Omega Products TRCIII-A-230VAC User Manual

ice point™ Calibration Reference Chamber
High Precision Thermoelectric Chamber Uses Water to Maintain 0°C (32°F) Continuously
U Stylish Design
TRCIII-A
Standard
230 Vac
model
only
U Portable Rugged Case for Use in the
Factory, Calibration Laboratory, or Instrument Shop
U Built-in Temperature Readout
with 0.1 (°F/°C) Resolution
Thermometers and Sensors
TRCIII-A shown smaller than actual size.
The ice pointTM TRCIII-A Calibration Reference Chamber is part of OMEGA’s fine line of calibration reference instrumentation.
The TRCIII-A ice pointTM Reference Chamber relies on the equilibrium of ice and distilled, deionized water at atmospheric pressure to maintain 6 reference wells at precisely 0°C (32°F). Each well extends into a sealed cylindrical chamber containing the distilled deionized water. The outer walls of the chamber are cooled by thermoelectric cooling elements. The increase in volume produced by the creation of ice crystals within the cell is sensed by the expansion of a bellows, which operates a microswitch and controls the cooling elements. The alternate freezing and thawing of the ice accurately maintains a 0°C (32°F) environment around the reference wells.
K
Probes sold separately; please next page.
Features
U 230 Vac CE Marked Models U NIST Calibration Certificate Included U Bench-Top Design with Tilt Handle
Standard
U Illuminated Power Switch U Companion hot point® Dry Block
Probe Calibrator Available (Model CL900A/950A)
K-14
Specifications
Reference Wells: 6 wells, 4.0 mm
5
(
32") ID, 95.25 mm (33⁄4") deep; wells are thermally and electrically grounded to each other; accepts
1
3 mm (
Reference Temperature: 0°C (32°F) Accuracy: ±0.1°C (±0.18) Stability: ±0.03°C (±0.07) for
constant ambient Environmental Ambient:
2 to 32°C (35 to 90°F)
Power: 115/230 Vac, 50 Hz, 104 W Stabilization Time: 2 hours after
initial power up Dimensions:
203 H x 419 W x 305 mm D (8 x 16
Weight: 11.8 kg (26 lb)
8") OD probes
1
2 x 12")
OMEGACARESM extended
SM
warranty program is available for models shown on this page. Ask your sales
Extended Warranty
Program
Unit must not be allowed to
freeze or damage will result.
We make running changes when
technical advances allow. Check at time
of ordering for additional features.
representative for full details when placing an order. OMEGACARE labor and equivalent loaners.
Caution:
SM
covers parts,
TRP Applications
Thermocouple leads
Connectors
TRP Applications
Copper leads
Series TRP Reference Probes
– up to 6 can be used at the same time
Model TRCIII-A
Working thermocouple under test in a known temperature environment
White-coded copper-copper
Supplied with male and female color-coded thermocouple connectors
connectors
To Order Visit omega.com/trciii-a for Pricing and Details
Model No. Description TRCIII-A ice point reference, 120 Vac operation TRCIII-A-230VAC* ice point reference, 230 Vac operation
Comes complete with operator’s manual and NIST calibration certificate.
Note: Only 230 Vac model is available as CE marked.
*
Ordering Examples: TRCIII-A, ice point™ reference cell, TRP-K, TRP-J, TRP-T, TRP-E, Type K, J, T and E temperature reference probes.
OCW-3, OMEGACARESM extends standard 1-year warranty to a total of 4 years.
TRCIII-A Accessories
TRP-T shown with additional wires attached to mating connectors.
Thermocouple TRP Probes Thermocouple Material Type
Thermocouple TRP Features
U Made with Special Limits
of Error Material
U All Standard
Calibrations Available
U Comes Complete with
Mating Connectors
U Can Be Ordered with
12, 24 or 36" Leads
Model No. + -
TRP-K CHROMEGA® ALOMEGA®
TRP-T Copper Constantan
TRP-E CHROMEGA® Constantan
TRP-J Iron Constantan
TRP-S Platinum Platinum 10% Rh
TRP-R Platinum Platinum 13% Rh
TRP-B Platinum 30% Rh Platinum 6% Rh
TRP-C W 5% Re W 26% Re
TRP-D W 3% Re W 25% Re
Refrigeration required for below ambient cooling.
*
Available as an option where not supplied as standard as HCTB-COIL.
**
††
For 240 Vac power, add suffix “-240V” to model number, no additional charge.
Note: For 4-point NIST calibration, points: -20°C, 60°C, 120°C, 180°C, add suffix “-NIST” to
model number for additional cost.
K-15