Pfeiffer Vacuum DPG 202 Operating Instructions Manual

OPERATING INSTRUCTIONS
DPG 202
Controller
EN
Translation of the original instructions
PG 0030 BEN/E (1605)
Table of contents
Table of contents
1 About this manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3 Transport and storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4 Product description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
6 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
7 Malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
9 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
10 Technical data and dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
1.1 Validity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2 Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.1 Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 Proper use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 Improper use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4.1 Product identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.2 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.3 Range of application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
5.1 Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
5.2 Connecting to the mains power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
5.3 Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.4 Configuring the data exchange . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.5 Pfeiffer Vacuum Protocol for "RS-485" . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
6.1 Switching on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6.2 Measuring mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.3 Menu overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.4 Configuration mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
7.1 Rectifying malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
10.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
10.2 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
10.3 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Declaration of conformity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
About this manual

1 About this manual

1.1 Validity

This operating manual is for customers of Pfeiffer Vacuum. It describes the functioning of the designated product and provides the most important information for safe use of the unit. The description follows applicable EU guidelines. All information provided in this operating manual refers to the current state of the product's development. The documen­tation remains valid as long as the customer does not make any changes to the product.
Up-to-date operating instructions can also be downloaded from www.pfeiffer-vacuum.com.

1.2 Conventions

Safety instructions The safety instructions in Pfeiffer Vacuum operating instructions are the result of risk
evaluations and hazard analyses and are oriented on international certification stan­dards as specified by UL, CSA, ANSI Z-535, SEMI S1, ISO 3864 and DIN 4844. In this document, the following hazard levels and information are considered:
DANGER
Pictographs
Imminent danger
Indicates an imminent hazardous situation that will result in death or serious injury.
WARNING
Possibly imminent danger
Indicates an imminent hazardous situation that can result in death or serious injury.
CAUTION
Possibly imminent danger
Indicates an imminent hazardous situation that can result in minor or moderate injury.
NOTICE
Command or note
Command to perform an action or information about properties, the disregarding of which may result in damage to the product.
Warning of a displayed source of danger in connection with operation of the unit or equipment
Command to perform an action or task associated with a source of dan­ger, the disregarding of which may result in serious accidents
Instructions in the text
Work instruction: here you have to do something.
Safety
Abbreviations CPT: Digital Piezo gauge
PPT: Digital Pirani gauge RPT: Digital Piezo/Pirani gauge HPT: Digital Pirani/Bayard-Alpert gauge MPT: Digital Pirani/Cold Cathode gauge DPG: Control unit CC sensor: Cold cathode sensor BA sensor: Bayard-Alpert sensor

2Safety

2.1 Safety precautions

Duty to inform
Each person involved in the installation or operation of the unit must read and observe the safety-related parts of these operating instuctions.
The operator is obligated to make operating personnel aware of dangers originating
from the unit or the entire system.
Installation and operation of accessories
Pfeiffer Vacuum pumps can be equipped with a series of adapted accessories. The in­stallation, operation and maintenance of connected devices are described in detail in the operating instructions of the individual components.
For information on order numbers of components, see "Accessories".Use original accessory parts only.
Observe the safety and accident prevention regulations.
Check regularly that all safety precautions are being complied with.
The unit has been accredited with protection class IP 20. Take necessary measures
when installing into ambient conditions, which afford other protection classes.
Do not modify or alter the unit yourself.
Ensure specified mains voltage.
Note the shipping instructions, when returning the unit.
Communicate the safety instructions to other users.

2.2 Proper use

2.3 Improper use

Transport and storage
NOTICE
EC conformity
The manufacturer's declaration of conformity becomes invalid if the operator modifies the original product or installs additional components.
Following installation into a plant and before commissioning, the operator must check
the entire system for compliance with the valid EU directives and reassess it accord­ingly.
The DPG 202 controler is used exclusively for connecting DigiLine gauges, e. g. CPT 100/200, RPT 100/200, PPT 100/200, HPT 100/200 and MPT 100/200.
Improper use will cause all claims for liability and warranties to be forfeited. Improper use is defined as usage for purposes deviating from those mentioned above, especially:
connection to mains supplies, which do not confirm to the regulations IEC 61010 or IEC 60950
connection to units which have exposed voltage-carrying parts
connection to pumps or units which are not suitable for this purpose according to their
operating instructions
operation of the devices in areas with ionizing radiation

3 Transport and storage

Units without external protection must not come into contact with electrostatically charge­able materials and must not be moved within electrical or magnetic fields.
In rooms with moist or aggressive atmospheres, the unit must be airproof shrink-
wrapped in a plastic bag together with a bag of desiccant.
Product description
set
DPG 202
V
A
UU
M
C
4
8
1
2
3
10
9
6
7
Sensor RS-485
SP1
SP2
USB
S1

4 Product description

The DPG 202 controller is used in conjunction with the Pfeiffer Vacuum DigiLine gauges for measurement and control total pressures. The unit is ready to connect the gauges CPT 100/200, RPT 100/200, PPT 100/200, HPT 100/200 and MPT 100/200.

4.1 Product identification

To correctly identify the product when communicating with Pfeiffer Vacuum, always have the information from the rating plate available.
D-35614 Asslar
Mod.:
DPG 202
Mod. No.:
PT G12 020
Input: Output:. Weight:
Fig. 1: Product identification on the rating plate
95 - 265 V, 50/60Hz, 25 VA
= 24 V, 1.0 A
0.75 kg
Scope of delivery The following positions are included in the delivery consignment:
– DPG 202 – Mating plugs for relay outputs – Operating instructions – Fixing screws for rack installation – USB cable
Ser. No.: 42000303

4.2 Function

Fig. 2: Frontpanel and rear view DPG 202
1 Display for presssure values
2 Menu key 3Set key
(mbar, bar, mTorr, Torr, hPa, Pa),
4 Up-/Down keys 6 Mains switch S1 7 Mains connection 8 Gauge input
9 USB port 10 Relay output (SP1, SP2)

4.3 Range of application

The unit DPG 202 must be installed and operated in the following ambient conditions:
Installation location weather protected (indoor) Protection class IP 20 Installation altitude max. 2000 m Ambient temperature +5°C to +50°C Relative humidity 5 ... 85 %, non-condensing Atmospheric pressure 860 hPa - 1060 hPa
Product description
Installation

5 Installation

5.1 Assembly

NOTICE
Ensure free convection
Incorrect installation can cause damage by over-heating or create a fire hazard. Maintain a minimum distance of 50 mm from cooling vents to adjacent components
or boundaries.
Install the device upright.
Rack installation The unit is fitted in a casing that is suitable for installation in 19"/3HE rack modules.
Insert the unit in the guiderails of a 19"/3HE rack and screw on the front panel secure-
ly with four fixing screws provided as part of the scope of supply.

5.2 Connecting to the mains power supply

The DPG 202 is equipped with an universal power supply with an input voltage range of
95 ... 265 V AC.
CAUTION
Excess voltage!
Danger of destroying the unit.
Power connections must comply with local regulations. Voltage information given on
the rating plate must correspond to the mains voltage values.
Use only cable with protective earth conductor.To protect the electronics and the power supply cable in case of failure provide an
appropriate network safety device.
WARNING
Danger of unsafe electrical installation
Safe operation after installation is the responsibility of the operator.
Do not independently modify or change the pump and electrical equipment.Make sure that the system is integrated in an emergency off safety circuit.Consult Pfeiffer Vacuum for special requirements.
WARNING
Danger of electric shock
In case of defect, the parts connected to the mains supply are under voltage.
Always keep the mains connection freely accessible so you can disconnect it at any
time.
Mains connection
S1
S
1
RS-485
adr
RS-485
3
5
13
15
9
11
7
device status
RS-485
adr
socket

5.3 Connections

Installation
If the unit is installed in a rack, the mains voltage must be supplied via a switched power distributor.
Order the mains cable separately (see "accessories").Switch off switch S1 on the power pack (position "0").Connect the mains cable to the mains.Always ensure a safe connection to the protective earthing conductor (PE, protection
class III).
NOTICE
Damage to the product
Only connect cables when de-energized. Never establish a connection using a live cable.
Setting the address selection switch
Communication is effected via Serial Interface RS-485 (address settable from 1...2).
Fig. 3: Setting the address selection switch
Remove the rubber plug (not shown in the illustration) from the address selector
switch, and set the required address according to the relevant connection situation.
Reinsert the rubber plug.
Installation
USB
RS-485
set
DPG 202
V
A
UU
M
C
12 12
12
34
5
Connecting the gauge to the DPG 202 controller
Set the address selector switch at the gauge to the required address (default setting:
"1").
Connect the gauge to the control unit using the connection cable.Switch on the control unit.
Fig. 4: Connections diagram gauge/DPG 202
Connect unit interface to the PC’s USB interface.
Note! Drivers and installation instructions can be downloaded from www.pfeiffer-Vacu-
um.com.
RS-485
M12 socket with screw coupling for the connection of a DigiLine gauge. The use of a Y­connector enables the connection of a second gauge.
Pin Assignment
1 RS-485: D+ 2+24 V 3GND 4 RS-485: D­5 not connected
Relay output The relay output provides two switching functions for external control.
NOTICE
Damage to the product
Only connect cables when de-energized.
Never establish a connection using a live cable.
Representation of the contacts in a state of rest, i.e.: Switch function = "off"
SP1
SP2
Pin Assignment
SP1 Relay output 1/max. 2 A SP2 Relay output 2/max. 2 A
Table 1: Phoenix Combicon, 6 pole (male)
10

5.4 Configuring the data exchange

9600 baud
8 data bits
1 stop bit
no parity

5.5 Pfeiffer Vacuum Protocol for "RS-485"

Installation
Telegram frame The telegram frame of the Pfeiffer Vacuum protocol contains only ASCII code characters
[32; 127], the exception being the end character of the message
C
. Basically, a master
R
(e.g. a PC) sends a telegram, which is answered by a slave (e.g. electronic drive unit or gauge).
a2 a1 a0 * 0 n2 n1 n0 l1 l0 dn ... d0 c2 c1 c0
a2 - a0 Unit address for slave
– Individual address of the unit ["001";"015"] *Action n2 - n0 Pfeiffer Vacuum parameter numbers l1 - l0 Data length dn ... d0 dn - d0 Data in data type concerned c2 - c0 Checksum (sum of ASCII values of cells a2 to d0) modulo 256 C
R
Carriage return (ASCII 13)
C
R
Telegrams Data request ?
a2 a1 a0 0 0 n2 n1 n0 0 2 = ? c2 c1 c0
Control command !
a2 a1 a0 1 0 n2 n1 n0 l1 l0 dn ... d0 c2 c1 c0
Data response / control command understood 
a2 a1 a0 1 0 n2 n1 n0 l1 l0 dn ... d0 c2 c1 c0
C
R
C
R
C
R
Error message 
a2 a1 a0 1 0 n2 n1 n0 0 6 N O _ D E F c2 c1 c0
_RANGE
_LOGI C
NO_DEF The parameter n2 - n0 does not exist _RANGE Data dn - d0 are outside the permitted range _LOGIC Logic access violation
C
R
11
Installation
 0011074006100023025
ASCII 48 48 49 49 48 55 52 48 48 54 49 48 48 48 50 51 48 50 53 13
Telegram examples
Read actual pressure value (data query) (Parameter [P:740], Slave device address: "001")
? 0010074002=?106
ASCII 48 48 49 48 48 55 52 48 48 50 61 63 49 48 54 13
C
C
Activate/send parameter for atmospheric pressure (high pressure) (control com­mand)
Atmospheric pressure adjustment (Parameter [P:741/740], Slave device address: "001")
? 0011074103001130
ASCII 48 48 49 49 48 55 52 49 48 51 48 48 49 49 51 48 13
 0011074103001130
ASCII 48 48 49 49 48 55 52 49 48 51 48 48 49 49 51 48 13
C
C
Activate/send parameter for low pressure (control command) Low pressure adjustment (Parameter [P:741/740], Slave device address: "001")
? 0011074103000129
ASCII 48 48 49 49 48 55 52 49 48 51 48 48 48 49 50 57 13
 0011074103000129
ASCII 48 48 49 49 48 55 52 49 48 51 48 48 48 49 50 57 13
C
C
R
R
R
R
R
R
Activating the key­board lock [P:008]
Reading the software version [P:312]
Reading the compo­nent names [P:349]
The keyboard lock for the set key can be activated/deactivated using the parameter
[P:008].
Send telegram according table " Telegram examples" to the DPG 202. The unit ad-
dress of the DPG 202 is factory set to 016 and cannot be changed.
Handling Telegram to DPG 202 Telegram from DPG 202 Unit adress
Keyboard lock "on"
Keyboard lock "off"
Table 2: Telegram examples
01610000806111111028
01610000806000000022
C
01610000806111111028
R
C
01610000806000000022
R
C
016
R
C
016
R
The software version can be read from the connected device using parameter 312:
Example: 010100
Parameter 349 contains a token of the component name:
DPG 202
12

6Operation

V010100
V010100
hPa
PRESSURE
REL: 1 2
DPG 202
Scan...
CH1: 1000 CH2: noSensor

6.1 Switching on

Maintain safe work conditions
The device described here is operated with mains voltage.
Before opening the device, cut off the voltage supply and secure it to prevent it from
After switching on the power supply, the connected gauge is ready for operation. It is ad­visable to wait for a stabilization period of 5-10 minutes before measurement. Operate the gauge during the measurement continuously independent of the applied pressure.
Switch on power supply using switch S1 on the mains pack.
being switched on unintentionally.
The DPG 202 carries out a self test and the following appears on the display: – software version number,
– the connected total pressure gauges are automatically iden­– the start menu PRESSURE with the actual pressure value
Operation
WARNING
tified,
appears in the display.
– The current gauge parameters are imported.
DPG 202 is ready – On channel 1, a gauge has been detected and a pressure of
1000 hPa is displayed.
– On channel 2, no gauge has been detected.
13
Operation
hPa
PRESSURE
REL:
1 2
CH1: noSensor CH2: noSensor
hPa
PRESSURE
REL:
1 2
CH1: ur/or CH2: ur/or
hPa
PRESSURE
REL:
1 2
CH1: Err1 CH2: Err1
hPa
PRESSURE
REL:
1 2
< 1.0·10
- 4
hPa
PRESSURE
REL: 1 2
CH1:
1005
CH2:
noSensor
More possible dis­plays
noSensor
The unit was not able to communicate with a gauge. Possible causes:
No gauge is connected.
The address selection switch of the gauge has not been set
on a RS-485 address 1 or 2.
The data line between the DPG 202 and gauge is not cor­rectly connected or is faulty.
The serial interface in the DPG 202 is defective.
ur/or
The measured pressure is outside the measurement range of the gauge.
ur = under range; pressure lower than lowest limit – or = overrange; pressure higher than highest limit
Err1
The connected gauge is defective.
Switching off
-4
< 1·10
This display indicates the filament of the BA sensor or the CC sensor is switched off in case of the HPT or MPT gauge, re­spectively.
The measured pressure with the Pirani sensor is currently
-4
<1.0·10 the Pirani sensors measuring range, the value is measured and displayed.
Switching off
The DPG 202 can be switched off in any operating mode. The keyboard lock is activated (indicated with a lock sign) until
the DPG 202 is switched off.
(ur of the Pirani sensor). If the pressure rises up to
14

6.2 Measuring mode

In measuring mode the DPG 202 shows the pressure of the connected gauge currently being measured. The display appears, depending on the measurement range, either in fixed-point or exponential notation. The switch outputs are controlled in accordance with the set point parameters and actual pressure.
If the communication is interrupted, noSensor appears on the display.
In such a case reinitialize the unit by a scan (COMMON MENU).
Operation
15
Operation
hPa
PRESSURE
REL: 1 2
CH1:
1005
CH2:
notr
ch:
1
CHANNEL MENU
type: RPT 200 V010100 gas: N2/O2/CO/Air corr: 1.00PI adjust
ch:
1
SENSOR MENU
type: RPT 200 V010100
SwMode: switch
rel: 1 2
ch:
1
RELAY MENU
chann: ch1 press: 10 hPa hyst: 2.0 hPa
pressure unit: hPa
backlight: On
scan: O
COMMON MENU
ch:
1
SET POINTS
SP1: 10 hPa SP2: 2.0 hPa

6.3 Menu overview

Press the Menu key to access the configuration mode. The following menus can be se­lected:
PRESSURE MENU After switching on, the DPG 202 is in the start menu PRES-
SURE. Here, the pressure value with the respective channel
number, the pressure unit and the status of the two switching relays are displayed.
CHANNEL MENU
This menu shows specific data for the selected channel (gauge type, gas type and correction factor). In addition there is also the function to select adjust for the respective sensors.
SENSOR MENU
Here the specific sensor data of the current channel is dis­played (type of gauge, sensor setup and the crossover setting of the combi gauges). The function degas is only possible on total pressure gauges using the hot cathode sensor.
RELAY MENU
This menu shows the respective parameters for the selected relay (channel assignment, set value and hysteresis with pres­sure unit). The current setting of the two switching relays is also displayed.
– big digit = relay on – small digit = relay off
16
SET POINTS
In this menu the two set points for the connected gauge can be set. This menu is only active when a transmitter with analog/re­lay option is connected.
COMMON MENU
Here the pressure unit can be selected. In addition, there is the option to switch on and off the display backlight (the lighting is
always activated when the unit is switched on). In AutoOff
mode the display backlight is automatically switched off if no key is pressed for 20 s.
By scanOn a new recognition of the connected gauges can be
forced. During the scan the status of the relays does not change, until the pressure values have been updated.

6.4 Configuration mode

set
ch: 1
CHANNEL MENU
type: ... gas: user corr: 2.00 adjust
set
ch: 1
CHANNEL MENU
type: ... gas: N2/O2/CO/Air corr: 1.00PI 1.00BA adjust
set
ch: 1
CHANNEL MENU
type: HPT 200 V010100 gas: N2/O2/CO/Air corr: 1.00PI 1.00BA adjust
set
set key: Settings are accepted, cursor jumps to next parameter.
Up/Down keys: Values are selected.
Menu key: Next menu will be displayed.
Operation
Gas type correction factor
For total pressure gauges employing a measuring method which is dependent on the type of gas, a correction factor can be assigned to the gauge. This will ensure, for a se­lected gas, the pressure display is corrected in a broad pressure range. On a combi­gauge such as the HPT 200 and the MPT 200 with two different measurement methods two correction factors are necessary (see the operating instructions for the gauge).
CHANNEL MENU ch: 1: Channel number type: Identified gauge type on channel 1 (e.g. HPT 200) gas: Set gas type for the selected gauge corr: Preset correction factor for the selected gauge adjust: Adjustment of range limits
CHANNEL MENU
1. Preprogrammed gas types:
The relevant correction factor will be displayed automatically at the parameter corr:
gas:N2/O2/CO/Air, CO2, H2, He, Ar, user Select the required gas type with Up/Down at the parameter
gas.
Confirm with set key.
CHANNEL MENU
2. Individual correction factor:
Select user specific correction factor with Up/Down at the
parameter corr. After altering a preset correction factor at the parameter gas the display will change to user.
To set a gas specific correction factor, enter a value in the
range of 0.2 – 8.0.
17
Operation
Re-adjusting For zero adjustment, ensure the pressure in the vacuum chamber is at least one decade
below the measuring range of the gauge and is not subject to large pressure fluctuations.
Bring gauge up to operating temperature (≥ 5 minutes).
set
ch: 1
type: ... gas: ... corr: ... adjust
ch: 1
type: ... gas: ... corr: ... adjust: HI/LO
ch: 1
type: ... gas: ... corr: ... adjust o.k. 995 mbar
CHANNEL MENU
LO
CHANNEL MENU
LO
CHANNEL MENU
set
HI
HI
set
set
CHANNEL MENU Press set key until the cursor flashes at the parameter ad-
just.
Press the Up-key; HI is displayed at the cursor position. Note: Within the limits of 800 to 1200 hPa the gauge can be ad-
justed to atmospheric pressure.
After pressing the set key the display shows the preset ref-
erence value (*)/(**).
Set the actual atmospheric pressure with the Up-/Down
keys (**).
By pressing the set key the new value is taken into account,
and the cursor jumps back to the channel number ch.
Successful adjustment is prompted by o.k.; if not o.k. is dis-
played the gauge could not be adjusted.
Set the zero adjustment LO in the same way. Note: When adjusting zero LO, ensure that the reference pres-
sure in the vacuum chamber is at least one decade below the measuring range of the gauge.
(*) The reference pressure can only be entered in the measurement units mbar, hPa or Torr. If, for
example the unit bar is selected, for reference pressure input purposes there is a temporary switch over to mbar, with mTorr to Torr and with Pa to hPa. When adjustment has been completed the orig­inally selected pressure unit is again active.
(**) These points are only required for atmospheric pressure adjustment of CPT100/200 and the
RPT100/200.
18
Operation
ch: 1
SENSOR MENU
type: HPT200 V010100 BAFil[1/2/auto]: On/O SwMode: trans_hi degas: O
SENSOR MENU
... BAFil[auto]: On/O ... ...
SENSOR MENU
... ... SwMode: trans_hi/lo/switch ...
set
set
set
set
SENSOR MENU
... ... ... degas: On/O
Sensor menu Total pressure gauges with hot cathode sensor (e.g. HPT 200 ) may need baking in ultra
high vacuum to desorp the gasload from surfaces. The degas process lasts for three minutes by default but can be stopped manually at any time. Pressure reading with re-
duced accuracy is possible during degas, the PRESSURE menu shows degas for the
concerned channel as long as the operation proceeds.
In the sensor menu the type dependent functions of the combination gauges RPT 100/ 200, HPT 100/200, and MPT 100/200 can be selected. The settings are described here­under taking the HPT 200 as an example.
SENSOR MENU
The high vacuum sensors of the combined gauges HPT 200 and MPT 200 can be switched manually: – BAFil = Filament of the BA sensor
CCSens = Cold cathode sensor
Press set key until the cursor flashes at the parameter BA-
Fil; the status of the function will be displayed (On or Off).
Switch filament On or Off using the Up/Down keys.Confirm with set key.
For the HPT 200 gauge the value in brackets define the select­ed filament:
 Select 1, 2 or auto.  Confirm with set key.
If auto is selected, an automatic switchover to the other filament
takes place when necessary (in case of wear).
For the MPT 200 gauge CCSens appears instead of BAFil.
Switch sensor on or off using the Up/Down keys.Confirm with set key.
Select switching mode for combination gauges: Press set key until the cursor flashes at the parameter
SwMode; the status of the function will be displayed
(trans_Hi, trans_Lo or switch, depending on the type of
gauge).
Use Up/Down key to select the desired function.Confirm with set key.
Activate the degas function: The function degas is only possible for total pressure gauges
with hot cathode sensor (HPT 100/200).
Press set key until the cursor flashes at the parameter de-
gas. The status of the degas function will be displayed (
or On).
Use the Up key to select degas On.  Start the degas process with the set key.
The degas process can be manually deactivated at any time by
switching the degas function to Off:
Press the set key until the cursor flashes at the parameter
degas.
Select the function Off with the Down key.Switch off the degas process with the set key.
Off
19
Operation
t
H
SP
0
1
p
Relay menu Set the thres hh ol d an d hyst ere si s va lu e s
The two relays are controlled by the two set points SP1 and SP2 (parameter press) and associated hysteresis H1 and H2 (parameter hsyt), as depicted in the illustration.
Fig. 5: Pressure and relay status curves illustrating threshold values and hysteresis
20
Operation
rel: 1
set
set
set
set
rel: 1 2
RELAY MENU
chann: ch1 press: 10 hPa hyst: 2.0 hPa
rel: 1
rel: 1 2
RELAY MENU
chann: ch1/ch2/o/on press: ... hPa hyst: ... hPa
rel: 1
rel: 1 2
RELAY MENU
chann: ... press: 10 hPa
hyst: ... ...
rel: 1
rel: 1 2
RELAY MENU
chann: ...
press: ... ...
hyst: 2.0 hPa
RELAY MENU
After selection of this menu via the Menu key the cursor flashes on the position rel: (relay number).
Select the required relay number with the Up/Down keys.Confirm relay number with the set key.
Assign channel number with Up/Down keys at the parame-
ter chann:
ch1/ch2: Relay switches according to the pressure value of
the selected channel.
off: Relay switches off independently of the set/actual com-
parison.
on: Relay switches on independently of the set/actual com-
parison.
Confirm channel setting with set key. Note: Each relay can be assigned to each of both gauge chan-
nels. It is also possible to link both relays with only one channel.
Alter the required set point with the Up-/Down keys in the
menu press.
Confirm set point with set key.Proceed in the same manner for the parameter hyst.
Note: The hysteresis is asymmetric to the set point, i.e.the re­lay switches on at the set point and off at set point+hysteresis. Altering the set point will cause the related hysteresis to be matched automatically by a percentage relation to the new set point. The preset value can be altered at will.
Example
Previous setting SP1: – press = 20 hPa (set point "on" 20 hPa)
– hyst = 2 hPa (set point "off" 22 hPa)
new set point SP1: – press = 400 hPa (set point "on" 400 hPa)
– hyst = 40 hPa (set automatically to 10 % of the threshold
value)
21
Operation
ch: 1
set
set
set
SET POINTS
SP1: 10 hPa SP2: 2.0 hPa
ch: 1
SET POINTS
SP1: 10 hPa SP2: 2.0 hPa
ch: 1
SET POINTS
SP1: 10 hPa SP2: 2.0 hPa
Set Point menu Setting the set points using DPG 202
The two relays are controlled by the two presettable set points SP1 and SP2. The hys­teresis H is fixed at 15% of the threshold values (see Fig. 5).
SET POINTS
After selecting this menu using the Menu key, the cursor flash­es at the position ch: (measurement channel).
Select the required channel number with the Up/Down keys. Use the set key to apply the channel number.
The cursor jumps to the pressure value for SP1.
Select the required pressure value for SP1 with the Up/
Down keys.
Confirm with the set key.
The cursor jumps to the pressure value for SP2.
Select the required pressure value for SP2 with the Up/
Down keys.
Confirm with the set key.
22
Operation
COMMON menu This menu offers the possibility to select the pressure unit and to switch on and off the
backlight (the backlight is always activated when the unit is switched on). The settings in this menu are independent of the channel.
COMMON MENU Pressure unit
The pressure unit of the display is shown on the right hand side
COMMON MENU
pressure unit: hPa/mbar..
backlight: On
scan: O
set
COMMON MENU
pressure unit: hPa
backlight: On/Auto O/O
scan: O
set
of the display. The settable pressure units are: hPa, mbar, bar, mTorr, Torr, Pa.
Press set key until the cursor flashes at the parameter pres-
sure unit.
Change to the desired unit using the Up-/Down keys.Confirm with set key; the cursor jumps to the next parameter.
Backlight
The backlight is always activated when the DPG 202 is switched on.
Press set key until the cursor flashes at the parameter back-
light.
Select the desired backlight setting between On/Auto Off/Off
using Up-/Down keys.
Confirm with Menu key; the display changes to the screen
PRESSURE.
In the Auto Off mode, the backlight is switched off if no key is pressed for 20 s. By pressing any key again the backlight is au­tomatically switched on again.
CH1: CH2:
REL: 1 2
PRESSURE
1005 noSensor
hPa
Perform scan:
By scanOn a new recognition of the connected gauges can be
forced. During the scan the status of the relays does not change, until the pressure values have been updated.
23
Malfunctions

7 Malfunctions

Voltage-bearing elements!
Danger to life from electric shock. Work on electrical installations may be carried out by trained and authorised electri-
cians only.
Disconnect the power supply and secure it against being switched back on.

7.1 Rectifying malfunctions

Errors Possible cause: Remedy
No display No power supply
The unit has not been able to communicate with a gauge.
Display: "noSensor"
Display: "ur" or "or": ● The measured pressure is outside
Display: "Err1" The connected gauge is defective.
DANGER
Mains voltage absent or to high/low
No gauge is connected
The address selection switch on
the gauge has not been set to the correct address (1 or 2)
The connected gauge or the serial interface on the DPG 202 is defec­tive
The data line between the DPG 202 and the gauge is not correctly connected or is faulty
the measuring range of the gauge
Note: If the error "Err1" occures be­cause of the defective BA filament of the HPT, "ur" briefly appears in the display. Afterwards "Err1" is dis­played
Check mains cable.Check mains voltage.
Check connection.
Check setting.
Exchange gauge.Exchange unit.
Check cable, exchange if
necessary.
Use suitable gauge.
Reset the error by discon-
necting from the mains. The Pirani sensor of the HPT gauge is then again ready for measurement.
24

8 Service

Service
Do make use of the Pfeiffer Vacuum service facilities. In the event that repairs are nec­essary a number of options are available to ensure any system down time is kept to min­imum:
Repair in the nearby Service center
Send unit and have it replaced with a new unit
Sending of units (Service Request)
For a quick and smooth handling of the service process, Pfeiffer Vacuum recommends the following steps:
Download the forms "Service Request" and "Declaration on Contamination".Fill out the "Service Request" form and send it by fax or e-mail to your local Pfeiffer
Vacuum service contact.
Include the confirmation on the "Service Request" from Pfeiffer Vacuum with your
shipment.
Fill out the "Declaration of Contamination" and include it in the shipment. This docu-
ment is mandatory to protect our service engineers. – Fill out and send one declaration for each device.
If possible, send unit in the original packaging.
1)
In the absence or incompleteness of the "Declaration on Contamination" and/or the use of unsuitable transport packaging, Pfeiffer Vacuum reserves the right to make a decon­tamination and/or to send the product back at the shipper’s expense.
Service orders
All service orders are carried out exclusively according to our repair conditions for vacu­um units and components. Detailed information, addresses and forms at:
http://www.pfeiffer-vacuum.com/service/repair-services/container.
1)
Forms under www.pfeiffer-vacuum.com
25
Accessories

9 Accessories

Designation DPG 202
Y-Connector M12 to RS-485 P 4723 010 Termination resistor for RS-485 PT 348 105 -T M12 m plug 4-pole with screw terminals RS-485 PT 348 106 -T Interface cable, M12 m straight/M12 m straight, 3 m PM 061 283 -T Connector M12 to RS-485 PM 061 270 -X Mains cable 230 V AC, CEE 7/7 to C13, 3 m P 4564 309 ZA Mains cable 115 V AC, NEMA 5-15 to C13, 3 m P 4564 309 ZE Mains cable, U.K. plug, 2.5 m P 4564 309 Y1 Mains cable, Swiss plug, 2.5 m P 4564 309 YR Software DokuStar, 2 channels PT 882 500

10 Technical data and dimensions

10.1 General

10.2 Technical data

Conversion table: pressure units
mbar bar Pa hPa kPa Torr
mbar 1 1 · 10
bar 1000 1 1 · 10
Pa 0.01 1 · 10 hPa 1 1 · 10 kPa 10 0.01 1000 10 1 7.5 Tor r
mm Hg
1.33 1.33 · 10
-3
-5
-3
-3
100 1 0.1 0.75
5
1 0.01 1 · 10
100 1 0.1 0.75
133.32 1.33 0.133 1
1 Pa = 1 N/m
1000 100 750
-3
2
Parameter DPG 202
[eng] Beschreibung The controller DPG 202 with integrated power supply can
Dimensions w x h x d (19"-rack, 14TE/3HE); 71 x 128 x 178 mm Connection for gauge M12 Protection category IP20 Power consumption max. 25 (gauges included) W Mains requirement: frequency (range) 50/60 Hz Mains requirement: voltage (range) 95-265 V AC Display LCD, backlit Display rate 2 Hz Feature USB Weight 0.75 kg Unit of measure mbar, bar, Pa, hPa, mTorr, Torr Measuring channel 2 Measurement rate 10 Hz Set point: Voltage max. 250 V AC, 4 A; 45 V DC, 2 A Set point: Changeover contact, potential-
free Temperature: Operating 5-50 °C Temperature: Storage -20-+60 °C
be used to connect up to two DigiLine gauges
2pieces
mm Hg
7.5 · 10
-3
26

10.3 Dimensions

Technical data and dimensions
128.5
71
66
106
172
189
27

Declaration of conformity

We hereby declare that the product cited below satisfies all relevant provisions accord-
ing to the following EC directives:
Electromagnetic Compatibility 2014/30/EU
Low Voltage 2014/35/EU
Restriction of the use of certain Hazardous Substances 2011/65/EU
DigiLine DPG 202
Harmonised standards and national standards and specifications which have been ap­plied:
EN 61326-1: 2013 Group 1 / Class B
EN 50581: 2012
EN 61010-1: 2010
Signature:
(Dr. Ulrich von Hülsen) Managing Director
Pfeiffer Vacuum GmbH Berliner Straße 43 35614 Asslar Germany
2016-05-30
Notizen / Notes:
Notizen / Notes:
Notizen / Notes:
VACUUM SOLUTIONS FROM A SINGLE SOURCE
Pfeiffer Vacuum stands for innovative and custom vacuum solutions worldwide, technological perfection, competent advice and reliable service.
COMPLETE RANGE OF PRODUCTS
From a single component to complex systems: We are the only supplier of vacuum technology that provides a complete product portfolio.
COMPETENCE IN THEORY AND PRACTICE
Benefit from our know-how and our portfolio of training opportunities! We support you with your plant layout and provide first-class on-site service worldwide.
Are you looking for a perfect vacuum solution? Please contact us:
www.pfeiffer-vacuum.com
Pfeiffer Vacuum GmbH
Headquarters • Germany T +49 6441 802-0 info@pfeiffer-vacuum.de
Loading...