PCE Instruments PCE-889 Operating Manual

Infrared Thermometer
PCE-889
Operating Manual
PCE Americas Inc. 711 Commerce Way Suite 8 Jupiter FL-33458 USA From outside US: +1 Tel: (561) 320-9162 Fax: (561) 320-9176 info@pce-americas.com
www.pce-instruments.com/english
www.pce-instruments.com
PCE Instruments UK Ltd.
Units 12/13
Southpoint Business Park
Ensign way
Hampshire / Southampton
United Kingdom, SO31 4RF
From outside UK: +44
Tel: (0) 2380 98703 0
Fax: (0) 2380 98703 9
info@pce-instruments.com
TABLE OF CONTENTS
INTRODUCTION…………………………………………………………………………. FEATURES……………………………………………………………………………….
WIDE RANGE APPLICATION
…………………………………………………………….
SAFETY
…………………………………………………………………………………..
DISTANCE & SPOT SIZE
…………………………………………………………………
SPECIFICATIONS
…………………………………………………………………………
FRONT PANEL DESCRIPTION
……………………………………………………………
INDICATOR
………………………………………………………………………………..
BUTTONS………………………………………………………………………………… MEASURMENT OPERATION………………………………………………………………
BATTERY REPLACEMENT
………………………………………………………………..
NOTES
…………………………………………………………………………………….
MAINTENANCE
…………………………………………………………………………..
INTRODUCTION
Thank you for purchase of the IR Thermometer. This is capable of non-contact (infrared) temperature measurements at the touch of a button. The built-in laser pointer increases target accuracy while the backlight LCD and handy push-buttons combine for convenient, ergonomic operation. The Non-contact Infrared Thermometers can be used to measure the temperature of objects’ surface that is improper to be measured by traditional (contact) thermometer (such as moving object, the surface with electricity current or the objects which are uneasy to be touched.) Proper use and care of this meter will provide years of reliable service.
FEATURES:
z Precise non-contact measurements z High distance to target ratio measures smaller surface areas at greater distances z Widest temperature range z Unique flat surface, modern housing design z Built-in laser pointer z Automatic Data Hold z / switch z Emissivity Digitally adjustable from 0.10 to 1.0 z MAX,MIN,DIF, AVG temperature displays z Backlight LCD display z Built-in laser pointer z Automatic selection range and Display Resolution 0.1ºC(0.1ºF) z Trigger lock z Set high and low alarms
WIDE RANGE APPLICATION:
Food preparation, Safety and Fire inspectors, Plastic molding, Asphalt, Marine and Screen printing, measure ink and Dryer temperature, HVAC/R, Diesel and Fleet maintenance.
1. SAFETY
z Use extreme caution when the laser beam is turned on. z Do not let the beam enter your eye, another person’s eye or the eye of an animal. z Be careful no to let the beam on a reflective surface strike your eye. z Do not allow the laser light beam impinge on any gas which can explode.
Distance & Spot Size
As the distance (D) from the object increases, the spot size (S) of the area measured by the unit becomes larger. The relationship between distance and spot size for each unit is listed below. The focal point for each unit is 914mm (36”). The spot sizes indicate 90% encircled energy.
Fig: 1
2. SPECIFICATIONS
Temperature range D: S
-50 to 1000 oC (-58 to 1832) 50:1
Display resolution 0.1
o
C(0.1)
Accuracy for targets:
Assumes ambient operating temperature of 23 to 25 oC (73 to 77)
-50 to -20℃(-58 to -4℉): ±5
o
C(
±9℉)
-20 to 200(-4 to 392℉):
± 1.5% of reading + 2℃(±3.6);
200 to 538 (392 to 1000℉):
± 2.0% of reading +2(±3.6);
538 to 1000(1000 to 1832):
± 3.5% of reading ± 5(±9)
Response time less than 1 second Spectral response 8~14um Emissivity Digitally adjustable from 0.10 to 1.0 Over range indication LCD will show “-0L”,”0L” Polarity Automatic (no indication for positive polarity);
Minus (-) sign for negative polarity Diode laser output <1mW,Wavelength 630~670nm, Class 2(II) laser product Operating temp. 0 to 50℃(32 to 122℉) Storage temp. –20 to 60 (-4 to 140℉) Relative humidity 10%~90%RH operating, <80%RH storage Power supply 9V battery, NEDA 1604A or IEC 6LR61, or equivalent Weight 290g (10.2 oz.) Size 100 x 56 x 230mm (3.9 x 2.2 x 9.0”) Safety “ CE ” Comply with EMC
Note:
z Accuracy: Given at 18 to 28 oC (64
to 82 ℉), less than 80 % RH.
z Field of View: Make sure that the target is larger than the unit’s spot size. The smaller the target,
the closer you should be to it. When accuracy is critical, make sure the target is at least twice as large as the spot size.
3. FRONT PANEL DESCRIPTION
IR sensor Laser pointer beam LCD Display down button up button mode button laser/backlight button Measurement TriggerHandle Grip Battery Cover
4. INDICATOR
Data hold Measuring indicationEmissivity symbol and valueLow battery, lock and laser” on” symbols Backlight “on” symbol ⑥ ℃/ symbolHigh alarm and low alarm symbolTemperature values for the MAX, MIN, DIF,
AVG HAL and LAL.
Symbols for EMS MAX, MIN, DIF, AVG,
HAL and LAL.
Current temperature value
5. Buttons
Up button (for EMS,HAL,LAL)
MODE button
(for cycling through the mode loop)
Down button (for EMS,HAL,LAL)
Laser/Backlight on/off button
(pull trigger and press button to
activate laser/backlight)
MOD E
MODE Button Function
The infrared thermometer measures Maximum (MAX), Minimum
(MIN), Differential (DIF), and Average (AVG) Temp. Each time you
take a reading. This data is stored and can be recalled with the
MODE button until a new measurement is taken. When the trigger is
pulled again, the unit will begin measuring in the last mode
selected.
Pressing the MODE button also allows you to access the High
MODE Alarm(HAL), Low Alarm(LAL),Emissivity(EMS).Each time you press
MODE ,you advance through the mode cycle.
The diagram shows the sequence of functions in the MODE cycle.
Switching C/F ,LOCK ON/OFF and SET ALARM.
C/F LOCK ON/OFFSET ALARM
Select the temperature units (oC or oF) using the oC/oF switch
To lock the unit on for continuous measurement, slide the middle switch LOCK ON/OFF right. If the trigger is pulled while the unit is lock on, the laser and backlight will turn on if they have been activated. When the unit is locked on, the backlight and laser will remain on unless it is turned off using the Laser/Backlight button on the keypad.
To activate the alarms, please slide the bottom switch SET ALARM right.
To set values for the High Alarm (HAL), Low Alarm (LAL) and Emissivity (EMS), firstly active the display by pulling the trigger or press the MODE button, then press the MODE button until the appropriate code appears in the lower left corner of the display, press the UP and down buttons to adjust the desired values.
6. MEASURMENT OPERATION
Hold the meter by its Handle Grip and point it toward the surface to be measured. Pull and hold the Trigger to turn the meter on and begin testing. The display will light if the
battery is good. Replace the battery if the display does not light.
While measuring, the SCAN display icon will appear ins the upper left hand corner of the LCD.
MAX
MIN
DIF
AVG
HAL
LAL
EMS
Rele ase the Trigger and the HOLD display icon will appear on the LCD indicating that the
reading is being held.
The meter will automatically power down after approximately 7 seconds after the trigger is
released.(Unless the unit is locked on)
Note: Measurement considerations
Holding the meter by its handle, point the IR Sensor toward the object whose temperature is to be measured. The meter automatically compensates for temperature deviations from ambient temperature. Keep in mind that it will take up to 30 minutes to adjust to wide ambient temperatures are to be measured followed by high temperature measurements, some time (several minutes) is required after the low (and before the high) temperature measurements are made. This is a result of the cooling process, which must take place for the IR sensor.
7. BATTERY REPLACEMENT
As battery power is not sufficient, LCD will display “ ” replacement with one new battery type 9V is required.
Open battery cover, then take out the
battery from instrument and replace with a new 9-Volt battery and place the battery cover back.
8. NOTES:
z How it Works
Infrared thermometers measure the surface temperature of an object. The unit’s optics sense emitted, reflected, and transmitted energy, which is collected and focused onto a detector. The unit’s electronics translate the information into a temperature reading, which is display on the unit. In units with a laser, the laser is used for aiming purposes only.
z Field of View
Make sure that the target is larger than the unit’s spot size. The smaller the target, the closer you should be to it. When accuracy is critical, make sure the target is at least twice as large as the spot size.
z Distance & Spot Size
As the distance (D) from the object increases, the spot size (S) of the area measured by the unit becomes larger. See: Fig: 1.
z Locating a hot Spot
To find a hot spot aim the thermometer outside the area of interest, then scan across with an up and down motion until you locate hot spot.
z Reminders
Not recommended for use in measuring shiny or polished metal surfaces ( stainless steel,
aluminum, etc.).See Emissivity
The unit cannot measure through transparent surfaces such as glass. It will measure the surfac e
temperature of the glass instead. Steam, dust, smoke, etc., Can prevent accurate measurement by obstructing the unit’s optics. z Emissivity
Emissivity is a term used to describe the energy-emitting characteristics of materials. Most (90% of typical applications) organic materials and painted or oxidized surfaces have an emissivity of 0.95 (pre-set in the unit). Inaccurate readings will result from measuring shiny or polished metal surfaces. To compensate, cove the surface to be measured with masking tape or flat black paint. Allow time for the tape to reach the same temperature as the material underneath it. Measure the temperature of the tape or painted surface.
Emissivity Values
Substance Thermal emissivity Substance Thermal emissivity Asphalt 0.90 to 0.98 Cloth (black) 0.98 Concrete 0.94 Human skin 0.98 Cement 0.96 Lather 0.75 to 0.80 Sand 0.90 Charcoal (powder) 0.96
Earth 0.92 to 0.96 Lacquer 0.80 to 0.95 Water 0.92 to 0.96 Lacquer (matt) 0.97 Ice 0.96 to 0.98 Rubber (black) 0.94 Snow 0.83 Plastic 0.85 to 0.95 Glass 0.90 to 0.95 Timber 0.90 Ceramic 0.90 to 0.94 Paper 0.70 to 0.94 Marble 0.94 Chromium oxides 0.81 Plaster 0.80 to 0.90 Copper oxides 0.78 Mortar 0.89 to 0.91 Iron oxides 0.78 to 0.82 Brick 0.93 to 0.96 Textiles 0.90
9. MAINTENANCE :
z Repairs or service are not covered in this manual and s hould only be carried out by qualified
trained technician.
z Periodically ,wipe the body with a dry cloth. Do not use abrasives or solvents on this instrument.
z For service , use only manufacturer’s specified parts.