Orphee Mythic 22 User manual

mythic 22
User’s Manual
Orphée SA
19, chemin du Champ des Filles - 1228 CH-Plan Les Ouates / SWITZERLAND Tel : +41.(0)22.706.18.40 Fax : +41.(0)22.794.43.91 - Web Site : www.orphee-medical.com
Cod. M22/UM-EN/006
Page 2/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
Revision Nb
Date
Author
Software
Comments
01
16/11/05
HC
> V1.0
Creation
02
27/03/06
OM/HC
> V1.0.0-008
Sections 3.4.6 / 4.2 / 5.6.3 / 9.8 modified
03
25/10/06
HC/SH
> V.1.1.0
Sections 3.4.4 / 5.7 / 8.5 / 9.6 / 9.1.1 modified
04
03/05/07
HC
> V 2.0.0
Management of USB key (sections 3.3/5.10/6.2/7.2)
05
06
07/12/09
HC/CM
OM
> V2.2.0
Add flags information. (section 8.5.5 / 8.5.6) Modify (QC & CALIBRATION ) Update Hydraulic diagram and tubing list Update of Installation Kit Update of Maintenance Kit Update Hydraulic diagram and tubing list
M
Manufactured in France for
ORPHEE SA
19, chemin du champ des filles CH-1228 Geneva / Plan-les-Ouates SWITZERLAND
Tel : +41 (0) 22 706 1840 Fax : +41 (0) 22 794 4391
http://www.orphee-medical.com
CONTACT ADDRESS
MANUFACTURER
REVISIONS
LOCAL AGENT
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 3/99
READ THIS BEFORE USING THE EQUIPMENT
WARNING! RISK OF DANGER! Indicates a procedure to be strictly respected in order to avoid any risks for the operator (user) or damages on the instrument or on the quality of results.
Indicates that wearing gloves is mandatory before performing the described operation due to risk of contact with materials that may be infectious.
Indicates hot temperatures surfaces and risk of burns
NOTA
Indicates important additional information
KEEP THESE INSTRUCTIONS
DANGER
Misuse of electrical equipment may cause electrocution, burns, fire and other hazards.
Check that the voltage setting matches the supply voltage. Protective earthing is required, plug the MYTHIC 22 into a supply outlet which has an earth connection. Preserve a good access to the supply outlet to be able to unplug the MYTHIC 22 in emergency case. Do not place the power supply adapter in liquid, nor put it where it could fall into liquid. If the power supply adapter
becomes wet, unplug it before touching it. Do not use the MYTHIC 22 if it is not working properly, or if it has suffered any damage ( damage to the supply cord or its plug; damaged caused by dropping the power supply adapter).
Do not let the power supply adapter or its flexible cord come into contact with surfaces which are too hot to touch. Do not place anything on top of the MYTHIC 22  Do not use the MYTHIC 22 where aerosol sprays are being used, or where oxygen is being administred. Do not use the MYTHIC 22 out of doorsAlways switch off the MYTHIC 22 and disconnect the power adaptor before dismantling any part.
The MYTHIC 22 is an automated hematology analyzer for in vitro diagnostic use in clinical laboratories by an
authorized people.
- Only human blood or artificial control blood should be run.
- Only the reagents mentioned in this manual are permitted to use.
- The optimum performances can be only achieved if the cleaning and maintenance procedures are carefully followed. Due to the use of this equipment, all parts and surfaces of the MYTHIC 22 are potentially infective. Wearing rubber gloves is highly recommended and after completion of work, wash hands with disinfectant.
Always replace or use parts of the equipment by parts supplied by ORPHEE distributor. Basic safety precautions should always be taken. If the equipment is not used according to the manufacturer’s
instructions, the protective by the equipment may be impaired. The treatment of waste and the elimination of a part or the complete instrument must be done in compliance with the local legislation. Any output or input connections (except the printer and the barcode reader supplied by ORPHEE) cannot be done without the ORPHEE representative authorization. Do not open the door located on the right side of the instrument (see section 1.1.3) when an hydraulic cycle is in progress for it would lead to an immediate stop. To re-start, shut the door and run a Control cycle (see section 9.3.1)
Page 4/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
Guidance and manufacturer’s declaration – Electromagnetic emmissions
The MYTHIC 22 is intended for use in the electromagnetic environment specified below. The customer or the user of the MYTHIC 22 should assure that it is used in such an environment.
Emmissions test
Compliance level
Electromagnetic environment - guidance
Harmonic emissions
IEC 61000-3-2
Class A
The MYTHIC 22 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Voltage fluctuations/flicker emissions IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – Electromagnetic immunity
The MYTHIC 22 is intended for use in the electromagnetic environment specified below. The customer or the user of the MYTHIC 22 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment - guidance
Electrostatic discharge (ESD) IEC 61000-4-2
±6 kV contact ±8 kV air
Complies
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Electrical fast transient/burst IEC 61000-4-4
±2 kV for power supply lines ±1 kV for input/output lines
Complies
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
±1 kV differential mode ±2 kV common mode
Complies
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
<5 % UT (>95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles <5 % UT (>95 % dip in UT) for 5 sec
Complies
Mains power quality should be that of a typical commercial or hospital environment. If the user of the MYTHIC 22 requires continued operation during power mains interruptions, it is recommended that the MYTHIC 22 be powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
3 A/m
Complies
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. NOTE UT is the a.c. mains voltage prior to application of the test level.
This equipment needs special precautions regarding general requirements for safety.
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 5/99
Guidance and manufacturer’s declaration – Electromagnetic immunity
The MYTHIC 22 is intended for use in the electromagnetic environment specified below. The customer or the user of the MYTHIC 22 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the MYTHIC 22, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF IEC 61000-4-6
3 Vrms 150Khz to 80Mhz
3 Vrms
d
= 1,2√P
Radiated RF IEC 61000-4-3
3 Vrms 80Mhz to 2,5Ghz
3 Vrms
d
= 1,2√P 80MHz to 800MHz
d
= 2,3√P 800MHz to 2,5GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80Mhz and 800MHz, the higher frequency range applies.
NOTE 2 Theses guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM an FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should considered. If the measured field strength in the location in which the MYTHIC 22 is used exceeds the applicable RF compliance level above, the MYTHIC 22 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the MYTHIC 22.
b
Over the frequency range 150KHz to 80MHz, field strengths should be less than 3V/m.
Page 6/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
The symbol on the product indicates that this product may not be treated as household waste. Instead it shall be handed over the applicable collection point for the recycling of electrical and electronic equipment. By ensuring this product is disposed of correctly, you will help prevent potential negative consequences for the environment and human health, which could otherwise be caused by inappropriate waste handling of this product. For more detailed information about recycling of this product, please contact your local city office or your distributor of this product.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
The user may find the following booklet, prepared by the Federal Communications Commission, helpful: How to identify and Resolve Radio/TV Interference Problems. This booklet is available from the U.S. Government Printing Office, Washington, D.C. 20402, Stock No. 004-000-00345-4. Pursuant to Part 15.21 of the FCC Rules, any changes or modifications to this equipment not expressly approved by C2
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 7/99
Name und Adresse der Firma Nom et adresse de l’entreprise Nome e indirizzo della ditta Name and address of the firm
Orphée S.A. 19 Chemin du Champ des Filles 1228 Plan Les Ouates
Wir erklären in alleiniger Verantwortung, dass Nous déclarons sous notre propre responsabilité que Dichiariamo sotto nostra responsabilità che We declare under our sole responsibility that
das Medizinprodukt für die In-vitro-Diagnostik le dispositif médical de diagnostic in vitro il dispositivo medico-diagnostico in vitro the in vitro diagnostic medical device
Mythic 22 Ref. M22OT
mit folgender Klassifizierung nach der Richtlinie über In-vitro-Diagnostika 98/79/EG avec la classification selon la directive relative aux dispositifs médicaux de diagnostic in vitro 98/79/CE con la classificazione secondo la direttiva relativa ai dispositivi medico-diagnostici in vitro 98/79/CE classified as follows according to the directive on in vitro diagnostic medical devices 98/79/EC
Produkt der Liste A, Anhang II / Dispositif de la liste A, annexe II /
Dispositivo dell’elenco A, allegato II / Device of List A, Annex II
Produkt der Liste B, Anhang II / Dispositif de la liste B, annexe II /
Dispositivo dell’elenco B, allegato II / Device of List B, Annex II
Produkt zur Eigenanwendung, das nicht in Anhang II genannt ist / Dispositif destiné à l’autodiagnostic non listé dans l’annexe II / Dispositivo per test autodiagnostico non elencato nell’allegato II / Device for self-testing not listed in Annex II
Sonstiges Produkt / Autre dispositif / Altro dispostivo / Other device
allen Anforderungen der Richtlinie über In-vitro-Diagnostika 98/79/EG entspricht, die anwendbar sind.
remplit toutes les exigences de la directive relative aux dispositifs médicaux de diagnostic in vitro 98/79/CE qui le concernent.
soddisfa tutte le disposizioni della direttiva relativa ai dispositivi medico-diagnostici in vitro 98/79/CE che lo riguardano.
meets all the provisions of the directive on in vitro diagnostic medical devices 98/79/EC which apply to it.
KONFORMITÄTSERKLÄRUNG /
DECLARATION DE CONFORMITE
DECLARATION OF CONFORMITY /
DICHIARAZIONE DI CONFORMITA
Page 8/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
Angewandte Gemeinsame Technische Spezifikationen, harmonisierte Normen, nationale Normen oder andere normative Dokumente
Spécifications techniques communes, normes harmonisées, normes nationales et autres documents normatifs appliqués
Specifiche tecniche comuni, norme armonizzate o nazionali applicate, altri documenti normativi applicati
Applied common technical specifications, harmonised standards, national standards or other normative documents
IEC 60601-1-2 (2001) EN 61000-3-2 EN 61000-3-3 EN 61000-4-2 (95) A1 (98) A2 (01) EN 61000-4-3 (02) EN 61000-4-4 (95) A1(01) EN 61000-4-5 A1 (01) EN 61000-4-6 (96) A1 (01) EN 61000-4-11 (94) A1 (01) EN 55011 Class B EN 55022 Class B IEC 61010-1 (2001) IEC 61010-2-081 (2001) IEC 61010-2-101 (2002)
Konformitätsbewertungsverfahren Procédure d’évaluation de la conformité Procedimentodi valutazionedellaconformità Conformity assessment procedure
Konformitätsbewertungsstelle (falls beigezogen) Organe respons. de l'évaluat.de la conformité(si consulté) Organo incaric. della valutaz. della conform. (se consultato) Notified Body (if consulted)
N/A
Ort, Datum / Lieu, date / Luogo, data / Place, date
Genève le 06 Juin 2008
Name und Funktion / Nom et fonction /Nome e funzione / Name and function
Philippe Daire
RA & QA
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 9/99
TABLE OF CONTENTS
READ THIS BEFORE USING THE EQUIPMENT .......................................................................... 3
1. INSTALLATION ....................................................................................................... 12
1.1 UNPACKING ................................................................................................................................................................................................ 12
1.1.1 Introduction ............................................................................................................................................................................................ 12
1.1.2 Unpacking Procedure ............................................................................................................................................................................. 12
1.1.3 Visual checking ........................................................................................................................................................................................ 13
1.2 INSTALLATION CONSTRAINTS ............................................................................................................................................................... 13
1.2.1 Installation place.................................................................................................................................................................................... 13
1.2.2 Installation environment ...................................................................................................................................................................... 13
1.3 ELECTRICAL CONNECTIONS ..................................................................................................................................................................... 14
1.3.1 Power supply block ................................................................................................................................................................................. 14
1.4 PRINTER CONNECTION ............................................................................................................................................................................. 14
1.5 CONNECTION, CHANGE AND PRIMING REAGENTS ................................................................................................................................. 15
1.5.1 Connection ............................................................................................................................................................................................... 15
1.5.2 Priming ..................................................................................................................................................................................................... 16
1.6 TRANSPORTATION AND STORAGE .......................................................................................................................................................... 18
2. GENERAL OVERVIEW ................................................................................................. 19
2.1 GENERALITIES .......................................................................................................................................................................................... 19
2.2 OVERVIEW ............................................................................................................................................................................................... 20
2.3 MAIN PART DESCRIPTION ...................................................................................................................................................................... 20
2.3.1 Display / Keyboard ................................................................................................................................................................................ 20
2.3.2 Dilution hydraulic part ......................................................................................................................................................................... 22
2.3.3 Mono electronic board ......................................................................................................................................................................... 23
2.3.4 Power Supply Block ............................................................................................................................................................................... 23
2.3.5 Reagent tray .......................................................................................................................................................................................... 24
3. INSTRUMENT SET UP ................................................................................................ 25
3.1 USER’S IDENTIFICATION ...................................................................................................................................................................... 25
3.1.1 Start Up ................................................................................................................................................................................................... 25
3.1.2 In process ............................................................................................................................................................................................... 25
3.2 SYSTEM STATUS ..................................................................................................................................................................................... 26
3.3 SET UP ..................................................................................................................................................................................................... 27
3.4 ADVANCED SET-UP .................................................................................................................................................................................. 28
3.4.1 Printer set up : ....................................................................................................................................................................................... 28
3.4.2 Communication: ...................................................................................................................................................................................... 28
3.4.3 Analysis options:.................................................................................................................................................................................... 29
3.4.4 Lab. parameters: ................................................................................................................................................................................... 29
3.4.5 Calibration factor: ................................................................................................................................................................................ 33
3.4.6 Other Setting: ....................................................................................................................................................................................... 33
3.4.7 Storage options: .................................................................................................................................................................................... 33
3.4.8 Version release: ..................................................................................................................................................................................... 34
4. SPECIFICATIONS ..................................................................................................... 35
4.1 ANALYTICAL SPECIFICATIONS .............................................................................................................................................................. 35
4.2 PHYSICAL SPECIFICATIONS .................................................................................................................................................................. 37
4.3 REAGENTS SPECIFICATIONS ................................................................................................................................................................ 39
4.3.1 Diluent...................................................................................................................................................................................................... 39
4.3.2 Lytic reagent “OnlyOne” ...................................................................................................................................................................... 40
4.3.3 Cleaning solution .................................................................................................................................................................................... 41
4.4 ANALYTICAL LIMITATIONS ................................................................................................................................................................... 42
4.4.1 Recommendations .................................................................................................................................................................................. 42
4.4.2 Interferences ........................................................................................................................................................................................ 42
Page 10/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
5. SAMPLE ANALYSIS ................................................................................................... 47
5.1 VERIFICATIONS BEFORE STARTING ...................................................................................................................................................... 47
5.2 START UP ................................................................................................................................................................................................. 47
5.3 REAGENT REPLACEMENT ......................................................................................................................................................................... 48
5.4 START UP RINSING ................................................................................................................................................................................. 49
5.5 PREPARATIONS BEFORE ANALYSIS ....................................................................................................................................................... 49
5.6 ANALYSIS ............................................................................................................................................................................................... 49
5.6.1 Introduction ........................................................................................................................................................................................... 49
5.6.2 Sample Identification .......................................................................................................................................................................... 50
5.6.3 Sample run .............................................................................................................................................................................................. 51
5.7 RESULTS .................................................................................................................................................................................................. 52
5.8 PRINTING ................................................................................................................................................................................................ 54
5.8.1 Model report (A4) – external printer ............................................................................................................................................... 54
5.8.2 Model report – Thermal printer ........................................................................................................................................................ 55
5.9 LOGS ........................................................................................................................................................................................................ 56
5.10 ARCHIVE ................................................................................................................................................................................................ 56
5.10.1 Results ................................................................................................................................................................................................... 57
5.10.2 View ........................................................................................................................................................................................................ 58
5.11 STAND BY AND SHUT DOWN................................................................................................................................................................. 59
6. QUALITY CONTROL .................................................................................................. 60
6.1 INTRODUCTION ....................................................................................................................................................................................... 60
6.2 QC ........................................................................................................................................................................................................... 60
6.2.1 Change ...................................................................................................................................................................................................... 61
6.2.2 Run control blood .................................................................................................................................................................................. 62
6.2.3 Levey-Jennings graph .......................................................................................................................................................................... 63
6.2.4 Restore.................................................................................................................................................................................................... 63
6.3 REPEATABILITY ....................................................................................................................................................................................... 63
7. CALIBRATION ......................................................................................................... 65
7.1 RESULTS ................................................................................................................................................................................................... 66
7.1.1 Calibration blood analysis ..................................................................................................................................................................... 66
7.1.2 Calibration ............................................................................................................................................................................................... 67
7.2 TARGET VALUE MODIFICATIONS ........................................................................................................................................................... 67
8. TECHNOLOGY.......................................................................................................... 69
8.1 DETECTION PRINCIPLE ............................................................................................................................................................................ 69
8.1.1 WBC, RBC, PLT Counting ....................................................................................................................................................................... 69
8.1.2 Five part diff measurement ................................................................................................................................................................ 70
8.1.3 Hemoglobin measurement .................................................................................................................................................................... 71
8.2 LEUCOCYTE ANALYSIS ............................................................................................................................................................................. 71
8.3 ERYTHROCYTE ANALYSIS ....................................................................................................................................................................... 72
8.4 ANALYSIS OF PLATELETS ....................................................................................................................................................................... 73
8.5 FLAGS ...................................................................................................................................................................................................... 73
8.5.1 General Flags .......................................................................................................................................................................................... 73
8.5.2 Instrument Flags .................................................................................................................................................................................. 74
8.5.3 Leucocytes Flags ................................................................................................................................................................................... 75
8.5.3 Erythrocyte and HGB Flags ................................................................................................................................................................ 76
8.5.4 Platelet Flags ......................................................................................................................................................................................... 76
8.5.5 QC Flags .................................................................................................................................................................................................. 76
8.5.6 STARTUP Flags ..................................................................................................................................................................................... 76
8.6 HYDRAULIC DESCRIPTION...................................................................................................................................................................... 77
8.6.1 Sampling module ..................................................................................................................................................................................... 77
8.6.2 Counting bath module ........................................................................................................................................................................... 77
8.6.3 Syringes module .................................................................................................................................................................................... 77
8.6.4 Optical manifold .................................................................................................................................................................................... 77
8.6.5 Optical bench ......................................................................................................................................................................................... 78
8.8 SOFTWARE .............................................................................................................................................................................................. 78
8.8.1 Windows .................................................................................................................................................................................................. 78
8.8.2 Menu tree ............................................................................................................................................................................................... 78
9. SERVICE ............................................................................................................... 80
9.1 MAINTENANCE ........................................................................................................................................................................................ 80
9.1.1 Maintenance table .................................................................................................................................................................................. 80
9.1.2 Concentrate cleaning ............................................................................................................................................................................ 81
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 11/99
9.1.3 Piston greasing ....................................................................................................................................................................................... 82
9.2 HYDRAULIC CYCLES ................................................................................................................................................................................. 83
9.3 MECHANICS............................................................................................................................................................................................. 84
9.4 REPAIRING............................................................................................................................................................................................... 85
9.4.1 Emergency stop ...................................................................................................................................................................................... 85
9.4.2 Needle or o-ring replacement ............................................................................................................................................................ 86
9.4.3 Baths dismantling .................................................................................................................................................................................. 88
9.4.4 Baths o-ring replacement .................................................................................................................................................................... 91
9.4.5 Aperture block replacement ............................................................................................................................................................... 91
9.5 TROUBLESHOOTING ............................................................................................................................................................................... 92
9.5.1 Analytical problems ............................................................................................................................................................................... 92
9.5.2 Other problems ..................................................................................................................................................................................... 93
9.6 TROUBLESHOOTING MESSAGE .............................................................................................................................................................. 93
9.7 LOGS ERRORS .......................................................................................................................................................................................... 97
9.8 HYDRAULIC DIAGRAM ............................................................................................................................................................................. 98
1. INSTALLATION
Page 12/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
If the MYTHIC 22 has been stored at a temperature less than 10°C it must be left at room temperature during 24 hours. It must be calibrated at each displacement and installation (See section 7).
INSTALLATION KIT
QTY
Designation
1
Tubing DILUENT
1
Tubing WASTE
1
150W switching adapter
1
European Power line cord
1
MYTHIC 22 User manual
1
MYTHIC 22 Installation Report
1
MYTHIC 22 Quality Control Certificate
1
Screwdriver Slot 1/4"
MAINTENANCE KIT
QTY
Designation
1
Tygon tubing L=500mm 1.02mm
1
Tygon tubing L=500mm 1.3mm
1
Tygon tubing L=1000mm 1.52x3.2mm
1
Tygon tubing L=500mm 2.06x4mm
1
Tubing 10
1
Tubing 50
5
Cables Ties
1
Rinsing Head O-ring
1
Silicon grease (3gr)
1
Short Arm TORX T10 Tool
1
Short Arm TORX T20 Tool
1. INSTALLATION
1.1 UNPACKING
1.1.1 Introduction
The MYTHIC 22 is an automated hematology analyzer for in vitro diagnostic use in clinical laboratories by an authorized people.
- Only human blood or artificial control blood should be run.
- Only the reagents mentioned in this manual are permitted to use.
- The optimum performances can be only achieved if the cleaning and maintenance procedures are carefully followed.
1.1.2 Unpacking Procedure
Before unpacking the instrument, we recommend to check the box of the instrument and notify any damage to the carrier.
- Open the box on the top, remove the starter kit.
- Remove the MYTHIC 22 from the box.
Starter kit contents:
1. INSTALLATION
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 13/99
Open the door on the right side with the key provided in the
kit.
To be checked :
1- Counting chambers perfectly locked in their manifold
locations.
2- Needle’s dismountable system located in the rocker. 3- Rocker in front position at the maximum course.
HAZARDOUS MOVING PARTS, BEWARE TO STAY AWAY FROM THESE PARTS WHEN THE MACHINE IS SWITCH ON.
1
2
3
1.1.3 Visual checking
1.2 INSTALLATION CONSTRAINTS
1.2.1 Installation place
To ensure that the MYTHIC 22 fulfills its function, place the instrument on a table which supports the weight of the instrument, printer and reagents (around 40 Kg). Leave a space of 10 cm in the rear of the instrument to ensure a well-ventilated place. Avoid a place that can be exposed to direct sunlight.
1.2.2 Installation environment
a) Indoor use; b) Altitude up to 2 000 m; c) Temperature 18 °C to 34 °C; d) Maximum relative humidity 80 % for temperatures up to 31 °C decreasing linearly to 50 % relative humidity at 40 °C; e) MAINS supply voltage fluctuations up to ±10 % of the nominal voltage; f) Transient over voltages typically present on the MAINS supply. g) Rated pollution degree II.
1. INSTALLATION
Page 14/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
CONNECTION :
SYMBOL :
- 2 USB ports (section 3.4.7) :
- Ethernet connection (TCP/IP) :
- External barcode reader (RS232C) :
- Host connection (RS232C) :
- Printer connection (Centronics) :
- Power supply cord connection :
- Equipotentiality :
Any output or input connections (except the printer and the bar code reader supplied by ORPHEE) cannot be done without the ORPHEE representative authorization.
- In the case of replacement of the main power wire supplied with the MYTHIC 22 the new one must comply with the local regulation (31.5mm cable and 250V 10A plug).
- The MYTHIC 22 has been certified with the power supply box provided with the machine. The use of another external power supply box is not guaranteed. Please contact your Orphée’s representative.
Please contact Orphée’s representative if you want to use the instrument in special conditions (height higher
than 2000 m or special power supply conditions).
1.3 ELECTRICAL CONNECTIONS
All the connectors are in the rear of the MYTHIC 22
1.3.1 Power supply block
MYTHIC 22 must be connected to the power with the power supply block provided with the starter kit. Choose a well-ventilated place for the block and be sure to connect this power supply in a socket-outlet with a correct earth connection. The power supply block must be placed at the rear of the MYTHIC 22 and, if possible in an upper position to avoid the contact with any liquid. To disconnect electrically the MYTHIC 22, remove the power supply plug from the main circuit.
1.4 PRINTER CONNECTION
Connect the printer cable in conformity with the printer user’s manual.
Use the parallel rear plug of the MYTHIC 22 ( ) or the USB plug ( ) to connect the printer cable. Select the printer driver (section 3.3).
1. INSTALLATION
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 15/99
Before handling the reagents, read carefully their specifications described in section 4.3.
- Remove the door on the left side of the instrument.
- Put the reagent bottles in the dedicated location.
- Remove the caps of the bottles
- Tighten the red caps on the Lyse bottle (red sticker) and the blue one on the cleaning solution bottle (blue sticker).
- Connect the diluent tube (male connector) on the outlet on the bottom and tighten the cap on the diluent container.
- To use a 20 liter diluent container adds the tubing straw adaptor supplied with the installation kit.
- Connect the waste tube (female connector) on the outlet on the top and tighten the cap on an empty container.
Do not modify the type and the length of the diluent and waste tubes. The diluent must be placed at the same level as the MYTHIC 22.
It is mandatory to collect the waste in a container and treat it in compliance with your local legislation.
Diluent
Waste
1.5 CONNECTION, CHANGE AND PRIMING REAGENTS
MYTHIC 22 works with the reagents described in section 4.3. Some commercial reagents can also be used. Please contact your Orphée’s representative.
1.5.1 Connection
Lysis reagent and cleaning solution:
Diluent and waste:
1. INSTALLATION
Page 16/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
Before starting, be sure that all the reagent and waste tubes are properly connected. The reagents must be stored 24 hours minimum at room temperature before use.
1.5.2 Priming
When first starting the MYTHIC 22, it is necessary to perform a complete prime of the fluidic circuit.
Priming procedure :
Switch on:
- Connect the power supply block.
- Press the ON/OFF button.
- The cycle LED turns red. No cycle can be performed before it turns green.
- The information window could stay up to 3 mn to enable the update of all the files.
Login:
- The operator’s identification display appears.
- Enter the user’s identification, the password (see section 3.1) and press
to validate.
- AUTONOMY (run) indicates the number of samples (runs) you can perform (calculated with the smaller quantity of reagents).
- If this window appears, it means that several results in memory have not been sent before the MYTHIC 22 was switched off.
- Press YES to send them immediately, or press LATER to wait at another time or NO if you do not want to send them.
- No USB key is available, connect an USB key then press YES or see section 3.4.7 to change the archive mode.
1. INSTALLATION
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 17/99
NOTA
To do an emergency stop push shortly on the switch on/off
button
System priming:
- The main menu is displayed.
- Press on .
- Press : The MYTHIC 22 performs a complete priming cycle.
- The cycle LED turns red. No cycle can be performed before it turns green.
- AUTONOMY (run) indicates the number of samples (runs.
- To prime or to know the quantity of reagent press the dedicated button.
- Press to visualize the cycle counters.
- To reset the counter with the button , please contact your Orphée’s representative.
1. INSTALLATION
Page 18/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
MYTHIC 22 IS NOW READY TO WORK.
DILUENT PRIME:
- From the MAIN MENU press then
- Enter lot number, the expiry date and the container capacity.
- Press to validate the new entry or after changing a new container with the same information.
- After the replacement a new container or to prime the diluent, press
- A new entry is automatically done in the logs (see section 5.9)
LYSE AND CLEANER PRIME: Proceed as described above for the diluent.
WASTE:
- Enter only the capacity of the container.
- After replacement of the waste container press
to have access to this screen.
to reset to initialize the waste calculation.
1.6 TRANSPORTATION AND STORAGE
Storage temperature: -10°C to +50°C. If the MYTHIC 22 has been stored at a temperature less than 10°C it must be left at room temperature during 24 hours. Before transportation outside the laboratory, perform a complete cleaning with a disinfectant in compliance with the local legislation.
2. GENERAL OVERVIEW
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 19/99
2. GENERAL OVERVIEW
2.1 GENERALITIES
MYTHIC 22 is a fully automated analyzer performing hematological analysis on whole blood collected on EDTA K2 or K3 tubes.
- Sample volume: 15,7 µl (inside the needle, the total volume could be upper in function of the blood
remaining outside of the needle).
- Throughput: > 45 samples/hour
- 22 analysis parameters in DIF mode and 12 parameters in CBC mode:
Leukocyte parameters: WBC White Blood Cells LYM Lymphocytes in % & # (DIF mode only) MON Monocytes in % & # (DIF mode only) NEU Neutrophils in % & # (DIF mode only) EOS Eosinophils in % & # (DIF mode only) BAS Basophiles in % & # (DIF mode only)
Erythrocyte parameters: RBC Red Blood Cells HGB Hemoglobin HCT Hematocrit MCV Mean Corpuscular Volume MCH Mean Corpuscular Hemoglobin MCHC Mean Corpuscular Hemoglobin Concentration RDW Red Blood cells Distribution Width
Thrombocyte parameters: PLT Platelet MPV Mean Platelet Volume PDW* Platelet Distribution Width PCT* Thrombocrit
* For Investigation Use only in the United States of America
2. GENERAL OVERVIEW
Page 20/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
Stand By and ON/OFF button
Numeric keyboard
Delete (DEL)
Exit (ESCAPE)
Cursors
Enter (ENTER)
Cycle in process Led
Touch screen LCD display 380*240
2.2 OVERVIEW
2.3 MAIN PART DESCRIPTION
2.3.1 Display / Keyboard
MYTHIC 22 consists of 8 main parts:
1. Display / Keyboard.
2. Dilution hydraulic part.
3. Mono electronic board.
4. Reagent tray.
5. Connection.
6. External power supply block.
7. Printer.
8. Barcode reader (option).
2. GENERAL OVERVIEW
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 21/99
Back to previous screen
Print, Send, Select Options
Direct access to main menu
Start Up rinsing and blank control
Cleaning and standby mode
Patient Archive (Section 5.10).
Quality Control (Section 6).
Calibration (Section 7).
Mythic setting parameters (Section 3).
Maintenance and Service Menu
Analysis performing (Section 5.5).
Mythic Events Logs (Section 5.9).
Reagents replacement (Section 5.3).
Log In and Log Out (Section 3.1).
Date and time. System status (Section 3.2).
Main Menu description
2. GENERAL OVERVIEW
Page 22/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
Sampling module
Syringe module
Counting module
Optical manifold
Optical bench
2.3.2 Dilution hydraulic part
All the fluidic part is on the right side of the instrument and consists of only five modules:
- Sampling module : o Rocker (patent pending): Manages the rise and descent of the needle.
- Syringe module (patent pending) consists of one block :
o Reagent syringes (Diluent, lysis), sampling and air syringes. o Liquid valve manifold assembly and tubing.
- Counting chambers :
o WBC and RBC counting chambers and hemoglobin measurement. o Liquid valve manifold assembly and tubing.
- Optical Manifold :
o Liquid valve manifold assembly and tubing.
- Optical bench :
o Optical bench (patented) with his flow cell (patented).
2. GENERAL OVERVIEW
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 23/99
To avoid all deterioration risks, only the service people can touch this electronic board.
- In the case of replacement of the main power wire supplied with the MYTHIC 22 the new one must comply with the local regulation.
- The MYTHIC 22 has been certified with the power supply box provided with the machine.
- The use of another external power supply box is not guaranteed. Please contact your Orphée’s representative.
Switch on/off
2.3.3 Mono electronic board
The mono electronic board is located between the hydraulic part and the reagent tray. The board, driven by a 32-bit processor, manages the following parts:
- Sample needle, rocker, syringe block motors.
- Display and keyboard.
- Connexion mode (RS232, Ethernet, …).
- Printer.
- Measurement (Counting, hemoglobin measurement).
- Data processing.
- External barcode reader.
2.3.4 Power Supply Block
MYTHIC 22 is supplied with an external power supply block.
2. GENERAL OVERVIEW
Page 24/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
2.3.5 Reagent tray
The reagent tray is dedicated to the OnlyOne lysing reagent and cleaning solution bottles.
3. INSTRUMENT SET UP
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 25/99
3. INSTRUMENT SET UP
3.1 USER’S IDENTIFICATION
3.1.1 Start Up
- After the instrument’s initialization, the identification window is displayed.
- In the window , the last operator’s identification appears.
- Either the identification is yours, press and enter your
password or the identification is not, press to enter your login.
- The window enables to change the language. Press to validate it.
- AUTONOMY (run) indicates the number of samples (runs) you can perform (calculated with the smaller quantity of reagents).
- Enter your identification name with the alphabetic keyboard.
- Place the cursor in the Password window.
- Enter your password for identification.
- For the first login, MYTHIC 22 proposes 3 access levels :
- Biologist Password can be modified in section 3.3.6.
o User : No password o Biologist : Password by default 1- 2- 3 o Service people
3.1.2 In process
- To change operator during the process, press to return to the
main menu, and then press on
- To change identification, proceed as described above (section 3.1.1).
3. INSTRUMENT SET UP
Page 26/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
- If this window appears, it means that several results in memory have not been sent before the MYTHIC 22 was switched off.
- Press YES to send them immediately, or press LATER to wait at another time or NO if you do not want to send them.
- No USB key is available, connect an USB key then press YES or see section 3.4.7 to change the archive mode.
3.2 SYSTEM STATUS
- Press on the date and hour to have access to the system status window.
- Different system status information is displayed.
- To change the screen luminosity press L, N or H
- To return to the MAIN MENU press
3. INSTRUMENT SET UP
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 27/99
3.3 SET UP
- From the MAIN MENU press on
- This menu is available for all users.
- The DATE & TIME window enables to modify the time and the date
- To select the language of the Mythic menu, choose the right one in the
Language combo box.
- : Select the printer or no printing.
- : Select the paper size depending on the number of results per page.
- : Two SID are available; Start day SID enables to select the first SID for each new day.
- : If you want to select a new SID number
- : Biologist reserved for complete settings.
(See section 3.4).
- Once modifications are done, press either to valid or to exit keeping the previous setting.
- Press to print, save or load from an USB key all the set up.
- To load new printer drivers plug the USB key then press on in the previous screen.
- Select the printer and its connection mode.
- Then press
- Press then, the driver is loaded in the MYTHIC 22
- This prompt appears if the release is failed. Check the USB connection or change the USB key or call your Orphée’s representative.
This prompt appears if the release is done successfully.
3. INSTRUMENT SET UP
Page 28/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
Any modification can affect the quality of the results. We
recommend modifying these values only after an Orphée’s
training.
3.4 ADVANCED SET-UP
3.4.1 Printer set up:
- This menu is reserved to biologist (see section 3.1).
- Please refer below for the description of each key.
- Printer set up menu is intended to present the printing report
- To select an option on the report, press on the corresponding case.
- To enter a header, press key.
- To exit the menu, either press to keep the last setting, or to save the last modifications.
3.4.2 Communication:
- Reserved for technical Service people.
- To set up the connection between MYTHIC and Host.
3. INSTRUMENT SET UP
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 29/99
3.4.3 Analysis options:
- OPTIONS box:
o ALWAYS ID and ALWAYS PID: To run a sample, user mandatory
needs to enter a ID and/or a PID.
o RUO: With tag the PCT and PDW parameters are displayed,
printed and send.
o US MODE: The Research Use Only message is printing below the
printing report.
o ABSOLUTE DIF: With tag absolute values for sub-populations of
leucocytes are displayed. In the other case, percentages are displayed.
o QC ALARMS: The message QC_F “QC failed” appear below the
printing report when the QC result is out of tolerance or expired (see section 8.5.5); The message QC_ND “QC not done” appear below the printing report when it is not run (see section 8.5.5).
- : Gives a choice of three unit systems: Standard, International System, and mmol.
- Barcode Reader box: The reading of the bar code is dedicated to the PID or ID or SID.
- BOLD DISPLAY box: display and print in bold-faced type the different choices in this box.
Once modifications are done, press either to validate your choices or to exit keeping the previous setting.
3.4.4 Lab. parameters:
- Select the blood type in the combo box then press:
to adjust the normal and panic limits (see section 3.4.4.1).
to adjust the parameters thresholds (see section 3.4.4.2).
to adjust the flags level (see section 3.4.4.3).
to adjust the corrections factors (see section
3.4.4.4).NOTA: No correction factors with type STANDARD.
- To enter a new blood type, press NOTA: The name of the first type STANDARD cannot be change.
- Press to print the blood type set up.
- allow to print all the blood type set up (about 20 pages are printed)
- only the blood type in the combo box is printed.
3. INSTRUMENT SET UP
Page 30/99
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
3.4.4.1 Limits:
To rename a blood type press then select it
then press to accede at the alphabetic keyboard then press
to validate or to exit without any modification.
- This display enables to enter normal and panic limits for every 22 parameters given by the MYTHIC 22 (see section 8).
- To validate the new values, press VALID key in the next page (see below).
- Once modifications are done, press to validate or to exit without any modification.
- Press to return to the parameter setting of the standard type.
Loading...
+ 69 hidden pages