Omron compA-I-R NE-C28P Instruction Manual

TM
Compressor Nebulizer
Model
NE-C28P
• Instruction Manual
• Mode d’emploi
• Gebrauchsanweisung
• Manuale di istruzioni
• Gebruiksaanwijzing
РУКОВОДСТВО ПО ЭКСПЛУАТАЦИИ
EN
FR
DE
IT
ES
NL
RU
AR
IM-NE-C28P-E-01-09/2011
5334069-0A
Before using the unit
Contents
Before using the unit
Introduction .................................................................................................. 3
Intended use ................................................................................................. 3
Important safety instructions ...................................................................... 4
1. Know your unit ....................................................................................... 7
2. Guide to using the COMP AIR .............................................................. 9
Operating instructions
3. Preparing the nebulizer for use .......................................................... 10
4. Taking medication ............................................................................... 14
Care and maintenance
5. Cleaning and daily disinfecting .......................................................... 17
5.1 Cleaning ............................................................................................ 17
5.2 Disinfecting the nebulizing parts ....................................................... 18
6. Maintenance and storage .................................................................... 21
7. Changing the air filter .......................................................................... 22
8. Troubleshooting ................................................................................... 23
9. Technical data ...................................................................................... 24
10. Accessories .......................................................................................... 27
2
Introduction
Intended use
Thank you for purchasing the OMRON COMP AIR. This product was developed in conjunction with respiratory therapists for the
successful treatment of asthma, chronic bronchitis, allergies and other respiratory disorders. The compressor forces air to the nebulizer kit. When the air enters the nebulizer kit, it converts the prescribed medication into an aerosol of microscopic droplets that can easily be inhaled.
This is a medical device. Operate the device only as instructed by your doctor and/ or respiratory therapist.
Medical Purpose This product is intended to be used for inhaling medication
for respiratory disorders.
Intended User • Legally certified medical experts, such as doctor, nurse and
therapist, or healthcare personnel or patient under the guidance of qualified medical experts.
• The user should also be capable of understanding general operation of NE-C28P and the content of instruction manual.
Intended Patients This product should not be used by patients, who are
unconscious or are not breathing spontaneously.
Environment This product is intended for use in a medical facility, such as
Hospital, clinic and doctor’s office, and in a room of general household.
Durable period Durable periods are as follows, provided the product is used
to nebulize saline 3 times a day for 10 minutes each time at room temperature (23°C).
Durable period may vary depending on usage environment. Compressor 5 years
Nebulizer Kit 1 year Mouthpiece 1 year Nosepiece 1 year Air Tube (PVC, 200cm) 1 year Air Filter 60 days Adult Mask (PVC) 1 year Child Mask (PVC) 1 year
EN
Precautions for use Warnings and cautions described in the instruction manual
should be observed.
3
Important safety instructions
Read all the information in the instruction manual and any other literature included in the box before using the device.
When using an electrical product, especially when children are present, basic safety precautions should always be followed including the following:
Warning:
Indicates a potentially hazardous situation which, if not avoided, could result in serious injury.
(Usage)
• For type, dose, and regime of medication follow the instructions of your doctor or respiratory therapist.
• If you feel anything unusual during use, stop using the device immediately and consult your doctor.
• Do not use only water in the nebulizer, for inhaling purposes.
• Clean and disinfect the nebulizer kit, mouthpiece and nosepiece or masks before using them for the first time after purchase, if the device has not been used for a long period of time, or if more than one person uses the same device.
• Be sure to wash the parts after use, and ensure that they are thoroughly disinfected and dried, and stored in a clean location.
• Keep the device out of the reach of unsupervised infants and children. The device may contain small pieces that can be swallowed.
• Store the device and accessories in a clean location.
• Do not store the air tube while there is moisture or medication remaining inside it.
• Always rinse the nebulizing parts with clean hot tap water after disinfecting them. Never leave the cleaning solution in the medication tank or mouthpiece.
• Do not use or store the device where it may be exposed to noxious fumes or volatile substances.
• Do not use the device where it may be exposed to flammable gas.
• Do not block the ventilation slots. Never place the device where the ventilation slots may be obstructed during operation. Do not cover the compressor with a blanket or towel etc., during use.
• Make sure that the nebulizer kit is clean before use.
• Always dispose of any remaining medication after use, and use fresh medication each time.
• Do not use in anaesthetic or ventilator breathing circuits.
(Risk of electrical shock)
• Do not use the compressor or power cord while they are wet, or connect them to a power outlet or other devices with wet hands.
• The compressor and power cord are not waterproof. Do not spill water, or other liquids, on these parts. If liquid does spill on these parts, immediately unplug the power cord and wipe off the liquid with gauze or other absorbent material.
4
Important safety instructions
• Do not immerse the compressor in water or other liquid.
• Do not use or store the device in humid locations, such as a bathroom.
• Do not operate the device with a damaged power cord or plug.
• Keep the power cord away from heated surfaces.
Read and follow the “Important information regarding Electro Magnetic Compatibility (EMC)” (page 26).
Caution:
Indicates a potentially hazardous situation which if not avoided, may result in minor or moderate injury, or physical damage.
(Usage)
• Provide close supervision when this device is used by, on, or near children or invalids.
• Make sure that the parts are attached correctly.
• Make sure that the vaporiser head is correctly installed before use.
• Make sure that the air filter is correctly attached.
• Make sure that the air filter is clean. If the air filter has changed colour, or has been used on average for more than 60 days, replace it with a new one.
• Do not spill liquid or medication on the compressor.
• Do not tilt the nebulizer kit at an angle of greater than 45° in all directions, or shake it while in use.
• Do not use or store the device while the air tube is creased.
• Use only original nebulizing parts, air tube, air filter and filter cover.
• Do not add more than 7 ml of medication to the medication tank.
• Do not carry or leave the nebulizer kit while the medication tank contains medication.
• Do not leave the device unattended with infants or persons who cannot express their consent.
• Do not subject the device or the parts to any strong shocks such as dropping the device on the floor.
• Do not distort the vaporiser head, or jab the nozzle of the medication tank with a pin etc.
• Do not insert fingers or objects inside the compressor.
• Do not disassemble or attempt to repair the compressor or power cord.
EN
• Do not leave the device or its parts in extreme hot or cold temperature, or under direct sunlight.
• Do not block the air filter cover.
• When disinfecting parts by boiling, make sure that the container does not boil dry.
• Never place, or attempt to dry the compressor, the power cord or nebulizer kit in a microwave oven.
• Do not use the device while sleeping or if drowsy.
5
Important safety instructions
• To avoid injury to the nose mucosa, do not squeeze the nosepiece into the back of the nose.
• Do not block the slit between the cap and the inhalation air inlet.
• Do not use a microwave oven or hair dryer to dry the nebulizing parts.
• Approved for human use only.
• When using the device, the main unit may become hot.
• Do not touch the main unit for other than necessary operation such as turning off the power while nebulizing.
• To avoid the medication residue on the face, be sure to wipe the face after removing the Mask.
• Do not use an autoclave, EOG gas disinfection or low temperature plasma sterilizer to disinfect the device.
(Risk of electrical shock)
• Always unplug the device from the power outlet after use and before cleaning.
• Plug the device into the appropriate voltage outlet. Do not overload power outlets or use extension cords.
• Do not wind the power cord around the compressor.
• Changes or modifications not approved by OMRON HEALTHCARE will void the user warranty.
• Do not pull the power cord strongly.
General Safety Precautions:
• Inspect the device and parts before using them each time, and check that there are no problems. In particular, be sure to check the following:
- That the vaporiser head, nozzle or air tube are not damaged.
- That the nozzle is not blocked.
- That the compressor operates normally.
• When using this device, there will be some noise and vibration caused by the pump in the compressor. There will also be some noise caused by the emission of compressed air from the nebulizer kit. This is normal and does not indicate a malfunction.
• Operate the device only as intended. Do not use the device for any other purpose.
• Do not use the device at temperatures greater than +40°C.
• Make sure that the air tube is securely attached to the compressor (main unit) and nebulizer kit, and does not come loose. Twist the air tube plug slightly when inserting it into the connectors to avoid the air tube disconnecting during use.
• To completely isolate the device from the power source unplug the plug from the power source.
Read and follow the “Correct Disposal of This Product” (page 26) when disposing of the device and any used accessories or optional parts.
Save these instructions for future reference.
6
1. Know your unit
A7
A6
D
E
A5
D1 D4
E1
E3
E2
E6
E4
E5
E7
A C
A1
A2
A3
B1
A4
A5
B2
B
D2 D3
Rear View
A. Compressor (Main Unit)
A1.Power Switch A2.Air Tube Connector A3.Air Filter Cover A4.Nebulizer Kit Holder A5.Ventilation Slots A6.Power Cord A7.Power Plug
B. Air Tube (PVC, 200cm)
B1.Air Tube Plug B2.Air Tube Plug
C. Nebulizing Parts
(* Air filter inside.)
D. Inhalation Attachments
D1.Mouthpiece D2.Adult Mask (PVC) D3.Child Mask (PVC) D4.Nosepiece
E. Nebulizer Kit
E1.Cap E2.Inhalation Air Inlet E3.Inhalation Top E4.Vaporiser Head E5.Medication Tank E6.Nozzle E7.Air Tube Connector
EN
7
1. Know your unit
• Instruction Manual
• Mode d’emploi
• Gebrauchsanweisung
• Manuale di istruzioni
• Manual de instrucciones
• Gebruiksaanwijzing
TM
EN
FR
DE
IT
ES
IM-NE-C28P-E-01-09/2011
5334069-0A
NL
AR
RU
Compressor Nebulizer
Model
NE-C28P
F
G H I
Accessories
F. Spare Air Filters x 5 G. Carrying Bag
H. Instruction Manual I. Warranty Card
8
2. Guide to using the COMP AIR
Preparing the Nebulizer for Use - Section 3 (page 10-13)
Warning:
Clean and disinfect the nebulizer kit, mouthpiece and nosepiece or masks before using them for the first time after purchase, if the device has not been used for a long period of time, or if more than one person uses the same device.
Taking Medication - Section 4 (page 14-16)
Caution:
• Do not tilt the nebulizer kit at an angle of greater than 45
° in all directions.
• Do not crease or fold the air tube.
Cleaning Parts After Use - Section 5 (page 17-20)
Note: Failure to clean the nebulizing parts
after use may lead to infection.
How the nebulizer kit works
The medication that is pumped up through the medication channel is mixed with compressed air which is generated by a compressor pump. The compressed air mixed with medication is tuned into fine particles and sprayed by being impacted to the baffle.
Baffle
Aerosol
Medication Channel
Medication
Compressed Air
Aerosol
Medication
Nozzle
EN
9