Introduction
Thank you for purchasing the OMRON COMP AIR.
This product was developed in conjunction with respiratory therapists for the successful
treatment of asthma, chronic bronchitis, allergies and other respiratory disorders. The
compressor forces air to the nebulizer kit. When the air enters the nebulizer kit, it converts
the prescribed medication into an aerosol of microscopic droplets that can easily be inhaled.
This is a medical device. Operate the device only as instructed by your doctor and/or
respiratory therapist.
NE-C801KD includes the kids accessories to enable children to feel more comfortable while
inhaling. It improves compliance with drug treatment.
How the nebulizer kit works
The medication that is pumped up through the medication
Aerosol
Baffle
Aerosol
channel is mixed with compressed air which is generated
by a compressor pump. The compressed air mixed with
medication is turned into fine particles and sprayed by
being impacted to the baffle.
Nozzle
Medication
Compressed Air
Medication
Channel
Medication
4
Intended use
Medical PurposeThis product is intended to be used for inhaling medication for
respiratory disorders.
Intended User• Legally certified medical experts, such as doctor, nurse and
therapist.
• Caregiver or patient under the guidance of qualified medical
experts for home treatment.
• The user should also be capable of understanding general
operation of NE-C801KD and the content of instruction manual.
Intended PatientsThis product should not be used on patients, who are unconscious
or are not breathing spontaneously.
EnvironmentThis product is intended for use in a medical facility, such as
Hospital, clinic and doctor’s office, and in a room of general
household.
Durable periodDurable periods are as follows, provided the product is used to
nebulize 2 ml of medication 2 times a day for 6 minutes each time
at room temperature (23°C).
Durable period may vary depending on usage environment.
Compressor (Main unit)5 years
Nebulizer Kit1 year
Kids Accessory 1 year
Mouthpiece1 year
Air Tube (PVC, 100cm)/(Silicone, 100cm) 1 year
Air Filter60 days
Mask (PVC)/(SEBS)1 year
Precautions for useWarnings and cautions described in the instruction manual should
be observed.
EN
5
Important safety instructions
Read all the information in the instruction manual and any other literature included in the
box before using the device.
When using an electrical product, especially when children are present, basic safety
precautions should always be followed including the following:
Warning:
Indicates a potentially hazardous situation which, if not avoided, could result in serious injury.
(Usage)
• For type, dose, and regime of medication follow the instructions of your doctor or
respiratory therapist.
• If you feel anything unusual during use, stop using the device immediately and consult
your doctor.
• Do not use only water in the nebulizer, for inhaling purposes.
• Keep the device and kids accessory out of the reach of unsupervised infants and children.
The device may contain small pieces that can be swallowed.
• Do not let infants and children play with a kids accessory. Kids accessory is not a toy, but a
component of nebulizer kit.
• Do not store the air tube while there is moisture or medication remaining inside it.
• Do not use or store the device where it may be exposed to noxious fumes or volatile
substances.
• Do not use the device where it may be exposed to flammable gas.
• Do not cover the compressor with a blanket, towel, or any other type of cover during use.
• Make sure that the nebulizer kit and kids accessory are clean before use.
• Always dispose of any remaining medication after use, and use fresh medication each time.
• Do not use in anaesthetic or ventilator breathing circuits.
• Store the parts in a clean location for avoiding infection.
(Risk of electrical shock)
• Do not use the compressor or AC adapter while they are wet, or connect them to a power
outlet or other devices with wet hands.
• The compressor and AC adapter are not waterproof. Do not spill water, or other liquids, on
these parts. If liquid does spill on these parts, immediately unplug the power cord and wipe
off the liquid with gauze or other absorbent material.
• Do not immerse the compressor in water or other liquid.
• Do not use or store the device in humid locations, such as a bathroom.
• Use only original OMRON AC adapter. Use of unsupported AC adapter may damage the
device.
• Do not operate the device with a damaged power cord or plug.
• Keep the power cord away from heated surfaces.
• Read and follow the “Important information regarding Electro Magnetic Compatibility
(EMC)” (page 23) in the Technical Data Section.
6
(Cleaning and disinfecting)
Observe the rules below when cleaning or disinfecting parts. Failure to observe these rules
may result in damage, inefficient nebulization or infection. For the instructions, refer to
“4. Cleaning and disinfecting” section.
• Clean and disinfect the nebulizer kit, kids accessory, mask, mouthpiece or nosepiece
before using them when as follows:
- the first time after purchase.
- if the device has not been used for a long period of time
- if more than one person uses the same device
• Be sure to wash or wipe the parts after use, and ensure that they are thoroughly
disinfected and dried, and stored in a clean location.
• Do not leave the cleaning solution in the parts.
Caution:
Indicates a potentially hazardous situation which if not avoided, may result in minor or
moderate injury, or physical damage.
(Usage)
• Provide close supervision when this device is used by, on, or near children or invalids.
• Make sure that the parts are attached correctly.
• Make sure that the vaporizer head is correctly installed before use.
• Make sure that the air filter is correctly attached.
• Make sure that the air filter is clean. If the air filter has been used for more than 60 days,
replace it with a new one.
• Do not spill liquid or medication on the compressor.
• Do not tilt the nebulizer kit so that it is at an angle of greater than 45° in all directions, or
shake it while in use.
• Do not use or store the device while the air tube is creased.
• Use only original nebulizing parts, air tube, air filter and air filter cover.
• Do not add more than 7 ml of medication to the medication tank.
• Do not carry or leave the nebulizer kit while the medication tank contains medication.
• Do not leave the device unattended with infants or persons who cannot express their
consent.
• Do not subject the device or the parts to any strong shocks such as dropping the device on
the floor.
• Do not distort the vaporizer head, or jab the nozzle of the medication tank with a pin or any
sharp object.
• Do not insert fingers or objects inside the compressor.
• Do not disassemble or attempt to repair the compressor or power cord.
• Do not leave the device or its parts where it will be exposed to extreme temperatures or
changes in humidity, such as in a vehicle during the summer, or where it will be exposed to
direct sunlight.
• Do not block the air filter cover.
• Do not use the device while sleeping or if drowsy.
• Do not block the slit between the cap and the inhalation air inlet.
• Approved for human use only.
• If the device is used continuously, the service life of the device may be shortened.
• Limit use to 20 minutes at a time, and allow a 40 minute interval before using the device
again.
• When using the device, the main unit may become hot.
EN
7
• Do not touch the main unit for other than necessary operation such as turning off the
power while nebulizing.
• To avoid the medication residue on the face, be sure to wipe the face after removing the
Mask.
(Risk of electrical shock)
• Always unplug the AC adapter from the power outlet after use and before cleaning.
• Plug the device into the appropriate voltage outlet. Do not overload power outlets or use
extension cords.
• Always remove the AC adapter from the device after use and before cleaning.
• Do not misuse the cord of the AC adapter.
• Do not wind the cord of the AC adapter around the compressor or AC adapter.
• Changes or modifications not approved by OMRON HEALTHCARE will void the user
warranty.
• Do not pull the power cord of AC adapter strongly.
(Cleaning and disinfecting)
Observe the rules below when cleaning or disinfecting parts. Failure to observe these rules
may result in damage, inefficient nebulization or infection. For the instructions, refer to
“4. Cleaning and disinfecting” section.
• Do not use a microwave oven, dish dryer or hair dryer to dry the device or the parts.
• Do not use an autoclave, EOG gas sterilizer or low temperature plasma sterilizer.
• When disinfecting parts by boiling, make sure that the container does not boil dry.
Otherwise it may also result in fire.
General Safety Precautions:
• Inspect the device and parts before using them each time, and check that there are no
problems. In particular, be sure to check the following:
- That the vaporizer head, nozzle, kids accessory or air tube are not damaged.
- That the nozzle is not blocked.
- That the compressor operates normally.
• When using this device, there will be some noise and vibration caused by the pump in the
compressor. There will also be some noise caused by the emission of compressed air from
the nebulizer kit. This is normal and does not indicate a malfunction.
• Operate the device only as intended. Do not use the device for any other purpose.
• Do not use the device at temperatures greater than +40°C.
• Make sure that the air tube is securely attached to the compressor and nebulizer kit, and
does not come loose. Twist the air tube plug slightly when inserting it into the air tube
connectors to avoid the air tube disconnecting during use.
• The AC adapter is provided so that the device can be completely isolated from the power
source. To completely isolate the device from the power source unplug the plug from the
power source.
• Read and follow the “Correct Disposal of This Product” in the Technical Data Section
(page 23) when disposing of the device and any used accessories or optional parts.
Save these instructions for future reference.
8
1. Know your unit
Air Filter Cover
*Air filter inside
Power
Connector
Compressor (Main Unit)
Cap
Inhalation Air Inlet
Inhalation Top
Vaporizer Head
Medication Tank
Air Tube Connector
Nozzle
Rear View
Nebulizer Kit
Nebulizer Kit
Air Tube Plug
Nebulizer Kit Holder
Air Tube
Kids Accessory
(Smile Bear)
Air Tube Plug
Air Tube Connector
Power Switch
Kids Accessory
(Happy Bunny)
EN
9
1.Know your unit
Other included items
MouthpieceChild Mask (PVC)Infant Mask (PVC)
Spare Air Filters × 5AC adapterStorage Bag
Instruction ManualWarranty card
10
Compressor Nebulizer
Model
NE-C801KD
• Instruction Manual
• Mode d’emploi
• Gebrauchsanweisung
• Manuale di istruzioni
• Manual de instrucciones
• Gebruiksaanwijzing
•
TMTM
IM-NE-C801S-KDE(V)-02-06/2014
EN
FR
DE
IT
ES
NL
RU
AR
1887051-1B
Operating instructions
2. Preparing the nebulizer for use
Warning:
Clean and disinfect the nebulizer kit, kids accessory, mask, mouthpiece or nosepiece
before using them when as follows:
- the first time after purchase.
- if the device has not been used for a long period of time
- if more than one person uses the same device
For directions on cleaning and disinfecting refer to Section 4 (page 16-17).
Note: Make sure that the air filter is changed regularly. Refer to Section 6 “Changing the air filter”
(page 19).
2.1Connecting AC adapter
1.Make sure that the power switch is turned off.
Check that the power switch is in the off () position.
2.Inserting the power plug of the AC
adapter into the power connector
on the compressor.
3.Plug the AC adapter into a power
outlet.
Note: When disconnecting the AC
adapter, unplug the AC adapter
from the power outlet before
removing the power connector
on the compressor.
Power Plug
Power Outlet
Power Connector
EN
11
2.Preparing the nebulizer for use
2.2Adding medication and assembling the parts
1.Remove the mask or mouthpiece, kids accessory and
cap from the nebulizer kit.
2.Remove the inhalation top from the medication tank.
1) Rotate the inhalation top anti-clockwise.
2) Lift the inhalation top out of the medication tank.
3.Add the correct amount of prescribed medication to
the medication tank.
Caution:
The capacity of the medication tank is 2 to 7 ml.
Note: Scales on the nebulizer kit are for reference only. The
scale on the outside of the tank is with no vaporizer
head in place. Please use the scale on your syringe or
vial for accurate measurement of medication.
4.Make sure that the vaporizer head is securely
attached to the inhalation top.
5.Put the inhalation top back inside the medication
tank.
1) Align the protrusion on the inhalation top with the
indent on the medication tank, as shown.
2) Rotate the inhalation top clockwise and align with
the medication tank until they click in place.
12
2.Preparing the nebulizer for use
6.Attach the cap, kids accessory, mask or mouthpiece
to the nebulizer kit.
1) Attach the cap to the inhalation air inlet.
2) Attach the kids accessory to the inhalation top.
3) Attach the mask or mouthpiece to the inhalation
top.
Note: If the amount of aerosol is too small, it is not
necessary to place the cap.
7.Attach the air tube.
1) Twist the air tube plug slightly and push it firmly into the air tube connector on
the compressor.
2) Twist the air tube plug slightly and push it firmly into the air tube connector on
the medication tank.
Notes:
• Make sure that the air tube is securely attached to the compressor and nebulizer kit, and
does not come loose. Twist the air tube plug slightly when inserting it into the air tube
connectors to avoid the air tube disconnecting during use.
• Take care not to spill the medication when attaching the air tube. Keep the nebulizer kit
upright.
EN
13
3. Taking medication
1.Hold the nebulizer kit as indicated on the right.
Caution:
Do not tilt the nebulizer kit so that it is at an angle of greater
than 45° in all directions. Medication may flow into the mouth
or it may result in not nebulizing effectively.
Right angle
45q45q
2.Press the power switch to the on () position.
The compressor starts and nebulization starts.
Note: Observe whether aerosol is generated.
To stop nebulization
Press the power switch to the off () position.
Warning:
Do not cover the compressor with a blanket or towel etc.,
during use.
Caution:
• Do not block the air filter cover.
• Limit use to 20 minutes at a time, and allow a 40 minute interval before using the device
again.
• When using the device, the main unit may become hot.
• Do not touch the main unit for other than necessary operation such as turning off the power
while nebulizing.
Note: Do not leave the compressor on for a long period of time. This could result in the
compressor overheating or malfunctioning.
3.Inhale medication according to your doctor’s and/or health care adviser’s
instructions.
Attach the mask so that it covers your nose and
mouth and inhale the medication.
Exhale through the mask.
Using the mouthpiece
Insert the mouthpiece into your mouth and inhale the
medication breathing normally.
Exhale as normal through the mouthpiece.
14
3.Taking medication
4.When treatment is complete, turn the power off. Check that there is no
condensation or moisture in the air tube and unplug the device from the power
outlet.
Warning:
Condensation may occur in the air tube. Do not store the air tube while it contains
condensation or moisture. This could result in infection via bacteria.
If there is moisture or liquid remaining in the air tube, be sure to follow the
procedure below to remove any moisture from within the air tube.
1) Make sure that the air tube is still connected to the air connector on the
compressor.
2) Unplug the air tube from the nebulizer kit.
3) Turn on the compressor and pump air through the air tube to expel the
moisture.
Use the nebulizer kit holder as a temporary holder for the nebulizer kit.
Note: Align the mark on the unit with the mark on the nebulizer kit as shown
EN
15
Care and maintenance
4. Cleaning and disinfecting
4.1Cleaning
Clean the parts after each use to remove residual medication. This will avoid inefficient
nebulization or infection.
Wash them in warm water and mild detergent (neutral detergent). Rinse them thoroughly
with clean hot tap water and allow to air dry in a clean place.
■ Non-washable parts
• Compressor, Air Tube (PVC)
Firstly, make sure that the power plug is unplugged from the power outlet.
Wipe clean with a soft cloth moistened with water or mild detergent (neutral detergent).
• Air Filter
Do not wash or clean the air filter. If the air filter becomes wet, replace it.
Damp air filters can cause blockages.
4.2Disinfecting
Always disinfect the parts after the last treatment of the day. If the parts are heavily stained,
replace them with new ones.
To select a method for disinfection, refer to the table on page 17.
Caution:
Do not use an autoclave, EOG gas sterilizer or low temperature plasma sterilizer.
Note: The printing may come off the surfaces of kids accessories as a result of repeated washing and
disinfecting.
Handling the Vaporizer Head
The vaporizer head is an important part that is used to
vaporize medication.
Be sure to observe the following precautions when handling it.
Notes:
• Always wash the vaporizer head after each use.
• Do not use a brush or pin etc., to clean the parts.
• When disinfecting the parts by boiling, be sure to boil them in plenty
of water.
• Do not boil vaporizer head together with other objects, other than
applicable nebulizer accessories.
• Do not use a microwave oven, dish dryer or hair dryer to dry parts.
16
Baffle
Vaporizer Head
Disinfecting
method
A
BBoiling
* an example of commercially available disinfectant. (The concentration and residence time
4.Cleaning and disinfecting
o: applicable X: not applicable
Parts
Mouthpiece
Nebulizer Kit
Air Tube
(Silicone, 100cm)
(optional)
Adult Mask (PVC)
Materials
Alcohol
Disinfecting
ethanol
Sodium
hypochlorite
Milton*
(0.1%, 15min.)
Quaternary
ammonium
Osvan*
(0.1%, 10min.)
Chlorhexidine
Hibitane*
(0.5%, 30min.)
Amphoteric
Surfactant
Tego 51*
(0.2%, 15min.)
PP
PP
Tube: Silicone
Connector: PP
(Vaporizer Head: PC)
oo ooX
oo ooo
oo ooo
oo ooo
oo ooo
Child Mask (PVC)
Mask: PVC
Band: Rubber
(Phthalate free)
(non-latex)
oo oXX
specified in the table are under the conditions where the service life of the parts is tested with
each disinfectant used as described in its instruction manual. Please note that the testing was
not performed for the purpose of ensuring the effectiveness of the disinfectants. There is no
intention of suggesting that such disinfectants be used. Conditions of use and ingredients of
disinfectant vary for manufacturers.
Please read the instruction manuals carefully before use and apply disinfectant to each part in
an appropriate way. Please notice that service life of the parts may be shorter depending on
conditions, environments and frequency of use.)
Kids Accessory
PP, PVC
(Phthalate free)
Air Tube
(PVC, 100cm)
PVC
(Phthalate free)
XX
XX
XX
XX
XX
XX
Air Filter
Polyester
EN
A. Use a commercially available disinfectant. Follow the instructions provided by the
disinfectant manufacturer.
Note: Never clean with benzene, thinner or a flammable chemical.
B. Parts may be boiled between 15 to 20 minutes.
After boiling, carefully remove the parts, shake off excess water
and allow to air dry in a clean environment.
17
5. Maintenance and storage
To keep your device in the best condition and protect the device from damage follow these
directions:
Caution:
• Do not leave the device unattended with infants or persons who cannot express their consent.
• Do not subject the device or the parts to any strong shocks such as dropping the device on the floor.
• Do not carry or leave the nebulizer kit while the medication tank contains medication.
Notes:
• Do not store the device in extreme hot or cold temperatures, high humidity or under direct sunlight.
• Do not crease or fold the air tube.
• Dispose of the device and any used accessories or optional parts according to applicable local
regulations.
• Never clean with benzene, thinner or a flammable chemical.
Always transport the device in its storage bag, and store the device in
the bag if it will not be used for a long period.
Thinner
Benzene
Note: Nebulizer kit should be replaced after using for one year.
18
6. Changing the air filter
If the air filter has been used for more than 60 days, replace it with a new one.
1.Pull the air filter cover off
the compressor.
2.Change the air filter.
3.Put the air filter cover back
in place.
Make sure it is attached
properly.
EN
Notes:
• Use only OMRON air filters designed for this device. Do not operate without a filter.
• There is no front or back orientation for the air filters.
• Check that air filters are clean and free of dust before inserting them.
• Do not wash or clean the filter. If the air filter becomes wet, replace it. Damp air filters can
cause blockages.
• Please contact your OMRON dealer for additional air filters, or if the parts supplied with the
device are damaged or missing.
19
7. Troubleshooting
Check the following if your unit should fail during operation. You can also refer to the pages
of this manual for complete instructions.
ProblemCauseRemedy
Nothing happens when the
power switch is pressed.
No nebulization or low
nebulization rate, when the
power is on.
The device is abnormally
loud.
The device is very hot.Is the compressor covered?Do not cover the compressor
Note: If the suggested remedy does not solve the problem, do not try to repair the device - no parts of
the unit are user serviceable.
Return the unit to an authorized OMRON retail outlet or distributor.
Is the AC adapter connected
correctly to the an electrical
outlet and the compressor?
Is there medication in the
medication tank?
Is there too much/little
medication in the medication
tank?
Is the vaporizer head missing
and not assembled correctly?
Is the nebulizer kit
assembled correctly?
Is the nozzle blocked?Make sure that the nozzle is
Is the nebulizer kit tilted at a
sharp angle?
Is the air tube connected
correctly?
Is the air tube folded or
damaged?
Is the air tube blocked?Make sure that the air tube is
Is the air filter dirty?Replace the air filter with a
Is the air filter cover attached
correctly?
Is the device operating
continuously over 20
minutes?
Check that the plug is
inserted in an electrical
outlet. Unplug then reinsert
the plug if necessary.
Add the correct amount of
medication to the medication
tank.
Attach the vaporizer head
correctly.
Assemble the nebulizer kit
correctly.
free of blockages.
Make sure that the nebulizer
kit is not tilted at an angle of
more than 45 degrees.
Make sure that the air tube is
correctly connected to the
compressor and nebulizer kit.
Make sure that the air tube
does not contain kinks.
free of blockages.
new one.
Attach the air filter cover
correctly.
with any type of cover during
use.
Limit use to 20 minutes at a
time, and allow a 40 minute
interval before using the
device again.
Classification:Class ll equipment, Type B applied part
COMP AIR
100 - 240V 350mA, 50/60Hz
12V0.8A
+10
°C to +40°C 30% to 85% RH
-20
°C to +60°C 10% to 95% RH 700 - 1060 hPa
Bunny), Air Tube (PVC, 100cm), 5 pcs Spare Air Filters,
Mouthpiece, Infant Mask (PVC), Child Mask (PVC), AC adapter,
Storage Bag, Instruction Manual, Warranty Card.
= Class ll equipment
= Type B
applied part
Read the instruction
manual carefully
= No operation= Power on
Notes:
• Subject to technical modification without prior notice.
• This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co., Ltd.,
Japan.
• The device may not work if the temperature and voltage conditions are different to those defined in
the specifications.
• Do not use the device where it may be exposed to flammable gas.
• This device fulfils the provisions of the EC directive 93/42/EEC (Medical Device Directive) and the
European Standard EN13544-1:2007, Respiratory therapy equipment - Part1: Nebulizing systems
and their components.
EN
21
8.Technical data
0
10
20
30
40
50
60
70
80
90
100
1000110.1
Technical data for the OMRON NE-C801KD Compressor with the OMRON V.V.T.
(Virtual Valve Technology) Nebulizer kit:
Particle Size:
**MMAD Approx. 3μm
MMAD = Mass Median Aerodynamic Diameter
Result of cascade impactor measurements for particle size**
Cumulative % particle mass of sodium fluoride under size
Individual
Mean
MMAD 3.0 μm
Particle Size Dp (μm)
Cumulative Undersize %
* Measured by OMRON HEALTHCARE Co., Ltd.
**Independently measured by Prof. Dr. Hiroshi Takano, Department of Chemical
Engineering and Materials Science, Faculty of Science and Engineering, Doshisha
University, Kyoto, Japan, according to EN 13544-1:2007
Notes:
• Performance may vary with drugs such as suspensions or high viscosity. See drug supplier’s data
sheet for further details.
• See web site of OMRON HEALTHCARE EUROPE to update technical information.
URL: www.omron-healthcare.com
22
8.Technical data
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular)
telephones, medical devices in use may be susceptible to electromagnetic interference
from other devices. Electromagnetic interference may result in incorrect operation of
the medical device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the
aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has been
implemented. This standard defines the levels of immunity to electromagnetic
interferences as well as maximum levels of electromagnetic emissions for medical
devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate strong
electrical or electromagnetic fields, near the medical device. This may result in
incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the
device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON
HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it should not be
disposed of, with other household wastes at the end of its working life. To prevent
possible harm to the environment or human health from uncontrolled waste disposal,
please separate this from other types of wastes and recycle it responsibly to promote
the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this product,
or their local government office, for details of where and how they can take this item for
environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions of the
purchase contract. This product should not be mixed with other commercial wastes for
disposal.
This product does not contain any hazardous substances.
EN
23
9. Accessories and replacement parts
Nebulizer Kit No. : N-01
Mouthpiece
Optional Medical Accessories
(within the scope of EC Medical Device Directive 93/42/EEC)
V.V.T. Nebulizer Kit N-01 &
Mouthpiece Set
Model: NEB-NEST3-81E
V.V.T. Nebulizer Kit N-01 &
Smile Bear Set
Model: NEB-NKKD1-81E
V.V.T. Nebulizer Kit N-01 &
Happy Bunny Set
Model: NEB-NKKD2-81E
Mouthpiece
Model: NEB-MP-E
Air Tube (PVC, 100cm)
Model: NEB-TP-L10E
Child Mask (PVC)
Model: NEB-MSMP-E
Air Tube (Silicon, 100cm)
Model: NEB-TSC-L10E
Other Optional/Replacement Parts
Air Filters (x 5)
Model: NEB-AFR-30E
Infant Mask (PVC)
Model: NEB-MSSP-E
AC Adapter
Model: C30-E-AC
24
9.Accessories and replacement parts
25
EN
Manufacturer
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN
EU-representative
Production facility
Subsidiaries
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
www.omron-healthcare.com