HEM-7200-AP3_A_M03_120823.pdf
Automatic Blood
Pressure Monitor
Model JPN1
Instruction Manual
1883264-4A
Introduction
Thank you for purchasing the OMRON JPN1 Automatic Blood Pressure
Monitor.
The OMRON JPN1 is a compact, fully automatic blood pressure monitor,
operating on the oscillometric principle. It measures your blood pressure and
pulse rate simply and quickly. For comfortable controlled inflation without the
need of pressure pre-setting or re-inflation the devices uses its advanced
“IntelliSense” technology.
Intended Use
This product is designed to measure the blood pressure and pulse rate of
people within the range of the designated arm cuff, following the instructions in
this instruction manual.
It is mainly designed for general household use. Please read the Important
Safety Information in this instruction manual before using the unit.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
Important Safety Information
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
or arteriosclerosis.
Please read this section carefully before using the unit.
Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result
in death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a
doctor before using the unit, as cuff inflation can cause internal bleeding.
(AC Adapter Usage)
• Never plug in or unplug the power cord from the electric outlet with wet
hands.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury to the user or patient or damage to the equipment or
other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express
their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result
in incorrect measurement results.
• Make sure that the air tube is not wrapped around other parts of your body
when taking measurements. This could result in injury when the air pressure
in the air tube is increased.
• Do not leave the cuff wrapped on the arm if taking measurements during the
night. This could result in injury.
• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
• Do not operate the unit in a moving vehicle (car, airplane).
• To inflate the cuff manually, refer to Section 3.3. If the cuff is over inflated, it
can cause internal bleeding.
(AC Adapter Usage)
• Use only the original AC adapter (optional) designed for this unit. Use of
unsupported adapters may damage and/or may be hazardous to the unit.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a
multiple-tap.
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off
the power and unplug the power cord immediately.
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with
plenty of clean water.
• Use only four “AA” alkaline or manganese batteries with this unit. Do not use
other types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four batteries at
the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol or caffeine,
smoking, exercising or eating.
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with the
main unit not the tube itself.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
• Read and follow the “Important information regarding Electro Magnetic
Compatibility (EMC)” in the Technical Data Section.
• Read and follow the “Correct Disposal of This Product” in the Technical Data
Section when disposing of the device and any used accessories or optional
parts.
1. Overview
Main unit:
A
F
E
A. Display
B. START/STOP button
C. Memory button
D. Up/Down ( ) buttons
E. Buzzer and Date/Time
setting ( ) button
B
C
D
D
F. Air Jack
G. Battery compartment
H. AC adapter jack (for optional AC
adapter)
G
Arm cuff:
I J K
L
Display:
M. Systolic blood pressure
N. Diastolic blood pressure
O Heartbeat symbol
(Flashes during measurement)
P. Memory symbol
Q. Deflation symbol
R. Average value symbol
M
N
O
P
Q
R
S
T
I. Arm cuff (Medium cuff: arm
circumference 22-32cm)
J. Marker
K. Air tube
L. Air plug
U
V
W
X
S. Movement error symbol
T. Date/Time display
U. Blood pressure level indicator
V. Pulse display
W. Batterylowsymbol
X. Irregularheartbeatsymbol
2. Preparation
2.1 Installing/Replacing the Batteries
Remove the battery cover.
1.
2.
Insert four “AA” batteries
as indicated in the battery
compartment and then
replace the battery cover.
Notes:
• If the battery low symbol ( ) appears on the display, turn the unit off
then replace all batteries at the same time.
• The measurement values continue to be stored in memory even after
the batteries are replaced.
Disposal of used batteries should be carried out in accordance with the
national regulations for the disposal of batteries.
2.2 Setting the Buzzer/ Date and Time
Press the button.
1.
Set the monitor to the correct date and time before
2.
taking a measurement for the first time.
1)
Push or buttons to
change.
2)
Push the button to confirm.
Next setting appears.
Year
Month
Day
Hour
Minute
Press the START/STOP button to store the setting.
3.
Notes:
• If the batteries have been removed for 30 seconds or more, the date
and time setting will need to be reset.
• If the date and time are not set, “-:--” appears during or after
measurement.
3. Using the Unit
3.1 Applying the Arm Cuff
Note: Wait 2-3 minutes before taking another blood pressure
measurement. Waiting between readings allows the arteries to
return to the condition prior to taking the blood pressure
measurement.
Remove tight-fitting clothing or tight rolled up sleeve from your
upper arm.
H
Do not place the cuff over thick clothes.
1. Insert the air plug into the air jack.
2. Put your arm through the cuff loop.
Warning:
Self-diagnosis of measured results and treatment are dangerous.
Please follow the instructions of your doctor.
Remove the arm cuff.
2.
Press the START/STOP button to turn off the monitor.
3.
The monitor automatically stores the measurement in its
memory.
It will automatically turn off after five minutes.
Important:
• Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
3. Position the arm correctly.
Systolic Blood Pressure Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg
This criteria is for home blood pressure measurement.
• Your blood pressure monitor includes an irregular
heartbeat feature. Irregular heartbeats can influence
the results of the measurement. The irregular
The bottom edge of the cuff should be 1 to 2 cm above the elbow.
Marker (arrow under tube) is centred on the middle of your inner arm.
Close the fabric fastener FIRMLY.
Notes:
• When you take a measurement on the right
arm, air tube will be at the side of your
elbow. Be careful not to rest your arm on
the air tube.
• The blood pressure can differ between the right arm and the left arm,
and therefore also the measured blood pressure values can be
different. OMRON recommends to always use the same arm for
measurement. If the values between the two arms differ substantially,
please check with your doctor which arm to use for your
measurement.
3.2 How to Sit Correctly
To take a measurement, you need to be
relaxed and comfortably seated, under
comfortable room temperature. No eating,
smoking or exercising 30 minutes before
taking a measurement.
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• The cuff should be at the same level as your
heart.
3.3 Taking a Reading
Notes:
• To cancel a measurement, press the START/STOP button to turn off
the unit and to release the air in the arm cuff.
• Remain still while taking a measurement.
Press the START/STOP button.
1.
The cuff will start to inflate automatically.
START
INFLATE DEFLATE END
1 to 2 cm
3.4 Using the Memory Function
COMPLETE
heartbeat algorithm automatically determines if the
measurement is usable or needs to be repeated. If the
measurement results are affected by irregular
heartbeats but the result is valid, the result is shown
together with the irregular heartbeat symbol ( ). If the
irregular heartbeats cause the measurement to be
invalid, no result is shown. If the irregular heartbeat
symbol ( ) is shown after you have taken a measurement, repeat
the measurement. If the irregular heartbeat symbol ( ) is shown
frequently, please make your doctor aware of it.
The monitor automatically stores the result up to 60 sets. It can also
calculate an average reading based on the measurements from the last
three readings taken within 10 minutes. If there are only two readings in
memory for that period, the average will be based on two readings. If
there is one reading in memory for that period, the average will be
based on one reading.
Notes:
• If the memory is full, the monitor will delete the oldest readings.
• When viewing the reading taken without setting the date and time,
“-:--” is displayed instead of the date and time.
Average Value
View the Average
Value
• If the previous
reading was taken
without setting the
date and time, the
average value is not
calculated.
View the Previous
Reading
• The Memory number
appears for a second
before the pulse rate
is displayed. The
newest set is
numbered “1”.
Previous Reading
Alternating
date/time display
RecentOld
View the Readings
Stored in Memory
To delete all the values stored in memory
When the memory symbol ( ) appears, first press the Memory
button. Then while holding it down, press the START/STOP button
simultaneously for about 2-3 seconds.
Note: You cannot partially delete the stored readings.
4. Troubleshooting and Maintenance
4.1 The Icons and Error Messages
Cause RemedyError Display
Remove the arm cuff. Wait
2-3 minutes and then take
Irregular heartbeats are
detected.
Movement during
measurement.
Blink
The batteries are low.
The batteries are
Lit
exhausted.
Cuff is under inflated.
Movement during
measurement.
Air plug disconnected.
Arm cuff not applied
correctly.
Clothing is interfering with
the arm cuff.
Air is leaking from the arm
cuff.
The arm cuff was inflated
above 299 mmHg when
inflating the cuff manually.
Device error.
Note: The irregular heartbeat symbol ( ) may also be displayed with error
messages.
another measurement.
Repeat the steps in section
3.3. If this error continues to
appear, contact your doctor.
Carefully read and repeat
the steps in section 3.3.
You should replace them
with new ones ahead of
time.
Refer to section 2.1.
You should replace them
with new ones at once.
Refer to section 2.1.
Carefully read and repeat
the steps listed under
section 3.3.
Repeat measurement.
Remain still and do not talk
during measurement.
Refer to section 3.3.
Insert the air plug securely.
Refer to section 3.1.
Apply the arm cuff correctly.
Refer to section 3.1.
Remove any clothing
interfering with the arm cuff.
Refer to section 3.1.
Replace cuff with new one.
Refer to Chapter 5.
Do not inflate the arm cuff
above 299 mmHg.
Refer to section 3.3.
Contact your local OMRON
representative.
4.2 Troubleshooting
Problem
The reading is
extremely low (or
high).
Arm cuff pressure
does not rise.
Arm cuff deflates too
soon.
Cannot measure or
readings are too low
or too high.
Nothing happens
when you press the
buttons.
Other problems.
Cause Remedy
Arm cuff not applied
correctly.
Movement or talking
during measurement.
Clothing is interfering
with the arm cuff.
The air tube is not
securely connected
into the main unit.
Air is leaking from the
arm cuff.
The arm cuff is loose.
The arm cuff has not
been inflated
sufficiently.
The batteries are
empty.
The batteries have
been inserted
incorrectly.
• Press the START/STOP button and repeat
measurement.
• If the problem continues, try replacing the
batteries with new ones.
If this still does not solve the problem, contact
your local OMRON representative.
Apply the arm cuff
correctly.
Refer to section 3.1.
Remain still and do not
talk during
measurement.
Refer to section 3.3.
Remove any clothing
interfering with the arm
cuff.
Refer to section 3.2.
Make sure that the air
tube is connected
securely.
Refer to section 3.1.
Replace the arm cuff
with a new one.
Refer to Chapter 5.
Apply the cuff correctly
so that it is firmly
wrapped around the arm.
Refer to section 3.1.
Inflate the cuff so that it
is 30 to 40 mmHg above
your previous
measurement result.
Refer to section 3.3.
Replace the batteries
with new ones.
Refer to section 2.1.
Insert the batteries with
the correct (+/-) polarity.
Refer to section 2.1.
If your systolic pressure is more than
220 mmHg
After the cuff starts to inflate, press and hold the
START/STOP button until the monitor inflates 30 to 40 mmHg
higher than your expected systolic pressure.
Notes:
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
4.3 Maintenance
To protect your unit from damage, please observe the following:
• Do not subject the main unit and the cuff to extreme
temperatures, humidity, moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
• Do not inflate the arm cuff over 299 mmHg.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations
(for example, dropping the unit on the floor).
• Do not use volatile liquids to clean the main unit.
• Do not wash the arm cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to clean
the arm cuff.
• Do not carry out repairs of any kind yourself. If a defect
occurs, consult your local OMRON representative.
• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and soap to clean the arm
cuff.
Calibration and Service
• The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
• It is generally recommended to have the unit inspected
every two years to ensure correct functioning and
accuracy. Please consult your local OMRON
representative.
4.4 Storage
1. Unplug the air tube from the air connector.
2. Gently fold the air tube into the arm cuff.
Note:
Do not bend the air tube excessively.
Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight,
dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at an
angle.
5. Optional Parts
Medium Arm Cuff
Arm circumference
22 - 32 cm
Note: Please check with your local OMRON representatives for the
appropriate optional part models.
Using the Optional AC Adapter
1.
Insert the AC adapter plug
into the AC adapter jack on
the rear side of the main
unit.
2.
Plug the AC adapter into an
electrical outlet.
To disconnect the AC adapter, unplug the AC adapter from the
electrical outlet first and then remove the AC adapter plug from the
main unit.
Large Arm Cuff
Arm circumference
32 - 42 cm
AC Adapter
6. Technical Data
Product Description
Model
Display
Measurement Method
Measurement Range
Accuracy
Inflation
Deflation
Memory
Rating
Power Source
Battery Life
Applied Part
Protection Against
Electric Shock
Operating
temperature/ Humidity
Storage temperature/
Humidity/ Air pressure
Console Weight
Cuff Weight
Outer Dimensions
Cuff Dimensions
Cuff Material
Package Content
Note: Subject to technical modification without prior notice.
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
• This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood
pressure measuring systems.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices. Electromagnetic interference
may result in incorrect operation of the medical device and create a potentially
unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
standard has been implemented. This standard defines the levels of immunity to
electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation
of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at
OMRON HEALTHCARE EUROPE at the address mentioned in this instruction
manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
of its working life. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate
this from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
take this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
of the purchase contract. This product should not be mixed with other commercial
wastes for disposal.
This product does not contain any hazardous substances.
Manufacturer
EU-representative
Production facility
Asia Pacific HQ
Automatic Blood Pressure Monitor
JPN1
LCD Digital Display
Oscillometric method
Pressure: 0 to 299 mmHg
Pulse: 40 to 180/min.
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Fuzzy-logic controlled by electric pump
Automatic pressure release valve
60 Measurements with date and time
DC6V 4W
4 “AA” batteries 1.5V
or AC adapter (optional, INPUT AC100-240V
50/60Hz 0.12A)
Approx. 1500 measurements with new alkaline
batteries at a room temperature of 23°C
= Type B
Internally powered ME equipment (When using only
the batteries)
= Class II ME equipment (Optional AC adapter)
+10 to +40°C / Maximum: 30 to 90% RH
-20 to +60°C / Maximum: 10 to 95% RH
/ 700-1060 hPa
Approx. 340g without batteries
Approx. 120g
Approx. 123 (w) mm × 85 (h) mm × 141(l) mm
Approx. 146 mm × 446 mm (Medium cuff: arm
circumference 22 to 32 cm)
Nylon and polyester
Main unit, cuff, instruction manual, battery set
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp
THE NETHERLANDS
www.omron-healthcare.com
OMRON HEALTHCARE Co., Ltd.
Matsusaka Factory
Mie, JAPAN
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08
Alexandra Technopark, Singapore 119967
www.omron-healthcare.com.sg
Made in Japan