These instruments photograph the subject of observation using a solid-state image sensor located at the endoscope tip under the
light transmitted from the processor/light source. The target of the observation is monitored by the physician using the endoscopic
image displayed on the video monitor. The endoscopic procedure is performed by inserting biopsy forceps and other endoscopic
accessories into the instrument channel inlet on the control body.
The bending section angulates in the intended direction and angle by operating the Angulation Control Lever; and air or fluids can
be suctioned from the distal end of the endoscope by operating the Suction Control Valve.
Indication for Use
The PENTAX Video Bronchoscopes (EB Family) have been designed to be used with a PENTAX Video Processor (including Light
Source), documentation equipment, video monitor, endo-therapy accessories (such as biopsy forceps) and other ancillary
equipment for endoscopy and endoscopic surgery within the airways and tracheobronchial tree.
Application
Medical purpose: Provide images for optical visualization, recording, and/or diagnostic aid.
Patient populations: Adults and lowercase pediatrics who have been determined by the physician to be appropriate candidates for
the use of this instrument.
Intended anatomical area: Airways and tracheobronchial tree
User: Medical doctors (expert approved by the medical safety officer to perform endoscopic examinations at each medical facility)
Place of Use: Medical facillity
Functions Used Frequently
The frequently used functions in these endoscope models are as follows:
Angulation capability using control lever
•
Remote control operation using remote buttons
•
Suctioning function
•
Removable Components
OF-B179
OF-B190
Suction Control Valve
Inlet Seal
Notes
Read this Instructions for Use (IFU) before reprocessing the endoscope, and save this book for future reference. Failure to read and
thoroughly understand the information presented in this IFU, as well as those developed for ancillary endoscopic equipment and
accessories, may result in serious injury, including infection by cross contamination to the patient and/or user. Furthermore, failure
to follow the instructions in this IFU or the companion Instructions for Use (operation) may result in damage to, and/ or malfunction
of, the equipment.
It is the responsibility of each medical facility to ensure that only well hyphen educated and appropriately trained personnel, who are
competent and knowledgeable about the endoscopic equipment, antimicrobial agents/processes, and hospital infection control
protocol be involved in the use and the reprocessing of these medical devices. Known risks and/or potential injuries associated with
flexible endoscopic procedures include, but are not limited to, the following: perforation, infection, hemorrhage, burns, and electric
shock.
This IFU describes the procedures for reprocessing and maintenance of the equipment after its use.
For inspection and preparation prior to its use, please refer to the separate Instructions for Use (Operation).
The text contained in this IFU is common to various types/models of PENTAX endoscopes, and users must carefully follow only
those sections and instructions pertaining to the specific instrument model in question.
If you have any questions regarding any of the information in this IFU or concerns pertaining to the safety and/or use of this
equipment, please contact your local PENTAX representative.
Sterility Statement
NOTE
Symbols on MarkingSymboles distinctifs
The endoscopes identified in this IFU are reusable semicritical devices. Since they are packaged non-sterile, they must be highlevel disinfected or sterilized BEFORE inital use. Prior to each subsequent procedure, they must be subjected to appropriate
cleaning and either high-level disinfection or sterilization processes.
Contraindication
Please consult regional and national health authority recommendations and requirements regarding protocols to follow in order to
reprocess and/or destroy endoscopes that will be used or have been determined to have been used (post procedure) on patients
afflicted with Creutzfeldt-Jacob Disease (CJD or vCJD).
Conventions
Throughout this IFU, the following conventions will be used to indicate a potentially hazardous situation which, if not avoided;
WARNING
CAUTION
: could result in death or serious injury.
: may result in minor or moderate injury or property-damage.
: may result in property-damage. Also advises owner/operator about important information on the use of
this equipment.
Prescription Statement
Federal (U.S.A) law restricts this device to sale by or on the order of a physician or other appropriately licensed medical
professional.
Symbols on Marking
Symboles distinctifs
Symbol for “MANUFACTURER”
Symbol for “DATE OF MANUFACTURE”
Symbol for “Authorised Representative in the European Union”
このCEマ ーキ ングは EC指令への適合宣言マークです。
The CE marking assures that this product complies with the requirements of the EC directive for safety.
Das CE Zeichen garantiert, daß dieses Produkt die in der EU erforderlichen Sicherheitsbestimmungen erfüllt.
Le logo CE certie que ce produit est conforme aux normes de sécurité prévues par la Communauté Européenne.
II marchio CE assicura che questo prodotto è conforme alle direttive CE relative alla sicurezza.
La marca CE asegura que este producto cumple todas las directivas de seguridad de la CE.
Attention, consult instructions for use
Attention, consulter le manuel d’utilisation
Type BF applied part (Safety degree specified by IEC 60601-1)
Partie appliquée de type BF (niveau de sécurité spécifié par la norme CEI 60601-1)
1 CARE AFTER USE ..................................................................................................................................... 6
1-1. General ............................................................................................................................................... 6
This symbol denotes an endoscope model's material compatibility with
the STERRAD
Endoscopes that do not have this symbol are incompatible with
the STERRAD
EB-1970TK)
®
NX™ system material compatibility identification symbol
®
®
– 1 –
Model Name
NX™ system. (EB-1575K, EB-1975K, EB-1990i)
NX™ system. (EB-1170K, EB-1570K, EB-1970K,
Minimum Instrument Channel Width
PVE CONNECTOR
STRAIN RELIEF BOOT
VENTING CONNECTOR
PVE SOAKING CAP OE-C9
ELECTRICAL
CONTACTS
VENTILATION CAP OF-C5
INSERTION TUBE
BENDING SECTION
Table of Minimum Instrument Channel Width
Video BronchoscopesEB-1170KEB-1570KEB-1970KEB-1970TKEB-1575KEB-1975KEB-1990i
Minimum Instrument
Channel Width (I.D.φmm)
1.22.02.83.22.02.81.2
– 2 –
NOMENCLATURE
Endoscope Components
・Suction Valve (OF-B179)
・Inlet Seal (OF-B190)
Accessories
・Bite Bock (OF-Z5)
Endoscope Components and Accessories
– 3 –
Accessories for Reprocessing
・PVE Soaking Cap (OE-C9)
・Ventilation Cap (OF-C5)
・Cleaning Brush (CS-C3S)
・Cleaning Brush (CS6002SN/CS6015ST/CS3010S)
・Cleaning Adapter (OF-B155)
Model NumberAppearance of BristlesLength of ShaftDiameter of Bristles
CS6002SN20cm
CS6015ST150cm
CS3010S100cm
– 4 –
φ
φ
φ
6mm
6mm
3mm
ENDOSCOPE REPROCESSING PROCEDURE FLOW
Pre-Cleaning
• Preparation
• Wiping of insertion tube
• Aspiration of detergent solution through suction channel
• Transport to cleaning room
Leak Testing
Cleaning
• Preparation
• Cleaning of all external surfaces
• Brushing of suction channel
• Filling of detergent solution into suction channel
• Soaking in detergent solution
• Rinsing
• Drying
High-Level Disinfection
• Preparation
• Filling of disinfecting solution into suction channel
• Soaking in disinfecting solution
• Rinsing
• Drying
Optional Sterilization
Sterilization using STERRAD® NX™ system (EB-1575K,
EB-1975K, EB-1990i only)
• Preparation
• Wrapping
• Sterilization Parameter Selection
CAUTION
STERRAD® NX™ Validation data has not been generated for endoscope
models in this manual other than EB-1575K, EB-1975K, and EB-1990i.
– 5 –
NOTE
1 CARE AFTER USE
1-1. General
This Instructions for Use (IFU) has been written in accordance with 21CFR
Part801, ISO 17664, and national guidelines on reprocessing of medical
products.
1-1-1. Application
WARNING
Reprocessing may affect device functionality. Prior to use, always inspect the
endoscope, components, and accessories for proper function to determine
that they are appropriate for patient use.
Components and Accessories for Video Bronchoscopes
MODELVideo Bronchoscopes
TYPE
Endoscope
Component
AccessoryBite Block - Adult SizeOF-Z5YY
Reprocessing
Accessory
Suction Control ValveOF-B179YY
Inlet SealOF-B190YY
PVE Soaking CapOE-C9YY
Ventilation CapOF-C5YY
Cleaning AdapterOF-B155YY
Cleaning BrushCS6002SNYY
Cleaning BrushCS6015STYN
Cleaning BrushCS3010SNY
Cleaning BrushCS-C3SYY
NameNumber
EB-1570K
EB-1970K
EB-1970TK
EB-1575K
EB-1975K
1
General
EB-1170K
EB-1990i
Y: YES
N: NO
– 6 –
1-1-2. Important Instructions
WARNING
1
General
• Reusable Medical Devices that are initially supplied non-sterile require
the end user to disinfect or sterilize them prior to initial use and to
subsequently reprocess them after each subsequent use.
• Proper care of the device after each procedure is extremely important.
Immediately (within one hour) after the completion of a procedure,
the endoscope and its removable components, and accessories should
be both pre-cleaned and mechanically cleaned with detergent solution.
Generally, if these endoscopes and accessories are not precleaned
within 15 minutes and mechanically cleaned within one hour after the
conclusion of the procedure, dried blood, mucus, or other patient debris
may cause damage to the devices or interfere with the ability of the user
to properly reprocess them.
• The use of detergent immediately after each procedure to dissolve and
remove organic contaminants and proteinaceous debris is essential to
the proper care and maintenance of the endoscope. Prior to disinfection
or sterilization, all instruments and components must be meticulously
cleaned. Failure to do so can result in incomplete or ineffective
disinfection or sterilization.
• Always inspect reprocessed endoscopes and accessories prior to use
according to their respective Instructions for Use (IFU).
• During the reprocessing process, always wear protective equipment
(e.g., gloves, gowns, face masks, etc.) to minimize the risk of cross
contamination.
• Contact the manufacturer and follow local regulations regarding safe use,
appropriate handling, and disposal of cleaning and disinfection solutions,
including alcohol and rinse water. Material Safety Data Sheets available
from the cleaning and disinfection solution manufacturer should be
consulted to provide guidance to end users about formulation, hazards,
chemical and physical properties, first aid, handling and storage,
stability, precautions, disposal, etc..
WARNING
Endoscopes are semicritical devices that require cleaning and at least highlevel disinfection. Use only legally marketed solutions and/or automated
endoscope reprocessors (AERs) for which validation testing with PENTAX
products has been performed by their manufacturers. A list of legally
marketed solutions/systems that have been determined to be compatible with
PENTAX brand products is contained in this manual.
– 7 –
CAUTION
To avoid damaging the endoscope, do NOT twist, rotate or excessively bend
any of the strain reliefs [(1), (2)] during inspection, clinical use, reprocessing,
or any handling activity. Be particularly cautious regarding the insertion tube
strain relief [(1)]. When wiping the insertion tube and the umbilical cable, use
a slow back and forth motion to wipe them along the tube/cable. Never apply
excessive force or torque to these strain reliefs or tubes/cables.
Do NOT Twist or RotateDo NOT Bend
1
General
(2)
(1)
Figure 1.1
CAUTION
• EB-1575K, EB-1975K, EB-1990i can be sterilized using the STERRAD®
NX™ system. For more detail, refer to sections 1-2-5-1 and 1-3-3-2 of
this instruction for use.
• Video bronchoscopes other than EB-1575K, EB-1975K, EB-1990i are
incompatible with sterilization using STERRAD® NX™ system.
• After every 100 cycles of STERRAD® NX™ exposure, the endoscope
should be returned to an authorized PENTAX service facility for repair.
Replacement of the insertion tube and bending section will be necessary
and other components may also require service.
• Although STERRAD® NX™ compatible endoscopes are generally capable
of withstanding up to 100 cycles of exposure to STERRAD® NX™, repair/
replacement of components might be necessary prior to 100 cycles,
depending upon the condition of the endoscope.
• Be sure to attach the ventilation cap (OF-C5) to the venting connector
before performing STERRAD® NX™ sterilization.
– 8 –
1
NOTE
NOTE
NOTE
General
This IFU contains detailed recommendations on the manual reprocessing of
PENTAX endoscopes using PENTAX supplied cleaning/disinfecting adapters.
AERs may also be used to reprocess flexible endoscopes. However, only
those AERs should be used whose manufacturers provide device-specific
instructions and have validation data to support each AER claim with respect
to PENTAX instruments. AER manufacturers should be consulted for their
specic claims including, but not necessarily limited to:
a) the ability of the AER to provide a cleaned and high-level disinfected (or
sterilized) endoscope and endoscope components (e.g., valves),
b) the identification of any special feature (internal channel) or endoscope
component that cannot be reprocessed and therefore requires manual
reprocessing,
c) the microbial quality of the rinse water,
d) the inclusion of an “automated” alcohol rinse cycle,
e) the inclusion of a terminal drying cycle that removes the majority of water
from within endoscope channels,
f) maintenance procedures for water filter replacement and/or
decontamination of the filtration system to ensure water of suitable
quality,
g) compliance with local regulations and/or guidelines.
PENTAX exible endoscopes should not be exposed to temperatures in excess
of 140oF (60oC) during either reprocessing or storage. During reprocessing
depending upon the detergent used, the endoscope may be damaged even if
the temperature does not exceed 140oF (60oC). A list of detergents that are
compatible with PENTAX endoscopes is contained in this manual.
All of the steps in the validated reprocessing protocol described in this
manual are intended to be performed in rapid succession and as a single,
continual procedure. There should be no breaks in between steps of the
protocol that are of sufficient duration to permit the endoscope to dry to
such an extent that dislodged debris and/or microbial contaminants would be
permitted to dry onto any endoscope surface. In the event that drying of the
endoscope occurs due to an excessive break in the reprocessing procedure,
the procedure should be completely repeated, beginning with the rst pre-
cleaning step.
– 9 –
1-1-3. Internal Channels of Video Bronchoscopes
The following internal schematic is designed to help users better understand the intricate
construction of PENTAX endoscopes. Please note that all solution entrance ports and flow