Water level/pressure · temperature
Parameter number sensor number, 4 characters (alphanumeric)
Parameter name sensor name, max. 40 characters (alphanumeric)
Measurement type/Meas.range Water level: depth or level · pressure
– Conductivity
0.001 … 2.000 mS/cm · 0.10 … 100.00 mS/cm
– Salinity: 2 … 42.00 PSU · 2 … 60.00 ppt
Unit + – Pressure sensor: m, cm, feet, inch / bar, psi
– Temperature: °C · °F
– Conductivity: mS/cm · µS/cm
– Salinity: PSU · ppt
– TDS: mg/l
Decimal places fixed specification, depending on the units and measuring range
– m: 3 or 2; cm: 1 or 0; feet: 2 or 1; inch: 1 or 0
– bar: 4 or 3; psi: 3 or 2
– °C: 2 or 1, °F: 2 or 1
– mS/cm: 3 or 2; µS/cm: 3
– PSU: 2; ppt: 2
– mg/l: 2
For the "Water level · depth" measurement type: Enter contact gauge value
For the "Water level · level" measurement type: Enter staff gauge value
(reference to level zero)
For the "Pressure" measurement type: Enter reference pressure
Sample interval 5 s … 24 h: time delay at which the OTT ecoLog 800 (3G) records measured val-
ues and stores them (store only, if sample interval = mean interval). The sample
interval is set to a fixed time pattern (e.g. sample interval of 10minutes, time pattern …, 00:10, 00:20, 00:30, … .
Storage interval 5 s … 24 h: time period for which the OTT ecoLog 800 (3G) calculates an arith-
metic mean from the sample interval values and stores it. The storage interval must
be equal to or larger than the sample interval. The sample interval must be contained in the storage interval as an integer number (e.g. sample interval: 10 minutes; storage interval: 1hour ➝ Every hour, the OTT ecoLog 800 (3G) stores an
arithmetic mean from 6 sample interval values).
Authorize Roaming Allows the cellular modem to book into other, external cellular networks (not only
into the cellular network specified by the SIM card). This might be of importance in
regions that are located close to a border, in which a foreign cellular network may
have better coverage but is more expensive.
SMS data transmission active Refer to "Setting up the cyclic data transmission via SMS".
GPRS data transmission active Refer to "Setting up the cyclic data transmission via GPRS(/3G)"; 3G data trans-
GPRS/3G data transmission active mission (UMTS/HSPA+) is only available with the OTT ecoLog 800 3G
Store GSM Signal strength Stores the GSM signal strength in a separate channel. The sample interval in this
channel is 24 hours (factory setting: Storage delta = 1); irrespective of this, the
OTT ecoLog 800 (3G) saves 1 value per day in any case.
Additional settings for conductivity, salinity and TDS
the mathematical algorithm on which the calculation of the specific conductivity at
a defined reference temperature is based. Possibilities: "Freshwater"; Saltwater";
Standard method 2510"; "ISO 7888/EN 27888"; "---" (none). For the "Standard
method 2510", the reference temperature can be selected: 20 °C or 25 °C.
the mathematical algorithm on which the calculation of the salinity is based. Possi-
bilities: "Standard method"; "USGS 2311". With the calculation method USGS
2311, the unit is "ppt"; with the standard method the unit "PSU" is fixed.
With the check box activated, the datalogger saves the values. With the check box
deactivated, the instantaneous values are still visible in the observer window.
Depth to water – set newly
Water level – set newly
Pressure value – set newly
Temperature compen sation
(Conductivity)
Calculation method
(Salinity)
Store salinity
Store TDS
30