Note: Specifications subject to change without notice.
Printed in Japan.
0405-0.1M (1002) (M)
Operation Manual
Cat. No. Z166-E1-02A
Introduction
Thank you for purchasing an OMRON ZX-E-series Smart Sensor (Inductive
Displacement Type). We hope you will fully utilize this product and its performance
for many years to come.
This manual describes the functions, performance, and application
methods of a ZX-E Smart Sensor. To ensure safety, read this manual carefully before
using the Sensor. In addition, keep this manual in an easily accessible location
for quick reference when needed.
ÇÕǹÇ?ÇëÊ
Preface
Section 1
Section 2
Section 3
Section 4
Section 5
Contents, Warranty and Application Considerations,
Precautions for Safe Use/Correct Use, and How to Use this Manual
Setting Number of Samples to Average80
Using Hold Functions81
Comparing Measured Values (Previous Value Comparisons)88
Changing Display Scales90
Entering Threshold Values97
Linear Output103
Setting Judgement Output Timing (Timer)112
4
ZX-E
Operation Manual
Preface
Section 6Auxiliary Functions115
Measuring with Multiple Amplifier Units116
Changing the Number of Display Digits123
Reversing the Display124
Adjusting Display Brightness (ECO Display)126
Using the Zero Reset Function127
Key Lock Function132
Initializing Settings Data133
Appendices135
Troubleshooting136
Error Messages and Countermeasures137
Q&A138
Glossary139
Specifications and Dimensions140
Characteristic Data148
Quick Reference for Displays157
1
èÕëÊ
2
èÕëÊ
3
èÕëÊ
4
èÕ
AppendicesIndex
Index 161
Revision History 165
ZX-E
Operation Manual
5
PrefaceContents
Preface
6
ZX-E
Operation Manual
PREFACE
READ AND UNDERSTAND THIS DOCUMENT
Please read and understand this document before using the products. Please consult your OMRON
representative if you have any questions or comments.
WARRANTY
OMRON’s exclusive warranty is that the products are free from defects in materials and workmanship for
a period of one year (or other period if specified) from date of sale by OMRON.
OMRON MAKES NO WARRANTY OR REPRESENTATION, EXPRESS OR IMPLIED, REGARDING
NON-INFRINGEMENT, MERCHANTABILITY, OR FITNESS FOR PARTICULAR PURPOSE OF THE
PRODUCTS. ANY BUYER OR USER ACKNOWLEDGES THAT THE BUYER OR USER ALONE HAS
DETERMINED THAT THE PRODUCTS WILL SUITABLY MEET THE REQUIREMENTS OF THEIR
INTENDED USE. OMRON DISCLAIMS ALL OTHER WARRANTIES, EXPRESS OR IMPLIED.
LIMITATIONS OF LIABILITY
OMRON SHALL NOT BE RESPONSIBLE FOR SPECIAL, INDIRECT, OR CONSEQUENTIAL
DAMAGES, LOSS OF PROFITS OR COMMERCIAL LOSS IN ANY WAY CONNECTED WITH THE
PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CONTRACT, WARRANTY, NEGLIGENCE, OR
STRICT LIABILITY.
In no event shall responsibility of OMRON for any act exceed the individual price of the product on which
liability is asserted.
Preface
IN NO EVENT SHALL OMRON BE RESPONSIBLE FOR WARRANTY, REPAIR, OR OTHER CLAIMS
REGARDING THE PRODUCTS UNLESS OMRON’S ANALYSIS CONFIRMS THAT THE PRODUCTS
WERE PROPERLY HANDLED, STORED, INSTALLED, AND MAINTAINED AND NOT SUBJECT TO
CONTAMINATION, ABUSE, MISUSE, OR INAPPROPRIATE MODIFICATION OR REPAIR.
SUITABILITY FOR USE
THE PRODUCTS CONTAINED IN THIS DOCUMENT ARE NOT SAFETY RATED. THEY ARE NOT
DESIGNED OR RATED FOR ENSURING SAFETY OF PERSONS, AND SHOULD NOT BE RELIED
UPON AS A SAFETY COMPONENT OR PROTECTIVE DEVICE FOR SUCH PURPOSES. Please refer
to separate catalogs for OMRON's safety rated products.
OMRON shall not be responsible for conformity with any standards, codes, or regulations that apply to
the combination of products in the customer’s application or use of the product.
At the customer’s request, OMRON will provide applicable third party certification documents identifying
ratings and limitations of use that apply to the products. This information by itself is not sufficient for a
complete determination of the suitability of the products in combination with the end product, machine,
system, or other application or use.
The following are some examples of applications for which particular attention must be given. This is not
intended to be an exhaustive list of all possible uses of the products, nor is it intended to imply that the
uses listed may be suitable for the products:
• Outdoor use, uses involving potential chemical contamination or electrical interference, or conditions or
uses not described in this document.
Operation Manual
ZX-E
7
PREFACE
Preface
• Nuclear energy control systems, combustion systems, railroad systems, aviation systems, medical
equipment, amusement machines, vehicles, safety equipment, and installations subject to separate
industry or government regulations.
• Systems, machines, and equipment that could present a risk to life or property.
Please know and observe all prohibitions of use applicable to the products.
NEVER USE THE PRODUCTS FOR AN APPLICATION INVOLVING SERIOUS RISK TO LIFE OR
PROPERTY WITHOUT ENSURING THAT THE SYSTEM AS A WHOLE HAS BEEN DESIGNED TO
ADDRESS THE RISKS, AND THAT THE OMRON PRODUCT IS PROPERLY RATED AND INSTALLED
FOR THE INTENDED USE WITHIN THE OVERALL EQUIPMENT OR SYSTEM.
PERFORMANCE DATA
Performance data given in this document is provided as a guide for the user in determining suitability and
does not constitute a warranty. It may represent the result of OMRON’s test conditions, and the users
must correlate it to actual application requirements. Actual performance is subject to the OMRON
Warranty and Limitations of Liability.
CHANGE IN SPECIFICATIONS
Product specifications and accessories may be changed at any time based on improvements and other
reasons.
It is our practice to change model numbers when published ratings or features are changed, or when
significant construction changes are made. However, some specifications of the product may be
changed without any notice. When in doubt, special model numbers may be assigned to fix or establish
key specifications for your application on your request. Please consult with your OMRON representative
at any time to confirm actual specifications of purchased products.
DIMENSIONS AND WEIGHTS
Dimensions and weights are nominal and are not to be used for manufacturing purposes, even when
tolerances are shown.
ERRORS AND OMISSIONS
The information in this document has been carefully checked and is believed to be accurate; however, no
responsibility is assumed for clerical, typographical, or proofreading errors, or omissions.
PROGRAMMABLE PRODUCTS
OMRON shall not be responsible for the user’s programming of a programmable product, or any
consequence thereof.
COPYRIGHT AND COPY PERMISSION
This document shall not be copied for sales or promotions without permission.
This document is protected by copyright and is intended solely for use in conjunction with the product.
Please notify us before copying or reproducing this document in any manner, for any other purpose. If
copying or transmitting this document to another, please copy or transmit it in its entirety.
ZX-E
Operation Manual
8
PREFACE
Preface
Operation Manual
ZX-E
9
PREFACE
Preface
10
ZX-E
Operation Manual
Precautions for Safe Use
Always observe the following precautions to ensure safety.
■ Environment
• Do not use the Smart Sensor in locations subject to explosive or flammable gases.
• To ensure safety in operation and maintenance, do not install the Smart Sensor near
high-voltage equipment or power devices.
■ Power Supply and Wiring
• Do not impose voltages exceeding the rated voltage (12 to 24 VDC ±10%).
• When supplying power to the Sensor, make sure that the polarity of the power is correct, and do not connect to an AC power supply.
• Do not short-circuit the load for the open collector output.
• Do not lay the power supply cable for the Smart Sensor together with high-voltage
lines or power lines. Doing so, or placing them into the same duct, can cause induction and lead to malfunction or damage.
• Always turn OFF the power supply before wiring and before connecting or disconnecting connectors.
PREFACE
PrefacePrecautions for Safe Use
■ Settings
• When setting the threshold value with the Smart Sensor connected to an external
device, turn ON the Amplifier Unit's judgement output hold input to prevent the
judgement from being output to the external device.
■ Others
• The ZX-L-series Smart Sensors (Laser Type), ZX-W-series Smart Sensors (Microwave Type), and ZX-T-series Smart Sensors (High-precision Contact Type) are not
compatible. Do not use ZX-L-series, ZX-W-series, or ZX-T-series Smart Sensors
together with ZX-E-series Smart Sensors.
• Do not attempt to disassemble, repair, or modify the Smart Sensor.
• When disposing of the Smart Sensor, treat it as industrial waste.
Operation Manual
ZX-E
11
PREFACE
PrefacePrecautions for Correct Use
Precautions for Correct Use
Always observe the following precautions to prevent operation failures, malfunctions, and
adverse effects on performance and equipment.
Smart Sensor Installation
■ Environment
Do not install the Smart Sensor in the following locations:
• Locations where the ambient temperature exceeds the rated temperature range
• Locations subject to rapid changes in temperature (causing condensation)
• Locations where the relative humidity exceeds the range of 35% to 85%
• Locations subject to corrosive or flammable gases
• Locations where dust, salt, or metallic powder accumulate on the Sensor
• Locations subject to direct vibration or impact
• Locations subject to direct sunlight
• Locations subject to exposure to water, oil, chemicals, etc.
• Locations subject to strong electromagnetic or electrical fields
• Locations subject to water vapor
Installation and Handling of Components
■ Power Supply and Wiring
• Do not extend the Sensor Head cable by more than 8 m. Use a ZX-XC@A Extension
Cable (order separately) to extend the cable from the Sensor.
• Use a shielded cable to extend the Amplifier cable. The shielded cable must have the
same specifications as that of the Amplifier cable.
• When using a commercially available switching regulator, ground the FG (frame
ground) terminal.
• If the power supply line is subject to surges, connect a surge absorber that meets the
conditions of the application environment.
• When connecting multiple Amplifier Units, connect the linear grounds of all the Amplifier Units.
■ Warm-up
After turning ON the power, allow the Smart Sensor to warm up for 30 minutes minimum prior to use. The circuitry is not stable immediately after turning the power ON,
and the values gradually change until the Sensor is completely warmed up.
■ Maintenance and Inspection
• Always turn OFF the power supply before adjusting or removing the Sensor Head.
• Do not use thinners, benzine, acetone, or kerosene to clean the Sensor Head or
Amplifier Unit.
12
ZX-E
Operation Manual
How to Use This Manual
Page Format
Section 3
Basic Operations
Adjusting Linearity
ZX-E Smart Sensor linearity is adjusted before shipment, however
more accurate linearity can be obtained by adjusting linearity again
for the actual sensing objects and operating environment.
Selecting Sensing Object Material
This section describes how to set the sensing object material.
To use the default linearity adjustment after changing materials,
select the material and then select Adjust.
PREFACE
PrefaceHow to Use This Manual
Section Title
Main Heading in
a Section
Indicates page contents.
Outline
Provides an outline or gives a
flowchart of the operation described under the main heading.
Sub-heading
Index Label
Gives the section number and
subject matter.
Changing to T Mode
Set the mode switch to T.
Set the switch to the threshold value to
be set.
Press any Cursor Key.
The first digit of the threshold value displayed on
the sub-display will flash and direct input will be
enabled.
Change the threshold value with the Cursor
Keys.
To cancel the selected setting, use the LEFT
Key to move the cursor to the leftmost digit
and press the LEFT Key again. he display will
return to the current setting.
Procedure and Additional Explanations
Information useful during the operation and reference
pages are provided here with special marks to indicate the
kind of information being provided.
Function Outline and
Suggestions
Operation
Indicates the operation to be
performed next.
ZX-E
Operation Manual
Display Area
Shows the display status after
a step in an operation.
Keys and Switches
Provides an illustration of the keys or
switches used during the operation.
* This page does not actually exist in
this manual.
Operation Manual
ZX-E
13
PREFACE
PrefaceHow to Use This Manual
Notation
■ Menus
Items that appear on the digital displays are set in ALL-CAPS.
■ Procedures
The order for the procedures is indicated by numbered steps.
■ Visual Aids
Provides information on important operating procedures, gives advice on how to use functions, and
highlights important performance information.
Indicates pages with relevant information.
Indicates useful information for when problems arise.
14
ZX-E
Operation Manual
Section 1
FEATURES
ZX-E Features16
Section 1FEATURES
Operation Manual
ZX-E
15
Section 1
FEATURES
ZX-E Features
Section 1 ZX-E Features
The ZX-E Smart Sensor measures the distance between the Sensor Head and the sensing
object.
Example: Detecting the Bottom Deadpoint on a Press Machine
Measurement detail
Sensing object
Measurement output
Judgement outputs
Distance
16
ZX-E
Operation Manual
Useful Notification Function
Section 1
FEATURES
■ Warming Up Display
The display shows the warming-up status when the power is turned ON. This enables
measurements to be started when the status has stabilized after warming up has been
completed.
See page 41.
Below specified
resolution
The sub-display flashes
during warm-up.
W-UP
Fluctuation of measured value
Warming up
in progress
Time lapsed after power up (min.)
Warming up
completed
Section 1 ZX-E Features
Many, Simple Functions
■ Measurement Ready at Power ON
The Smart Sensor can be used simply by installing and wiring it. Simply turn ON the
power and it’s ready to operate.
The measurement distance is displayed on the Amplifier Unit.
Operation Manual
ZX-E
17
Section 1
t
FEATURES
■ Simple Linearity Adjustment
Place the sensing object at specified distances and simply press the ENT Key to exe-
Section 1 ZX-E Features
cute precise linearity adjustment. Time-consuming offset and range adjustments are
not required.
Precise adjustment is also possible for non-ferrous sensing objects.
p. 54
100%
0%
50%
Rated measurement
distance
■ Simple Calculation Settings
Use a Calculating Unit to simply measure thickness and sum and difference calculations between two measurements.
p. 116
18
Calculating Uni
Thickness
ZX-E
Operation Manual
Section 1
FEATURES
Mutual Interference Prevention for Closely Mounted
Sensor Heads
The Smart Sensor has a mutual interference prevention function which allows multiple
Sensor Heads to be mounted close to each other. This function is supported for up to
five Sensor Heads by using ZX-CAL2 Calculating Units.
p. 26 and p. 120
Calculating Units
Compatibility between Sensor Heads and Amplifier Units
Section 1 ZX-E Features
Amplifier Units do not need to be changed when Sensor Heads are changed for maintenance or to switch to new products.
Operation Manual
ZX-E
19
Section 1
FEATURES
Extendable Sensor Head Cables
An extension cable with a maximum length of 8 m can be connected. The ZX-XC-A
Section 1 ZX-E Features
Extension Cable is required to extend the Sensor Head cable.
p. 24
Extension Cable
Up to 8 m
20
ZX-E
Operation Manual
Monitoring Measurement Status
Section 1
FEATURES
■ Resolution Display for Sensing Object
The resolution can be displayed, allowing judgements to be made about detection margins while viewing the resolution value.
p. 51
Resolution
Resolution
Deviation to
be detected
■ Confirm Measurement Status on a Personal Computer
Use an Interface Unit and Smart Monitor V2 to view measurement waveforms and log
measurement data on a personal computer. This function is useful for making on-site
measurement adjustments and for day-to-day quality control.
Section 1 ZX-E Features
p. 24
Smart Monitor V2
Interface Unit
Amplifier Unit
Operation Manual
ZX-E
21
Section 1
FEATURES
Section 1 ZX-E Features
22
ZX-E
Operation Manual
Section 2
PREPARATIONS FOR
MEASUREMENT
Basic Configuration24
Part Names and Functions25
Installing the Amplifier Unit28
Installing Sensor Heads30
Connections34
Wiring Output Cables38
Section 2PREPARATIONS FOR MEASUREMENT
Operation Manual
ZX-E
23
Section 2 Basic Configuration
Section 2
PREPARATIONS FOR MEASUREMENT
Basic Configuration
The basic configuration of the ZX-E-series Smart Sensors is shown below.
ZX-L-series Smart Sensors (Laser Type) and ZX-E-series Smart Sensors (Inductive Displacement Type) are
not compatible. Do not use ZX-L-series and ZX-E-series Smart Sensors together.
Basic Configuration
Smart Monitor
(software)
ZX-SW11EV2
(Version 2.0 or later)
Controls Amplifier
Units from a
personal computer
and monitors
measured values.
Personal
computer
Sensor
Heads
ZX-ED@@T
ZX-EM@@T
ZX-EV@@T
Sensor Heads
detect the
sensing objects.
p. 30
Interface Unit
ZX-SF11
(Version 2.0 or later)
Used when
connecting a
personal computer.
p. 37
Amplifier Units
ZX-EDA11 or
ZX-EDA41
Process measurements and output
measurement results.
p. 25
Power Supply
12 to 24 V DC (±10%)
Extension
Cables
ZX-XC1A (1 m)
ZX-XC4A (4 m)
ZX-XC8A (8 m)
Used between
Sensor Heads and
Amplifier Units.
Calculating Units
ZX-CAL2
Used when connecting
multiple Amplifier
Units.
• Perform calculations.
• Prevent mutual
interference
p. 35
24
ZX-E
Operation Manual
PREPARATIONS FOR MEASUREMENT
(12)
y(
Part Names and Functions
Amplifier Units
Controls (❋)
Display area (❋)
(4) Output cable
(3) Connectors (one on each side, two total)
(1) Input cable
(2) Current/voltage switch (back of Unit)
Section 2
Section 2 Part Names and Functions
❋ Details of Controls
and Display Area
(8) HIGH indicator
(9) PASS indicator
(10) LOW indicator
(11) Main display
Sub-displa
(5) Power ON indicator
(6) Zero reset indicator
(7) ENABLE indicator
(15) Operating Keys
(14) Mode switch
13) Threshold switch
(1) The input cable connects the Sensor Head.
(2) The current/voltage switch selects either a current or voltage linear output.
Current/voltage
output selector
Monitor focus settings are also required when switching the output.
Voltage output
Current output
p. 103
(3) The connectors connect Calculation and Interface Units.
(4) The output cable connects to the power supply and external devices, such as sync sensors or
programmable controllers.
(5) The Power ON indicator lights when the power is turned ON.
(6) The Zero Reset indicator lights when the zero reset function is enabled.
(7) The ENABLE indicator lights when the measurement result is within the measurement dis-
tance.
(8) The HIGH indicator lights when the judgement result is HIGH.
(9) The PASS indicator lights when the judgement result is PASS.
(10) The LOW indicator lights when the judgement result is LOW.
(11) The main display shows measured values and function names.
(12) The sub-display shows additional information and function settings for measurements.
Reading Displays, p. 47
(13) The threshold switch selects whether to set (and display) the HIGH or LOW threshold.
(14) The mode switch selects the operating mode. Switching Modes, p. 46
(15) The Control Keys set measurement conditions and make other settings.
Key Operations, p. 48
Operation Manual
ZX-E
25
Section 2
PREPARATIONS FOR MEASUREMENT
Sensor Heads
Section 2 Part Names and Functions
Sensor head
Calculating Units
Display (❋)
❋ Display Detail
Preamplifier
Connectors (one on each side, two total)
Connects to Amplifier Unit.
Connector indicators
Light when Calculating Unit is connected
to Amplifier Units.
Connector
Connects to Amplifier Unit.
26
ZX-E
Operation Manual
Interface Units
(2) Amplifier Unit Connector
(1) Communications Connector
Section 2
PREPARATIONS FOR MEASUREMENT
Display (❋)
Section 2 Part Names and Functions
❋Display Detail
(3) Power supply indicator
(4) Sensor communications indicators (BUSY and ERR)
(5) External terminal communications indicators (BUSY and ERR)
(1) The communications connector connects the communications cable to the computer.
(2) The Amplifier Unit connector connects to the Amplifier Unit.
(3) The power supply indicator lights when the power is turned ON.
(4) BUSY: Lights during communications with the Smart Sensor.
ERR:Lights if an error occurs during communications with the Smart Sensor.
(5) BUSY: Lights during communications with the personal computer.
ERR:Lights if an error occurs during communications with the computer.
Operation Manual
ZX-E
27
Section 2
PREPARATIONS FOR MEASUREMENT
Installing the Amplifier Unit
Amplifier Units can be easily mounted to 35-mm DIN Track.
Section 2 Installing the Amplifier Unit
DIN Track (order separately)
PFP-100N (1 m)
PFP-50N (0. 5 m)
PFP-100N2 (1 m)
End Plates (order separately)
PFP-M
■ Installation
Hook the connector end of the Amplifier Unit on the DIN Track and press in at the bottom until the Unit locks into place.
Hook on the connector end
Always hook the connector end of the Amplifier Unit on the DIN Track first. Mounting strength
may decrease if the output cable end is hooked on the DIN Track first.
28
ZX-E
Operation Manual
PREPARATIONS FOR MEASUREMENT
■ Removal Method
Push the Amplifier Unit up and pull out from the connector end.
Section 2
Section 2 Installing the Amplifier Unit
Operation Manual
ZX-E
29
Section 2
PREPARATIONS FOR MEASUREMENT
Installing Sensor Heads
This section describes how to install Sensor Heads and Preamplifiers.
Sensor Heads
Section 2 Installing Sensor Heads
■ Installation
❚ ZX-ED@@T Sensor Heads (Non-threaded Type)
Mounting Bracket (order separately)
Y92E-F5R4 (for 5.4 mm dia.)
When using a set screw, tighten the screw to a torque of 0.2 N⋅m or less.
Mount the Sensor Head as shown in the following diagram.
(Unit: mm)
Set screw hole
ModelA
ZX-EDR5T9 to 18
ZX-ED01T9 to 18
ZX-ED02T11 to 22
❚ ZX-EM@@T Sensor Heads (Threaded Type)
The tightening torque for the threaded type (ZX-EM@@T) is shown in the following
table.
Model
ZX-EM02T
ZX-EM07MT
ZX-EM02HT5.9 N·m
When using the ZX-EM02HT, consider the thermal expansion caused by the rise in temperature
of the sensing object, and make sure that the sensing object does not touch the sensing surface.
Even when operated within the specified temperature range, using in locations subject to
extreme temperature fluctuations may result in deterioration of Sensor characteristics.
When using the ZX-EM02HT, make sure that the bending radius of the cable from the Sensor
Head is 8 mm or greater .
Strength
(torque)
15 N·m
30
ZX-E
Operation Manual
Section 2
PREPARATIONS FOR MEASUREMENT
■ Installation Distance
Mount the Sensor Head so that the distance between the Sensor Head and the sensing object is approximately half of the measurement distance.
Example: ZX-ED01T Sensor Head
Measurement distance: 0 to 1 mm
0.5 mm
Sensing object: Ferrous object 18 × 18 mm or larger
Use a ferrous sensing object larger than a standard sensing object. If a smaller-than-standard
sensing object or a non-ferrous object is used, the predetermined characteristics may not be
obtained.
Characteristic Data, p. 148
■ Influence of Surrounding Metal
Separate the Sensor Head from surrounding metals by at least the distances shown in
the following diagram.
Dia.
(Unit: mm)
ModelDia. AB
ZX-EDR5T89
ZX-ED01T109
ZX-ED02T129
ZX-EM02T129
ZX-EM07MT5520
ZX-EV04T16
ZX-EM02HT189
×324.8
Section 2 Installing Sensor Heads
Operation Manual
ZX-E
31
Section 2
PREPARATIONS FOR MEASUREMENT
■ Mutual Interference
When using multiple Sensor Heads, separate each Sensor Head by the minimum distances shown in the following diagram.
The distance between Sensor Heads can be further reduced when the Sensor Heads are side by
side if the mutual interference prevention function is used.
Performing Calculations, p. 116
Section 2 Installing Sensor Heads
• Face to face
• Side by side
B
ModelA
ZX-EDR5T53.120
ZX-ED01T105.450
ZX-ED02T20850
ZX-EM02T201050
ZX-EM07MT10030150
ZX-EV04T801450
ZX-EM02HT201250
Mutual Interference Prevention Function
UsedNot used
(Unit: mm)
32
ZX-E
Operation Manual
Preamplifiers
■ Installation
Use the enclosed Preamplifier Mounting Bracket.
Section 2
PREPARATIONS FOR MEASUREMENT
The Preamplifier can also be mounted to 35-mm DIN Track.
Use the ZX-XBE2 Preamplifier DIN Track Mounting Bracket (order separately) when mounting
the Preamplifier to DIN Track.
1. Use M3 screws to fix the enclosed Preampli-
fier mounting bracket.
2. Snap one end of the Preamplifier into the
bracket.
Mounting Dimensions
Section 2 Installing Sensor Heads
(Unit: mm)
Two, M3 holes
3. Then snap the other end of the Preamplifier
into the bracket.
■ Removal Method
Hold the center of the Preamplifier and lift.
Operation Manual
ZX-E
33
Section 2
PREPARATIONS FOR MEASUREMENT
Connections
This section describes how to connect component parts of the Smart Sensor.
Section 2 Connections
Turn OFF the power supply to the Amplifier Unit before connecting or removing components. The Smart Sensor may malfunction if components are connected or removed while the power is ON.
Sensor Heads
Do not touch the terminals inside the connector.
■ Connection Method
Push the Sensor Head connector into the Amplifier Unit connector until it locks.
34
■ Removal Method
When disconnecting the Sensor Head, hold the connector ring and the Amplifier Unit
connector and pull them straight out.
Do not pull only on the connector ring, because the input cable of the Amplifier Unit may be damaged.
Connector ring
All settings on the Amplifier Unit will be cleared when the Sensor Head is replaced with a different model.
ZX-E
Operation Manual
Section 2
PREPARATIONS FOR MEASUREMENT
Calculating Units
Use a Calculating Unit to connect Amplifier Units when making calculations between
Amplifier Units and to prevent mutual interference between Sensor Heads.
The number of Amplifier Units that can be joined depends on the functions being used.
FunctionNo. of Connectable Amplifier Units
Calculation2
Mutual interference prevention5
Provide power to all connected Amplifier Units.
■ Connection Method
Section 2 Connections
1. Open the connector covers on the Amplifier Units.
Open the connector covers by lifting and sliding them open.
2. Mount the Calculating Unit to the DIN Track.
3. Slide and connect the Calculating Unit to the Amplifier Unit connector.
4. Slide and connect the second Amplifier Unit to the Calculating Unit connector.
Perform the above operation in the reverse order to remove Calculating Units.
Operation Manual
ZX-E
35
Section 2
PREPARATIONS FOR MEASUREMENT
■ Channel Numbers of Amplifier Units
The following diagram shows the channel numbers when multiple Amplifier Units are
connected.
Section 2 Connections
36
ZX-E
Operation Manual
Section 2
PREPARATIONS FOR MEASUREMENT
Interface Units
Use an Interface Unit to connect a personal computer to the Smart Sensor system.
■ Connection Method
Section 2 Connections
1. Open the connector cover on the Amplifier Unit.
Open the connector cover by lifting and sliding it open.
2. Mount the Interface Unit to the DIN Track.
3. Slide and connect the Interface Unit to the Amplifier Unit connector.
Perform the above operation in the reverse order to remove Interface Units.
When multiple Amplifier Units are used, connect the Interface Unit to the Amplifier Unit with the
highest channel number.
Operation Manual
ZX-E
37
Section 2
PREPARATIONS FOR MEASUREMENT
Wiring Output Cables
The following diagram shows the wires in the output cable.
Wire the output cable correctly. Incorrect wiring may damage the Smart Sensor.
Section 2 Wiring Output Cables
Brown
Blue
White
Green
Gray
Black
Shield
Pink
Orange
Purple
Red
(1) A 12- to 24-VDC (±10%) power supply is connected to the power supply terminals. When
using an Amplifier Unit with a PNP output, the power supply terminal is also the common I/O
terminal for all I/O except for the linear output.
Use a stabilized power supply separate from other devices and power systems for the Amplifier
Unit, particularly when high resolution is required.
(2) The GND terminal is the 0-V power supply terminal. When using an Amplifier Unit with an
NPN output, the GND terminal is also the common I/O terminal for all I/O except for the linear
output.
(3) The HIGH judgement output outputs HIGH judgement results.
(4) The PASS judgement output outputs PASS judgement results.
(5) The LOW judgement output outputs LOW judgement results.
(6) The linear output outputs a current or voltage output in accordance with the measured value.
(7) The linear output GND terminal is the 0-V terminal for the linear output.
(1) Power supply
(2) GND
(3) HIGH judgement output
(4) PASS judgement output
(5) LOW judgement output
(6) Linear output
(7) Linear output GND
(8) Judgement output hold input
(9) Zero reset input
(10) Timing input
(11) Reset input
38
• Use a different ground for the linear output from the normal ground.
• Always ground the linear output terminal even when linear output is not used.
(8) When the judgement output hold input is turned ON, the judgement outputs are held and not
output to the external devices. Turn the judgement output hold input ON when setting threshold values.
When setting threshold values while connected to external devices, turn ON the Amplifier Unit’s
judgement output hold input to prevent the outputs to external devices from changing.
(9) The zero reset input is used to execute and clear zero reset.
(10) The timing input is for signal input from external devices. Use it for hold function timing.
(11) The reset input resets all measurement processing and outputs.
ZX-E
Operation Manual
I/O Circuit Diagrams
■ NPN Amplifier Unit
Brown
Section 2
PREPARATIONS FOR MEASUREMENT
Section 2 Wiring Output Cables
12 to 24 V DC
Internal circuit
Current/voltage
output selector
Current output
(4 to 20 mA)
100 Ω
Voltage
output
(±4 V)
HIGH judgement
White
Green
Gray
BlueGND (0 V)
PinkJudgement output hold input
PurpleTiming input
Orange Zero reset input
RedReset input
Black
ShieldLinear ground
output
PASS judgement output
LOW judgement output
Linear output
Load
Load
Load
Load
12 to 24 V DC
Current output: 300 Ω max.
Voltage output: 10 kΩ min.
Operation Manual
ZX-E
39
Section 2
PREPARATIONS FOR MEASUREMENT
■ PNP Amplifier Unit
Brown
White
Section 2 Wiring Output Cables
Green
Gray
BlueGND (0 V)
PinkJudgement output hold input
Purple Timing input
Internal circuit
Current/voltage
output selector
Current output
(4 to 20 mA)
100 Ω
Voltage
output
(±4 V)
Orange Zero reset input
RedReset input
Black
ShieldLinear ground
12 to 24 V DC
HIGH judgement
output
PASS judgement output
LOW judgement output
Load
Linear output
Load
12 to
24
V DC
Load
Load
Current output: 300 Ω max.
Voltage output: 10 kΩ min.
40
ZX-E
Operation Manual
Section 2
PREPARATIONS FOR MEASUREMENT
Confirming Warm-up Completion
When the power is turned ON in RUN or T Mode, the sub-display will flash W-UP to show that
the Sensor is warming up. Warm-up requires approximately 5 to 15 minutes. When warming
up has been completed, the normal display will be shown.
SUB
Measurement operations can be performed while in warm-up display status, but the precision of
measurements before warming up has been completed will be low. For high-precision measurements, wait until warming up has been completed.
Comparing Measured Values
(Previous Value Comparison)
Changing Display Scales
p. 80
p. 81
p. 88
p. 90
44
ZX-E
Operation Manual
Setting Judgement Conditions
Entering Threshold Values
Setting Output Data
Linear Output
Setting Judgement Output Timing (Timer)
If required
Correcting the Measurement
Reference Point
Using the Zero Reset Function
p. 97
p. 103
p. 112
p. 127
If Problems Occur
Section 3
BASIC OPERATION
Abnormal Operation
Unknown Terms
Measuring with Multiple
Amplifier Units
Applied Settings
Performing Calculations
Preventing Mutual
Interference between
Saving Zero Reset Level
Troubleshooting
Glossary
p. 139
p. 129
p. 136
p. 116
p. 120
Error Messages
Error Messages and
Countermeasures
Meaning of Digital Display
Quick Reference for Digital
Displays
Using ECO Display Function
p. 126
Additional Functions
Reversing the Display
p. 124
p. 137
Section 3 Flow of Operation
p. 157
Setting Offset Values
for Zero Reset
p. 128
Changing the Number
of Display Digits
p. 123
Key Lock Function
p. 132
Initializing Settings Data
p. 133
Changing Setttings
Operation Manual
ZX-E
45
Section 3
BASIC OPERATION
Basic Knowledge for Operation
Switching Modes
The ZX-E has three modes. Use the Mode Switch on the Amplifier Unit to switch
between modes. Switch to the desired mode before starting operation.
Section 3 Basic Knowledge for Operation
ModeDescription
RUNNormal operation mode
TMode for setting the threshold values
FUNMode for setting measurement conditions
Function Transition Charts, p. 51
46
ZX-E
Operation Manual
Section 3
BASIC OPERATION
Reading Displays
The data displayed on the main and sub-displays depends on the mode currently
selected. When the power is first turned ON after shipment, RUN mode data is displayed.
Main display
Sub-display
ModeMain DisplaySub-display
RUNDisplays the measured value
(the value after measurement
conditions have been reflected.)
For example, when the hold function is set, the held value will be
displayed.
Changes between displaying the present value (actual measured value), threshold value, output value, and resolution in
order when the Control Keys are pressed.
Threshold Value Display
Displays either the HIGH or LOW threshold value, depending
on the position of the threshold switch.
Section 3 Basic Knowledge for Operation
The monitor focus setting determines whether the value is
output as voltage or current.
Output Settings (Monitor Focus), p. 103
TDisplays the measured value
(the value after the measurement
conditions have been reflected).
For example, when the hold function is set, the held value will be
displayed.
FUNDisplays the function names in
order when the Control Keys are
pressed.
Displays the threshold value for the threshold being set.
Displays either the HIGH or LOW threshold value, depending
on the position of the threshold switch.
Displays the setting for the function displayed on the main display.
Function Transition Charts, p. 51
■ Alphabet Display Format
The alphabet appears on the main and sub-displays as shown in the following table.
Operation Manual
ZX-E
47
Section 3
s
BASIC OPERATION
Key Operations
Use the Control Keys to change the display and set measurement conditions.
Control Key
Section 3 Basic Knowledge for Operation
The mode currently selected determines the key functions.
Switching Modes, p. 46
Key
Function
RUN ModeT ModeFUN Mode
LEFT Key
RIGHT Key
Cursor
Keys
ENT KeyPerforms zero reset.Function changes depend-
UP KeyPerforms timing input.Used when changing
DOWN Key Resets input.
Changes sub-display
content.
Used when selecting
numeral digits.
numerals.
ing on operation.
• Confirms threshold value.
• Executes teaching.
Function changes depending on setting.
• Switches function display.
• Selects numeral digit.
• Stops setting.
Function changes depending on setting.
• Switches between
selections.
• Changes numerals.
Confirms the set condition or
value.
48
ZX-E
Operation Manual
Section 3
BASIC OPERATION
Setting Conditions
Display the target function on the main display and select the desired value from the
sub-display to set measurement conditions.
This section uses the example of setting a peak hold as the hold condition to explain
how to set measurement conditions.
Changing to FUN Mode and HOLD
1. Set the mode switch to FUN.
2. Use the LEFT and RIGHT Keys to display
HOLD on the main display.
Setting Hold Conditions
3. Press either the UP or DOWN Key.
The present set value will flash on the sub-display.
4. Use the UP and DOWN Keys to select P-H.
Press either the LEFT or RIGHT Key to cancel the
selected option. The display will return to the current setting (OFF in this example).
Section 3 Basic Knowledge for Operation
5. When you finish selecting the set value, press
the ENT Key to confirm the setting.
The setting will be registered.
Operation Manual
ZX-E
49
Section 3
BASIC OPERATION
Inputting Numerals
This section describes how to input numeric values for threshold and output settings.
The example of direct input of the low threshold value will be used.
Changing the low threshold from 0.2000 to 0.1900
Changing to T Mode
Section 3 Basic Knowledge for Operation
1. Set the mode switch to T.
Setting Threshold Value
2. Set the switch to L.
The measured value will be displayed on the main
display. The current setting will be displayed on the subdisplay.
3. Press any Cursor Key.
The first digit on the sub-display will flash and direct input
will be enabled.
4. Use the LEFT or RIGHT Key to move the cur-
sor to the first decimal place.
5. Use the UP or Down Key to display 1.
50
6. Repeat steps 4 and 5 to move the cursor to
the second decimal place and display 9.
To cancel the selected setting, use the LEFT Key to move
the cursor to the leftmost digit and press the LEFT Key
again. Alternatively, use the RIGHT Key to move to the
rightmost digit and press the RIGHT Key again. The display will return to the current setting (0.2000 in this example).
7. When you finish adjusting the numeric value,
press the ENT Key to confirm the value.
The display will change from flashing to being lit continuously and the numeric value will be registered.
ZX-E
Operation Manual
BASIC OPERATION
p
Function Transition Charts
Reading Transition Charts
The upper section is the main display and the lower section is the sub-display.
Main display
Sub-display
RUN Mode
Measured value (See note.) (The main display always shows the measured value.)
Section 3
Section 3 Function Transition Charts
Present value (See note.)
Note: In FUN mode, the measured value and
Threshold value
Output value
resent value are displayed first.
Resolution
The numerals shown in the above diagram are an example only. The actual display may
be different.
Present Values and Measured Values p. 139
T Mode
There is no function transition in T mode.
Measured value
Threshold value
The numerals shown in the above diagram are an example only. The actual display may
be different.
p. 97
In RUN and T modes, the position of the threshold switch will determine whether the HIGH or LOW threshold
will be displayed.
Threshold switch
ZX-E
Operation Manual
51
Section 3
BASIC OPERATION
FUN Mode
Section 3 Function Transition Charts
Linearity adjustment
(See note 1.)
Number of samples
to average
p. 54
Note 1: Linearity adjustment is dis-
played first when you enter
FUN mode.
p. 80
Hysteresis
p. 102
Hold
Trigger mode
p. 83
p. 81
When CLOSE is selected, the special functions will not be displayed and
Special functions
(See note 2.)
the display will return to LINER.
Previous value
comparison
p. 88
Note 2: Previous value compari-
son is displayed when
hold is not set to OFF.
Mutual interference prevention
(See note 3.)
p. 120
Note 3: Mutual interference prevention is
displayed only on the CH1 Amplifier
Unit.
Display during
zero reset
p. 127
When ALL is selected, all special functions are displayed.
Linear output
correction
p. 107
Limited number
of display digits
p. 123
Zero reset
memory
p. 130
Monitor focus
p. 103
ECO mode
p. 126
Clamp value
setting
Scaling
p. 90
Display reverse
p. 124
Settings for nonmeasurement
p. 110
52
ZX-E
Operation Manual
BASIC OPERATION
Timer
Section 3
Self trigger level
p. 83
Special
Delay hold
p. 86
Self trigger hysteresis width
p. 83
Delay time
p. 86
Settings
initialization
p. 133
Sampling period
p. 86
p. 112
2-sensor
operation
(See note 4.)
Timer time
setting
Section 3 Function Transition Charts
p. 112
Thickness
setting mode
p. 116
p. 116
Note 4: When multiple Amplifier Units are con-
nected, 2-sensor operation is displayed
on all but the CH1 Amplifier Unit.
This symbol indicates that after the ENT Key has been
pressed to confirm the selection, the LEFT and RIGHT
Keys should be pressed to move through the menu.
ZX-E
Operation Manual
53
Section 3 Adjusting Linearity
Section 3
BASIC OPERATION
Adjusting Linearity
ZX-E Smart Sensor linearity is adjusted before shipment, however more accurate linearity can
be obtained by adjusting linearity again for the actual sensing objects and operating
environment.
Adjust linearity before setting measurement conditions. Linearity should also be adjusted
again when Sensor Heads are replaced.
❚ Flow of Operation
Selecting Sensing
Object Material
Linearity p. 139
Entering Adjustment
Val ues
Executing Adjustment
Selecting Sensing Object Material
This section describes how to set the sensing object material.
SelectionMaterial
FE (default)Iron
SUSStainless steel (SUS304)
ALAluminum
Linearity According to Material
54
Characteristic Data, p. 148
To use the default linearity adjustment after changing materials, select the material and then perform
Executing Adjustment.
ZX-E
Operation Manual
Changing to FUN Mode and LINER
1. Set the mode switch to FUN.
2. Use the LEFT or RIGHT Key to display
LINER on the main display.
Setting Sensor Object Material
3. Press the ENT Key.
METAL will be displayed.
Section 3
BASIC OPERATION
Section 3 Adjusting Linearity
4. Press the UP or DOWN Key.
The sub-display will flash.
5. Use the UP or DOWN Key to select the mate-
rial.
6. Press the ENT Key to confirm the selection.
The setting will be registered.
Operation Manual
ZX-E
55
Section 3 Adjusting Linearity
e
Section 3
BASIC OPERATION
Entering Adjustment Values
Set the sensing object at the positions for 0%, 50%, and 100% of the rated measurement distance and register those measured values as the adjustment values.
100%
50%
0%
Rated measurement distanc
Measurement distance, p. 143
Perform the registration in order, starting with 0%, then 50% and 100% of the measurement distance.
To simply replace the sensing object material and use the default linearity adjustment, skip this operation and perform Executing Adjustment.
Entering Adjustment Value for 0% Position
1. Set the sensing object at the 0% position.
2. Use the LEFT and RIGHT Keys to display
D000.
3. Press the ENT Key.
The sub-display will show OK and the adjustment value
will be registered.
0%
Rated measurement distance
56
Press the ENT Key again to reinput the adjustment value.
ZX-E
Operation Manual
Entering Adjustment Value for 50% Position
Section 3
BASIC OPERATION
4. Move the sensing object to the 50% position.
5. Use the LEFT and RIGHT Keys to display
D050.
6. Press the ENT Key.
The sub-display will show OK and the adjustment value
will be registered.
Press the ENT Key again to reinput the adjustment value.
Entering Adjustment Value for 100% Position
50%
Rated measurement distance
Section 3 Adjusting Linearity
7. Move the sensing object to the 100% posi-
tion.
8. Use the LEFT or RIGHT Keys to display
D100.
9. Press the ENT Key.
The sub-display will show OK and the adjustment value
will be registered.
Press the ENT Key again to reinput the adjustment value.
100%
Rated measurement distance
ZX-E
Operation Manual
57
Section 3 Adjusting Linearity
Section 3
BASIC OPERATION
Executing Adjustment
This section describes how to execute adjustment based on the settings made in steps
and .
Executing Adjustment
1. Use the LEFT and RIGHT Keys to display
TABLE on the main display and START on the
sub-display.
2. Press the ENT Key.
Linear adjustment will be executed.
When the adjustment data has been registered, the subdisplay will show OK.
If the sub-display shows ERRTB, there has been a mistake in the settings. Check the material selected for the
sensing object and the set positions and register the settings again.
■ Cancelling Adjustment
The adjustment settings will be cleared if the adjustment is cancelled.
1. Use the LEFT and RIGHT Keys to display
TABLE on the main display and CANCL on
the sub-display.
OK
NG
58
2. Press the ENT Key.
The settings for the current adjustment will be cancelled
and the display will change to LINER.
ZX-E
Operation Manual
Section 3
BASIC OPERATION
Initializing Adjustment Settings
Initialize the linearity adjustment settings to return to the default settings.
This section describes how to initialize the linearity adjustment settings only. To initialize other settings, use the INIT function.
Initializing Settings Data, p. 133
Moving to FUN Mode and LINER
1. Set the mode switch to FUN.
2. Use the LEFT and RIGHT Keys to display
LINER on the main display.
Initializing Settings
3. Press the ENT Key.
The display will show METAL.
4. Use the LEFT and RIGHT Keys to display
LINIT.
Section 3 Adjusting Linearity
5. Press and hold down the ENT Key.
One dash (−) at a time will appear on the sub-display.
6. Release the ENT Key once OK is displayed
on the sub-display.
The adjustment settings have been initialized.
Operation Manual
ZX-E
59
Section 3 Adjusting Linearity
Section 3
BASIC OPERATION
60
ZX-E
Operation Manual
Section 4
MAIN APPLICATIONS AND
SETTING METHODS
Detecting Bottom Deadpoint62
Measuring Height67
Measuring Eccentricity and Vibration71
Measuring Thickness74
Section 4MAIN APPLICATIONS AND SETTING METHODS
Operation Manual
ZX-E
61
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Detecting Bottom Deadpoint
This section describes, as an example, how to detect the bottom deadpoint of a press
machine.
Section 4 Detecting Bottom Deadpoint
When making settings while still connected to an external device, set the Amplifier Unit’s judgement output
hold input to ON so that the output to the external device remains unchanged.
Adjust linearity before executing this operation.
Adjusting Linearity, p. 54
❚ Flow of Operation
Mounting Sensor
to Press
Adjusting
Detection Position
Setting
Measurement
Timing
Setting Bottom
Deadpoint Position
Setting Tolerance
Judgement Values
62
ZX-E
Operation Manual
Section 4
t
MAIN APPLICATIONS AND SETTING METHODS
Mounting Sensor to Press
Mount the Sensor Head and the sensing object to the press machine. Refer to the following diagram for the required mounting jig.
Installing Sensor Heads, p. 30
Use a ferrous sensing object and one that is as large as or larger than the standard sensing object.
Sensing object, p. 143
Sensing objec
Section 4 Detecting Bottom Deadpoint
Adjusting Detection Position
Adjust the position of the Sensor Head so that when the press machine is set at the
bottom deadpoint position, the distance between the Sensor Head and the sensing
object is about half of the measurement distance. Refer to the Amplifier Unit display
while adjusting the Sensor Head position.
Measurement Distance, p. 143
Operation Manual
ZX-E
63
Section 4
MAIN APPLICATIONS AND SETTING METHODS
1. Put the press machine in inching mode and
lower the stripper (or top mold) to the bottom
deadpoint.
Stripper or top mold
Bottom mold
2. Adjust the position of the Sensor Head so that
this position is at about half of the measurement distance.
The measured value will be displayed on the
Section 4 Detecting Bottom Deadpoint
Amplifier Unit. Refer to this display while
adjusting the Sensor Head.
1/2 the
measurement
distance
64
ZX-E
Operation Manual
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Setting Measurement Timing
The hold function is used to detect the bottom deadpoint.
To ignore bounding when measuring during press operation, specify a time delay from
the timing signal to when sampling starts.
When the timing signal cannot be input from the device, set a self-down trigger.
Measured value
Self-trigger level
Section 4 Detecting Bottom Deadpoint
The bottom value is held.
Timing input
Delay
time
Sampling
period
Refer to Section 5 Detailed Settings for details on settings.
Using Hold Functions,p. 81
Operation Manual
ZX-E
65
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Setting Bottom Deadpoint Position
Set the bottom deadpoint as reference value 0.
Bottom deadpoint
Refer to Section 6 Auxiliary Functions for details on settings.
Using the Zero Reset Function, p. 127
Section 4 Detecting Bottom Deadpoint
Set the previous value comparison to ignore slight bottom deadpoint fluctuation at press startup and
the influence of temperature drift.
Comparing Measured Values (Previous Value Comparisons), p. 88
A value other than 0 can be set.
Setting Offset Values, p. 128
Setting Tolerance Judgement Values
Set the upper and lower limits (the HIGH and LOW threshold values) for the PASS (OK)
range for the reference value set in step .
SettingDescription
HIGH thresholdEnter the upper threshold for lift caused by scraps or residue.
LOW thresholdEnter the lower threshold for overshooting caused by pressing with no workpiece.
The HIGH, PASS, and LOW judgement results will be output based on the threshold
values set here.
Measurement resultJudgement
Measurement result > HIGH thresholdHIGH
LOW threshold
LOW threshold > Measurement resultLOW
≤ Measurement result ≤ HIGH threshold PASS
66
Refer to Section 5 Detailed Settings for details on operation.
Entering Threshold Values, p. 97
ZX-E
Operation Manual
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Measuring Height
This section describes how to measure the height of an object, using a rivet as an example.
Rivet: Metal
Plate: Resin
Section 4 Measuring Height
When making settings while still connected to an external device, set the Amplifier Unit’s judgement output
hold input to ON so that the output to the external device remains unchanged.
Adjust linearity before performing measurement.
Adjusting Linearity, p. 54
❚ Flow of Operation
Place an actual sensing object in position. Have a reference sample ready beforehand.
Mounting to
Device
Adjusting
Measurement
Distances
Setting
Measurement
Timing
Measuring
Reference
Samples
Setting Tolerance
Judgement Values
Operation Manual
ZX-E
67
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Mounting to Device
Mount the Sensor Head to the inspection device.
Refer to the following diagram and prepare a mounting jig.
Installing Sensor Heads, p. 30
Section 4 Measuring Height
Set screw
Mounting jig
(resin or metal)
Adjusting Measurement Distances
Place the reference sample in position and adjust the Sensor Head position. Refer to
the Amplifier Unit display and adjust the Sensor Head position so that the upper and
lower limits of the height (H) to be measured fall within the measurement distance.
Measurement distance, p. 143
68
Measured Value Display
The Amplifier Unit display shows the distance (H) from the reference sample (default).
The display can also be set to show the height of the reference sample.
Changing Display Scales, p. 90
Height of reference sample
ZX-E
Operation Manual
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Setting Measurement Timing
The hold function is used for height measurement. The minimum value (bottom) during
the sampling period is held.
When the timing signal cannot be input from the device, set a self-down trigger.
Measured value
Self-trigger level
The bottom value is held.
Sampling
Section 4 Measuring Height
The following settings are required when the reference sample height is displayed using the scaling
function:
Measurement trigger: Self-up trigger
Hold condition:Peak hold
Refer to Section 5 Detailed Settings for details on settings.
Using Hold Functions, p. 81
Operation Manual
ZX-E
69
Section 4 Measuring Height
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Measuring Reference Samples
The height of the reference sample is measured using position teaching and the measurement result is registered as the HIGH threshold value.
The registered value becomes the reference for the threshold value set in step .
Refer to Section 5 Detailed Settings for details on settings.
Position Teaching, p. 99
The reference sample height can also be set to 0.
Using the Zero Reset Function, p. 127
Setting Tolerance Judgement Values
Refer to the HIGH threshold registered in step and set the upper and lower limits
(HIGH and LOW thresholds) for a PASS (OK) judgement.
The HIGH, PASS, and LOW judgement results will be output based on the threshold
values set here.
Measurement resultJudgement
Measurement result > HIGH thresholdHIGH
LOW threshold
LOW threshold > measurement resultLOW
Refer to Section 5 Detailed Settings for details on operation.
≤ measurement result ≤ HIGH threshold PASS
70
Inputting Threshold Values Directly, p. 98
ZX-E
Operation Manual
MAIN APPLICATIONS AND SETTING METHODS
Measuring Eccentricity and Vibration
This section describes, as an example, how to measure the eccentricity of a shaft.
Section 4
When making settings while still connected to an external device, set the Amplifier Unit’s judgement output
hold input to ON so that the output to the external device remains unchanged.
Adjust linearity before performing this operation.
Adjusting Linearity, p. 54
❚ Flow of Operation
Metal shaft
Section 4 Measuring Eccentricity and Vibration
Mounting to
Device
Adjusting
Measurement
Distances
Measuring
Deflection
Setting Tolerance
Judgment Values
Operation Manual
ZX-E
71
Section 4 Measuring Eccentricity and Vibration
(
)
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Mounting to Device
Mount the Sensor Head to the inspection device.
Refer to the following diagram and prepare a mounting jig.
Installing Sensor Heads, p. 30
Set screw
Mounting jig
resin or metal
Adjusting Measurement Distances
Adjust the position of the Sensor Head so that the distance (H) between the Sensor
Head and the sensing object is about half the measurement distance, as shown in the
diagram. Refer to the Amplifier Unit display while adjusting the Sensor Head position.
Measurement distance, p. 143
72
ZX-E
Operation Manual
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Measuring Deflection
Use the peak-to-peak hold function to measure the normal deflection. Rotate the shaft,
input a timing signal from an external device, and measure the deflection. The difference between the maximum and minimum measurement results (the deflection) will be
used as a reference when setting tolerances.
Measured value
Deflection
Timing input
Refer to Section 5 Detailed Settings for details on settings.
Using Hold Functions, p. 81
Setting Tolerance Judgment Values
Refer to the deflection measured in step and set either the upper limit (HIGH
threshold) or lower limit (LOW threshold) for a PASS (OK) judgement.
The judgement result will be output based on the threshold value set here. The output
will depend on the type of threshold set.
Output when upper limit is set: PASS or HIGH
Output when lower limit is set: PASS or LOW
Measurement resultJudgement
Measurement result > HIGH thresholdHIGH
LOW threshold
LOW threshold > Measurement resultLOW
≤ Measurement result ≤ HIGH threshold PASS
Section 4 Measuring Eccentricity and Vibration
Refer to Section 5 Detailed Settings for details on settings.
Inputting Threshold Values Directly, p. 98
Operation Manual
ZX-E
73
Section 4
e
MAIN APPLICATIONS AND SETTING METHODS
Measuring Thickness
This section describes how to measure thickness, using the thickness of a steel plate as an
example.
Steel plat
Section 4 Measuring Thickness
When making settings while still connected to an external device, set the Amplifier Unit’s judgement output
hold input to ON so that the output to the external device remains unchanged.
Adjust linearity before performing this operation.
Adjusting Linearity, p. 54
❚ Flow of Operation
Mounting to
Device
Adjusting Setting
Distances
(If required)
Preventing Mutual
Interference
Setting
Expressions
Setting Tolerance
Judgement Values
74
ZX-E
Operation Manual
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Mounting to Device
■ Connecting Amplifier Units
Connect two Amplifier Units by placing a Calculating Unit between them as shown in
the diagram.
Calculating Unit
The calculation result is displayed on (i.e., output to) the CH2 Amplifier Unit. Connect
the CH2 output cable to the external device to enable external control.
Section 4 Measuring Thickness
Connections, p. 34
The CH1 Amplifier Unit will display (output) the measurement result for the CH1 Sensor Head only.
■ Mounting Sensor Heads to Inspection Device
Refer to the following diagram and prepare mounting jigs. Mount the Sensor Heads facing each other.
Installing Sensor Heads, p. 30
Set screws
Mounting jigs
(resin or metal)
Operation Manual
ZX-E
75
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Adjusting Setting Distances
Set a reference sample with a known thickness (T).
Adjust the Sensor Heads so that the distances between the reference sample and the
Sensor Heads (A and B) are each about half the measurement distance. Refer to the
Amplifier Unit display when adjusting the Sensor Heads.
Measurement distance, p. 143
Section 4 Measuring Thickness
Preventing Mutual Interference
Mutual interference prevention settings are required when the distance between Sensor Heads is less than the mutual interference distance.
These settings are not required if the Sensor Heads are further apart than the mutual
interference distance.
Setting distance
For information on mutual interference distance, refer to Mutual Interference, p. 32.
The settings are made on the CH1 Amplifier Unit. Refer to Section 6 Auxiliary Func-tions for details on settings.
Preventing Mutual Interference between Sensors, p. 120
76
ZX-E
Operation Manual
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Setting Expressions
Position the reference sample and set the expression to calculate the thickness of the
reference sample.
The settings are made on the CH2 Amplifier Unit. Select THICK as the expression type
and enter the thickness (T) of the reference sample.
Section 4 Measuring Thickness
When the thickness is entered, the positional relationship between the Sensor Heads
at that point will be registered. The thickness is measured based on the positional relationship of the Sensor Heads.
Refer to Section 6 Auxiliary Functions for details on settings.
Performing Calculations, p. 116
Setting Tolerance Judgement Values
Set the upper and lower limits (HIGH and LOW thresholds) for the thickness for a PASS
(OK) judgement.
The HIGH, PASS, and LOW judgement results will be output based on the threshold
values set here.
Measurement resultJudgement
Measurement result > HIGH thresholdHIGH
LOW threshold
LOW threshold > Measurement resultLOW
Refer to Section 5 Detailed Settings for details operations.
≤ measurement result ≤ HIGH threshold PASS
Inputting Threshold Values Directly, p. 98
Operation Manual
ZX-E
77
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Section 4 Measuring Thickness
78
ZX-E
Operation Manual
Section 5
DETAILED SETTINGS
Setting Number of Samples to Average80
Using Hold Functions81
Comparing Measured Values
(Previous Value Comparisons)88
Changing Display Scales90
Entering Threshold Values97
Linear Output103
Setting Judgement Output Timing (Timer)112
Section 5 DETAILED SETTINGS
Operation Manual
ZX-E
79
Section 5
DETAILED SETTINGS
Setting Number of Samples to Average
The number of samples to average is the number of data points used to average data
measured by the Sensor. The average value will be output.
Use the number of samples to average function to ignore sudden variations in measured
values. If the number of samples is increased, however, the response time of the judgement
outputs and linear output will be increased.
Selection for No. of
samples to average
10.3 ms
20.5 ms
40.8 ms
81.5 ms
162.5 ms
325 ms
64 (default)10 ms
12820 ms
Section 5 Setting Number of Samples to Average
Moving to FUN mode and AVE
25640 ms
51275 ms
1024150 ms
2048300 ms
4096600 ms
Response time
1. Set the mode switch to FUN.
2. Use the LEFT and RIGHT Keys to display
AVE on the main display.
80
Selecting Number of Samples to Average
3. Press the UP or DOWN Key.
The sub-display will flash.
4. Use the UP and DOWN Keys to select the
number of samples to average.
5. Press the ENT Key to confirm the selection.
The setting will be registered.
ZX-E
Operation Manual
Section 5
p
DETAILED SETTINGS
Using Hold Functions
The hold functions hold data for specific points during the measurement period, such as the
maximum or minimum value, and output those values at the end of the measurement period.
❚ Flow of Operation
(If required)
1
Selecting Hold Condition
for Measured Values
2
Setting the Trigger for
Measurement in Hold
Mode
Selecting Hold Condition for Measured Values
The time period from the start of hold measurements to the end of hold measurements
is called the sampling period.
The value to be held during that sampling period is selected here.
3
Setting Delay Time
Section 5 Using Hold Functions
The CLAMP value is output until the first sampling period is finished.
CLAMP value, p. 110
Any of the 5 settings shown in the table can be selected as the value to hold.
SelectionDetails
OFF
(Default)
P-H (Peak hold)Holds the maximum value during the sampling period. The output changes when
B-H (Bottom hold)Holds the minimum value during the sampling period. The output changes at the
Hold measurement is not performed. The measured value is always output.
the sampling period is finished and is held then until the end of the next sampling
period.
Maximum
Current measured value
end of the sampling period and is held until the end of the next sampling period.
Current measured value
value
ling period
Sam
Minimum
value
Sampling period
Output
Output
Operation Manual
81
ZX-E
Section 5
t
g
DETAILED SETTINGS
SelectionDetails
PP-H (Peak-to-peak
hold)
Holds the difference between the maximum and minimum values. This option is
selected mainly when detecting vibration. The output changes at the end of the
sampling period and is held until the end of the next sampling period.
Maximum
Current measured value
value
Minimum
value
Sampling period
Output (difference between
maximum and minimum)
S-H (Sample hold)Holds the measured value at the start of the sampling period. The output
AVE-H (Average hold)Holds the average measured value during the sampling period. The output
Section 5 Using Hold Functions
Moving to FUN mode and HOLD
changes at the start of the sampling period and is held until the start of the next
sampling period.
Current measured value
Sampling period
changes at the end of the sampling period and is held until the end of the next
sampling period.
Current measured value
Samplin
period
Outpu
Output (average)
1. Set the mode switch to FUN.
2. Use the LEFT and RIGHT Keys to display
HOLD on the main display.
82
Selecting Hold Condition
3. Press the UP or DOWN Key.
The sub-display will flash.
4. Use the UP and DOWN Keys to select the
hold condition.
5. Press the ENT Key to confirm the selection.
The setting will be registered.
ZX-E
Operation Manual
Section 5
p
p
DETAILED SETTINGS
Setting the Trigger for Measurement in Hold Mode
Select the input method for the timing of the start and end of the measurement period.
SelectionDetails
TIMIG
(Timing input)
(Default)
Enter the trigger for the start of sampling by using the timing input. The period
that the timing signal is ON is the sampling period.
Timing input
Sampling period
When a delay time is set, the input OFF timing and the end of the sampling period will not be synchronous. Sampling will end after the specified sampling period has expired.
UP
(Self-up trigger)
DOWN
(Self-down trigger)
The sampling period is the period that the measured value is greater than the
specified self-trigger level. Hold measurement is possible without a sync input.
Self-trigger
level
Measured
value
Sampling
eriod
When a delay time is set, the timing when the measured value becomes
smaller than the self-trigger level and the end of the sampling period will
not be synchronous. Sampling will end after the specified sampling
period has expired.
The sampling period is the period that the measured value is lower than the
specified self-trigger level. Hold measurement is possible without a sync input.
Measured
value
Self-trigger
level
Sampling
period
Sampling
eriod
Hysteresis width
(for self-trigger)
Operating value
Release value
Hysteresis width
(for self-trigger)
Operating value
Release value
Section 5 Using Hold Functions
When a delay time is set, the timing when the measured value becomes
greater than the self-trigger level and the end of the sampling period will
not be synchronous.
Hysteresis (hysteresis width)
Set the hysteresis based on the fluctuations in the measured values around the trigger level. The
hysteresis will be applied from the start of the sampling period and will prevent timing input chattering.
ZX-E
Operation Manual
83
Section 5
DETAILED SETTINGS
Selecting Trigger Types
1. Use the LEFT and RIGHT Keys to display H-
TRG on the main display.
H-TRG will not be displayed if the hold condition is set to
OFF.
2. Press the UP or DOWN Key.
The sub-display will flash.
3. Use the UP or DOWN Key to select the trig-
ger type.
4. Press the ENT Key to confirm the selection.
Section 5 Using Hold Functions
The selected trigger type will be registered.
Setting Trigger Levels (when UP or DOWN Is Selected)
5. Use the RIGHT or LEFT Key to display H-
LVL on the main display.
H-LVL will not be displayed if timing input (TIMIG) is
selected as the trigger.
6. Press the UP or DOWN Key.
The leftmost digit of the sub-display will flash.
7. Use the Cursor Keys to set the trigger level.
8. Press the ENT Key to confirm the settings.
The trigger level will be registered.
Move between digits.
Increment and decrement the numeric value.
84
ZX-E
Operation Manual
Setting Hysteresis Width (when UP or DOWN Is Selected)
9. Use the LEFT and RIGHT Keys to display H-
HYS on the main display.
H-HYS will not be displayed if timing input (TIMIG) is set
as the trigger.
10. Press the UP or DOWN Key.
The leftmost digit of the sub-display will flash.
Section 5
DETAILED SETTINGS
11. Use the Cursor Keys to set the hysteresis
width for the trigger level.
12. Press the ENT Key to confirm the setting.
The hysteresis width will be registered.
Move between digits.
Increment and decrement the numeric value.
Section 5 Using Hold Functions
Operation Manual
ZX-E
85
Section 5
g
DETAILED SETTINGS
Setting Delay Time
Delay time is set to ignore measurements immediately after the timing input. This is
useful for avoiding bounding during device startup and the influence of machine vibration.
The delay time (the delay between timing input and the start of sampling) and the sampling period can be set.
The default delay time setting is OFF.
Make the sum of the delay time and sampling period less than the timing input ON interval.
If the next timing input for measurement is received before the delay and sampling period have
passed, that timing input will be ignored and will not be reflected in the sampling.
Delay time
Timing input
Section 5 Using Hold Functions
Moving to Delay Hold (H-DLY)
Smpling period
Timing input ignored.
Timin
Delay time
input used.
1. Use the LEFT and RIGHT Keys to display H-
DLY on the main display.
H-DLY will not be displayed if the delay conditions are set
to OFF.
2. Press the UP or DOWN Key.
The sub-display will flash.
86
3. Use the UP and DOWN Keys to display ON.
4. Press the ENT Key.
The settings mode for the H-DLY function is now
enabled.
ZX-E
Operation Manual
Setting Delay Time
5. Use the LEFT and RIGHT Keys to display H-
D-T on the main display.
H-D-T will not be displayed if the H-DLY function is set to
OFF.
6. Press the UP or DOWN Key.
The leftmost digit of the sub-display will flash.
Section 5
DETAILED SETTINGS
7. Enter the delay time (ms).
8. Press the ENT Key to confirm the setting.
The delay time will be registered.
Setting Sampling Period
9. Use the LEFT and RIGHT Keys to display H-
S-T on the main display.
H-S-T will not be displayed if the H-DLY function is set to
OFF.
10. Press the UP or DOWN Key.
The leftmost digit of the sub-display will flash.
11. Use the Cursor Keys to enter the sampling
period (ms).
Move between digits.
Increment and decrement the numeric value.
Section 5 Using Hold Functions
Move between digits.
12. Press the ENT Key to confirm the setting.
The sampling period will be registered.
Increment and decrement the numeric value.
Operation Manual
ZX-E
87
Section 5
DETAILED SETTINGS
Comparing Measured Values
(Previous Value Comparisons)
Use the previous value comparison function to ignore gradual changes in measured values
over time, due to factors such as temperature drift, and only detect and judge sudden
changes.
The hold function must be set before previous value comparison can be set. The difference
from the previous hold value with a PASS judgement becomes the measured value. For
example, if the judgement for that previous measurement is HIGH or LOW, the comparison is
performed with the hold value before that.
Sudden change
Hold value
Section 5 Comparing Measured Values (Previous Value Comparisons)
Measured value
(after comparison
with previous value) X(n)-X(n-1)
The hysteresis width setting will be disabled if the previous value comparison function is used.
Hysteresis Setting, p. 102
Moving to FUN mode and SPCL
Gradual change
HIGH threshold
1. Set the mode switch to FUN.
88
2. Use the LEFT and RIGHT Keys to display
SPCL on the main display.
ZX-E
Operation Manual
Moving to COMP
3. Press the UP or DOWN Key.
The sub-display will flash.
4. Use the UP and DOWN Keys to display SET
or ALL.
5. Press the ENT Key.
6. Use the LEFT and RIGHT Keys to display
COMP on the main display.
Section 5
DETAILED SETTINGS
Section 5 Comparing Measured Values (Previous Value Comparisons)
Setting Previous Value Comparison
7. Press the UP or DOWN Key.
The sub-display will flash.
8. Use the UP and DOWN Keys to display ON.
9. Press the ENT Key to confirm the setting.
The previous value comparison function can now be
used.
Operation Manual
ZX-E
89
Section 5
DETAILED SETTINGS
Changing Display Scales
Change the display scale when you want to display a value different on the main display to the
actual measured value.
Place the sensing object in position and make the settings for either one-point or two-point
scaling.
Section 5 Changing Display Scales
The scaling set here is reflected in the display only. The output does not change.
The minimum display value is
the minimum display value after scaling is executed, the display will be
greater than the maximum, the display value will be 59,999.
Scaling Application Example
PurposeSettings Method
To display distance from Sensor HeadOffset the display value.One-point scaling
Sensor
Head
Value to display:
2.5 mm
Sensing
object
Actual
measured
value:
1.0 mm
−19,999 and the maximum is 59,999. If the measurement result is less than
Display value (mm)
−19,999. If the measurement result is
with no inverted
display values
p. 92
Actual measured
value (mm)
90
To display the height of the sensing objectOffset the display value and invert the relation-
Sensor
Head
Actual measured value:
1.0 mm
Sensing
object
Value to display:
1.5 mm
ship between display and reference values.
Display value (mm)
Actual measured
value (mm)
ZX-E
Operation Manual
One-point scaling
with inverted display values
p. 92
PurposeSettings Method
To correct display values to match actual distances
Sensor Head
Actual distance: 1.0 mm
Value to display: 0.8 mm
Sensing
object
Slide gauge
DETAILED SETTINGS
Offset the display value and change the display value range.
Display value (mm)
Actual measured
value (mm)
Section 5
Two-point scaling
p. 95
Section 5 Changing Display Scales
Actual distance: 0.2 mm
Value to display: 0.3 mm
Sensing
object
The settings listed below return to the default settings when scaling is set.
Make the settings for these items after scaling settings have been completed.
Self-trigger level, p. 83
Output Settings (Monitor Focus), p. 103
Using the Zero Reset Function, p. 127
Finding Thicknesses, p. 118
Operation Manual
ZX-E
91
Section 5
DETAILED SETTINGS
One-point Scaling
With one-point scaling, measurement is performed for one position and an offset value
is set for that measurement. Both offset and display inversions (changing the relationship between increasing and decreasing values) can be set.
This section describes how to make settings for one-point scaling, using an different
example.
Example: Displaying the Height of the Sensing Object
Display value (mm)
Sensor Head
Actual distance: 1.0 mm
Section 5 Changing Display Scales
Sensing
object
Moving to FUN mode and SPCL
Value to display: 1.5 mm
1. Set the mode switch to FUN.
2. Use the LEFT and RIGHT Keys to display
SPCL on the main display.
Moving to SCALE
3. Press the UP or DOWN Key.
The sub-display will flash.
Actual measured
value (mm)
92
4. Use the UP and DOWN Keys to display
SET or ALL.
5. Press the ENT Key.
ZX-E
Operation Manual
6. Use the LEFT and RIGHT Keys to display
SCALE on the main display.
Executing Scaling
7. Press the UP or DOWN Key.
The sub-display will flash.
8. Use the UP and DOWN Keys to display ON.
9. Press the ENT Key to confirm the setting.
The sub-display will show P1SCL.
Section 5
DETAILED SETTINGS
Section 5 Changing Display Scales
10. Set the sensing object at the position where
the display value change is required.
Set the sensing object within the measurement distance. The ENABLE indicator will be lit when the
sensing object is within the measurement distance. Scaling is not possible if the sensing object is
not within the distance.
11. Press one of the Cursor Keys.
The current measured value will be displayed on the
main display.
The leftmost digit of the sub-display will flash
.
12. Use the Cursor Keys to set the offset for the
measured value on the sub-display.
The position of the decimal point can be changed using
the next steps.
13. Press the ENT Key to confirm the settings.
The decimal point will flash.
14. Use the LEFT and RIGHT Keys to change the
position of the decimal point, if necessary.
Move between digits.
Increment and decrement the numeric value.
Operation Manual
ZX-E
93
Section 5
DETAILED SETTINGS
15. Press the ENT Key to confirm the settings.
The inverted display settings will appear on the sub-display.
Selecting Inverted Display
16. Use the UP and DOWN Keys to inverted dis-
plays.
OptionsDetails
D-FWDNot inverted.
D-INVInverted
(Inverts the increase/decrease relationship
between display and measured values)
Section 5 Changing Display Scales
17. Press the ENT Key to confirm settings.
The sub-display will be P2SCL.
18. Press the ENT Key.
Confirming that Scaling Is Completed
If scaling has been completed correctly, the display will
be OK.
If scaling was unable to be completed, the display will be
NG.
Check that the sensing object is within the measurement
distance and execute scaling again.
94
ZX-E
Operation Manual
Section 5
DETAILED SETTINGS
Two-point Scaling
Measurement is performed at two positions and offset values are set for those measurements. Both an overall offset can be set and the range can be changed.
This section describes how to set two-point scaling, using an example of correcting display values to match actual distances.
Example: Correcting Display Values to Match Actual Distances
Display value (mm)
Sensor Head
Actual distance: 1.0 mm
Value to display: 0.8 mm
Sensing
object
Actual measured value (mm)
Slide gauge
Actual distance: 0.2 mm
Value to display: 0.3 mm
Sensing
object
Separate the two specified points by at least 1% of the rated measurement distance for the connected Sensor.
For example, the rated measurement distance for the ZX-ED01T Sensor is 1 mm. Therefore, the
two specified points must be separated by 10
Setting the First Point
1. Set the first point by following steps 1. to 15.
of the one-point scaling procedure.
p. 92
Section 5 Changing Display Scales
µm min.
The range for two-point scaling is automatically
set based on the values entered for the two
points. Inverted display settings are ignored.
Operation Manual
ZX-E
95
Section 5 Changing Display Scales
Section 5
DETAILED SETTINGS
Setting the Second Point
2. Place the sensing object in the position for which the display is to be changed (the sec-
ond point).
The sensing object must be set at a distance at
least 1% of the rated measurement distance
away from the first point and also at a distance
within the measurement distance range.
3. Press one of the Cursor Keys.
The current measured value will be displayed on the
main display. The leftmost digit of the sub-display will
flash.
4. Use the Cursor Keys to set the offset for the
measured value.
The position of the decimal point can be changed using
the next steps.
Move between digits.
Increment and decrement the numeric value.
5. Press the ENT Key to confirm the settings.
The decimal point will flash.
6. Use the LEFT and RIGHT Keys to move the
decimal point.
7. Press the ENT Key to confirm this setting.
Confirming the Completion of the Scaling Settings
If scaling has been completed correctly, the display will
be OK.
If scaling was unable to be completed, the display will be
NG. Check the following points and then execute scaling
again.
• Is the sensing object within the measurement distance?
• Are the two points separated by at least 1% of the rated
measurement distance?
96
ZX-E
Operation Manual
Section 5
DETAILED SETTINGS
Entering Threshold Values
Threshold values are set to determine the range for PASS judgements. Both HIGH and LOW
threshold values are set. There are three judgement outputs: HIGH, PASS, and LOW.
HIGH threshold value
Measured value
LOW threshold value
(ON when measured value >
HIGH threshold value)
Output
(ON when LOW threshold value ≤
Measured value ≤ HIGH threshold
(ON when measured value <
LOW threshold value)
The following table outlines the three methods for setting the threshold values.
MethodDetails
Direct inputSets threshold values by direct numerical value input.
Direct input is useful when you know the dimensions for an OK judgement or when
you want to fine-tune threshold values after teaching.
Position teachingPerforms measurement and uses the measurement results to set threshold values.
Position teaching is useful when threshold samples, i.e., with the upper and lower
limits, can be obtained beforehand.
Automatic teachingPerforms measurement continuously while the keys are held down and sets the
Hysteresis (hysteresis width) can also be set for threshold values. Set hysteresis when judgements are
unstable to prevent chattering.
p. 102
maximum and minimum measurements during that period as the threshold values.
Automatic teaching is useful when you want to set threshold values by starting the
device and obtaining real measurements.
Section 5 Entering Threshold Values
When making settings while still connected to an external device, set the Amplifier Unit judgement’s output
hold input to ON so that the output to the external device remains unchanged. The judgement outputs in T
mode will be the same as in RUN mode, i.e., HIGH, PASS, and LOW.
ZX-E
Operation Manual
97
Section 5
DETAILED SETTINGS
Inputting Threshold Values Directly
The threshold values can be set by directly inputting the numeric values.
Direct input is useful when the dimensions for an OK judgement are known beforehand
or when fine-tuning threshold values after teaching.
Moving to T Mode
1. Set the mode switch to T.
Setting Threshold Values
2. Move the threshold switch to either H or L,
i.e., the threshold to be set.
The current measured value is shown on the main dis-
Section 5 Entering Threshold Values
play and the threshold being set (value for the threshold
switch selection) will be displayed on the sub-display.
3. Press one of the Cursor Keys.
The leftmost digit of the threshold value will flash on the
sub-display.
4. Use the Cursor Keys to set the threshold
value.
5. Press the ENT Key to confirm the settings.
The threshold values will be registered.
Countermeasures for Setting Errors
One of the following two errors has occurred if the display reads ERRLH or ERRHL.
HIGH threshold value < LOW threshold value
HIGH threshold value
sis width
Hysteresis Setting, p. 102
− LOW threshold value < hystere-
Move between digits.
Increment and decrement the numeric value.
98
If an error is displayed, the threshold value has not been
updated. Set the threshold values again.
ZX-E
Operation Manual
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.