Omron ZX-GT28S11, ZX-GT2840S11, ZX-GT28S41, ZX-GT2840S41 Datasheet

Smart Laser Micrometer
ZX-GT
• High accuracy: 5-10 µm
• All surfaces
• Long sensing distance: < 500 mm
• Line width up to 28 mm
• Fast sampling time: 0.5 ms
• PC software for setup
Ordering Information
ZX-GT
Sensors
Appearance Optical system Measuring
Separate type
Integrated type
Through-beam 28 mm 0 to 500 mm 10 µm NPN ZX-GT28S11
Controller
Appearance Power supply Output type Model
DC NPN ZX-GTC11
Accessories (Order Separately)
Set of Interface Unit and Setup software PCs
Output type Model
NPN ZX-GIF11A
PNP ZX-GIF41A
Interface Unit(RS-232C/Binary output)
Appearance Power supply Output type Model
DC NPN ZX-GIF11
width
PNP ZX-GTC41
Sensing distance
40 mm NPN ZX-GT2840S11
Resolution Output type Model
PNP ZX-GT28S41
PNP ZX-GT2840S41
PNP ZX-GIF41
1ZX-GT
Setup software PCs
Name Model
Smart Monitor GT ZX-GSW11
Calculating Units
Appearance Model
ZX-CAL2
Receiver-Controller Extension Cable
Cable length Model Quantity
Standard cable Flexible cable
1 m ZX-XGC1A ZX-XGC1R 1 m
2 m ZX-XGC2A ZX-XGC2R
5 m ZX-XGC5A ZX-XGC5R
8 m ZX-XGC8A ZX-XGC8R
20 m ZX-XGC20A ZX-XGC20R
Up to two extension cables can be connected. However, be sure to limit the total extension cable length between the receiver and the Controller to 30 meters (including the receiver cable).
Specifications
Sensor
Item ZX-GT28S11 ZX-GT2840S11 ZX-GT28S41 ZX-GT2840S41
Output type NPN PNP
Appearance Separate type Integrated type Separate type Integrated type
Light source Visible semiconductor laser diode (wavelength 650 nm, CLASS 1 of EN60825-1/IEC60825-1, CLASS of
Measuring width 28 mm
Sensing distance 0 to 500 mm 40 mm 0 to 500 mm 40 mm
Minimum sensing object
Linearity
Resolution
Temperature characteristic
Indicators (emitter) Laser ON indicator (green), laser alarm indicator (red)
Indicator (receiver) Optical axis setting indicator (green)
Laser OFF input/sync input ON: Short-circuited with 0 V or 1.5 V max.
Laser deterioration alarm output NPN open-collector output
Power consumption (emitter) 30 mA max.
Power supply voltage (emitter) 24 VDC +10%, -15% ripple (p-p) 10% max.
Dielectric strength 1,000 VAC, 50/60 Hz for 1 min
Insulation resistance 20 MΩ (at 500 VDC megger)
Operating ambient illumination (emitter)
Operating ambient illumination (receiver)
Ambient temperature Operating: 0 to +40°C Storage: -15 to +50°C(with no icing or condensation)
Ambient humidity Operating and storage: 35 to 85% (with no condensation)
Vibration resistance (durability) 10 to 150 Hz Single-amplitude: 0.75 mm for 80 min each in X, Y and Z directions
Degree of protection IEC60529 IP40
Cable length 2 m
Material Case: aluminum die-cast, Lens: glass
Weight (packed state) Approx. 550 g Approx. 570 g Approx. 550 g Approx. 570 g
Accessories Laser warning labels, Instruction Sheet
F. S.: 28 mm measuring range of receiver *1: Distance between emitter and receiver: 500 mm, measurement object at 250 mm from receiver. Glass ends of chamfer 0.1 mm or more can be detected in glass edge measurement mode.
(at binary level 70%)
*2: Linearity is given to be a typical error with respect to an ideal straight line when the distance between the emitter and receiver is 100 mm and light is blocked at a distance of 50 mm from the
receiver. (On the ZX-GT2840@@, the measurement object is measured at a distance of 20 mm from the receiver.)
*3: The amount of fluctuation (±3σ) in the analog output when the distance between the emitter and receiver is 100 mm and a ZX-GTC@@ is connected *4: Change in the light cutoff value on one side when the distance between the emitter and receiver is 100 mm and the light is half-cutoff at a distance of 50 mm from the receiver (On the ZX-
GT2840@@, the measurement object is measured at a distance of 20 mm from the receiver.)
*5: Standard mode (NORM) used
FDA(21CFR 1040.10 and 1040.11)
0.5mm dia.
±0.1%F.S.
(*1)
(*2)
0.2 mm dia.
10 µm (number of process values to average: 16)
±0.01%F.S/C
(*4)
OFF: Open (leakage current: 0.1 mA max.)
30 VDC 20 mA max. Residual voltage 1.2 V max.
3,000 lx (incandescent light)
1,000 lx (incandescent light)
(*5)
0.5 mm dia.
(*3)
ON: Short-circuited with power supply voltage or power supply voltage -1.5 V max. OFF: Open (leakage current: 0.1 mA max.)
PNP open-collector output 30 VDC 20 mA max. Residual voltage 2 V max.
(*1)
0.2 mm dia.
2 Displacement sensors / Width-measuring Sensors