Pfeiffer Vacuum MVP 040-2, MVP 070-3, MVP 070-3C Operating Instructions Manual

Betriebsanleitung • Operating Instructions
Translation of the Original Operating Instructions
Diaphragm Pumps
MVP 040-2, MVP 070-3/3C
PU 0050 BE/A (0809)
Table of contents
Table of contents
1.1 Validity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2 Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.1 Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 Proper use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 Improper use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3Transport and storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.1 Transport. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.2 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4.1 Product identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4.2 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
5Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.1 Setting up the pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.2 Connecting the vacuum side. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.3 Connecting the exhaust side . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.4 Connecting to the mains power supply . . . . . . . . . . . . . . . . . . . . . . 11
6Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6.1 Before switching on the pump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6.2 Switching on the pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6.3 Pumping condensable vapours. . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.4 Switching off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
7Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
7.1 Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
7.2 Cleaning and replacing diaphragm and valves . . . . . . . . . . . . . . . . 16
7.3 Replacing the device fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
8Decommissioning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
8.1 Shutting down for longer periods. . . . . . . . . . . . . . . . . . . . . . . . . . . 26
9Malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
9.1 Rectifying malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
10 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
11 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
12 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
13 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
13.1 Substances in contact with the media . . . . . . . . . . . . . . . . . . . . . . . 33
13.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
2
1About this manual
1.1 Validity
This operating manual is for customers of Pfeiffer Vacuum. It describes the func­tioning of the designated product and provides the most important information for safe use of the unit. The description follows applicable EU guidelines. All informa­tion provided in this operating manual refer to the current state of the product's de­velopment. The documentation remains valid as long as the customer does not make any changes to the product.
Up-to-date operating instructions can also be downloaded from www.pfeiffer-vacuum.net.
Applicable docu­ments
MVP 040-2, MVP 070-3/3C Operating instructions
Safe ty informat ion f or vac uum pum ps "Safety Guide" PT 030 0 BN*
Declaration of Conformity Part of this document
Operating instructions for accessories (order-specifically) see section "accessories"*
*also available via www.pfeiffer-vacuum.net
For information about other certifications, if applicable, please see the signet on the pump or:
About this manual
•www.tuvdotcom.com
•TUVdotCOM-ID 0000021320
1.2 Conventions
Safety instructions The safety instructions in Pfeiffer Vacuum operating manuals are the result of risk
evaluations and hazard analyses and are oriented on international certification standards as specified by UL, CSA, ANSI Z-535, Semi-S1, ISO 3864 and DIN 4844. In this document, the following hazard levels and information are considered:
DANGER
Immediate danger
Death or very severe injuries occur.
WARNIN G
Possible danger
Death or injuries may occur.
CAUTION
Possible danger
Medium to slight injuries may occur.
NOTE
Command or note
Command to perform an action or information about properties, the disregarding of which may result in damage to the product.
3
Safety
Pictograph definitions
Instructions in the
! Work instruction : here y ou have to do some thin g .
Prohibition of an action or activity in connection with a source of danger, the disregarding of which may result in serious accidents.
War ning of a d isp layed s ource of d a nger in co nnection with operation of the unit or equipment.
Command to perform an action or task associated with a source of danger, the disregarding of which may result in serious accidents.
text
Abbreviations used C version: Corrosive gas version
Symbols used The following symbols are used consistently throughout in all illustrations:
V
Vacuum connection
Exhaust
G
Gas ballast valve
Power connection
2Safety
2.1 Safety precautions
Duty to inform
Each person involved in the installation, operation or maintenance of the vacuum pump must read and observe the safety-related parts of these operating instructions.
! The operator is obligated to make operating personnel aware of dangers originating
from the vacuum pump, the pumped medium and the entire system.
•Before pumping dangerous, corrosive or environmentally hazardous media, take suitable precautions:
–Test the compatibility with substances in contact with the media. –Prevent the release of process gases and their reaction products and by-pro-
–Safety measures (e.g. wearing protective clothing and safety goggles) to pre-
•Before pumping gases which could form ignitable mixtures, take suitable pre­cautions:
–By implementing the required safety measures, prevent potentially explosive
NOTE
ducts or dispose of them according to the relevant regulations.
vent inhalation and skin contact.
mixtures from occurring in the housing and from being ignited in the event of a diaphragm crack by mechanically produced sparks, hot surfaces or static electricity.
4
2.2 Proper use
Safety
–If necessary, use inert gas for gas ballast supply or ventilation.
•Connect the vacuum pump to a shockproof socket only. –Use only undamaged network cables which comply with the regulations. –Make sure that the grounding is sufficient.
•Do not expose any body parts to the vacuum.
•Observe the safety and accident prevention regulations.
•Check regularly that all safety precautions are being complied with.
•Do not carry out any unauthorised modifications or conversions to the pumps.
•Depending on the operating and ambient conditions, the surface temperature of the pumps may rise above 70 °C. Use suitable finger guards if necessary.
• When returning the pumps to us please note the instructions in the Service sec­tion.
NOTE
CE conformity
The manufacturer's declaration of conformity becomes invalid if the operator modifies the original product or installs additional components.
! Following inst allation into a plant and before commissioning, the operator must
check the entire system for compliance with the valid EU directives and reassess it accordingly.
2.3 Improper use
•The vacuum pump may only be used to generate a vacuum.
•Installation, operating and maintenance regulations must be complied with.
•Other accessories than those described in this manual must not be used without the agreement of Pfeiffer Vacuum.
• When pumping gases which could form explosive or ignitable mixtures, take suitable precautions:
–If necessary, connect inert gas for ventilation or gas ballast supply.
Improper use will cause all claims for liability and guarantees to be forfeited. Im­proper use is deemed to be all use for purposes deviating from those mentioned above, especially:
•Pumping of corrosive gases (exception: pumps in C version).
•Pumping of explosive media.
•Operation of the pump in potentially explosive areas.
•Pumping of gases containing impurities such as particles, dusts and condensate; note the vapour compatibility levels of the pump.
•Pumping of substances that tend to sublime.
•Use of the vacuum pump to generate pressure.
•Pumping of liquids.
•Connection to pumps or units which are not suitable for this purpose according to their operating instructions.
•Connection to units which have exposed voltage-carrying parts.
Improper use of the equipment automatically invalidates all warranty and liability claims.
5
Transport and storage
3Transport and storage
3.1 Transport
! Remove the locking cap from the vacuum and fore-vacuum flange immediately
before connecting!
! Use only the handles provided for that purpose to lift the pump.
–Do not use the interhead connections on the top side of the pump to carry the
pump.
3.2 Storage
Fig. 1: Lifting and transporting the pump
! Check that all the openings on the pump are securely closed. ! Store the pump in a cool, dry place; preferably at room temperatures (approx.
20°C). –For a longer period of storage, seal the pump in a PE bag with drying agents
enclosed.
6
Product description
4Product description
4.1 Product identification
To co rr ec tl y id en ti fy t he p ro du ct w he n co mm un ic at in g wit h Pf e if fe r Va cu um , al ­ways have the information from the rating plate available.
•Pump model and model number
•Serial number
•Date of manufacture
Scope of delivery •Pump with mains power supply (switchable) via rubber socket
•Locking caps for vacuum and exhaust flange
•Operating manual
Differences between the pump versions
Pump type Pump designs
MVP 040-2
MVP 070-3
MVP 070-3C
Standard version:
•G 1/4" elbow union with enclosed hose 10/8, 1000 mm with an elbow union in G 1/4" at the end
Standard version:
•G 1/4" elbow union + enclosed hose 10/8, 1000 mm with a elbow union in G 1/4" at the end
C version:
•For DN 16 ISO-KF
7
Product description
4.2 Function
The diaphragm vacuum pumps of the series MVP 070-3 are three stage and those of the MVP 040-2 series are two stage, dry compressor vacuum pumps. The pumps are positive displacement pumps with a periodic change of size of the suction chamber produced by t he movement of the diaphragm. The gas f low causes the valves to open and close automatically. The pump units are directly connected to the drive motor.
23 U N 2.1 40
A A1 B
Fig. 2: MVP 040-2
2.1 U N
A1 Vacuum connection
2.1 Exhaust (with silencer) ADiaphragm head 1 BDiaphragm head 2 NPower switch UVoltage selector switch 23 Hose connection 40 Gas ballast valve
A1 Vacuum connection
2.1 Exhaust (with silencer) NPower switch UVoltage selector switch
A1
Fig. 3: MVP 070-3
8
Product description
2.1
Fig. 4: MVP 070-3C
40
U N
A1
A1 Vacuum connection
2.1 Exhaust NPower switch UVoltage selector switch 40 Gas ballast valve
9
Installation
5Installation
5.1 Setting up the pump
Installation location Observe the following requirements when setting up the pump:
•Note the load-bearing capacity of the mounting surface.
•Maximum installation altitude 2000 m (above mean sea level)
•Permissible ambient temperature: +12 ... 40°C
•Maximum relative humidity 85%
! Always place the pump on a firm, even surface.
– Where stationary installation is involved, anchor the pump on site.
! When installing the pump in a closed housing, ensure there is sufficient air cir-
culation. –Voltage and frequency information given on the motor rating plate must be
visible.
–If vacuum pump is installed above 1000 m above mean sea level check com-
patibillity with applicable safety requirements, e.g. DIN EN 61010 (motor may overheat due to insufficient cooling).
5.2 Connecting the vacuum side
! Remove locking cap on intake connection and connect vacuum pump to the ap-
paratus.
! The connection between the pump and the recipient should be kept as short as
possible. –Depending on the pump type, use metallic hoses or PVC hoses with flange
connections.
–Separators, filters etc. may be installed upstream to protect the pump (see ac-
cessories). However, please observe the loss of pumping capacity due to the conductivity of the accessories.
5.3 Connecting the exhaust side
High pressure in the exhaust line!
Danger of damage to the seals and danger of the pump bursting.
! Install the line without shut-off valves on the exhaust side. ! Pumpe nicht mit Überdruck am Einlass betreiben; max. zulässige Drücke und Druck-
differenzen beachten.
! Choose the cross-section of the exhaust line to be at least the size of the nominal
connection diameter of the vacuum pump's exhaust connection.
! Piping to the pump must be suspended or supported.
–Physical forces from the piping system must not be allowed to act on vacuum
pumps.
! Lay piping from the pump sloping downward so that no condensate can flow
back into the pump; otherwise fit a condensate separator. –If an air trap is created in the system, then a device for draining condensation
water must be provided at the lowest point.
CAUTION
10
Emission of toxic substances from the exhaust!
Danger of poisoning from emitted gases or vapours, which can be detrimental to health and/or can pollute the environment, depending on the particular application.
! Comply with the applicable regulations when working with toxic substances. ! Only officially approved filter systems may be used to separate and remove these
substances.
5.4 Connecting to the mains power supply
The pump is driven by single-phase extended voltage range motors with revers­ible voltage ranges.
Excess voltage!
Danger of destroying the motor. ! Power connections must comply with local regulations. Voltage and frequency infor-
mation given on the motor rating plate must correspond to the mains voltage and frequency values.
! To p ro t e c t t h e m o t or an d su p pl y ca b l e in c as e of m al f u n ct i on , m a in s fu s e p r o t e ct i o n
must be implemented.
Installation
WARNIN G
CAUTION
11
Installation
Single phase motors ! The mains voltage must be determined on-site each time before the pump is in-
stalled or moved to a different location.
NOTE
Overvoltage!
An incorrect voltage range setting can damage the motor.
! Disconnect the pump from the power supply. ! Only change the voltage range when the pump is disconnected from the power
mains.
Changing the voltage range
! Disconnect the pump from the power supply. ! Set the desired vo ltage range on the voltage selector switch using a suitable
screwdriver.
115
230
Intermittend opera­tion with TC via relay box (accessory)
230
115
Switch position: "115" "230"
Volta ge ran ge s:
100-115 V, 50/60 Hz , 120 V, 60 Hz
200-230 V, 50/60 Hz
Motor protection
A self-locking thermal winding protector switches off the pump motor in the event of overheating.
! Allow the pump to cool off several minutes and do not switch it back on until it
has cooled off.
Connection of diaphragm pumps in the pumping station according wiring diagram in operating instructions PT 0030 BN of the Backing Pump Relay Boxes:
L N
3
TC
Fig. 5: Connection with relay box (PM 061 372-T/374-T) to TC 110/400
4
accessory A
PE PE N L´
MVP-Reihe
MVP series
12
6Operation
6.1 Before switching on the pump
! Compare the voltage information on the rating plate with the mains voltage. ! Check that the exhaust connection allows free flow (max. permissible pressure
1.1 b ar a bso lu te) . –Activate the shut-off valves in such a way that they open before or at the same
time as the pump is started.
! Protect the p ump sufficien tly from taking in contaminants by means of suitable
precautions (e.g. dust filters).
Dangerous overpressure overload!
Mixing up the connections leads to a dangerous overpressure overload in the pump, and the motor could be damaged.
! Before commissioning, it is imperative that you ensure that the pressure does rise
above the maximum permissible pressure on the pressure side.
! Start pumps at a maximum pressure differential of 1 bar between inlet and outlet.
Operation
CAUTION
6.2 Switching on the pump
The pump can be switched on in any pressure range.
No special precautions are necessary when pumping dry gases. In order to attain the lowest possible final pressures, the gas ballast valve should be closed.
The pump attains the stated values for throughput rates and final pressure levels only once the operating temperature is reached (after approximately 15 minutes).
! Switch on the pump with the vacuum flange closed and allow to warm up for 15
minutes.
Hot surface!
Danger of burns if hot parts are touched. Depending on the operating and ambient conditions, the surface temperature of the pump may rise above 70 °C.
! In this case, use suitable finger guards.
CAUTION
13
Operation
6.3 Pumping condensable vapours
Should the process gases contain condensable gases present at high percentages, the vacuum pump must be operated with a gas ballast (i.e. with an open gas ballast valve).
Reactive, explosive or otherwise dangerous mixtures!
Uncontrolled gas inlet at the gas ballast valve can result in dangerous mixtures. ! By implementing the required safety measures, the user must prevent potentially ex-
plosive mixtures from occurring in the inside of the pump and from being ignited in the event of a diaphragm crack by mechanically produced sparks, hot surfaces or static electricity.
! If necessary, use inert gas for ventilation and gas ballast supply.
Bad final vacuum and damage to the pump!
Danger of condensation and a reduced final vacuum during operation without a gas ballast or in case of insufficient supply of flushing gas.
! Only pump vapors when the pump is warm and the gas ballast valve is open. ! When the process has been completed, allow the pump to continue running for
about 30 minutes at atmospheric pressure with the gas ballast open.
WARNIN G
NOTE
Gas ballast valve Letting in gas ballast improves the discharge of condensate, and the pump
achieves the specified final vacuum more quickly.
40Gas ballast valve 43 Cap
43
40
Fig. 6: Operation with gas ballast valve
! Open gas ballast valve; to do so, turn cap 43 on the gas ballast valve 40 so that
the two holes line up.
6.4 Switching off
14
The pump can be switched off in any pressure range.
7Maintenance
7.1 Precautions
Pump parts may be contaminated from pumped media!
Danger of poisoning due to contact with harmful substances.
! Decontaminate the pump before carrying out any maintenance work. ! In the event of contamination, take suitable safety precautions to prevent your health
from being harmed by any dangerous substances.
The valves and the diaphragms are wear parts. If the rated ultimate vacuum is no longer achieved, the pump interior, the diaphragms and the valves must be cleaned and the diaphragms and valves must be checked for cracks or other dam­age.
Depending on individual cases it may be efficient to check and clean the pump heads on a regular basis. In case of normal wear the lifetime of the diaphragms and valves is > 10000 operating hours.
! Allow the pump to cool to a safe temperature. ! Only dismantle the pump as far as necessary in order to repair defects. ! Use only alcohol or similar agents for cleaning pump parts. ! Re-assemble pump in reverse order.
Maintenance
WARNING
NOTE
Service work should be carried out by qualified personal only!
Pfeiffer Vacuum is not liable for any damage to the pump resulting from work carried out improperly.
! Ta ke ad va nt ag e of o ur se rv ic e tr ai nin g pr ogr am s; ad di ti on al in fo rm at io n at
www.pfeiffer-vacuum.net.
! Please state all the information on the pump rating plate when ordering spare parts.
15
Maintenance
Checklist for inspec­tion, maintenance and overhaul
Certain repair and overhaul work should only be performed by Pfeiffer Vacuum Service (PV). Pfeiffer Vacuum will be released from all warranty and liability claims if the required intervals for inspection, maintenance, or overhaul are exceeded or inspection, maintenance, repair or overhaul procedures are not performed proper­ly. This also applies if replacement parts other than Pfeiffer Vacuum OEM replace­ment parts are used.
Activity
Check silencer for contamination X
Clean, change valves and diaphragms X
Change silencer X
Depending on the process, the required intervals for inspection and maintenance can exceed the
typical values specified in the table. Please consult Pfeiffer Vacuum if necessary.
daily
7. 2 C l e a n i n g a n d r e p l a c i n g d i a p h r a g m a n d v a l v e s
as required;
at least once every six months
as required;
at least annually
as required;
at least every 2 years
MVP 040-2 Cleaning and repla-cing the valves
! Tu rn o ff t h e va cu um p um p, ve nt t o at m os phe ri c pr es sur e an d al lo w to co ol, i f
necessary.
! Disconnect the drive motor from the mains and secure it so that it cannot be
switched on.
! Unscrew the hose connection between the pump stages by loosening the hol-
low screw with a wrench (size 17).
! Remove the cheesehead screws 26 (fou r per membrane head). ! Remove housing cover G with head cover K; prevent the head cover from falling
out of the housing cover.
! Remove head cover K from housing cover G to check valve seals 11. ! Ta k i n g c a r e wi t h t h e p o s i ti o n o f v a l v e s e a l s 11 , r e m ov e a nd c he c k fo r d a ma g e ,
and replacing if necessary.
! Check diaphragm pair 10 for damage and replace if necessary.
Damage to the valve seals
Valve seals can be destroyed by gluing to after wrong installation. ! Pay attention to the fitting position of the inlet and outlet valve seals. The valve flap
of the valves may not cover the groove in the head and housing cover K/G.
! Clean all parts and inspect for wear. ! Re-assemble pump in reverse order. ! Check the other membrane head in the same way.
NOTE
16
40 A1
Maintenance
2.3 K1
2.1
2.3
11
G1 K1 17 6 D 9 10 3 16 K2 G2 26
Fig. 7: MVP 040-2
A1 Vacuum connection with
compression coupling in G
1/4" DSpacer disks G1 Housing cover, inlet side G2 Housing cover, outlet side K1 Head cover, inlet side K2 Head cover, outlet side
2.1 Silencer
2.3 Sealing ring 3Diaphragm spring was-
her 6Housing panel 9Diaphragm support disk 10 Dia ph rag m p air
11 In l e t a n d o u t l e t v a l v e
seal 16 Countersink screw 17 Countersink screw 26 Socket cap screw 40 Gas ballast valve
Changing the Diaphragm
! Dismantle membrane heads as described before. ! Remove screw 16; this screw has been stuck to make secure against loosening.
NOTE
Damage to the pu mp and reduced final vacuum!
A changed dead centre (TDC) leads in the most unfavorable case to a bearing damage.
! Check for washers D under diaphragm support washer. ! Make sure that the original number is reassembled at the individual membrane head.
! Dismantle diaphragm spring washer 3, diaphragm pair 10 and diaphragm sup-
port washer 9. –If it is difficult to separate the old diaphragm from support washer 9, apply e.g.
alcohol or petroleum ether.
Assembly
! When existing, re-fit spacers D under the diaphragm support washer 9. ! Put two single diaphragms with their rough side together so that the labelled
side of the two diaphragms are readable.
! Arrange diaphragm support washer 9, diaphragm pair 10 and diaphragm spring
washer 3 onto screw 16.
! Place some threadlocker (low strength) under the head of screw 16 and screw in
con-rod.
! Refit head cover, valve seals and housing cover in reverse order, taking care with
the position of valve seals 11.
17
Maintenance
! First gently and then firmly tighten diagonally-offset cheesehead screws26;
torque: 12 Nm.
NOTE
Dissimilar valves!
Pump not achieving the specified end pressure ! Make sure the valves have the correct mounting orientation; the outlet valve in head
cover K1 and the inlet valve in head cover K2 are from PTFE (white). The two other valves are from FPM (brown).
V
N
K1 K2
Fig. 8: Position of the valve seals 11 in the head covers K1/K2
N
18
MVP 070-3 Cleaning and repla-cing the valves
! Tu rn o ff t h e va cu um p um p, ve nt t o at m os phe ri c pr es sur e an d al lo w to co ol, i f
necessary.
! Disconnect the drive motor from the mains and secure it so that it cannot be
switched on.
! Mount pump to the side, provide suitable support as necessary. ! Remove the cheesehead screws 26 (fou r per membrane head). ! Remove housing cover G with head cover K; prevent the head cover from falling
out of the housing cover.
! Remove head cover K from housing cover G to check valve seals 11. ! Ta k i n g c a r e wi t h t h e p o s i ti o n o f v a l v e s e a l s 11 , r e m ov e a nd c he c k fo r d a ma g e ,
and replacing if necessary.
! Check diaphragm pair 10 for damage and replace if necessary.
! Replace the two O-rings 12 enclosing the valve seat. ! Clean all parts and inspect for wear. ! Re-assemble pump in reverse order. ! Follow the same procedure for th e other diaphragm heads.
12
Maintenance
G2
K4
11
2.1
Fig. 9: MVP 070-3
A1
D 9 10 3 K1 11 26 G1
2.3
G1 Housing cover 1 G2 Housing cover 2 K4 Head cover membrane head 4
2.1 Silencer
2.3 Seal ring
3Diaphragm spring was-
her
9Diaphragm support
washer
10 Diaphragm pair
11 I n l e t a n d o u t l e t v a l ve
seal (standard) 12 O-ring 26 Cheesehead screw
19
Maintenance
Changing the Diaphragm
Attach the diaphragm key ! Carefully raise the diaphragm at the side without causing it any damage; do not
use sharp-edged tools.
! Slide diaphragm key under the diaphragm until it reaches the support disk. ! Use the diaphragm key to loosen the diaphragm support disk and unscrew to-
gether with diaphragm and diaphragm spring washer.
! If the old diaphragm pair is difficult to separate from the diaphragm support disk
9, loosen in alcohol or petroleum.
Assembly
! Assembling is carried out in reverse order. ! Put two single diaphragms with their rough side together so that the labelled
side of the two diaphragms are readable.
! Position new diaphragm pair 10 between diaphragm spring washer 3 with
square head screw and diaphragm support washer 9; make sure that the square head screw of the diaphragm clamping disc is correctly seated in the guide hole of the diaphragm support disc.
NOTE
Damage to the pu mp and reduced final vacuum!
A changed dead centre (TDC) leads in the most unfavorable case to a bearing damage.
! Check for washers D under diaphragm support washer. ! Make sure that the original number is reassembled at the individual membrane head.
Installing the diaphragm ! Raise the diaphragm at the side and carefully slide the diaphr agm key through
to the diaphragm support disk.
! Screw diaphragm spring washer, diaphragm, diaphragm support washer and
spacers (if applicable) to connecting rod; optimum torque: 6 Nm. –It is recommended to use a torque key, attach hex key to diaphragm key (he-
xagonal bolt 6 mm wide).
! Bring the diaphragms into a position in which they are in contact with the hou-
sing and centred with respect to the bore.
! Mount head cover; ensure correct orientation:
–Align the nose (N) with the notch in the housing cover.
! Insert the valve seals and O-rings into their respective positions; ensure the val-
ves are correctly aligned.
! Mount housing cover; insert the connecting pipe for the two housing covers on
the housing cover; –ensure the sealing rings are correctly positioned in the pipe grooves; replace
O-rings as necessary.
! Refit head cover, valve seals and housing cover in reverse order, taking care with
the position of valve seals 11.
! First gently and then firmly tighten diagonally-offset cheesehead screws26;
torque: 12 Nm.
20
Maintenance
1 ... 4 Diaphragm heads NNose on head cover 11 Va l v e s e a l
N
11
N
4
N
3
MVP 070-3
Fig. 10: Positions required for installing the valve seals 11
2
N
1
21
Maintenance
MVP 070-3C Cleaning and repla-cing the valves
Service the pump heads of only one pump side at the same time.
! Tu rn o ff t h e va cu um p um p, ve nt t o at m os phe ri c pr es sur e an d al lo w to co ol, i f
necessary.
! Disconnect the drive motor from the mains and secure it so that it cannot be
switched on.
! Open the hinged cover of the connection fastener 17. ! Loosen the fillister head screw with a Torx screw driver TX20; do not detach the
screw completely.
! Use a TX20 torx screwdriver to undo the 4 screws on the head cover hood; pay
attention to the spring washers beneath the screws.
! Pull head cover hood forward and remove. ! Carefully move the diaphragm pump into a lateral position so that the pump
heads to be serviced are pointing upwards; support the pump below the motor as necessary.
40.2 40
17 15 19
22 6
D 9 10 3 8 K 11/13 18 20 26 G
Fig. 11: MVP 070-3C
DSpacers GHead cover hood KHead cover 3Diaphragm spring washer 6Housing panel 9Diaphragm support washer 10 Diaphragm ( indi vidual part )
11 In l e t a n d o u t l et v a l v e
seal II 13 O-ring 15 Valve head 17 Connectio n holder with
hinged cover 18 Disc spring
19 C lam pin g cl aw 20 Countersunk screw 22 Connecting ho se 26 Fillister head screw 40 Gas ballast valve
40.2 Rubber cap for pos. 40
Loosen hose clamp. ! Loosen U-hose connection (hose connection to the o pposite side of the pump);
–do this by levering out the hose clamp with a flat-blade screwdriver and re-
moving the hose.
22
Maintenance
Loosen clamping claws
on the valve heads
! Using a TX20 torx screwdriver, undo two countersunk screws on each side and
remove the clamping claws; –the head covers remain assembled.
! Completely remove valve heads together with disc springs, where applicable
with connecting hose, hose clips and connection holders; –ensure the valves are correctly seated.
! Check valves and O-rings for damage and contamination; clean or replace parts
as necessary.
Assembly
! Insert valves and O-rings; ensure valves are correctly seated.
Fig. 12: Valve orientation
•Inlet side (IN): –Identified by the designation "IN" next to the valve seat. The valve flap is poin-
ting towards the pocket in the valve seat.
•Outlet side (EX): –Identified by the designation "EX" next to the valve seat. The valve has the
same orientation as the inlet valve.
! Install valve heads together with disc springs; mount disc springs with curvature
uppermost.
! Centre valve head on the valve seat; valve head must lie flat inside the valve seat
nose.
! Thread the square nut of the connection holder into the groove on the head co-
ver and attach the screw.
! Mount the clamping claws. ! Gently insert the countersunk screws; correct their alignment on the valve heads
as necessary.
! Tighten countersunk screws using a TX20 torx screwdriver, torque: 3 Nm.
23
Maintenance
Changing the Diaphragm
Attach the diaphragm key ! Loosen U-hose connection (hose connection to the opposite side of the pump);
–do this by levering out the hose clamp with a flat-blade screwdriver and re-
moving the hose.
! Undo 4 cylinder screws from the two head covers K and remove both head co-
vers together with the valve heads and connections.
! Carefully raise the diaphragm at the side without causing it any damage; do not
use sharp-edged tools.
! Slide diaphragm key under the diaphragm until it reaches the support disk. ! Use the diaphragm key to loosen the diaphragm support disk and unscrew to-
gether with diaphragm and diaphragm spring washer.
! If the old diaphragm pair is difficult to separate from the diaphragm support disk
9, loosen in alcohol or petroleum.
NOTE
Damage to the pu mp and reduced final vacuum!
A changed dead centre (TDC) leads in the most unfavorable case to a bearing damage.
! Check for washers D under diaphragm support washer. ! Make sure that the original number is reassembled at the individual membrane head.
Installing the diaphragm ! Assembling is carried out in reverse order.
! Position new diaphragm 10 between diaphragm spring washer 3 with square
head screw and diaphragm support washer 9; make sure that the square head screw is correctly seated in the guide hole of the diaphragm support disc.
–Position diaphragm with pale side towards diaphragm spring washer.
! Raise the diaphragm at the side and carefully slide the diaphragm key through
to the diaphragm support disk.
! Screw diaphragm spring washer, diaphragm, diaphragm support washer and
spacers (if applicable) to connecting rod; optimum torque: 6 Nm. –It is recommended to use a torque key, attach hex key to diaphragm key (he-
xagonal bolt 6 mm wide).
! Bring the diaphragms into a position in which they are in contact with the hou-
sing and centred with respect to the bore.
! Put on head cover with valve heads and connections. ! Screw in the socket head screws diagonally first slightly with a hex key size 5;
maximum torque: 12 Nm.
! Mount the U-hose and hose clamp; close the hose clamp with dimensioning
tool.
! Install head cover hood and tighten screws using TX20 torx screwdriver. ! Tighten countersunk screws of connection holder using TX20 torx screwdriver. ! Close film hinge on the connection holder.
24
7. 3 R e p l a c i n g t h e d e v i c e f u s e s
The safety fuses (PU E22 016 -T) are located in the terminal box and are integrated into a cable.
Vo lt ag e- b ea ri ng e l em e nt s
Danger to life from electric shock.
! The fuses can be changed only by trained and authorised electricians. ! Before opening the terminal box, switch off the pump and pull the power plug.
Maintenance
WARNIN G
6,3 AT
Fig. 13: Replacing the fuses
! Disconnect the drive motor from the mains and secure it so that it cannot be
switched on.
! Wait for two min u tes until the capaci tors have dischar ged. ! Establish the cause of the fault and rectify before restarting. ! Open the terminal box lid. ! Remove the entire cable with integrated fuse from the flat pin bushing and re-
place.
! Remount the terminal box lid.
25
Decommissioning
8Decommissioning
8.1 Shutting down for longer periods
Before shutting down the pump, observe the following procedure and adequately protect the pump system against corrosion:
Shortly after conden­sate has formed:
! Let the vacuum pump continue to run for several minutes with the intake po rt
open.
! Should media get into the pump which could corrode the pump materials or
form deposits, clean and check the diaphragm heads.
In the long term: ! Carry out the measures described for brief shutdowns.
! Disconnect the pump from the equipment. ! Close the manual gas ballast valve. ! Close the inlet and outlet opening (e.g. with transport caps). ! Store the pump in a dry place.
26
9Malfunctions
Please note the following instructions should the pump malfunction:
Hot surface!
Danger of burns if hot parts are touched. The surface temperature of the pump may rise above 105 °C in case of malfunction.
! Carry out work on the pump only after it has cooled to a safe temperature.
Malfunctions
CAUTION
27
Malfunctions
9.1 Rectifying malfunctions
Problem Possible causes Remedy
Pump will not start up
Pump switches off after a while after being started
Pump not achieving the end pressure
Pumping speed of pump too low
Unusual noises during operation
No mains voltage or voltage does not correspond to the motor data
Pump temperature too low Warm up pump to > 12°C
Thermal protection switch of the motor has responded
Phase failure Check fuse
Diaphragms or valves dirty Clean pump(see p. 15, chap. 7)
Overpressure in the exhaust lead
One of the integrated fuses is defective
Thermal protection switch of the motor has responded
Mains fuse protection triggered due to overload (e.g. cold start)
Exhaust pressure too high
Condensate in the pump
Gas ballast valve open Close gas ballast valve
Valves o r d iap hragm s dirty or defective
Leak in the system Fix leak
Intake line not well-dimensio­ned
Exhaust pressure too high
Diaphragms or valves defective
Suction chamber dirty Clean suction chamber
Silencer loose or missing Check silencer; replace if necessary
Valves d irty or def ect ive
Motor fan defective Replace motor fan
Connection rod or motor bea­ring defective
Check mains voltage and mains fuse protection; check motor switch
Detect and fix cause of overheating; allow pump to cool off if necessary.
Check exhaust lead
Check fuses and replace if necessary
Detect and fix cause of overheating; allow pump to cool off if necessary.
Warm up pu mp
Check opening of exhaust line and exhaust accessories
Operate pump for a longer period of time at atmospheric pressure; if neces­sary, open the gas ballast valve
Clean or change valves and diaphragms
(see p. 15, chap. 7)
Keep connections as short as possible and see that cross-sections are suffi­ciently dimensioned
Check opening of exhaust line and exhaust accessories
Clean or change valves and diaphragms
(see p. 15, chap. 7)
Clean or change valves and diaphragms
(see p. 15, chap. 7)
Contact Pfeiffer Vacuum Service
28
NOTE
Service work should be carried out by qualified personal only!
Pfeiffer Vacuum is not liable for any damage to the pump resulting from work carried out improperly.
! Ta ke ad va nt ag e of o ur se rv ic e tr ai nin g pr ogr am s; ad di ti on al in fo rm at io n at
www.pfeiffer-vacuum.net.
! Please state all the information on the pump rating plate when ordering spare parts.
10 Service
Service
Pfeiffer Vacuum offers first-class service!
•Maintenance/repairs on the spot by Pfeiffer Vacuum field service
•Maintenance/repairs in the nearby service center or service point
•Fast replacement with exchange products in mint condition
•Advice on the most cost-efficient and quickest solution
Detailed information and addresses at: www.pfeiffer-vacuum.net (Service).
Maintenance and repairs in the Pfeiffer Vacuum ServiceCenter
The following steps are necessary to ensure a fast, smooth servicing process:
! Download the forms "Service Request" and "Declaration on Contamination". ! Fill in the "Service Request" form and send it by fax or e-mail to your service
address.
! Include the confirmation on the service request from Pfeiffer Vacuum with your
shipment.
! Fill in the contamination declaration and enclose it in the shipment (required!). ! Dismantle all accessories. ! Send the pump in its original packaging if at all possible.
1)
Sending of contaminated pumps or devices
No units will be accepted if they are contaminated with micro-biological, explosive or radioactive substances. “Hazardous substances” are substances and com­pounds in accordance with the hazardous goods directive (current version). If pumps are contaminated or the declaration on contamination is missing, Pfeiffer Vac uum per form s decontamin ation at the sh ipper's expe nse.
! Neutralise the pump by flushing it with nitrogen or dry air. ! Close all openings airtight. ! Seal the p ump or unit in suitable protective film. ! Retur n the pump/u nit only in a suitable and sturdy transport con tainer an d send
it in while following applicable transport conditions.
Service orders
All service orders are carried out exclusively according to our repair conditions for vacuum units and components.
1)
Forms under www.pfeiffer-vacuum.net
29
Spare parts
11 Sp are pa rt s
Spare part pack­age/ Spare parts
Set of wearing parts
Silencer
Seal ing ring MVP 040 -2 P 3 529 143-C 1 (von 13 ) 2.3
Hose connection, complete
Hose, Polyethylene 10 x 8 mm
Gas ballast valve MVP 040-2 P 0920 634 E 1 40
Set of fus es
Diaphragm key
Pump type No. Pieces consisting of
the parts
MVP 040-2 PU E22 013 -T 1 10, 11
MVP 070-3 PU E22 014 -T 1 10, 11, 11a, 11b
MVP 070-3C PU E22 015 -T 1 10, 11
MVP 040-2
MVP 070-3 1
MVP 040-2 P 0920 542 E
MVP 040-2
MVP 070-3
MVP 040-2 MVP 070-3 MVP 070-3C
MVP 070-3 MVP 070-3C
P 0920 412 E
P 0988 088 1 m
PU E22 016 -T 2
PK 050 186 1
1
2.1
2 fuses, each 6,3 AT
30
12 Accessories
Designation MVP 040-2
Mains cable 115 / 230 V without plug, right angle IEC 320/C13 socket, 3 m
Mains cable 230 V with safety plug CEE 7, right angle IE C 320/C13 socket, 2 m
Mains cable 115 V with UL-plug, right angle IEC 320/C13 socket, 2 m
Scre w-in f lan ge DN 16 ISO-K F / G 1/4" inc l. seal (for MVP 040-2, MVP 070-3 inlet/outlet)
Backing pump relay box, single phase 5 A, for TC 11 0/ T C P 3 5 0
Backing pump relay box, single phase 5 A, for TC 400/TCP 350
PK 050 111
PK 050 109
PK 050 110
PK 050 114-T
PM 061 372-T
PM 061 374-T
Designation MVP 070-3
Mains cable 230 V with safety plug CEE 7, right angle IE C 320/C13 socket, 2 m
Mains cable 115 V with UL-plug, right angle IEC 320/C13 socket, 2 m
Mains cable 115 / 230 V without plug, right angle IEC 320/C13 socket, 3 m
Scre w-in f lan ge DN 16 ISO-K F / G 1/4" inc l. seal (for MVP 040-2, MVP 070-3 inlet/outlet)
Backing pump relay box, single phase 5 A, for TC 11 0/ T C P 3 5 0
Backing pump relay box, single phase 5 A, for TC 400/TCP 350
PK 050 109
PK 050 110
PK 050 111
PK 050 114-T
PM 061 372-T
PM 061 374-T
Accessories
Designation MVP 070-3 C
Mains cable 115 / 230 V without plug, right angle IEC 320/C13 socket, 3 m
Mains cable 115 V with UL-plug, right angle IEC 320/C13 socket, 2 m
Mains cable 230 V with safety plug CEE 7, right angle IE C 320/C13 socket, 2 m
Backing pump relay box, single phase 5 A, for TC 11 0/ T C P 3 5 0
Backing pump relay box, single phase 5 A, for TC 400/TCP 350
PK 050 111
PK 050 110
PK 050 109
PM 061 372-T
PM 061 374-T
Further detailed accessories are contained in the Pfeiffer Vacuum printed or Online Catalogue.
31
Technical data
13 Technical data
MVP 040-2
Parameter MVP 04 0-2, Diaph ra gm pump
Flange (in)
Flange (out) G 1/4" + silen cer
Nominal pumping speed at 50 Hz 2.3 m
Nominal pumping speed at 60 Hz 2.5 m
Ultimate pressure with gas ballast 5 mbar
Ultimate pressure without gas ballast 4 mbar
Exhaust pressure max. 1100 mbar
Leak rate 1·10
Sound pressu re level 51 (50Hz) 53 (60H z) dB (A)
Ambient temperature 12-40 °C
Motor rating 180 W
Rotation speed at 50 Hz 1500 1/min
Rotation speed at 60 Hz 1800 1/min
Mains requirement: voltage (selectable)
Current consumption
Switch Yes
Wei ght 11.4 k g
G 1/4" elbow union with enclosed hose 10/8, 1000 mm with an elbow union in G 1/4" at the end
3
/h
3
/h
-2
mbar l/s
10 0-115 V ( +/- 10%) 50 /60 Hz, 120 V (+/- 10%) 60 Hz, 200-230 V (+/- 10%) 50/60 Hz
10 0-115 V 5 0/60 Hz 3 ,4 A 120 V 60 Hz 3,4 A 200-230 V 50/60 Hz 1,8 A
MVP 070-3/-3C
Parameter MVP 07 0-3,
Diaphragm pump
G 1/4" elbow union + enc-
Flange (in)
Flange (out) G 1/4" + silencer Hose wave 10 mm
Nominal pumping speed at 50 Hz 3.8 m
Nominal pumping speed at 60 Hz 4.3 m
Ultimate pressure with gas ballast - 3 mbar
Ultimate pressure without gas ballast 1 mbar 1.5 mbar
Exhaust pressure max. 1100 mbar 1100 mbar
Leak rate 1·10
Sound pressu re level 52 (50 Hz); 55 (6 0 Hz) dB (A) 45 (50 Hz); 50 (60 Hz) dB (A)
Ambient temperature 12-40 ºC 12-40 ºC
Motor rating 250 W 250 W
Rotation speed at 50 Hz 1500 1/min 150 0 1/m in
Rotation speed at 60 Hz 1800 1/min 180 0 1/m in
Mains requirement: voltage (selec­table)
Current consumption
Switch Yes Yes
Wei ght 16. 4 k g 14. 3 k g
losed hose 10/8, 1000 mm with a elbow union in G 1/4" at the end
3
/h 3.4 m3/h
3
/h 3.8 m3/h
-2
mbar l/s 1·10-2 mbar l/s
10 0-115 V ( +/- 10 %) 5 0/6 0 Hz ; 120 V (+/- 10 %) 60 Hz ; 200-230 V (+/- 10 %) 50/60 Hz
10 0-115 V 5 0/60 Hz 5 ,7 A 120 V 60 Hz 5,7 A 200-230 V 50/60 Hz, 3,0 A
MVP 070-3 C, Diaphragm pump
For DN 16 ISO-K F
100-115 V (+/- 10 %), 50/6 0 Hz ; 120 V (+/- 10 %) , 60 Hz ; 200-230 V (+/- 10 %) , 50/60 Hz
100-115 V 50/60 Hz 5,7 A 120 V 60 Hz 5,7 A 200-230 V 50/60 Hz 3,0 A
32
13.1 Substances in contact with the media
MVP 040-2, MVP 070-3 MVP 070-3C
Aluminium, FPM, PE PTFE (carbon fibre reinforced), ETFE-CF, FFKM, ECTFE-CF
Technical data
33
Technical data
/
13.2 Dimensions
212
65
125
238 (W)
Fig. 14: MVP 040-2
106
EX - Schalldämpfer / silencer
) H
( 8
9 1
EX - Schalldämpfer / silencer
) H
( 8
9 1
Schlauchlänge 1000 mm tube lenght 1000 mm
IN - G1/4"
242 (L)
173
Schlauchlänge 1000 mm tube lenght 1000 mm
Gasballast gas ballast
22.5
IN - G1/4"
Gasballast / gas ballas t
65
125
238 (L)
65
125
242 (L)
Fig. 15: MVP 070-3
108
Fig. 16: MVP 070-3C
IN
für Kleinflansch DN 16/ for small flange DN 16
) H
( 8
9 1
EX - Schlauchwelle / hose nozzle DN 10
260
325 (W )
325 (W)
260
18
51
18
34
Declaration of conformity
according to the EC directive:
Machinery 2006/42/EC (Annex II, no. 1 A)
We hereby declare that the product cited below satisfies all relevant provisions of EC directive "Machinery" 2006/42/EC.
In addition, the product cited below satisfies all relevant provisions of EC directive "Electromagnetic Compatibility" 2004/108/EC .
The agent responsible for compiling the technical documentation is Mr. Sebastian Oberbeck, Pfeiffer Vacuum GmbH, Berliner Straße 43, 35614 Aßlar.
MVP 040-2, MVP 070-3/3C
Guidelines, harmonised standards and national standards and specifications which have been applied:
DIN EN ISO 12100-2 : 2004 DIN EN 61010-1 : 2002 DIN EN 61326-1 : 2006
DIN EN 1012-2 : 1996
Signatures:
(M.Bender) Managing Director
(Dr. M. Wiemer) Managing Director
Pfeiffer Vacuum GmbH Berliner Straße 43 35614 Asslar Germany
CE/2010
Vacuum i s nothing, b u t everything to us !
Tu r b o p u m p s
Rotary vane pumps
Roots pumps
Dry compressing pumps
Leak detectors
Val ve s
Components and feedthroughs
Vac uu m mea surem ent
Gas analysis
System engineering
Service
Pfeiffer Vacuum Technology AG · Headquarters/Germany
Te l. + 49 - (0 ) 6 4 4 1- 8 0 2- 0 · F a x + 4 9- ( 0) 6 4 4 1- 8 0 2- 2 0 2 · i n fo @ pf e if fe r- va cu um .d e · w w w. p fe i ff e r- va c uu m. n et
Loading...