Omron IntelliSense HEM-907 Instruction Manual

OMRON IntelliSense™ Blood Pressure Monitor
Model
HEM-907
Table of Contents
Be Sure to Read This Section
Exemptions ........................................2
Know Your Unit
Features of the Product......................7
Components of the Product ...............8
Options...........................................9
Names and Functions of the Parts...10
Names of the Parts.......................10
Functions and setting
while the unit is in use ..................12
Preparations before Measurement
How to Apply the Cuff ......................14
How to Use the Power Source.........16
How to Use the AC Adapter............16
Installation and Replacement
of Battery Pack.............................17
How to Measure Blood Pressure
List of Measurement Modes.............19
SINGLE Mode ..............................20
AVG Mode....................................22
MANU Mode.................................24
How to Check Pressure Accuracy..26
How to Clean the Unit after Use ........27
List of Error Codes...........................28
Troubleshooting ...............................29
Specifications
Caution.............................................30
Five Year Limited Warranty..............31
Thank you very much for purchasing the OMRON IntelliSense™ Blood Pressure Monitor.
Please thoroughly read this Instruction Manual before using this monitor to insure safe and accurate use.
Please keep this manual near the monitor all the time for future reference.
This Instruction Manual contains the Quality Guarantee Card.
INSTRUCTION MANUAL
2
Exemptions
OMRON does not accept liability and warranty becomes void under the following circumstances:
1. When persons, not authorized by OMRON, perform repairs or modifications of this product.
2. When use and/or operation of this device is adversely effected by a device not manufactured by OMRON.
3. When use and/or operation of this device is adversely effected by use of parts, not authorized by OMRON, to repair or modify this product.
4. When Notes on Safety or Instructions for Use contained in this manual are not followed.
5. When use and/or operation of this device is caused by an act of God, such as fire, earthquake, flood or other natural disasters.
1. The product and contents of this Instruction Manual may be changed without prior notice.
2. We have prepared the contents of this Instruction Manual thoroughly. However, if an inadequate description or error is found, please let us know.
3. Reproducing or copying any or all of this Instruction Manual without OMRON’s written consent is prohibited.
Specifications
3
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
Notes on Safety
• The warning signs and the sample icons shown here are listed to insure safe and accurate use.
• The icons and meanings are as follows.
Warning sign Contents
Indicates matters in which death or severe bodily damage may arise as a result of incorrect handling.
Indicates matters in which bodily harm or material damage* may arise as a result of incorrect handling.
* Material damage refers to a wide range of damage involving your house, household goods, domestic animals, and pets.
Examples of signs
The icon indicates caution (including warning and danger). Matters involving actual caution are indicated by text or pictures in or near . The pictured icon refers to "caution for flammability".
The icon indicates prohibitions (what you cannot do). Matters involving actual prohibitions are indicated by text or pictures in or near . The pictured icon refers to "prohibition to disassemble".
The icon indicates something that is compulsory (always follow). Matters involving actual compulsory actions are indicated by text or pictures in or near . The pictured icon refers to "unplugging the power source plug".
Do not wrap the cuff over an arm to which intravenous injection or transfusion is being conducted, or when otherwise contraindicated.
Do not connect the air tube or the cuff to other equipment which is connected to an intracorporeal organ. Air embolisms may result.
Self diagnosis of measured results or treatment is dangerous. Please follow the instruction of the doctor or healthcare provider.
If cuff inflation does not stop, remove the cuff or pull out the air tube from the main unit.
If battery fluid gets into your eye or comes in contact with skin, wash the effected area with water repeatedly. Immediately consult a doctor for treatment.
Do not use this unit in the presence of flammable gas or anesthetics or in a high pressure oxygen room or oxygen tent.
Do not use the battery pack (option) for devices other than for this unit. Do not disassemble the battery pack.
Do not touch the AC adapter with wet hands.
4
Unplug the AC adapter from the electric outlet if this unit is unused for an extended period of time.
Unplug the AC adapter from the electric outlet when installing, removing, or cleaning the unit.
Confirm readings with a stethoscope when an irregular pulse wave is displayed or when the measured value is questionable or erratic.
Use an AC adapter indicated for use with a power supply of 110 VAC. Do not share an electric outlet with other unit or electric appliance. After cleaning this unit, dry it well before plugging the AC adapter in the
electric outlet. If this unit fails to perform as indicated, discontinue use, turn off the unit,
unplug the AC adapter from the electric outlet, and contact OMRON’s repair department.
Do not disassemble or modify this unit.
Notes on Safety
Do not use any cuff other than the models exclusive for this unit. Do not use this unit on infants. Do not use this unit on patients using a pump oxygenator. Do not use an AC adapter or battery pack not specified for this unit.
Specifications
5
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
Notes on Safety
Do not use a cellular phone near this unit. Do not use this unit in a vehicle. Do not install the parts and/or instruments not specified for this unit. Do not use a broken power cord or AC adapter.
Do not install or store this unit where it may come in contact with water or liquid medication.
General advice
Do not place or put anything on this unit. Do not drop this unit. Turn off power to the unit and unplug the AC adapter from the electric outlet before moving
the unit. Read the instruction manual of the other devices to be used at the same time with this unit, to
understand and be aware of the interaction between the devices. When using the unit
• Do not inflate the cuff without being wrapped over the arm.
• Do not use a damaged cuff.
• Be sure that patients do not touch the Buttons of this unit.
After using the unit
• Do not disinfect this unit by autoclave or gas sterilization (EtO, gluteraldehyde, or high concentration ozone).
Do not install or store this unit in the following places.
• Under the direct sunlight,
• Dusty or salty environment,
• Places having slope, vibration, and/or shock,
• Storage of chemicals or where combustable gas may be generated,
• Under high temperature and high humidity.
6
Notes on Safety
Maintenance and inspection
1. Check the unit operation on regular basis.
2. If this unit has not been used for more than three months, be sure to check that this unit operates normally and safely before use.
Troubleshooting
If device error 9 (Er9) occurs, take the following procedure promptly: (1) Remove the cuff from the patient's arm. (2) Turn off the power of the unit and unplug the AC adapter from the electric outlet. (3) Display "Out of use" on this unit so that it cannot be used. (4) Contact Omron repair service (1-877-216-1336).
7
Features of the Product
OMRON IntelliSense™ Blood Pressure unit, Model HEM-907 is developed to measure blood pressure and pulse rate accurately and simply in a doctor's office, examination room, or patient bedside.
• One-button operation
Simply wrap the cuff and push the START Button. Blood pressure and pulse rate are automatically measured by the oscillometric method.
• Automatic pressure setting
When the P-SET (Pressure Setting) Knob is set to "AUTO," the unit will automatically inflate the cuff to the optimal pressure according to each patient's blood pressure. Pre-setting inflation level is not necessary.
• Noiseless operation
This unit operates so quietly that it can be used in the patient room at night.
• Average Mode (AVG Mode)
In the AVG Mode, this unit will automatically measure for two or three times. The average of systolic and diastolic blood pressures and pulse rate are displayed. Each measurement can also be shown individually. The number of measurements, waiting time before first measurement, and the interval can be changed.
• Auscultation Mode (MANU Mode)
You can measure auscultatory blood pressure by using a stethoscope, with automatic cuff inflation and deflation by this unit. Because the cuff pressures during deflation are displayed digitally and synchronized with the heart beat, they can be read with accuracy. After taking systolic reading you can accelerate cuff deflation to shorten measurement time.
• Large and easy to read display
Large and easy to read figures are displayed on the LCD display.
INTELLISENSE™ blood pressure monitor is a global brand name of blood pressure monitor equipped with bio-information sensing and high performance fuzzy logic technology, which are the key technologies of OMRON. With this excellent sensing technology, the monitor can achieve “accurate” and “fast” measurement.
8
Components of the Product
Main unit
Accessories
Air tube
1.0m (39 1/3") *
Model: HEM-TUBE-100
AC adapter (80100H907S)
Model: HEM-AC-J
Instruction Manual
Items identified with an asterisk (*) are consumables and not covered by the guarantee.
IntelliSense™ Blood Pressure Monitor, Model HEM-907
(Included and also available separately)
Cuff / bladder set Large (with built-in bladder) * Model: HEM-907-CL19 Applicable arm circumference: 13
" to 17" (
32 to 42 cm)
Cuff / bladder set Small (with built-in bladder) * Model: HEM-907-CS19 Applicable arm circumference: 7” to 9
" (
17 to 22 cm)
Cuff / bladder set Medium (with built-in bladder) * Model: HEM-907-CR19 Applicable arm circumference: 9
" to 13"
(22 to 32 cm)
Specifications
9
Be Sure to Read This Section
Preparations before Measurement
How to Measure Blood Pressure
Know Your Unit
Components of the Product
Exclusive Stand Model: HEM-907-STAND
Wall-hanging kit Model: HEM-907-WKIT
Bladders *
Pole-mounting kit Model: HEM-907-PKIT
Cuffs without bladder *
Options
Large Cuff Model:HEM-907-CUFFL
Medium Cuff Model: HEM-907-CUFFM
Small Cuff Model: HEM-907-CUFFS
Large Bladder Model: HEM-907-BLDRL
Medium Bladder Model: HEM-907-BLDRM
Small Bladder Model: HEM-907-BLDRS
Battery pack (48H907N) * Model: HEM-907-PBAT
Air tube 1.3 m *(51 1/6") Model: HEM-TUBE-130
Items identified with an asterisk (*) are consumables and not covered by the guarantee.
Accessories and options are subject to change without prior notice.
(not included)
10
Features and Functions
Main unit
1.Display: Displays blood pressure and, pulse rate readings, and oscillation pulse level.
2.HIDE Button: Switches display and non-display of measured results.
3.DC jack: Connects the AC adapter.
4.P-SET (pressure setting) Knob: In the AUTO position, inflation level is automatically set.
Otherwise, inflation level can optionally be set manually between 100 and 280 mmHg.
5.MODE Selector: Selects the operation mode.
• One-time Measurement Mode (SINGLE Mode): Measurement with automatic inflation.
• Average Mode (AVG Mode): Automatically measures two (or three) times consecutively.
• Auscultation Mode (MANU Mode): Automatic inflation, automatic deflation, and pressure display for auscultation (does not measure blood pressure).
• Check Mode (CHECK Mode): Checks the accuracy of pressure display. Displays only pressure.
6.ON/OFF (power) Button: Turns on or off the unit.
7.START Button: Starts the measurement.
8. DEFLATION (deflation control) /Measurement Result Display Switch Button:
• In the MANU Mode, deflates the cuff by approximately 5 to 10 mmHg with each push during deflation.
• In the AVG Mode, switches the display of average values and the measurement results with each push.
9.Air Connector: Connects the air tube.
10.STOP Button: Stops the measurement and deflates air rapidly.
1. Display
2. HIDE (non-display) Button
3. DC jack
4. P-SET (pressure setting) Knob
5. MODE Selector
6. ON/OFF (power) Button
7. START Button
8. DEFLATION (deflation control) /Measurement Result Display Switch Button
9. Air Connector
10. STOP Button
Names of the Parts
(See Page 11 for detail)
Specifications
11
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
Features and Functions
Display
Systolic blood pressure
Diastolic blood pressure
Pulse rate
Contents when the AVG Mode is selected
AVG.: Mean value 1st : First reading 2nd : Second reading 3rd : Third reading
Ready to Measure / Pulse synchronization
Pulse level / Number of irregular pulses
Number of irregular pulse waves: The pulse wave having a difference of more than 25% from the average interval of pulse waves is determined irregular. Number of irregular pulse waves is displayed up to three times.
Battery level
Displays for the operable level.
Blinks for a low charged
level
Displays for the inoperable
level.
Charging
Displays when the battery pack is being charged.
External power source
Displays AC when the unit is connected to the external power source via the AC adapter.
Example of display
No. of irregular pulses
123
12
Features and Functions
Functions setting
(1) Inflation level setting
AUTO (Automatic setting): Can be used when the SINGLE, AVG, or MANU Mode is selected.
The unit estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure (approximately 30-40 mmHg above the patient’s systolic pressure).
Manual level setting: Inflation level can be set manually between 100 and 280 mmHg. Set the level to 30 to 40 mmHg higher than the expected systolic pressure.
• To set the P-SET to "AUTO," turn the P-SET Knob counterclockwise as far as it goes until you can hear a click.
• In the "AUTO" setting, inflation level may not be set automatically when the systolic blood pressure is more than 220 mmHg. Use the unit in the manual setting.
• If the cuff has not been inflated to the necessary level, it may be re-inflated automatically.
(2) Non-display function
Use to prohibit the display of measurement results. However, the cuff pressure during measurement is displayed. This function can be used in the SINGLE and AVG Modes.
By pushing the Button, display or non-display of status is switched alternatingly.
(3) Manual deflation control
Accelerate deflation by pushing the DEFLATION (deflation control) / Measurement Result Display Switch Button during deflation in the measurement by the MANU Mode.
With each push of the Button, cuff is deflated rapidly in increments of 5 to 10 mmHg.
Display Non-display
Specifications
13
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
Features and Functions
List of Functions for each Mode
Inflation level
setting
Non-display
function
Manual deflation
control
Single (SINGLE)
Mode
Average
(AVG) Mode
Auscultation
(MANU) Mode
Check
(CHECK) Mode
Function
Measurement
Mode
(4) AVG Function setting
You can set the number of measurements, the waiting time until the 1st measurement, and the measurement interval for the AVG Mode.
Items to set Set value
F1 Number of measurements 2 times or 3 times F2
Waiting time until the start of 1st measurement
0 sec, 3 min, 5 min, or 10 min.
F3 Measurement interval 5 sec, 30 sec, 1 min, or 2 min.
Note: The bold letters represent the factory-set values.
Procedure to change the set values
1) When the power is OFF, push the ON/OFF (power) Button for more than three seconds while holding the START Button; F1 is displayed.
2) Push the START Button and select the function to set from F1 to F3. Each time you push the START Button, the functions change in the order of .
3) Push the DEFLATION (deflation control)/Measurement Result Display Switch Button and change the set values.
4) When the setting is finished, push the ON/OFF (power) Button to turn off the power. The setting is changed.
min sec
Function #
14
How to Apply the Cuff
General advice
• Do not inflate the cuff without being wrapped over the arm.
• Do not use a damaged cuff.
1. Select the cuff according to the arm size
Arm circumference Name of the cuff
(7” - 9”) 17-22 cm HEM-907-CS19(Small)
(9” - 13”) 22-32 cm HEM-907-CR19 (Medium)
(13” - 17”) 32-42 cm HEM-907-CL19(Large)
• Check the following before applying the cuff:
1) The bladder is correctly installed in the cuff.
2) The bladder is not twisted inside the cuff.
3) The bladder tube is protruding from the cuff as shown in the Figure on the right.
2. Connect the air tube securely.
• Connect the air tube to the main unit by securing the air plug to the base of the air connector.
• Securely connect the air tube and the cuff/bladder set by rotating the luer connection as shown in the Figure on the right.
Do not wrap the cuff over on arm to which intravenous injection or transfusion is being conducted, or when otherwise contraindicated.
Do not connect the air tube or the cuff to other equipment which is connected to intracorporeal organ. Air embolisms may result.
The cuff of OMRON IntelliSense™ Blood Pressure Monitor HEM-907 plays an important role of collecting the information on blood vessels. Please wrap the cuff according to the procedure, below.
Cuff
Air tube
Air connector
Air plug
Cuff side Air tube side
Extract the air tube from .
Enlarged figure of
the circled part
Specifications
15
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
How to Apply the Cuff
4. Align the Artery Position Mark with the brachial artery.
5. Wrap the cuff snugly using both hands and securely fasten it with the Velcro™ tape. At this time, the lower edge of the cuff must be placed 1/2" to 1" above the inner side of elbow joint.
• If the INDEX is positioned outside the RANGE, select the cuff suitable for the patient's arm circumference and wrap it again.
• Wrap the cuff so that you can insert only one finger between the cuff and arm.
7. Keep the level of the cuff at the same level as the heart during the measurement.
3. Place the hand of the patient with the palm of hand facing upward.
1/2"-1"
16
How to Use the Power Source
General advice
• Read the instruction manual of the other devices to be used at the same time with this unit to understand and be aware of the interaction between the devices.
Do not use this unit in the presence of flammable gas, or anesthetics, or in a high pressure oxygen room or oxygen tent.
Do not touch the AC adapter with wet hands.
Be sure to use the AC adapter from the power supply of 110 VAC.
Do not install or store this unit where it may come in contact with water or liquid medication.
[1]
[2]
How to use the AC adapter
Connect the AC adapter to the DC jack of the main unit [1] and the electric outlet [2].
NOTE: When the AC adapter is connected and the unit is turned off, the AC adapter
charges the installed rechargeable battery.
Specifications
17
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
How to Use the Power Source
1. Remove both screws on the upper portion of the battery cover of this unit, and remove the cover.
2. Disconnect the old battery pack from the connector and replace with a new one.
3. Install the battery cover and fasten it with both screws. Be careful not to pinch the lead wire.
4. Connect the main unit to the AC adapter to charge the new battery. The battery is not charged when you purchase the monitor. When you use the battery for the first time, charge it for more than twelve hours before use.
If battery fluid gets into your eye or comes in contact with skin, wash the affected area with water repeatedly. Immediately consult a doctor for treatment.
Do not use the battery pack for devices other than for this unit.
Do not disassemble the battery pack.
Installation and Replacement of Battery Pack (option)
18
Battery life
• You can use the unit for approximately three hundred measurements with one charge.
• Approximate life of battery is two years. However the battery life from each charging may be shortened depending on the state of using. If the interval between charging becomes short and the mark appears frequently, replace it.
Charging time
• At approximately five seconds after connecting the AC adapter, the unit will start battery charging automatically.
• While the battery is being charged, the mark turns on.
• The battery can be completely charged in approximately twelve hours.
Battery low
• When the mark starts to blink, twenty to thirty measurements remain on the battery.
However if the Battery Low Mark starts to blink, charge it soon.
• If the mark is displayed, the battery is low and the unit cannot operate. Please charge the battery.
Automatic Power Off
• When using the unit with the battery, the unit will turn off automatically after five minutes
of inactivity.
• While the AC adapter is connected, the Auto Power Off function does not work.
How to Use the Power Source
Specifications
19
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
How to Measure Blood Pressure
General advice
• Patients should not touch the unit.
If cuff inflation does not stop, remove the cuff or pull out the air tube from the main unit.
List of Measurement Modes
To measure
only once
SINGLE Mode
Refer to Page
19.
To measure two
times
(or three times)
automatically and
consecutively
AVG Mode
Refer to Page
21
.
To measure by
using a
stethoscope
MANU Mode
Refer to Page
23.
Confirm readings with the stethoscope when an irregular pulse wave is displayed or when the measured value is questionable or erratic.
Do not use a cellular phone near this unit. Do not use this unit in a vehicle.
20
How to Measure Blood Pressure (in SINGLE Mode)
1. Push the ON/OFF (power) Button to turn on the power.
2. Set the MODE Selector to "SINGLE".
3. Set the P-SET (inflation level) Knob to "AUTO" or to the target pressure value.
4. Measure the patient’s arm size, and wrap appropriate cuff over the patient's arm. (Refer to Pages 14 and 15.)
Specifications
21
Be Sure to Read This Section
Know Your Unit
How to Measure Blood Pressure
How to Measure Blood Pressure (in SINGLE Mode)
5. Push the START Button to start the measurement.
• Do not push the START Button without wrapping the cuff.
• If you want to stop measurement, push the STOP Button. The cuff will rapidly deflate.
6. The measurement results are displayed.
• While the battery pack is in use, the monitor will turn off automatically after five minutes of inactivity and the display (measurement results) will disappear. (Automatic Power Off)
7. Push the ON/OFF (power) Button to turn off the power.
If the monitor determines that the pressure value is not correct, an error display appears (Er1 to 9). In this case, refer to Page 28 and start the measurement again.
Preparations before Measurement
22
How to Measure Blood Pressure (in AVG Mode)
1. Push the ON/OFF (power) Button to turn on the power.
2. Set the MODE Selector to "AVG".
The factory-set values are set as follows:
• Number of measurements: 2
• Waiting time until the 1st measurement: 0 sec.
• Interval: 1 min.
To change these factory-set values, refer to Page 12.
3. Set the P-SET (inflation level setting) Knob to "AUTO" or the target pressure value.
4. Measure the patient’s arm size and wrap appropriate cuff over the patient's arm. (Refer to Pages 14 and
15.)
Specifications
23
Be Sure to Read This Section
Know Your Unit
How to Measure Blood Pressure
How to Measure Blood Pressure (in AVG Mode)
Preparations before Measurement
5. Push the START Button to start the measurement.
After the pre-select waiting time, the unit takes the 1st measurement.
After displaying the results of 1st measurement, subsequent measures occur automatically at the specified intervals.
6. The measurement results are displayed.
After all the measurements are finished, average values will be displayed.
Each time the DEFLATION (deflation control) / Measurement Result Display Switch Button is pushed, the measurement results for each reading and the average value will be displayed.
7. Push the ON/OFF (power) Button to turn off the power.
If the monitor determines that the pressure value is not correct, an error display appears (Er1 to 9). In this case, refer to Page 28 and start the measurement again.
• For setting the number of measurements, the waiting time until the 1st measurement, and the interval, refer to Page 14.
• If you want to stop measurement, push the STOP Button. The unit will rapidly deflate.
• If an error occurs during measurement, the monitor will automatically start measurement again. If a second error occurs, measurement will automatically stop.
• Do not push the START Button without wrapping the cuff.
• While the battery is in use, the monitor will turn off after five minutes of inactivity and the display (measurement results) will disappear. (Automatic Power Off)
24
How to Measure Blood Pressure (in MANU Mode)
1. Push the ON/OFF (power) Button to turn on the power.
3. Set the P-SET (inflation level setting) Knob to "AUTO" or the target pressure value.
4. Measure the patient’s arm size and wrap appropriate cuff over the patient's arm. (Refer to pages 14 and 15.)
2. Set the MODE Selector to "MANU".
Specifications
25
Be Sure to Read This Section
Know Your Unit
How to Measure Blood Pressure
How to Measure Blood Pressure (in MANU Mode)
Preparations before Measurement
5. Place the stethoscope on the patient's arm.
6. Push the START Button to start the measurement.
• Do not push the START Button without wrapping the cuff.
• Do not squeeze or press the cuff during the measurement.
• If you want to inflate again after the start of
deflation, push the START Button.
• If you want to accelerate deflation after the
start of deflation, push the DEFLATION (deflation control) / Measurement Results Display Switch Button. Each time the Button is pushed, cuff is deflated rapidly in increments of 5 to 10 mmHg.
9. Push the ON/OFF (power) Button to turn off the power.
7. Take the readings.
8. Push the STOP Button to remove air inside the cuff.
• The unit does not automatically deflate in the MANU Mode.
If the monitor determines that the pressure value is not correct, an error display appears (Er1 to 9). In this case, refer to Page 28 and start the measurement again.
26
How to Check Pressure Accuracy (CHECK Mode)
Accuracy of pressure display can be checked in the CHECK Mode.
What you need to prepare
(1) Calibrated mercury manometer (including inflation bulb), (2) T-tube, (3) two air tubes, and (4) a sturdy cylindrical shaped object on which the cuff is wrapped.
Check result
Accuracy of the monitor is controlled by ±3 mmHg or 2% of standard manometer reading. If your result shows a difference exceeding the tolerance, contact Omron repair department (1-877-216-1336).
How to check
1. Connect the manometer, inflation bulb, cuff, and the monitor with the T-tube as shown in the figure on the right.
2. Tightly wrap the cuff over a sturdy cylinder.
3. Release the valve of inflation bulb to remove the air inside the cuff completely.
4. Push the ON/OFF (power) Button to turn on the monitor.
5. Set the MODE Selector to “CHECK”.
6. Close the valve of inflation bulb and inflate the cuff to the pressure to be checked, based on the manometer read.
7. Compare the pressure values displayed on the monitor to the one on the manometer.
Specifications
27
Be Sure to Read This Section
Know Your Unit
How to Measure Blood Pressure
Preparations before Measurement
How to Clean the Unit after Use
1. Wipe the monitor with a soft, damp cloth diluted with disinfectant alcohol, or diluted detergent.
2. Complete cleaning by wiping the monitor with a soft, dry cloth.
General advice
• Do not clean this unit with gasoline, paint thinner, or high concentration alcohol.
• Do not disinfect this unit by autoclave or gas sterilization (EtO, gluteraldehyde, or high concentration ozone.)
When cleaning this unit, unplug the AC adapter from the electric outlet.
After cleaning this unit, dry it well, before plugging the AC adapter in the electric outlet.
28
List of Error Codes
Error code Explanation
Inflation error
• When the pressure does not exceed 12 mmHg within the set time after the start of inflation
• When the inflation does not reach the set cuff pressure within the specified time after the start of inflation
Deflation error
• When the deflation speed is too fast during the measurement
• When the deflation speed is too slow during the measurement
• When the measurement does not finish within the specified time after starting the measurement
Overpressure error
• The cuff pressure exceeded 299 mmHg.
Insufficient inflation error
• Blood pressure could not be measured due to insufficient inflation level.
Indeterminable blood pressure error
• Blood pressure could not be measured even when the cuff pressure reached the specified pressure.
Low pulse level error
• Pulse wave was too small.
Blood pressure error
• Relationship between systolic and diastolic pressures was abnormal.
Pulse rate error
• Pulse rate did not stay within the range of 30 to 199 beats/min.
Device error
• This is a trouble of the main unit of the monitor.
How to correct
• Confirm that the air tube connecting the cuff and the main unit is connected securely.
• Confirm that the air flow in the air tube connecting the cuff and the main unit isn’t being restricted.
• Confirm that the cuff is wrapped correctly (refer to pages 14 and 15).
• Check bladder for leaks and, if necessary, replace the bladder with new one (option).
• Confirm that air flow in the air tube connecting the cuff and the main unit isn’t being restricted.
• If the measurement is made by setting the P­SET to "AUTO", ask the patient not to move during the inflation.
• Confirm that the P-SET is securely set to "AUTO". Turn the Knob counterclockwise as far as it goes until you can hear a click sound.
• If the measurement is made by manual inflation level setting, set the value to 30 to 40 mmHg higher.
• Confirm that the cuff is wrapped correctly (refer
to pages 14 and 15).
• Confirm that the cuff is wrapped correctly (refer
to pages 14 and 15).
• Ask the patient not to move during the measurement.
• Check the patient for arrhythmia.
• Contact Omron’s repair department toll-free at (1-877-216-1336).
Specifications
29
Be Sure to Read This Section
Know Your Unit
How to Measure Blood Pressure
Preparations before Measurement
Troubleshooting
Trouble
The unit inflates to abnormally high (low) pressure.
The monitor cannot measure blood pressure.
Measured values are abnormally high (low).
What to inspect
Is the cuff wrapped correctly?
Is the patient moving during inflation?
Does this patient have arrhythmia?
Check the patient's condition.
Is the patient moving during the measurement?
Does the patient have an arrhythmia?
Is the size of the cuff correct and is it wrapped correctly?
Is the level of the brachium to which the cuff is wrapped at the same level as the heart?
How to correct
Wrap the cuff correctly, and measure again.
(Refer to Page 14 and 15.)
Ask the patient not to move during measurement, and measure again.
Set the P-SET to 30 to 40 mmHg higher than estimated systolic pressure of the patient, then measure.
After checking the patient with the stethoscope refer to the "list of error codes".
(Refer to Page 27.)
Ask the patient not to move during measurement, and measure again.
Set the P-SET to 30 to 40 mmHg higher than estimated systolic pressure of the patient, then measure.
Select the cuff according to the patient's arm circumference, wrap it correctly, then measure again.
(Refer to Pages 14 and 15.)
Keep the level of the brachium to which the cuff is wrapped at the same level as the heart, then measure again.
If the unit malfunctions during use, please check the following:
Are the patient’s clothes restricting normal blood flow to the arm?
Remove the clothing and measure again.
30
Caution
CAUTION:
Changes or modifications not expressly approved by Omron Healthcare, Inc. could void the user’s authority to operate this product.
NOTE:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only) This product has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. The product generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If the product does cause harmful interference to radio or television reception, which can be determined by turning the product on and off, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and receiver.
• Connect the product into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only) This digital apparatus does not exceed the Class B limits for radio noise emissions from digital apparatus as set out in the interference-causing equipment standard entitled "Digital Apparatus", ICES-003 of the Canadian Department of Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux appareils numériques de Classe B prescrites dans la norme sur le matériel brouilleur: "Appareils Numériques", NMB-003 èdictée par le ministre des communications.
Five Year Limited Warranty
31
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
Specifications
Your HEM-907 IntelliSense™ Automatic Blood Pressure Monitor is warranted to be free from manufacturing defects for a period of five years under normal use. The five year warranty excludes the monitor cuff. The cuff is warranted for a one year period. This warranty extends only to the original retail purchaser.
Should repair be needed within the warranty period, ship the unit prepaid to Omron Healthcare, Inc., 300 Lakeview Parkway, Vernon Hills, IL 60061, Attn: Service Dept., together with $5.00 for return shipping and insurance. Be sure to include the model number of your unit and your phone number on any correspondence.
We will either repair or replace (at our option) free of charge any parts necessary to correct defects in the materials or workmanship.
The above warranty is complete and exclusive. The warrantor expressly disclaims liability for incidental, special, or consequential damages of any nature. (Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above warranty may not apply to you.)
Any implied warranties arising by the operation of law shall be limited in duration to the term of this warranty. (Some states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you.)
This warranty gives you specific legal rights, and you may have other rights which vary from state to state. As a condition to operation of your warranty, the enclosed registration card must be completed and sent to us within 10 days from the date of purchase.
FOR CUSTOMER SERVICE CALL TOLL FREE:
1-877-216-1336
Specifications
Name: OMRON Digital Automatic Blood Pressure Monitor Model: HEM-907 Display: Digital display Measurement: Oscillometric method Measurement Range: Pressure; 0 to 299 mmHg
Pulse rate; 30 to 199 beats/min
Accuracy: Pressure; Within ±3 mmHg
Pulse rate; Within ±5% of reading Inflation: Automatic inflation with pumping Deflation: Automatic deflation by electromagnetic control valve Air Release: Automatic rapid air release by electromagnetic control valve Pressure Detection: Electrostatic capacity semi-conductor pressure sensor Power supply: AC adapter (110 VAC, 50/60 Hz, 20 VA) or
Battery pack (4.8 VDC, 6W) Electric Shock Protection Method
: Class II B type
Operating Temperature and Humidity
: 50°F to 104°F (10 to 40°C), 30 to 85% RH Weight of Main Unit: Approx.32 oz (910 g) External Dimensions: Approx.5 1/2" (W)x 8"(H) x5 1/6"(D)
139 (W) x 203 (H) x 131 (D) mm
Accessories: Cuff / bladder set Large, Cuff / bladder set Medium, Cuff / bladder set Small,
AC adapter, air tube (1.0 m), Instruction Manual (with guarantee card)
Options: Cuff Large (with no bladder), cuff Medium (with no bladder), cuff Small (with no
bladder), Bladder Large, bladder Medium, bladder Small, air tube 1.3 m (51 1/6"), Battery pack Stand exclusive for this unit, wall-hanging kit, pole-mounting kit
* Please note that specifications may be changed without prior notice.
For Customer Service
Call Toll Free
1-877-216-1336
Manufactured by OMRON Distributed by:
OMRON HEALTHCARE, INC.
300 Lakeview Parkway Vernon Hills, Illinois 60061 www.omronhealthcare.com
Copyright © 2000 OMRON HEALTHCARE, INC. HEM-907INST