![](/html/dc/dc3d/dc3d0d48d3af2fdac6be516a8f306897a8db6e4afe0813923992be46e0ae9c36/bg1.png)
innovative infrared technology
CTvideo 3M
Non-contact temperature measurement
of metals and ceramics from 50 °C to 1800 °C
(122 °F to 3272 °F) with adjustable focus,
patented crosshair laser and video sighting
![](/html/dc/dc3d/dc3d0d48d3af2fdac6be516a8f306897a8db6e4afe0813923992be46e0ae9c36/bg2.png)
optris
®
CTvideo 3M
TECHNICAL DATA
Features:
• Parallel use of video sighting and crosshair laser
for easy sensor alignment (measuring spots up
from 0.5 mm (0.02 in) under all viewing conditions
possible
• Manual focusing for measurement distances
from 90 mm (3.5 in)
• Response times up from 1 ms
Measurement specications
Temperature range
(scalable via software)
Spectral range 2.3 μm
Optical resolution
(90 % energy)
System accuracy
(at T
= 23 ± 5 °C
amb
[at
= 73 ±9 °F])
Tamb
Repeatability
(at T
= 23 ± 5 °C
amb
[at
= 73 ±9 °F])
Tamb
Temperature resolution
(digital)
Exposure time
(90 % signal)
Emissivity / Gain
(adjustable via software)
Transmissivity / Gain
(adjustable via program-ming
keys or software)
Signal processing
(parameter adjustable
via software)
Software (incl.) optris
1)
50 °C ... 400 °C (122 °F ... 752 °F) (3ML)
100 °C ... 600 °C (212 °F ... 1112 °F) (3MH)
150 °C ... 1000 °C (302 °F ... 1832 °F) (3MH1)
200 °C ... 1500 °C (392 °F ... 2732 °F) (3MH2)
250 °C ... 1800 °C (482 °F ... 3272 °F) (3MH3)
60:1 (3ML)
100:1 (3MH)
300:1 (3MH1-3MH3)
3)
± (0.3 % of reading + 2 °C)
± (0.3 % of reading + 36 °F)
± (0.1 % of reading + 1 °C)
± (0.1 % of reading + 34 °F)
0.1 K
4)
1 ms
0.100 – 1.100
0.100 – 1.000
Peak hold, valley hold, average;
extended hold function with
threshold and hysteresis
®
Compact Connect
(Sensor setup, video sighting and
process monitoring)
• Usable in up to 70 °C (158 °F) ambient temperature without
cooling and automatic laser switch o at 50 °C (122 °F)
• Short measuring wavelength of 2.3 μm decreases
measuring mistakes on surfaces with low or unknown
emission rate
• optris Compact Connect software for fast on-site
sensor setup, video alignment and real-time
process monitoring
General specications
Environmental rating IP 65 (NEMA-4)
2)
2)
2)
Ambient temperature –20 °C to 70 °C (50 °C with laser ON)
Storage temperature –40 °C to 85 °C (–40 °F to 185 °F)
Relative humidity 10 – 95 %, non-condensing
Vibration IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Shock IEC 68-2-27: 50 G, 11 ms, any axis
Weight 600 g (21.2 oz) (sensing head)
Electrical specications
Output / analog 0/4 – 20 mA, 0-5/10 V, thermocouple J, K
Alarm output 24 V/50 mA (open collector)
Output / digital USB 2.0
Video sighting Digital (USB 2.0), 640 x 480 px,
Output impedances mA max. 500 Ω (with 8-36 V DC)
Inputs Programmable functional inputs for external emissivity
Cable length
(sensor + elecronics)
Cable length (USB) 5 m, extendable up to 100 m over Ethernet
Current draw (laser) max. 160 mA
Power supply
Laser 635 nm
1)
T
> T
Objekt
2)
Specication valid at T
3)
Ɛ = 1, response time 1 s
4)
With dynamic adaptation at low signal levels
5)
The functionning of the LCD display may be limited in ambient temperatures
below 0 °C
sensing head
+ 25 °C
Object
(–4°F to 158 °F (122 °F with laser ON)
420 g (14.8 oz) (electronics)
Ethernet (via optional USB server)
FOV 3.1° x 2.4°
mV min. 100 kΩ load impedance, thermocouple 20 Ω
adjustment, ambient temperature compensation,
trigger (reset of hold functions)
3 m (standard), 5 m, 10 m
(9.8 ft [standard], 16.4 ft, 32.8 ft)
(16.4 ft, expandable up to 328.1 ft over Ethernet)
8
–
36 V DC
1 mW, ON/OFF via electronic box or software
≥ start of measurement range + 50 °C