ice point™ Calibration
Reference Chamber
High Precision—Uses Freezing
Water to Maintain 0°C (32°F)
TRCIII
Optional
U A 0°C (32°F)
Thermoelectric
“Refrigerator”
U Eliminates Old
Fashioned
“Ice Bath”
U For Versatile Use
in the Factory,
Laboratory,
Use Up to
6 Probes at
One Time
with TRCIII
Instrument Shop
U Calibration of All
Temperature Instruments
and Sensors
U Rugged Outer Case for
Safe Portability
U Accepts Up to 6 Probes
U Available in 120 or
220 Vac Models
U NIST Traceable
Calibration Available
U Companion Dry Block
Probe Calibrator
Available (Model
CL900/950)
The TRCIII ice pointTM Calibration
Reference Chamber relies on the
equilibrium of ice and distilled,
deionized water at atmospheric
pressure to maintain 6 reference
wells at precisely 0°C (32°F). The
wells extend into a sealed cylindrical
chamber containing the distilled,
deionized water. The outer walls
of the chambers are cooled by
thermo-electric cooling elements.
The increase in volume produced by
the creation of ice crystals within the
TRCIII
cell is sensed by the expansion of a
bellows, which operates a microswitch
and controls the cooling elements.
The alternate freezing and thawing
of the ice accurately maintains a
0°C (32°F) environment around the
reference wells.
The automatic operational features
of this instrument eliminate the need
for frequent attention, required of
common ice baths. On/off cycling
of a pilot light indicates proper
operation of the unit at ice point
temperature.
Any combination of thermocouples
may be used with this instrument by
simply inserting the reference probes
in the reference wells. Calibration of
other types of temperature sensors
at 0°C (32°F) may also be performed.
When accurate temperature
measurements are required, the
TRCIII ice pointTM Chamber delivers
the precision you need. The reference
wells are maintained at 0°C (32°F)
to within ±0.1°C and a stability
of ±0.04°C for constant ambient.
Continuous cycling operation
maintains constant equilibrium
conditions within the cell, eliminating
the slow drift often experienced
in common ice baths as melting
K
TRCIII ice point™ shown with companion
CL950 dry block probe calibrator.
See omega.com for details.
occurs. TRP reference probes, used
with the TRCIII, provide a junction
from thermocouple alloys to copper
wire, at the probe tip. When inserted
into the TRCIII, TRP probes enable
the change from thermocouple alloy
to copper at 0°C (32°F), for accurate
cold junction compensation.
K-12
Thermocouple
TRP Features
U Made with Special
Limits of Error Material
U All Standard
Calibrations Available
U Comes Complete with
Mating Connectors
U Can Be Ordered with
12, 24 or 36" Leads
TRP Applications
Thermocouple leads
Connectors
TRP Applications
Copper leads
Series TRP Reference Probes
up to six can be used at
the same time
Model TRCIII
–
Sturdy handle for easy portability.
Specifications
Reference Wells: 6 wells, 5⁄32" ID, 33⁄4"
deep; wells are thermally and electrically
grounded to each other; accepts
probes
Reference Temperature: 0°C
Accuracy: ±0.1°C or better
Stability: ±0.04°C for constant ambient
Environmental Ambient: 2 to 32°C
(35 to 90°F)
Power Required: 115/220 V, 60 Hz,
100 Watts
Stabilization Time: 2 or 3 hours
depending on ambient temperature
and thermal load
Dimensions:
305 H x 159 W x 322 mm D
(12.0 x 6.25 x 12.7")
Weight: 9.5 kg (21 lb)
Extended Warranty
Program
SM
OMEGACARE
is available for models shown on this page.
Ask your sales representative for full details
when placing an order. OMEGACARE
covers parts, labor and equivalent loaners.
extended warranty program
1
⁄8" OD
SM
SM
Working
Thermocouple
under test in a
known temperature
environment
White-coded
Copper-Copper
Supplied with male
and female
color-coded
thermocouple
connectors
connectors
TRP-T shown with additional wires attached to mating connectors.
Unit must not be allowed to
freeze or damage will result.
Caution:
Thermocouple TRPs
Thermocouple Material Type
Model No. + TRP-K CHROMEGA
®
ALOMEGA
®
TRP-T Copper Constantan
TRP-E CHROMEGA
TRP-J Iron
TRP-S Platinum
®
Constantan
Constantan
Platinum 10% Rh
TRP-R Platinum Platinum 13% Rh
TRP-B Platinum 30% Rh Platinum 6% Rh
TRP-C W 5% Re W 26% Re
TRP-D W 3% Re W 25% Re
For thermocouple Types J, K, T, E, standard length is 12". For 24" or 36" long leads add
suffix “-24” or “-36” to model number, for an additional cost.
For thermocouple Types R, S, B, C, G, D, standard length is 6". Add appropriate suffix for
each additional 6", for an additional cost.
To Order Visit omega.com/trciii for Pricing and Details
Model No. Description
TRCIII ice point™ reference 120 Vac operation
TRCIII-220 ice point™ reference 220 Vac operation
CAL-3 NIST-traceable calibration
Comes complete with operator’s manual, mineral oil and power cord 1.8 m (6').
Ordering Example: TRCIII, ice point™ reference cell, TRP-K, TRP-J, TRP-T, TRP-E,
Type J, K, T and E temperature reference probes.
OCW-3, OMEGACARESM extends standard 1-year warranty to a total of 4 years.
K-13