Wireless Relative Humidity/
Temperature Transmitter
UWRH-2-NEMA
U NEMA 4X (IP65) Weather
Resistant Enclosure
U Measures Both
Relative Humidity and
Temperature
in One Unit
U Free Software Converts
Your PC Into a MultiChannel Chart Recorder
or Data Logger
U Low Power Operation
and Sleep Mode For
Long Battery Life
U Transmits Relative
Humidity, Ambient
Temperature, Signal
Strength and Battery
Status in Real Time
U Works with Any
UWTC Series Wireless
Receivers and
Transceivers
OMEGA’s new Wireless Relative
Humidity transmitter features a
stand-alone, compact, battery
powered NEMA design that
transmits measurements back to
a host receiver up to 90 m (300')
away. When activated the wireless
transmitter will transmit readings
continuously at pre set time interval
that was programmed by the user
during the initial setup. Each unit
measures and transmits: Relative
Humidity Reading, Ambient
Temperature, RF Signal Strength
and Battery Condition to the
receiver and is displayed on your
PC screen in real time using the
provided software. When used
with host receiver model UWTCREC1 data from up to 96 wireless
transmitters can be received and
displayed. Each unit includes free
software that converts your PC into
a strip chart recorder or data logger
so readings can be saved and later
printed or exported to a spread
sheet file. When used with host
transceiver model UWTC-REC2
wireless data from one unit can be
re-transmitted out of the receiver
by a wired connection as a analog
voltage or current signal to interface
with a controller, PLC or data
acquisition board.
Specifications
Temperature Range: -17 to 49ºC
(2 to 120ºF)
Accuracy: ±1ºC (±1.8°F)
Relative Humidity
Accuracy: ±2.5% RH from 20 to 80%
RH; ±3.5% RH below 20 and above
80% RH@ 25°C (76°F)
Operating Environment: -10 to 70°C
(14 to 158°F)
Computer Interface: USB (one
interface cable included with receiver)
Transmit Sample Rate: Programmable
from 2 seconds to 2 minutes
Radio Frequency (RF) Transceiver
Carrier: ISM 2.4 GHz, direct sequence
spread spectrum.
RF Output Power: 10 dBm (10 mW)
Range of RF Link:
Up to 90 m (300'): Outdoor line of sight
Up to 45 m (130'): Indoor/urban
To Order Visit omega.com/uwrh-2-nema for Pricing and Details
Model No. Description
UWRH-2-NEMA Wireless RH/temperature transmitter
UWTC-BATT-C
UWTC-CABLE Spare programing cable
Comes complete with one 3.6V lithium battery, programming software, measurement and
logging software and operator’s manual.
Ordering Example: UWRH-2-NEMA, wireless RH/temperature transmitter, UWTC-REC1,
96-channel receiver and UWTC-BATT-C, spare battery.
Visit omega.com/wireless for more information on UWTC-REC1.
UWRH-2-NEMA wireless
relative humidity
temperature transmitter,
shown smaller than
actual size.
Range: 2 to 98% RH
Replacement 3.6V lithium “C” battery assembly
RF Data Packet Standard:
IEEE 802.15.4, open communication
architecture
Software (Included Free): Requires
Windows
Battery: One 3.6V lithium, (C) (included)
Battery Life (Typical): 3 year at 1 sample/
minute reading rate @ 25°C (77°F)
Data Transmitted to Host: Relative
humidity, ambient temperature, RF transmit
strength and battery condition
Enclosure: NEMA 4X (IP65)
Polycarbonate
Dimensions: 80 L x 82 W x 50 mm H
(3.15 x 3.22 x 1.97")
For the Latest
Features and
Specifications Visit
omega.com/wireless!
®
2000 or higher operating system
H
u
Hu-6