Omega Products OS136-1-MA User Manual

Miniature Low-Cost Non-Contact Infrared Temperature Sensor/Transmitter
OS136 miniature stainless
steel sensor/transmitter
with NEMA 4 housing
shown actual size.
Each unit includes 2 hex mounting nuts,
1.8 m (6') shielded cable and operator’s manual.
(INCH)
(INCH)
OS136
Optional
U 2 Temperature Ranges
Available: OS136-1: -18 to 202°C (0 to 400°F) OS136-2: 149 to 538°C (300 to 1000°F)
U Sensor and Transmitter
Combined
U Packaged into a 19 mm
3
(
4") OD by 89 mm (3.5") Long Stainless Steel Housing
U Optical Field of View:
6 to 1
U Fixed Emissivity of 0.95 U NEMA 4 (IP66)
Rated Housing
U 4 to 20 mA, 0 to 5 Vdc,
0 to 10 Vdc, Type K Thermocouple, 10 mV/ Degree Analog Outputs Available
U Makes Fast and Accurate
Measurements
The OMEGA® miniature, low-cost, high-performance infrared sensor/ transmitter model OS136 features a 19 mm (3⁄4") OD by 89 mm (3.5") long stainless steel NEMA 4 (IP66) housing. This mini transmitter is ideal for applications that require measuring temperature in hard­to-reach, confined, or harsh environments. The OS136 comes with a 1.8 m (6') shielded cable for power and output connections. Models feature industry standard outputs that provide easy interfacing directly to all meters, controllers, data loggers, recorders, computer boards, and PLCs. The fixed emissivity of 0.95 allows for quick and easy measurement requiring no adjustment during installation and use.
Refer to Accessories chart on next page for
NIST calibration ordering information.
Specifications
Temperature Range: OS136-1: -18 to 202°C (0 to 400°F)
OS136-2: 149 to 538°C (300 to 1000°F)
J-61
Accuracy @ 22°C (72°F) Ambient:
OS136-1: 3% of rdg or 4.4°C (8°F), whichever is greater OS136-2: 3% of rdg or 5.5°C (10°F), whichever is greater from 185 to 510°C (365 to 950°F)
Repeatability: 1% of rdg Optical Field of View: 6 to 1
(distance to spot size)
Spectral Response: 5 to 14 microns Response Time: 150 msec,
0 to 63% of final value
Emissivity: Fixed at 0.95 Analog Output:
MA: 4 to 20 mA V1: 0 to 5 Vdc V2: 0 to 10 Vdc K: Type K thermocouple, compensated MVC: 10 mV/°C MVF: 10 mV/°F Output Load Requirements: Min Load (0 to 5 Vdc): 1 kΩ Min Load (0 to 10 Vdc): 2 kΩ Max Load (4 to 20 mA): (Power supply – 4) / 20 mA Min Load (10 mV/Deg): 10 kΩ Min Load (K T/C): 100 kΩ
Operating Ambient Temperature:
No Water Cooling: 0 to 70°C
(32 to 158°F) With Water Cooling (OS136-WC): 0 to 200°C (32 to 392°F) With Air Cooling (OS136-WC): 0 to 110°C (32 to 230°F) Operating Relative Humidity: Less than 95% RH, non-condensing Water Flow Rate for OS136-WC:
0.25 GPM, room temperature, minimum Air Flow Rate for OS136-WC: 5 CFM (2.4 liters/sec)
Warm-Up Period: 1 to 2 minutes Thermal Shock: About 30 minutes for
25°C abrupt ambient temperature change
Air Flow Rate for Air Purge Collar:
1 CFM (0.5 liters/sec) Transmitter Housing: 316 SS, NEMA 4 (IP66) rated
Power: 12 to 24 Vdc @ 50 mA Dimensions: 19 OD x 89 mm L
(0.75 x 3.5")
Weight: 181 g (0.40 lb)
Laser Sight Accessory Specifications
Laser Wavelength (Color):
630 to 670 nm (red)
Operating Distance: Up to 9.1 m (30') Max Laser Power Output:
Less than 1 mW @ 22°C ambient
Safety Classification:
Class 2 EN60825-1/11.2001 FDA Classification: Class II laser product; complies with 21 CFR 1040.10
Laser Beam Diameter:
Less than 5 mm (0.2")
Beam Divergence: Less than 2 mrad Operating Temperature:
0 to 50°C (32 to 122°F)
Operating Relative Humidity:
Less than 95% RH, non-condensing Power Switch: On/off, slide switch on the battery pack
Power Indicator: Red LED
Connect your OS136 to a precision controller CNi8DH, sold separately. For additional meters see omega.com.
CNi8DH shown actual size.
Power: Battery pack, 3 Vdc (included) Caution and Certification Label: Located on the head sight
circumference Identification Label: Located on the head sight circumference Aperture Label: Located on the head sight circumference
Dimensions:
38 D x 50.8 mm L (1.5 x 2")
OMEGACARESM extended warranty program is available for models shown on this page. Ask your sales representative for full details when placing an order. OMEGACARE covers parts, labor and equivalent loaners.
To Order Visit omega.com/os136 for Pricing and Details
Model No. Description OS136-(*)-K Infrared transmitter with Type K thermocouple output OS136-(*)-MV-F Infrared transmitter with 10 mV/°F output OS136-(*)-MV-C Infrared transmitter with 10 mV/°C output OS136-(*)-MA Infrared transmitter with 4 to 20 mA output OS136-(*)-V1 Infrared transmitter with 0 to 5 Vdc output OS136-(*)-V2 Infrared transmitter with 0 to 10 Vdc output
Comes complete with 2 hex mounting nuts, 1.8 m (6’) shielded cable and operator’s manual.
Insert “1” for -18 to 202°C (0 to 400°F); insert “2” for 149 to 538°C (300 to 1000°F).
*
Ordering Examples: OS136-1-MA, infrared transmitter with temperature range of -18 to 202°C
(0 to 400°F) and 4 to 20 mA output.
OCW-3, OMEGACARE OS136-1-V1 and CNi8DH33, infrared transmitter with temperature range of -18 to 202°C
(0 to 400°F), 0 to 5 Vdc output, and
SM
extends standard 1-year warranty to a total of 4 years.
1
8 DIN temperature controller.
Accessories Model No. Description OS100-MB Mounting bracket OS100-AP Air purge collar OS136-WC Water/air cool jacket OS100-LS†† Laser sighting TX4W-50 4-conductor shielded high-temp cable, 15 m (50') TX4-100 4-conductor shielded cable, 30 m (100') PSR-24S
PSR-24L PSR-24L-230
Regulated power supply, US plug, 90 to 264 Vac input, 24 Vdc output, 400 mA, screw terminals, UL Regulated power supply, US plug, 90 to 264 Vac input, 24 Vdc output, 400 mA, stripped leads, UL Regulated power supply, European plug, 230 Vac input, 24 Vdc output, 400 mA, stripped leads, CE
PSU-93 Unregulated power supply, 16 to 23 Vdc, 300 mA max, screw terminal CAL-3-IR† NIST-traceable calibration
††
OS136-LS laser sighting: 1 unit suitable for aligning multiple heads. Used during installation only.
Ordering Example: OS100-MB, mounting bracket.
Extended Warranty
Program
SM
SM
J
J-62
Loading...