Omega Products OS1513-L1-R3-4 User Manual

J-115
Fiber Optic Sensors
For Use with DP1511, DP1521 and DP1531 Monitors
±1.0%Accuracy Emissivity AdjustmentUnaffected by
Electrical Noise
Chambers at up to 10-7Torr
OS1513 General Purpose Sensor
10 msec Time Constant160 to 1535°C
(320 to 2800°F) Temperature Range
0.020" at 6" Minimum
Spot Size Available
OS1533 High Speed Sensor
0.3 msec Time Constant425 to 1370°C
(800 to 2500°F) Temperature Range in Subranges of No More than 2:1 (exam) 400-800°
0.010" at 3" Minimum
Spot Size Available
OS1513/OS1533 Series
Starts at
$
1715
The OS1513 and OS1533 fiber optic detectors are designed to be used specifically with the DP1511, DP1521 and DP1531 monitors and are appropriate for a variety of applications.
These fiber optic detectors are capable of temperatures measurements from 160 to 1535°C (320 to 2800°F).
These detectors are designed to be used with the appropriate monitor when conventional line of sight infrared measurements are not possible. The advantages of these fiber optic sensors include the ability to bypass opaque obstacles and to
DP1531 19" single channel high speed monitor, $4995, sold separately.
Custom Assemblies Available
Small Spot Sizes Down to 0.004"
Temperatures to 427°C (800°F) at Pressures
to 30,000 PSI
Ceramic/Metal Tip Probe for Induction Heating
Where Probe is Close to Surface
Polymer Bolt Probe for Extrusion Applications
Ejector Pin Probe for Injection Molding
OS1513-L1-R1-4, $1775, shown smaller than actual size.
Mounting bracket included.
DP1511 general purpose single channel monitor, $3040, sold separately, see page J-121.
OMEGACARESMextended warranty program is available for models shown on this page. Ask your sales representative for full details when placing an order. OMEGACARE
SM
covers parts, labor and
equivalent loaners.
J-116
J
Specifications
Ambient Temperature:
10 to 50°C (50 to 122°F)
Accuracy: ±1.0% rdg Repeatability: ±0.5% rdg Spectral Response:
0.91 to 2.3 microns
Time Constant: 10 msec, OS1513; 0.3 msec, OS1533
Dimensions: 6.4 H x 10 W x 13 cm L (2.5 x 4 x 5")
Weight: 1.8 kg (4 lb) Fiber Optic Cable: Glass or quartz fiber
with heavy duty stainless steel jacket standard, 0.020"/0.070 cable Dia., 149°C (300°F) ambient max
To Order
(Specify Model Number)
Base Temperature Spot Size @ Cable
Model No. Price Range Target Dist Length
OS1513-L1-R1-4 $1775 540 to 1090°C 0.020" @ 6" 1.2 m General Purpose 1000 to 2000°F (4')
OS1513-L1-R2-1 1715 160 to 340°C 0.075" @ 6" 0.3 m General Purpose 320 to 650°F (1')
OS1513-L1-R3-4 1775 260 to 540°C 0.22" @ 8" 1.2 m General Purpose 500 to 1000°F (4')
OS1513-L1-R4-4 1775 760 to 1535°C 1.0" @ 26" 1.2 m General Purpose 1400 to 2800°F (4')
OS1533-L1-R5-4 2400 675 to 1370°C 0.10" @ 12" 1.2 m High Speed 1250 to 2500°F (4')
OS1533-L1-R6-5 2420 815 to 1370°C 0.040" @ 6" 1.5 m High Speed 1500 to 2500°F (5')
OS1533-L1-R7-4 2700 425 to 870°C 0.010" @ 3" 1.2 m High Speed 800 to 1600°F (4')
Comes with connection cable to a monitor, mounting bracket, and operator’s manual. Consult OMEGA for ordering and pricing on custom OS1513 or OS1533 fiber optic systems. Notes: Other temperature ranges and spot size @ target distance and vacuum
bushing available; consult factory. The OS1513/OS1533 fiber optic sensor do not operate
stand-alone. Must be used in conjunction with DP1511, DP1512 or DP 1531 monitors.
Ordering Example: OS1533-L1-R7-4, high speed sensor with temperature range of 425 to 870°C, spot size of 0.010" @ 3", and 1.2 m (4') cable, $2700. OCW-3, OMEGACARESMextends standard 2-year warranty to a total of 5 years ($350), $2700 + 350 = $3050.
reach into closed areas for viewing. They can take measurements unattainable by conventional infrared pyrometers, as they are unaffected by media in the environment, such as smoke, water vapor, etc. The OS1513 and OS1533 are recommended for measurements in high energy fields or of small objects, down to 0.004" in diameter. For remote mounting of the electronics out of hazardous environments, the fiber optic lens and cable may be remotely mounted up to 9 m (30') from the infrared detector head. They can also be used in vacuum chamber applications, where the fiber optic assembly can be placed in vacuums to 10-7Torr. Also available is a backlighting capability, where a beam of light illuminates the target, providing accurate indication of the target area.
Important factors which determine the fiber optic assembly best suited for your application are: the distance from the fiber optic lens to the surface being measured, the size of the object being measured; the temperature range being measured; the ambient temperature around the fiber optic cable; the length of the fiber optic cable; how the surface is being heated; and the atmospheric conditions between the sensor and the product.
Typical Fiber Optic Probes shown. L2, L3 and L4 are custom assemblies, Consult OMEGA for details. L1 shown with optional bifurcated cable for backlighting.
L1
L4
L2
L3
Available Probe Types L1 Lens Probe (Standard) L2 Ceramic/Metal Tip L3 Polymer Bolt L4 Ejector Pin Probe
Backlight source,
see OS1500-BLS for specifications.
OS1512 infrared detector shown with the DP1500 monitor.
MOST POPULAR MODELS HIGHLIGHTED!