Pfeiffer Vacuum DUO 6, DUO 6 M, DUO 1.6, DUO 1.6 M, DUO 3 Operating Instructions Manual

...
OPERATING INSTRUCTIONS
DUO 6 / DUO 6 M
Rotary Vane Pump
EN
Translation of the original instructions
PD 0062 BEN/F (1608)
Table of contents
1 About this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 Transport and storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4 Product description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
6 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
7 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
8 Decommissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
9 Malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
10 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
11 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
12 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
13 Technical data and dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
1.1 Validity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2 Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.1 Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Protective equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 Proper use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.4 Improper use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.1 Transport. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.2 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4.1 Product identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4.2 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.1 Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.2 Connecting the vacuum side. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.3 Connecting the exhaust side. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.4 Connecting to the mains power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.5 Filling up the operating fluid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.1 Before switching on the pump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
6.2 Switching on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
6.3 Pumping condensable vapours. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
6.4 Betriebsmittel nachfüllen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6.5 Switching off the pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
7.1 Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.2 Changing the operating fluid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.3 Changing the kind of operating fluid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
8.1 Shutting down for longer periods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
8.2 Re-starting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
8.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
9.1 Rectifying malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
11.1 Spare parts packages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
13.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
13.2 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
13.3 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
2

1 About this manual

1.1 Validity

This operating manual is for customers of Pfeiffer Vacuum. It describes the functioning of the designated product and provides the most important information for safe use of the unit. The description follows applicable EU guidelines. All information provided in this operating manual refers to the current state of the product's development. The documen­tation remains valid as long as the customer does not make any changes to the product.
Up-to-date operating instructions can also be downloaded from www.pfeiffer-vacuum.com.
Applicable docu­ments
Duo 6 / Duo 6 M Operating instructions
Declaration of Conformity Part of this document Operating instructions for accessories (order-specifically) see section "accessories"*
*also available via www.pfeiffer-vacuum.com
For information about other certifications, if applicable, please see the signet on the prod­uct or:
www.tuvdotcom.comTUVdotCOM-ID 0000021320
About this manual

1.2 Conventions

Safety instructions The safety instructions in Pfeiffer Vacuum operating instructions are the result of risk
evaluations and hazard analyses and are oriented on international certification stan­dards as specified by UL, CSA, ANSI Z-535, SEMI S1, ISO 3864 and DIN 4844. In this document, the following hazard levels and information are considered:
DANGER
Imminent danger
Indicates an imminent hazardous situation that will result in death or serious injury.
WARNING
Possibly imminent danger
Indicates an imminent hazardous situation that can result in death or serious injury.
CAUTION
Possibly imminent danger
Indicates an imminent hazardous situation that can result in minor or moderate injury.
NOTICE
Command or note
Command to perform an action or information about properties, the disregarding of which may result in damage to the product.
3
About this manual
Pictographs
Instructions in the
Work instruction: here you have to do something.
Prohibition of an action to avoid any risk of accidents, the disregarding of which may result in serious accidents
Warning of a displayed source of danger in connection with operation of the unit or equipment
Command to perform an action or task associated with a source of dan­ger, the disregarding of which may result in serious accidents
Important information about the product or this document
text
Abbreviations C version: Corrosive gas version
Symbols used The following symbols are used consistently throughout in all illustrations:
V
Vacuum flange
Exhaust flange
G
Gas ballast valve
Power connection
4

2Safety

2.1 Safety precautions

Duty to inform
Each person involved in the installation, operation or maintenance of the vacuum pump must read and observe the safety-related parts of these operating instructions.
The operator is obligated to make operating personnel aware of dangers originating
from the vacuum pump, the pumped medium and the entire system.
Installation and operation of accessories
Pfeiffer Vacuum pumps can be equipped with a series of adapted accessories. The in­stallation, operation and maintenance of connected devices are described in detail in the operating instructions of the individual components.
For information on order numbers of components, see "Accessories".Use original accessory parts only.
Do not expose any body parts to the vacuum.
Observe the safety and accident prevention regulations.
Check regularly that all safety precautions are being complied with.
Do not carry out any unauthorised modifications or conversions to the pumps.
Depending on the operating and ambient conditions, the surface temperature of the
pumps may rise above 70 °C. Use suitable finger guards if necessary.
When returning the pumps to us please note the instructions in the Service section.
Safety
The following safety instructions are only valid for the disassembly of the drive system for a vacuum pump with a magnetic coupling:
When disassembling the drive system from the pump housing, the strong magnetic field may influence the function and operational reliability of electrical and electronic devices.
Persons with cardiac pacemakers must keep away from the magnetic coupling. Danger to life!
Minimum distance: 2 m!
Disassembled magnetic couplings must be kept away from computers, data storage media and other electronic components.
Keep the disassembled components of the magnetic coupling separate at all times. Danger of crushing!
Do not allow any magnetised parts into the vicinity of the magnetic coupling. Danger of injury!

2.2 Protective equipment

Determined situations concerning the handling of vacuum pumps require wearing of per­sonal protective equipment. The owner, respectively the employer are obligated to pro­vide an adequate equipment to any operating persons.
Danger to health by hazardous substances during maintenance or installation
Depending on the process vacuum pumps, components or operating fluids can be con­taminated by toxic, reactive or radioactive substances.
Wear adequate protective equipment during maintenance and repairs or in case of
reinstallation.
DANGER
5
Safety

2.3 Proper use

CAUTION
Risk of injury through hot surfaces
Vacuum pumps can become hot during operation.
Allow the pump to cool before maintenance and repairs.If necessary wear protective gloves according to EN 420.
WARNING
Increased noise emission!
Increased noise emission can occur within a limited area surrounding the vacuum pump.
Provide noise protection orwear hearing protection.
NOTICE
EC conformity
The manufacturer's declaration of conformity becomes invalid if the operator modifies the original product or installs additional components.
Following installation into a plant and before commissioning, the operator must check
the entire system for compliance with the valid EU directives and reassess it accord­ingly.

2.4 Improper use

The vacuum pump may only be used to generate a vacuum.
Only use the vacuum pump for applications with oxygen concentration 21%.
Installation, operating and maintenance regulations must be complied with.
Other accessories, than those described in this manual, must not be used without the
agreement of Pfeiffer Vacuum.
Improper use will cause all claims for liability and warranties to be forfeited. Improper use is defined as usage for purposes deviating from those mentioned above, especially:
pumping of corrosive gases
pumping of explosive media
operation in potentially explosive areas
pumping of gases containing impurities such as particles, dusts and condensate; note
the vapour compatibility levels of the pump
pumping of substances that tend to sublime
use of the vacuum pump to generate pressure
pumping of liquids
the use of operating fluids not specified by Pfeiffer Vacuum
connection to pumps or units which are not suitable for this purpose according to their
operating instructions
connection to units which have exposed voltage-carrying parts
6

3 Transport and storage

3.1 Transport

Transport instructions
Remove the locking cap from the vacuum and exhaust flange immediately before con-
necting! – Check the cone strainer, paying attention to the O-ring.
Use only the handle on the top side of the pump to lift the pump.
Transport and storage

3.2 Storage

Fig. 1: Transporting the pump
Check that all the openings on the pump are securely closed.Fill up the pump with new operating fluid to the top edge of the sight glass.Store the pump only dry and dust-free indoors within the specified environmental con-
ditions. – In rooms with moist or aggressive atmospheres, the pump must be airproof shrink-
wrapped in a plastic bag together with a bag of desiccant.
– After storage periods longer than two years, it is recommended to carry out main-
tenance and change the operating fluid before using the pump.
7
Product description

4 Product description

4.1 Product identification

To correctly identify the product when communicating with Pfeiffer Vacuum, always have the information from the rating plate available.
Pump model and model number
Type and amount of operating fluid
Date of manufacture
Please find the voltage range and motor-related data on the separately attached motor rating plate.
D-35641 Asslar
Mod.: DUO 3 Mod.-Nr.: PK D57 XXX
Ser.- Nr.: 0014440
S : max. 2.5/2.9 m /h
(N )
2
Oil : P3 0.4 l Mass : 11,2 kg
n : XXXX 1/min
max.
Made in Germany 2005/01
3
Fig. 2: Product identification on the rating plate
Scope of delivery Pump with drive unit
Operating fluid (except F4 and F5)
Cone strainer and centering ring/centering ring with nozzle with O-rings
Locking cap for vacuum and exhaust flange
Operating instructions
8

4.2 Function

1
23
6
2
304/306
15
318 308/310 302 60 406
V
G
Installation
Vacuum pumps of the Duo series are oil-sealed, two-stage rotary vane vacuum pumps. The vacuum pumps are equipped with a safety valve which, when the pump is at a stand­still, closes the vacuum chamber vacuum tight and at the same time vents the pump.
Fig. 3: Duo 6 / Duo 6 M
1 Vacuum flange 2 Exhaust flange 6 Connector IEC/C14 23 Inlet port operating fluid
return line
60 Gas ballast valve 302 Casing 318 Sight glass 406 Motor 308 Operating fluid drain screw

5 Installation

5.1 Set-up

Installation location Observe the following requirements when setting up the pump:
Consider the load-bearing capacity of the installation site.
Maximum installation altitude 2000 m (above mean sea level)
Permissible ambient temperature: +12 ... 40 °C
Maximum relative humidity 85%
304 Operating fluid filler
screw 306 O-ring 310 O-ring
9
Installation
max. 10
°
> 35 mm
Fig. 4: Setting up the pump
Fill up with operating fluid before operating the first time (see p. 17, chap. 5.5).
– Amount and type according to rating plate
Always place the pump on a firm, even surface.Where stationary installation is involved, anchor the pump on site.When installing the pump in a closed housing, ensure there is sufficient air circulation.
– Sight glass and gas ballast valve must be visible and readily accessible. – Voltage and frequency information given on the motor rating plate must be visible.

5.2 Connecting the vacuum side

Remove locking cap from the vacuum flange;
– pay attention to the cone strainer and the respective O-ring in the intake port.
The connection between the pump and the vacuum chamber should be kept as short
as possible. – Depending on the pump type, use metallic hoses or PVC hoses with flange con-
nections.
– Separators, filters etc. may be installed upstream to protect the pump (see acces-
sories). However, please observe the loss of pumping capacity due to the conduc­tivity of the accessories.

5.3 Connecting the exhaust side

High pressure in the exhaust line!
Danger of damage to the seals and danger of the pump bursting.
Observe the maximum permissible pressure of 1500 hPa (absolute), activate shut-
off valves in such a way that they open before or at the same time as the pump is started.
Reduced pressure in the exhaust line!
Reduced pressure in the exhaust line can cause malfunctions and damage the pump. It is only allowed in pumps with magnetic coupling.
Ensure that when discharging gases the exhaust pressure is at least 250 hPa higher
than the suction pressure.
CAUTION
NOTICE
10
Remove the protective cap from the connection flange.Choose the cross-section of the exhaust line to be at least the size of the nominal
connection diameter of the vacuum pump's exhaust connection.
Piping to the pump must be suspended or supported.
Installation
X
S
ONF/OME
Z
9
8
10/11
29
28
– Physical forces from the piping system must not be allowed to act on vacuum
pumps.
Lay piping from the pump sloping downward so that no condensate can flow back into
the pump; otherwise fit a condensate separator. – If an air trap is created in the system, then a device for draining condensation water
must be provided at the lowest point.
WARNING
Emission of toxic substances from the exhaust!
Danger of poisoning from emitted gases or vapours, which can be detrimental to health and/or can pollute the environment, depending on the particular application.
Comply with the applicable regulations when working with toxic substances.Only officially approved filter systems may be used to separate and remove these
substances.
Fitting the OME and the oil return line (op­tion)
Turn off the vacuum pump, vent to atmospheric pressure and allow to cool.Remove the protective cap from the connection flange.Place OME on the exhaust side of the pump with flange pointing downwards and fit
with clamping ring (accessories), pay attention to centering ring.
Fig. 5: Duo 6 with operating fluid return line
8 Locking screw 9 Fitting 10 Operating fluid return hose
Note: As the hexagon socket of the locking screw is not metric, a special spanner (3/
11 Spring (inside the hose) 28 Fitting 29 Locking cap
S Clamping ring Z Centering ring
16") (provided as part of the scope of supply) is required.
Unscrew locking screw 8.Drain off operating fluid if so and fill in the pump.Screw in fitting 9 in place of the locking screw 8; take care with seal ring.Loosen fitting 28 and remove cap 29.Insert spring 11 into hose 10 (anti-kink device).Fit operating fluid return hose 10 at both sides,
– keep the hose as short as possible and protect it from bending.
11
Installation
Tighten the union nuts of both screw fittings.

5.4 Connecting to the mains power supply

Depending on the pump type, different motor versions or mains voltages are possible:
Single phase motor for fixed voltage with – built-in thermal protection switch, – mains switch and – mains connection socket (C14)
Single phase motor with switchable voltage range, – thermal protection switch, – mains switch and – mains connection socket (C14)
Three phase motor (without switch and mains cable).
Voltage-bearing elements
Danger to life from electric shock.
The electrical connection can be carried out only by trained and authorised electri-
cians.
Disconnect the power supply and secure it against being switched back on.Ensure the system is adequately earthed.
DANGER
NOTICE
Excess voltage!
Danger of destroying the motor.
Power connections must comply with local regulations. Voltage and frequency infor-
mation given on the motor rating plate must correspond to the mains voltage and fre­quency values.
To protect the motor and supply cable in case of malfunction, mains fuse protection
must be implemented.
WARNING
Danger of injury from moving parts!
After power failure or motor shutdown due to overheating, the motor may restart auto­matically.
Secure the motor so that it cannot be switched on while any work is being performed
on the pump.
If necessary, dismantle the pump from the installation for inspection.
Single phase motors The vacuum pumps are equipped with single-phase motors with built-in thermal protec-
tion switch. This interrupts the motor current in case of overheating, but provides no per­manent shutdown of the motor.
12
Installation
0
1
115
230
115
230
Fig. 6: Motor circuit diagram with switch
Fuse protection
The transmission power of the pump’s magnetic coupling is so great that the cou­pling is no overload protection for the motor.
To protect the motor in case of malfunction, additionally provide a fuse protection in
accordance with the regional regulations. – Select a fuse with slow characteristics according to the table below.
Motor voltage
[V], ± 10 %
100 ... 110 100 ... 120
100/200 100/200
115/230 115/230
230 ... 240 230 ... 240
208 ... 240 208 ... 240
Frequency
[Hz]
50 60
50 60
50 60
50 60
50 60
Nominal current
[A]
4.8
5.2
3.8/2.2
5.2/3
3.8/1.9
5.2/2.6
2
2.2
2.2
2.5
Recommended fuse, slow
[A]
16 16
16/6 16/6
16/6 16/6
6 6
6 6
Changing the voltage range
Only valid for pumps with reversible motor:
The mains voltage must be determined on-site each time before the pump is installed
or moved to a different location.
Disconnect the pump from the power supply.Set the desired voltage range on the voltage selector switch using a suitable screw-
driver.
Switch position: "115" "230" Voltage range
115 V 10%, 50/60 Hz 230 V 10%, 50/60 Hz Motor 115/230 V Voltage range
100 V 10%, 50/60 Hz 200 V 10%, 50/60 Hz Motor 100/200 V
13
Installation
W2
U1
L1
V2
V1
W1
U2
L3
L2
W2
U2
V2
U1
V1
W1
L1
L2
L3
W2
U1
L1
V2
V1
W1
U2
L3
L2
W2
U2
V2
U1
V1
W1
L1
L2
L3
NOTICE
Overvoltage!
An incorrect voltage range setting can damage the motor.
Always check the set voltage range before switching on the pump.Only change the voltage range when the pump is disconnected from the power
mains.
Three-phase motor The three-phase current motor circuit
The connections U1 - L2, V1 - L1 and W1 - L3 result in a clockwise rotation of the motor shaft as seen looking towards the motor fan.
Delta Connection
The three coils are connected in series with the connection point connected to the mains. The voltage of each coil is the same as the mains voltage whereas the mains current is the cube root of the coil current. Delta connections are denoted by the symbol . The voltage between the mains supply lines is called mains voltage. The mains current is the current which flows in the supply lines.
Fig. 7: Motor coil and connecting plate of Delta Connection (for low voltage)
Star Connection
The ends of the three coils are connected at the star center. The terminal voltage is the cube root of the coil voltage; the mains and the coil current are the same. Star connec­tions are denoted by the symbol Y.
Fig. 8: Motor coil and connecting plate of Star Connection (for high voltage)
NOTICE
Do not start with star/delta connection.
Always start motor directly.
Inspection of the direction of rotation
For pumps with three-phase motors, it is necessary to check the direction of rotation!
14
Motor protection
Installation
CAUTION
Operating fluids may leak out!
If the direction of rotation is incorrect, there is a danger that operating fluids may leak at the vacuum flange.
Always check the direction of rotation before filling in operating fluid.
Remove the locking cap from the exhaust flange (if existing).Switch the pump on briefly (from 2 to 3 sec.).
– The motor and motor fan must turn in a clockwise direction (see the arrow on the
support stand).
If the direction of rotation is incorrect: Swap two phase contacts at the connecting ca-
ble.
Fill up the operating fluid.
The transmission power of the pump’s magnetic coupling is so great that the cou­pling is no overload protection for the motor.
With PTC temperature sensors (3PTC)
Pump motors equipped with PTC temperature sensors (3PTC) in the stator windings can be connected to a PTC resistor tripping device for protection against overload. Other ap­proved motor temperature monitoring can be used also by the operator.
Tripping devices store the shutdown event and need to be manually switched back on again via the integrated RESET button or via the external RESET S3. Mains-ON is de­tected as an automatic RESET.
Set up the connections so that the directional rotation indicated on the pump is main-
tained, regardless of the representations in the current flow diagram.
L1 L2 L3
F1 - F3
K1
A1 A2 T1 T2 Y2Y1 24 21 22 14 1211
Us
3)
M
W
AC 220 ... 240 V
11
V
U
3
10
T1...T3
N
2)
2)
2)
S3
1)
1)
1)
F4
S1
k1
S2
K1
Fig. 9: Connection example for a three-phase AC motor with PTC resistor tripping device
US Control voltage S
OFF button
1
S
ON button
2
S
RESET button, external
3
K1 Contactor F1 ... F4 Fuses T1... T3 PTC resistor sensor H1 Tripping indicator M Motor, 3-phase
1)
H1
Only for devices with two
relay outputs
2)
Only for MSR type
3)
Only for order no.:
P 4768 051 FQ
With motor protection switch
Suitable are protection switches with slow triggering characteristics. The drive motor can have a power consumption that is higher than the rated current I 60034-1 it is permissible to exceed the rated current I
1.5 times for a period of 2 min-
N
. According to DIN EN
N
utes. The setting must permit the overload ability of the motor and can be found in the following table.
15
Installation
Voltage [V] Frequency [Hz] Motor rating [kW] IN [A] I
230 400 265 460
50 0.15 1.0 4.3 50 0.15 0.6 2.6 60 0.18 1.0 4.5 60 0.18 0.6 2.7
Motor control system Frequency converter (valid for three phase motors)
Operation of rotary vane pumps with variable rotation speeds is possible in the mains fre­quency range between 35 and 60 Hz. The start-up can use a ramp (run-up time: max. 30 s); the shutdown can occur directly.
max
[A]
16

5.5 Filling up the operating fluid

The type and amount of operating fluid should be visible on the pump's rating plate for every rotary vane pump.
Permissible operating fluid
P3 (standard operating fluid)
Operating fluid for special applications on request
Use approved operating fluids only!
The use of operating fluids that have not been approved by Pfeiffer Vacuum shall result in a limited warranty. In such cases, it is not possible to guarantee that product-specific performance data will be achieved.
Prior consultation is required before using other application-specific operating fluids.
Installation
NOTICE
Filling up the operat­ing fluid
Unscrew operating fluid filler screw 304.Fill up the operating fluid.
– First fill when the pump is cold: Maximum 3/4 of the min./max. range.
304
max.
min.
Fig. 10: Filling up the operating fluid
Screw in operating fluid filler screw 304.
WARNING
Toxic vapours!
Danger of poisoning when igniting and heating synthetic operating fluids (e.g. F4/F5) above 300 °C.
Observe the application instructions.Do not allow operating fluid to make contact with tobacco products; observe safety
precautions when handling chemicals.
17
Operation

6Operation

6.1 Before switching on the pump

Check the operating fluid level in the sight glass.Compare the voltage and frequency information on the rating plate with the mains volt-
age and frequency values.
Check that the exhaust connection allows free flow (max. permissible pressure 1500
hPa absolute). – Activate the shut-off valves in such a way that they open before or at the same time
as the pump is started.
Protect the pump sufficiently from taking in contaminants by means of suitable precau-
tions (e.g. dust filters); if necessary, check operating fluid regularly or replace at short­er intervals.

6.2 Switching on

The pump can be switched on in any pressure range between atmospheric and
ultimate pressure.
The ideal operating condition of the pump is achieved during continuous operation. Cy­clic operation is possible, but 10 cycles per hour should not be exceeded and the oper­ating phase should always be longer than the downtime (non-operation time).
No special precautions are necessary when pumping dry gases. In order to attain the lowest possible ultimate pressures, the gas ballast valve should be closed.
Hot surface!
Danger of burns if hot parts are touched. Depending on the operating and ambient con­ditions, the surface temperature of the pump may rise above 70 °C.
In this case, use suitable finger guards.
Switch on pump at main switch 15.Switch on the pump with the vacuum flange closed and allow to warm up for 30 min-
utes.
Check operating fluid level only when the pump is warm and running; therefore
– close vacuum flange and gas ballast valve, – correct filling level during operations: within the markings at the sight glass frame, – check operating fluid daily in non-stop operation, otherwise whenever the pump is
switched on. Refilling is possible when the pump is in final vacuum operation.

6.3 Pumping condensable vapours

Should the process gases contain condensable gases, the rotary vane pump must be operated with gas ballast (i.e. with an open gas ballast valve).
Bad final vacuum and damage to the pump!
Danger of condensation and corrosion due to exceeding the water vapour compatibility during operation without gas ballast or in case of insufficient supply of flushing gas.
Only pump vapours when the pump is warm and the gas ballast valve is open.When the process has been completed, allow the pump to continue running for
about 30 minutes with the vacuum flange closed and the gas ballast open for oper­ating fluid regeneration purposes.
CAUTION
NOTICE
18
Operation
7.3
7.1
7.2
7.4
Gas ballast valve, standard version
To avoid condensation in the pump when pumping condensable vapours, air is periodi­cally fed into the working chamber at the beginning of the compression phase via the gas ballast valve 42.
The gas ballast valve is closed when turning to the right to position 0 and open when turn­ing to the left to position 1. Intermediate settings are not possible.
The gas ballast valve 42 will only be contaminated when airborne dust is in the intake air.
1
0
1
0
geöffnet/ open
1
1
0
0
geschlossen/ closed
Fig. 11: Standard version 42 of gas ballast valve
Operation with acces­sories
Gas ballast valve, corrosive gas version
If the pumping process requires the connection of flushing gas, the C version of the gas ballast valve with the flushing gas connection must be used.
Fig. 12: Corrosive gas version of gas ballast valve
7.1 O-ring
7.2 O-ring
Connect flushing gas at the flushing gas connection 7.4.Set flushing gas pressure; maximum pressure 1500 hPa (absolute).
7.3 Proportioning screw
7.4 Flushing gas connection (for DN 6 mm hose)
19
Operation
– Select the type and amount of flushing gas depending on the process; consult Pfei-
ffer Vacuum if necessary.
Use the proportioning screw 7.3 to set the desired amount of gas.
– Closed when fully turned to the right; open when fully turned to the left.
Gas ballast valve with solenoid valve
To control the flow of the flushing gas externally, an electromagnetic valve can also be used as an alternative to the versions described above. The valve makes it easier to op­erate the gas ballast and allows clean air or other gas to be let in in a process-controlled manner.
NOTICE
Flushing gas pressure higher than allowed endangers the operational reliability of the pump.
The power input of the pump, the temperature and the ejection of operating fluid will in­crease.
Observe the maximum permissible flushing gas pressure of 1500 hPa (absolute).Set the amount of flushing gas on site; dosing is not possible when using a solenoid
valve!
Fig. 13: Assembling the solenoid valve at the gas ballast inlet
Performance data of the solenoid valve
2/2 way valve closed when disconnected Supply voltage 24 V DC, +/- 10 % Power input 4 W Socket Type 2506 Threaded connection of flushing gas 1/8" inside Flushing gas pressure max. 1500 hPa (absolute) Amount of flushing gas max. 180 l/h
20

6.4 Betriebsmittel nachfüllen

If the operating fluid has reached its minimum filling level, the operating fluid must be topped up. The fluid can be topped up during operation in the final vacuum.
Operation
Filling up the operat­ing fluid
Unscrew operating fluid filler screw 304.When the pump is at operation temperature, top up the operating fluid up to the "max."
marking.
304
max.
min.
Fig. 14: Filling up the operating fluid
Screw in operating fluid filler screw 304.
21
Operation

6.5 Switching off the pump

The pump can be switched off in any pressure range.
Rotary vane pumps have an integrated safety valve on the intake side. If the differential pressure between the exhaust side and the intake side is 250 hPa, then the valve clos­es automatically and vents the pump when the pump is switched off.
Switch the pump off at the mains switch or disconnect from the mains in a secure man-
ner.
Venting the vacuum chamber
Danger of backflow of operating fluid into the intake line!
Contamination of the connected vacuum system!
Vent the vacuum chamber within 30 s, regardless of the chamber size.For a longer venting process, use an additional shut-off valve and shut off the intake
line after switching off the pump.
NOTICE
Maintaining the vac­uum in the chamber
NOTICE
Danger of backflow of operating fluid into the intake line!
Contamination of the connected vacuum system!
Because the safety valve of the pump is not suitable for longer-term sealing, install
an additional shut-off valve in the intake line.
Shut off the intake line immediately after switching off the pump.
22

7 Maintenance

7.1 Precautions

Danger of injury from moving parts!
After power failure or motor shutdown due to overheating, the motor may restart auto­matically.
Secure the motor so that it cannot be switched on while any work is being performed
If necessary, dismantle the pump from the installation for inspection.
Pump parts may be contaminated from pumped media!
Danger of poisoning due to contact with harmful substances.
Decontaminate the pump before carrying out any maintenance work.In the event of contamination, take suitable safety precautions to prevent your health
Maintenance
WARNING
on the pump.
WARNING
from being harmed by any dangerous substances.
Turn off the vacuum pump, vent to atmospheric pressure and allow to cool.Disconnect the drive motor from the mains and secure it so that it cannot be switched
on.
Only dismantle the pump as far as necessary to carry out maintenance.Dispose of used operating fluid in compliance with local regulations.When using synthetic operating fluids or working with toxic substances or substances
contaminated with corrosive gases, the relevant instructions governing their use must be observed.
Use only alcohol or similar agents for cleaning pump parts.
Magnetic coupling The following safety instructions are only valid for the disassembly of the drive system
for vacuum pumps with magnetic coupling!
DANGER
Strong magnetic field in the vicinity of the drive system!
Danger to life for persons with cardiac pacemakers when the drive system is disassem­bled.
Persons with cardiac pacemakers must not enter the area (≤ 2m) of the magnetic
field.
Rooms in which open couplings are accessible must be identified: ’’ No trespassing
for persons with heart pacemaker’’!
Disassembled magnetic couplings must be kept away from computers, data storage
media and other electronic components.
23
Maintenance
Checklist for inspec­tion, maintenance and overhaul
Certain maintenance and overhaul work should only be performed by Pfeiffer Vacuum Service (PV). Pfeiffer Vacuum will be released from all warranty and liability claims if the required, below listed, intervals are exceeded or maintenance or overhaul procedures are not performed properly. This also applies if replacement parts other than Pfeiffer Vac­uum OEM replacement parts are used.
Activity K/I/
daily as required;
1
W/R
at least annually
Check operating fluid level K X Visual inspection (leak-tightness/oil leaks) K X Check filter insert of external oil mist filter (if
existent) Change operating fluid I X Disassemble casing, sight glass and pump-
ing system and clean casing outside (with­out cleaning agent)
Clean gas ballast valve and silencer I X Clean the motor fan cap I X Disassemble the pump, clean and exchange
seals and wearing parts:
Radial shaft seal ringCoupling half at motor sideHydraulic valveExhaust valve tongueValve flap of the gas ballast valveVane springs
Carry out work in accordance the mainte­nance and additionally:
Change vanesChange valves, springs and sight glassChange silencer nozzleCheck coupling and exchange if neces-
sary
Depending on the process, the required replacement intervals for the operating fluid and the intervals for inspection, maintenance and overhaul may be shorter than the guide values specified in the table. Consult Pfeiffer Vacuum Service, if necessary.
KX
IX
W X (PV)
R X (PV)
as required; at least every 2 years
1. K: Checking, I: Inspection, W: Maintenance, R: Overhaul
as required; at least every 4 years
24

7.2 Changing the operating fluid

The service life of the operating fluid is dependent on the application area for the pump. It must be changed if:
The specified ultimate pressure is no longer reached
The operating fluid in the sight glass is visibly contaminated, milky, or cloudy
The operating fluid is thermally aged, identifiable by its color ID value (applies to min-
eral oils only).
Depending on the applications, Pfeiffer Vacuum recommends determining the ex­act service life of the operating fluid during the first year of operation.
The replacement interval may vary from the guide value specified by Pfeiffer Vacuum depending on the thermal and chemical loads, and the accumulation of suspended par­ticles and condensation in the operating fluid.
Hot operating fluid!
Danger of burns when draining due to contact with skin.
Wear suitable protective clothing.Use a suitable collecting vessel.
Maintenance
WARNING
Draining the operat­ing fluid
WARNING
Operating fluid may contain toxic substances from the pumped media!
Danger of poisoning from the emission of harmful substances from the operating fluid.
Wear suitable protective clothing and respirators.Dispose of operating fluid according to the local regulations
Request safety data sheets for operating fluids and lubricants
from Pfeiffer Vacuum or download at www.pfeiffer-vacuum.com.
Dispose of operating fluid according to the local regulations.
Turn off the vacuum pump, vent to atmospheric pressure and allow to cool.Unscrew operating fluid filler screw 304.Unscrew operating fluid drain screw 308.Drain the operating fluid while still quite hot;
– to empty the pump fully, tip it forward slightly.
25
Maintenance
308/ 310
304/306
Determining the level of deterioration
Flushing and clean­ing
Fig. 15: Draining the operating fluid
Screw in operating fluid drain screw 308; pay attention to O-ring.Screw in operating fluid filler screw 304.Allow pump to run for a maximum of 5 seconds with the vacuum flange open.Drain off remaining operating fluid.
– In case of serious contamination, the operating fluid will have to be changed sev-
eral times (flushing):
Fill up with operating fluid and check the filling level (see p. 17, chap. 5.5).
The level of deterioration of operating fluid P3 can be determined for clean processes
with the colour scale (in accordance with DIN 51578); supplementary sheet PK 0219 BN on request or at www.pfeiffer-vacuum.com.
Suck off operating fluid from the pump through the operating fluid filler opening.Fill the specimen in a test tube or some similar vessel and test by holding against the
light.
Where discolouration is red brown (equivalent to 5 on the scale) change operating flu-
id at the latest.
If the interior of the pump is heavily contaminated with process residues, we recommend performing several changes of operating fluid to flush away the contamination:
Operate the pump with the gas ballast open until the pump has warmed up.Drain the operating fluid again and check for contamination, flush again if necessary.Take off the cap and clean the sight glass and pump system externally (without clean-
ing agent).
Replace the filter elements in the accessories.Screw the operating fluid drain screw back in.
Fill up with operating fluid and check the filling level (see p. 17, chap. 5.5).
Screw in operating fluid filler screw 304.
26

7.3 Changing the kind of operating fluid

When filling up, topping up or changing the operating fluid, always use the type of oper­ating fluid indicated on the pump type plate. If, for example, amended process conditions require the use of a different operating fluid, the fluid can be changed as follows:
Changing the type of operating fluid.
A change of operating fluid type can be only be made between mineral (P3) and syn­thetic operating fluid (D1). It is not possible to change from these two types to F4/F5 or the other way round!
For the two flushing processes and final fill, the pump needs to be filled up three times
with fresh operating fluid, and this is the amount of operating fluid required.
Perform two flushing processes with the new operating fluid.Clean any accessories present such as the ONF/OME or ORF/ODK and replace their
filter elements; pay attention to whether more operating fluid is required.
Fill the pump for the final time with the third filling.Note down the current type of operating fluid in an appropriate place on the pump
(preferably on the type plate).
Maintenance
NOTICE
27
Decommissioning

8 Decommissioning

8.1 Shutting down for longer periods

Before shutting down the pump, observe the following procedure and adequately protect the pump system against corrosion:
Switch off pump.Change operating fluid.Start the pump and allow the pump to warm up.Switch off the pump.Fill up the pump with new operating fluid to the top edge of the sight glass.Close vacuum flange and exhaust flange with locking caps.Store the pump only dry and dust-free indoors within the specified environmental con-
ditions. – In rooms with moist or aggressive atmospheres, the pump must be airproof shrink-
wrapped in a plastic bag together with a bag of desiccant.
– After storage periods longer than two years, it is recommended to carry out main-
tenance and change the operating fluid before using the pump.
Do not store pump in the vicinity of machines, lanes, etc., because strong vibrations
can damage the rotor bearings.

8.2 Re-starting

8.3 Disposal

Emission of operating fluid!
Danger of the operating fluid being emitted at the exhaust flange if overfilled.
Drain the operating fluid to the normal level before restarting the pump.
NOTICE
Re-starting
The serviceability of the operating fluid without operation is a maximum of 2 years. Be­fore restarting after a shut-down of 2 years or longer, carry out the following work.
Replace the operating fluid.Replace the radial shaft sealing rings and further elastomer parts.Replace bearings at pumps with anti-friction bearings.Follow the maintenance instructions and inform Pfeiffer Vacuum.
Products or parts thereof (mechanical and electrical components, operating fluids, etc.) may cause environmental burden.
Safely dispose of the materials according to the locally applicable regulations.
28

9 Malfunctions

Please note the following instructions should the pump malfunction:
Hot surface!
Danger of burns if hot parts are touched. The surface temperature of the pump may rise above 105 °C in case of malfunction.
Carry out work on the pump only after it has cooled to a safe temperature.
Motor overload!
Depending on the malfunction (e.g. blocking during cold start), the motor may not be sufficiently protected by the built-in thermal protection switch from damage through overheating.
Implement an additional network safety device.

9.1 Rectifying malfunctions

Problem Possible causes Remedy
Pump will not start up No mains voltage or voltage does
Pump switches off af­ter a while after being started
Pump does not attain ultimate pressure
Pumping speed of pump too low
Loss of operating fluid Swivel gasket leaky Check tightness; replace gasket if neces-
Malfunctions
CAUTION
NOTICE
Check mains voltage and mains fuse protec-
not correspond to the motor data Pump temperature too low Warm up pump to > 12 °C Thermal protection switch has re-
sponded Pump system dirty Clean pump; contact Pfeiffer Vacuum Ser-
Pump system damaged Clean and overhaul pump; contact Pfeiffer
Motor defective Replace motor Thermal protection switch of the
motor has responded Mains fuse protection triggered
due to overload (e.g. cold start) Exhaust pressure too high Check opening of exhaust line and exhaust
Measurement reading is false Check gauge, check ultimate pressure with-
Pump or connected accessories are dirty
Operating fluid dirty Operate pump for a longer period with gas
Leak in system Repair leak Operating fluid filling level too low Top off operating fluid Pump damaged Contact Pfeiffer Vacuum Service. Intake line not well-dimensioned Keep connections as short as possible and
Exhaust pressure too high Check opening of exhaust line and exhaust
Radial shaft seal ring leaky Replace seal ring and check bushing Operational loss of operating fluid,
without ONF/OME Operational loss of operating fluid,
with ONF/OME
tion; check motor switch
Detect and fix cause of overheating; allow pump to cool off if necessary.
vice if necessary.
Vacuum Service if necessary.
Detect and fix cause of overheating; allow motor to cool off if necessary.
Warm up pump
accessories
out installation connected. Clean pump and check components for con-
tamination.
ballast valve open or change operating fluid
ensure that cross-sections are sufficiently dimensioned
accessories
sary
Install oil mist filter and oil return unit
Clean oil mist filter and oil return unit, change filter respectively
29
Malfunctions
Problem Possible causes Remedy
Unusual operating noises
Silencer dirty Clean or replace the silencer. Damage to the pump system Clean and overhaul pump; contact Pfeiffer
Vacuum Service if necessary.
Motor bearing defective Replace motor; contact Pfeiffer Vacuum
Service if necessary
NOTICE
Service work should be carried out by a qualified person only!
Pfeiffer Vacuum is not liable for any damage to the pump resulting from work carried out improperly.
Take advantage of our service training programs; additional information at www.pfei-
ffer-vacuum.com.
Please state all the information on the pump rating plate when ordering spare parts.
30

10 Service

Service
Pfeiffer Vacuum offers first-class service!
Maintenance/repairs on site by Pfeiffer Vacuum field service
Maintenance/repairs in a nearby service center or service point
Fast replacement with exchange products in mint condition
Advice on the most cost-efficient and quickest solution
Detailed information and addresses at: www.pfeiffer-vacuum.com (Service).
Maintenance and repairs in Pfeiffer Vacuum ServiceCenter
The following steps are necessary to ensure a fast, smooth servicing process:
Download the forms "Service Request" and "Declaration on Contamination".Fill out the "Service Request" form and send it by fax or e-mail to your Pfeiffer Vacuum
service address.
Include the confirmation on the service request from Pfeiffer Vacuum with your ship-
ment.
Fill in the contamination declaration and enclose it in the shipment (required!).Dismantle all accessories.Drain operating fluid/lubricant.Drain cooling medium, if used.Send the pump or unit in its original packaging if possible.
1)
Sending of contaminated pumps or devices
No units will be accepted if they are contaminated with micro-biological, explosive or ra­dioactive substances. “Hazardous substances” are substances and compounds in ac­cordance with the hazardous goods directive (current version). If pumps are contaminat­ed or the declaration on contamination is missing, Pfeiffer Vacuum performs decontamination at the shipper's expense.
Neutralise the pump by flushing it with nitrogen or dry air.Close all openings airtight.Seal the pump or unit in suitable protective film.Return the pump/unit only in a suitable and sturdy transport container and send it in
while following applicable transport conditions.
Service orders
All service orders are carried out exclusively according to our repair conditions for vacu­um units and components.
1)
Forms under www.pfeiffer-vacuum.com
31
Spare parts

11 Spare parts

11.1 Spare parts packages

The spare parts packages listed here are only applicable for standard models.
Please state all information on the rating plate when ordering spare parts. Other spare parts than those described in this manual must not be used without the agreement of Pfeiffer Vacuum.
Set of radial shaft seal ring (RSSR)
RSSR and felt ring
Coupling, motor side
Set of coupling for pump with magnetic coupling
Can
Coupling half, drive side
Coupling half, pump side
Maintenance kit and set of seals
The kit contains all the critical wearing parts that should be replaced after disassembly and cleaning the pump.
Set of seals with all seals (O-rings) of the assembly groups and the subassemblies
Radial shaft seal ring, coupling and felt ring
Wearing parts pumping system
Wearing parts vacuum safety valve
Set of vanes
Vanes
Vane springs
Overhaul kit and set of seals
The kit contains all the wearing parts that should be replaced after disassembly and cleaning the pump:
Set of seals with all seals (O-rings) of the assembly groups and the subassemblies
Wearing parts of the pumping system (incl. vanes and springs)
Wearing parts of the valves
Coupling spider.
Spare parts package No.
Radial shaft ring kit incl. motor coupling PK E06 100 -T Coupling kit, M version PK E06 011 -T Maintenance kit+ set of seals, standard PK E01 042 CT Maintenance kit+ set of seals, M version PK E01 043 CT Set of vanes PK E08 031 -T Overhaul kit+set of seals, standard PK E02 042 CT Overhaul kit+set of seals, M version PK E02 043 CT
32

12 Accessories

Benennung Duo 6
KAS 16, condensate separator for pumping speeds from 1.6 to 12 m3/h PK Z10 003 ONF 16 S, oil mist filter for pumping speeds up to 12 m OME 16 M, oil mist filter for pumping speeds of up to 12 m ZFO 16, zeolite trap PK Z70 003 Operations monitoring unit 3 for Duo 1.6/3/6/11 and Duo 5/10/20 M PK 196 141 -T Operations monitoring unit 2 for Duo 1.6/3/6/11 and Duo 5/10/20 M PK 196 142 -T Operations monitoring unit 1 for Duo 1.6/3/6/11 and Duo 5/10/20 M PK 196 157 -T Mains cable 230 V with safety plug CEE 7, right angle IEC 320/C13 socket, 2 m PK 050 109 Mains cable 115 V with NEMA-plug, right angle IEC 320/C13 socket, 2 m PK 050 110 Mains cable 115 / 230 V without plug, right angle IEC 320/C13 socket, 3 m PK 050 111 SAS 16, dust separator, DN 16 ISO-KF, polyester filter PK Z60 506 Oil return unit from OME 16 S to Duo 1.6 / Duo 3 PK 005 986 -T Oil return unit from OME 16 M to Duo 1.6, 3, 6, 11 PK 006 080 -T Gas ballast valve - corrosive gas version PK 194 144 -U Gas ballast valve with magnet valve, 24 V DC PK 194 343 -U P3, mineral oil, 0.5 l PK 001 136 -T P3, mineral oil, 1 l PK 001 106 -T
3
/h PK Z40 001
3
/h PK Z40 003
Accessories

13 Technical data and dimensions

13.1 General

Conversion table: pressure units
mbar bar Pa hPa kPa Torr
mbar 1 1 · 10
bar 1000 1 1 · 10
Pa 0.01 1 · 10 hPa 1 1 · 10 kPa 10 0.01 1000 10 1 7.5 Tor r
mm Hg
1.33 1.33 · 10
Conversion table: gas throughput units
mbar·l/s Pa·m3/s sccm Torr·l/s atm·cm3/s
mbar·l/s 1 0.1 59.2 0.75 0.987
3
Pa·m
/s 10 1 592 7.5 9.87
sccm 1.69 · 10
Torr·l/s 1.33 0.133 78.9 1 1.32
3
atm·cm
/s 1.01 0.101 59.8 0.76 1
-2
-3
-5
-3
-3
1.69 · 10
1 Pa = 1 N/m
-3
mm Hg
100 1 0.1 0.75
5
1 0.01 1 · 10
100 1 0.1 0.75
133.32 1.33 0.133 1
1000 100 750
2
1 1.27 · 10
-2
-3
7.5 · 10
1.67 · 10
-3
-2
33
Technical data and dimensions

13.2 Technical data

Parameter Duo 6 Duo 6 M
Flange (in) DN 16 ISO-KF DN 16 ISO-KF Flange (out) DN 16 ISO-KF DN 16 ISO-KF Pumping speed at 50 Hz 5 m Pumping speed at 60 Hz 6 m Ultimate pressure with gas ballast 3 · 10 Ultimate pressure without gas ballast 3 · 10 Exhaust pressure, min. Atmospheric pressure 250 hPa Exhaust pressure, max. 1500 hPa 1500 hPa Rotation speed at 50 Hz 3000 min Rotation speed at 60 Hz 3600 min Leak rate safety valve 1 · 10-5Pa m3/s 1 · 10-5Pa m3/s Emission sound pressure level without gas ballast at 50 Hz 51 dB (A) 51 dB (A) Ambient temperature 12-40 °C 12-40 °C Protection category IP 40 IP 40 Rated power 50 Hz 0.2 kW 0.2 kW Rated power 60 Hz 0.24 kW 0.24 kW Switch Yes Yes Mains cable No No Shipping and storage temperature -25-+55 °C -25-+55 °C Operating fluid filling 0.5 l 0.5 l Weight 16 kg 17.5 kg Cooling method, standard Air Air
Typical ultimate pressure according to PNEUROP
3
/h 5 m3/h
3
/h 6 m3/h
-2
hPa 3 · 10-2hPa
-3
hPa 3 · 10-3hPa
-1
-1
3000 min 3600 min
-1
-1
34

13.3 Dimensions

Technical data and dimensions
+2
B
7,2()
3
C
+2
E
DN16 ISO-KF
31,5
+1
D
32,5
70
190
+2
A
Fig. 16: Duo 6
Di­men­sions
A 381 mm 381 mm 371 mm B 378 mm 391 mm 371 mm C 145 mm 145 mm 145 mm D 174 mm 182 mm 177 mm E 107 mm 107.5 mm 107.5 mm
Duo 6 Duo 6 Duo 6
Duo 6, 1-phase motor, 115/230 V, 50/60 Hz
Duo 6, 1-phase motor, 100-110 V, 50 Hz; 100-120 V, 60 Hz
Duo 6, 1-phase mo­tor, 230-240 V, 50/60 Hz
61
88 95
123
127
+1
4
±1
172
±1
35
Technical data and dimensions
(7,2)
70
3
DN 16 ISO-KF
D
+1
88
95
123
±1
127
±1
172
±1
B
+2
A
+2
E
+2
190
33,5
C
31,5
61
Fig. 17: Duo 6 M
Di­men­sions
A 423 mm 423 mm 413 mm B 420 mm 433 mm 413 mm C 145 mm 145 mm 145 mm D 176 mm 182 mm 177.5 mm E 108.5 mm 108.5 mm 108.5 mm
Duo 6 M Duo 6 M Duo 6 M
Duo 6 M, 1-phase mo­tor, 115/230 V, 50/60 Hz
Duo 6 M, 1-phase motor, 100-110 V, 50 Hz; 100-120 V, 60 Hz
Duo 6 M, 1-phase mo­tor, 230-240 V, 50/60 Hz
36

Declaration of conformity

We hereby declare that the product cited below satisfies all relevant provisions accord­ing to the following EC directives:
Machinery 2006/42/EC (Annex II, no. 1 A)
Electromagnetic Compatibility 2014/30/EU
Restriction of the use of certain Hazardous Substances 2011/65/EU
The agent responsible for compiling the technical documentation is Mr. Sebastian Ober­beck, Pfeiffer Vacuum GmbH, Berliner Straße 43, 35614 Aßlar.
DuoLine
Duo 6 / Duo 6 M
Harmonised standards and national standards and specifications which have been ap­plied:
DIN EN ISO 12100 : 2010 DIN EN 61010-1 : 2010 DIN EN 61000-6-3 : 2007 + A1: 2011
DIN EN 1012-2 : 2011-12 DIN EN 61000-6-1 : 2007 DIN EN 61000-6-4 : 2007 + A1: 2011
DIN EN ISO 13857 : 2008 DIN EN 61000-6-2 : 2006 DIN EN ISO 2151 : 2: 2008
ISO 21360-1, 2 : 2012
Signature:
Pfeiffer Vacuum GmbH Berliner Straße 43 35614 Asslar Germany
(Dr. Ulrich von Hülsen) Managing Director
2016-08-05
VACUUM SOLUTIONS FROM A SINGLE SOURCE
Pfeiffer Vacuum stands for innovative and custom vacuum solutions worldwide, technological perfection, competent advice and reliable service.
COMPLETE RANGE OF PRODUCTS
From a single component to complex systems: We are the only supplier of vacuum technology that provides a complete product portfolio.
COMPETENCE IN THEORY AND PRACTICE
Benefit from our know-how and our portfolio of training opportunities! We support you with your plant layout and provide first-class on-site service worldwide.
Are you looking for a perfect vacuum solution? Please contact us:
www.pfeiffer-vacuum.com
Pfeiffer Vacuum GmbH
Headquarters • Germany T +49 6441 802-0 info@pfeiffer-vacuum.de
Loading...