PendoTECH PressureChecker User Guide
Warning potential shock hazard. Do not submerge this product.
Protect the product before cleaning with any liquids by covering
openings that expose the internal components.
Revision 1
PendoTECH PressureChecker™ User Guide
Revision 1
Copyright © 2009-2014, PendoTECH
All rights reserved. No part of this publication may be reproduced, stored in an electronic
retrieval system, or transmitted, in any form or by any means, whether electronic,
mechanical, by photocopying, or otherwise, without the written consent of PendoTECH.
The information in this User Guide is believed to be accurate and reliable for use and
operation of the monitor, however, PendoTECH assumes no responsibility for the use of
this product except for what is covered in the Limited Warranty and Terms and Condition
of Sale.
WARNING: “WARNING” is used to indicate the presence of a hazard which can
cause severe personal injury, death, or substantial property damage if the warning is
ignored
NOTE: “NOTE” is used to notify the user of installation or operation information
which is important but not hazard related.
Goods and software are not designed, intended or authorized for use as components
in life support or medical devices. They are not designed for any application in which
the failure of the product could result in personal injury, death or property damage.
Page 2/15

PendoTECH PressureChecker User Guide
Revision 1
Table of Contents
1. Overview of the PressureChecker ........................................................................................... 4
1.1. Key Features ................................................................................................................................. 5
1.2. Device Details ............................................................................................................................... 5
1.2.1. Hardware Details ...................................................................................................................... 5
1.2.2. Specifications ............................................................................................................................ 6
1.2.3. Theory of Operation .................................................................................................................. 7
2. Using the Device ..................................................................................................................... 8
2.1. Testing sensors non-invasively ..................................................................................................... 8
2.2. Using as a monitor simulator to test a monitor ........................................................................... 11
3. Key System Components ...................................................................................................... 13
3.1. Power Supply and Low Battery Indicator ................................................................................... 13
3.2. Precision Pressure Transducer .................................................................................................... 13
3.3. Display ........................................................................................................................................ 13
3.4. Pressure Generating Cylinder ..................................................................................................... 13
APPENDIX A: PRODUCT WARRANTY ................................................................................. 14
PENDOTECH LIMITED WARRANTY ..................................................................................... 14
APPENDIX B: Service Information ............................................................................................ 15
Page 3/15

PendoTECH PressureChecker User Guide
Revision 1
1. Overview of the PressureChecker
The PressureChecker™ is a calibrated device that can be used to verify proper
functioning and accuracy of PendoTECH Single Use Pressure Sensors and monitors
non-invasively. It can be used to test the sensors in-situ without the requirement to
access the fluid path. In its small, hand-held enclosure are all the devices necessary to
perform the required function. PendoTECH Pressure Sensors are 100% tested at
several points in the manufacturing process, however, this device gives the ability to
verify the performance of the sensor by the end-user. PendoTECH Pressure Sensors
are also frequently used on single use bioreactors to measure pressure in the bag. The
required resolution is often less than 1 psi and the PressureChecker has the readability
and ability to regulate pressure in this range. It also has an internal pressure sensor
simulator and can be used to verify proper functioning of monitors such as the
PressureMAT™, PendoTECH Process Control Systems, or 3rd party monitors.
There are numerous applications in biopharmaceutical production processes where the
device can be used.
System Components Supplied:
PDKT-650-950 PressureChecker pressure sensor and monitor verification tool
(with 9V alkaline battery)
PDKT-650-905 4 foot connecting tube with luer slip
Calibration Certificate
Page 4/15

PendoTECH PressureChecker User Guide
Monitor
Connector
Pneumatic
Port
ON/OFF
Switch
Adjustment
Knob
Zero
Adjust
LCD
Monitor
connector
Pneumatic Port
with cap in place
Revision 1
1.1. Key Features
The PressureChecker incorporates a specially designed pressure cylinder and precision solid-state
transducer for generating and measuring static pressure from -7 to +10.5 psi. The adjustment knob is used
to regulate the pressure, and the pressure value is shown on the liquid crystal display. A built-in
pneumatic port provides the external connection for fittings to connect to the sensor test port.
Verify proper functioning of sensor and monitors
Verify proper function of sensor in-situ by accessing the atmospheric reference side of sensor
NIST Traceable
Can be returned for annual calibration
Battery operated for portability
1.2. Device Details
1.2.1. Hardware Details
Device details are as shown:
The connections on the left side panel are as shown below:
Page 5/15