Command Response Description
aM2! atttn<CR><LF>
and
a<CR><LF> after 1 second
Beam 1: Read signal amplitudes of the last measurement
a – Sensor address
ttt – Time in seconds until the sensor has
provided the result.
OTT SLD response = 001
n – Number of measured values
OTT SLD response = 9
a<CR><LF> – Service request
aD0!
aD1!
aD2!
a<value1><value2><value3><CR><LF>
a<value4><value5><value6><CR><LF>
a<value7><value8><value9><CR><LF>
Send data
a – Sensor address
<valuex> – Beam 1: Signal amplitude
Measuring cell 1 … 9 [counts]
Measured value format: pbbb
Range: +0 … +255 counts
aM3! atttn<CR><LF>
and
a<CR><LF> after 1 second
Start measurement: Pitch, roll, pressure
a – Sensor address
ttt – Time in seconds until the sensor has
determined the measurement result.
OTT SLD response = 001
n – Number of measured values
OTT SLD response = 3
a<CR><LF> – Service request
aD0! a<value1><value2><value3>…
…<CR><LF>
Send data
<value1> – Pitch (Position of the instrument in transverse axis)
[0.1°]
Measured value format: pbb.ee
Range: –25.0 … +25.0°
<value2> – Roll (Position of the instrument in longitudinal
axis) [0.1°]
Measured value format: pbb.ee
Range: –25.0 … +25.0 °
<value3> – Water coverage for pressure measurement [dbar]
Measured value format: pb.eee
Range: +0 … +9.999 dbar
aM4! atttn<CR><LF>
and
a<CR><LF> after 1 second
Beam 2: Read signal amplitudes of the last measurement
a – Sensor address
ttt – Time in seconds until the sensor has
provided the result.
OTT SLD response = 001
n – Number of measured values
OTT SLD response = 9
a<CR><LF> – Service request
aD0!
aD1!
aD2!
a<value1><value2><value3><CR><LF>
a<value4><value5><value6><CR><LF>
a<value7><value8><value9><CR><LF>
Send data
a – Sensor address
<valuex>
– Beam 2: signal amplitude
Measuring cell 1 … 9 [counts]
Measured value format: pbbb
Range: +0 … +255 counts
aM5! atttn<CR><LF>
and
a<CR><LF> after 1 second
Read unfiltered velocity in x direction
a – Sensor address
ttt – Time in seconds until the sensor has
provided the result.
OTT SLD response = 001
n – Number of measured values
OTT SLD response = 9
a<CR><LF> – Service request
28