Omega Products ORETX-600 User Manual

omega.com
Differential pH or Orp Probes With 4 to 20 mA Outputs
PHETX-600/
ORETX-600 Series
PHETX-600
1
1
NPT differential
2
pH electrode.
PHETX-620, hot tap
pH electrode with
PHE-620-BVA, PVC
hot tap ball valve.
All models shown
smaller than actual size.
PHETX-610 compression fitting pH electrode.
U Integral 2-Wire 4 to 20 mA Transmitter U Accurate Differential Measurement U Eliminates Ground Loop Interference U Built-In Pre-Amp Supports Up to 914 m (3000')
Sensor-to-Analyzer Distance
U 4.6 m (15') Standard Cable Length U Automatic Temperature Compensation on pH Versions
Applications
U Process Control, Industrial and Municipal Water Treatment U Industrial and Municipal Waste Treatment,
and Neutralization
U Fume Scrubbers, Plating, and Circuit Board Manufacturing U Food and Beverage U Chemical Processing U Pulp and Paper U Mining and Power Generation U Pharmaceuticals
OMEGA provide accurate measurements under conditions that often render conventional pH probes inoperable. These probes, come with an integral encapsulated 4 to 20 mA, 2‑wire transmitter to feed directly to a PLC or a DCS.
The PHETX‑600 and ORETX‑600 probes employ a differential measurement technique. Unlike conventional combination probes, the differential probe has two high impedance measurement circuits containing a common metallic return electrode. One circuit includes the process measurement electrode which generates a potential E1 proportional to the process pH. The second circuit
®
pH and ORP differential probes stay in service and
includes an internal measurement electrode immersed in a stable buffer solution which generates a standard reference potential, E2. Both circuit share a common potential E3 developed at the return electrode. The two circuits are fed into amplifiers which provide an output representing the differential between them: (E1 ‑ E 3)– (E2 ‑ E3). The common potential E3 is cancelled out electronically, greatly reducing inaccuracies caused by ground loops which may exist between process and instrument grounds. Ground loop current will flow through the low impedance path of the return electrode, affecting the potential E3, but not the differential measurement.
The differential probe maintains its accuracy and stability in aggressive process applications long after a combination‑style probes performance begins to deteriorate. Maintenance costs are reduced and the life of the probe is increased. The internal reference electrode is electrically connected to the process solution by means of a field‑replaceable double junction salt bridge which greatly reduces the possibility of contamination of the buffer solution in the reference circuit. Although seldom required, the reference solution may be easily replaced by removing the screw‑out salt bridge. A salt bridge and buffers are available for this purpose.
Another advantage of the PHETX/ORETX‑600 series are the semi‑flush face which is easily cleaned and avoids solution materials gathering on protrusions found in other probes. The domed glass electrode, the protective metal electrode and the temperature sensor protrude only about 1⁄8” while the salt bridge is flush. A flat‑face version of the pH probe is also available.
The PHETX/ORETX‑600 series is available in: 1 body style (‑600), “easy‑in, easy‑out” variable insertion depth version with 1 version (‑620).
1
NPT compression fitting (‑610), or the hot tap
4
1
NPT threaded
2
Order Online
Over 100,000 Products Available!
®
To download information and to order Industrial pH Instrumentation and Electrode products online,
D-1
Specifications
4.6 m (15') Cable
PHETX/ORETX-600
4.6 m (15') Cable
180
(7.10)
228
(9)
394
(15.5)
29
(1.14)
29
(1.14)
¾ NPT
Compression
Fitting
25 (1)
25 (1)
1½ NPT
1½ NPT
37 (1.44)
61 (2.4) Sensor
removal
clearance
for front
thread mount
Min 22 (7/8) Max 127 (5)
PHETX/ORETX-610
PHETX/ORETX-620
394
(15.5)
29
(1.14)
25 (1)
PHETX/ORETX-620
Measuring Range: pH: 0 to 14 pH (consult factory for applications below
2 or above 12)
ORP: ‑500 to 500 mV or 0 mV to 1000 mV, field selectable
Flow Rate: 10 ft/sec max (3 meters/sec) flow should be as low as
possible in low conductivity water and in solutions with high suspended solids
Wetted Materials: CPVC, kynar, glass, titanium palladium alloy and EPDM (platinum for ORP probe); opt FKM and ceramic
Transmission Distance: Dependant upon transmission distance and supply voltage
Dimensions: mm (inch)
4.6 m (15') Cable
1½ NPT
180
(7.10)
Sensitivity: pH: 0.001 pH ORP: 0.1 mV
Stability pH: 0.03 pH per day, non‑cumulative
ORP: 2 mV per day, non‑cumulative Temperature Compensation (Automatic): ‑5 to 95°C (23 to 203°F) Pressure Limit: 100 psig at 65°C (149ºF) max Temperature Limits (CPVC): ‑5 to 95°C (23 to 203°F) the
temperature limit of probes in flow through applications is limited by pressure and by the pipe fitting material.
Probe Cable: 5 conductor plus shield, 4.6 m (15') long
1½ NPT
61 (2.4) Sensor
removal
clearance
for front
thread mount
PHETX/ORETX-600
37 (1.44)
To Order Visit omega.com/phetx-600 for Pricing and Details
Model No. Description
PHETX-600 1¹⁄₂ NPT differential pH electrode with 4 to 20 mA outout ORETX-600 1¹⁄₂ NPT differential ORP electrode with 4 to 20 mA outout PHETX-610 Compression fitting pH electrode with 4 to 20 mA output for ⁷⁄₈ to 5" variable insertion ORETX-610 Compression fitting ORP electrode with 4 to 20 mA output for ⁷⁄₈ to 5" variable insertion PHETX-620 Hot tap pH electrode with 4 to 20 mA output for ⁷⁄₈ to 5" variable insertion ORETX-620 Hot tap ORP electrode with 4 to 20 mA output for ⁷⁄₈ to 5" variable insertion
Comes complete with electrode, salt bridge and operator’s manual.
Accessories Model No. Description
PHE-600-SB Replacement salt bridge kit with PVDF outer junction (3 pack) for PHETX/ORETX‑600 series PHE-612-SB Replacement salt bridge kit with PVDF outer junction (3 pack) for PHETX/ORETX‑610/620 series PHE-600-EP Electrode protector for PHETX/ORETX‑600 PHE-610-EP Electrode protector for PHETX/ORETX‑610/620 PHE-620-BVA PVC hot tap ball valve assembly for PHETX/ORETX‑620 PHA-4 4.00 pH buffer solution 500 mL (1 pint) bottle PHA-7 7.00 pH buffer solution 500 mL (1 pint) bottle PHA-10 10.00 pH buffer solution 500 mL (1 pint) bottle
Ordering Examples: PHETX-600, 1 PHETX-620, hot-tap differential pH electrode with 4 to 20 mA output, and PHE-620-BVA, PVC hot tap ball valve assembly.
1
NPT diffiential pH electrode with 4 to 20 mA output, PHA-4, pH 4 buffer, and PHA-7, pH 7 buffer.
2
D-2
Our Industrial pH Instrumentation and Electrodes product line continues to expand, visit omega.com for new details!