Omega Products TRP-C User Manual

Supplied with male
Series TRP Reference Probes up to six can be used at the same time
Copper leads
Connectors
TRP Application-Probe Calibration
"Working"
Model TRC III-A
Thermocouple leads
Temperature Reference Probes
With Standard Size Connectors for Thermocouple Calibration
Dimensions: mm (inch)
3 (1⁄8) diameter
140 (51⁄2)
TRP Series
U All Standard Calibrations Available U Comes Complete with Mating
Connectors and Locking Clips
U Can Be Ordered with 300 mm (12"),
600 mm (24") or 900 mm (36") Leads
U 304 Stainless Steel Sheath Material
Accurate thermocouple temperature measurement requires a stable reference junction. It is common practice to create a transition reference junction by attaching a copper lead wire to each thermocouple leg. When the transition reference junction is held at the ice point 0°C (32°F) the output of the thermocouple appearing across the copper leads attached to the readout instrument is stable and predictable. For accuracy and convenience, OMEGA offers its series TRP REFERENCE PROBE which locates the thermocouple­copper transition reference junction at the end of an 1⁄8" diameter sheath for easy insertion into the TRCIII-A ice point calibration chamber. Up to six series TRP Probes can be used at one time in OMEGA’s Ice Point Chamber. Use the temperature­millivolt tables to convert readings into °C or °F when using a voltmeter with the TRCIII-A ice point™ reference cell shown below.
Model TRCIII-A ice point™
OMEGA’s TRP’s are used in conjunction with ice point™ reference cells such as TRCIII-A, see omega.com
To Order Visit omega.com/trp-k for Pricing and Details
Model Thermocouple Material Type
Number +
TRP-K* CHROMEGA
TRP-N* OMEGA-P® OMEGA-N TRP-T* Copper Constantan TRP-E* CHROMEGA TRP-J* Iron TRP-S TRP-R TRP-B TRP-C TRP-D
Note: 12" leads are standard. For 24 or 36" long leads add suffix “-24” or “-36” to model
*
number for an additional cost. † Standard length is 6", for each additional 6" length there is an additional cost. Ordering Examples: TRP-J-24. TRP-R-12.
A. To readout instrument “Copper-Copper” B. To measuring junction “Thermocouple Alloys”
TRP Application-Probe Calibration
Thermocouple leads
"Working" Thermocouple under test in a known temperature environment
Platinum
Platinum Platinum 13% Rh
Platinum 30% Rh Pt 6% Rh
W 5% Re W 26% Re
W 3% Re W 25% Re
B
A
300 (12)
To Recorder or Readout
TRP Application-Meter Calibration
TRCIIIA
Copper Wire
Cu
Cu
Supplied with male and female color-coded thermocouple connectors
ALOMEGA
Constantan Constantan
Platinum 10% Rh
Thermocouple Wire
a (+)
b (-)
Thermocouple
Meter
Copper leads
Connectors
Model TRC III-A
Color-coded Copper-Copper connectors
Series TRP Reference Probes up to six can be used at the same time
®
®
To Probe
Precision
Voltage
Source
®
®
A-37