Instruction Manual and Experiment Guide for the PASCO scientific

Model SE-9634, 9635, and 9636

012-04273D 10/95

ESR APPARATUS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

© 1990 PASCO scientific

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

$7.50

ESR Apparatus

012-04273D

 

 

The lightning flash with arrowhead, withinanequilateraltriangle,isintended to alert the user of the presence of uninsulated “dangerous voltage” within the product’s enclosure that may be of sufficient magnitude to constitute a risk of electric shock to persons.

CAUTION

RISK OF ELECTRIC SHOCK

DO NOT OPEN

CAUTION:

TO PREVENT THE RISK OF ELECTRIC SHOCK, DO NOT REMOVE BACK COVER. NO USER SERVICEABLE PARTS INSIDE. REFER SERVICING TO QUALIFIED SERVICE PERSONNEL.

The exclamation point within an equilateral triangle is intended to alert the user of the presence of important operating and maintenance (servicing) instructions in the literature accompanying the appliance.

®

2

012-04273D

ESR Apparatus

 

 

Table of Contents

Section

Page

Copyright, Warranty, and Equipment Return

.................................................. ii

Introduction .....................................................................................................

1

Equipment ......................................................................................................

2-4

Setup with the Complete ESR Equipment .....................................................

5-6

Setup with the ESR Basic Equipment ..........................................................

7-8

Making ESR Measurements ............................................................................

9

Appendix ........................................................................................................

10

®

i3

 

ESR Apparatus

012-04273D

 

 

Copyright, Warranty and Equipment Return

Please— Feel free to duplicate this manual subject to the copyright restrictions below.

Copyright Notice

The PASCO scientific Model SE-9634, SE-9635, and SE-9636 Complete ESR System manual is copyrighted and all rights reserved. However, permission is granted to non-profit educational institutions for reproduction of any part of this manual providing the reproductions are used only for their laboratories and are not sold for profit. Reproduction under any other circumstances, without the written consent of PASCO scientific, is prohibited.

Limited Warranty

PASCO scientific warrants this product to be free from defects in materials and workmanship for a period of one year from the date of shipment to the customer. PASCO will repair or replace, at its option, any part of the product which is deemed to be defective in material or workmanship. This warranty does not cover damage to the product caused by abuse or improper use. Determination of whether a product failure is the result of a manufacturing defect or improper use by the customer shall be made solely by PASCO scientific. Responsibility for the return of equipment for warranty repair belongs to the customer. Equipment must be properly packed to prevent damage and shipped postage or freight prepaid. (Damage caused by improper packing of the equipment for return shipment will not be covered by the warranty.) Shipping costs for returning the equipment, after repair, will be paid by PASCO scientific.

Credits

This manual edited by: Dave Griffith

Equipment Return

Should the product have to be returned to PASCO scientific for any reason, notify PASCO scientific by letter, phone, or fax BEFORE returning the product.

Upon notification, the return authorization and shipping instructions will be promptly issued.

ä NOTE: NO EQUIPMENT WILL BE ACCEPTED FOR RETURN WITHOUT AN AUTHORIZATION FROM PASCO.

When returning equipment for repair, the units must be packed properly. Carriers will not accept responsibility for damage caused by improper packing. To be certain the unit will not be damaged in shipment, observe the following rules:

The packing carton must be strong enough for the item shipped.

Make certain there are at least two inches of packing material between any point on the apparatus and the inside walls of the carton.

Make certain that the packing material cannot shift in the box or become compressed, allowing the instrument come in contact with the packing carton.

Address:

PASCO scientific

 

10101 Foothills Blvd.

 

Roseville, CA 95747-7100

Phone:

(916) 786-3800

FAX:

(916) 786-3292

email:

techsupp@pasco.com

web:

www.pasco.com

ii

®

4

 

Figure 2 ESR on the Oscilloscope

012-04273D

ESR Apparatus

 

 

Introduction

ESR in Theory

 

 

 

 

 

 

 

 

Test Sample

 

 

 

The basic setup for electron spin

 

 

 

 

B

 

 

 

 

 

 

 

 

 

 

 

 

 

 

resonance is shown in Figure 1.

 

 

 

 

 

 

 

 

 

 

 

A test sample is placed in a

 

 

 

 

 

 

 

 

 

 

 

 

 

 

uniform magnetic field. The

 

 

 

 

 

 

 

 

 

 

 

 

 

sample is also wrapped within a

 

 

 

 

 

 

 

 

 

 

 

coil that is connected to an RF

 

 

 

 

 

 

 

 

 

 

 

 

 

oscillator. The smaller magnetic

 

 

 

 

 

 

 

 

 

 

 

field induced in the coil by the

 

 

 

 

 

 

RF

 

 

 

 

 

oscillations of the oscillator is at

 

 

 

 

 

 

 

 

 

 

 

 

Oscillator

 

 

 

 

 

right angles to the uniform

 

 

 

 

 

 

 

 

 

 

 

 

 

 

magnetic field.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

i

 

 

 

Consider, for the moment, a

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

single electron within the test

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

sample. The electron has a

 

 

 

 

 

 

 

 

 

 

 

 

 

 

magnetic dipole moment (μs) that is related to its intrinsic

 

 

angular momentum, or spin, by the vector equation:

 

 

 

 

 

 

 

 

 

 

 

 

 

Ammeter

 

 

 

 

 

 

 

 

 

 

Figure 1 ESR Diagram

 

 

μ

S

=g

S

μ

S

; (equation 1)

 

 

where:

 

 

Bh

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

gs =a constant characteristic of the electron, the g-factor

 

 

μ =the Bohr magneton =

 

eh

=5.788 x 10–9 eV

 

 

 

2me

B

 

 

 

 

 

 

 

 

 

 

G

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

s =the spin of the electron

 

 

 

 

 

 

 

 

 

 

h =Planck's constant = 6.582 x 10–16eV-sec.

 

 

The magnetic dipole moment of this electron interacts with

 

 

the uniform magnetic field. Due to its quantum nature, the

 

 

electron can orient itself in one of only two ways, with

 

 

energies equal to E0 ± gsμB/2; where E0 is the energy of the

 

 

electron before the magnetic field was applied. The energy

 

 

difference between these two possible orientations is equal to

 

 

gsμBB; where B is the magnitude of the magnetic field.

 

 

Resonance occurs when the RF oscillator is tuned to a

 

 

frequency ν, such that the energy of the irradiated photons,

 

 

hν, is equal to the difference between the two possible energy

 

 

states of the electron. Electrons in the lower energy state can

 

 

then absorb a photon and jump to the higher energy state.

 

 

This absorption of energy effects the permeability of the test

 

 

sample, which effects the inductance of the coil and thereby

 

 

the oscillations of the RF oscillator. The result is an observ-

 

 

able change in the current flowing through the oscillator.

 

 

The condition for resonance, therefore, is that the energy of

 

 

the photons emitted by the oscillator match the energy

 

 

difference between the spin states of the electrons in the test

 

 

sample. Stated mathematically:

 

 

 

 

 

 

 

 

 

 

®

 

 

 

hν=gSμBB

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

ESR in Practice

For an electron with only two energy states, in a magnetic field of a given magnitude, it would be necessary to set the RF frequency with considerable accuracy in order to observe resonance. In practice, this difficulty is solved by varying the magnitude of the magnetic field about some constant value. With the PASCO ESR Apparatus, this is done by supplying a small AC current, superimposed on a larger DC current, to a pair of Helmholtz coils. The result is a magnetic field that varies sinusoidally about a constant value.

If the RF frequency is such that equation 2 is satisfied at some point between the minimum and maximum values of the sinusoidally varying magnetic field, then resonance will occur twice during each cycle of the field. Resonance is normally observed using a dual trace oscilloscope. The oscilloscope traces, during resonance, appear as in Figure 2. The upper trace is a measure of the current going to the Helmholtz coils, which

is proportional to the magnetic field. The lower trace shows the envelope of the voltage across the RF oscillator, which dips sharply each time the magnetic field passes through the resonance point.

ESR in Research

In research, ESR measurements are considerably more complicated than equation 2 would indicate. The electrons and protons in an atom or molecule form a complicated electromagnetic environment, which is affected by the externally applied magnetic field. The various energy splittings and shifts that show up in ESR measurements can therefore provide sensitive information about the internal structure of the atoms and molecules.

The test sample included with the PASCO ESR Apparatus, DPPH*, is a particularly simple substance for ESR measurements. It has an orbital angular momentum of zero, and only one unpaired electron. Therefore, for a given value of the external magnetic field, it has only a single resonant frequency. This makes it possible to investigate some of the basic principles of electron spin resonance, without (or before) getting into the more complex world of ESR analysis.

(* Diphenyl-Picryl-Hydrazyl)

Pasco scientific SE-9636 User Manual

ESR Apparatus

012-04273D

 

 

The ESR Equipment

Included Equipment

The ESR Apparatus is available in three separate packages (see Figure 3):

• The ESR Probe Unit (SE-9634)includes:

The Probe Unit with base-

Three RF Probes and a DPPH sample in a vial The Passive Resonant Circuit

The Current Measuring Lead for the Probe Unit

The ESR Basic System (SE-9635)includes: The ESR Probe Unit (SE-9634)

A pair of Helmholtz Coils with bases The ESR Adapter (SE-9637)

The Complete ESR System (SE-9636)includes:

The ESR Probe Unit (SE-9634)

A pair of Helmholtz Coils with bases

The Control Unit

Additional Equipment Needed:

If you're using the SE-9636 Complete ESR System, you'll need the following additional equipment:

a DC ammeter capable of measuring up to 3 A (such as PASCO's SB-9599)

a dual trace oscilloscope (such as PASCO's SE-9533)

connecting wires with banana plug connectors

If you're using the SE-9635 Basic ESR System, you'll need the following additional equipment:

Frequency Meter(to 130 kHz) (such as PASCO's SB-9599 Universal Multimeter)

SF-9584 Low Voltage AC/DC Power Supply

(or an equivalent supply providing a 10 volt, 3 amp DC output and a 0 - 4 volt, 1 amp AC output)

Power Supply providing ± 12 VDC

DC Ammeter(0 - 3 amp) (PASCO Model SB-9599)

 

 

Helmholtz Coils

 

 

 

 

 

 

 

with bases

 

 

 

 

 

 

 

 

 

 

f

ESR Adapter (SE-9637)—included

RF Probes (3):

 

 

 

1000

 

+12V

 

 

 

 

0

 

V

in the ESR Basic System, but not in

 

 

 

 

the Complete ESR System

 

 

–12V

 

 

 

 

 

 

 

 

 

 

 

 

13 - 30 MHz

Probe Unit

 

 

 

 

 

 

 

 

 

 

with base

 

 

 

 

 

 

 

 

 

30 - 75 MHz

 

 

 

 

 

 

 

 

 

 

75 - 130 MHz

 

 

 

 

 

 

 

 

 

 

DPPH sample

 

 

 

 

 

Control Unit

 

 

 

 

 

 

 

 

B

 

 

 

 

 

 

 

 

 

1

 

2

 

 

 

 

 

 

 

 

1

 

2

f/MHz

 

 

 

 

 

 

 

 

 

 

 

 

Passive

 

 

0

10V

0

5V

0

ϕ

90°

 

 

 

 

U0

 

Umod

 

Y

 

Resonant

 

 

 

 

 

 

 

 

 

 

1

 

 

 

 

 

ϕ

 

 

 

 

 

 

 

 

 

X

 

Circuit

 

2

I max3A

 

 

 

 

 

 

0…10V–

 

 

 

0…90°

 

 

 

 

 

 

 

 

 

 

 

 

 

Umod

0…5V

 

 

 

 

 

 

 

 

 

 

Current Measuring Lead for

 

 

 

 

 

 

 

 

 

the Probe Unit

 

 

 

 

 

 

 

 

 

 

ESR Probe Unit (SE-9634)

 

 

 

 

 

 

 

 

 

 

ESR Basic System (SE-9635)

 

 

 

 

 

 

 

 

 

 

Complete ESR System (SE-9636)

 

 

 

Figure 3 The ESR Apparatus

®