Omega Products ORE-5551-10 User Manual

General Purpose 1⁄2 MNPT pH Sensor
Dimensions: mm (inch)
PHE-7357-10
Applications
The PHE-7357 sensor is designed for use in most water and waste water applications. The large reservoir of reference electrolyte, coupled with the non-fouling nature of the Porous PTFE liquid junction, assures a long life in general service.
The sensor body reduces the risk of accidental breakage during installation and maintenance by the inclusion of protective lobes around the sensing element. Also available for ORP measurement.
Use of the high temperature version is recommended where service temperatures exceed 80°C (176°F).
PHE-5551 is a double-junction construction with solution ground in a robust PPS housing, being highly resistant to electrode poisoning solutions. Standard on these elctrodes is a 110°C (230°F) rating.
Specifications
pH Range: 0 to 14 pH Temperature Range:
0 to 80°C (32 to 176°F) (standard) 0 to 110°C (32 to 230°F) (high temperature)
Pressure Range: 0 to 100 psig Accuracy: ±0.1% over full range Impedance: 80 MΩ (standard version)
150 MΩ (high temperature version) Reference Cell: Single junction, KCl/AgCl
Reference Junction: Porous PTFE Wetted Materials:
PPS, PTFE, glass, 316 SS Response Time: 95% of reading in 10 seconds
Drift: Less than 2 mV per week
½ NPT
140
(5.50)
½ NPT
16.5
(0.650)
12.2
(0.480)
PHE-7357
PHE-7357-10
PHE-5551-10
Both shown smaller
than actual size.
To Order Visit omega.com/phe735710_555110 for Pricing and Details
Model No. Description
PHE-7357-10-( PHE-5551-10-(*) General purpose pH sensor with solution ground ORE-7357-10 ORP probe for clean water ORE-5551-10 ORP probe for clean water with solution ground
Comes complete with 3 m (10') cable and operator’s manual. Non standard cable lengths available for additional cost.
* Specify ATC sensor for PHE-5551-10-(*) only: “-PT100” for 100 Ω Pt RTD or “-PT1K’ for 1000 Ω Pt RTD, for additional cost.
Ordering Examples: PHE-5551-10-PT1K, general purpose sensor with 1 KΩ Pt RTD ATC and 3 m (10') cable.
ORE-5551-10, ORP electrode.
Option Code (For Additional Cost)
Order Suffix Description
-HF HF/Fluoride resistant
-HT High temperature reference cell, standard on PHE/ORE-5551
) General purpose pH sensor
*
1
2 MNPT pH sensor with
D-39
Our Industrial pH Instrumentation and Electrodes product line continues to expand, visit omega.com for new details!
For Sales & Service