
PRESSURE
In S mode, the Bi-Flex attribute adjusts therapy by inserting a small amount of pressure relief during the latter stages of inspiration and during the beginning part of exhalation. In the following diagram, the bold lines represent Bi-Flex in comparison to the dashed line representing normal BiPAP therapy. Bi-Flex levels of 1, 2, or 3 progressively reect increased pressure relief that will take place at the end of inspiration and at the beginning of expiration.
31
Note: Bi-Flex is only available up to 25 cm H2O in S mode.
IPAP
EPAP
Bi-Flex
BiPAP
1
2
3
TIME
Ramp
The device is equipped with a linear ramp function. In CPAP, S, S/T, T, and PC modes, the Ramp feature will reduce the pressure and then gradually increase (ramp) the pressure to the prescription pressure setting so patients can fall asleep more comfortably. The gure below illustrates how the ramp function works.
Bi-Flex in Comparison to Traditional Bi-Level Therapy
Note: IPAP will not ramp below 4 cm H2O.
The Ramp Function
TIME
Chapter 3 Modes, Features, and Alarms
32
Rise Time
Rise Time
In S, S/T, PC, T, PC-SIMV, and SIMV modes, rise time is the amount of time it takes the device to change from the expiratory pressure setting to the inspiratory pressure setting. Rise time levels of 1, 2, 3, 4, 5, or 6 progressively reect slowed response of the pressure increase that will take place at the beginning of inspiration. Adjust the rise time to nd the most comfortable setting for the patient.
PRESSURE
IPAP
EPAP
TIME
AVAPS Feature
Average Volume Assured Pressure Support (AVAPS) is a feature available in the S, S/T, PC, and T modes. It helps patients maintain a tidal volume (VT) equal to or greater than the target tidal volume (Volume setting in the ventilator) by automatically controlling the pressure support (PS) provided to the patient. The AVAPS feature adjusts PS by varying the IPAP level between the minimum (IPAP Min) and maximum (IPAP Max) settings. AVAPS averages VT and changes the PS value gradually. This occurs over several minutes. The rate of change is slow, so that the patient is not aware of breath-to-breath pressure changes.
As patient eort decreases, AVAPS automatically increases PS to maintain the target tidal volume. The IPAP level will not rise above IPAP Max, even if the target tidal volume is not reached. Conversely, as patient eort increases, AVAPS will reduce PS. IPAP will not fall below IPAP Min, even if the target tidal volume is exceeded.
Rise Time
Note: AVAPS is only available if you are using a passive circuit.
Note: C-Flex, Bi-Flex and AVAPS are intended for use by adult patients.
If the Ramp function has been activated, it will take precedence over the AVAPS feature. Once ramp is complete, AVAPS will resume.
Trilogy200 clinical manual
IPAP Max
PRESSURE
IPAP Min
EPAP
VOLUME
Target Volume
15 30 45 60
33
TIME
AVAPS Feature
15 30 45 60
TIME
Flow Pattern Types
Two ow patterns are available in Volume Control ventilation therapy modes:
• Square
• Ramp
Square
With a square wave pattern, airow is generally constant throughout inspiration of the breath.
FLOW
TIME
Square Flow Pattern
Chapter 3 Modes, Features, and Alarms
34
Ramp
With a ramp ow pattern, the airow starts high and decreases throughout inspiration of the breath.
FLOW
50% of Peak Flow
TIME
For the active circuit in volume modes, peak ow is required to be a minimum of 20 l/min. The wave form may be attened when the combination of Inspiratory Time and Tidal Volume set points would result in a ow of less than 20 l/min. Therefore, for some settings, a Ramp ow pattern may provide a pattern that more closely resembles a Square ow pattern.
Ramp Waveform Pattern
Trilogy200 clinical manual
Sigh Feature
FLOW
The sigh feature is available for Volume ventilation modes only. When the sigh feature is enabled, the ventilator delivers a sigh breath in place of every 100th mandatory or assisted breath delivered regardless of the mode of operation (i.e., AC, CV, and SIMV). The Sigh breath is delivered using a volume equal to 150% of the set volume that was in eect when the breath was initiated.
Dual Prescription Feature
35
Example of Sigh Feature
The device provides a dual prescription feature that allows you to enter a primary prescription and a secondary prescription for the patient if needed. For example, you can set a primary daytime prescription and secondary nighttime prescription. See Chapter 5 for more information on the dual prescription feature.
Chapter 3 Modes, Features, and Alarms
Note: Both prescriptions must use the same circuit type.
36
Triggering
The device can be set to trigger breaths using the Auto-Trak or Flow Trigger sensitivity features.
Digital Auto-Trak Sensitivity
An important characteristic of the device is its ability to recognize and compensate for unintentional leaks in the system and to automatically adjust its trigger and cycle algorithms to maintain optimum performance in the presence of leaks. This feature is known as Digital Auto-Trak Sensitivity. The following sections examine this function in detail by describing the leak tolerance function and sensitivity.
Leak Tolerance
A microprocessor monitors the total ow of the patient circuit and calculates patient ow values.
A. Leak Estimation: Average and Parabolic
The device uses two leak estimation algorithms. A conservation of mass algorithm is used to compute the average leak for a given pressure support relationship. This average leak is used when large leak variations are present in the system. Average leak is a high estimate during EPAP pressure and a low estimate during IPAP pressure. A better leak estimate, enabled by the digital system, is the parabolic leak algorithm. Parabolic leak is proportional to the square of the patient pressure; therefore, the leak estimate is correlated to the changing patient pressure. Both algorithms include unintentional circuit leak and are averaged over several breaths.
B. Patient Flow
The total circuit ow is comprised of the circuit leak and the patient ow. The calculated patient ow is the total ow minus the circuit leak. Patient ow is a primary input into the triggering and cycling mechanisms.
Trilogy200 clinical manual
Auto-Trak Sensitivity
An essential feature of the device while operating in all modes is its ability to eectively sense spontaneous breathing eorts, which causes the ventilator to trigger to inspiration and cycle to expiration. Because no preset sensitivity threshold can assure patient and machine synchrony with changing breathing eorts and circuit leaks, the device continuously tracks patient breathing patterns and automatically adjusts sensitivity thresholds to ensure optimum sensitivity as breathing patterns change or as circuit leaks change. The algorithms used to ensure optimum sensitivity are the Volume Trigger, Shape Signal, Spontaneous Expiratory Threshold (SET), Flow Reversal, Maximum IPAP Time, and Volume Control Cycle.
Volume Trigger (Expiration to Inspiration):
The volume trigger is one method used to trigger inspiration during spontaneous breathing in all modes except T and CV. The volume trigger threshold is 6 ml of accumulated patient inspiratory volume. When patient eort generates inspiratory flow causing 6 ml of volume, inspiration is triggered.
37
Note: Auto-Trak is only available if you are using a passive circuit.
Shape Trigger/Shape Cycle (Expiration to Inspiration) (Inspiration to Expiration):
The shape trigger/cycle is another method used to trigger inspiration and/or cycle from inspiration to expiration during spontaneous breathing in all modes except T and CV. This method continuously tracks patient inspiratory and expiratory flow and adjusts the spontaneous trigger and cycle thresholds for optimum sensitivity. The Shape Signal appears as a shadow image of the patient’s actual flow. The shape signal functions as a sensitivity threshold at either inspiration or expiration. When the patient’s flow rate crosses the shape signal the unit changes pressure levels. The following gure illustrates how the shape signal is super-imposed onto the actual waveform to trigger and cycle o IPAP. The shape signal is created by osetting the signal from the actual patient flow by 15 l/min and delaying it for a 300 msec period. This intentional delay causes the shape signal to be slightly behind the patient’s flow rate.
Chapter 3 Modes, Features, and Alarms
38
IPAP
A sudden change in patient flow will cross the shape signal, causing the pressure level to change.
Pressure
EPAP
Cycle to
Shape Signal
EPAP Crossover Point
Flow
Estimated Patient Flow
Trigger to IPAP Crossover Point
Tracking the patient’s flow pattern with the Shape Signal provides a sensitive mechanism to trigger to inspiration or cycle to expiration in response to changing breathing patterns and circuit leaks.
Shape Signal
Trilogy200 clinical manual
Spontaneous Expiratory Threshold (Inspiration to Expiration):
IPAP
A second method used to cycle to expiration during spontaneous breathing in all modes except T, CV, AC, and SIMV, is called Spontaneous Expiratory Threshold (SET). The SET rises in proportion to the inspiratory flow rate on each breath. When the SET and actual patient flow value are equal, the unit cycles to expiration.
Pressure
EPAP
Spontaneous Expiratory Threshold
Flow
39
Spontaneous Expiratory Threshold
Flow Reversal (Inspiration to Expiration):
As ow begins to decrease during inspiration, a ow reversal can occur due to a large leak around the mask or because the patient’s mouth is open. When the device senses this ow reversal, it automatically cycles to expiration.
Maximum IPAP Time (Inspiration to Expiration):
The maximum inspiratory time is determined by the adjustment of the Inspiratory time setting. A maximum IPAP time of 3.0 seconds acts as a safety mechanism to limit the time spent in inspiration during spontaneous breathing. For mandatory or assisted breaths, the maximum inspiratory time will equal the Inspiratory time setting up to 5.0 seconds.
Chapter 3 Modes, Features, and Alarms
40
Volume Control Cycle (Inspiration to Expiration) (Only available during Volume Control Therapy)
An Inspiratory Time setpoint limits the time spent in inspiration during breathing in all modes. Once the time limit is reached, the unit automatically cycles to expiration.
Flow Trigger
Flow trigger provides a manual setting that allows for breath initiation and termination based on a set ow trigger sensitivity and ow cycle sensitivity.
Flow Trigger Sensitivity (Expiration to Inspiration):
The ow trigger initiates when the patient’s inspiratory eort creates a ow equal to or greater than the ow trigger sensitivity setting. The method of the ow trigger is dependent upon the circuit type that is chosen.
Leak Compensation:
When using the Passive Circuit conguration, compensation for both the intentional and unintentional leak is included in the triggering method.
When using the Active PAP Circuit conguration, leak compensation is not available.
When using the Active Flow Circuit conguration, ow trigger with leak compensation may be enabled. The default setting when using the Active Flow Circuit is Leak Compensation On. The clinician has the option to turn o leak compensation; however, unintentional leak will not be compensated. Both options measure the ow at the proximal ow sensor.
Flow Cycle Sensitivity (Inspiration to Expiration):
This cycling method is only active if the Flow Trigger has been selected for the Trigger Type. As ow begins to decrease during inspiration, if the patient ow is less than the ow cycle sensitivity setting, the device will cycle to expiration.
For example: if the ow cycle sensitivity setting is set to 75%, when the ow has decreased by 25% of the peak ow, the device will cycle to the EPAP/PEEP level.
Trilogy200 clinical manual
Note: Enabling Leak Compensation when using the Active Flow Circuit conguration only aects triggering and does not aect tidal volume delivery or Vte measurement.
BTPS Compensation
All ows and volumes used in Trilogy are expressed in BTPS - Body Temperature atmospheric Pressure Saturated with H2O.
All pressures are expressed relative to atmospheric pressure.
Ventilator Alarms
This section describes all of the ventilator alarms and informational messages in order of priority, from high priority alarms to low priority alarms and nally informational messages. Refer to Chapter 6 for more information on alarms.
Loss of Power Alarm
This is a high priority alarm. It may occur when a complete power failure has occurred and power was lost while the device was providing therapy. This may happen if the internal battery was the only power source in use and was completely depleted.
41
Ventilator Inoperative Alarm
This is a high priority alarm. It occurs when the ventilator detects an internal error or a condition that may aect therapy. The device will shut down if the cause of the failure indicates that the device cannot deliver therapy safely. If the device can deliver therapy at a limited level, the device will continue to deliver limited therapy.
Ventilator Service Required Alarm
This is a high priority alarm. It occurs when the device cannot perform to specication, a backup safety feature is compromised, or the delivery of therapy is compromised. The device continues to operate (possibly in a reduced capacity mode). If the problem is not corrected, the device will generate a reminder message once per hour until the issue is corrected. Additionally, if therapy is stopped, a reminder message will immediately appear when therapy is turned on again.
Chapter 3 Modes, Features, and Alarms
42
Check Circuit Alarm
This is a high priority alarm. It occurs when the device detects a problem with the patient circuit, such as pinched or detached tubing, water condensation in the proximal pressure lines, or problems with the active exhalation device.
Low Circuit Leak Alarm
This is a high priority alarm that only occurs with the passive circuit. It occurs when the system detects a problem with the leak device in the passive circuit.
High Expiratory Pressure Alarm
This is a high priority alarm. It occurs when the delivered pressure exceeds the target patient pressure during the expiratory phase by 5 cmH2O. This may be due to pinched tubing or the patient having a fast breath rate. The device continues to operate. The alarm will automatically terminate when the delivered pressure comes within 5 cmH2O of the target patient pressure during the expiratory phase.
Low Expiratory Pressure Alarm
This is a high priority alarm. It occurs when the delivered pressure is 5 cmH2O or more below the target patient pressure during the expiratory phase. The device continues to operate. The alarm will automatically terminate when the delivered pressure comes within 5 cmH2O of the target patient pressure during the expiratory phase.
High Internal Oxygen Alarm
This is a high priority alarm. It occurs when there is a leak in the internal air delivery system that allows oxygen to build up inside the device. The alarm is generated when the internal oxygen concentration reaches 5% above ambient levels.
Trilogy200 clinical manual
Circuit Disconnect Alarm
This is a high priority alarm. It occurs when the breathing circuit is disconnected or has a large leak. The device continues to operate. The alarm will automatically terminate when the circuit is reconnected or the leak is xed.
Apnea Alarm
This is a high priority alarm. It occurs when the patient has not triggered a breath within the time specied in the apnea alarm setting. The device continues to operate. The alarm will automatically terminate when two consecutive patient breaths are detected that meet the apnea alarm time setting.
High Vte Alarm
This is a high priority alarm. It occurs when the estimated exhaled tidal volume is greater than the High Vte alarm setting for three consecutive breaths. The device continues to operate. The alarm will automatically terminate when a breath occurs in which the exhaled tidal volume does not reach the High Vte alarm setting.
43
WARNING
You should not rely on any single alarm to detect a circuit disconnect condition. The Low Tidal Volume, Low Minute Ventilation, Low Respiratory Rate, and Apnea alarms should be used in conjunction with the Circuit Disconnect alarm. The Apnea alarm is only intended for spontaneously breathing patients.
Low Vte Alarm
This is a high priority alarm. It occurs when the estimated exhaled tidal volume is lower than the Low Vte alarm setting for three consecutive breaths. The device continues to operate. The alarm will automatically terminate when a breath occurs in which the exhaled tidal volume exceeds the Low Vte alarm setting.
When AVAPS is on, this alarm will occur when the calculated tidal volume is less than 90% of the target tidal volume setting. The alarm will automatically terminate when a breath occurs in which the exhaled tidal volume is equal or greater to 90% of the target tidal volume setting.
Chapter 3 Modes, Features, and Alarms
44
High Vti Alarm
This is a high priority alarm. It occurs when the delivered tidal volume is greater than the High Vti alarm setting for three consecutive breaths. The device continues to operate. The alarm will automatically terminate when a breath occurs in which the delivered tidal volume does not reach the High Vti alarm setting.
Low Vti Alarm
This is a high priority alarm. It occurs when the delivered tidal volume is less than the Low Vti alarm setting for three consecutive breaths. The device continues to operate. The alarm will automatically terminate when a breath occurs in which the delivered tidal volume exceeds the Low Vti alarm setting.
High Respiratory Rate Alarm
This is a high priority alarm. It occurs when the respiratory rate is greater than the High Respiratory Rate alarm setting. The device continues to operate. The alarm will automatically terminate when the measured respiratory rate is less than the High Respiratory Rate alarm setting.
Low Respiratory Rate Alarm
This is a high priority alarm. It occurs when the respiratory rate is less than the Low Respiratory Rate alarm setting. The device continues to operate. The alarm will automatically terminate when the measured respiratory rate is greater than the Low Respiratory Rate alarm setting.
High Inspiratory Pressure Alarm
This alarm occurs in several stages and escalates from an audible beep for the rst two occurrences to a high priority alarm if the problem continues. It is detected dierently for volume and pressure therapy modes.
For volume modes, the alarm will sound if the measured patient pressure exceeds the High Inspiratory Pressure setting specied by the clinician. The alarm will automatically terminate when the peak inspiratory pressure is less than or equal to the High Inspiratory Pressure alarm setting.
Trilogy200 clinical manual
For pressure modes, the alarm occurs when the delivered pressure exceeds the target patient pressure by 5 cmH2O or more during the inspiratory phase. The device will automatically cycle to the expiratory phase and continue to operate. The alarm will automatically terminate when the delivered pressure falls within 5 cmH2O of the target patient pressure during the inspiratory phase.
Low Inspiratory Pressure Alarm
This is a high priority alarm. It is detected dierently for volume and pressure therapy modes.
For volume modes, the alarm will sound if the measured patient pressure is less than the Low Inspiratory Pressure setting specied by the clinician. The alarm will automatically terminate when the peak pressure at the end of the breath is greater than or equal to the Low Inspiratory Pressure alarm setting.
For pressure modes, the alarm occurs when the delivered pressure is 5 cmH2O or more below the target patient pressure during the inspiratory phase. The alarm will automatically terminate when the delivered pressure comes within 5 cmH2O of the target patient pressure during the expiratory phase.
45
High Minute Ventilation Alarm
This alarm is a high priority alarm. It occurs when the patient’s minute ventilation is greater than the High Minute Ventilation alarm setting. The device continues to operate. The alarm will automatically terminate when the calculated minute ventilation is less than the High Minute Ventilation alarm setting.
Low Minute Ventilation Alarm
This alarm is a high priority alarm. It when the patient’s minute ventilation is less than the Low Minute Ventilation alarm setting. The device continues to operate. The alarm will automatically terminate when the calculated minute ventilation is greater than the Low Minute Ventilation alarm setting.
Chapter 3 Modes, Features, and Alarms
46
Low Battery Alarm
The Low Battery alarm occurs when the last battery available is low or nearly depleted. This alarm occurs in two stages. When approximately 20 minutes of battery run time remains, a medium priority alarm is generated, and the device continues to operate. If no action is taken and the battery continues to deplete, the alarm escalates to a high priority alarm when approximately 10 minutes of battery run time remains.
High Temperature Alarm
This alarm occurs when the estimated patient airstream temperature or the ventilator internal temperature is too high. The alarm occurs in several stages. The ventilator continues to operate. Internal fans are started when the medium priority alarm is generated. If the condition causing the high temperature is not corrected and the temperature continues to rise, the alarm will escalate to the high priority alarm.
Replace Detachable Battery Alarm
The Replace Detachable Battery alarm occurs when the detachable battery is nearing the end of its useful life or a failure in the detachable battery that prevents it from charging or discharging has been detected.
WARNING
Immediately seek an alternate power source when the “Low Battery” message appears. Complete power failure and loss of therapy is imminent.
WARNING
The ventilator has a two-stage low battery alarm. The medium priority alarm indicates that approximately 20 minutes of operation remain, and the high priority alarm indicates that less than 10 minutes of operation remain. Actual run time may be more or less than this and varies with battery age, environmental conditions, and therapy.
The alarm occurs in several stages, from low to high priority. The device continues to operate when the alarm is the low priority alarm. If the alarm is reset without removing the battery, the alarm will be regenerated once every hour until the detachable battery is removed. The device continues to operate, the detachable battery is not used, and the power source is switched to the next available power source if the alarm is the high priority alarm.
Ventilator Service Recommended Alarm
This is a medium priority alarm. It occurs when the device has detected an error, but the error will not aect device performance or safety. The device continues to operate. Therapy and safety are not compromised.
Trilogy200 clinical manual
If the problem is not corrected, the device will generate a reminder message once per day, or whenever power is cycled, until the issue is corrected. Additionally, if the device is powered o, a reminder message will immediately appear when the device is turned on again.
AC Power Disconnected Alarm
This is a medium priority alarm. It occurs when the AC power source was lost, and the device has switched to an alternate power source (either a detachable or external battery, if connected, or the internal battery if no other source is available). The device continues to operate. If AC power returns, the ventilator will beep, but no message will appear on the display.
Keypad Stuck Alarm
This is a low priority alarm. It occurs when a key becomes lodged inside the case of the device.
Battery Discharging Stopped due to Temperature
47
Info Message
This info message occurs when the detachable or internal battery becomes overheated while providing power for the device. The device continues to operate. The battery is not used and the power source is switched to the next available power source.
Battery Not Charging due to Temperature Info Message
This info message occurs when the detachable or internal battery becomes too hot while charging or the device was in too cold an environment before charging started. The device continues to operate. Battery charging stops until the battery cools or warms suciently.
Battery Not Charging Info Message
This info message occurs when the device has detected an error condition with the battery that prevents it from accepting a charge. The device continues to operate. Battery charging stops.
Chapter 3 Modes, Features, and Alarms
48
Check External Battery Info Message
This info message occurs when a bad connection exists to the external battery or the external battery failed. The device continues to operate using power from the detachable battery, if available, or the internal battery.
Battery Depleted Info Message
This info message occurs when the aected battery is fully depleted. The device continues to operate using the next available power source.
External Battery Disconnected Info Message
This info message occurs when the external battery power source is lost and the device has switched to an alternate power source (either a detachable battery, if connected, or the internal battery if no other source is available). If external battery power returns, the ventilator will beep, but no message will appear on the display.
Detachable Battery Disconnected Info Message
This info message occurs when the detachable battery power source is lost and the device has switched to an alternate power source (the internal battery if no other source is available). If detachable battery power returns, the ventilator will beep, but no message will appear on the display.
Start On Battery Info Message
This info message indicates that the ventilator has started on battery power and no AC power is available. The device operator should verify that this is what is wanted.
Card Error Info Message
This info message occurs when an unusable SD card is inserted into the ventilator. The device continues to operate but data cannot be logged onto the SD card.
Trilogy200 clinical manual
Trilogy200
clinical manual
4. Ventilator Setup
49
This chapter provides instructions on how to assemble the ventilator. It includes the following setup information:
• Properly positioning the device
• Installing the air lter
• Supplying power to the device
• Connecting the breathing circuit
• Connecting a water trap (optional)
• Connecting a remote alarm (optional)
WARNING
Ventilator dependent patients should be continuously monitored by qualied personnel. These personnel should be prepared to provide alternate therapy in the event of ventilator failure or inoperative equipment.
WARNING
Do not use the ventilator on a patient until a system checkout has been performed. See Chapter 10 of this manual.
WARNING
For ventilator dependent patients, always have alternate ventilation equipment, such as a back-up ventilator, manual resuscitator, or similar device, available.
CAUTION
Do not operate the ventilator at temperatures below 5° C (41° F) or above 40° C (104° F).
Chapter 4 Ventilator Setup
50
Position the Device
Place the ventilator on a at, level surface. Or, as an alternative, you can place the device in the optional in-use bag. The in-use bag is required for portable use in order to properly secure the ventilator.
Do not operate the ventilator while on its side, upside-down, or in any other orientation. Make sure that the air inlet on the back of the device is not blocked. If you block the air ow around the device, the ventilator may not work properly.
Install the Air Filter
The device uses a gray foam filter that is washable and reusable. The reusable filter screens out normal household dust and pollens. The filter must be in place at all times when the device is operating. Two reusable gray foam filters are supplied with the device. If the filter is not already installed when you receive the device, you must install the filter before using the ventilator.
CAUTION
The reusable foam inlet lter is required to protect the ventilator from dirt and dust. Wash periodically and replace when damaged for proper operation.
To install the filter, insert the gray foam filter into the filter area as shown.
Supply Power to the Device
The device can operate on AC or DC power. The ventilator accesses power from potential sources in the following order:
• AC Power
• External Battery
• Detachable Battery Pack
• Internal Battery
Trilogy200 clinical manual
Installing the Filter
Note: See Chapter 7 for
information on how to clean and replace the air lter.
51
Using AC Power
An AC power cord is provided with the device.
1. Plug the socket end of the AC power cord into the AC power inlet on the side of the device. To prevent accidental removal of the AC power cord from the device, you may secure the AC power cord to the device using the AC power cord retainer located on the back of the device.
2. Plug the pronged end of the cord into a wall outlet not connected to a wall switch.
3. Ensure that all connections are secure. If AC power is connected correctly and the device is operating properly, the green AC Power LED should be on.
Note: This device is activated (i.e., ‘powered on’) when the AC power cord is connected or any of the three battery sources is available. Pressing the Start/Stop button turns the airflow on or o.
CAUTION
The device may only be operated at temperatures between 5˚ C and 40˚ C (41˚ F and 104˚ F).
WARNING
Periodically inspect the power cord for damage or signs of wear. Discontinue use and replace if damaged.
Note: To remove AC power, disconnect the power supply cord from the electrical outlet.
Connecting the AC Power Cord
Chapter 4 Ventilator Setup
52
4. If desired, secure the power cord using the cord retainer on the back of the device, as shown below. To secure the power cord, use a screwdriver to remove the screw on the cord retainer. Insert the cord as shown into the cord retainer, and then re­attach the cord retainer to the device by re-attaching the screw.
Using DC Power
You can operate the ventilator using an external battery, detachable battery, or the internal battery.
Securing the Power Cord Using the Cord Retainer
External Battery
The ventilator can operate from a 12 VDC deep cycle marine-type (lead acid) battery using the Philips Respironics Trilogy External Battery Cable. This cable is pre-wired and properly terminated to ensure safe connection of an external battery to the ventilator. Battery operating time depends on the characteristics of the battery and usage of the device.
Due to a variety of factors, including battery chemistry, battery age, and use prole, the capacity of the external battery as shown on the device display is only an estimate of the actual remaining capacity.
Refer to the instructions supplied with the Philips Respironics Trilogy External Battery Cable for detailed information on how to operate the device using an external battery.
Trilogy200 clinical manual
CAUTION
Do not use the same external battery to operate both the ventilator and any other equipment such as power chairs.
CAUTION
An external battery should only be connected to the ventilator using the Philips Respironics Trilogy External Battery Cable. This cable is fused, pre-wired and properly terminated to ensure safe connection to a standard deep cycle lead acid battery. Use of any other adapter or cable may cause improper operation of the ventilator.
Detachable Battery Pack
Philips Respironics oers a detachable Lithium-Ion battery pack. To use the detachable battery pack, snap the battery into place on the back of the ventilator, as shown in the following illustration. When the device is not connected to an AC power source or an external battery, the detachable battery will power the device, if attached. The length of time the ventilator will operate on battery power depends on many factors such as device settings, battery charge level, and condition or age of the battery. When fully charged, a new battery can power the ventilator for approximately three hours under typical patient conditions.
Whenever the ventilator is connected to AC power, it will automatically recharge the detachable battery pack. A completely discharged detachable battery will reach 80% charge status within 8 hours, when charging at approximately 23˚ C ambient temperature.
Insert and securely latch the detachable battery into the device as shown below.
CAUTION
The detachable and internal batteries wear out based on the amount of use (hours or full charge-discharge cycles). The battery capacity and life are also reduced by operation at higher temperatures.
CAUTION
Only use the Philips Respironics Trilogy Detachable Battery with the device.
53
Insert
Detachable
Battery
Properly Installed
Detachable
Battery
Attaching the Detachable Battery
CAUTION
Prolonged operation or storage at elevated temperatures may reduce the service life of the detachable or internal battery and other internal components of the ventilator.
Chapter 4 Ventilator Setup
54
One side of the detachable battery has a set of LEDs that indicate the amount of charge left on the battery. You can press the button below the LEDs to view how much charge remains:
LED Battery Capacity
All 5 LEDs are lit 80-100% capacity 4 LEDs are lit 60-79% capacity 3 LEDs are lit 40-59% capacity 2 LEDs are lit 20-39% capacity 1 LED is lit 10-19% capacity 1 LED ashes 1 to 9% capacity 0 LEDs are lit 0% capacity
Internal Battery
The device contains an internal battery that can be used as a back­up power source. It is intended for use during short periods while switching between external power sources, emergency situations, or short durations when the user needs to be mobile. The length of time the ventilator will operate on internal power depends on many factors such as device settings, battery charge level, and condition or age of the battery. When fully charged, a new battery can power the ventilator for approximately three hours under typical patient conditions.
Whenever the ventilator is connected to AC power, it will automatically recharge the internal battery. A completely discharged internal battery will reach 80% charge status within 8 hours when charging at approximately 23˚ C ambient temperature.
Device Power Source Indicators
There are many power source indicators on the device and the display screen. These indicators are described in detail below.
AC Power Indicator
Detachable Battery LEDs
WARNING
The internal battery is NOT intended to serve as a primary power source. It should only be used when other sources are not available, or briey when necessary; for example when changing power sources.
When AC power is applied to the device, the green AC LED indicator (~) on the front of the device lights.
Trilogy200 clinical manual
55
DC Power Indicators
The internal, detachable, and external battery symbols that will display on the Monitoring screen are shown below. The detachable and external battery symbols will only appear on-screen if a detachable or external battery is attached to the device.
Battery Symbol
Internal Battery
Note: Under normal operating conditions, the internal battery symbol will always display on the Monitoring screen. If the symbol appears as an empty red battery on your screen, contact Philips Respironics or an authorized service representative to have your
device serviced.
Detachable Battery
External Battery
There are several DC power indicators that will display on the Monitoring screen to indicate which battery is in use (if applicable), if the batteries are low, charging, or discharged, etc. The following table explains all of the DC power indicators.
DC Power Indicator Description
Battery In Use Indicator A black box will appear around the battery that is in use. For instance, if
the external battery is currently in use, the symbol appears on the Monitoring screen.
Green Fully Charged Battery Indicator
Partially Charged Battery Indicator
When a battery is charged to greater than 90% of its capacity, all of the bars in the battery symbol will appear in green.
When a battery is partially charged, some of the bars in the battery symbol will appear in green, while others will be clear. For instance, if the external battery is 50% charged, the following symbol displays on­screen:
Yellow Low Battery Indicator (Medium Priority)
When the device detects that an in-use battery’s charge is low (has approximately 20 minutes of charge left), the inside of the box surrounding the battery symbol turns yellow. (In addition to the battery indicator on the Monitoring screen, a medium priority alarm message will display indicating “Low Battery.” See Chapter 6 for more information. The yellow indicator is for the last available battery source.
Chapter 4 Ventilator Setup
56
DC Power Indicator Description
Red Low Battery Indicator
Yellow Battery Recharging Symbol ( )
When the device detects that an in-use battery’s charge is nearly depleted (has approximately 10 minutes of charge left), the inside of the box surrounding the battery symbol turns red. In addition to the battery indicator on the Monitoring screen, a high priority alarm message will display indicating “Low Battery.” See Chapter 6 for more information. The red indicator is for the last available battery source.
Whenever AC power is applied to the device, the internal and detachable batteries will recharge as needed. If the internal battery is being recharged, the symbol displays. If the detachable battery is being recharged, the symbol displays.
Battery Disposal
Dispose of the batteries in accordance with local regulations.
First Time Use
When setting up the device for the rst time or after annual service, apply AC power to the ventilator before turning on the blower. Attempting to use the ventilator without rst applying AC power, such as installing a detachable battery pack and starting the blower, will cause the internal battery to be displayed in red as an empty battery. When in this state, the internal battery will not be usable until AC power is applied.
Trilogy200 clinical manual
57
Connect the Breathing Circuit to the Ventilator
Complete the following steps to set up your patient circuit.
1. If you are using a bacteria lter, connect one end of the exible tubing to the outlet of the bacteria lter, and connect the bacteria lter inlet to the breathing circuit connection located on the side of the ventilator. If you are not using a bacteria lter, connect the tubing directly to the device’s breathing circuit connection.
Trilogy Breathing Circuit
Connection
Flexible Tubing
Bacteria Filter
2. Connect the other end of the exible tubing to a separate exhalation device.
a. If your circuit type is Passive and you are using a Philips
Respironics Passive Exhalation Device:
1. Connect the exible tubing to the rotating end of the Passive Exhalation Device.
WARNING
Philips Respironics recommends that a main line outlet bacteria lter (Part Number 342077) be used whenever the device is used for invasive therapy or if the ventilator may be used on multiple patients.
Connecting the Breathing Circuit to the Device
Note: The device can be used
with reusable or disposable circuits. For detailed instructions on how to set up your device using a disposable circuit, refer to the instructions included with the disposable circuit.
2. Connect the other end of the Passive Exhalation Device to the patient interface (e.g., the mask).
Flexible Tubing
Whisper Swivel II
Bacteria Filter
Passive Exhalation Device
Flexible Tubing
Bacteria Filter
Refer to the instructions included with the Passive Exhalation Device for more detailed setup information.
Connecting the Passive Exhalation Device
Note: The device does not
provide circuit compliance compensation in volume modes.
Chapter 4 Ventilator Setup
58
b. If your circuit type is Active PAP:
Active Exhalation Device with Proximal Pressure
1. Connect the exible tubing to an active exhalation device with proximal pressure.
A. Connect the active exhalation device with proximal
pressure to the exible tubing that attaches to the ventilator and to the patient interface (e.g., the tracheostomy tube).
2. Connect the proximal pressure line and the exhalation
valve line to the active exhalation device with proximal pressure and the Universal Porting Block on the device as described below.
A. Attach the proximal pressure line (.476 cm or 3/16”
line) to the proximal pressure port as shown in the next illustration.
B. If not attached, connect the other end of the proximal
pressure line to the Trilogy Universal Active PAP Tube Adapter as shown in the next illustration.
C. Connect the Trilogy Universal Active PAP Tube Adapter
to the top, striped port on the Universal Porting Block on the ventilator.
D. Connect the exhalation valve line (.317 cm or 1/8” line)
to the exhalation valve port on the top of the active exhalation device with proximal pressure.
E. Connect the other end of the exhalation valve line
to the exhalation valve port on the Universal Porting Block.
Note: Passive circuit ventilation provides leak compensation in both volume and pressure modes. Active PAP circuit ventilation does not provide leak compensation. Particular care should be used at low tidal volumes to ensure adequate monitoring of exhaled tidal volume.
WARNING
For ventilator dependent patients, always have alternate ventilation equipment, such as a back-up ventilator, manual resuscitator, or similar device, available.
Note: Make sure that components marked with an arrow are oriented properly. When attaching an active exhalation device with proximal pressure to the patient, make sure the proximal pressure port faces away from the patient.
Note: The symbol appears next to the Exhalation Valve port on the Universal Porting Block to indicate where the active exhalation valve line connects.
Trilogy200 clinical manual
59
Universal Porting Block
Exhalation Valve Port
Proximal Pressure Port
Exhalation Valve Line
Proximal Pressure Line
Trilogy Universal Active PAP Tube Adapter
Exhalation Valve Port on Universal Porting Block
Refer to the instructions included with the Active Exhalation Device for more information.
c. If your circuit type is Active Flow:
1. Connect the Flow Sensor to an active exhalation device as shown below.
Flow Sensor
2. Connect the other end of the exible tubing to the active exhalation device.
A. Connect the active exhalation device to the exible
tubing that attaches to the ventilator.
B. Connect the ow sensor to the patient interface (e.g.,
the tracheostomy tube).
Connecting an Active Device with Proximal Pressure
Connecting the Flow Sensor
Note: If the active exhalation device has the proximal pressure line connected when you are setting up an active ow circuit type, remove the proximal pressure line and cap the proximal pressure port before attaching to the exible tubing.
Note: Make sure that components marked with an arrow are oriented properly.
Chapter 4 Ventilator Setup
60
C. Connect the ow lines and the exhalation valve line to
the Universal Porting Block on the ventilator as shown below.
• Connect the White Striped Flow Line to
the top, striped port on the Universal Porting Block on the device.
• Connect the other Flow Line to the
middle port on the Universal porting Block.
• Connect the Exhalation Valve Line to the
Exhalation Valve port on the top of the active exhalation device, and connect the other end of the line to the Exhalation Valve port on the Universal Porting Block.
White Striped Flow Line
Exhalation Valve Port
Universal Porting Block on Trilogy Device
Note: When using an active exhalation device with proximal ow, ensure that any additional proximal ports are capped.
Note: The symbol appears next to the Exhalation Valve port on the Universal Porting Block to indicate where the active exhalation valve line connects.
Flow Line
Exhalation Valve Line
Exhalation Valve Port on Porting Block
Refer to the instructions included with the active exhalation device for more information.
Connect a Water Trap
If you are using an optional water trap, connect it to the patient circuit according to the manufacturer’s instructions.
Trilogy200 clinical manual
Connecting the Active Exhalation Device with
Proximal Flow Sensor
Connect Supplemental Oxygen (Optional)
t
To attach supplemental oxygen to the device:
1. Connect the oxygen tubing to the Oxygen Inlet Quick Connect. Then, connect the Quick Connect to the Oxygen Inlet on the back of the ventilator.
Oxygen Inlet
Oxygen Inlet
Quick Connec
61
Oxygen Source
Tubing
2. Make sure you press the Oxygen Inlet Quick Connect rmly into the inlet. You should hear a click, and the latch at the top of the inlet will pop up.
This device is only compatible with a low ow oxygen source
providing up to 15 l/min.
Attaching the O2 Tubing
Press Tubing in and Latch Pops Up
CAUTION
Do not connect an unregulated or high pressure oxygen source to this connector on the device.
Chapter 4 Ventilator Setup
62
Connect the Remote Alarm (Optional)
You can use a remote alarm as a nurse call system or in-house remote alarm system. It can generate an alarm at a distance of up to 91 meters (300 feet) from the ventilator. An audible tone sounds and a red light blinks to indicate that an alarm condition exists. The remote alarm will sound when any of the following conditions exist:
• The ventilator is o.
• Any alarm occurs and is not silenced or reset.
Refer to the Accessories chapter of this manual for information on which remote alarm systems are compatible with the device. Refer to the instructions included with the Remote Alarm Adapter Cable for information on how to connect a remote alarm to the ventilator.
WARNING
When using a remote alarm, make sure you fully test the remote alarm connector and cable by verifying that:
- Annunciated alarms on the ventilator are also annunciated on the remote alarm.
- Disconnecting the remote alarm cable from the ventilator or from the remote alarm results in an alarm notication at the remote alarm.
WARNING
The remote alarm should be tested daily.
Trilogy200 clinical manual
Trilogy200
clinical manual
5. Viewing and Changing Settings
63
This chapter explains how to scroll through the ventilation screens and change ventilation settings. It also describes how to connect the ventilator to the patient once the settings are complete.
Keypad Lock Feature
There is a Keypad Lock feature that users can enable from the Options menu. It is intended to prevent accidental changes to device settings. This feature will lock the navigation keys (Up, Down, Stop, Left, and Right keys). If the Keypad Lock is enabled, a Keypad Unlock message will display on the bottom of the screen any time you press one of the navigation keys.
If the keypad is locked, you must unlock it before you can enter the Menu. To unlock the keypad and enter the menu, you must rst hold the Right button for 5 seconds to unlock the keypad. An audible indicator sounds when the keypad is successfully unlocked. Once the display is unlocked, you can enter the Menu as you normally would by pressing the Up button.
Note: There is a keypad lock inactivity time-out period. After you have unlocked the keypad as indicated, the keypad will re-lock after ve minutes of inactivity to prevent someone from accidentally pressing a button and changing any of the settings.
Note: When Keypad Lock is enabled, the Left, Right, and Up/ Down buttons are locked while the ventilator is turned on. The
Alarm Indicator/Audio Pause continues to function normally. The Start/Stop button is locked only when this button is used to stop therapy.
Note: The keypad will automatically unlock if an alarm or informational message occurs and will remain unlocked the entire time alarms are active.
Note: Pressing the Left (Cancel) button will cancel the Keypad Unlock action.
Chapter 5 Viewing and Changing Settings
64
Accessing the Startup and Monitor Screens
1. After you press the button to begin therapy, the Startup screen appears momentarily, indicating the device name and the software version.
2. The next screen that appears is the Monitor screen. The appearance of this screen will vary, depending on how you set up the device. If Detailed View is turned o in the Options menu, your screen will look like the screen shown below.
Monitor Panel
Date and Time
Panel
Status Panel
– The top section of the screen, called the Monitor panel,
shows the therapy mode and, if you set up a dual prescription for the patient, the Prescription indicator appears, indicating Primary or Secondary prescription. The patient breath symbol also displays during a patient-triggered breath, and a bar graph displays the current pressure level.
Note: The symbols that appear on the Monitor screen are described in detail later in this
chapter.
Monitor Screen – Detailed View O
Note: No Soft Buttons display on the Monitor screen if Keypad Lock is enabled.
– The center section of the screen displays the current date
and time.
– The bottom section, called the Status panel, displays certain
symbols that indicate features being used, such as Ramp, as well as battery status.
Trilogy200 clinical manual
If Detailed View is turned on in the Options menu, the Monitor
Measured Settings
screen will look like the screen shown below.
Monitor Panel
65
Panel
Status Panel
This screen contains more detailed information about the
therapy.
– The top Monitor panel contains the Prescription indicator
if a dual prescription exists, the therapy mode, a graph displaying the current pressure, and the current date and time. Additionally, this panel also displays patient pressure, respiratory rate (RR), exhaled tidal volume (Vte), and leak.
– The second panel in Detailed view is the Measured
Settings panel. It provides patient-related data including Peak Inspiratory Pressure (PIP), Minute Ventilation, Peak Inspiratory Flow, Mean Airway Pressure (MAP), and I:E Ratio.
– The third panel is the Status panel and shows the same
information displayed in the Detailed View O screen, including features in use such as Ramp and battery status.
Monitor Screen – Detailed View On
Chapter 5 Viewing and Changing Settings
66
Monitor Screen Indicators
This section describes the following indicators:
• Monitor Panel Indicators
• Measured Settings Panel Indicators
• Status Panel Indicators
Monitor Panel Indicators
Note: Dashes on the display screen indicate that the device was unable to compute the parameters. For example, when the device is rst connected to a patient, the tidal volume, minute ventilation, and leak may be dashed until the device is able to accurately calculate these patient parameters.
All of the indicators that may appear on the Monitor Panel are described in detail in the following table.
Indicator Description
Prescription If you set up a dual prescription for the patient, the words
“Primary” or “Secondary” appear in the top left corner of the panel to indicate which prescription is active.
Therapy Mode The current therapy mode displays at the top of the panel (for
example, CPAP, S, S/T, etc.). If a special feature such as Flex, AVAPS, or Sigh is active, this feature will appear next to the therapy mode.
Date and Time If you are in Detailed view, the current date and time appear
in the top right corner of the panel. (In Detailed View O, they appear in the center panel.)
Patient Breath This symbol displays during a patient-triggered breath.
Airway Pressure Manometer and Peak Pressure Symbol
The manometer (bar graph) displays the airway pressure in the patient circuit at all times. The manometer bar moves to the right as pressure increases during inhalation, and moves to the left as pressure decreases during exhalation. The peak pressure is also indicated on this bar. It is positioned according to the maximum patient pressure reached during each breath. The Peak Pressure symbol appears as a blue bar on the manometer. If a High Inspiratory Pressure alarm occurs, the Peak Pressure symbol changes from blue to red.
Trilogy200 clinical manual
Indicator Description
67
Low Pressure Indicator
L
If you enable a volume therapy mode, this indicator appears below the manometer bar, indicating the low pressure alarm setting.
High Pressure Indicator
H
If you enable a volume therapy mode, this indicator appears below the manometer bar, indicating the high pressure alarm setting.
Pressure This indicator displays the current patient pressure. This only
appears in detailed view.
Respiratory Rate (RR) This indicator displays the measured respiratory rate in Breaths
Per Minute (BPM). This only appears in detailed view.
Exhaled Tidal Volume (Vte)
This indicator displays the estimated exhaled tidal volume in milliliters and reects compensation for BTPS. This only appears in detailed view when Passive Circuit is selected.
Inhaled Tidal Volume (Vti)
This indicator displays delivered tidal volume in milliliters and reects compensation for BTPS. This only appears in detailed view when the Active with PAP Circuit is selected.
Leak This indicator displays the total leak (non-returned ow), between
the unit outlet and the patient, averaged over the previous breath. This only appears in detailed view when the Passive Circuit is selected.
Chapter 5 Viewing and Changing Settings
68
Measured Settings Panel
All of the indicators that may appear on the Measured Settings panel (available only in Detailed view), are described in the following table.
Indicator Description
PIP Peak Inspiratory Pressure displays the maximum pressure
delivered to the patient during the previous breath.
I:E Ratio Displays a comparison of the time spent in inspiration to the time
spent in expiration during the previous breath.
Peak Flow Displays the maximum inspiratory ow delivered to the patient
during the previous breath in l/min BTPS.
MAP Displays the Mean Airway Pressure, which is the weighted
average of pressure in the patient’s airway over 6 breaths.
MinVent Minute ventilation displays the amount of air delivered to the
patient over the last minute in l/min BTPS.
Trilogy200 clinical manual
Status Panel Indicators
All of the indicators that may appear on the Status Panel are described in the following table.
Indicator Description
Indicates that the device is in Full Menu Access mode, which means you can adjust all prescription settings. Philips Respironics recommends that you change the device to Limited Menu Access mode before giving the device to the patient, so patients cannot adjust their prescription settings. Only trained health care
professionals and clinicians should adjust prescription settings.
Displays when a Secure Digital (SD) memory card is inserted in the ventilator.
Displays when the ventilator detects an error with the SD card.
Displays at all times when an external battery is attached to the ventilator. The level of green shading shown in the symbol indicates the battery capacity and will go down as the battery charge level decreases. When the entire symbol is green, the battery is fully charged.
69
Displays at all times when a detachable battery is attached to the ventilator. The level of green shading shown in the symbol indicates the battery capacity and will go down as the battery charge level decreases. When the entire symbol is green, the battery is fully charged.
Displays at all times, indicating the status of the internal battery. The level of green shading shown in the symbol indicates the battery capacity and will go down as the battery charge level decreases. When the entire symbol is green, the battery is fully charged.
Chapter 5 Viewing and Changing Settings
70
Indicator Description
A black box displays around the battery that is currently supplying power to the ventilator when AC power is not available. (In the status panel shown above, the external battery is in use, so the symbol displays.)
A yellow lightning bolt symbol displays with the Detachable or Internal battery symbol to indicate when the battery is charging. (In the status panel shown on the previous page, the detachable battery is being charged, so the symbol displays.)
Displays when the Alarm Indicator/Audio Pause button has been pressed and Audio Pause is active. The alarm is silenced for one minute when the Alarm Indicator/Audio Pause button is pressed.
Displays when the Ramp feature is active.
Note: If a battery in use is very low (less than 20 minutes remaining), the inside of the box surrounding the battery symbol will change to yellow and all of the bar indicators in the battery will be empty. If a battery in use is near depletion (less than 10 minutes remaining), the inside of the box surrounding the battery symbol will change to red, and all of the bar indicators in the battery will be empty. These color changes only occur for the last available battery source.
Trilogy200 clinical manual
On-Screen Button Panel
The illustration below shows the on-screen button panel on the Main Menu screen, in relation to the buttons on the front of the device.
On-Screen
Button
Panel
At the very bottom of the display screen is the on-screen button panel. This panel corresponds with the control buttons on the ventilator:
• The left on-screen button species the action for the Left
button on the device.
• The center on-screen button species the action for the
Up/Down buttons on the device.
• The right on-screen button species the action for the
right button on the device.
71
Example On-Screen Button Panel
Note: The on-screen buttons will vary depending on which screen is displayed and what settings are enabled on your device.
Navigating the Menu Screens
To navigate through all of the menu screens and settings:
• Use the Up/Down button to scroll through the menu
options and settings.
• Use the Left and Right buttons to perform the actions
specied on the screens’ left and right on-screen buttons.
Chapter 5 Viewing and Changing Settings
72
Changing and Viewing Settings in Full Menu Access Mode
Clinicians can view and change settings using the Menu screens when the device is in Full Menu Access mode. To enter the Menu screens from the Monitor screen, press the Up button on the ventilator. The Main Menu screen shown below appears.
Choose from the following selections on the Main Menu screen:
• Safely Remove SD Card - This option will appear if an SD
card is inserted in the ventilator. Select this option when you want to remove the SD card. When the “Remove SD Card” conrmation message appears, remove the card. If you press the left (cancel) button or don’t remove the card within 30 seconds, the conrmation message will close and the ventilator will continue writing to the card.
• Settings and Alarms: View and change prescription
settings and alarms.
• Options: View and change device settings, such as Full or
Limited Access mode, Detailed View, Language, etc.
• Alarm Log: View a list of the 20 most recent alarms that
have occurred.
• Event Log: View a list of all events that have occurred,
such as ventilator setting changes, ventilator inoperative conditions, alarms, etc.
• Information: View detailed information about the device,
such as the device’s software version and serial number.
Note: For some therapy settings, once you reach the highest or lowest setting available, pressing the Up/ Down button again will cycle back through the settings. For the parameters that do not wrap, when you reach the highest or lowest setting possible, a “Limit Reached” message appears in the Menu Banner on-screen.
Main Menu Screen Example
Note: In the example Main Menu screen shown, the 2/6 that appears in the Menu banner indicates that item 2 is highlighted from a total of 6 items in the menu.
Note: If you change a setting but decide you do not want to save it, you can press the Left button to cancel the change.
Trilogy200 clinical manual
73
Changing the Device Settings and Alarms
From the Main Menu screen, use the Up/Down button to highlight the Settings and Alarms menu, and press the Right button to select the menu.
Device Settings Common to All Therapy Modes
Some of the settings on this menu will vary depending on the therapy mode you select. The section below describes all of the settings that are common to all therapy modes.
Dual Prescription Setting
You can turn the dual prescription setting on or o. Enable the setting if you want to create two separate prescriptions for the patient. For instance, you may want to set up a daytime prescription and then a separate nighttime prescription. If you enable this setting, then the menu options on the Main Menu will change to include three new options:
• Switch to Primary/Secondary Settings
• Primary Settings and Alarms
• Secondary Settings and Alarms
Note: In the Options menu described later in this chapter, you can specify the Pressure Units displayed by the device, choosing either cm H2O, hPa, or mBar. The default setting is cm H2O, so this manual uses cm H2O throughout.
The Main Menu screen will look like the screen below.
Main Menu Screen with Dual Prescription Enabled
Chapter 5 Viewing and Changing Settings
74
Mode Setting
You can change the Mode setting on the Settings and Alarms screen to one of the following therapy modes:
• CPAP
• S
• S/T
• T
• PC
• PC-SIMV
• CV
• AC
• SIMV
Circuit Type
There are three circuit types you can select:
• Passive
• Active PAP
• Active Flow
The Passive circuit type uses the Whisper Swivel II passive exhalation device. The Active PAP circuit type uses an active exhalation device with a proximal air pressure sensing connection. The Active Flow circuit type uses an active exhalation device with a proximal ow sensor.
When using the Passive circuit, the ventilator displays estimated patient pressures based on the resistance of the standard patient circuit (Whisper Swivel II with 1.8 meter tubing). Adding accessories to the patient circuit (humidier, water trap, etc.) may cause an increase in circuit resistance and cause the device to display slightly higher pressures than what is actually delivered to the patient.
Note: See Chapter 3 for detailed information about each therapy mode.
Note: Refer to the chart in Chapter 3 to easily review all of the settings available in each therapy mode.
Note: This chapter describes how to enable all of the device settings, including device alarms. For detailed information about each alarm, please refer to Chapter 6.
Note: To change the circuit type, you must be in the Setup screen with the airow turned o. See the Full Menu Access Mode section for more information.
Note: When the Circuit Type setting is set to Passive Circuit, all Ramp Start Pressure settings in all modes will maintain the minimum range.
Note: When the Circuit Type setting is set to Active PAP or Active Flow, the Flex and AVAPS features are unavailable.
With the Active PAP or Active Flow circuit type selected, patient pressure is measured directly and is not aected by any change in circuit resistance.
Trilogy200 clinical manual
The Passive circuit provides leak compensation. When using the Passive circuit in Volume Ventilation, the set Vti is delivered to the patient above the calculated circuit and cu (or mask) leak. This is dierent from traditional active circuit ventilation where the cu (or mask) leak reduces the tidal volume delivered to the patient. Volume ventilation with the Passive circuit delivers an inspiratory tidal volume close to the device setting regardless of leak; this should be considered when transitioning a patient from an active to a passive circuit. With a Passive circuit, Vte is estimated based on the calculated sum of circuit and cu (or mask) leak.
The Active Flow circuit monitors proximal ow and proximal pressure. When using the Active Flow circuit conguration, ow trigger with leak compensation may be enabled. The default setting when using the Active Flow Circuit is Leak Compensation On. The clinician has the option to turn o leak compensation; however, unintentional leak will not be compensated. Both options measure the ow at the proximal ow sensor. Leak compensation is not available in the Active with PAP circuit conguration.
Circuit Disconnect
This setting enables or disables the circuit disconnect alarm. If enabled, an audible alarm will sound when a large, continuous air leak (such as mask removal) has been detected in the circuit.
You can choose O to disable the alarm. Or, you can increase or decrease the setting from 5 to 60 seconds in 5 second increments. For example, a setting of 10 means that the alarm will sound after the circuit has been disconnected for 10 seconds.
75
WARNING
You should not rely on any single alarm to detect a circuit disconnect condition. The Low Tidal Volume, Low Minute Ventilation, Low Respiratory Rate, and Apnea alarms should be used in conjunction with the Circuit Disconnect alarm.
Apnea
This setting enables or disables the apnea alarm. If enabled, an audible alarm will sound when an apnea is detected.
You can choose O to disable the alarm. Or, you can increase or decrease the setting from 10 to 60 seconds in 5 second increments. For example, a setting of 10 means that the alarm will sound if the time between spontaneous breaths exceeds 10 seconds.
Chapter 5 Viewing and Changing Settings
76
Low Vte
This setting enables or disables the Low Vte alarm. The alarm activates when the estimated exhaled tidal volume is less than or equal to this setting. You can choose O to disable this alarm, or you can increase or decrease the setting from 40 ml to 2000 ml in 5 ml increments. It cannot be set higher than the High Vte setting.
When AVAPS is On, the alarm activates when the calculated tidal volume is less than 90% of the target tidal volume setting. This alarm can be set to on or o.
High Vte
This setting enables or disables the High Vte alarm. The alarm activates when the estimated exhaled tidal volume is greater than or equal to this setting. You can choose O to disable this alarm, or you can increase or decrease the setting from 50 ml to 2000 ml in 5 ml increments. It cannot be set lower than the Low Tidal Volume setting, except to be turned o.
Low Vti
This setting enables or disables the Low Vti alarm. The alarm activates when the measured inhaled tidal volume is less than or equal to this setting. You can choose O to disable this alarm, or you can increase or decrease the setting from 40 ml to 2000 ml in 5 ml increments. It cannot be set higher than the High Vti setting.
Note: The High and Low Vte alarms are available when the Passive or Active Flow Circuit is selected.
Note: The High and Low Vti alarms are only available when the Active PAP Circuit is selected.
High Vti
This setting enables or disables the High Vti alarm. The alarm activates when the measured inhaled tidal volume is greater than or equal to this setting. You can choose O to disable this alarm, or you can increase or decrease the setting from 40 ml to 2000 ml in 5 ml increments. It cannot be set lower than the Low Vti setting, except to be turned o.
Trilogy200 clinical manual
Low Minute Ventilation
This setting enables or disables the Low Minute Ventilation alarm. The alarm activates when the calculated minute ventilation is less than or equal to this setting. You can choose O to disable this alarm, or you can increase or decrease the setting from 0.1 l/min to 99 l/min.
High Minute Ventilation
This setting enables or disables the High Minute Ventilation alarm. The alarm activates when the calculated minute ventilation reaches or exceeds this setting. You can choose O to disable this alarm, or you can increase or decrease the setting from 1 l/min to 99 l/min in 1 l/min increments. It cannot be set lower than the Low Minute Ventilation setting except to be turned o.
Low Respiratory Rate
This setting enables or disables the Low Respiratory Rate alarm. The alarm activates when the measured respiratory rate is less than or equal to this setting. You can choose O to disable this alarm, or you can increase or decrease the setting from 4 BPM to 80 BPM in 1 BPM increments. It cannot be set higher than the High Respiratory Rate.
77
Note: You can increase or decrease the Low Minute Ventilation setting in
0.1 l/min increments from
0.1 to 0.99 l/min and 1 l/min increments from 1 to 99 l/min.
High Respiratory Rate
This setting enables or disables the High Respiratory Rate alarm. The alarm activates when the measured respiratory rate reaches or exceeds this setting. You can choose O to disable this alarm, or you can increase or decrease the setting from 4 BPM to 80 BPM in 1 BPM increments. It cannot be set lower than the Low Respiratory Rate except to be turned o.
Chapter 5 Viewing and Changing Settings
78
Low Inspiratory Pressure
This setting congures the Low Inspiratory Pressure alarm. It is only user-settable in CV, AC, and SIMV modes. It cannot be set lower than PEEP + 2 cm H2O or higher than the High Inspiratory Pressure. For passive circuits, you can increase or decrease the Low Inspiratory Pressure from 6 to 40 cm H2O in increments of 1 cm H2O. For active circuits, you can increase or decrease the setting from 2 to 40 cm H2O in increments of 1 cm H2O. For pressure modes, this alarm is not user­settable.
High Inspiratory Pressure
This setting enables or disables the High Inspiratory Pressure alarm. It is only user-settable in CV, AC, and SIMV modes. The High Inspiratory Pressure cannot be set lower than the Low Inspiratory Pressure. You can increase or decrease the High Inspiratory Pressure from 10 to 80 cm H2O in increments of 1 cm H2O. For pressure modes, this alarm is not user-settable.
Additional Settings Specic to Therapy Modes
The Settings and Alarms menu also contains many additional settings specic to the various therapy modes. The specic settings for each therapy mode are described below.
Continuous Positive Airway Pressure (CPAP) Mode
In addition to the general settings described in the previous section of this manual, you can also set the following settings in CPAP mode.
1. CPAP
You can increase or decrease the CPAP pressure setting from 4 to
20 cm H2O in increments of 1.
2. Trigger Type
The device can be set to trigger breaths based on automatic ow
thresholds or specic ow settings. You can select Auto-Trak or Flow Trigger as the Trigger Type. When Auto-Trak is selected, the Auto-Trak trigger initiates based on automatic ow thresholds.
Trilogy200 clinical manual
Note: If the CPAP pressure is set to 4 (the minimum setting), the Ramp Length setting will be unavailable.
Note: Trigger Type is not available when an Active PAP or Active Flow circuit is selected. Flow trigger is the triggering method used for Active PAP and Active Flow circuits.
79
When Trigger Type is set to Flow Trigger, Flow Trigger Sensitivity Leak Compensation, and Flow Cycle Sensitivity become active, and the trigger initiates based on the Flow Trigger Sensitivity setting.
3. Flow Trigger Sensitivity
If you set the Trigger Type to Flow Trigger, the Flow Trigger
Sensitivity setting displays. You can increase or decrease the setting from 1 to 9 l/min in 1 l/min increments.
The ow trigger initiates when the patient’s inspiratory eort
creates a ow equal to or greater than the ow sensitivity setting.
4. Leak Compensation
If you are using an Active Flow circuit, you can turn Leak
Compensation On or O.
5. Flow Cycle Sensitivity
If you set the Trigger Type to Flow Trigger, the Flow Cycle
Sensitivity setting displays. You can increase or decrease the setting from 10 to 90 percent (%) in 1% increments.
Note: Auto-Trak is only available if the Passive Circuit is selected.
Note: Flow Trigger with leak compensation is only available if the Active Flow circuit is selected.
Note: Enabling Leak Compensation when using the Active Flow Circuit conguration only aects triggering and does not aect tidal volume delivery or Vte measurement.
Note: A ow cycle sensitivity set at 90% will result in the most sensitivity. If the ow cycle sensitivity is set at 10%, this will result in the least sensitivity.
As ow begins to decrease during inspiration,
if the patient ow is less than the ow cycle set point, the device will cycle to expiration. For example: if the ow cycle is set to 75%, when the ow has decreased by 25% of the peak ow, the device will cycle to the EPAP/PEEP level.
6. Ramp Length
The Ramp Length allows you to set the ramp time.
You can disable Ramp by selecting O, or you can increase or decrease the Ramp Length setting from 5 to 45 minutes in 5-minute increments.
Flow Cycle Sensitivity
Chapter 5 Viewing and Changing Settings
80
7. Ramp Start Pressure
You can increase or decrease the ramp start pressure in
increments of 1 from 4 cm H2O to the CPAP pressure setting. The patient also has access to this setting, unless the ramp length is set to O.
8. Flex
You can enable or disable the Flex setting. O disables the
setting and prevents the patient from using Flex. To enable the setting, set Flex to 1, 2, or 3. The patient also has access to this setting, if Flex is enabled. The Flex feature is not available when using an active circuit type.
Spontaneous (S) Mode
The following settings, described in the CPAP mode section of this chapter, also are available in S mode:
• Flex
• Trigger Type
• Flow Trigger Sensitivity
• Flow Cycle Sensitivity
• Ramp Length
• Ramp Start Pressure
In addition to those settings, the settings below are also available in S mode.
1. AVAPS
AVAPS is only available if Flex is not enabled.
You can disable AVAPS by selecting O, or you can enable AVAPS
by selecting On. If you select O, the IPAP setting displays. If you select On, the IPAP Max Pressure and IPAP Min Pressure display.
Note: The Ramp Start Pressure setting will not display if the Ramp Length is set to O or if the CPAP pressure is set to 4 cm H2O.
Note: Ramp Start Pressure is less than or equal to CPAP - 1 cm H2O in CPAP mode.
Note: In CPAP mode, Flex is only available when CPAP is greater than 4 cm H2O.
Note: In S mode, Flex is only available when EPAP is greater than or equal to 4 cm H2O and IPAP is less than or equal to 25 cm H2O.
Note: Flex is only available when Auto-Trak is enabled.
Note: Ramp is not available in the Passive Circuit when IPAP = EPAP = 4 cm H2O or when IPAP Min = EPAP = 4 cm H2O.
Note: Ramp Start Pressure is less than or equal to EPAP - 1 cm H2O in S, S/T, T, and PC modes.
Note: Ramp Start Pressure is greater than or equal to 0 cm H2O when the circuit type is Active PAP or Active Flow in S, S/T, T, and PC modes.
Note: Ramp Start Pressure is greater than or equal to 4 cm H2O when the circuit type is Passive in S, S/T, T, and PC modes.
Note: AVAPS is only available if the Passive Circuit is selected.
Trilogy200 clinical manual
2. IPAP
The IPAP setting displays if AVAPS is O. You can increase or
decrease the Inspiratory Positive Airway Pressure (IPAP) from 4 to 50 cm H2O in increments of 1. IPAP is limited to a maximum of 25 cm H2O when Flex is enabled. You cannot set the IPAP setting lower than the EPAP setting.
3. IPAP Max Pressure
The IPAP Max Pressure setting displays if AVAPS is enabled.
You can increase or decrease the setting from 4 to 50 cm H2O in increments of 1. The IPAP Max Pressure must be equal to or greater than the IPAP Min value.
4. IPAP Min Pressure
The IPAP Min Pressure setting displays if AVAPS is enabled.
You can increase or decrease the setting from 4 to 50 cm H2O in increments of 1. The IPAP Min Pressure must be equal to or greater than the EPAP value, and it must be less than or equal to the IPAP Max Pressure.
5. EPAP
You can increase or decrease the Expiratory Positive Airway
Pressure (EPAP) from 4 to 25 cm H2O in increments of 1. For active circuits, EPAP can be set to zero.
81
Note: IPAP, IPAP Max, or IPAP Min cannot be set to more than 30 cm H2O above EPAP.
Note: EPAP cannot be set to more than 30 cm H2O below IPAP, IPAP Max, or IPAP Min.
When AVAPS is disabled, the EPAP setting must be less than
or equal to the IPAP setting. When AVAPS is enabled, the EPAP pressure must be less than or equal to the IPAP Min Pressure.
6. Tidal Volume
The Tidal Volume setting displays if AVAPS is enabled. You can
increase or decrease the setting from 50 to 2000 ml in 5 ml increments. Use this setting to establish the target volume of gas which the ventilator will produce and deliver during each Spontaneous breath.
Chapter 5 Viewing and Changing Settings
Note: In CV, AC, and SIMV modes, the tidal volume setting is limited by the Inspiratory Time, to maintain the system’s minimum and maximum peak ows.
82
7. Rise Time
You can adjust the rise time to nd the most comfortable setting
for the patient. Increase or decrease the setting from 1 to 6 until you nd the right setting. The rise time levels from 1 to 6 progressively reect slowed response of the pressure increase that will take place at the beginning of inspiration.
8. Apnea Rate
If the Apnea alarm is enabled, you can set the Apnea Rate from
4 to 60 BPM in 1 BPM increments. In S mode, the Apnea Rate is greater than or equal to 1:2 I:E ratio.
Spontaneous/Timed (S/T) Mode
All of the settings described in the S Mode section are also available in S/T mode, except for the Flex setting. In addition to those settings, the settings below are also available in S/T mode.
1. Breath Rate
In AC mode, you can increase or decrease the Breath Rate setting
from 0 to 60 BPM, while in all other modes, you can increase or decrease the setting from 1 to 60 BPM in 1 BPM increments. Use the Breath Rate setting to establish the minimum rate of mandatory breaths that the ventilator will deliver per minute.
2. Inspiratory Time
You can adjust the Inspiratory Time setting from 0.3 to 5.0
seconds in 0.1 second increments. Inspiratory Time is the duration for the inspiratory phase of a mandatory breath.
Timed (T) Mode
All of the settings available in S/T mode are available in T mode, except for the Trigger Type setting. Please refer to the descriptions in the S and S/T Mode sections of this chapter for detailed information on the T mode settings.
Note: The Rise Time setting only displays if Flex is disabled. If Flex is enabled, the device will use a rise time of 3.
Note: In S/T, T, PC, PC-SIMV, SIMV, CV, and AC modes, the Apnea Rate is greater than or equal to the Breath Rate and is limited by the current Inspiratory Time setting to maintain a minimum 1:1 I:E ratio.
Note: In volume modes, the Breath Rate range is limited by the current Inspiratory Time setting to maintain a minimum 1:1 I:E ratio.
Note: In pressure modes, the inspiratory time range is limited by the current Breath Rate setting to maintain a minimum 1:1 I:E ratio.
Note: In volume modes, the inspiratory time range is limited by the current Tidal Volume and Breath Rate settings to maintain a minimum 1:1 I:E ratio and the system’s minimum and maximum peak ow.
Trilogy200 clinical manual
Pressure Control (PC) Mode
All of the settings available in S/T mode are available in PC mode, except for the Flow Cycle Sensitivity setting (when Flow Trigger is enabled). Please refer to the descriptions in the S and S/T Mode sections of this chapter for detailed information on the PC mode settings.
Pressure Control Synchronized Intermittent Mandatory Ventilation
(PC-SIMV) Mode
The following settings, described in the S and S/T mode sections of this chapter, also are available in PC-SIMV mode:
• Breath Rate
• Inspiratory Time
• Trigger Type
• Flow Trigger Sensitivity (if Trigger Type is set to Flow
Trigger)
• Flow Cycle Sensitivity (if Trigger Type is set to Flow
Trigger)
• Rise Time
83
In addition to these, the following settings are also available in PC­SIMV mode.
1. Pressure
You can increase or decrease the Pressure setting from 4 to 50
cm H2O in increments of 1. This is the pressure the ventilator will deliver during the inspiratory phase of a mandatory or assist breath.
2. Pressure Support
You can increase or decrease the Pressure Support setting from
0 to 30 cm H2O in increments of 1. This is the pressure support the ventilator will deliver during the inspiratory phase of a Spontaneous breath.
Chapter 5 Viewing and Changing Settings
Note: In PC-SIMV mode, you cannot set up Pressure Support for Mandatory and Assist breaths (Pressure - PEEP) greater than 30 cm H2O.
Note: The Pressure Support and PEEP settings together cannot exceed 50 cm H2O.
84
3. PEEP
The Positive End Expiratory Pressure (PEEP) setting can be
increased from 0 to 25 cm H2O in active circuits and 4 to 25 cm H2O in passive circuits, in increments of 1. PEEP is the positive pressure maintained in the patient circuit during exhalation. The PEEP must be less than or equal to the pressure setting.
Control Ventilation (CV) Mode
The following settings, described in the previous sections of this chapter, are also available in CV mode:
• Tidal Volume
• Breath Rate
• Inspiratory Time
• PEEP
• High Inspiratory Pressure
• Low Inspiratory Pressure
In addition to these, the following settings are also available in CV mode.
1. Flow Pattern
You can choose either Ramp or Square for the Flow Pattern setting.
2. Sigh
You can enable or disable the Sigh setting by selecting On or O.
A Sigh is a breath that is delivered every 100 breaths at 150% of the normal volume.
Trilogy200 clinical manual
Note: The Flow Pattern setting might be limited to only Ramp or Square based on the Tidal Volume, Inspiratory Time, and Breath Rate settings to maintain the minimum and maximum peak ows.
Assist Control (AC) Mode
The AC mode contains the following settings described in the S, S/T, PC-SIMV, and CV mode sections in this chapter. Please refer to the descriptions in those sections for detailed information.
• Tidal Volume
• Breath Rate
• Inspiratory Time
• Flow Pattern
• PEEP
• Trigger Type
• Flow Trigger Sensitivity
• Sigh
• High Inspiratory Pressure
• Low Inspiratory Pressure
Synchronized Intermittent Mandatory Ventilation (SIMV) Mode
The SIMV mode contains the following settings described in the S, S/T, PC-SIMV, and CV mode sections in this chapter. Please refer to the descriptions in those sections for detailed information.
85
Note: The Low Inspiratory Pressure is limited to PEEP +2 in CV, AC, and SIMV modes.
Note: Flow Cycle Sensitivity is not available in AC mode.
• Tidal Volume
• Breath Rate
• Inspiratory Time
• Pressure Support
• Flow Pattern
• PEEP
• Trigger Type
• Sigh
• Rise Time
• High Inspiratory Pressure
• Low Inspiratory Pressure
Chapter 5 Viewing and Changing Settings
86
Viewing and Changing Options Menu Items
From the Main Menu screen, select the Options item.
The Options menu appears, shown in the screen below.
The following settings are available on the Options menu.
Menu Access – You can select Full or Limited menu
access. Full menu access allows operators to access all ventilator and prescription settings. Limited menu access allows operators to access only certain settings and does not allow them to change prescription settings. To prevent patients from changing prescription settings, do not give them Full menu access.
Detailed View – You can turn Detailed View on or o
using this setting. Detailed view displays additional settings and therapy information on the Monitor screen.
Options Menu
Language – The next item on the Options menu allows
you to select the Language that the software will appear in (English, French, German, etc.). The information on the screens will display in the language selected here.
Pressure Units – The next item allows you to select the
pressure units that will display on the screens. You can choose either:
- cm H2O
- hPa
- mBar
Trilogy200 clinical manual
All pressure units on the screens will display in the unit of
measure selected here.
87
Alarm Volume – You can adjust the volume of the device
alarms using this setting. Select either Loud or Soft as the alarm volume options.
Keypad Lock – You can enable or disable the Keypad
Lock feature, which is described in detail earlier in this chapter. Enabling the Keypad Lock feature can prevent someone from accidentally pressing a button and changing any of the settings. Select On to enable the feature or O to disable it.
Keypad Backlight – The next item you can set is the
Keypad Backlight. You can turn the backlight On or O using this setting. Whenever you press the button to begin therapy, the keypad backlight temporarily lights up. Once therapy is being provided, the keypad will be lit according to this Keypad Backlight setting. If the setting is On, the backlight remains on while therapy is provided. If the setting is O, the backlight remains o while therapy is provided.
LCD Brightness – The LCD display is lit by a backlight.
The backlight turns on when the initial Startup screen displays. You can adjust the brightness of the LCD backlight from 1 – 10, with 1 being the dimmest setting and 10 being the brightest.
Screen Saver – You can change the screen saver to
reduce power consumption or dim the screen in a dark room. You can choose the following settings:
- O: No screen saver displays and the LCD backlight remains lit at your brightness setting.
WARNING
Make sure the alarm volume is set loud enough to be heard by the caregiver. Consider the use of a remote alarm.
Note: Setting the screen saver to Black allows the device to run for a longer period of time on battery power.
- Breath: The display appears as a black screen, with only the patient breath indicator and manometer visible.
- Black: The display’s backlight is turned o and the display is black with no information visible.
Chapter 5 Viewing and Changing Settings
88
- Dim: The display’s backlight is decreased, so that the display is still visible but not as bright.
If enabled, the screen saver will display after 5 minutes of no
keypad activity. Pressing any button on the device will exit the screen saver. And, any alarm or informational message will also exit the screen saver.
Date Format – You can choose either mm/dd/yyyy or
dd/mm/yyyy as the date format that will display on the device screens.
Time Format – You can choose to display either an AM/
PM time format or 24 Hour time format (for example, 2:49 PM or 14:49).
Month – The month defaults to the current month. The
adjustable range is from 1 (January) – 12 (December).
• Day – The day defaults to the current day. The adjustable
range is from 1 – 31. The maximum value is based on the selected month.
Year – The year defaults to the current year. The
adjustable range is from 2000 – 2099.
Hour – The hour defaults to the current hour. The
adjustable range is from 12 AM – 12 PM or 0-23, depending on the selected Time Format.
Minute – The minute defaults to the current minute. The
adjustable range is from 0 – 59.
IP Address Mode – You can change the IP address
mode to either DHCP or Static, depending on the type of network you are using (if applicable).
Trilogy200 clinical manual
Operational Hours – The operational hours displays
the total number of hours that the device blower has been on since the last time this value was reset. You can reset this value to zero (0) if desired (e.g., each time you give the device to a new patient). This value helps you determine how often the patient is using the device. The Operational Hours shown here diers from the Blower Hours shown on the Information screens. The Blower Hours displayed in the Information screen is the total number of hours that the blower has been working over the life of the device. You cannot reset this value.
Viewing the Alarm Log
From the Main Menu screen, you can select Alarm Log to access the Alarm Log screen. An example is shown below.
89
Note: In the Alarm Log screen, the 1/2 shown in the Menu banner indicates that page 1 of 2 alarm log pages is being viewed at this time.
The alarm log displays the alarms in chronological order with the most recent events displayed rst. It lists the 20 most recent alarms or messages that appeared on the device display. When the device is in Limited Menu access mode, the alarm log cannot be cleared. It can be cleared when in Full Menu access mode. Depending on how many alarms have occurred, the alarm log may be several pages long. The entries in the alarm log use the same names that displayed when the alarm initially occurred and was displayed in the Alarm View.
Chapter 5 Viewing and Changing Settings
Alarm Log Screen
Note: In Full Menu access mode, you can press the Right (Clear) button to clear the alarm log if desired.
Loading...