Omega Products UWPH-2-NEMA User Manual

Wireless pH/Temperature Transmitter
with Automatic Temperature Compensation
UWPH-2-NEMA
Weather Proof/ Wall Mounted
U 0 to 14 pH and 0 to 100°C
(32 to 212°F) Ranges
U
Transmit Data in Real-Time, Up to 120 m (400’)
U Free Software Converts Your PC into
a Multi-Channel Chart Recorder or Data Logger
U
Low Power Operation and Sleep Mode
U Works with UWTC-REC Receivers for
a Complete Wireless System
OMEGA’s new wireless pH/temperature transmitter features a high performance microprocessor based wireless radio transmitter built into a NEMA enclosure. Compatible with most pH probes with a BNC connector, the UWPH provides fast, accurate readings. For automatic temperature compensation, the UWPH accepts a Pt100 RTD probe through a standard (series T) connector. Configured through a standard USB port, the UWPH can transmit data at rates from every 2 seconds to every 2 minutes. Each transmitted reading includes the pH and temperature data, along with RF signal strength and battery condition to the host. Using the standard software (included), this data can be displayed on screen in real time. The software allows the user to use your PC as a meter, chart recorder or data logger, so data can be saved, or exported to a spreadsheet file.
Works with Wireless Receivers:
UWTC-REC1 48-Channel Receiver
UWTC-REC2 48-Channel Receiver with
Analog Output and Alarm
UWTC-REC4 DIN Rail Mount 4-Channel Receiver with 4-Channel Analog Output and Alarms
UWTC-REC6 1-Channel Receiver with Analog Output
UWPH-2-NEMA,
wireless
pH/temperature
transmitter.
PHE-1311 gel-filled pH electrode.
FCC/Industry
PRTF-10-2-100-1/8-6-E-TA3F RTD probe.
USB
connection
(cable
UWTC-REC2 models are available with 4 to 20 mA, 0 to 5 Vdc, 0 to 10 Vdc or Type K thermocouple output. Mating connector and cable included.
pH
pH
Laptop not included.
Canada
Approved
All models shown smaller than actual size.
Data
Logging and
Recording Software
Included
14 pH
W-13
Recommended RTD and pH Electrode
PRTF-10-2-100-1/8-6-E-TA3F
RTD probe shown smaller than actual size.
PHE-1311 gel-filled
pH electrode shown
smaller than actual size.
W
SPECIFICATIONS
Input Range: 0 to 14 pH Accuracy: ±0.1 pH Resolution: 0.01 pH Response Time: 2 second max Input Connection: BNC Temperature Compensation:
Automatic, 0 to 100°C (32 to 212°F)
RTD Temperature Input: Input Type: Pt100 Ω, 0.00385 curve Range: 0 to 100°C (32 to 212°F) Accuracy: ±1°C (1.8°F) Resolution: 0.1°
Input Connection: TA3M receptacle;
TA3F mating connector (included)
Computer Interface: USB Transmit Sample Rate: Programmable
from 2 seconds to 2 minutes
Radio Frequency (RF) Transceiver Carrier: ISM 2.4 GHz, direct sequence
spread spectrum
RF Output Power: 10 dBm (10 mW) Range of RF Link: Outdoor Line of Sight:
Up to 120 m (400') Indoor/Urban: Up to 40 m (130')
RF Data Packet Standard: IEEE 802.15.4, open communication architecture
Software (Included Free): Requires Windows operating system
Power: One 3.6V, lithium “C” cell (included)
Battery Life (Typical): 3 years; 1 sample/minute reading rate @ 25°C
Enclosure: NEMA 4X polycarbonate Enclosure Dimensions:
80 L x 82 mm W (3.15 x 3.23")
Note: pH and RTD probes sold separately.
®
2000, XP or Vista (32-bit)
To Order Visit omega.com/uwph-2-nema for Pricing and Details
Model No. Description
UWPH-2-NEMA Wireless pH/temperature transmitter
Receivers/Accessories
Model No. Description
UWTC-REC1 USB-powered 48-channel transmitter receiver UWTC-REC2-(*) 48-channel receiver with analog output UWTC-REC2-D-(*) 48-channel receiver with analog output and display
UWTC-REC4-(*) 48-channel DIN rail receiver with 4 analog outputs and alarms UWTC-REC6-(*) 1-channel transceiver with analog output UWTC-CABLE Spare USB programming/communication cable UWTC-BATT-C Replacement 3.6V, lithium “C” cell battery assembly PHE-1311 Gel-filled pH electrode, general purpose
PRTF-10-2-100-1/8-6-E-TA3F
PHA-4 4.00 pH buffer solution 500 mL (1 pint) bottle PHA-7 7.00 pH buffer solution 500 mL (1 pint) bottle PHA-10 10.00 pH buffer solution 500 mL (1 pint) bottle
Comes complete with measurement/data logging software, one 3.6V lithium “C” cell assembly, TA3F RTD mating connector, and operator’s manual. UWTC-REC2 units also include DC power adaptor.
Specify analog output signal: “V1” for 0 to 5 Vdc; “V2” for 0 to 10 Vdc, or “MA” for 4 to 20 mA.
*
Ordering Example: UWPH-2-NEMA, wireless pH/temperature transmitter, UWTC-REC1, USB powered 48-channel transmitter receiver, PHE-1311, gel-filled pH electrode and PRTF-10-2-100-1/8-6-E-TA3F, Pt100 RTD probe.
Pt100 RTD probe, ¹⁄₈" dia, 6” L, with 40' PFA insulated cable with TA3F connector
W-14