Operating Instructions 7585 Operating Instructions

Page 1
Q-TRAK™ XP
INDOOR AIR QUALITY MONITOR
MODEL 7585

OPERATION AND SERVICE MANUAL

P/N 6013907, REVISION A
NOVEMBER 2020
Page 2
ii Q-Trak™ XP IAQ Monitor Model 7585
Page 3
Operation and Service Manual iii
Copyright©
TSI Incorporated / 2020 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824 Limitation of Warranty and Liability (effective February 2016)
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.) Seller warrants the goods, excluding software, sold hereunder, under normal use and
service as described in the operator's manual, to be free from defects in workmanship and material for 24 months, or if less, the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and exceptions:
a. The following gas sensors are warranted for 12 months from the date of shipment -
carbon dioxide, carbon monoxide, chlorine, hydrogen sulfide, nitric oxide, nitrogen dioxide, ozone, and TVOC high and low sensors;
b. The following gas sensors are warranted for 6 months from the date of shipment -
ammonia and formaldehyde sensors;
c. Parts repaired or replaced as a result of repair services are warranted to be free
from defects in workmanship and material, under normal use, for 90 days from the date of shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on
any fuses, batteries or other consumable materials. Only the original manufacturer's warranty applies;
e. This warranty does not cover calibration requirements, and seller warrants only that
the instrument or product is properly calibrated at the time of its manufacture. Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a
factory authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to
accidental or intentional damage, or is not properly installed, maintained, or cleaned according to the requirements of the manual. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO
SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT,
SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS.
BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S
OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS. TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER
OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE
Page 4
iv Q-Trak™ XP IAQ Monitor Model 7585
REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call Customer Service department at (800) 680-1220 (USA) or (1) 651-490-2860 (International).
Trademarks
TSI and TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other country’s trademark registrations. Q-Trak and TrakPro are trademarks of TSI Incorporated. Wi-Fi is a registered trademark of the Wi-Fi Alliance.
Page 5
v

CONTENTS

CONTENTS ......................................................................................... V
SAFETY .............................................................................................. IX
Description of Caution/Warning Symbols ........................................ ix
Caution ......................................................................................... ix
Warning ........................................................................................ ix
Caution and Warning Symbols ......................................................... x
Battery Safety and Disposal ............................................................. x
Laser Safety ..................................................................................... xi
Labels .............................................................................................. xii
Reusing and Recycling ...................................................................xiv
CHAPTER 1 OVERVIEW ................................................................... 1
CHAPTER 2 UNPACKING AND PARTS IDENTIFICATION ............. 3
Standard Equipment ......................................................................... 3
Optional Gas Sensors ....................................................................... 5
Optional Accessories ........................................................................ 6
Other Replacement Parts ................................................................. 6
CHAPTER 3 SETTING UP THE Q-TRAK™ XP MONITOR ............... 9
Installation of Gas Sensors ............................................................... 9
Providing Power to the Q-Trak™ XP .............................................. 10
Operating the Instrument with the AC Adapter ........................... 10
Operating the Instrument with the Lithium-Ion Battery ................ 11
Charging the Battery and Calibrating the Power Gauge ............. 12
Installing and Connecting to Available Wi-Fi® (optional
Wi-Fi® dongle ............................................................................... 13
Connecting to a Computer and Installing TrakPro™
Ultra Software .............................................................................. 13
CHAPTER 4 OPERATIONAL OVERVIEW ...................................... 15
Powering ON the Q-Trak™ XP Monitor .......................................... 15
Navigating the Dashboard Page+ .................................................. 17
Main Menu ...................................................................................... 18
Dashboard ................................................................................... 19
Settings ........................................................................................ 19
Manage Data ............................................................................... 20
Calibration ................................................................................... 20
Workflows .................................................................................... 21
Device Information ....................................................................... 22
Page 6
vi Q-Trak™ XP IAQ Monitor Model 7585
CHAPTER 5 SETTINGS ................................................................... 23
General Settings ............................................................................. 24
Date/Time ....................................................................................... 25
Display/Power Management ........................................................... 26
Sensors ........................................................................................... 27
Network Settings ............................................................................. 28
Units of Measure ............................................................................. 29
Notifications/Alarms ........................................................................ 30
CHAPTER 6 CALIBRATION ............................................................ 31
Gas Sensors ................................................................................... 31
Gas Sensor Verification .................................................................. 31
Gas Calibration Procedure.............................................................. 33
Step 1 – Select a Gas Sensor for Calibration .............................. 33
Step 2 – Place the calibration cap onto the gas sensor
you want to calibrate ................................................................ 34
Step 3 - Select the sensor to be calibrated on the
instrument display .................................................................... 34
Step 4 - Perform ZERO Capture ................................................. 35
Step 5 – Perform SPAN Capture ................................................. 35
Step 6 – Save or Discard Calibration .......................................... 36
Gas Calibration Unsuccessful ..................................................... 37
Built-in Sensor Calibration: Temperature, Relative Humidity,
and Barometric Pressure Sensors ............................................... 37
Reset Sensors to Default Calibration .............................................. 39
CHAPTER 7 LOGGING DATA ......................................................... 41
Logging Data ................................................................................... 42
Logging Profiles .............................................................................. 43
Locations ......................................................................................... 43
Settings ........................................................................................... 45
Logging Mode .............................................................................. 45
Sample Interval ............................................................................ 46
Schedules ....................................................................................... 46
CHAPTER 8 MANAGE DATA .......................................................... 49
Selecting Log Files .......................................................................... 49
View Log Files ................................................................................. 50
Delete Log Files .............................................................................. 50
Export Log Files .............................................................................. 51
Edit Log File Names ........................................................................ 52
Scrolling Through Data List ......................................................... 52
CHAPTER 9 WORKFLOWS ............................................................ 53
Percent Outside Air (%OA) Calculation Procedure ........................ 53
Page 7
Operation and Service Manual vii
CHAPTER 10 TRAKPRO™ ULTRA SOFTWARE ........................... 57
CHAPTER 11 UPDATE INSTRUMENT SOFTWARE ..................... 59
CHAPTER 12 MAINTENANCE ........................................................ 61
Replacing Sensors .......................................................................... 61
Recalibration ................................................................................... 61
Removing the Sensor Module from the Handle of the
Q-Trak™ XP Monitor. ................................................................... 61
Storage ........................................................................................... 63
CHAPTER 13 TROUBLESHOOTING .............................................. 65
Error Messages .............................................................................. 68
CHAPTER 14 HELP .......................................................................... 71
TSI® Technical Support/Service ...................................................... 71
APPENDIX A SPECIFICATIONS ..................................................... 73
APPENDIX B DEFINITIONS ............................................................. 77
Page 8
viii Q-Trak™ XP IAQ Monitor Model 7585
(This page intentionally left blank)
Page 9
ix

Safety

This section provides instructions to ensure safe and proper operation of the Q-Trak™ XP Indoor Air Quality Monitor Model 7585.

WARNING

• The instrument must be used in the manner described in this manual. Failure to follow all of the procedures described in this manual can result in serious injury to you or can cause irrevocable damage to the instrument.
• There are no user-serviceable parts inside the instrument. Refer all repairs to a qualified factory­authorized technician.
• The Q-Trak™ XP monitor is not rated for intrinsic safety. DO NOT operate the Q-Trak™ XP monitor under conditions where there is a risk of fire or explosion.

Description of Caution/Warning Symbols

Appropriate caution/warning statements are used throughout the manual and on the instrument that require you to take cautionary measures when working with the instrument.

Caution

CAUTION
Failure to follow the procedures prescribed in this manual might result in irreparable equipment damage. Important information about the operation and maintenance of this instrument is included in this manual.
Warning
WARNING
Warning means that unsafe use of the instrument could result in serious injury to you or cause damage to the instrument. Follow the procedures prescribed.
Page 10
x Q-Trak™ XP IAQ Monitor Model 7585

Caution and Warning Symbols

The following symbols may accompany cautions and warnings to indicate the nature and consequences of hazards:
Warns that the instrument contains a laser and that important information about its safe operation and maintenance is included in the manual.
Warns that the instrument is susceptible to electro­static discharge (ESD) and ESD protection should be followed to avoid damage.
Indicates the connector is connected to earth ground and cabinet ground.

Battery Safety and Disposal

This instrument uses a rechargeable Lithium ion battery with built-in protection against safety hazards. Always dispose of Li-ion
batteries and transport Li-ion batteries in compliance with regional regulations.
WARNING
• Use only TSI® supplied batteries in this instrument (P/N 800123).
• DO NOT abuse the battery in any way as the battery may rupture or catch fire.
• DO NOT use a substitute or non-rechargeable battery in this instrument.
• DO NOT short-circuit, incinerate, dismantle or mutilate Lithium ion batteries.
• DO NOT expose to water or heat.
• DO NOT use any battery which shows signs of
damage, such as bulging, swelling, a swollen plastic wrap, liquid in the plastic wrap, etc.
Page 11
Operation and Service Manual xi

Laser Safety

The Q-Trak™ XP Model 7585 is a Class I laser-based instrument. During normal operation, the user WILL NOT be exposed to laser radiation.
The following precautions should be taken to avoid exposure to hazardous radiation in the form of intense, focused, visible light.
• DO NOT remove any parts from the Q-Trak™ XP monitor unless you are specifically told to do so in this manual.
• DO NOT disassemble the Q-Trak™ XP monitor. There are no user-serviceable components inside the instrument.
WARNING
The use of controls, adjustments, or procedures other than those specified in this manual may result in exposure to hazardous optical radiation.
WARNING
If the Q-Trak™ XP monitor is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
When operated according to the manufacturer’s instruction, this device is a Class 1 laser product as defined by U.S. Department of Health and Human Services standards under the Radiation Control for Health and Safety Act of 1968.
D I S C L A I M E R
The CO2 measurement sensor is sensitive to radio frequency interference. Electromagnetic fields produced by items including, but not limited to, microwave ovens, Wi-Fi® routers, cellular phones, and personal communication devices, i.e., walkie-talkie, two-way radios, etc. have the potential to adversely affect the CO2 measurement. To ensure a proper measurement these devices should be TURNED OFF or placed in a mode that does not transmit. If these devices are not turned off or placed in a mode that does not transmit, the accuracy of the CO2 measurement may be compromised if these devices are located within four meters of the CO2 measurement sensor.
Page 12
xii Q-Trak™ XP IAQ Monitor Model 7585

Labels

Advisory and identification labels or markings are attached to the instrument.
1. Multi-sensor module
2. Base handle
3. Carbon Dioxide sensor
4. Chlorine sensor
5. Carbon Monoxide sensor
6. Hydrogen Sulfide sensor
7. Ammonia sensor
8. Nitric Oxide sensor
Page 13
Operation and Service Manual xiii
9. Nitrogen Dioxide sensor
10. Ozone sensor
11. TVOC ppm sensor
12. TVOC ppb sensor
13. Formaldehyde sensor
14. European symbol for non­disposable item. Item must be recycled.
15. Battery pack
Page 14
xiv Q-Trak™ XP IAQ Monitor Model 7585

Reusing and Recycling

As part of TSI® Incorporated’s effort to have a minimal negative impact on the communities in which its products are manufactured and used:
DO NOT dispose of use batteries in the trash.
Follow local environmental requirements for battery recycling.
If instrument becomes obsolete, return to TSI®
for disassembly and recycling.
Page 15
1
CHAPTER 1
Overview
The Q-Trak™ XP Indoor Air Quality (IAQ) Monitor is designed for Indoor Air Quality and Industrial Hygiene professionals to address a wide range of indoor air quality assessments and analysis. The Q-Trak™ XP monitor combines multiple-gas and particle measurements into a single lightweight, handheld instrument that is easy to configure, maintain, and calibrate in the field. The Model 7585 includes sensors for simultaneous measurements of temperature, relative humidity, barometric pressure, mass concentration, particle concentration, carbon dioxide, and room for up to five additional pluggable gas sensors. With enough on-board memory to record data from all sensors for 30 days when sampling data once a minute.
Also included with the Q-Trak™ XP monitor is TrakPro™ Ultra software application used for posttest analysis and report generation. Refer to Chapter 10 for more information.
These Application Notes for the Q-Trak™ can be found under TSI’s web site at www.tsi.com/7585.
Page 16
2 Q-Trak™ XP IAQ Monitor Model 7585
(This page intentionally left blank)
Page 17
3
CHAPTER 2
Unpacking and Parts Identification
Carefully unpack the instrument and accessories from the shipping container. Check the individual parts against the list of components below. If anything is missing or damaged, notify TSI® immediately.
All standard equipment can be purchased separately if needed.
NOTE
Optional gas sensors are shipped in individual boxes but are included in the main shipping container.
NOTE
The lithium-ion battery is shipped in the same box as the instrument; however, it cannot be pre-installed in the Q-Trak™ XP instrument prior to shipment as required by law.

Standard Equipment

Part No.
Description
Picture
7585
Includes Q-Trak™ XP Base Meter with Multi-Sensor IAQ Module and the following sensors:
• PM 2.5 sensor
• Temperature and Relative
Humidity sensor
• Barometric Pressure sensor*
• Carbon Dioxide (CO2) sensor
(801399)
IAQ module includes six configurable gas sensor slots
7580
Base Meter only – Handle portion
Page 18
4 Q-Trak™ XP IAQ Monitor Model 7585
Part No.
Description
Picture
801430*
IAQ Multi-Sensor Gas Module with Built-in Sensors: Temperature, Relative Humidity, Barometric Pressure, Particle
801399
CO2 (Carbon Dioxide), NDIR (Nondispersive Infrared sensor)
800121
Carrying Case
800123
Lithium Ion Rechargeable Battery Pack
804001
USB Cable
800122
AC Adapter/Power Supply
N/A
Calibration certificates
N/A
6013907
Q-Trak™ XP User Manual (English: Included on USB drive)
6013908
Q-Trak™ XP Quick Start Guide
(includes French, German, Simplified Chinese, and Traditional Chinese)
Page 19
Operation and Service Manual 5
Part No.
Description
Picture
800120
Gas Sensor Calibration Cap 7004280
USB Flash Drive with Q-Trak™ XP manuals and other literature. Included with 7585.

Optional Gas Sensors

Part No.
Description
Picture
801399
CO2 (Carbon Dioxide), NDIR (Nondispersive Infrared sensor)
(801403)
801400
Cl2 (Chlorine), EC (Electrochemical Sensor)
801401
CO (Carbon Monoxide), EC (Electrochemical Sensor)
801402
H2S (Hydrogen Sulfide), EC (Electrochemical Sensor)
801403
NH3 (Ammonia), EC (Electrochemical Sensor)
801404
NO (Nitric Oxide), EC (Electrochemical Sensor)
801405
NO2 (Nitrogen Dioxide), EC (Electrochemical Sensor)
801406
O3 (Ozone), EC (Electrochemical Sensor)
801407
TVOCH (VOC ppm), PID (Photoionization Detection Sensor)
801408
TVOCL (VOC ppb), PID (Photoionization Detection Sensor)
801409
CH2O (Formaldehyde), EC (Electrochemical Sensor)
Page 20
6 Q-Trak™ XP IAQ Monitor Model 7585

Optional Accessories

Part No.
Description
Picture
800124
Wi-Fi® dongle
800129
Q-Trak™ XP Battery Cover with Tripod Mount and Tabletop Tripod

Other Replacement Parts

Part No.
Description
Picture
800120
Q-Trak™ XP Gas Sensor calibration cap
800121
Q-Trak™ XP Case: hard sided carry case
800126
Battery Cover
Page 21
Operation and Service Manual 7
Part No.
Description
Picture
800127
Sensor Module Cover
800125
Battery cover with tripod mount
800128
Tabletop Tripod
N/A
Page 22
8 Q-Trak™ XP IAQ Monitor Model 7585
(This page intentionally left blank)
Page 23
9
CHAPTER 3
Setting Up the Q-Trak™ XP Monitor

Installation of Gas Sensors

In addition to the standard built-in sensors, the device can accept up to six gas sensors.
Install the gas sensors as follows:
1. Ensure the instrument is powered OFF.
2. Remove the sensor module cover.
3. Position the sensor as shown, making certain the pins make good contact with the device. Gas sensors can be installed in any available slot. Listen for a click to confirm sensor is connected properly.
4. Attach cover to the sensor module.
5. To configure which sensors are shown on the display, refer to Sensors section in Chapter 5.
Page 24
10 Q-Trak™ XP IAQ Monitor Model 7585

Providing Power to the Q-Tra k™ XP

The Q-Trak™ XP monitor can be powered with provided lithium-ion battery or with the provided A/C adapter.

Operating the Instrument with the AC Adapter

The instrument can be operated with or without a battery using the A/C adapter.
To operate the instrument with AC adapter:
1. Plug the AC adapter into an electrical outlet.
2. Plug the other end into the instrument.
NOTE
In general, TSI® recommends operating the Q-Trak™ XP monitor with a battery even when powering the instrument with an AC power supply. Having a battery installed shortens the warm-up time needed for sensors to make accurate measurements.
NOTE
For best results, power on the instrument for 10 minutes before verifying gas sensor performance. This allows the instrument and sensors to thermally stabilize and reduces the amount of measurement drift during verification or calibration.
Page 25
Operation and Service Manual 11

Operating the Instrument with the Lithium-Ion Battery

To install the battery:
1. Remove the battery compartment cover.
2. Connect the battery to the handle connector.
NOTE
The battery pack connector is designed to prevent improper connection.
3. Insert the battery as shown in the picture, keeping the battery wires to the side and clear of the battery cover screw.
Page 26
12 Q-Trak™ XP IAQ Monitor Model 7585
4. Attach the battery compartment cover, ensuring battery wires are clear of the screw.
IMPORTANT
Always calibrate the battery Power Gauge after installing a battery pack.

Charging the Battery and Calibrating the Power Gauge

1. With the battery installed, turn on the instrument by pressing the power button and run until the battery is fully discharged. The instrument will automatically turn off when the battery is fully
discharged.
NOTE
The battery is shipped with approximately 25% to 30% charge.
2. Connect the AC adapter to the instrument and fully charge the battery. This may take up to 3 hours. Note the lightning bolt
indicator when charging vs fully charged .
NOTE
It is normal for the battery to feel warm as it is charging.
3. The instrument is now ready for use with the battery. This Power Gauge calibration procedure does not need to be repeated unless the battery is replaced or disconnected.
Page 27
Operation and Service Manual 13
Installing and Connecting to Available Wi-Fi®
(optional Wi-Fi® dongle
To connect your device to a Wi-Fi® network, proceed as follows:
1. Install the dongle into the USB port located inside the battery compartment.

Connecting to a Computer and Installing Tra kPro™ Ultra Software

To connect the Q-Trak™ XP monitor to a computer for downloading data files or when using the TrakPro™ Ultra Software application:
1. Connect the USB-C cable provided with the Q-Trak™ XP monitor to a computer.
2. From a computer, navigate to www.tsi.com/7585 and select the TrakPro™ Ultra application exe.
3. Download and install the application.
4. Launch the TrakPro™ Ultra application.
Page 28
14 Q-Trak™ XP IAQ Monitor Model 7585
5. Refer to of the user guide in “help” menu within TrakPro™ Ultra software for operation information.
Page 29
15
CHAPTER 4
Operational Overview

Powering ON the Q-Trak™ XP Monitor

Press and release the power button . A progress bar will appear as the instrument boots.
To power down the instrument, press the power button and press shutdown from the
pop-up screen The first time the Q-Trak XP
monitor is powered ON the
Welcome to Q-TrakTM XP
page is presented. Select the appropriate Language and Time Zone then select DONE.
After selecting DONE the Dashboard page is displayed in Survey mode.
Page 30
16 Q-Trak™ XP IAQ Monitor Model 7585
Survey mode displays real-time readings of sensors selected from the Sensors page discussed later in Chapter 5. It does not log measurements or statistics.
Selecting Graph presents real-time readings in a graphical format while selecting Information present data in a numerical format. To view sensors not shown on the page, swipe up and down.
Page 31
Operation and Service Manual 17
NOTE
Sensors will only appear in the Dashboard page once they have been selected in the Sensors page discussed later in Chapter 5.

Navigating the Dashboard Page+

The following Dashboard diagram dissects the layout of the Dashboard page.
Page 32
18 Q-Trak™ XP IAQ Monitor Model 7585

Main Menu

To display the Main Menu, select the icon in the upper left corner of the header on any page
of the display. The Main Menu has seven
options:
• Dashboard
• Settings
• Manage Data
• Calibration
• Workflows
• Device Information
• Help
Page 33
Operation and Service Manual 19

Dashboard

This is the main page for viewing live readings, live graphs and logging data.
Select Dashboard from the menu in the header any time to return to the Dashboard (Home) page.

Settings

Select Settings from the Main Menu, to view the Settings page options. Refer to
Chapter 5, Settings for detailed
information about the device setting options.
Page 34
20 Q-Trak™ XP IAQ Monitor Model 7585

Manage Data

Select Manage Data to display logged data stored in the device. Refer to Chapter 8, Manage
Data for detailed information.

Calibration

Sensors are calibrated before they are shipped from TSI. If a sensor requires a field calibration, select Calibration from the Main Menu. The Calibration page presents the sensors (except the particle sensor) installed in the device.
NOTE
A user calibration cannot be applied to particle measurements.
Page 35
Operation and Service Manual 21

Workflows Percent of Outside Air: Select Workflows from the Main Menu,

and then % Outside Air to begin performing a % Outside Air Calculation. Refer to Chapter 9, Workflows for details.
Page 36
22 Q-Trak™ XP IAQ Monitor Model 7585

Device Information

To display general information about the Q-Trak™ XP monitor, select Device Information from the Main Menu. The Device Information page lists the device name, IP address, model number, and many other characteristics of the device as well as information about the installed sensors.
To view sensors not shown on the page, swipe up and down. This information is important for troubleshooting issues related to the operation of the device. Furthermore, the Device Information page includes the ability to update the software version for the Q-Trak™ XP monitor from a flash drive or computer.
To update the Q-Trak™ XP monitor software, select the down arrow from the Device Information page and select the latest software version then select the Update button. Refer to Chapter 11, Update
Instrument Software for more information.
Page 37
23
CHAPTER 5
Settings
Navigate to the Settings page by selecting Settings from the Main Menu.
The Settings page options are:
• General Settings
• Date/Time
• Display/Power Management
• Sensors
• Network Settings
• Units of Measure
• Notifications/Alarms
These options are described in the following subsections.
Page 38
24 Q-Trak™ XP IAQ Monitor Model 7585

General Settings

The General Settings page is used to configure the following functions:
• Device Name is used to customize the name of the instrument. This feature is helpful when associating data to a specific instrument.
NOTE: The device will display up to 14 characters of the device name in the header of the main Dashboard page.
• Language is used to select the desired language shown on the instrument.
• Time Constant is used to adjust the averaging period for the readings shown on the display.
• Sound is used to enable or disable the beeper. A beep is emitted at the end of a Sample Interval when logging data or when a Notification/Alarm occurs.
• Density Correction enables or disable the application of density correction to the measurement of CO2 and VOC.
• Show Out-of-Range Readings enables or disables out- of- range readings shown on the device display. An out-of­range reading is a measured value that is either less than
a sensor’s Minimum Detection Limit or greater than the
upper recommended range. For example, a negative gas concentration reading is considered an out-of-range reading. It is recommended that this setting is Enabled when performing a Bump Test.
Page 39
Operation and Service Manual 25
NOTE
Time Constant is the display averaging period. The display will update every second; however, the displayed reading will be the average over the time constant period. For example, if the time constant is 5 seconds, the display will update every second, but the displayed reading will be the average of the last 5 seconds.

Date/Time

The Date/Time page is used to set the current date, time, and Time Zone. This information is critical for logging data and creating log schedules.
Page 40
26 Q-Trak™ XP IAQ Monitor Model 7585

Display/Power Management

The Display/Power Management page is used to
configure the following functions
• Display Brightness adjusts the brightness of the display.
• Display Sleep sets the time when the instrument enters into sleep mode. Sleep mode is a power-saving state that darkens the display. To exit sleep mode touch the display. The selectable sleep time options are 1, 5, and 15 minutes. By default, Display Sleep is 1 minute.
• Enable Auto Sleep enables and disables sleep mode.
• Require Passcode enables and disables the passcode protection feature. Once the instrument enters sleep mode a passcode is required to regain access (passcode is 7585).
Page 41
Operation and Service Manual 27

Sensors

The Display section of the Sensors page is used to
configure what sensors are displayed on the Dashboard page, and the order they are displayed. To view sensors not shown on the page, swipe up and down. The list of sensors includes the following information:
• A sensor icon and name.
• A toggle button to enable
or disable the Visibility of the sensor on the Dashboard page.
• Up and down arrow buttons to set the
order of the sensors on the Dashboard page.
NOTE
The order of sensors on the Sensors page is the order shown on the Dashboard page.
Page 42
28 Q-Trak™ XP IAQ Monitor Model 7585
The VOC Configuration section of the Sensors page is used to select a specific Volatile Organic Compound (VOC). This is done by selecting an already pre­defined VOC in the drop-down list or by selecting Custom and entering a Response Factor and Molecular Weight.
• The Response Factor and Molecular Weight are used
to calculate the actual concentration of a specific VOC.

Network Settings

The Network Settings page is used to connect the Q-Trak™ XP monitor to a wireless network.
1. Select the Connect to Network radio button.
2. Select a Wireless Network down arrow to view the list of available networks.
3. Select the desired network.
Page 43
Operation and Service Manual 29
4. Enter the network password in the Network Password field.
5. Select the CONNECT button to connect the Q-Trak™ XP monitor to the network.
6. Once connected the Connection Status, IP Address, Subnet Mask, Connection Type, Default Gateway, and MAC Address will populate.
7. Select the Done button to exit the Network Settings page.

Units of Measure

The Units of Measure page is used to set the units of measure for the installed sensors.
1. Select icon to view the list of available units of measure.
2. Select the desire unit of measure.
3. Select SAVE button to save changes.
Page 44
30 Q-Trak™ XP IAQ Monitor Model 7585
NOTE
The units of measure for all PM mass concentration measurements (PM 1, PM 2.5, and PM 10) are set by the PM drop-down list. Similarly, the units of measure of all particle concertation measurements (PC 0.3, PC 0.5, PC 1.0, PC 2.5, PC 5.0, PC 10.0) are set by the Particle Conc. Drop-down list.

Notifications/Alarms

The Notifications/Alarms page is used to enable and disable notifications/alarms of the list of sensors as well as setting their trigger thresholds. The following is included in this page.
• Toggle buttons to enable or disable Notifications / Alarms.
• Text fields to enter
Notifications / Alarms
thresholds.
• Lastly, to view sensors not shown on the page, swipe up and down.
NOTE
If two or more of the same sensor type is installed, only one will trigger an alarm.
Page 45
31
CHAPTER 6
Calibration

Gas Sensors

TSI® recommends testing the accuracy of gas sensor measurements each time the instrument is used. This is a common practice within the gas measurement community which includes IAQ consultants and industrial hygienists. Due to the nature of the sensors along with other environmental factors such as altitude, temperature, and relative humidity, the zero value and sensor sensitivity can shift over time requiring you to perform a field calibration.

Gas Sensor Verification

Verifying the accuracy and performance of a sensor can easily be accomplished in the field by comparing it to certified ZERO and SPAN calibration gases (also referred to as a “Bump Test”). To do this, attach a calibration cap to a bottle of the recommended ZERO gas and place the cap onto the gas sensor being checked. If the probe reads a number outside of the sensor accuracy after the measurement has stabilized, the sensor has drifted and needs to be recalibrated. Perform the same procedure with the recommended SPAN gas to check the sensor measurement against the certified SPAN bottle concentration and determine if a user calibration is necessary. See Table 1 for recommended ZERO and SPAN gases.
Page 46
32 Q-Trak™ XP IAQ Monitor Model 7585
NOTE
For best results, power on the instrument for 10 minutes before verifying gas sensor performance. This allows the instrument and sensors to thermally stabilize and reduces the amount of measurement drift during verification or calibration.
NOTE
Specific instructions for the calibration of each sensor are available during calibration. To access the help for a sensor, select and a pop-up appears with information to guide you through specific sensor calibrations.
The following table shows the recommended calibration gases and their respective concentrations (in ppm) for each gas sensor.
Calibration gases are available globally through local distributors.
Table 1: Calibration Gases and Concentrations
Carbon Dioxide (CO2)
Carbon Dioxide (CO2)
5000
Nitrogen
TVOC – High
Isobutylene (C4H8)
100 2000
Zero Air or Nitrogen
100 ppm is the first span
2000 ppm is the second span
TVOC - Low
Isobutylene (C4H8)
20
Zero Air or Nitrogen
Carbon Monoxide (CO)
Carbon Monoxide (CO)
400
Zero Air
Nitrogen Dioxide (NO2)
Nitrogen Dioxide (NO2)
20
Nitrogen
Nitric Oxide (NO)
Nitric Oxide (NO)
20
Nitrogen
Chlorine (Cl2)
Nitrogen Dioxide (NO2)
20
Zero Air
Nitrogen Dioxide (NO2) is used as a surrogate SPAN calibration gas
Ozone (O3)
Nitrogen Dioxide (NO2)
20
Zero Air
Nitrogen Dioxide (NO2) is used as a surrogate SPAN calibration gas
Page 47
Operation and Service Manual 33
Hydrogen Sulfide (H2S)
Hydrogen Sulfide (H2S)
50
Zero Air Ammonia (NH3)
Ammonia (NH3)
100
Zero Air
Formaldehyde (CH2O)
Formaldehyde (CH2O)
10
Zero Air

Gas Calibration Procedure

To perform a gas calibration, required accessories and material include a calibration cap, ZERO calibration gas, SPAN calibration gas, gas regulator, and tubing. The gas regulator used to control the flow should be capable of providing 0.3 L/min.
Select Calibration from the Main Menu and the Calibration page will appear showing the available sensors for calibration. All sensors appear grayed out until you select the sensor they would like to calibrate.

Step 1 – Select a Gas Sensor for Calibration

Remove the sensor module cover.
Page 48
34 Q-Trak™ XP IAQ Monitor Model 7585

Step 2 – Place the calibration cap onto the gas sensor you want to calibrate

Step 3 - Select the sensor to be calibrated on the instrument display

NOTE
You have the ability to “SKIP” either the ZERO and/or SPAN gas calibration steps during the calibration process.
Page 49
Operation and Service Manual 35

Step 4 - Perform ZERO Capture

1. Connect the ZERO gas regulator to the calibration cap and start the flow of ZERO gas. Wait for the measurement to stabilize.
2. Once the ZERO concentration looks stable, select Next to perform a ZERO concentration capture. An illuminated dial will appear on the sensor icon showing the time remaining.

Step 5 – Perform SPAN Capture

1. Enter the SPAN gas bottle concentration into the reference concentration field using the numeric keypad.
2. Connect the SPAN gas regulator to the calibration cap and start the flow of SPAN gas. Wait for the measurement to stabilize.
3. Once the span concentration looks stable, select Next to perform span concentration capture. An illuminated dial will appear on the sensor icon showing the time remaining.
Page 50
36 Q-Trak™ XP IAQ Monitor Model 7585

Step 6 – Save or Discard Calibration

A CALIBRATION ADJUSTMENT % is displayed after the SPAN
capture that informs you how
much the sensor’s calibration
slope has changed from the factory calibration slope.
1. Select Save to accept the calibration.
2. Select Cancel to quit the calibration procedure and discard calibration.
3. A successful calibration will be indicated by a near
the sensor icon.
4. If the calibration reading completes but the Reference Calibration gas concentration is out-of-range than what is recommened by the device, the page displays a near the sensor icon and a warning notice is sent to the Notice list in the heading bar of the instrument.
An Out-of-Range error during the calibration of a sensor means
the sensor’s offset or calibration slope adjustment has drifted
outside of the TSI® recommended specification for continued use. Sensor should be replaced or sent to TSI® for re-calibration.
Page 51
Operation and Service Manual 37

Gas Calibration Unsuccessful

If a sensor has drifted beyond the TSI® recommended calibration adjustment specification, a is displayed by the sensor icon and a
warning notice is sent to the Notice list in the heading bar of the instrument. TSI® recommends replacing the sensor. However, the sensor is calibrated and can still be used. See Out of Range in
Appendix B, Definitions for more information.

Built-in Sensor Calibration: Temperature, Relative Humidity, and Barometric Pressure Sensors

To calibrate the Temperature, Relative Humidity, and Barometric Pressure sensors:
Select either the Temperature, Relative Humidity, or Barometric Pressure sensor icon on the Calibration page. A Reference Value for the sensor will appear.
Page 52
38 Q-Trak™ XP IAQ Monitor Model 7585
The Reference Value is the current reading of the sensor. This reading can be changed by selecting on the Reference Value text
field and entering the desired value and select < >. Select Next then Save and the new value is displayed as the Reference Value.
The Total Offset is the adjustment from the original factory calibration for that sensor. This lets you track the overall drift of the sensor.
If the drift exceeds the factory allowable tolerance, a is displayed by the sensor icon and a warning notice is sent to the Notice list in the heading bar of the instrument that instructs you to return the
unit to TSI® for service. Calibration that is within the factory allowable tolerance, a is
displayed by the sensor icon.
Page 53
Operation and Service Manual 39

Reset Sensors to Default Calibration

To reset the sensors to their factory defaults,
1. Return to the Calibration page and select Reset to Factory Calibration at the bottom of the page.
2. Select the sensor(s) to be reset.
3. When the selected sensor(s) are highlighted, select RESET.
4. Select Yes or No. Selecting Yes resets the sensor(s) to the factory calibration.
Page 54
40 Q-Trak™ XP IAQ Monitor Model 7585
(This page intentionally left blank)
Page 55
41
CHAPTER 7
Logging Data
From the Main Menu select Dashboard then select > Logging at the top of the
page. This leads to the Dashboard page with the Start button as shown below.
From this page either select Start to begin logging data based on the selected Logging
Profile or define a Logging Profile by selecting .
NOTES
When selecting START before creating a Logging Profile the Q-Trak™ will log data based on a default Logging Profile and log a single reading from all available sensors using a sample interval of 1 second.
Page 56
42 Q-Trak™ XP IAQ Monitor Model 7585

Logging Data

To begin logging data, select START.
When logging is initiated, the Dashboard page displays second by second measurement readings next to the sensor icons while the Avg value is the average of all readings in the log file.
NOTE
On-board memory is capable of storing data from all available sensors for 30 days when sampling data once a minute.
You can log data with Manual
Save, Continuous Save, or Scheduled Save (see figure). To
stop logging at any time, select
STOP then select Yes.
NOTE
The Q-Trak™ always logs data from all available sensors.
Page 57
Operation and Service Manual 43

Logging Profiles

Selecting the from the Dashboard page leads to the Logging Profile page where
you define and select a logging profile. A logging profile is composed of a Locations and
Settings such as Logging Mode. If Scheduled Save Logging Mode (see figure) is
selected then a schedule must be defined.
The following subsections describe how to add/change
Locations, Settings, and Schedules.

Locations

A location is composed of a Location Name, typically a building, and an Area/Room in the building. The highlighted location is the currently selected location. To change the current location, select another location in the list.
NOTE
The list of locations can exceed the number of locations that are visible on the display. To view additional locations swipe up or down.
On the Logging Profile page select Locations.
Page 58
44 Q-Trak™ XP IAQ Monitor Model 7585
To create a new location:
1. Select to create a new location.
2. Enter a Building Name.
3. Enter Area/Room.
4. Select and select Done to add a location.
5. Select Save to save the location.
To edit an existing location:
1. Select and select Edit for the location.
2. Edit the Building Name and/or Area/Room for the Location.
3. Select and select Done to complete editing.
4. Select Save to save the change. To delete a location:
1. Select and select Delete for the Location.
2. Select SAVE to delete the location.
Page 59
Operation and Service Manual 45

Settings

Settings include Logging Mode, Sample Interval, Test Duration, and other items such as Select Logging Schedule.
Logging Mode
There are three logging modes:
• Manual Save
• Continuous Save
• Scheduled Save
Manual Save logging mode averages measurements collected during the sample interval and records the average value in the log file. Once the sample is complete the application prompts you to either select START or DONE. Select START to take another sample or DONE to end the logging session.
Selecting the View Before Save toggle button to ON, Manual Save prompts you to either SAVE or DISCARD once the sample is complete. Swipe up or down to view the collected data then select
SAVE or DISCARD to save or discard the data. After selecting SAVE or DISCARD the page will display START or DONE. Select START to take another sample or DONE to end the logging session.
Continuous Save logging mode averages measurements collected
during the sample interval and records the average value in the log file. Unlike Manual Save mode, Continuous Save mode continues logging data until the Time Duration time has elapsed. In this mode, the sample data is automatically saved to memory at the end of the sampling interval and you are not given the option to view the data before saving.
Page 60
46 Q-Trak™ XP IAQ Monitor Model 7585
Schedule Save collects data according to a user-defined schedule with a specific start and end date and time. Moreover, this logging mode averages measurements collected during the sample interval and records the average value in the log file like the previous logging modes. Lastly, the data is automatically saved without user intervention.

Sample Interval

Sample interval is used in all three logging modes to set the time between samples. It can be as short as one second or as long as 99 hours, 59 minutes, and 59 seconds.

Schedules

To set up schedules for logging, Select Schedules from the Logging Profile page.
Page 61
Operation and Service Manual 47
To create a new logging schedule:
1. Select to create a logging schedule.
2. Enter a schedule name.
3. If this is a recurring logging profile, select Recurring and select the days of the week to log data.
4. Enter start times and end times as desired.
5. If an additional time segment is desired, select Add New Segment and enter a start time and end time.
6. Repeat step 5 to add more time segments.
7. Add Notes to the schedule if desired.
8. Select DONE to finish editing and select SAVE to save changes.
NOTE
The list of time segments can exceed what is shown on the display. To view additional time segments swipe up and down.
To edit an existing logging schedule:
1. Select and then select Edit.
2. Modify the schedule as desired.
3. Select DONE to finish editing and select SAVE to save changes.
To delete an existing logging schedule:
1. Select and select Delete.
2. Select Yes.
Page 62
48 Q-Trak™ XP IAQ Monitor Model 7585
(This page intentionally left blank)
Page 63
49
CHAPTER 8
Manage Data
From the Main Menu select Manage Data. The Manage Data page contains all logged files on the device. You
can select a log file for viewing, deleting, exporting, and editing. As well as scrolling through a list of log files.

Selecting Log Files

Select a log file by selecting a checkbox. Multiple log files can be selected by selecting additional checkboxes while all log files can be selected by selecting the checkbox at the top of the page.
Page 64
50 Q-Trak™ XP IAQ Monitor Model 7585

View Log Files

To view a log file select the checkbox then select the
icon. To view all the data in a log file swipe up and down and to view data over time select Graph. If more than one log file is selected for viewing, select the up and down arrows to view the other log files.

Delete Log Files

Select a log file then select the icon to delete it. Next select YES.
NOTE
Deleted log files cannot be recovered.
Alternatively, when viewing a log data file select then select the
icon.
Page 65
Operation and Service Manual 51

Export Log Files

Logged data files can be exported to a USB flash drive or to a PC via TrakPro™ Ultra Software.
Exporting to USB Drive
Attach a USB flash drive (thumb drive) to the Q-Trak™ monitor via the USB port next to the power jack. Select a log file from the menu and then select icon.
Alternatively, when viewing a logged data file, select the menu and then select to export the file to the flash drive.
Files will be transferred to the USB drive in .CSV format for use with spreadsheets on a computer.
Exporting Files to a PC
Connect the instrument to a computer via Wi-Fi® or the USB-C Cable supplied with the instrument.
Open TrakPro™ Ultra Software on the PC (see TrakPro™ Ultra Software user manual) and select a log file from the menu and then select icon.
Alternatively, when viewing a logged data file, select the menu and then select to export the file to the PC in a .db File format for use with TrakPro™ Ultra Software. Files can then be exported as .csv or other file formats for use with other software.
NOTE
If exporting data to a USB flash drive, and there is not enough space on the drive to house the file, or if some other error occurs during the export, the following warning message will appear: “Data Export Failed – Please Check USB Drive.” This warning will persist unit it is dismissed.
Page 66
52 Q-Trak™ XP IAQ Monitor Model 7585

Edit Log File Names

When viewing a log file, select the icon and then select Edit.
Enter a new Log File Name and add a comment if desired.
Scrolling Through Data List
When the list of log files exceeds what can be viewed on the display, select icons to scroll through the list of log files.
Page 67
53
CHAPTER 9
Workflows
The following workflows are available:
• % Outside Air (%OA) calculation procedure.

Percent Outside Air (%OA) Calculation Procedure

For those unfamiliar with Percent Outside Air (%OA) Calculation and its use, please read TSI® Application Note TI-138, which is found on the TSI® website at
https://www.tsi.com. It is
important to understand what % Outside Air is, why it is important, and how to use this instrument to calculate %OA.
After selecting Workflows from the Main Menu, select
%Outside Air to view the % Outside Air room.
Page 68
54 Q-Trak™ XP IAQ Monitor Model 7585
Three measurements are necessary for % Outside Air calculations - Outside Air, Supply Air, and Return Air. The measurements may be taken in any order. The calculation is performed once the final measurement is taken.
Before taking a measurement select icon to create a
% Outside Air Profile. From the % Outside Air Profile
page, you can:
• Select a CO2 or Temperature as the Measurement Type. If
necessary, review the application note TI-138 to determine which option to use.
• Add a new location by selecting icon and enter
a Building Name and Area/Room. Similar to what was done in Chapter 7
Logging Data.
• Edit or Delete existing logging profile locations by
selecting icon. Similar to what was done in Chapter 7
Logging Data.
• Select Save or Cancel as desired to return to the %Outside Air page.
Page 69
Operation and Service Manual 55
As mentioned above, three measurements are necessary for % Outside Air calculations - Outside Air, Supply Air, and Return Air. The measurements may be taken in any order. The calculation is performed once the final measurement is taken.
IMPORTANT
The Q-Trak™ XP instrument is intended for indoor use only. Care should be taken when taking % of Outdoor Air measurements so that the instrument is not exposed to rain, sleet, hail, snow or other inclement weather conditions as exposure to these elements will void the factory warranty.
1. Position the Q-Trak™ monitor outside to perform the Outside Air measurement, select START, and wait until the measurement is complete.
2. The % Outside Air page shows the number of outstanding readings needed before a calculation can be made.
3. Position the Q-Trak™ monitor in the supply vent to perform the Supply Air measurement, select START, and wait until the measurement is complete.
4. Lastly, position the Q-Trak™ monitor by the return vent to perform the Return Air measurement, select START, and wait until the measurement is complete.
5. If a measurement (Outside Air, Supply Air, or Return Air) needs repeating select CLEAR then select START.
6. Once the measurements are complete, the % Outside Air is calculated and shown on the bottom of the page.
Page 70
56 Q-Trak™ XP IAQ Monitor Model 7585
7. Select DISCARD to discard all measurements or select SAVE to save the % Outdoor Air Calculation.
8. After saving the % Outdoor Air Calculation you can perform another % Outdoor Air calculation or select the
icon to exit the %
Outside Air page.
Instead of making measurements, you have the option of entering values for
Outside Air, Supply Air, or Return Air by selecting --- and
typing the appropriate value. Select DONE when finished. The
calculation is then made using entered value(s).
Lastly, if the % Outside Air calculation cannot be made using the measurements / entered values, an error message is shown along with a DISCARD button.
Selecting DISCARD clears all measurements/entered values and you may start over.
Alternatively, you can select Clear to clear a single measurement/entered value and repeat taking the measurement/enter a new value.entered value(s).
Page 71
57
CHAPTER 10
TrakPro™ Ultra Software
Post-test analysis and report generation is performed using the TrakPro™ Ultra software application.
To install TrakPro™ Ultra Software:
1. Using a web browser, navigate to https://tsi.com/support/tsi-
software-and-firmware/ and enter the search term Q-Trak XP.
2. Download the TrakPro™ Ultra software application by following the on-screen instructions.
NOTE
The manual for TrakPro™ Ultra Software is embedded in the application.
Page 72
58 Q-Trak™ XP IAQ Monitor Model 7585
(This page intentionally left blank)
Page 73
59
CHAPTER 11
Update Instrument Software
The Device Information page allows you to update the Q-Trak™ XP software without having to return the instrument to TSI® Incorporated.
To update the firmware:
1. Access the TSI® website site at www.tsi.com/7585 using a computer.
2. Select the latest software version and download it to a USB thumb drive.
3. Insert the USB drive into the USB port on the end of the Q-Trak™ XP instrument and turn the instrument on.
4. Select Device Information from the Main Menu.
5. Select the drop-down arrow to view all available software versions.
6. Select the latest version of the software.
7. Select the UPDATE button.
8. The instrument will shut down while updating. Allow the update to complete before attempting to perform any operations.
9. The instrument will reboot after updating the software and is ready to use with the updated software.
Page 74
60 Q-Trak™ XP IAQ Monitor Model 7585
(This page intentionally left blank)
Page 75
61
CHAPTER 12
Maintenance

Replacing Sensors

Replacement sensors are sold separately and available through TSI® Incorporated. Refer to Chapter 2 for replacement parts.

Recalibration

See Chapter 6. Calibration for information on how to calibrate sensors.

Removing the Sensor Module from the Handle of the Q-Trak™ XP Monitor.

Page 76
62 Q-Trak™ XP IAQ Monitor Model 7585
To remove the sensor module from the handle of the instrument, proceed as follows:
1. Turn the instrument OFF.
2. Turn the instrument over and press upward on the release tab just above the back-cover screw as shown in the picture.
3. Turn the instrument over and hold it as shown while continuing to press on the release tab as previously described.
4. Press downward with both thumbs while holding the release tab above the back­cover. This requires some steady force.
Page 77
Operation and Service Manual 63
5. Continue to press down until the sensor module is free from the handle.
6. Separate the sensor module from the handle.
7. Separation complete.

Storage

When storing the instrument for extended periods, remove the battery after fully charging before storage. Also, remove the sensors and store them in their original “snap cap” containers. When re-installing the battery, refer to the initial battery instillation process for calibrating the battery status gauge.
Page 78
64 Q-Trak™ XP IAQ Monitor Model 7585
(This page intentionally left blank)
Page 79
65
CHAPTER 13
Troubleshooting
Table 2 lists the symptoms, possible causes, and recommended solutions for common problems encountered with the Model 7585. If your symptom is not listed, or if none of the solutions solves your problem, please contact TSI® Incorporated.
Table 2: Troubleshooting the Model 7585
Symptom
Possible Causes
Corrective Action
Blank display Depleted battery
Plug in AC adapter to charge battery
Damaged display unit
Return to TSI® for service
Battery will not charge
Damaged battery
Replace battery
Battery disconnected
Connect battery in battery compartment
Touch screen display not responding
Instrument software issue
Press the power button for 10 to 12 seconds to turn off the instrument. Then press the power button to turn the instrument back on.
Damaged display unit
Return to TSI® for service.
Gas sensor is installed, but not visible in the Sensors page
Sensor was installed while the instrument was powered.
Power cycle the instrument.
Damaged sensor
Install sensor in another location in the sensor module. If the sensor continues not be detected. Contact TSI® to service or replace the sensor. If the warranty period has expired it may be necessary to purchase a new sensor.
Page 80
66 Q-Trak™ XP IAQ Monitor Model 7585
Symptom
Possible Causes
Corrective Action
Gas sensor is installed, but not visible in the Sensors page
(cont.)
Damaged gas sensor module
Install sensor in another location in the sensor module. If the sensor is detected. Contact TSI® to service or replace the sensor module. If the warranty period has expired it may be necessary to purchase a new sensor module.
Gas sensor fails calibration
Wrong gas is being used to calibrate sensor.
Verify the correct gas is being used.
Calibration cap is attached to wrong sensor.
Verify the calibration cap is attached to the correct sensor.
Fan no longer rotating
Fan is jammed by dust, dirt, or other contaminant
Clean the fan while instrument is powered off with compressed air
Damaged fan or module
Return to TSI® for service
Sensors not visible on the Dashboard
Sensor not turned on in the Sensors page.
From the Main Menu select Settings then select Sensors. Once in the Sensors page select the appropriate toggle button to make the sensor visible on the Dashboard.
Density correction warning
Faulty Barometric Pressure or Temperature sensor(s)
Return to TSI® for Service
From the Main Menu select
Settings then select General Settings. Once in the General Setting page select Density Correction toggle button to
turn off Density Correction.
Page 81
Operation and Service Manual 67
Symptom
Possible Causes
Corrective Action
While powering the Q-Trak™ XP monitor the instrument unexpectedly powers off.
Battery power is fully discharged.
Charge battery with supplied power supply.
Replace battery.
Battery temperature sensor measures a temperature greater than 60°C.
Place the Q-Trak™ monitor in a cooler environment and wait approximately 15 minutes. Then power instrument.
Replace battery.
Battery temperature sensor measures a temperature less than -20°C.
Place the Q-Trak™ monitor in a warmer environment and wait approximately 15 minutes. Then power instrument.
Replace battery.
Unable to connect
TrakPro™ Ultra
software to Q--Trak™ XP monitor via USB-C cable.
USB-C cable is damaged.
Use another USB-C cable.
Q-Trak™ XP monitor is connected to a Wi-Fi® network.
Disconnect Q-Trak™ XP monitor from Wi-Fi® network
then connect TrakPro™ Ultra
to Q-Trak™ XP monitor via USB-C cable.
Alternatively, use Wi-Fi®
network to connect TrakPro™
Ultra to the Q-Trak™ XP monitor.
Unable to connect TrakPro™ Ultra software to Q-Trak™ XP monitor via Wi-Fi®.
Router assigned the same IP address to the Q-Trak™ XP monitor and another device.
Reboot the Q-Trak™ XP monitor.
Page 82
68 Q-Trak™ XP IAQ Monitor Model 7585
Unexplained rapid changes in CO
2
measurement.
Radio frequency interference.
Identify whether the following electronic devices capable of transmitting radio waves are nearby: microwave ovens, Wi-Fi® routers, cellular phones, or personal communication devices, i.e., walkie-talkie, two-way radios.
If one of these devices is present, turn them off or place them in a mode that does not transmit. Alternatively, increase the distance between the Q-Trak™ monitor and transmitting device unitl the problem is no longer present.
NOTE: Other electronic devices capable of transmitting radio waves may also adversely impact the CO2 measurement.

Error Messages

Message
Definition
<SensorType> is in error. ERRCODE: Corrupt Reading
Cannot calculate measurement value due to erroneous User Calibration or missed measurement.
<SensorType> is in error. ERRCODE: No Reading
Cannot execute sensor due to sensor being removed or faulty sensor electronics.
Density correction warning
Cannot compensate for density changes due to faulty Barometric Pressure or Temperature sensor(s).
PID electronics error. Service or replace VOC sensor
VOC sensor's electronics (Bulb and/or Stack) is faulty.
Page 83
Operation and Service Manual 69
Message
Definition
Sensor error. Service or replace <SensorType> sensor
Cannot initialize sensor due to corrupt sensor memory or faulty electronics.
Battery is faulty. Please replace battery.
Battery has been depleted to unsafe level.
The battery is too cold to run. Saving data and shutting down…
Battery temperature has dropped below -20°C and is unsafe to operate.
The battery is too hot to run. Saving data and shutting down…
Battery temperature has exceeded 60°C and is unsafe to operate.
Battery too cold to charge
Battery temperature is between
-20°C and 0°C and cannot
safely charge.
Battery critically low. Plug in charging cable. Device will shut down in <30 Second Timer>
Apply AC power and allow battery to charge.
Battery low. Plug the power cord into the instrument or auto shutdown will occur soon.
Battery has less than 15 minutes of runtime remaining. Fan error. Service sensor module.
Sensor module fan is blocked or faulty.
Temperature sensor error. Service sensor module.
Temperature sensor is faulty.
Barometric pressure sensor error. Service sensor module.
Barometric Pressure sensor is faulty.
Measurement Value = 88888888
Cannot calculate measurement value possibly due to erroneous User Calibration, faulty sensor, or sensor removed.
Page 84
70 Q-Trak™ XP IAQ Monitor Model 7585
Message
Definition
<SensorType> sensor calibration out of range. Please refer to the user manual.
Sensor has drifted beyond the TSI® recommended calibration adjustment specification.
Entered value is out of range. Please refer to the user manual for recommended ranges.
Reference value used for calibrating temperature, relative humidity or barometric pressure is outside of sensor’s detectable limits.
Page 85
71
CHAPTER 14
Help

TSI® Technical Support/Service

Please contact one of TSI® Incorporated’s offices or your local distributor to make service arrangements and to receive a Service Request number. To fill out an online Service Request form, visit TSI® Incorporated’s website at http://service.tsi.com.

United States:

TSI Incorporated Shoreview, MN USA
Tel: ............. +1 800-680-1220
.............. +1 651-490-2860
Fax: ............ +1 651-490-3824

Europe:

TSI Instruments Ltd. High Wycombe, UNITED KINGDOM
Telephone: .. +44 (0) 149 4 459200
E-mail: .......... [email protected]

TSI GmbH:

Aachen, GERMANY
Telephone: .. +49 241-52303-0
Fax: ............ +49 241-52303-49
E-mail: ........ [email protected]

Asia:

TSI Instruments Singapore Pte Ltd SINGAPORE
Telephone: .. +65 6595-6388
Fax: .............. +65 6595-6399
E-mail: .......... [email protected]

China:

TSI Instrument (Beijing) Co., Ltd. Haidian District, Beijing, CHINA
Telephone: .. +86-10-8219 7688
Fax: .............. +86-10-8219 7699
E-mail: .......... [email protected]
Page 86
72 Q-Trak™ XP IAQ Monitor Model 7585

TSI Regional Office Locations:

*TSI Incorporated – Visit our website www.tsi.com for
more information.
USA Tel: .......... +1 800 874 2811
UK Tel: ............. +44 149 4 459200
France Tel: ...... +33 1 41 19 21 99
Germany Tel: .. +49 241 523030
India Tel:.......... +91 80 67877200
China Tel: ........ +86 10 8219 7688
Singapore Tel: +65 6595 6388
Page 87
73
APPENDIX A
Specifications
Specifications are subject to change without notice. The Q-Trak™ XP Indoor Air Quality Monitor Model 7585 arrives from
the factory with the following preset technical specifications.
Logging Capability .....................
73 million data points
Operating Temperature .............
40 to 105°F (5 to 40°C)
Storage Temperature ................
-4 to 60°F (-20 to 60°C)
Humidity ....................................
0 to 95% RH (non-condensing)
Altitude ......................................
Up to 4000 meters when using AC adapter
Meter Dimensions .....................
3.8 in. × 8.3 in. × 2.1 in.
(9.7 cm × 21.1 cm × 5.3 cm)
Weight with Batteries .................
1.2 lbs./0.55 kg
Power Requirements .................
Li-ion battery – 7.4V 3200mAh AC Adapter – Input: 100-24 VAC, 50-60 Hz Output: 12 VDC 3A
Battery Runtime .........................
8+ hours with 7585 base configuration
Languages Supported ...............
English, German, French, Traditional and Simplified Chinese
Installed Sensors
Particulate Mass
Sensor Type ..............................
Optical Particle Counter (OPC)
Particulate Mass (PM) ...............
PM1.0, PM2.5, PM10
Range ........................................
0 to 500 µg/m3
Accuracy ...................................
±10 μg/m³ (0 – 100 μg/m³), ±10% of reading 100 to 500 μg/m³
PM Resolution ...........................
1 µg/m3, 0.001 mg/m3
PM Units ....................................
µg/m3, mg/m3
Particle Counts (PC) ..................
PC0.3, PC0.5, PC1.0, PC2.5, PC5.0, PC10.0
PC Resolution ...........................
0.01/cm3
PC Units ....................................
#/cm3
Response Time .........................
<10 seconds
Page 88
74 Q-Trak™ XP IAQ Monitor Model 7585
Installed Sensors
Carbon Dioxide
Sensor Type ..............................
Nondispersive Infrared (NDIR)
Range .......................................
0 to 5000 ppm
Accuracy ...................................
± 50 ppm
Resolution .................................
1 ppm
Response Time (t90).................
< 40s @ 20°C ambient
Temperature
Sensor Type ..............................
Thermistor
Range .......................................
32 to 140°F (0 - 60°C)
Accuracy ...................................
± 1.0°F (0.5°C)
Resolution .................................
0.1°F (0.1°C)
Response Time (t90).................
< 15 seconds to 90% of the actual temperature
Relative Humidity
Sensor Type ..............................
Capacitive
Range .......................................
0 to 100% RH
Accuracy ...................................
± 3% RH
Resolution .................................
0.1 % RH
Response Time (t63).................
8 Seconds to 63% of the actual relative humidity
Barometric Pressure
Sensor Type ..............................
Piezoresistive
Range .......................................
7.7 to 37.2 in hg (196.0 to 945.0 mmHg)
Accuracy ...................................
±0.12 in. Hg (±3.0 mmHg)
Resolution .................................
0.01 in. Hg (0.1 mmHg)
Response Time (t63).................
>2 seconds to 63% of the actual barometric pressure
Optional Gas Sensors
All sensors are pre-calibrated and include a certificate of calibration.
Carbon Dioxide (CO2) sensor - P/N 801399
Sensor Type ..............................
NDIR (Non-dispersive Infrared)
Range .......................................
0 - 5000 ppm
Accuracy
2 ....................................................
± 50 ppm
Resolution .................................
1 ppm
Response Time (t90).................
< 40s @ 20°C ambient
Page 89
Operation and Service Manual 75
Optional Gas Sensors
Carbon Monoxide (CO) sensor - P/N 801401
Sensor Type ..............................
Electrochemical
Range ........................................
0 to 400 ppm
Accuracy ...................................
± 2% of reading ± 1 ppm
Resolution .................................
0.1 ppm
Response Time (t90) .................
< 30 seconds from zero to 10 ppm
TVOC Sensor (ppb) sensor - P/N 801408
Sensor Type ..............................
PID (Photo Ionization Detector)
Range ........................................
0 to 20 ppm
Accuracy
1,2 .................................................
± 3% of reading ± 1 ppm
Resolution .................................
0.01 ppm
Response Time (t90) .................
<3 seconds
TVOC Sensor (ppm) sensor - P/N 801407
Sensor Type ..............................
PID (Photo Ionization Detector)
Range ........................................
0 to 2000 ppm
Accuracy
1,2 .................................................
0-20 ppm: ±3% of reading ± 1ppm; 21-250 ppm: ± 1ppm; 251-2000 ppm: ± 20 ppm
Resolution .................................
0.1 ppm
Response Time (t90) .................
<3 seconds
Formaldehyde (CH2O) sensor - P/N 801409
Sensor Type ..............................
Electrochemical
Range ........................................
0-10 ppm
Accuracy ...................................
± 0.3 ppm
Resolution .................................
0.01 ppm
Response Time (t90) .................
< 90 seconds
Ozone (O3) sensor - P/N 801406
Sensor Type ..............................
Electrochemical
Range ........................................
0 to 20 ppm
Accuracy ...................................
± 0.3 ppm
Resolution .................................
0.01 ppm
Response Time (t90) .................
< 80 seconds from zero to 1 ppm
Page 90
76 Q-Trak™ XP IAQ Monitor Model 7585
Optional Gas Sensors
Chlorine (CL) sensor- P/N 801400
Sensor Type ..............................
Electrochemical
Range .......................................
0 to 20 ppm
Accuracy ...................................
± 0.3 ppm
Resolution .................................
0.01 ppm
Response Time (t90).................
< 60 seconds from zero to 10 ppm
Ammonia Sensor (NH3) sensor - P/N 801403
Sensor Type ..............................
Electrochemical
Range .......................................
0 to 100
Accuracy ...................................
± 1 ppm
Resolution .................................
0.1 ppm
Response Time (t90).................
< 45 seconds
Hydrogen Sulfide (H2S) sensor - P/N 801402
Sensor Type ..............................
Electrochemical
Range .......................................
0 to 50 ppm
Accuracy ...................................
± 0.5 ppm
Resolution .................................
0.01 ppm
Response Time (t90).................
< 60 seconds from zero to 2 ppm
Nitric Oxide (NO) sensor - P/N 801404
Sensor Type ..............................
Electrochemical
Range .......................................
0 to 20 ppm
Accuracy ...................................
± 0.2 ppm
Resolution .................................
0.1 ppm
Response Time (t90).................
< 25 seconds from zero to 2ppm
Nitrogen Dioxide (NO2) sensor - P/N 801405
Sensor Type ..............................
Electrochemical
Range .......................................
0 to 20 ppm
Accuracy ...................................
± 0.5 ppm
Resolution .................................
0.01 ppm
Response Time (t90).................
< 80 seconds from zero to 2ppm
t90: Time to achieve 90% of the actual concentration
1
TVOC accuracy statement based on factory testing with isobutylene
2
CO2 and TVOC accuracies based on density correction activated
Page 91
77
APPENDIX B
Definitions
Out-of-Range (Calibration)
An Out-of-Range error during the calibration of a
sensor means the sensor’s offset or calibration
slope adjustment has drifted outside of the TSI® recommended specification for continued use. Sensor should be replaced or sent to TSI® for recalibration.
Test Duration
The time over which the data will be logged while in Continuous Save mode. The test duration can be set from 1 second to 99 hours: 59 minutes: 59 seconds.
Sample Interval
The sample interval is the time period where data is collected and averaged to produce a single value. For example, if the sample interval is set to 1 minute, each 1 second sample will be averaged over the minute and result in a single value. The sample interval can be set from 1 second to 99 hours: 59 minutes: 59 seconds. However, the sample interval cannot be greater than the test duration.
Time Constant
The time constant is an averaging period for the readings shown on the display. It is used to dampen fluctuations in the measurement to make it easier to read. If there are significant fluctuations in the readings, a longer time constant setting will reduce the fluctuations. The display will update ever second; however, the displayed reading will be the average over the time constant period. For example, if the time constant is 5 seconds, the display will update every second, but the displayed reading will be the average of the last 5 seconds.
Page 92
78 Q-Trak™ XP IAQ Monitor Model 7585
(This page intentionally left blank)
Page 93
Page 94
TSI Incorporated – Visit our website www.tsi.com for more information.
USA Tel: +1 800 680 1220 UK Tel: +44 149 4 459200 France Tel: +33 1 41 19 21 99 Germany Tel: +49 241 523030
India Tel: +91 80 67877200 China Tel: +86 10 8219 7688 Singapore Tel: +65 6595 6388
P/N 6013907 Rev A ©2020 TSI Incorporated Printed in U.S.A.
Loading...