Omega Products ORE-5460 User Manual

In Line Flat Surface pH/ORP Electrodes
PHE-5460 Series
U Designed For ¾,
1, or 2" Pipes or Pipe Tees
U CPVC
Construction with Acryl Gel Standard For High Temperature Applications
PHE-5460, and PHEH-54-10-PT100, shown smaller than actual size.
In Line
CABLE ASSEMB L Y
MOUNTING GLAND WITH 3/4 NPT MALE PIPE THREADS
TEE (N O T INCLUDED )
OPTIONAL ATC ELEMENT P O TTED IN 1/4 NPT NIPPL E (INCLUDES 3/4" PVC TEE)
In Line Electrodes for ¾ 1, 2" Pipe
Flat Surface In-Line Electrodes (CPVC)
In-Line Electrodes
The PHE/ORE-5460 in-line pH and ORP electrodes feature a flat, self-cleaning abrasion-free surface. When exposed to turbulent flow, the resulting scrubbing action keeps the electrodes clean. The electrodes can be used in applications with 0 to 87°C (32 to 190°F) temperatures, pressures up to 6.89 bar (100 psig) and a 0 to 14 pH measurement range. Having fresh material sweeping past the electrode’s surface improves response time as well as accuracy, especially for the measurement of oily waste fluids, lime slurries, flocculant and emulsions. The electrodes accept a cable assembly with a gland mounted in a
In-line electrodes are ideal for replacing existing threaded electrodes. They are used in applications where flow can be shut off for electrode maintenance.
The high temperature Acrylamide gel is supplied as standard with the PHE-5460 electrodes. These are only available with CPVC construction.
Maximum Temperature: 87°C (190°F) A vailable Options:
-HF” Fluoride resistant, additional cost “-LC” Low conductivity, additional cost “-GL” Ground loop, additional cost
Visit omega.com/phe-5460 for additional specifications and available options.
3
, 1 or 2 NPT tee.
4
To Order
Visit omega.com/phe-5460 for Pricing and Details
Model No. Description PHE-5460 pH electrode for ¾" pipe, CPVC
PHE-5460-1 pH electrode for 1" pipe, CPVC PHE-5460-2 pH electrode for 2" pipe, CPVC ORE-5460 ORP electrode for ¾" pipe, CPVC ORE-5460-1 ORP electrode for 1" pipe, CPVC ORE-5460-2 ORP electrode for 2" pipe, CPVC
Mounting Assemblies (CPVC) — Required for Installation PHEH-54-10 ¾ NPT mounting assembly and cable without ATC PHEH-54-10-1 1 NPT mounting assembly and cable without ATC PHEH-54-10-2 2 NPT mounting assembly and cable without ATC
PHEH-54-10-(*) ¾ NPT mounting assembly and cable with ATC PHEH-54-10-1-(*) 1 NPT mounting assembly and cable with ATC PHEH-54-10-2-(*) 2 NPT mounting assembly and cable with ATC
Note: Temperature compensation is not required with ORP electrodes. Order electrode and mounting assemblies separately. Materials of construction must be the same. Mounting Assemblies come with 3 m (10') cable with BNC connector. Cable lengths available
up to 15 m (50'). Consult Engineering for ordering information and per foot price of additional cable.
* Specify ATC Sensor
Order Code Description
PT100 100 Ω Pt RTD PT1K 1000 Ω Pt RTD TH700 Series 700 Thermistor R3K 3000 Ω Balco
Ordering Examples: PHE-5460-GL-HF ¾" in-line pH electrode with ground loop interrupt circuit and HF resistant glass.
PHEH-54-10-PT100 mounting assembly with ¾ NPT with 100 Ω Pt RTD ATC and 3 m (10') cable. PHE-5460,
3 m (10') cable.
3
" in-line pH electrode, PHEH-54-10, mounting assembly with
4
NOTE: At the time of initial purchase, please order
both electrode and mounting assembly. Mounting assembly includes insertion assembly and cable. If ordered with automatic temperature compensation (ATC), the temperature compensator is built into the insertion assembly. When electrode has expired, simply re-order electrode.
3
NPT with
4
D-31
Our Industrial pH Instrumentation and Electrodes product line continues to expand, visit omega.com for new details!
For Sales & Service
U.S.A. and Canada