Pentair EVERPURE CONSERV LT-S User Manual

Page 1
EVERPURE
®
CONSERV® LT-S HIGH EFFICIENCY REVERSE OSMOSIS (RO) PROCESSOR
Conserv LT-S EV9975-20
INTRODUCTION
The Everpure Conserv LT-S Processor is a pre-assembled solution designed to provide high purity water for steam applications. It combines a number of water treatment technologies into one easy-to-install package. The Processor provides reduction of sediment. It also limits corrosion* and scale* by reducing TDS, chlorine* and chloramines*.
INSPECTION
Before mounting the Conserv LT-S Processor, inspect the parts. Use the table and illustration provided to verify the parts provided. If any parts are missing, contact Customer Service at 800.942.1153 (US only)/630.307.3000 or by email at CSeverpure@pentair.com.
Parts
1. Wall Unit with GRO-75EN RO Membrane, Cartridge in Place
2. 4CC Mineral Cartridge
3. 4FC5 Filter Cartridge
4. Storage Tank
5. Parts Bag Containing: Bypass Plug Red Tubing with Connecting Elbow Power Cord Blue Tubing Angled Ball Valve Ball Valve TDS Meter White Tubing
1
2
3
5
4
* Not performance tested or certified by NSF
FOODSE RV IC E
EVERPURE® CONSERV® LT-S HIGH EFFICIENCY RO SYSTEM INSTALLATION AND OPERATION GUIDE
®
Page 2
Tools Required
12.24”
• Hand or Electric Drill • Phillips Screwdriver
• Air Pressure Gauge • Towel
• Pencil • Safety Glasses
• Tape Measure • Drill Bits: 1/8", 1/4" and
3/8"
• Utility Knife or Tube Cutter
(for plastic tubing)
NOTE: Not all tools listed may be necessary for installation.
Read installation procedures before starting to determine required tools.
• Air Compressor with Schrader Coupling
SELECTING A MOUNTING LOCATION
1. Give consideration to the weight of the unit when operating, i.e. when filled with water. Operating weight of the Everpure® Conserv® LT-S Processor is approx.:
• 25 lbs (11.3 kgs)
2. The location should allow for:
• The Conserv LT-S Processor to be wall mounted.
• Adequate space for “in” and “out” water line
connections.
• Proximity to a drain for flushing.
• Access to a 100 - 240 Vac, 50/60 Hz AC wall plug.
• Placement of tank within 10 feet of Processor
3. The inlet water supply must meet the requirements listed in the section "Operating Specifications".
DIMENSIONS
(310.8 mm)
21.14”
(537 mm)
Figure 1
7.84
(199.2 mm)
SAFETY INSTRUCTIONS
NOTE: This appliance can be used by children aged
from eight (8) years and above and persons with reduced physical, sensory or mental capabilities or lack of experience and knowledge if they have been given supervision or instruction concerning use of the appliance in a safe way and understand the hazards involved.
NOTE: The appliance is only to be installed in
locations where its use, maintenance and cleaning is restricted to trained adult personnel. No one should play with the appliance.
NOTE: Children should be supervised to ensure that
they do not play with the appliance.
NOTE: Cleaning and user maintenance shall not be
made by children without supervision.
NOTE: The appliance is not suitable for installation in
an area where a water jet could be used.
NOTE: If the supply cord is damaged, it must be
replaced by the manufacturer, its service agent or similarly qualified persons in order to avoid a hazard.
WARNING: Installation shall comply with applicable
local, state, and national plumbing and electrical regulations.
WARNING: This appliance must be grounded. This
appliance is equipped with a cord having an appliance-grounding conductor and a grounding plug. The plug must be plugged into an appropriate outlet that is installed and grounded in accordance with all local codes and ordinances. If the supply cord is damaged, it must be replaced by the manufacturer, its service agent or similarly qualified persons in order to avoid a hazard.
WARNING: Improper connection of the appliance-
grounding conductor can result in a risk of electric shock. Check with a qualified electrician or service representative if you are in doubt whether the appliance is properly grounded. Do not modify the plug provided with the appliance; if it will not fit in the outlet, have a proper outlet installed by a qualified technician.
WARNING: The appliance is intended to be
permanently connected to the water mains and not connected by a hose-set. The appliance is not suitable for installation in an area where a water jet could be used and must not be cleaned by a water jet.
2 WATER YOU CAN TRUST
®
Page 3
WARNING: Before installing the Processor, make
certain your water supply complies with the following operating specifications. Failure to do so may reduce the effectiveness of the system and will void the warranty. Consult your local water treatment utility or a certified water testing lab to determine the quality of your water.
WARNING: Do not use with water that is
microbiologically unsafe or of unknown quality without adequate disinfection before or after the system. The Everpure® Conserv® LT-S Processor will not protect against disease-causing bacteria or remove harmless, naturally-occurring bacteria.
WARNING: The Conserv LT-S Processor contains
replaceable components that are critical to the efficiency of the system. Replacement of the reverse osmosis component should be with one of identical specifications, as defined by the manufacturer, to assure the same efficiency and contaminant reduction performance.
WARNING: The Conserv LT-S Processor contains filter
cartridges and an RO membrane cartridge included with the system that have limited service lives. Product water should be tested periodically to verify that the Processor is working properly.
WARNING: The Conserv LT-S Processor shall only be
used for arsenic reduction on chlorinated water supplies containing detectable residual free chlorine at the system inlet. Water systems using an in-line chlorinator should provide one (1) minute of chlorine contact time before the unit.
CAUTION: The Conserv LT-S Processor must be
protected against freezing which can cause the filter housing to crack, resulting in water leakage. Do not use electrical heating tape or propane torch on this unit.
CAUTION: Make certain the unit is firmly attached to
wall to prevent it from falling and possibly becoming damaged.
NOTE: Substances listed as reduced are not
necessarily in your water. The Processor must be maintained according to manufacturer's instructions, including replacement of filter cartridges.
NOTE: Your water must be within required limits
for satisfactory operation. If not, the RO membrane cartridge’s life may be shortened and your warranty will be voided (see Operating Specifications).
NOTE: Install on cold water line only.
NOTE: Do not install where system will be exposed to
direct sunlight.
NOTE: Make certain that installation complies with all
state and local laws and regulations.
NOTE: Use only plumbers tape without adhesive
backing to seal joints. Do not use pipe compound (“pipe dope”), sticks, or similar compounds with this unit; they contain petroleum derivatives which can cause crazing and cracking of the plastic in the filter
housing. NOTE: Use only soap and water to clean components. NOTE: Do not use aerosol sprays (bug spray, cleaning
fluids, etc.) near the Conserv LT-S Processor.
They contain organic solvents which can cause
crazing and cracking of the plastic in the filter
housing. NOTE: During extended periods of non-use, turn OFF
the water supply to the unit and remove the
cartridges. Place the cartridges in a sealed
plastic bag. Store them in a refrigerator for
future use. When restarting the unit, the
cartridges are reinstalled and flushed per
instructions.
OPERATING SPECIFICATIONS
WARNING: Before installing the Processor, make
certain your water supply complies with the following operating specifications. Failure to do so may reduce the effectiveness of the system and will void the warranty. Consult your local water treatment utility or a certified water testing lab to determine the quality of your water.
Inlet Temperature: 32 - 100°F (0 - 37.7°C)
Dynamic Operating Pressure:
Maximum Static Pressure
Inlet Total Dissolved Solids:
Sulfide, Iron and Manganese:
Water Supply pH Limits: 4 to 11
Turbidity: 1.0 NTU maximum
For cold water use only.
40 - 85 psi (2.7 - 5.8 bar, 0.27 - 0.58 MPa), non-shock
125 psi (8.6 bar, 0.86 MPa)
1,000 ppm maximum
Less than 0.1 ppm
®
CONSERV® LT-S HIGH EFFICIENCY RO SYSTEM INSTALLATION AND OPERATION GUIDE
3 EVERPURE
Page 4
EVERPURE CONSERV® LT-S PROCESSOR
1
Secure Unit to Wall
2
2
Remove RO
1
Membrane
3
Approximately
25 lbs
(11.3 kgs)
Operating
Weight
3
4 WATER YOU CAN TRUST
®
Page 5
2
Connect Valve to Storage Tank
Precharge Storage Tank 5-7 psi (0.34 bar, 0.034 MPa)
3
Connect Tubing for Flushing
1
3/8" Tubing Supply
Water In
2
OPEN
4
3/8" Tubing
Product Water
3/8" Tubing
To Storage
Tank
3
CLOSE
Out
4
Prepare to Flush the Everpure® 4FC5 Cartridge
Connect
Electrical
Unplugged
2
3
Inlet Water
CLOSE
1
Install
Bypass
Plug
®
CONSERV® LT-S HIGH EFFICIENCY RO SYSTEM INSTALLATION AND OPERATION GUIDE
5 EVERPURE
Page 6
5
Install and Flush Everpure® 4FC5 Cartridge
Plug In Power Cord
Unplug Power Cord
Inlet Water
OPEN
2
Cartridge
Install
3
5
Closed Valve
As Shown
Plugged
1
2
1
4FC5
4
Flush to Drain Five (5) Minutes
6
Prepare the Pentair
®
GRO-75EN Cartridge for Use
2
Replace
Membrane
1
Remove Cover
6 WATER YOU CAN TRUST
®
Page 7
7
Connect Tubing to Pentair® GRO-75EN Cartridge
Push White 3/8" Tubing
2
into White Connector
Remove Plug
1
Storage Tank Not Shown
8
Flush Pentair
Push Blue 3/8" Tubing
into Blue Connector
Push Red 3/8" Tubing
into Red Connector
®
GRO-75EN Cartridge
Plug In Power Cord
Unplug Power Cord
3
4
Inlet Water
Open
2
Fully CLOSE - Turn To Right
1
4
Flush To Drain
3
for 24 Hours
®
CONSERV® LT-S HIGH EFFICIENCY RO SYSTEM INSTALLATION AND OPERATION GUIDE
7 EVERPURE
Page 8
9
Set Blending Valve
Plug In Power Cord
2
Unplug In Power Cord
6
Open
Press
OPEN One (1) Full CCW
1
Turn of the Valve
Closed
3
4
5
Repeat Until
2"
2"
Flashing Blue
1/4 Turn or Less
Solid Blue
Blending Valve
Press
1/4 Turn or Less
Solid Blue
Blending Valve
1/8 Turn
Blending Valve
8 WATER YOU CAN TRUST
®
Page 9
10
Install Everpure® 4CC Cartridge
1
2
1
4CC
Remove Port
Plug
Install 4CC Cartridge
2
®
CONSERV® LT-S HIGH EFFICIENCY RO SYSTEM INSTALLATION AND OPERATION GUIDE
9 EVERPURE
Page 10
11
Flush Storage Tank and 4CC Cartridge
Plug In Power Cord
3
OPEN
1
CLOSE
When Pump Turns
Off, OPEN valve
2
Repeat Steps 2 and 4
5
4
Four (4) Times
10 WATER YOU CAN TRUST
®
Page 11
HOW REVERSE OSMOSIS (RO) WORKS
The Everpure Conserv LT-S Processor uses a semi­permeable membrane to reduce scale and corrosion. The RO membrane cartridge contains multiple layers of micron-thin film wound around a hollow center core. Water molecules can pass through the cartridge, while dissolved salts are rejected.
Your water supply is prefiltered to reduce dirt and chlorine that may foul the membrane. The RO membrane cartridge separates this prefiltered water into RO WATER and REJECT WATER. Water pressure forces water through the membrane within the RO membrane cartridge then into the storage tank. This is the RO water. Dissolved salts cannot pass through the membrane and are sent to the drain as reject water.
The Conserv LT-S Processor features a blend-in-tank process which uses the RO water described above, with calcium carbonate, and adds filtered water from carbon filtration to create a consistent product water. When you use water from the storage tank, the pressure inside the tank drops and the pressure switch closes. The system then begins to operate, replenishing the water you took from the storage tank. The storage tank can hold up to 3 gals (11.3 L) of water at one time.
NOTE: When used under operating conditions specified
in this manual, the RO membrane cartridge of the Conserv LT-S Processor should last 12 to 24 months. The precise life span of the Pentair® GRO-75EN RO Membrane Cartridge will depend on the quality of the water entering the system and the frequency with which you use it. Frequent use prevents the dissolved salts from building up on the membrane as scale. The more water the system is required to produce, the longer the membrane will last.
1
Secure Manifold Wall Mount:
The Conserv LT-S Processor weighs approx. 25 lbs (11.3 kgs).
CAUTION: The Conserv LT-S Processor should be mounted
on a stud or firm surface. The mounting bracket will support the weight of the system and help prevent strain on the water lines.
1. Remove the Pentair GRO-75EN Membrane from the clip on the wall unit.
2. When mounting the Conserv LT-S Processor on a wall, four (4) screws are required to mount the unit. The two (2) top screws will be 10.5" (26.7 cm) apart and level. Screw the two (2) screws into the wall. Leave a gap between the screw head and the wall of 1/8". The backside of the unit has two (2) slots that will fit over the screw heads and slide down to lock (Figure 2).
3. Tighten both screws from the front.
4. Insert screws into the two (2) remaining mounting holes. Screw them in and tighten.
Mounting Screw Slots
10.50”
(26.7 cm)
4x
0.45”
(1.1 cm)
INSTALLATION
NOTE: Please read all instructions, specifications, and
precautions before installing and using your Conserv LT-S Processor.
NOTE: The Conserv LT-S Processor may be installed
hanging on a wall or set under a sink.
NOTE: The Conserv LT-S Processor must be installed
vertically.
WARNING: Improper connection of the appliance-
grounding conductor can result in a risk of electric shock. Check with a qualified electrician or service representative if you are in doubt as to whether or not the appliance is properly grounded. Do not modify the plug provided with the appliance; if it will not fit in the outlet, have a proper outlet installed by a qualified technician.
Figure 2
2
Assemble Storage Tank:
1. Locate the right angle ball valve. Screw the valve onto the top fitting of the Storage Tank until tight.
2. CLOSE the ball valve.
3. Apply precharge of 5-7 psi (0.34 bar, 0.034 MPa)
4. Position the Storage Tank near the Conserv LT-S Processor.
®
CONSERV® LT-S HIGH EFFICIENCY RO SYSTEM INSTALLATION AND OPERATION GUIDE
11 EVERPURE
Page 12
3
Connect Tubing for Cartridge Flushing:
1. Refer to Figure 3 for a general description of connection points.
2. Prepare the plumbing to accept the RO system.
NOTE: The blended water tubing/piping and associated
fittings connecting the RO blended water outlet to the equipment being serviced should be food-grade material that meets NSF® Standard 51, 61, or similar, with a minimum pressure rating of 125 psi (8.6 bar,
0.86 MPa). The blended water may react with metal piping, creating a corrosive condition. Plastic pipe or reinforced beverage tubing are generally very good material choices for RO water distribution piping. The size of the blended water tubing/piping should be 3/8" (9.5 mm) ID minimum. Distances of 25 feet (7.6 m) or greater from the RO processor to the equipment being serviced should be 1/2" (12.7 mm) ID minimum.
3. Locate an appropriate point in the hard water plumbing to tie in a new section of tubing/piping that can be routed to the RO processor. CLOSE the nearest shut-off valve that is upstream from the tie-in point and relieve the pressure in the line. Install a 3/8" (9.5 mm) tubing/piping section and connect it to the INLET - Supply Water fitting. See Figure 3.
4. Connect a length of 3/8" O.D. tubing to the connection for the Storage Tank. Route the other end to the Storage Tank and insert it into the ball valve. The ball valve should be closed.
5. Connect a length of 3/8" O.D. tubing to the Everpure Conserv LT-S Processor outlet connection. Secure an open ball valve to the other end and position it over a drain.
4
Prepare to Flush the Everpure 4FC5 Cartridge:
The Everpure 4FC5 Filter Cartridge is installed and flushed before the Pentair GRO-75EN RO and the Everpure 4CC Cartridges are installed.
1. Install the bypass plug into the filter head for the Everpure 4CC Cartridge. This head is centered and toward the front of the unit.
2. Connect the electrical cord to the socket located on the power supply.
3. The inlet water should be connected and turned off.
Power Cord
Inlet Water
Bypass
Plug
Water
Supply Inlet
Figure 3
Figure 4
CAUTION: The cartridges are identified on the label using
the model number: Pentair® GRO-75EN RO Membrane Cartridge Everpure® 4FC5 Filter Cartridge Everpure® 4CC Mineral Cartridge
Install the cartridges in the correct locations. If the cartridges are installed improperly, the processor will not operate correctly.
5
Install and Flush the Everpure 4FC5 Filter Cartridge:
Locate the 4FC5 Filter Cartridge and inspect the o-rings. They should be clean and lightly lubricated with silicone grease.
1. Line up the tabs with the slots in the opening on the left side of the head assembly. See Figure 5 and Figure 6.
2. Insert the cartridge to full depth.
3. Rotate the cartridge 900 to the right (clockwise) to lock it in place.
12 WATER YOU CAN TRUST
®
Page 13
Figure 5
4. OPEN the valve for the inlet water.
5. Plug the power cord into a 115 V wall outlet.
6. Flush to drain for five (5) minutes.
7. Unplug the power cord.
1
2
White Port
Figure 7
CAUTION: DO NOT remove the quick-connect elbow fitting
from the red tubing. This assembly contains a flow control that may become damaged if removed. The efficiency of the Everpure Conserv LT-S Processor may be affected.
1. Remove the GRO-75EN RO Membrane Cartridge from its packaging and position it with the single white port at the top.
2. Press the cartridge into the clip on the back plate. The cartridge may be moved up or down to line up with the tubing connections.
7
Connect Tubing to Pentair GRO-75EN:
4FC5
Flush to
Drain
Figure 6
6
Install the Pentair GRO-75EN RO Membrane
Cartridge:
The GRO-75EN RO Membrane Cartridge is installed in the vertical position. There are three (3) quick-connect ports that are labeled:
• Inlet - White
• Product (RO Water) - Blue
• Drain - Red
White
Blue
Red
Figure 8
®
CONSERV® LT-S HIGH EFFICIENCY RO SYSTEM INSTALLATION AND OPERATION GUIDE
13 EVERPURE
Page 14
1. Insert RO inlet connection into the white quick-connect
0-75 ppm (solid blue)
port of the RO filter.
2. Insert drain water connection into the red quick-connect port of the RO filter.
3. Insert RO water connection into the blue quick-connect port of the RO filter.
8
Flush the Pentair® GRO-75EN RO Membrane
Cartridge:
Blend Valve
Figure 9
1. Close the blend valve completely by turning it to the right (clockwise) until it stops.
2. Apply power.
3. Flush to drain for 24 hours.
4. Turn OFF power.
USING THE TDS METER TO TEST FOR TOTAL DISSOLVED SOLIDS
After the processor has been flushed, it is still necessary to test and set the Total Dissolved Solids (TDS) to an acceptable level. A TDS meter is included for testing purposes.
The included TDS Meter is designed for limited use. After six (6) tests, the meter may not be accurate and should be discarded.
Probe End
Figure 10
NOTE: For better accuracy and frequent testing for TDS, use
a standard pocket TDS meter.
9
Set Blending Valve
The water exiting the Everpure Conserv LT-S Processor should have a TDS reading of 80 ppm before the Everpure 4CC Cartridge is added.
NOTE: The blend adjustment valve will be opened one (1)
full turn to take the initial reading.
1. CLOSE the blend adjustment valve, then OPEN it one (1) full turn to the left (counterclockwise).
2. Turn power ON and collect 1" to 2" of RO blended water in a clean glass container.
3. Insert probe end of meter no more than 1" into water, and depress blue cap. Do not submerge entire unit.
4. One (1) or two (2) lights will light immediately. Match light(s) to the chart to get approximate range of Total Dissolved Solids (TDS). Release the blue cap, and write down the TDS range.
76-150 ppm (flashing blue)
100-250 ppm (flashing blue + solid green)
251-300 ppm (solid green)
301-400 ppm (flashing green)
401-500 ppm (solid green + flashing green)
501-600 ppm (flashing green + solid red)
601-750 ppm (flashing red)
751-900 ppm (solid red)
The Conserv LT-S Processor includes a blend system that can combine or "blend" water sources. OPENING the blend adjustment valve will add untreated water with higher TDS to the RO water.
To achieve the target TDS of 80 ppm, it is likely that the blend valve will need to be adjusted. Use the TDS reading from the previous steps as the starting point. If that number is above 80 ppm, the blend adjustment valve will need to be closed slightly. If that number is below 80 ppm, the blend adjustment valve will need to be opened slightly.
NOTE: Changes to the blend adjustment valve should be
made in small increments. Adjust the valve in 1/8 to 1/4 of a rotation for best results.
The desired reading of 80 ppm is at the high end of the solid blue range. To be certain the blended water is at or near 80 ppm, use the following procedure:
1. Run blended water into a glass and collect 1" to 2" of water.
2. Test this water for a TDS level. Based on the result, the mixing valve will be opened or closed.
• If the meter showed a solid blue light, then the mixing
valve will be OPENED more.
• If the meter showed any other indicator, then the
mixing valve will be CLOSED more.
14 WATER YOU CAN TRUST
®
Page 15
3. Adjust the mixing valve if needed and run blended water for two (2) minutes. Discard this water.
4. Repeat Steps 3, 4 and 5 until the light indicators change between solid blue and flashing blue. The blended water TDS should now be at 80 ppm.
10
Install the Everpure® 4CC Mineral Cartridge:
Locate the 4CC Cartridge and inspect the o-rings. They should be clean and lightly lubricated with silicone grease.
1. Turn power OFF.
2. Remove the port plug.
3. Line up the tabs on the 4CC Cartridge with the slots in the opening on the front side of the head assembly. See Figure 6 and insert the cartridge to full depth.
4. Rotate the cartridge 900 to the right (clockwise) to lock it in place.
1
2
4CC
Blended
Water Outlet
Figure 12
1. OPEN the ball valve on the top of the Storage Tank.
2. CLOSE the ball valve at the blended water out tubing.
3. Apply power. It will take approximately two (2) hours for the Storage Tank to fill.
4. The pump turns off when the Storage Tank is full. OPEN the ball valve for the blended water outlet. Allow the Storage Tank to empty.
5. Repeat Steps 2, 3 and 4 four (4) more times.
Figure 11
11
Flush Storage Tank and Everpure® 4CC Cartridge:
Flush the Everpure 4CC Mineral Cartridge:
The 4CC Mineral Cartridge will be flushed when the storage tank is flushed. No further flushing is required.
Storage Tank:
The Storage Tank will hold approximately 3 gals (11.3 L). The tank air pressure is preset to 5-7 psi (0.34 bar, 0.034 MPa) when shipped via ground transportation. For air shipments the precharge is completely removed. Check the tank air pressure. Ensure that the tank has 5-7 psi (0.34 bar, 0.034 MPa) of precharge upon installation. Add air if needed. Tank pressure may be checked using an air pressure gauge on the air valve on the side of the tank.
PLACE THE CONSERV LT-S PROCESSOR INTO OPERATION
1. Remove the ball valve from the end of the blended water outlet.
2. Connect the blended water outlet tubing to the application.
TESTING YOUR REVERSE OSMOSIS PROCESSOR
Total Dissolved Solids (TDS) Test
Under NSF/ANSI Standard 58, it is highly recommended that the blended water be tested at least every six (6) months to verify that the processor is performing satisfactorily.
CAUTION: Visually check the entire processor for leaks. If a
leak is present, see Troubleshooting.
NOTE: Initially, the water may appear cloudy. This is a result
of air trapped in the postfilter cartridge. It is not harmful and will disappear in a matter of minutes. It may take up to a week after installing a new postfilter cartridge for the trapped air to dissipate.
®
CONSERV® LT-S HIGH EFFICIENCY RO SYSTEM INSTALLATION AND OPERATION GUIDE
15 EVERPURE
Page 16
WHEN TO CHANGE THE CARTRIDGES
The life of the cartridges depends on the water volume used and the substances in the water. Normally, cartridges should be changed at intervals of 12 months. Replace the cartridges sooner if the water pressure at the application equipment begins to drop noticeably or if you notice changes in the flow of the blended water.
White
Tubing
CHANGING THE CARTRIDGES
Replacement Cartridges:
Everpure 4FC5 Filter Cartridge - EV9693-21
Everpure 4CC Mineral Cartridge - EV9627-45
Pentair® GRO-75EN RO Membrane Cartridge - 4002575
The 4FC5 and 4CC Cartridges need to be changed when replacing the GRO-75EN RO Membrane Cartridge. All cartridges should be changed at the same time.
NOTE: Only the replacement cartridges listed can be used
with this processor. Failure to use the recommended replacement cartridges will void your warranty.
1. Place a small pan or towel under the Everpure Conserv LT-S Processor to catch any water that may drip. Turn OFF incoming water.
2. Close the ball valve on the the Storage Tank and OPEN the application equipment valve to relieve pressure.
3. Remove the two (2) cartridges by twisting each cartridge a 1/4-turn to the left (counterclockwise). Then pull down to release.
4. Remove the GRO-75EN RO Membrane Cartridge by removing the tubing from all three (3) connections. Then pull the cartridge out of the clip.
5. At the bottom of the cartridge, remove the quick-connect elbow from the red tubing to gain access to the flow controller and the attached capillary tube. Replace the flow controller (Figure 13) in the red tubing.
6. Insert a new flow controller (Figure 13) with capillary tube into the tubing. Replace the quick-connect elbow.
Blue
Tubing
Flow
Controller
Red Tubing
Figure 13
7. Insert the new GRO-75EN RO Membrane Cartridge into position in the clip. Rotate the cartridge so the quick­connect ports line up with the tubing.
8. Push the red and blue tubing into the quick-connect ports at the bottom. Push the top white tubing into the connector at the top of the cartridge.
NOTE: The bottom of the GRO-75EN RO Membrane Cartridge
will be supported by the lower bracket.
9. Position the 4FC5 Cartridge under the left side head. Line up the tabs on the cartridge with the slots in the head.
10. Push the new cartridge into the head until it can go no further.
11. Twist the cartridge a 1/4-turn to the right (clockwise) until it locks in place. Pull gently down to ensure the cartridge is locked in place.
12. Flush the 4FC5 Cartridge following the procedure in the Installation Instructions in Step 5.
13. Repeat Steps 10, 11 and 12 for the 4CC Cartridge in the front center head.
CAUTION: The 4CC Cartridge may contain carbon fines that
will be removed during the initial flushing. Follow the proper flushing procedure to prevent damage to the GRO-75EN RO Membrane Cartridge.
16 WATER YOU CAN TRUST
INITIAL FLUSHING OF CARTRIDGES
The installation of new cartridges requires that a preliminary flushing of the cartridges be performed. This initial flushing is required to prevent damage to the RO cartridge.
Refer to the Installation section for the complete flushing procedure.
®
Page 17
EVERPURE CONSERV LT-S SANITIZING PROCEDURE ­CONTAMINATED WATER SUPPLY/BOIL ALERTS
When the water system has been contaminated or a boil water advisory has been discontinued, the Conserv LT-S Processor must be sanitized. If a boil water advisory has just been discontinued: Flush the water lines according to your water supplier's instructions. OPEN the delivery faucet and flush the processor for about 10 minutes, then follow the filter cartridge replacement instructions below.
1. Turn OFF the water supply and OPEN the application valve to relieve pressure.
2. Remove the filter cartridges and discard in trash. Wash hands with soap and warm water.
NOTE: If a drip pan is used, clean with dish soap and
warm water.
NOTE: The following steps require placing bleach into the
Conserv LT-S Processor. Pentair® offers a hollow cartridge, Everpure® 2JT Flushing/Sanitizing Cartridge (EV9608-10), that is designed for this application. The following steps describe how to use the Pentair Everpure 2JT Cartridge (EV9608-10).
CAUTION: DO NOT install the Pentair® GRO-75EN RO
Membrane Cartridge for the sanitization procedure. The solution will severely damage the membrane.
Sanitization Procedure with the Pentair Everpure 2JT Cartridge (EV9608-10)
3. Grasp the tip of the outlet tower center, at the top of the cartridge and pull out gently. The entire central plug with the long tube attached will come out, leaving a hole.
Figure 14
4. Add about a teaspoon of laundry bleach (5.25% sodium hypochlorite solution) to the empty flushing cartridge through the hole in the top.
5. Lubricate o-ring with a high quality silicone lubricant. Replace the central plug assembly and insert the flushing cartridge into the left side head for the Everpure 4FC5 Cartridge.
NOTE: Do not use petroleum jelly to lubricate the o-rings. It
will degrade the o-rings.
Connect White
and Blue
Tubing
Cartridge
with
Sanitizer
Plug 4CC
2JT
Port
Figure 15
6. Install a plug into the port for the 4CC Filter Cartridge.
NOTE: A plug was included with the Conserv LT-S Processor.
7. The tubing for the removed GRO-75EN Cartridge must be connected together. Make a length of tubing with angle quick-connectors. Use the length of tubing with two right-angled quick-connectors to connect the white tubing at the top to the blue tubing at the bottom. The red tubing at the bottom is waste water and does not need to be sanitized.
8. The tank valve should be open.
9. The application valve should be open.
10. The power switch should be on.
11. Slowly OPEN the water supply valve. Monitor the water flowing out of the Conserv LT-S Processor at the application valve. When the smell of chlorine is present, CLOSE the application valve. The Storage Tank will fill with chlorinated water.
12. Wait for a period of at least 30 minutes to allow adequate disinfection time.
13. Turn OFF the water supply valve.
14. OPEN the application valve. The Storage Tank will drain out.
15. When the Storage Tank is emptied, CLOSE the application valve and the tank valve. Remove the 2JT Cartridge.
16. Install a new 4FC5 Cartridge in the first (left side) head. Follow the instructions in the section "Installing the Cartridges" to flush the 4FC5 Cartridge.
17. OPEN the application valve and slowly OPEN the water supply valve. Allow water to run for five (5) minutes. This will flush carbon fines out of the new cartridge.
18. CLOSE the application valve and OPEN the Storage Tank valve, allowing the tank to fill three (3) to five (5) minutes.
®
CONSERV® LT-S HIGH EFFICIENCY RO SYSTEM INSTALLATION AND OPERATION GUIDE
17 EVERPURE
Page 18
19. Turn OFF the water supply valve.
20. OPEN the application valve to allow the tank to empty.
21. When the water flow has stopped, CLOSE the application valve and OPEN the water supply valve to refill the tank for three (3) to five (5) minutes.
22. Repeat Steps 18, 19 and 20 three (3) times.
23. Turn OFF the water supply valve and OPEN the application valve to remove pressure.
24. Remove the plug from the front center port. Install a new Everpure® 4CC Filter Cartridge.
25. With the electrical power on, OPEN the water supply valve. Flush water out of the application valve for five (5) minutes.
26. Remove power.
27. CLOSE the water supply valve and OPEN the application valve to remove pressure.
28. Remove the tubing that connects the Pentair® GRO-75EN RO Membrane Cartridge white (top) and blue (bottom) processor tubing.
29. Install a new GRO-75EN RO Cartridge. Connect the three (3) tubes to the proper ports. The tubing and the ports are color coded.
30. OPEN the water supply valve and the application valve.
31. Apply power and allow the application valve to drip for 12 hours.
32. CLOSE the application valve. The Storage Tank will fill. Check for leaks during this time.
The Conserv LT-S Processor is ready for operation.
4. Inspect the rubber seal inside the threaded area. Repair or replace the valve as needed.
5. Thread the tank valve onto the top of the tank opening by turning it clockwise until snug.
6. Wet the tubing and insert it 3/4" (1.9 cm) into the quick­connect fitting.
7. Turn water supply ON and CLOSE the application valve.
8. Allow processor to pressurize for several hours and check for leaks.
9. Check for leaks after tank is fully pressurized [one (1) to three (3) hours]. If leak persists, call Technical Support .
Leaks at Quick-Connect Fittings
CAUTION: The 1/4" red tubing connected to the bottom
drain connection of the RO cartridge has a flow restrictor installed. The processor will not operate correctly if the restrictor is removed.
1. CLOSE tank valve, turn OFF cold water supply to the processor, and OPEN the application valve.
2. Press collar around the quick-connect fitting while pulling the tubing with your other hand.
3. Tubing should be cut squarely. Internal and external burrs should be removed. Place a mark on tubing 5/8" (1.6 cm) from end on 1/4" tubing or 3/4" (1.0 cm) from end on 3/8" tubing.
4. Wet the end and insert tubing until the mark is flush with the quick-connect fitting.
5. OPEN the cold water supply until it comes to a stop. OPEN the tank valve and CLOSE the application valve. If leaks persist, call Technical Support.
TROUBLESHOOTING
NOTE: If leaks persist, or if there are other leaks in the
processor, turn OFF water supply and call Technical Support 800.942.1153 (US only)/630.307.3000.
Leaks Between Cartridge and Filter Housing
1. Turn OFF cold water supply to processor. CLOSE tank valve. OPEN the application valve to relieve water pressure.
2. Remove the cartridge, inspecting it for damage. Inspect o-rings to make sure they are seated and clean.
3. Insert and twist the cartridge back into the filter housing.
4. Turn ON water supply. OPEN tank valve. CLOSE the application valve after water begins to flow. If leaks persist, call Technical Support.
Leak Between Tank Valve and Storage Tank
1. Turn OFF water supply to processor. OPEN the application valve to drain storage tank. Run for three (3) to five (5) minutes until water drips.
2. Remove tubing from tank valve by pressing the collar around the fitting while pulling the tubing with your other hand.
3. Unscrew the tank valve from the storage tank.
No Flow or Slow Flow from the Brine (Reject) Line (Less than 6 fl. oz. or 180 mL per minute)
NOTE: Before checking brine (or reject) flow, make sure the
unit is producing water by turning the valve on the storage tank OFF and OPENING the application valve. Water should drip from valve.
1. Replace the prefilter cartridge according to the "Changing the Cartridge Instructions" and recheck the brine (or reject) flow rate.
2. If the prefilter cartridge is not at fault, the brine (or reject) flow controller could be clogged. Call Technical Support.
High TDS in Product Water
1. If high TDS (Total Dissolved Solids) is detected in the blended water, the RO cartridge may need to be replaced or the reject flow control tubing may be clogged. If this is a new installation, call Technical Support.
2. Otherwise, draw 1 gal (3.8 L) of water from the unit. After 10 minutes, run water from the blended water port and test the water again.
3. Determine when you last changed the RO cartridge and call Technical Support.
18 WATER YOU CAN TRUST
®
Page 19
Limited Flow at the Application Equipment Valve
1. Turn OFF water supply to processor.
2. OPEN the application valve to turn on flow.
3. Unscrew the blue cap on the side of the Storage Tank to expose air valve. Use a small air compressor or bicycle pump to slowly add air to the storage tank. This will force the water out of the storage tank through the valve. Continue to slowly add air until water stops coming out of the valve.
4. Turn OFF the application equipment valve.
5. Using an air pressure gauge, adjust the pressure in the storage tank to 5-7 psi (0.34 bar, 0.034 MPa).
6. Replace the blue cap.
7. OPEN the incoming water valve until it comes to a stop. Let the processor run one (1) to three (3) hours to fill the tank. A full tank weighs approximately 32 lbs (14.5 kgs) pounds. If performance has not improved, call Technical Support.
Sudden Return of Taste and Odor
If shortly after complete servicing, noticeable taste and odors return, contact Technical Support.
Daily Production Rate: 70 gpd (265 Lpd).
This reverse osmosis system contains a replaceable component critical to the efficiency of the system. Replacement of the reverse osmosis component should be with one of identical specifications, as defined by the manufacturer, to ensure the same efficiency and contaminant reduction performance.
REPLACEMENT PARTS
Contact your local Pentair® Everpure Dealer for filter cartridges or system replacement parts.
Description Part Number
GRO-75EN RO Membrane Cartridge
4FC5 Filter Cartridge
4CC Mineral Cartridge
Storage Tank
TDS Meter
Port Plug
Waste Water Restrictor
Conserv LT-S Cartridge Kit
4002575
EV9693-21
EV9627-45
EV3146-98
EV3125-98
EV3108-76
EV3128-31
EV9976-25
NOTES
If you are experiencing a problem not listed in this manual, shut OFF the water supply and CLOSE the tank valve. Call Technical Support.
Check for compliance with state and local laws and regulations.
Do not use with water that is microbiologically unsafe or of unknown quality without adequate disinfection before or after the system.
For installations in Massachusetts, the Commonwealth of Massachusetts Plumbing Code 248 CMR shall be adhered to. Consult your licensed plumber for installation of the system. This system and its installation must comply with state and local regulations.
Systems certified for cyst reduction may be used on disinfected waters that may contain filterable cysts.
Substances that may be reduced are not necessarily in your water. Filter must be maintained according to manufacturer’s instructions, including replacement of filter cartridges.
The efficiency rating for this system is 56.31%. Efficiency rating means the percentage of the influent water to the system that is available to the user as reverse osmosis treated water under operating conditions that approximate typical daily usage.
The tested recovery rating is 63.02%. Recovery rating means the percentage of the influent water to the membrane portion of the system that is available to the user as reverse osmosis treated water when the system is operated without a storage tank or when the storage tank is bypassed.
Conserv LT-S Tested and Certified by NSF International against NSF/ANSI 58 for the reduction of claims specified on the Performance Data Sheet and at NSF.org.
Water Treatment
Appliance - E359339
®
CONSERV® LT-S HIGH EFFICIENCY RO SYSTEM INSTALLATION AND OPERATION GUIDE
19 EVERPURE
Page 20
LIMITED WARRANTY
COMMERCIAL WATER TREATMENT EQUIPMENT
You have just purchased one of the finest water treatment units made. As an expression of confidence in this product, Pentair Filtration Solutions, LLC (“PFS”) offers the following product warranty. This product is warranted against material defects in materials and workmanship to the original end-user when installed in accordance with the PFS specifications. The warranty period commences on the date of purchase and is administered as follows:
For a period of ONE YEAR Replaceable elements (i.e., filter & water treatment cartridges)* For a period of ONE YEAR The entire system (excluding replaceable elements)
The unit must be used in operating conditions that conform to PFS’s recommended guidelines. This warranty will not apply if the unit has been modified, repaired or altered by someone not authorized by PFS.
If a part described above is found to have a material defect in materials or workmanship within the specified warranty period, you should notify Pentair® Everpure technical service at the phone number listed below. Any part found materially defective within the terms of this warranty will be repaired or replaced (at PFS’s discretion) by your local dealer or Pentair Everpure technical service. You pay only freight from our factory and local dealer charges. Any item repaired or replaced pursuant to this warranty will be covered under the original warranty terms of the system.
PFS is not responsible for damage caused by accident, fire, flood, freezing, Act of God, misuse, misapplication, neglect, oxidizing agents (such as chlorine, ozone, chloramines and other related components), alteration, installation or operation contrary to our printed instructions, or by the use of accessories or components which do not meet PFS’s specifications. Refer to the specifications section in the Installation and Operating manual for approved application parameters.
Our product performance specifications are furnished with each water treatment unit. TO THE EXTENT PERMITTED BY LAW, EVERPURE DISCLAIMS ALL IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION WARRANTIES OF MERCHANTABILITY AND FITNESS FOR PARTICULAR PURPOSE; TO THE EXTENT REQUIRED BY LAW, ANY SUCH IMPLIED WARRANTIES ARE LIMITED IN DURATION TO THE PERIOD SPECIFIED ABOVE FOR THE ENTIRE WATER TREATMENT UNIT.
As a manufacturer, we do not know the characteristics of your water supply or the purpose for which you are purchasing this product. The quality of water supplies may vary seasonally or over a period of time, and your water usage rate may vary as well. Water characteristics can also differ considerably if this product is moved to a new location. For these reasons, we assume no liability for the determination of the proper equipment necessary to meet your requirements, and we do not authorize others to assume such obligations for us. Further, we assume no liability and extend no warranties, express or implied, for the use of this product with a non­potable water source or a water source which does not meet the conditions for use described in the owner’s guide or performance data sheet for this product.
OUR OBLIGATIONS UNDER THIS WARRANTY ARE LIMITED TO THE REPAIR OR REPLACEMENT (AT PFS’S DISCRETION) OF THE FAILED PARTS OF THE WATER TREATMENT UNIT, AND WE ASSUME NO LIABILITY WHATSOEVER FOR DIRECT, INDIRECT, INCIDENTAL, CONSEQUENTIAL, SPECIAL, GENERAL OR OTHER DAMAGES.
Some states do not allow the exclusion of implied warranties or limitations on how long an implied warranty lasts, so the above limitation may not apply to you. Similarly, some states do not allow the exclusion of incidental or consequential damages, so the above limitation or exclusion may not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.
*Warranty applies to material defects in materials & workmanship only.
WATER FILTRATION
EVERPURE-SHURF LO WO RLD HEADQUARTERS, 1040 MUIRFIELD DRIVE, HANOVER PARK, IL 60133 USA • FOODSERVICE.PENTAIR.COM
800.942.1153 (US ONLY) • 630.307.3000 MAIN • 630.307.3030 FAX • CSEVERPURE@PENTAIR.COM EMAIL EVERPURE-SHURF LO AUSTRALIA, 1-21 MONASH DRIVE, DANDENONG SOUTH, VIC 3175, AUSTRALIA
011.1300 576 190 TEL • 011.61.39.562.7237 FAX • AU.EVERPURE@PENTAIR.COM EMAIL EVERPURE-SHURF LO CHINA, 21F CLOUD 9 PLAZA, NO 1118, SHANGHA I, 200052, CHINA
86.21.3211.4588 TEL • 86.21.3211.4580 FAX • CHINA.WATER@PENTAIR.COM EMAIL EVERPURE-SHURF LO INDIA, GREEN BOULEVAR D, B-9/A, 7TH FLOOR - TOWE R B SECTOR 62, NOIDA - 201301
91.120.419.9444 TEL • 91.120.419.9400 FAX • INDIACUSTOMER@PENTAIR.COM EMAIL EVERPURE-SHURF LO EUROPE, PENTAIR WATER BELGIUM BV BA, INDUSTRIEPARK WOLFSTEE, TOEKOMSTLAAN 30, B-2200 HERENTALS, BELGIUM
+32.(0).14.283.504 TEL • +32.(0).14.283.505 FAX • SALES@EVERPURE-EUROPE.COM EMAIL EVERPURE-SHURF LO JA PAN I NC., HASHIMOTO MN BLDG. 7F, 3-25-1 HASHIMOTO, MIDORI-KU, SAGAMIHARA-SHI KANAGAWA 252-0143, JAPAN
81.(0)42.775.3011 TEL • 81.(0)42.775.3015 FAX • INFO@EVERPURE.CO.JP EMAIL EVERPURE-SHURF LO SOUTHEAST ASIA, 390 HAVELOCK ROAD, #04-01 KING’S CENTRE, SINGAPORE 169662
65.6768.5800 TEL • 65.6737.5149 FAX • CSEVERPURE@PENTAIR.COM EMAIL
Pentair trademarks and logos are owned by Pentair plc or its affiliates. Third party registered and unregistered trademarks and logos are property of their respective owners. Because we are continuously improving our products and services, Pentair reserves the right to change specifications without prior notice. Pentair is an equal opportunity employer.
© 2018 Pentair Filtration Solutions, LLC. All Rights Reserved.
EV9975-20 Rev A JA18
Loading...