Long Distance Industrial Wireless
Transmitters and Receivers
UWXL Series
UWXL-24-TC shown smaller
U Thermocouple, RTD,
Infrared Temperature,
RH, pH, Process Voltage/
Current and Input Models
U J, K, T, E, R, S, B, and N
Thermocouples
U 385 and 392 Pt100 RTD
Probes
U IP68 Rated Enclosure–
Dust- and Water-Tight
U Free Software Converts
Your PC Into a MultiChannel Chart Recorder
or Data Logger
U Compatible with
Both UWTC-REC and
UWTC-REC-NEMA
Wireless Receivers
U Long Battery Life
The UWXL Series industrial wireless
transmitters from OMEGA offer
unparalleled flexibility for process
measurement applications. Capable
of ranges to 450 m (1500’) line of
sight (LOS), the UWXL transmitters
are designed for a variety of
applications, including temperature,
infrared temperature, relative
humidity, process transducers with
standard voltage or current outputs,
as well as pH. Thermocouple
models are user-configurable for
Dust- and
Water-Tight
any of 9 calibrations, while RTD
units are compatible with both
0.00385 and 0.00392 curve Pt100
sensors. UWXL transmitters are
programmable to transmit data at
intervals configurable from once
every 2 seconds to every 2 minutes.
Transmitted data includes the
process measurement along with
the ambient temperature, RF signal
strength and battery condition.
Software provided with each model
can display this information in real
time. This free software allows you
to use your PC as a multi-channel
meter, recorder or data logger; data
can also be exported to a standard
than actual size.
spreadsheet file.
UWXL transmitters are compatible
with a wide variety of OMEGA
wireless receivers. UWTC-REC
receivers are available with optional
analog output, alarms and display,
while the UWTC-REC-NEMA
receivers add a NEMA rated
enclosure. Receivers can accept up
to 48 transmitters. Receivers with
analog outputs can also retransmit
data through a wired connection,
as an analog voltage, current or
thermocouple signal, to interface
with a controller, PLC or data
acquisition interface.
Up to 450 m
(1500') Line of
Sight Range
IP68
Enclosure
Models shown smaller
than actual size.
Process Voltage
Transmitter
Humidity
Transmitter
Thermocouple
Transmitter
Typical Multi-Unit Wireless
Automation System
Up to 450 m (1500')
Infrared
Transmitter
UWTC-REC4
receiver.
W-31
Analog Output (Wired)
HE-XE102 PLC (Programmable Logic
Controller), visit omega.com/ocsxl
Transmitter model
UWXL-24-RH-RPI.
Humidity/temperature
sensor included.
UWXL-RH features a detachable
combination humidity/temperature
sensor that can be remotely mounted.
Temperature transmitter models
UWXL-TC-24, UWXL-RTD.
Thermocouple and RTD models feature
screw terminal connections that allow you to
connect a wide range of sensors and probes.
Models
with intergral probes are available
contact factory.
Transmitter model
UWXL-IR-1.
Infrared sensor
head included.
W
UWXL-IR-1 features a remotely mounted
sensor head to measure non-contact
infrared temperature.
Common Specifications
Computer Interface: USB
Transmit Sample Rate: Programmable
from 2 second to 120 second
Radio Frequency (RF) Transceiver
Carrier: ISM 2.4 GHz, direct sequence
spread spectrum, (2.450 to 2.490 GHz–
12 RF channels)
RF Output Power: 18 dBm (63 mW)
Range of RF Link: Up to 450 m (1500')
outdoor line of sight; up to 90 m (300')
indoor/urban
RF Data Packet Standard: IEEE
802.15.4, open communication
architecture
Software (Included Free): Requires
Windows
operating system
Power: One 3.6V, Lithium C Cell
(included)
Battery Life (Typical): 3 years;
at 1 sample/minute reading rate @ 25°C
Operating Temperature: -10 to 70°C
(14 to 158°F)
Storage Temperature: -40 to 80°C
(-40 to 176°F)
Model UWXL-24-TC
Input: J, K, T, E, R, S, B, C or N
thermocouple, software selectable
Measurement Range:
J: -100 to 760°C (-148 to 1400°F)
K: -100 to 1260°C (-148 to 2300°F)
T: -200 to 400°C (-328 to 752°F)
E: -200 to 1000°C (-328 to 1832°F)
R: 260 to 1760°C (500 to 3200°F)
S: 260 to 1760°C (500 to 3200°F)
B: 870 to 1820°C (1598 to 3308°F)
C: 0 to 2315°C (32 to 4200°F)
N: -100 to 1260°C (-148 to 2300°F)
®
2000, XP or Vista (32 bit)
Accuracy @ 25°C:
Types J and K: ±0.5% rdg or ±1.0°C
(1.8°F), whichever is greater
Types T, E, and N: ±0.5% rdg or
±2.0°C (3.6°F), whichever is greater
Types R, S, B and C: ±0.5% FS
Resolution: 1°C/1°F
Sensor Connection: Screw terminals
Model UWXL-24-RTD
Input: Pt100 RTD; 0.00385 or 0.00392
curve, software selectable
Measurement Range:
Pt100, 0.00385: -200 to 850°C
(-328 to 1562°F)
Pt100, 0.00392: -100 to 457°C
(-148 to 854°F)
Accuracy: ±0.5°C (1.0°F)
Resolution: 1°C/1°F
Sensor Connection: Screw
terminals
Model UWXL-24-IR-1
Temperature Range: -18 to 538°C
(0 to 1000°F)
Accuracy: ±2% rdg or 2.2°C (4°F),
whichever is greater; @ 22°C (72°F)
ambient temperature and emissivity
≥0.95
Optical Field of View: 6:1 (distance/
spot size)
Sensor Head Cable Extension:
Up to 15 m (50’) total
Repeatability: ±1% rdg
Spectral Response: 5 to 14 microns
Response Time: 100 ms (0 to 63% of
final value)
Emissivity: 0.1 to 1.00, adjustable
Operating Ambient:
Sensor Head: 0 to 70°C (32 to158°F)
Sensor Head (-HT Model): 0 to 85°C
(32 to 185°F)
Sensor head with OS100-WC
(Water Cooling Jacket): 0 to 200°C
(32 to 392°F)
Relative Humidity: Less than
95% RH, non-condensing
OS100-WC Water Flow Rate:
0.25 GPM, room temperature
Thermal Shock: About 30 minute
for 25°C (77°F) abrupt ambient
temperature change
Warm-Up Period: 3 minutes
OS100-AP Air Flow Rate: 1 CFM
(0.5 L/s) (1 x 2.5")
OS100-LA Laser Sight
Accessory
Wavelength (Color): 630 to 670 nm (red)
Operating Distance (Laser Dot):
Up to 9.1 m (30')
Max Output Optical Power: Less
than 1 mW @ -6°C (22°F) ambient
temperature
European Classification: Class 2,
EN60825-1/11.2001
Max Operating Current: 45 mA
@ 3 Vdc
FDA Classification: Complies with
21 CFR 1040.10, Class II laser product
Beam Diameter: 5 mm (0.20")
Beam Divergence: <2 mrad
Operating Ambient: 0 to 50°C (32 to
122°F); <95% RH, non-condensing
Power Switch: On/off slide switch on
the battery pack
Power Indicator: Red LED
Power: Battery pack, 3 Vdc (consists
of two 1.5 Vdc “AA” lithium batteries,
included)
W-32