The lightning flash with arrowhead,
within an equilateral triangle, is intended
to alert the user of the presence of
uninsulated “dangerous voltage” within
the product’s enclosure that may be of
sufficient magnitude to constitute a risk
of electric shock to persons.
CAUTION
RISK OF ELECTRIC SHOCK
DO NOT OPEN
CAUTION:
TO PREVENT THE RISK OF
ELECTRIC SHOCK, DO NOT
REMOVE BACK COVER. NO USER
SERVICEABLE PARTS INSIDE.
REFER SERVICING TO QUALIFIED
SERVICE PERSONNEL.
The exclamation point within an equilateral triangle is intended to alert the
user of the presence of important
operating and maintenance (servicing) instructions in the literature accompanying the appliance.
2
®
012-04273DESR Apparatus
T able of Contents
SectionPage
Copyright, Warranty, and Equipment Return..................................................ii
Please—Feel free to duplicate this manual
subject to the copyright restrictions below.
Copyright Notice
The PASCO scientific Model SE-9634, SE-9635, and
SE-9636 Complete ESR System manual is copyrighted
and all rights reserved. However, permission is granted
to non-profit educational institutions for reproduction of
any part of this manual providing the reproductions are
used only for their laboratories and are not sold for profit.
Reproduction under any other circumstances, without the
written consent of PASCO scientific, is prohibited.
Limited Warranty
PASCO scientific warrants this product to be free from
defects in materials and workmanship for a period of one
year from the date of shipment to the customer. PASCO
will repair or replace, at its option, any part of the product
which is deemed to be defective in material or workmanship. This warranty does not cover damage to the product
caused by abuse or improper use. Determination of
whether a product failure is the result of a manufacturing
defect or improper use by the customer shall be made
solely by PASCO scientific. Responsibility for the return
of equipment for warranty repair belongs to the customer.
Equipment must be properly packed to prevent damage
and shipped postage or freight prepaid. (Damage caused
by improper packing of the equipment for return shipment will not be covered by the warranty.) Shipping
costs for returning the equipment, after repair, will be
paid by PASCO scientific.
Equipment Return
Should the product have to be returned to PASCO
scientific for any reason, notify PASCO scientific by
letter, phone, or fax BEFORE returning the product.
Upon notification, the return authorization and
shipping instructions will be promptly issued.
ä
NOTE: NO EQUIPMENT WILL BE
ACCEPTED FOR RETURN WITHOUT AN
AUTHORIZATION FROM PASCO.
When returning equipment for repair, the units
must be packed properly. Carriers will not accept
responsibility for damage caused by improper
packing. To be certain the unit will not be
damaged in shipment, observe the following rules:
➀ The packing carton must be strong enough for the
item shipped.
➁ Make certain there are at least two inches of
packing material between any point on the
apparatus and the inside walls of the carton.
➂ Make certain that the packing material cannot shift
in the box or become compressed, allowing the
instrument come in contact with the packing
carton.
resonance is shown in Figure 1.
A test sample is placed in a
uniform magnetic field. The
sample is also wrapped within a
coil that is connected to an RF
oscillator. The smaller magnetic
field induced in the coil by the
oscillations of the oscillator is at
RF
Oscillator
right angles to the uniform
magnetic field.
Consider, for the moment, a
single electron within the test
Figure 1 ESR Diagram
i
Ammeter
sample. The electron has a
magnetic dipole
moment (µ
) that is related to its intrinsic
s
angular momentum, or spin, by the vector equation:
µS=gSµ
where:
=a constant characteristic of the electron, the g-factor
g
s
µB=the Bohr magneton =
;(equation 1)
B
h
eh
=5.788 x 10
2m
e
–9
eV
G
S
s=the spin of the electron
=Planck's constant = 6.582 x 10
h
–16
eV-sec.
The magnetic dipole moment of this electron interacts with
the uniform magnetic field. Due to its quantum nature, the
electron can orient itself in one of only two ways, with
energies equal to E0 ± gsµB/2; where E0 is the energy of the
electron before the magnetic field was applied. The energy
difference between these two possible orientations is equal to
gsµBB; where B is the magnitude of the magnetic field.
Resonance occurs when the RF oscillator is tuned to a
frequency ν, such that the energy of the irradiated photons,
hν, is equal to the difference between the two possible energy
states of the electron. Electrons in the lower energy state can
then absorb a photon and jump to the higher energy state.
This absorption of energy effects the permeability of the test
sample, which effects the inductance of the coil and thereby
the oscillations of the RF oscillator. The result is an observable change in the current flowing through the oscillator.
The condition for resonance, therefore, is that the energy of
the photons emitted by the oscillator match the energy
difference between the spin states of the electrons in the test
sample. Stated mathematically:
hν=gSµBB
®
ESR in Practice
For an electron with only two energy states, in a magnetic
field of a given magnitude, it would be necessary to set the
RF frequency with considerable accuracy in order to observe
resonance. In practice, this difficulty is solved by varying the
magnitude of the magnetic field about some constant value.
With the PASCO ESR Apparatus, this is done by supplying
a small AC current, superimposed on a larger DC current, to
a pair of Helmholtz coils. The result is a magnetic field that
varies sinusoidally about a constant value.
If the RF frequency is such that equation 2 is satisfied at
some point between the minimum and maximum values of
the sinusoidally varying magnetic field, then resonance will
occur twice during each cycle of the field. Resonance is
normally observed using a dual trace oscilloscope. The
oscilloscope traces, during resonance, appear as in Figure 2.
The upper trace is a measure of the current going to the
Helmholtz coils, which
is proportional to the
magnetic field. The
lower trace shows the
envelope of the voltage
across the RF oscillator,
which dips sharply each
time the magnetic field
passes through the
Figure 2 ESR on the
Oscilloscope
resonance point.
ESR in Research
In research, ESR measurements are considerably more
complicated than equation 2 would indicate. The electrons
and protons in an atom or molecule form a complicated
electromagnetic environment, which is affected by the
externally applied magnetic field. The various energy
splittings and shifts that show up in ESR measurements can
therefore provide sensitive information about the internal
structure of the atoms and molecules.
The test sample included with the PASCO ESR Apparatus,
DPPH*, is a particularly simple substance for ESR measurements. It has an orbital angular momentum of zero, and only
one unpaired electron. Therefore, for a given value of the
external magnetic field, it has only a single resonant frequency. This makes it possible to investigate some of the
basic principles of electron spin resonance, without (or
before) getting into the more complex world of ESR
analysis.
(* Diphenyl-Picryl-Hydrazyl)
1
ESR Apparatus012-04273D
The ESR Equipment
Included Equipment
The ESR Apparatus is available in three separate packages (see Figure 3):
•The ESR Probe Unit (SE-9634) includes:
The Probe Unit with baseThree RF Probes and a DPPH sample in a vial
The Passive Resonant Circuit
The Current Measuring Lead for the Probe Unit
•The ESR Basic System (SE-9635) includes:
The ESR Probe Unit (SE-9634)
A pair of Helmholtz Coils with bases
The ESR Adapter (SE-9637)
•The Complete ESR System (SE-9636) includes:
The ESR Probe Unit (SE-9634)
A pair of Helmholtz Coils with bases
The Control Unit
RF Probes (3):
13 - 30 MHz
Probe Unit
with base
Additional Equipment Needed:
If you're using the SE-9636 Complete ESR System,
you'll need the following additional equipment:
•a DC ammeter capable of measuring up to 3 A
(such as PASCO's SB-9599)
•a dual trace oscilloscope (such as PASCO's SE-9533)
•connecting wires with banana plug connectors
If you're using the SE-9635 Basic ESR System, you'll
need the following additional equipment:
•Frequency Meter (to 130 kHz) (such as PASCO's
SB-9599 Universal Multimeter)
•SF-9584 Low Voltage AC/DC Power Supply
(or an equivalent supply providing a 10 volt, 3 amp
DC output and a 0 - 4 volt, 1 amp AC output)
•Power Supply providing ± 12 VDC
•DC Ammeter (0 - 3 amp) (PASCO Model SB-9599)
Helmholtz Coils
with bases
f
1000
+12V
0
–12V
ESR Adapter (SE-9637)—included
in the ESR Basic System, but not in
V
the Complete ESR System
30 - 75 MHz
75 - 130 MHz
DPPH sample
Current Measuring Lead for
the Probe Unit
ESR Probe Unit (SE-9634)
Passive
Resonant
Circuit
ESR Basic System (SE-9635)
Complete ESR System (SE-9636)
Figure 3 The ESR Apparatus
2
Control Unit
B
1
2
1
010V
1
2
I 3A
max
0…10V–
U
0
05V
f/MHz
2
090°
U
mod
U
mod
ϕ
Y
X
ϕ
0…90°
0…5V
®
Loading...
+ 12 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.