Thank you for choosing a Pfeiffer Vacuum product. Your new turbopump is designed to
support you by its performance, its perfect operation and without interfering your individual
application. The name Pfeiffer Vacuum stands for high-quality vacuum technology, a
comprehensive and complete range of top-quality products and first-class service. With this
expertise, we have acquired a multitude of skills contributing to an efficient and secure
implementation of our product.
Knowing that our product must not interfere with your actual work, we are convinced that our
product offers you the solution that supports you in the effective and trouble-free execution of
your individual application.
Please read these operating instructions before putting your product into operation for the
first time. If you have any questions or suggestions, please feel free to contact info@pfeiffer-
vacuum.de.
Further operating instructions from Pfeiffer Vacuum can be found in the Download Center on
our website.
Disclaimer of liability
These operating instructions describe all models and variants of your product. Note that your
product may not be equipped with all features described in this document. Pfeiffer Vacuum
constantly adapts its products to the latest state of the art without prior notice. Please take
into account that online operating instructions can deviate from the printed operating
instructions supplied with your product.
Furthermore, Pfeiffer Vacuum assumes no responsibility or liability for damage resulting from
the use of the product that contradicts its proper use or is explicitly defined as foreseeable
misuse.
Copyright
This document is the intellectual property of Pfeiffer Vacuum and all contents of this
document are protected by copyright. They may not be copied, altered, reproduced or
published without the prior written permission of Pfeiffer Vacuum.
We reserve the right to make changes to the technical data and information in this document.
2/60
Table of contents
1About this manual7
1.1Validity7
1.2Target group7
1.3Conventions7
2Safety10
2.1General safety instructions10
2.2Safety instructions10
2.3Safety precautions14
2.4Limits of use of the product14
2.5Proper use15
2.6Foreseeable improper use15
Table of contents
1.1.1 Related documents7
1.1.2 Product variants affected7
1.3.1 Instructions in the text7
1.3.2 Pictographs8
1.3.3 Stickers on the product8
1.3.4 Abbreviations9
3Product description17
3.1Function17
3.1.1 Cooling17
3.1.2 Rotor bearing17
3.1.3 Drive18
3.2Shipment18
3.3Identifying the product18
3.3.1 Product types18
3.3.2 Product features18
4Transportation and Storage19
4.1Transport19
4.2Storage19
5Installation20
5.1Preparatory work20
5.2Connecting the high vacuum side20
5.2.1 Requirements for the dimensioning of a counter flange20
5.2.2 Considering earthquake protection21
5.2.3 Using a splinter shield or protective screen22
5.2.4 Using the vibration compensator22
5.2.5 Mounting orientations22
5.2.6 Attaching ISO-K flange onto ISO-K23
5.2.7 Attaching ISO-K flange to ISO-F23
5.2.8 Attach ISO-F flange to ISO-F25
5.2.9 Fastening CF flange to CF-F26
5.3Connecting the fore-vacuum side27
5.4Cooling water connection28
5.5Connecting accessories30
5.5.1 Connecting accessories to the turbopumps30
5.5.2 Connecting the venting valve31
5.5.3 Connecting the air cooling31
5.5.4 Connecting sealing gas32
5.5.5 Installing a heating jacket32
5.6Connecting the electrical supply33
5.6.1 Earthing the pump34
5.6.2 Create electrical connection34
3/60
Table of contents
6Operation36
6.1Commissioning36
6.2Operating modes37
6.2.1 Operation without operating unit37
6.2.2 Operation via multi-function connection "remote"37
6.2.3 Operation via connection "E74"37
6.2.4 Operation via Pfeiffer Vacuum display and control unit37
6.2.5 Operation via field bus37
6.3Switching on the turbopump38
6.4Operation monitoring38
6.4.1 Operating mode display via LED38
6.4.2 Temperature monitoring39
6.5Switching off and venting39
6.5.1 Switching off39
6.5.2 Venting40
7Maintenance41
7.1General maintenance information41
7.2Maintenance intervals and responsibilities41
7.3Replacing the operating fluid reservoir41
7.4Replacing the electronic drive unit43
7.5Confirming the rotation speed specification44
8Decommissioning46
8.1Shutting down for longer periods46
8.2Recommissioning46
8.3Disposing of the vacuum pump47
9Malfunctions48
10Service solutions from Pfeiffer Vacuum50
11Spare parts HiPace 300 P52
12Accessories53
13Technical data and dimensions54
13.1 General54
13.2 Technical data54
13.3 Dimensions58
Declaration of conformity59
4/60
List of tables
Tbl. 1:Abbreviations used in this document..........................................................................9
These operating instructions are for customers of Pfeiffer Vacuum. They describe the function of the
designated product and provide the most important information for safe usage of the product. The descriptions comply with applicable directives. All information provided in these operating instructions refer
to the current development status of the product. The documentation remains valid as long as the customer does not modify the product in any way.
1.1.1 Related documents
DocumentNumber
Operating instructions, Electronic drive unit TC 400. standardPT 0203 BN
Current documents are available from the Pfeiffer Vacuum Download Center.
1.1.2 Product variants affected
●
HiPace 300 P, DN 100 ISO-K, TC 400
●
HiPace 300 P, DN 100 CF-F, TC 400
●
HiPace 300 P, DN 100 ISO-F, TC 400
1.2 Target group
This operating instructions are aimed at all persons performing the following activities on the product:
●
transport,
●
setup (installation),
●
usage and operation,
●
decommissioning,
●
maintenance and cleaning,
●
storage or disposal.
The work described in this document is only permitted to be performed by persons with the appropriate
technical qualifications (expert personnel) or who have received the relevant training from Pfeiffer Vacuum.
1.3 Conventions
1.3.1 Instructions in the text
Usage instructions in the document follow a general structure that is complete in itself. The required action is indicated by an individual step or multi-part action steps.
Individual action step
A horizontal, solid triangle indicates the only step in an action.
7/60
CUS
TÜV Rheinland
Oil
S/N
Mod.
HiPace 300 P
PM P04 710 C
- - - - - - - -
P/N
n, f
MassMade in Germany 2018/11
PM 203 410 -T
60000 1/min, 1000 Hz
6.4 kg
D-35614 Asslar
Do not open!
before having read the
operating instructions
Nicht öffnen!
Erst Betriebsanweisung lesen
Prima d’aprire,
leggerele instruzioni d’uso
Pas ouvrir !
il faut lire d’abord
l’instruction de
service
warranty seal
About this manual
►
This is an individual action step.
Sequence of multi-part action steps
The numerical list indicates an action with multiple necessary steps.
1.
Step 1
2.
Step 2
3.
...
1.3.2 Pictographs
Pictographs used in the document indicate useful information.
Note
Tip
1.3.3 Stickers on the product
This section describes all the stickers on the product along with their meaning.
Rating plate
The rating plate of the turbopump is located on the lower part
of the vacuum pump.
Operating instructions note
This sticker indicates that this operating instructions must be
read before performing any tasks.
Protection class
This sticker describes protection class 1 for the product. The
positioning indicates the position for the ground connection.
Closure seal
The product is sealed ex-factory. Damaging or removing a closure seal results in loss of the warranty.
8/60
1
34
2
Fig. 2:Position of the labels on the product
1 Operating instructions note3 Note for ground connection
2 Closure seal4 Rating plate of the turbopump
About this manual
1.3.4 Abbreviations
Abbreviation Meaning in this document
CFFlange: Metal-sealed connector in accordance with ISO 3669
dDiameter value (in mm)
DCDirect current
DCUDisplay Control Unit (Pfeiffer Vacuum display and control unit).
DNNominal diameter as size description
fRotation speed value of a vacuum pump (frequency, in rpm or Hz)
PProcess execution
HPUHandheld Programming Unit. Aid for control and monitoring of pump parameters
HVHigh vacuum flange, high vacuum side
ISOFlange: Connector in accordance with ISO 1609 and ISO 2861
LEDIlluminating diode
PEEarthed conductor (protective earth)
[P:xxx]Electronic drive unit control parameters. Printed in bold as three-digit number in
remote26-pole D-Sub connecting socket on the turbopump electronic drive unit
S1S1 switch on power supply pack
WAFwidth across flats
TTemperature (in °C)
TCTurbopump electronic drive unit (turbo controller)
TPSVoltage supply (turbo power supply)
VVFore-vacuum flange, fore-vacuum connection
square brackets. Frequently displayed in conjunction with a short description.
Example: [P:312] software version
Tbl. 1:Abbreviations used in this document
9/60
Safety
2 Safety
2.1 General safety instructions
This document includes the following four risk levels and one information level.
DANGER
Imminent danger
Indicates a hazardous situation which, if not avoided, will result in death or serious injury.
►
Instructions on avoiding the hazardous situation
WARNING
Possibly imminent danger
Indicates a hazardous situation which, if not avoided, could result in death or serious injury.
►
Instructions on avoiding the hazardous situation
CAUTION
Possibly imminent danger
Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
►
Instructions on avoiding the hazardous situation
Danger of property damage
Notice is used to address practices not related to physical injury.
►
Instructions on avoiding property damage
2.2 Safety instructions
All safety instructions in this document are based on the results of the risk assessment carried out in
accordance with Machinery Directive 2006/42/EC Annex I and EN ISO 12100 Section 5. Where applicable, all life cycle phases of the product were taken into account.
Risks during transport
WARNING
Danger of serious injury due to falling objects
Due to falling objects there is a risk of injuries to limbs through to broken bones.
►
Take particular care and pay special attention when transporting products manually.
►
Do not stack the products.
►
Wear protective equipment, e.g. safety shoes.
NOTICE
Notes, tips or examples indicate important information on the product or on this
document.
10/60
Risks during installation
DANGER
Danger to life from electric shock
Power supply packs that are not specified or are not approved will lead to severest injuries up to
death.
►
Make sure that the power supply pack meets the requirements for double isolation between
mains input voltage and output voltage, in accordance with IEC 61010 and IEC 60950.
►
Make sure that the power supply pack meets the requirements in accordance with IEC 61010
and IEC 60950.
►
Where possible, use original power supply packs or only power supply packs that correspond
with the applicable safety regulations.
WARNING
Risk of danger to life through missing mains disconnection device
The vacuum pump and electronic drive unit are not equipped with a mains disconnection device
(mains switch).
►
Install a mains disconnection device according to SEMI-S2.
►
Install a circuit breaker with an interruption rating of at least 10,000 A.
WARNING
Risk of injury due to incorrect installation
Dangerous situations may arise from unsafe or incorrect installation.
►
Do not carry out your own conversions or modifications on the unit.
►
Ensure the integration into an Emergency Off safety circuit.
Safety
WARNING
Risk of cuts on moving, sharp-edged parts when reaching into the open high vacuum flange
With the high vacuum flange open, access to sharp-edged parts is possible. A manual rotation of the
rotor increases the danger situation. There is the risk of cuts, up to the separation of body parts (e.g.
fingertips). There is a risk of hair and loose clothing being drawn in. Objects falling in destroy the turbopump during subsequent operation.
►
Only remove the original protective covers immediately prior to connecting the high vacuum
flange.
►
Do not reach into the high vacuum connection.
►
Wear protective gloves during installation.
►
Do not start the turbopump with open vacuum connections.
►
Always carry out the mechanical installation before electrical connection.
►
Prevent access to the high vacuum connection of the turbopump from the operator side (e.g.
open vacuum chamber).
WARNING
Danger of cut injuries from unexpected start up.
The use of mating plugs of the electronic drive unit (accessories) enables the automatic run-up of the
vacuum pump as soon the power is turned on. Attaching mating plugs before or during the installation leads to the movement of parts hence the risk of cut injuries by sharp-edged in the exposed high
vacuum flange.
►
Only connect mating plugs after the mechanical installation.
►
Only switch on the vacuum pump immediately prior to operation.
11/60
Safety
WARNING
Risk of scalding from suddenly escaping hot cooling water
The turbopump water connections are open to both sides. When connecting the cooling water supply,
there is a risk of scalding from suddenly escaping, hot cooling water at over pressure.
►
Prior to installation, ensure that pressure is discharged from the cooling water system, and that it
is cooled down.
►
Wear protective equipment, e.g., safety goggles and gloves.
Risks during operation
WARNING
Risk of burns on hot surfaces when using additional equipment for heating during operation
The use of additional equipment for heating the vacuum pump or for optimizing the process generates very high temperatures on surfaces that can be touched. There is a risk of burning.
►
If necessary, set up a contact guard.
►
If necessary, apply the warning stickers provided for this at the danger points.
►
Ensure adequate cooling down before working on the vacuum pump or in its vicinity.
►
Wear protective equipment, e.g., gloves.
WARNING
Risk of serious injury in the event of vacuum pump destruction due to over pressure
Gas entry with very high over pressure results in destruction of the vacuum pump. There is a risk of
serious injury due to ejected objects.
►
Never exceed the permissible 1500 hPa (absolute) inlet pressure on the suction side or the venting and sealing gas connection.
►
Make sure that high, process-related over pressures cannot directly enter the vacuum pump.
CAUTION
Risk of injuries due to contact with vacuum when venting
While venting the vacuum pump there is a risk of minor injuries due to the direct contact of body parts
with the vacuum, e.g. hematomas.
►
Do not fully unscrew the venting screw out of the housing during venting.
►
Keep a distance from automatic venting device, such as venting valves.
Risks during maintenance
WARNING
Health hazard through poisoning from toxic contaminated components or devices
Toxic process media result in contamination of devices or parts of them. During maintenance work,
there is a risk to health from contact with these poisonous substances. Illegal disposal of toxic substances causes environmental damage.
►
Take suitable safety precautions and prevent health hazards or environmental pollution by toxic
process media.
►
Decontaminate affected parts before carrying out maintenance work.
►
Wear protective equipment.
WARNING
Risk of burns on hot surfaces when using additional equipment for heating during operation
The use of additional equipment for heating the vacuum pump or for optimizing the process generates very high temperatures on surfaces that can be touched. There is a risk of burning.
►
If necessary, set up a contact guard.
►
If necessary, apply the warning stickers provided for this at the danger points.
►
Ensure adequate cooling down before working on the vacuum pump or in its vicinity.
►
Wear protective equipment, e.g., gloves.
12/60
Safety
WARNING
Risk of cuts on moving, sharp-edged parts when reaching into the open high vacuum connection
Incorrect handling of the turbopump before maintenance work results in hazardous situations with
risk of injury. There is a risk of cuts from accessing sharp-edged, rotating parts when removing the
turbopump.
►
Wait until the turbopump comes to a standstill (rotation speed f=0).
►
Switch the turbopump off properly.
►
Secure the turbopump against re-start.
►
Close open connections immediately following removal, using the original protective cover.
CAUTION
Risk of injuries due to contact with vacuum when venting
While venting the vacuum pump there is a risk of minor injuries due to the direct contact of body parts
with the vacuum, e.g. hematomas.
►
Do not fully unscrew the venting screw out of the housing during venting.
►
Keep a distance from automatic venting device, such as venting valves.
CAUTION
Electric shock and damage to the vacuum pump and electronic drive unit due to improper disconnection of components
Even after the mains power is switched off, the turbopump continues to deliver electrical energy during its run-down period. If the turbopump and electronic drive unit are disconnected prematurely,
there is the risk of electric shock from contact with live components. There is a risk of electric body
contact and, as a result, the destruction of electronic components.
►
Never disconnect the turbopump and electronic drive unit from each other if power is still connected or if the rotor is running.
►
Monitor the turbopump rotation via the parameters available in the electronic drive unit (e.g., [P:398]).
►
Wait until the turbopump comes to a standstill (rotation speed f=0).
Risks in the event of malfunctions
WARNING
Danger to life from electric shock in the event of a fault
In the event of a fault, devices connected to the mains may be live. There is a danger to life from
electric shock when making contact with live components.
►
Always keep the mains connection freely accessible so you can disconnect it at any time.
WARNING
Danger to life from the turbopump breaking away in the event of a fault
Sudden blocking of the rotor generates high destructive torques in accordance with ISO 27892. If the
turbopump is not properly secured, it will tear off. The energy released in this way can propel the entire turbopump or shattered pieces from its interior through the surrounding area. Potentially dangerous gases may escape. There is a risk of serious injury, potentially even fatal, and significant equipment damage.
►
Follow the installation instructions for this turbopump.
►
Observe the requirements regarding stability and design of the counter flange.
►
Use only original accessories or fixing material approved by Pfeiffer Vacuum for the installation.
13/60
Safety
WARNING
Risk of injury caused by the turbopump breaking away with the vibration compensator in the
event of a malfunction
Sudden jamming of the rotor generates high destructive torques in accordance with ISO 27892.
When using a vibration compensator, this will probably lead to the turbopump being sheared off in
use. The energy that this would release could throw the entire turbopump or shattered pieces from its
interior through the surrounding space. Potentially dangerous gases can escape. There is a risk of
very serious injuries, including death, and extensive property damage.
►
Take suitable safety precautions on-site for the compensation of the occurring torques.
►
Before installing a vibration compensator, you must first of all contact Pfeiffer Vacuum.
WARNING
Danger to life from poisoning where toxic process media leak from damaged connections
Sudden twisting of the turbopump in the event of a fault causes fittings to accelerate. There is the risk
of damaging on-site connections (e.g., fore-vacuum line) and resulting leaks. This results in leakage
of process media. In processes involving toxic media, there is a risk of injury and danger to life due to
poisoning.
►
Keep masses connected to the turbopump as low as possible.
►
Use flexible lines to connect to the turbopump where necessary.
2.3 Safety precautions
General safety precautions when handling the product
►
Observe all applicable safety and accident prevention regulations.
►
Check that all safety measures are observed at regular intervals.
►
Do not expose body parts to the vacuum.
►
Always ensure a secure connection to the earthed conductor (PE).
►
Never disconnect plug connections during operation.
►
Observe the above shutdown procedures.
►
Before working on the high vacuum connection, wait until the rotor has stopped completely (rotation speed f = 0).
►
Never put the device into operation with the high vacuum connection open.
►
Keep lines and cables away from hot surfaces (> 70°C).
►
Never fill or operate the unit with cleaning agents or cleaning agent residues.
►
Do not carry out your own conversions or modifications on the unit.
►
Observe the unit protection class prior to installation or operation in other environments.
Duty to provide information on potential dangers
The product holder or user is obliged to make all operating personnel aware of dangers posed by this product.
Every person who is involved in the installation, operation or maintenance of the
product must read, understand, and adhere to the safety-related parts of this document.
Infringement of conformity due to modifications to the product
The Declaration of Conformity from the manufacturer is no longer valid if the operator changes the original product or installs additional equipment.
●
Following installation into a system, the operator is required to check and reevaluate as necessary the conformity of the overall system in the context of the
relevant european Directives before commissioning that system.
Ambient temperature5 °C to 30 °C with convection cooling with-
out gas throughput
5°C to 35°C with air cooling
5°C to 40°C with water cooling
Permissible surrounding magnetic field5.5 mT
Maximum irradiated thermal output2.4 W
Maximum permissible rotor temperature of the turbopump
Tbl. 2:Permissible ambient conditions
110 °C
Notes on ambient conditions
The specified permissible ambient temperatures apply to operation of the turbopump at maximum permissible backing pressure or at maximum gas throughput,
depending on the cooling type. The turbopump is intrinsically safe thanks to redundant temperature monitoring.
●
The reduction in backing pressure or gas throughput permits operation of the
turbopump at higher ambient temperatures as well.
●
If the maximum permissible operating temperature of the turbopump is exceeded, the electronic drive unit first reduces the drive output and switches it off
where necessary.
2.5 Proper use
►
Use the turbopump only for generating vacuum.
►
Use the turbopump only in combination with a suitable backing pump that can deliver up to the
required maximum fore-vacuum pressure.
►
Use the turbopump only in closed indoor areas.
►
Use the turbopump only for the evacuation of dry and inert gases.
►
Use the turbopump for the evacuation of corrosive gases only in conjunction with sealing gas.
2.6 Foreseeable improper use
Improper use of the product invalidates all warranty and liability claims. Any use that is counter to the
purpose of the product, whether intentional or unintentional, is regarded as misuse.
●
Transport, installation or operation of the vacuum pump in an impermissible spatial position
●
Establishing the voltage supply without correct installation
●
Installing the pump with unspecified mounting material
●
Pumping explosive media
●
Pumping corrosive media without using sealing gas
●
Pumping dust without using sealing gas
●
Pumping of fluids
●
Operation with impermissible high gas throughput
●
Operation with impermissible high fore-vacuum pressure
●
Operation with excessively high irradiated heat output
●
Operation in impermissible high magnetic fields
●
Operation in an incorrect gas mode
15/60
Safety
●
Venting with impermissible high venting rates
●
Use for pressure generation
●
Use in areas with ionizing radiation
●
Operation in explosion-hazard areas
●
Use in systems in which sporadic loads and vibrations or periodic forces act on the device
●
The causing of hazardous operating conditions by a presetting on the electronic drive unit that is
contrary to the process
●
Use of accessories or spare parts that are not listed in these instructions
16/60
3 Product description
1312
2
3
14
1
15
9
8
4
7
10
6
5
11
3.1 Function
The turbopump forms a compact unit with the electronic drive unit. The Pfeiffer Vacuum power supply
packs serve as voltage supply.
The feature description "P" designates turbopumps suitable for the process. They are insensitive
against dust and corrosive gases.
Product description
3.1.1 Cooling
3.1.2 Rotor bearing
Fig. 3:Design 300 P (example DN 100 ISO-K and DN 100 CF-F)
2 Turbo rotor10 Connecting panel of the electronic drive unit
3 High vacuum connection, DN 100
CF-F
4 Ground terminal12 LED operating mode display
5 Pump base13 Electronic drive unit TC 400
6 Sealing gas connection14 Protective cover for the high vacuum connection
7 Venting screw15 Pump housing, aluminum
8 Cooling water connection
●
Water cooling
●
Air cooling (optional)
The electronic drive unit automatically regulates the drive power down in the event of excessive temperatures.
Hybrid-bearing turbopump
●
High vacuum side: wear-free permanent magnetic bearing
●
Fore-vacuum side: ball bearing with ceramic balls
Permanent lubrication of the rotor bearing on the fore-vacuum side is ensured by an operating fluid reservoir.
11 Connecting plug for voltage supply DCin
17/60
Product description
3.1.3 Drive
●
Electronic drive unit TC 400
─
Operating voltage 24 V DC
●
Electronic drive unit TC 400
─
Operating voltage 48 V DC
3.2 Shipment
●
Turbopump with electronic drive unit
●
Protective cover for the high vacuum connection
●
Protective cover for the fore-vacuum connection
●
Mating plug for the "remote" connection to the TC 400 (type-dependent)
●
Mating plug for the "E74" connection to the TC 400 (type-dependent)
●
Operating instructions
3.3 Identifying the product
►
To ensure clear identification of the product when communicating with Pfeiffer Vacuum, always
keep all of the information on the rating plate to hand.
►
Learn about certifications through test seals on the product or at www.tuvdotcom.com with company ID no. 000021320.
3.3.1 Product types
The product designation of Pfeiffer Vacuum turbopumps from the HiPace series is composed of the
family name, the size (which is based on the pumping speed of the vacuum pump) and, if required, an
additional feature description.
FamilySize/modelProperty, attribute, feature
HiPace10 to 2800none = Standard version
Tbl. 3:Product designation of Pfeiffer Vacuum HiPace turbopumps
3.3.2 Product features
mini = Compact version
U = Overhead version
C = Corrosive gas version
P = Process
M = Active magnetic bearing
T = Temperature management
E = High efficiency
H = high compression
I = ion implantation
18/60
FeatureVersion
HV flangeDN 100 ISO-KDN 100 ISO-FDN 100 CF-F
Flange materialAluminiumAluminiumStainless steel
Tbl. 4:Turbopump features
4 Transportation and Storage
4.1 Transport
WARNING
Danger of serious injury due to falling objects
Due to falling objects there is a risk of injuries to limbs through to broken bones.
►
Take particular care and pay special attention when transporting products manually.
►
Do not stack the products.
►
Wear protective equipment, e.g. safety shoes.
Pfeiffer Vacuum recommends keeping the transport packaging and original protective cover.
Instructions for safe transport
1.
Transport the turbopump only within the permissible temperature limits.
2.
Observe weight specified on the rating plate.
3.
Where possible, always transport or ship the turbopump in its original packaging.
4.
Always carry the turbopump with both hands.
5.
Remove the protective cover only immediately prior to installation.
Transportation and Storage
4.2 Storage
Pfeiffer Vacuum recommends storing the products in their original transport packaging.
Storing the turbopump
1.
Close all flange openings with the original protective caps.
2.
Close all other connections (e.g. venting connection) with the respective original parts.
3.
Store the turbopump only indoors within the permissible temperature limits.
4.
In rooms with humid or aggressive atmospheres: Hermetically seal the turbopump together with a
drying agent in a plastic bag.
19/60
Installation
5 Installation
The installation of the turbopump and its fastening is of outstanding importance. The rotor of the turbopump revolves at very high speed. In practice it is not possible to exclude the risk of the rotor touching
the stator (e.g. due to the penetration of foreign bodies into the high vacuum connection). The kinetic
energy released acts on the housing and on the anchoring of the turbopump within fractions of a second.
Comprehensive tests and calculations conforming to ISO 27892 confirm the safety of the turbopump
both against crashes (destruction of the rotor blade) and against bursting (breakage of the rotor shaft).
The experimental and theoretical results are expressed in safety measures and recommendations for
the correct and safe fastening of the turbopump.
5.1 Preparatory work
WARNING
Risk of cuts on moving, sharp-edged parts when reaching into the open high vacuum flange
With the high vacuum flange open, access to sharp-edged parts is possible. A manual rotation of the
rotor increases the danger situation. There is the risk of cuts, up to the separation of body parts (e.g.
fingertips). There is a risk of hair and loose clothing being drawn in. Objects falling in destroy the turbopump during subsequent operation.
►
Only remove the original protective covers immediately prior to connecting the high vacuum
flange.
►
Do not reach into the high vacuum connection.
►
Wear protective gloves during installation.
►
Do not start the turbopump with open vacuum connections.
►
Always carry out the mechanical installation before electrical connection.
►
Prevent access to the high vacuum connection of the turbopump from the operator side (e.g.
open vacuum chamber).
General notes for the installation of vacuum components
►
Choose an installation location that permits access to the product and to supply lines at all times.
►
Observe the ambient conditions given for the area of use.
►
Provide the highest possible level of cleanliness during assembly.
►
Ensure that flange components during installation are grease-free, dust-free and dry.
Select the installation location
1.
Observe the instructions for transport to the installation location.
2.
Make sure that there are sufficient cooling options for the turbopump.
3.
Install suitable shielding if the surrounding magnetic fields exceed the permissible levels.
4.
Install suitable shielding so that the irradiated thermal output does not exceed the permissible values when high temperatures occur due to the process.
5.
Observe the permissible temperatures for the vacuum connection.
5.2 Connecting the high vacuum side
5.2.1 Requirements for the dimensioning of a counter flange
NOTICE
Risk of damage due to incorrect counter flange design
Unevenness on the operator-side counter flange results in stresses in the vacuum pump housing,
even when properly attached. This can produce leakage or negative changes in running characteristics.
►
Adhere to the shape tolerances for the counter flange.
►
Observe the maximum flatness deviations over the entire surface.
20/60
VACUUM CHAMBER
Installation
The assembly of vertical superstructural parts on the high vacuum connection is the
responsibility of the operating company. The loading capacity of the high vacuum
flange is specific for the turbopump used. The total weight of superstructural parts
must not exceed the maximum values specified.
If the rotor is suddenly blocked, the torques arising from the system and the high vacuum flange must
be absorbed. The installation elements for turbo pumps are special designs by Pfeiffer Vacuum.
Maximum
torque occurring in
the event of
a burst
2000 Nm500 N
Tbl. 5:Requirements for the dimensioning of customer-specific high vacuum connection
Important information for correct installation
1)
►
Only use the approved mounting kits of Pfeiffer Vacuum for the high vacuum connection of the
turbopump.
Maximum
permissible
axial load on
the high vac-
uum flange
(equivalent to
50 kg)
2)
FlatnessMinimum ten-
± 0.05 mm
5.2.2 Considering earthquake protection
Vacuum pump damage caused by external vibrations
In the event of earthquakes or other external vibrations, there is the risk of the rotor coming into contact with the safety bearings, or the housing wall touching the turbopump. This can produce mechanical loads up to and including destruction of the turbopump.
►
Make sure that all flange and safety connections absorb the resulting forces.
►
Secure the vacuum chamber against displacement or tipping.
sile strength
of the flange
material in all
operating
states
170 N/mm
270 N/mm
NOTICE
Engagement
depth of
the fixing
screws
2
2.5 × d5.5 mT2.4 W
2
1.5 × d
Maximum
permissible surrounding
magnetic
field
Maximum
permissible irradiated
thermal
output
Fig. 4:Example: Safety connections
The theoretically calculated torque in the event of a burst (rotor shaft breakage) according
1)
to ISO 27892 was not reached in any experimental test.
2)A one-sided load is not permitted.
21/60
VV
± 25° max
HV
VV
VV
Installation
Safety connection, customer-side
5.2.3 Using a splinter shield or protective screen
Pfeiffer Vacuum centering rings with splinter shield or protective screen in the high vacuum flange protect the Turbopump against foreign matter from the vacuum chamber. The pumping speed is reduced
according to the passage guide values and the size of the high vacuum flange.
Flange sizeReduced pumping speed in % by gas type
H
2
Splinter shield DN 100572424
Protective screen DN 10022108
Tbl. 6:Reduction of the pumping speed when using a splinter shield or protective screen
►
With ISO flanges, use centering rings with protective screen or splinter shield.
5.2.4 Using the vibration compensator
Pfeiffer Vacuum vibration compensators are suitable for use on vibration-sensitive systems.
HeN
2
Ar
WARNING
Risk of injury caused by the turbopump breaking away with the vibration compensator in the
event of a malfunction
Sudden jamming of the rotor generates high destructive torques in accordance with ISO 27892.
When using a vibration compensator, this will probably lead to the turbopump being sheared off in
use. The energy that this would release could throw the entire turbopump or shattered pieces from its
interior through the surrounding space. Potentially dangerous gases can escape. There is a risk of
very serious injuries, including death, and extensive property damage.
►
Take suitable safety precautions on-site for the compensation of the occurring torques.
►
Before installing a vibration compensator, you must first of all contact Pfeiffer Vacuum.
Installing the vibration compensator
1.
Install the vibration compensator only with vertical passage.
2.
Consider the flow resistance.
3.
Secure the turbopump additionally to the high vacuum flange.
4.
Observe the fastening of the ISO flanges.
5.2.5 Mounting orientations
Pfeiffer Vacuum turbopumps from the HiPace series are suitable for use with dry compressing backing
pumps for mounting in all orientations.
►
When using oil-sealed backing pumps, avoid backflow from the fore-vacuum range.
22/60
Fig. 5:Recommended alignment when using oil-sealed backing pumps
Specify a horizontal mounting orientation for oil-sealed backing pumps
1.
Always align the fore-vacuum connection downwards vertically.
–
Permissible deviation ± 25°
2.
Support the tube connections in front of the turbopump.
3.
Do not allow any forces from the piping system to act on the turbopump.
4.
Do not load the high vacuum flange of the turbopump on one side.
5.2.6 Attaching ISO-K flange onto ISO-K
ISO flange connections
For the connection of flanges in ISO-KF or ISO-K design, twisting may occur in the
event of sudden blockage of the rotor, despite correct installation.
●
Leak-tightness of the flange connection, however, is not jeopardized in this regard.
Fig. 6:Flange connection ISO-K to ISO-F, bracket screws
Connection with bracket screw
1.
For the connection of the turbopump, use only the approved mounting kits from Pfeiffer Vacuum.
2.
Connect the flange with the components of the mounting kit according to the figure.
3.
Use for all prescribed components for the turbopump.
4.
Tighten the bracket screws cross-wise in 3 steps.
–
Tightening torque: 5, 15, 25 ± 2 Nm
5.2.7 Attaching ISO-K flange to ISO-F
The connection types for the installation ISO-K flange with ISO-F flange are:
●
"Hexagon head screw and tapped hole"
●
"Stud screw with tapped hole"
●
"Stud screw with through hole"
ISO flange connections
For the connection of flanges in ISO-KF or ISO-K design, twisting may occur in the
event of sudden blockage of the rotor, despite correct installation.
●
Leak-tightness of the flange connection, however, is not jeopardized in this regard.
Fig. 13:Flange connection CF-F, hexagon head screw and through hole
Connection of the hexagon head screw and through holes
1.
For the connection of the turbopump, use only the approved mounting kits from Pfeiffer Vacuum.
2.
If used: Insert the protective screen or splinter shield with clamping lugs downwards in the turbopump high vacuum flange.
3.
Place the seal exactly in the hollow.
4.
Connect the flange with the components of the mounting kit according to the figure.
5.
Tighten the screw couplings all the way around.
–
Tightening torque: 22 ± 2 Nm
6.
Then check the torque, since flowing of the sealing material may make it necessary to re-tighten
the screws.
26/60
Fig. 14:Flange connection CF-F, stud screw and tapped hole
Connection of the stud screw and tapped hole
1.
For the connection of the turbopump, use only the approved mounting kits from Pfeiffer Vacuum.
2.
Screw in the required number of stud screws with the shorter end in the holes on the counter
flange.
3.
If used: Insert the protective screen or splinter shield with clamping lugs downwards in the turbopump high vacuum flange.
4.
Place the seal exactly in the hollow.
5.
Connect the flange with the components of the mounting kit according to the figure.
6.
Tighten the screw couplings all the way around.
–
Tightening torque: 22 ± 2 Nm
7.
Then check the torque, since flowing of the sealing material may make it necessary to re-tighten
the screws.
Fig. 15:Flange connection CF-F, stud screw and through hole
Installation
Connection of the stud screw and through hole
1.
For the connection of the turbopump, use only the approved mounting kits from Pfeiffer Vacuum.
2.
If used: Insert the protective screen or splinter shield with clamping lugs downwards in the turbopump high vacuum flange.
3.
Place the seal exactly in the hollow.
4.
Connect the flange with the components of the mounting kit according to the figure.
5.
Tighten the screw couplings all the way around.
–
Tightening torque: 22 ± 2 Nm
6.
Then check the torque, since flowing of the sealing material may make it necessary to re-tighten
the screws.
5.3 Connecting the fore-vacuum side
WARNING
Danger to life from poisoning where toxic process media leak from damaged connections
Sudden twisting of the turbopump in the event of a fault causes fittings to accelerate. There is the risk
of damaging on-site connections (e.g., fore-vacuum line) and resulting leaks. This results in leakage
of process media. In processes involving toxic media, there is a risk of injury and danger to life due to
poisoning.
►
Keep masses connected to the turbopump as low as possible.
►
Use flexible lines to connect to the turbopump where necessary.
Suitable backing pump
Use the turbopump only in combination with a suitable backing pump that can deliver up to the required maximum fore-vacuum pressure. To achieve the fore-vacuum
pressure, use a suitable vacuum pump or a pumping station from the Pfeiffer Vacuum range.
In this case, the backing pump is also controlled directly via the turbopump electronic drive unit interfaces (e.g., relay box or connection cable).
27/60
1
2
3
4
5
Installation
Fig. 16:Example of fore-vacuum connection on HiPace 300
With rigid pipe connections, include bellows to attenuate external vibrations.
2.
Install a fore-vacuum connection with small flange components, e.g. connection elements and
pipe components DN 16 ISO-KF from the Pfeiffer Vacuum Components Shop.
3.
Implement measures to counteract the backflow of operating fluids or condensate from the forevacuum area.
4.
Observe the information in the operating instructions of the backup pump or pumping station when
connecting and operating it.
5.4 Cooling water connection
The HiPace 300 turbopumps use water cooling as standard.
WARNING
Risk of scalding from suddenly escaping hot cooling water
The turbopump water connections are open to both sides. When connecting the cooling water supply,
there is a risk of scalding from suddenly escaping, hot cooling water at over pressure.
►
Prior to installation, ensure that pressure is discharged from the cooling water system, and that it
is cooled down.
►
Wear protective equipment, e.g., safety goggles and gloves.
Recommendations for cooling the turbopump
1.
Use additional cooling in the event of increased fore-vacuum pressure (> 0.1 hPa).
2.
Use additional cooling if operating with high gas throughput.
3.
Always use water cooling at ambient temperatures > 35 °C.
28/60
ParameterCooling water
1
2
3
4
Appearance
pH value7 to 9
Carbonate hardness, max.10 °dH
Chloride content, max.100 mg/l
Sulfate content, max.240 mg/l
Carbonic acid content, max.not detectable
Ammonia content, max.not detectable
Electrical conductivity, max.500 µS/cm
Particle size, max.150 µm
Cooling water temperatureSee "Technical data"
Cooling water consumptionSee "Technical data"
Feed line overpressure, max.6000 hPa
●
filtered
●
mechanically clear
●
visually clear
●
no turbidity
●
no sediment
●
free from grease and oil
12.53 °e
17.8 °fH
178 ppm CaC0
3
Installation
Tbl. 7:Requirements on the cooling water composition
Connection to the turbopumpExternal cooling water supply
Banjo fitting with insertion systemHose cable
External hose diameter 8 mm
Internal hose diameter 6 mm
Tbl. 8:Requirements on the cooling water connection
Requirements: The customer-side cooling system is closed and depressurized. The provided hose
connections correspond to the turbopump connections.
1.
If necessary, use a dirt trap in the feed line.
2.
Insert the hoses for the cooling water feed line and return line as far as they will go into respective
connections on the turbopump.
3.
Adjust the banjo fittings according to your requirements.
–
Tightening torque for the banjo fitting: 3 to 3.5 Nm
5.5 Connecting accessories
Installation and operation of accessories
Pfeiffer Vacuum offers a series of special, compatible accessories for its products.
●
Information and ordering options for approved accessories can be found online.
●
The following accessories are not included in the scope of supply.
5.5.1 Connecting accessories to the turbopumps
Electronic drive unit TC 400 accessory connection
The electronic drive unit of the turbopump offers space for the connection of maximum 4 accessory devices. M12 sockets with the designation "accessory" are available for this purpose.
●
The accessory connections have been preconfigured at the factory.
●
After connecting pre-configured accessory devices, these are immediately
ready for operation according to the factory settings.
●
The use of other accessories for turbopumps is possible and requires settings
in the configuration of the electronic drive unit.
●
The desired accessory output is configured via RS-485 using Pfeiffer Vacuum
display and control units or a PC.
●
You can find more detailed information in the “Electronic drive unit TC|400” operating manual.
Electronic drive
unit connection
Acc. AA1Y-1Fan (continuous opera-
Acc. BB1Y-1Venting valve
Accessory
connection
A2Y-2backing pump
B2Y-2Heater
Y-connector
Pre-set configuration
tion)
30/60
Tbl. 9:Factory pre-set accessory connections on the electronic drive unit
Connecting accessory devices
1.
Observe the installation instructions in the operating instructions for the relevant accessory.
2.
Note the existing configuration of existing connections and control lines.
3.
Connect only matching accessory devices to the electronic drive unit.
4.
Use the Y-connector from the scope of supply if you want to connect 3 or 4 devices.
Using additional accessories
1.
Observe the installation instructions in the operating instructions for the relevant accessory.
2.
Note the existing configuration of existing connections.
3.
Use the Pfeiffer Vacuum display and control unit DCU 002, or a DCU with integrated power supply
pack.
5.5.2 Connecting the venting valve
1
3
2
4
A Pfeiffer Vacuum venting valve serves the automatic venting of a turbopump after switching off or in
event of a power failure.
●
The permissible inlet pressure for venting gas is 1,500 hPa absolute.
Connecting the venting valve
1.
Unscrew the venting screw with sealing ring out of the turbopump by hand.
2.
Screw the venting valve with the outlet side into the venting connection of the turbopump.
3.
Screw the control line of the venting valve into the corresponding connection on the electronic
drive unit.
Establish the venting gas supply
In case you want to use dry room air as venting gas, the venting valve is ready for operation immediately.
In case you want to use another inert gas (e.g. nitrogen N2) as venting gas, follow the handling steps.
1.
Provide an external venting gas supply with a maximum inlet pressure of 1,500 hPa absolute.
2.
Connect the venting gas supply to the inlet side (G 1/8") of the valve.
3.
If required, use a connecting adapter from the accessory of the valve (not included in the scope of
supply).
5.5.3 Connecting the air cooling
Installation
Operational conditions for the air cooling
Air cooling is permitted up to a permissible ambient temperature of + 35 °C!
For processes with smaller gas loads and good fore-vacuum pressure, air cooling (optional) is possible
instead of water cooling.
Fig. 18:Connecting the air cooling
1 Pump base3 Air cooling
2 Allen head screw4 Countersunk screw
Required tools
●
Allen key, size 2.5
●
Allen key, size 3
Procedure
1.
Attach the air cooling to the intended bores on the pump base with 2 screws.
2.
Screw the control line of the sealing gas valve into a free accessory connection on the electronic
drive unit.
3.
If necessary, configure the accessory output for the sealing gas via the interfaces of the electronic
drive unit.
–
The specification for the air cooling is "accessory A1".
31/60
Installation
5.5.4 Connecting sealing gas
The use of sealing gas serves to protect the turbopump in dusty processes, corrosive processes or with
excessive gas throughputs. Sealing gas prevents the ingress of damaging substances into the motor
and bearing area. The supply is carried out either via a sealing gas valve or a sealing gas throttle without control.
●
Pfeiffer Vacuum recommends the use of sealing gas from 50 % of the maximum gas throughput.
●
The permissible inlet pressure for sealing gas is 1,500 hPa absolute.
●
The flow rate for sealing gas is at 12.5 to 15 sccm.
Connecting the sealing gas valve
1.
Unscrew the screw plug with sealing ring out of the sealing gas connection of the turbopump.
2.
Screw the sealing gas valve with the outlet side into the sealing gas connection of the turbopump.
3.
Screw the control line of the sealing gas valve into a free accessory connection on the electronic
drive unit.
4.
Configure the accessory output for the sealing gas via the interfaces of the electronic drive unit.
Connecting the sealing gas throttle without control valve
1.
Unscrew the screw plug with sealing ring out of the sealing gas connection of the turbopump.
2.
Screw the sealing gas throttle with the outlet side into the sealing gas connection of the turbopump.
Establishing the sealing gas supply
In case you want to use room air as sealing gas, the sealing gas valve is ready for operation immediately.
In case you want to use another inert gas (e.g. nitrogen N2) as sealing gas, follow the handling steps.
1.
Provide an external sealing gas supply with a maximum inlet pressure of 1,500 hPa absolute.
2.
Connect the sealing gas supply to the inlet side (1/8" thread) of the valve.
3.
If required, use a connecting adapter from the accessory of the valve (not included in the scope of
supply).
5.5.5 Installing a heating jacket
Vacuum pump destruction due to excessive energy input during operation
Simultaneous loading by means of high drive power (gas throughput, fore-vacuum pressure), high
heat radiation, or strong magnetic fields results in uncontrolled heating of the rotor and can destroy
the vacuum pump.
►
Consult Pfeiffer Vacuum before combining varying loads on the vacuum pump. Lower limit values apply.
Operational conditions for the housing heating
●
●
The turbopump and vacuum chamber must be baked out to reach the ultimate pressure more quickly.
The heating jacket may only be used in turbopumps with high vacuum flange in stainless steel design.
The heating duration depends on the degree of contamination in the process, as well as the final pressure to be reached, and should be at least 4 hours.
NOTICE
When using housing heating and water cooling, the max. permissible temperature at the connection flange of the vacuum chamber is 120°C.
The maximum permissible rotor temperature is 110 °C.
For high temperatures that occur due to the process, the thermal output irradiated must not exceed 2.4 W. If necessary, install suitable screening sheets (information on request).
32/60
(2x)
1
2
4
5
3
Fig. 19:Connection of the heating jacket
1 Pump housing4 Heating relay box
2 Allen head screw5 "Hot surface" warning sticker
3 Heating jacket
Required tools
●
Allen key, size 4 or 5
Installing a heating jacket
1.
Carefully bend out the heating jacket at the outer clamping strap.
2.
Push the jacket onto the cylindrical section of the pump body from the side.
3.
Care should be taken that the heater bands do not become deformed and that the jacket is resting
flush against the body.
4.
Secure the heating jacket onto the body with the Allen head screw.
5.
Apply the sticker "Hot surface" on the pump housing.
6.
Install suitable shielding plates as protection against the irradiated heat input exceeding the maximum values.
Installation
Fixing screw
thread
M56 Nm7 Nm7 Nm
M611 Nm12 Nm12 Nm
Tbl. 10:Tightening torque for heating jacket fastening screws
Connecting the control line
1.
Connect the accessory control line to the electronic drive unit using the corresponding accessory
connection, and tighten the plug securely.
2.
Check the specification in the configuration "accessory B1".
3.
Establish the power supply for the relay box in accordance with the operating instructions for the
accessory.
Tightening torque in
cold condition
5.6 Connecting the electrical supply
WARNING
Risk of danger to life through missing mains disconnection device
The vacuum pump and electronic drive unit are not equipped with a mains disconnection device
(mains switch).
►
Install a mains disconnection device according to SEMI-S2.
►
Install a circuit breaker with an interruption rating of at least 10,000 A.
Tightening torque for
bakeout
One-time retightening
after cooling
33/60
M4
Installation
WARNING
Risk of injury due to incorrect installation
Dangerous situations may arise from unsafe or incorrect installation.
►
Do not carry out your own conversions or modifications on the unit.
►
Ensure the integration into an Emergency Off safety circuit.
5.6.1 Earthing the pump
Pfeiffer Vacuum recommends connecting a suitable grounding cable to discharge applicative interferences.
Fig. 20:Example: Connecting the grounding cable
Procedure
1.
Use the turbopump ground terminal (M4 female thread).
2.
Route the connection in accordance with locally applicable provisions.
5.6.2 Create electrical connection
DANGER
Danger to life from electric shock
Power supply packs that are not specified or are not approved will lead to severest injuries up to
death.
►
Make sure that the power supply pack meets the requirements for double isolation between
mains input voltage and output voltage, in accordance with IEC 61010 and IEC 60950.
►
Make sure that the power supply pack meets the requirements in accordance with IEC 61010
and IEC 60950.
►
Where possible, use original power supply packs or only power supply packs that correspond
with the applicable safety regulations.
WARNING
Danger to life from electric shock in the event of a fault
In the event of a fault, devices connected to the mains may be live. There is a danger to life from
electric shock when making contact with live components.
►
Always keep the mains connection freely accessible so you can disconnect it at any time.
34/60
DC out
S1
TPS / DCU
Electronic
Drive Unit
DC in
PM 061 352 -T
Installation
WARNING
Danger of cut injuries from unexpected start up.
The use of mating plugs of the electronic drive unit (accessories) enables the automatic run-up of the
vacuum pump as soon the power is turned on. Attaching mating plugs before or during the installation leads to the movement of parts hence the risk of cut injuries by sharp-edged in the exposed high
vacuum flange.
►
Only connect mating plugs after the mechanical installation.
►
Only switch on the vacuum pump immediately prior to operation.
Fig. 21:Connecting the electronic drive unit to a power supply pack
Connecting the electronic drive unit
1.
Make sure that you have the correct supply voltage.
2.
Make sure that the power supply pack main switch is off prior to connection.
3.
Use a matching connecting cable from the Pfeiffer Vacuum accessories program.
4.
Insert the connecting cable into the connection "DCin" on the electronic drive unit and close the
bayonet lock.
5.
Insert the connecting cable into the connection "DCout" on the power supply pack and close the
bayonet lock.
35/60
Operation
6 Operation
6.1 Commissioning
Important settings and function-related variables are factory-programmed into the vacuum pump electronic drive unit as parameters. Each parameter has a three-digit number and a description. Parameterdriven operation and control is supported via Pfeiffer Vacuum displays and control units, or externally
via RS-485 using Pfeiffer Vacuum protocol.
WARNING
Danger of cut injuries from unexpected start up.
The use of mating plugs of the electronic drive unit (accessories) enables the automatic run-up of the
vacuum pump as soon the power is turned on. Attaching mating plugs before or during the installation leads to the movement of parts hence the risk of cut injuries by sharp-edged in the exposed high
vacuum flange.
►
Only connect mating plugs after the mechanical installation.
►
Only switch on the vacuum pump immediately prior to operation.
Vacuum pump destruction due to excessive energy input during operation
Simultaneous loading by means of high drive power (gas throughput, fore-vacuum pressure), high
heat radiation, or strong magnetic fields results in uncontrolled heating of the rotor and can destroy
the vacuum pump.
►
Consult Pfeiffer Vacuum before combining varying loads on the vacuum pump. Lower limit values apply.
NOTICE
NOTICE
Turbopump destruction due to gases with too high molecular masses
The pumping of gases with impermissible high molecular masses leads to the destruction of the turbopump.
►
Make sure that the gas mode is set correctly by [P:027] in the electronic drive unit.
►
Consult Pfeiffer Vacuum before you use gases with higher molecular masses (> 80).
Parameter NameDesignationAdjustment, setting
[P:027]GasModeGas mode0 = heavy gases
[P:035]CfgAccA1Accessory connection A10 = fan (continuous opera-
[P:720]VentSpdVenting at rotation speed, delayed venting50 %
[P:721]VentTimeVenting time, delayed venting3600 sec.
36/60
Tbl. 11:Factory setting of the electronic drive unit when delivered
Notes for turbopump commissioning
1.
Observe the cooling water flow and flow rate.
2.
When using sealing gas, observe the sealing gas flow and throughput.
3.
Observe the reduction of the drive power depending on the mains connection voltage provided.
4.
Provide the current supply for the product.
6.2 Operating modes
The turbopump can be operated in different modes.
●
Operation without control unit
●
Operation via connection "E74"
●
Operation via "remote" connection
●
Operation via interface RS-485 and Pfeiffer Vacuum display and control unit or PC
●
Operation via field bus
6.2.1 Operation without operating unit
Automatic start
When using the supplied mating plug on the electronic drive unit or bridging contacts according to the terminal lay-out, the turbopump is ready for operation. Once
the supply voltage is provided, the turbopump starts up immediately.
Instructions for operation without control panel
1.
Use only the approved Pfeiffer Vacuum mating plug with bridges on the connection of the electronic drive unit.
2.
Switch on the mains supply of the turbopump only immediately before operation.
After applying the operating voltage, the electronic drive unit carries out a self-test to check the supply
voltage. After completing the self-test successfully, the turbopump starts and activates connected additional equipment according to the configuration.
Operation
6.2.2 Operation via multi-function connection "remote"
Remote control is available via the 26-pole D-sub connector with the “remote” designation on the electronic drive unit. The accessible individual functions are mapped to "PLC levels".
Instructions for remote control operation
►
See the operating manual of the electronic drive unit for the standard version.
6.2.3 Operation via connection "E74"
Operation is possible via the 15-pole D-sub connection with the “E74” designation on the electronic
drive unit. Besides the signals defined in the Directive SEMI E74-0301, the connection is provided with
an inverted alarm signal and an analog output.
Instructions for operation with E74
►
See the operating manual of the electronic drive unit with E74 version.
6.2.4 Operation via Pfeiffer Vacuum display and control unit
The connection of a Pfeiffer Vacuum display and control panel permits the controlling of the turbopump
via the parameters fixed in the electronic drive unit.
Instructions for operation with display and control unit
1.
When handling the Pfeiffer Vacuum display and control unit, observe the associated operating instructions:
–
"DCU" operating instructions available from the Pfeiffer Vacuum Download Center.
–
"HPU" operating instructions available from the Pfeiffer Vacuum Download Center.
2.
Observe the electronic drive unit operating instructions from the turbopump scope of delivery.
3.
Connect the display and control unit to the connection "RS-485" on the electronic drive unit.
6.2.5 Operation via field bus
Integrating and operating Pfeiffer Vacuum turbopumps in the customer's field bus system is possible
when using an electronic drive unit with a corresponding connection panel.
37/60
Operation
The following are available:
●
Profibus
●
EtherCAT
●
DeviceNet
Instructions for field bus operation
►
See the operating manual of the electronic drive unit with corresponding connection panel.
6.3 Switching on the turbopump
WARNING
Danger of cut injuries from unexpected start up.
The use of mating plugs of the electronic drive unit (accessories) enables the automatic run-up of the
vacuum pump as soon the power is turned on. Attaching mating plugs before or during the installation leads to the movement of parts hence the risk of cut injuries by sharp-edged in the exposed high
vacuum flange.
►
Only connect mating plugs after the mechanical installation.
►
Only switch on the vacuum pump immediately prior to operation.
WARNING
Risk of burns on hot surfaces when using additional equipment for heating during operation
The use of additional equipment for heating the vacuum pump or for optimizing the process generates very high temperatures on surfaces that can be touched. There is a risk of burning.
►
If necessary, set up a contact guard.
►
If necessary, apply the warning stickers provided for this at the danger points.
►
Ensure adequate cooling down before working on the vacuum pump or in its vicinity.
►
Wear protective equipment, e.g., gloves.
WARNING
Risk of serious injury in the event of vacuum pump destruction due to over pressure
Gas entry with very high over pressure results in destruction of the vacuum pump. There is a risk of
serious injury due to ejected objects.
►
Never exceed the permissible 1500 hPa (absolute) inlet pressure on the suction side or the venting and sealing gas connection.
►
Make sure that high, process-related over pressures cannot directly enter the vacuum pump.
Switching on the turbopump
►
Connect the power supply pack to the mains power supply on the customer-side.
►
Switch on the power supply pack.
6.4 Operation monitoring
6.4.1 Operating mode display via LED
LEDs on the electronic drive unit indicate basic operating conditions of the turbopump. A differentiated
malfunction and warning display is possible only for operation with DCU or HPU.
38/60
LEDSymbol LED statusDisplayMeaning
Offwithout current
On, flashing"Pumping station OFF", rotation speed ≤ 60
Green
YellowOffno warning
RedOffno error, no warning
Tbl. 12:Behavior and meaning of the LEDS of the electronic drive unit
On, inverse flashing"Pumping station ON", set rotation speed not
On, constant"Pumping station ON", set rotation speed
On, flashing
On, constantWarning
On, constantMalfunction, error
-1
rpm
reached
reached
"Pumping station OFF", speed > 60 rpm
Operation
-1
6.4.2 Temperature monitoring
If threshold values are exceeded, output signals from temperature sensors allow the turbopump to be
switched to a safe condition. Depending on the type, temperature thresholds for warning and malfunction messages are stored immutably in the electronic drive unit. For information purposes, various status requests are set up in the parameter set.
●
If the excess temperature warning threshold is exceeded, the electronics reduce the turbopump
drive power with the aim of avoiding shutdown.
─
Examples are an impermissible motor temperature, or impermissibly high housing temperature.
●
Further reduction of drive power and thus decreasing speed can potentially lead to the set rotation
speed switchpoint not being met, and thus to the turbopump switching off.
●
If the error threshold for excess temperature is exceeded, the turbopump switches off immediately.
6.5 Switching off and venting
We recommend
Vent the turbopump after shutdown. By doing so, you prevent particles flowing back
into the vacuum system from the fore-vacuum area.
6.5.1 Switching off
Notes for switching off the turbopump
1.
Shut down the turbopump via the control unit or remote control.
2.
Close the fore-vacuum line.
3.
Switch off the backing pump, if necessary.
4.
Vent the turbopump (options see below).
5.
Close the supply lines (e.g. for cooling water or sealing gas).
39/60
Operation
6.5.2 Venting
NOTICE
Damage to the turbopump due to non-permissibly fast pressure rise during venting
Non-permissibly high pressure rise rates place a significant load on the rotor and the magnetic bearing of the turbopump. During venting very small volumes in the vacuum chamber or the turbopump,
there is a risk of uncontrollable pressure rises. This causes mechanical damage to the turbopump,
including potential failure.
►
Observe the prescribed maximum pressure rise speed of 15 hPa/s.
►
Avoid manual and uncontrolled venting of very low volumes.
►
Where necessary, use a venting valve from the Pfeiffer Vacuum range of accessories.
Manual venting
Manual venting describes the standard process for venting the turbo pumping station.
1.
Ensure that the vacuum system is shut down.
2.
Open the black venting screw on the turbopump by one revolution maximum.
3.
Wait for pressure equalization to atmospheric pressure in the vacuum system.
4.
Close the venting screw again.
Use a Pfeiffer Vacuum venting valve
The Pfeiffer Vacuum venting valve is an optional accessory for installation on the turbopump.
The venting valve is normally closed. Control is via the turbopump electronic drive unit, and configuration of parameters [P:012] and [P:030]. In the event of a power failure, the turbopump continues to deliver sufficient energy during its run-down period to initiate a proper venting process. When power is restored, the venting process is interrupted.
►
Switch off the turbopump.
–
The venting process starts automatically.
Venting speed [P:720] Venting duration [P:721] Venting duration in the event of a power
failure
50 % of rated speed3600 s3600 s
Tbl. 13:Factory settings for delayed venting of turbopumps
General information for fast venting
We recommend fast venting of larger volumes in 4 steps.
1.
Use a Pfeiffer Vacuum venting valve for the turbopump, or match the valve cross-section to the
size of the recipient and maximum venting rate.
2.
Vent the vacuum system with a maximum rate of pressure rise of 15 hPa/s for a duration of 20
seconds.
3.
Then vent the system with a second venting valve of any size, for example directly at the vacuum
chamber.
4.
Wait for pressure equalization to atmospheric pressure in the vacuum system.
40/60
7 Maintenance
7.1 General maintenance information
WARNING
Danger to life from electric shock during maintenance and service work
The device is only completely de-energized when the mains plug has been disconnected and the turbopump is at a standstill. There is a danger to life from electric shock when making contact with live
components.
►
Before performing all work, switch off the main switch.
►
Wait until the turbopump comes to a standstill (rotation speed =0).
►
Remove the mains plug from the device.
►
Secure the device against unintentional restarting.
WARNING
Health hazard through poisoning from toxic contaminated components or devices
Toxic process media result in contamination of devices or parts of them. During maintenance work,
there is a risk to health from contact with these poisonous substances. Illegal disposal of toxic substances causes environmental damage.
►
Take suitable safety precautions and prevent health hazards or environmental pollution by toxic
process media.
►
Decontaminate affected parts before carrying out maintenance work.
►
Wear protective equipment.
Maintenance
7.2 Maintenance intervals and responsibilities
Recommendations for performing maintenance measures
1.
Clean the turbopump exterior with a lint-free cloth and a little isopropanol.
2.
Replace the electronic drive unit as an independent unit.
3.
Replace the operating fluid reservoir as an independent unit.
4.
Pay attention to when the operating fluid must be changed.
5.
Change the operating fluid reservoir at least every 4 years.
6.
Have Pfeiffer Vacuum Service replace the rotor bearing of the turbopump at least every 4 years.
7.
Consult with Pfeiffer Vacuum Service about shorter maintenance intervals for extreme loads or impure processes.
8.
For all other cleaning, maintenance or repair work, contact the appropriate Pfeiffer Vacuum Service location.
7.3 Replacing the operating fluid reservoir
WARNING
Risk of poisoning from contact with harmful substances
The operating fluid reservoir and parts of the turbopump may contain toxic substances from pumped
media.
►
Decontaminate affected parts before carrying out maintenance work.
►
Prevent health hazards or environmental impacts with adequate safety precautions.
►
Observe the operating fluid safety data sheet.
►
Dispose of the operating fluid reservoir according to applicable regulations.
41/60
1
2
3
4
5
6
Maintenance
NOTICE
Damage to sealing surfaces from unsuitable tools
The use of unsuitable tools for removal or insertion of sealing rings damages the sealing surfaces,
causing vacuum pump leakage.
►
Never use sharp, metallic tools (e.g., tweezers).
►
Remove sealing rings by hand only, or with wooden or plastic sticks if necessary.
You can find the safety data sheet can in the Pfeiffer Vacuum Download Center.
Preparatory work
1.
Do not apply any mechanical loads to the electronic drive unit.
2.
Switch off the turbopump (see chapter “Switching off”, page 39).
3.
Vent the vacuum system to atmospheric pressure (see chapter “Venting”, page 40).
4.
Interrupt the electric supply.
5.
Remove all cables from the electronic drive unit.
6.
If dismantling the turbopump from the system: Close all openings with the original protective
covers and screw plugs.
Place the turbopump on the closed high vacuum flange.
2.
Unscrew all Allen head screws from the screw cap on the pump base.
3.
Remove the screw cap.
4.
Remove the O-ring from the groove using a toothpick.
–
5.
Remove the operating fluid reservoir from the bearing housing using the tweezers.
6.
Pull the old Poroplast rods out of the pump base using tweezers.
7.
Clean the screw cap with a clean, lint-free cloth.
Avoid damage caused by scratches.
–
Do not use cleaning agents.
Installing the operating fluid reservoir
1.
Insert all new Poroplast rods using tweezers.
2.
Insert the operating fluid reservoir into the bearing housing with the felt side in the direction of the
injection tip.
–
When doing so, do not exert pressure on the operating fluid reservoir.
3.
Insert the O-ring in the groove in the pump base.
4.
Fit the screw cap.
5.
Tighten all 3 Allen head screws uniformly.
–
Tightening torque: 2.5 Nm.
7.4 Replacing the electronic drive unit
The electronic drive unit of the turbopump cannot be repaired. In the event of a defect, replace the entire electronic drive unit with a replacement part.
Damage to the turbopump and electronic drive unit due to improper disconnection of components
Even after the mains power is switched off, the turbopump continues to deliver electrical energy during its run-down period. If the turbopump and electronic drive unit are disconnected prematurely,
there is the risk of body contact and consequently the destruction of electronic components.
►
Never disconnect the turbopump and electronic drive unit from each other if power is still connected or if the rotor is running.
►
Monitor the turbopump rotation speed via the parameters available in the electronic drive unit
(e.g., [P:398]).
►
Wait until the turbopump comes to a standstill (rotation speed f=0).
Maintenance
NOTICE
NOTICE
Property damage from electrostatic discharge
Neglecting the electrostatic hazard for electronic components results in their damage or destruction
►
Implement ESD safety measures at the workstation.
►
Observe EN 61340 "Protection of electronic devices from electrostatic phenomena".
Backing up settings made by the customer
The factory operating parameters are always preset in replacement units. All settings made by the customer to the original electronic drive unit are lost when it is
replaced. To preserve your custom settings, you have the following options:
1.
Back up all your settings as a parameter set in an HPU.
2.
Load a backup parameter set by means of HPU into the new electronic drive
unit.
3.
Program the individual settings into the new electronic drive unit by hand.
4.
Note the operating instructions of the electronic drive unit and the HPU.
Preparatory work
1.
Do not apply any mechanical loads to the electronic drive unit.
2.
Switch off the turbopump (see chapter “Switching off”, page 39).
3.
Vent the vacuum system to atmospheric pressure (see chapter “Venting”, page 40).
4.
Interrupt the electric supply.
5.
Remove all cables from the electronic drive unit.
6.
If dismantling the turbopump from the system: Close all openings with the original protective
covers and screw plugs.
43/60
1
2
3
4
Maintenance
Fig. 23:Installing and removing the electronic drive unit TC 400
1 Allen head screw3 Adapter plate
2 Electronic drive unit4 Original protective cover
Unscrew all 3 Allen head screws from the electronic drive unit.
3.
Remove the old electronic drive unit straight out from the pump.
4.
Place a new electronic drive unit straight onto the adapter plate connection of the turbopump.
5.
Screw the electronic drive unit to the turbopump with all 3 Allen head screws.
–
Tightening torque: 2.5 Nm
7.5 Confirming the rotation speed specification
The typical nominal rotation speed of a turbopump is preset at the factory in the electronic drive unit. If
the electronic drive unit is replaced or a different pump type is used, the reference set value of the nominal rotation speed is cleared. The manual confirmation of the nominal rotation speed is part of a redundant safety system as a measure for preventing excess rotation speed.
HiPaceNominal rotation speed
10 | 30 | 60 | 801500 Hz
3001000 Hz
4501100 Hz
400 | 700 | 800820 Hz
Tbl. 14:Characteristic nominal rotation speeds of the turbopumps
Required aids
●
A connected Pfeiffer Vacuum display and control unit.
Adjusting the nominal rotation speed confirmation
The redundant confirmation of the nominal rotation speed of a turbopump is possible by adjusting the
[P:777] NomSpdConf parameter in the electronic drive unit.
●
Knowledge of the configuration and setting of electronic drive unit operating parameters.
1.
Observe the display and control unit operating instructions.
2.
See the electronic drive unit operating instructions.
3.
Set the parameter [P:794] to "1" and activate the expanded parameter set.
4.
Open and edit the parameter [P:777].
44/60
Maintenance
5.
Set the parameter [P:777] to the required value of the nominal rotation speed in Hertz.
Alternative: A Pfeiffer Vacuum SpeedConfigurator for the one-time immediate setting of parameter [P:
777] is included with the replacement units.
45/60
Decommissioning
8 Decommissioning
8.1 Shutting down for longer periods
WARNING
Health hazard through poisoning from toxic contaminated components or devices
Toxic process media result in contamination of devices or parts of them. During maintenance work,
there is a risk to health from contact with these poisonous substances. Illegal disposal of toxic substances causes environmental damage.
►
Take suitable safety precautions and prevent health hazards or environmental pollution by toxic
process media.
►
Decontaminate affected parts before carrying out maintenance work.
►
Wear protective equipment.
Procedure for a longer downtime of the turbopump (> 1 year)
1.
Remove the turbopump from the vacuum system if necessary.
2.
Change the operating fluid reservoir of the turbopump if necessary.
3.
Close the high vacuum flange of the turbopump.
4.
Evacuate the turbopump via the fore-vacuum connection.
5.
Vent the turbopump via the venting connection with dry, oil-free air or inert gas.
6.
Close all flange openings with the original protective caps.
7.
Store the turbopump upright with the high vacuum flange upwards.
8.
Store the turbopump indoors only, within the specified temperature range.
9.
In rooms with humid or aggressive atmospheres: Hermetically seal the turbopump together with a
drying agent in a plastic bag.
8.2 Recommissioning
Risk of damage to the turbopump as a result of operating fluid aging after recommissioning
The shelf life of the operating fluid of the turbopump is limited. Aging of the operating fluid may lead
to the failure of the ball bearing and cause damage to the turbopump.
►
Pay attention to when the operating fluid must be changed:
●
after maximum 2 years without operation,
●
after maximum 4 years combined operation and downtimes.
►
Observe the maintenance instructions and inform Pfeiffer Vacuum Service.
Procedures for recommissioning the turbopump
1.
Check the turbopump for contamination and moisture.
2.
Clean the turbopump externally with a lint-free cloth and a little isopropanol.
3.
If necessary, arrange for Pfeiffer Vacuum Service to completely clean the turbopump.
4.
Observe the total running time of the turbopump and if necessary, arrange for Pfeiffer Vacuum Service to replace the bearing.
5.
Change the operating fluid reservoir of the turbopump.
6.
Install the turbopump according to these operating instructions (see chapter “Installa-
tion”, page 20).
7.
Recommission the turbopump according to these instructions (see chapter “Commission-
ing”, page 36).
NOTICE
46/60
8.3 Disposing of the vacuum pump
WARNING
Health hazard through poisoning from toxic contaminated components or devices
Toxic process media result in contamination of devices or parts of them. During maintenance work,
there is a risk to health from contact with these poisonous substances. Illegal disposal of toxic substances causes environmental damage.
►
Take suitable safety precautions and prevent health hazards or environmental pollution by toxic
process media.
►
Decontaminate affected parts before carrying out maintenance work.
►
Wear protective equipment.
Procedure
►
Dispose of all substances safely according to local regulations.
Decommissioning
47/60
Malfunctions
9 Malfunctions
Should malfunctions occur, you can find information about potential causes and how to fix them here.
The operating manual of the associated electronic drive unit contains more detailed error descriptions.
ProblemPossible causesRemedy
Pump will not start up; none of
the built-in LEDs on the electronic drive unit illuminate
Pump will not start up; green
LED on the electronic drive unit
is flashing
Pump does not attain the nominal rotation speed within the set
run-up time
●
Current supply interrupted
●
Incorrect operating voltage
●
No operating voltage
present
●
Electronic drive unit defective
●
For operation without
control unit: Pins 1-3
and 1-14 on the “remote” connection are
not connected
●
For operation via
RS-485: The bridge be-
tween pins 1 and 14 inhibits control commands
●
For operation via
RS-485: Parameters not
set in the electronic
drive unit
●
Voltage drop in the cable is too high
●
fore-vacuum pressure
too high
●
Pump leakage
●
Gas throughput too high
●
Rotor not running
smoothly, defective
bearing
●
Run-up time setpoint adjusted too low
Thermal load due to:
●
lack of ventilation
●
water flow too low
●
fore-vacuum pressure
too high
●
ambient temperature too
high
1.
2.
3.
1.
2.
1.
2.
1.
2.
1.
2.
1.
2.
1.
1.
2.
1.
2.
1.
2.
3.
1.
1.
2.
1.
1.
Check the plug contacts on the power supply pack.
Check the current supply lines.
Check the output voltage (24 or 48 V DC) at the
power supply pack “DC out” connection.
Observe the electronic drive unit rating plate.
Supply the correct operating voltage.
Supply the correct operating voltage.
Switch on the power supply pack.
Replace the electronic drive unit.
Contact Pfeiffer Vacuum Service.
Connect the connections according to the electronic drive unit connection diagram.
Check the bridges on the connection cable.
Remove the bridge at the “remote” connection.
Check the connection cable.
Set the parameters [P: 010] and [P: 023] via the interface RS-485 to 1 = "ON".
Check the connection cable.
Use a suitable connection cable.
Check backup pump compatibility (see technical
data).
Check that the backup pump is working.
Carry out leak detection.
Check seals and flange connections.
Eliminate leaks.
Reduce the process gas load.
Check the pump for noise development
Contact Pfeiffer Vacuum Service.
Extend the run-up time setpoint [P:700] via a display and control unit.
Reduce the thermal load.
─
Ensure adequate air supply.
─
Adjust the cooling water flow.
─
Reduce the fore-vacuum pressure.
─
Adapt the ambient conditions.
48/60
Malfunctions
Pump not achieving the ultimate
pressure
Unusual noises during operation
Red LED on the electronic drive
unit illuminates
●
Pump is dirty
●
Vacuum chamber, pipes
or pump leaky
●
Rotor bearing damaged
●
Rotor damaged
●
Splinter shield or protective screen loose
●
Group error
1.
Heat the pump if required.
2.
Have it cleaned.
3.
Contact Pfeiffer Vacuum Service.
1.
Carry out leak detection starting from the vacuum
chamber.
2.
Check seals and flange connections.
3.
Eliminate leaks in the vacuum system.
1.
Contact Pfeiffer Vacuum Service.
1.
Contact Pfeiffer Vacuum Service.
1.
Check and correct the seat of the splinter shield or
protective screen in the high vacuum flange.
2.
Follow the installation instructions.
1.
Reset the malfunction by switching the current supply off and on.
2.
Reset the malfunction with V+ on pin 13 on the “remote” connection.
3.
Set the parameter [P: 009] via the interface
RS-485 to 1 = Malfunction acknowledgment.
4.
Set the parameter [P: Set parameter [P:010] via
the interface RS-485 to 0 = off and then 1 = On and
Malfunction acknowledgment.
5.
Carry out a differentiated malfunction analysis with
a display and control unit.
6.
Contact Pfeiffer Vacuum Service.
Tbl. 15:Troubleshooting turbopumps
49/60
ANFORDERUNG SERVICE
ERKLÄRUNG KONTAMINIERUNG
Service solutions from Pfeiffer Vacuum
10 Service solutions from Pfeiffer Vacuum
We offer first class service
Long vacuum component service life, coupled with low downtimes, are clear expectations that you have
of us. We satisfy your needs with capable products and outstanding service.
We are consistently striving to perfect our core competence, service for vacuum components. And our
service is far from over once you’ve purchased a product from Pfeiffer Vacuum. It often enough really
just begins then. In proven Pfeiffer Vacuum quality, of course.
Our professional sales engineers and service technicians stand ready to provide hands-on support to
you worldwide. Pfeiffer Vacuum offers a complete portfolio of service offerings, ranging from genuine
spare parts right through to service agreements.
Take advantage of Pfeiffer Vacuum Service
Whether for preventative on-site service from our field service, fast replacement with as-new replacement products or repair in a Service Center close to you; you have various options for upholding your
equipment availability. Detailed information and addresses can be found on our website in the Pfeiff-
er Vacuum Service section.
Advice on the optimum solution is available from your Pfeiffer Vacuum contact partner.
For quick and smooth handling of the service process, we recommend the following steps:
1. Download the current form templates.
─ Declaration of Service Request
─ Service Request
─ Declaration of Contamination
a. Dismantle all accessories and keep them (all external mounted parts as valve, inlet
screen, etc.).
b. Drain the operating fluid/lubricant as necessary.
c. Drain the cooling medium as necessary.
2. Fill out the service request and the declaration of contamination.
3. Send the forms via email, fax or post to your local Service Center.
4. You will receive a response from Pfeiffer Vacuum.
Sending of contaminated products
No units will be accepted if they are contaminated with micro-biological, explosive or radioactive substances. If products are contaminated or if the declaration of contamination is missing, Pfeiffer Vacuum
will contact the customer before starting maintenance. In addition, depending on the product and the
level of contamination additional decontamination costs may be required.
50/60
ERKLÄRUNG KONTAMINIERUNG
Service solutions from Pfeiffer Vacuum
5. Prepare the product for transport in accordance with the details in the declaration of contamination.
a) Neutralize the product with nitrogen or dry air.
b) Close all openings with airtight blank flanges.
c) Seal the product in appropriate protective film.
d) Only pack the product in suitable, stable transport containers.
e) Observe the applicable transport conditions.
6 Affix the declaration of contamination to the outside of the packaging.
7 Then send your product to your local Service Center.
8 You will receive a confirmation message/a quotation from Pfeiffer Vacuum.
For all service orders, our General Terms and Conditions of Sales and Supply and General Terms and
Conditions of Repair and Maintenance apply to vacuum equipment and components.
51/60
1
3
2
Spare parts HiPace 300 P
11 Spare parts HiPace 300 P
Fig. 24:Spare parts HiPace 300 P
Position DesignationSizeOrder numberNotePieces Quantity
1Electronic drive unit TC
400
2Operating fluid reservoirPM 203 410 -Tincl. Poroplast rods and O-
3Banjo fittingP 4131 007 DWater cooling2
Tbl. 16:Overview of the spare parts available for the HiPace 300 P
refer to the rating plate depending on the connec-
tion panel
ring
1
1
52/60
12 Accessories
Accessories
View the line of accessories for Pfeiffer Vacuum turbopumps online at pfeiffer-vac-
uum.com.
53/60
Technical data and dimensions
13 Technical data and dimensions
13.1 General
This section describes the basis for the technical data of Pfeiffer Vacuum turbopumps.
Technical data
Maximum values refer exclusively to the input as a single load.
●
Specifications according to PNEUROP committee PN5
●
ISO 27892 2010:"Vacuum technology - Turbomolecular pumps - Measurement of rapid shutdown
torque"
●
ISO 21360 2012: “Vacuum technology - Standard methods for measuring vacuum-pump performance - Part 1: General description”
●
ISO 21360 2018: "Vacuum technology - Standard methods for measuring vacuum-pump performance - Part 4: Turbomolecular vacuum pumps"
●
Ultimate pressure with test dome after 48 h bake out duration
●
Gas throughput with water cooling; backing pump = rotary vane pump (10 m3/h)
●
Cooling water consumption at maximum gas throughput, cooling water temperature 25 °C
●
Integral leakage rate with 100 % helium concentration, 10 s measurement duration
●
Sound pressure level at distance to vacuum pump = 1 m
mbarbarPahPakPaTorr | mm Hg
mbar1
bar10001
Pa0.01
hPa1
1 · 10
1 · 10
1 · 10
-3
-5
-3
10010.10.75
5
1 · 10
10.01
1000100750
-3
1 · 10
7.5 · 10
10010.10.75
kPa100.0110001017.5
Torr | mm Hg1.33
1.33 · 10
-3
1 Pa = 1 N/m
133.321.330.1331
2
Tbl. 17:Conversion table: Pressure units
mbar l/s
Pa m3/s
sccmTorr l/s
atm cm3/s
mbar l/s10.159.20.750.987
Pa m3/s
sccm
1015927.59.87
1.69 · 10
-2
1.69 · 10
-3
1
1.27 · 10
-2
1.67 · 10
Torr l/s1.330.13378.911.32
atm cm3/s
1.010.10159.80.761
Tbl. 18:Conversion table: Units for gas throughput
-3
-2
13.2 Technical data
ClassificationHiPace® 300 PHiPace® 300 PHiPace® 300 P
Cooling water temperature15 – 35 °C15 – 35 °C15 – 35 °C
Sound pressure level50 dB(A)50 dB(A)50 dB(A)
Venting connectionG 1/8"G 1/8"G 1/8"
Connection pressure (abs.) for
venting/sealing gas valve
Integral leak rate
Relative humidity of air5 – 85 %, not con-
Shipping and storage temperature-25 °C – 55 °C-25 °C – 55 °C-25 °C – 55 °C
Mounting orientationAnyAnyAny
BearingHybridHybridHybrid
Weight6.4 kg8.4 kg6.7 kg
0.09 hPa0.09 hPa0.09 hPa
137/60000 W/min
180/49980 W/min
201/60000 W/min
221/49980 W/min
201/60000 W/min
221/49980 W/min
-1
137/60000 W/min-1137/60000 W/min
-1
180/49980 W/min-1180/49980 W/min
-1
201/60000 W/min-1201/60000 W/min
-1
221/49980 W/min-1221/49980 W/min
-1
201/60000 W/min-1201/60000 W/min
-1
221/49980 W/min-1221/49980 W/min
2.4 W2.4 W2.4 W
1500 hPa1500 hPa1500 hPa
1 · 10-8 Pa m³/s1 · 10-8 Pa m³/s1 · 10-8 Pa m³/s
densing
5 – 85 %, not condensing
5 – 85 %, not condensing
-1
-1
-1
-1
-1
-1
Tbl. 20:Technical data for HiPace 300 P | 48 V
57/60
131
195
131
162
118
184
163
DN 100 ISO-K
Ø 110
45°
45°
M8 (5x)
8
3
2.6°
18.6
20°
Ø 144
130
DN 16 ISO-KF / G1/4"
Ø 8
90°
131
195
131
162
Ø 110
45°
45°
M8 (5x)
8
3
2
.
6
°
118
18.6
20°
Ø 144
130
184
163
DN 16 ISO-KF / G1/4"
DN 100 ISO-F
Ø 8
90°
118
131
Ø 110
45°
45°
M8 (5x)
8
3
2.6°
18.6
20°
Ø 144
130
DN 16 ISO-KF / G1/4"
Ø 8
90°
143
174
196
207
175
DN 100 CF-F
Technical data and dimensions
13.3 Dimensions
Dimensions in mm
Fig. 25:HiPace 300 | TC 400 | DN 100 ISO-K
58/60
Fig. 26:HiPace 300 | TC 400 | DN 100 ISO-F
Fig. 27:HiPace 300 | TC 400 |DN 100 CF-F
Declaration of conformity
We hereby declare that the product cited below satisfies all relevant provisions of the following EU Directives:
●
Machinery 2006/42/EC (Annex II, no. 1 A)
●
Electromagnetic compatibility 2014/30/EU
●
Restriction of the use of certain hazardous substances 2011/65/EU
The authorized representative for the compilation of technical documents is Mr. Helmut Bernhardt, Pfeiffer Vacuum GmbH, Berliner Straße 43, 35614 Aßlar, Germany.
Turbopump
HiPace 300 P
Harmonized standards and national standards and specifications which have been
applied:
DIN EN ISO 12100:2011
DIN EN 1012-2:2011
DIN EN 61000-3-2:2014
DIN EN 61000-3-3:2014
DIN EN 61010-1:2010
DIN EN 61326-1:2013
DIN EN 62061:2013
Signature:
(Dr. Ulrich von Hülsen)
Managing Director
Pfeiffer Vacuum GmbH
Berliner Straße 43
35614 Aßlar
Germany
Aßlar, 2018-11-09
ed. A - Date 1812 - P/N:PT0633BEN
*PT0338*
Loading...