optris
®
CX hs
TECHNICAL DATA
Infrared thermometer
for measurement of
smallest temperature
differences of 0.025 K
Features:
• New benchmark in low noise infrared thermometry
•
Suitable for relative measurement of smallest temperature differences
of 0.025 K
• Touches the physical limits ensuring low ambient temperature drift
• Complete solid state design without any moving parts
• Assures usability of a new class of ultra low noise level infrared thermometers
General specications
Environmental rating IP 65 (NEMA-4)
Ambient temperature –20 °C ... 75 °C (–4 °F ... 167 °F)
Storage temperature – 40 °C ... 85 °C (–40 °F ... 176 °F)
Relative humidity 10 – 95 %, non condensing
Vibration (sensor) IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Shock (sensor) IEC 68-2-27: 50 G, 11 ms, any axis
Weight 350 g (12.3 oz)
Electrical Specications
Outputs / analog 0 / 4 – 20 mA
Output / alarm 0 – 30 V/ 500 mA (open collector)
Outputs / digital USB (optional)
Loop impedances Max. 1000 Ω
Cable length 8 m (26.2 ft)
Power Supply 5 – 30 V DC
1)
Measurement specications
Temperature range
(scalable via software)
Spectral range 8 – 14 μm
Optical resolution
(90 % energy)
CF-Optics (optional) 0.8 mm @ 10 mm (0.03 in @ 0.39 in)
System accuracy (at ambient
temp. 23 ±5 °C) (23 ±41 °F)
Repeatability (at ambient temp.
23 ±5 °C) (23 ±0.5 °F)
Temperature resolution 0.025 K (for temperatures >20 °C [>68 °F],
Response time (95 %) 150 ms (adjustable up to 999 s
Emissivity/ Gain
(adjustable via programming
keys or software)
Transmissivity/ Gain
(adjustable via programming
keys or software)
Signal processing (parameter
adjustable via programming
keys or software, respectively)
Software optris
1)
In dependence on supply voltage
2)
Whichever is greater
–30 °C ... 150 °C
(–22 °F ... 302 °F)
15:1
±1 % or ±1 °C
(±1 % or ±1.8 °F2))
±0.3 % or ±0.3 °C
(±0.3 % or ±0.6 °F1))
AVG >0.2 s)
via optional programming device)
0.100 – 1.100
0.100 – 1.100
Peak hold, valley hold, average; extended hold
function with threshold and hysteresis
2)
2)
®
Compact Connect
Optris Infrared Sensing, LLC · 200 International Drive, Suite 170 · Portsmouth · NH 03801 USA
Phone: 603-766-6060 · E-Mail: sales@optris-ir.com · www.optris.com
optris
®
CX hs
Optical specications
15:1 optics 15:1 optics with CF-lens
13
20
27
33
40
47
53
S7 (0.28)8 (0.31)
D 0
100
(0)
(3.94)
(0.51)
200
(7.87)
(0.79)
300
(11.8)
(1.06)
400
(15.7)
(1.30)
500
(19.7)
(1.57)
600
(23.6)
(1.85)
700
(27.6)
(2.09)
800
(31.5)
mm
(in)
mm
(in)
Dimensions
7
5
(0.20)
0.8
(0.03)5 (0.20)
10
(0.39)
(0.59)
15
S
(0.28)
D 0
(0)5 (0.20)
S= Spotsize / D= Distance
(0.43)
(0.79)
11
16
21
27
32
(0.63)
(0.98)
(0.83)
25
30
(1.18)
20
(1.06)
(1.38)
mm
(1.26)
(in)
mm
35
40
(in)
(1.57)
Accessories
Air purge collar CF-lens / Protective window
Compact Connect software
• Software for easy sensor setup and remote controlling, supports multi
tasking
• Graphic display for temperature trends and automatic data logging for
analysis and documentation with 1 ms response time
• Adjustment of signal processing functions and programming of outputs
of the sensor
• Automatic emissivity adjustment
• The software CompactConnect allows to customize the sensor to
application needs of the user
Specications are subject to change without notice · CX hs LT-US2017-10-A