Omega Products ORE-2715 Installation Manual

Page 1
User’s Guide
Shop online at
www.omega.com
e-mail: info@omega.com
PHE/ORE Series
pH/ORP Electrodes
Page 2
OMEGAnet
®
Online Service Internet e-mail
www.omega.com info@omega.com
Servicing North America:
USA: One Omega Drive, P.O. Box 4047
ISO 9001 Certified Stamford CT 06907-0047
TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
Customer Service: 1-800-622-2378 / 1-800-622-BEST Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
®
®
®
Mexico: En Español: (001) 203-359-7803 e-mail: espanol@omega.com
FAX: (001) 203-359-7807 info@omega.com.mx
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
TEL: +31 (0)20 3472121 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: nl@omega.com
Czech Republic: Rudé armády 1868, 733 01 Karviná 8
TEL: +420 (0)69 6311899 FAX: +420 (0)69 6311114 Toll Free: 0800-1-66342 e-mail: czech@omega.com
France: 9, rue Denis Papin, 78190 Trappes
TEL: +33 (0)130 621 400 FAX: +33 (0)130 699 120 Toll Free in France: 0800-4-06342 e-mail: france@omega.com
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
TEL: +49 (0)7059 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: germany@omega.com
United Kingdom: One Omega Drive, River Bend Technology Centre
ISO 9002 Certified Northbank, Irlam, Manchester
M44 5EX United Kingdom TEL: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-488 e-mail: sales@omega.co.uk
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient-connected applications.
page 2 OMEGA PHE/ORE Series pH/ORP Electrode Instructions
Page 3
OMEGA PHE/ORE Series pH/ORP Electrode Instructions
1.0 in.
(27 mm)
0.9 in.
(23 mm)
4.1 in.
(104 mm)
3.4 in.
(86 mm)
1.2 in.
(30.5 mm)
2.6 in.
(65 mm)
1.0 in.
(27 mm)
1.2 in.
(30.5 mm)
2.6 in.
(65 mm)
0.9 in.
(23 mm)
3.4 in.
(86 mm)
3.8 in.
(97 mm)
50
3
PSIBAR
50
12275167
25 77
0
10
32
50
25
2
75
5
100
7
°F
°C
PHE-3271
PHE-3271-HF
SAFETY INSTRUCTIONS
1. Depressurize and vent system prior to installation or removal.
2. Confirm chemical compatibility before use.
3. Do not exceed maximum temperature/pressure specifications.
4. Wear safety goggles or faceshield during installation/service.
5. Do not alter product construction.
6. When using chemicals or solvents care should be taken and appropriate eye, face, hand, body, and/or respiratory protection should be used.
1. Technical Data
1.1 General
Compatibility: OMEGA PH-2720-PA Twist-Lock Preamplifier
Pipe Size Range:
•1/2 in. and up.
Use OMEGA installation fittings to 4 in.
Use pipe adapter in pipes over 4 in.
Submersion with PH-2720-PA pre-amplifier requires 3/4 in. NPT or ISO 7-R 3/4 in. male threaded extensions
Efficiency: >97% @ 25°C (77° F) Wetted Materials: CPVC Body
Glass FPM O-rings Porous UHMW Polyethylene reference junctions ORP: Platinum sensing surface
Primary Functions: PHE-3271, ORE-2715: Flat surface
resists fouling PHE-2716, ORE-2717: Bulb surface for general use PHE-3271-HF: Extended use in applications with trace hydrofluoric acid (<2%) PHE-2716-DI: Extended use in pure waters (<100 µS)
1.4 Temperature and pressure specifications
ORP electrode operating Temperature/Pressure
6.89 bar @ 0°C to 65ºC (100 psi @ 32° to 149° F)
4.00 bar @ 66ºC to 85°C (58 psi @150° to 185°F)
PHE-2716 and PHE-2716-DI pH electrode operating Temperature/
Pressure
6.89 bar @ 0°C to 65ºC (100 psi @ 32° to 149° F)
4.00 bar @ 66ºC to 85°C (58 psi @150° to 185°F)
PHE-3271 pH electrode operating Temperature/Pressure
6.89 bar @ 10°C to 65ºC (100 psi @ 50° to 149° F)
4.00 bar @ 66ºC to 85°C (58 psi @150° to 185°F)
Secondary Junction: Nylon Filament Element: Ag/AgCl Shipping Weight: 0.2 kg (0.4 lb)
1.2 PHE-3271 and PHE-2716 pH Electrodes
Operating Range: 0 to 14 pH (PHE-3271-HF: 0 to 12 pH) Reference: Electrolyte: Solidified Acrylamide Gel
3.5M KCI (PHE-3271, PHE-3271-HF, PHE-2716)
Temperature Sensor: 3K Balco (3000 = 25°C) ResponseTime: <5 s for 95% of signal change Response Time, τ: 140 s (PHE-3271), 196 s (PHE-2716)
0.1 M KCl (PHE-2716-DI)
1.3 ORE-2715/ORE-2717 ORP Electrodes
Operating Range: -999 to +999 mV Reference: Electrolyte: Solidified Acrlyamide Gel
Response Time: Application dependent
3.5M KCL
OMEGA PHE/ORE Series pH/ORP Electrode Instructions
PHE-3271-HF pH electrode operating Temperature/Pressure
6.89 bar @ 0°C to 50ºC (100 psi @ 32° to 122° F)
Storage Temperature: > -12° C (11° F)
1.5 Dimensions
Flat electrode Bulb electrode
page 3
Page 4
2. Electrode care and application
pH/ORP electrodes are similar to batteries; they age with time and usage. The following information will help maximize electrode life.
2.1 Conditions to Avoid:
High temperatures, strong acids or caustics will elevate electrochemical reactions and speed electrode aging.
Coatings on the glass or junction surfaces (i.e. proteins) cause extended response time and inaccurate measurement.
Never store the electrode tip in deionized (DI) water.
Never expose electrode to temperatures below -12 °C (10 °F) or allow it to dehydrate. These conditions will damage the electrode.
Never scrape or sand the glass electrode surface.
Treat glass electrode surfaces with care. The glass is very thin and requires care to prevent accidental breakage.
2.2 Submersible Installation Tips:
Mount electrodes in a location with ample clearance for removal for periodic cleaning and recalibration. Choose a location that keeps the
YES
YES
electrode glass completely submerged at all times.
Place the electrode tip in pH 4 buffer during system maintenance or storage to avoid dehydration.
Mount the electrode near tank outlet away from reagent addition areas.
2.3 In-Line Installation Tips:
pH and ORP electrodes respond best in moving fluids.
30°
NO
NO
30°
The internal measuring electrode chamber contains a wire within a liquid and a slight amount of air. The electrode must be mounted at least 30° from horizontal to ensure proper sensing. Mounting angles less than 30° will impede performance.
NO
3. pH Electrode Calibration
All pH electrodes are designed to ensure linearity during their lifespan. The following sections define proper electrode operation.
3.1 Offset (STD)
Electrode offsets occur due to:
Clogged reference junction
Aged or contaminated reference solution/wire
Check offsets in a pH 7 buffer @ 25 °C. The theoretical output is 0 mV. Any deviation from 0 mV is the pH electrode offset.
pH Electrode Offset pH 7 buffer @ 25°C Theoretical: 7.0 pH (0.0 mV)
New: pH 7 ± 0.25 pH (±15 mV)
Reliable: pH 7 ± 0.85 pH (± 50 mV) Electrode offsets greater than 0.85 pH (50 mV) indicate the electrode requires cleaning or replacement. See section 5: Maintenance and cleaning.
°C pH Error
3.2 Slope (SLP)
Electrode slope is the number of mV per pH unit. At 25°C the theoretical slope is 59.16 mV per pH. Temperature has an appreciable affect on electrode slope. Reliable instrumentation includes temperature compensation. The graph below illustrates potential pH error when a temperature compensated instrument is not used.
23456789101112 15 0.15 0.12 0.09 0.06 0.03 0 0.03 0.06 0.09 0.12 0.15 25 00000000000 35 0.15 0.12 0.09 0.06 0.03 0 0.03 0.06 0.09 0.12 0.15 45 0.3 0.24 0.18 0.12 0.06 0 0.06 0.12 0.18 0.24 0.3 55 0.45 0.36 0.27 0.18 0.09 0 0.09 0.18 0.27 0.36 0.45
Theoretical mV Values @ 25°C
pH mV
2+ 295 mV 3+ 236 mV 4 +177 mV 5 +118 mV 6+ 59 mV 70 mV 8- 59 mV
9 - 118 mV 10 - 177 mV 11 - 236 mV 12 - 295 mV
Recommendations:
Calibrate temperature before calibrating the standard and slope.
The mV offset will track across the entire pH range. The slope is usually not affected by offset changes. (i.e. pH 7= +10 mV, pH 4= +187 mV); slope = 59 mV
Coatings on the glass may affect sensor slopes. See section 5 maintenance and cleaning.
•A constant output near 0 mV in all buffer solutions indicates a shorted electrode that must be replaced.
3.3 Response Time/Stability
Response time and stability are affected by the condition of the pH electrode's glass surface (ORP electrode - Platinum surface), reference junction, and reference solution. Restoration to acceptable levels can often be accomplished by cleaning the electrode's glass surface (ORP electrode - Platinum surface) and reference junction.
Electrode mV values should remain stable ±3 mV. Conditions that may cause fluctuations are:
1. Electrode coating
2. Ground fault: If proper operation is observed in the beaker, but is unstable in the application, a ground fault probably exists.
Using instrumentation with isolated inputs and outputs may restore stable operation.
Solution grounding may also restore stable operation.
page 4 OMEGA PHE/ORE Series pH/ORP Electrode Instructions
Page 5
4. ORP Electrode Calibration
All ORP electrodes are designed to ensure linearity during their lifespan. The following sections define proper electrode operation.
4.1 Offset (STD)
Electrode offsets occur due to:
Clogged reference junction
Aged or contaminated reference solution/wire
Offsets are easily checked in pH 7 buffer saturated with quinhydrone @ 25 °C; since the theoretical output is +87 mV. Any deviation from +87 mV is the ORP electrode offset (i.e. +90 mV). Quinhydrone is the oxidizer measured by the ORP electrode and is required for calibration. To guarantee buffer saturation, mix 1/8g quinhydrone per 50 mL of pH buffer.
ORP Electrode Offset: Solution: pH 7 or pH 4 buffer saturated with Quinhydrone @ 25 °C
pH 7 + Quinhydrone pH 4 + Quinhydrone
Theoretical mV: + 87 mV + 264 mV New: 87 ± 15 mV +264 ± 15 mV Reliable: 87 ± 50 mV +264 ± 50 mV
Electrode offsets greater than ±50 mV indicate the electrode requires cleaning or replacement, see section 5.2.
4.2 Slope (SLP)
ORP slope errors are generally caused by contamination of the platinum electrode surface. Cleaning the electrode surface will usually restore proper values, response time, and stability.
Common ORP Values
Reaction mV
3
-
+ e
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
- 913
- 440
- 407
- 400
- 370
- 250
- 126
- 37 0 + 771 + 799 + 800 + 1060 + 1066 + 1160 + 1188 + 1369
Cr Cr2+ + 2e Fe → Fe2+ + 2e Cr2+ Cr3+ + e 4OH- O2 + 2H2O + 4e - 401 2I- I2 + 2e Ti2+ Ti3+ + e Ni Ni2+ + 2e Pb Pb2+ + 2e Fe Fe3+ + 3e H2 2H+ + 2e Fe2+ Fe3+ + e Ag Ag+ + e Pb Pb4+ + 4e 3Br- Br 2Br- → Br2 + 2e
-
ClO
ClO2 + e
2
Pt Pt2+ + 2e Ag Ag2+ + 2e
Many systems require both pH and ORP calibration. To conserve calibration reference solutions, use pH 7 and 4 buffers for pH calibration first. ORP calibration can be performed with the same buffers by adding quinhydrone.
5. Maintenance and Cleaning
5.1 Maintenance
Variables can affect long term pH or ORP electrode life. For this reason, a maintenance log is recommended for trend analysis. When storing boxed sensors, lay the sensor flat to maximize hydration of the glass surface. Keep the glass surface wet at all times. Soak the sensor tip in pH 4.0 buffer during system maintenance intervals. In-line applications should be plumbed with a depression (trap) so liquid is maintained around the sensor tip. If the sensor dehydrates, soak the sensor tip in pH 4 buffer for 24 to 48 hours, then visually inspect the electrode for surface cracks, swelling, or discoloration. Severely dehydrated electrodes cannot be restored to normal operation.
OMEGA PHE/ORE Series pH/ORP Electrode Instructions
page 5
Page 6
5.2 Cleaning
Cleaning techniques vary depending on the type of coating present on the glass electrode surface or reference junction.
Soft coatings can be removed by vigorous stirring, or with directed spray of an applicable detergent or solvent onto the glass electrode surface. Chlorine bleach or mild detergent may be used to remove soft coatings. Always rinse electrode tip in clean water after cleaning.
Hard coatings can be chemically removed. Use the least harsh chemical which will remove the contaminant within two (2) minutes without attacking the materials of construction. e.g. calcium carbonate may be removed with a 5% HCL (muriatic acid) solution.
Oily or organic coatings can be removed with detergents or an appropriate solvent that does not attack the materials of construction e.g. isopropyl alcohol may be used but acetone must be avoided to prevent damage to the CPVC sensor body.
ORP electrode surface (platinum rod) can be gently sanded with 600 grit wet and dry silicone or carbide sandpaper, jewelers rouge, crocus cloth, or very fine steel wool.
WARNING!
Wear appropriate eye, face, hand, body, and respiratory protection when using chemicals or solvents.
6. Replacement parts and accessories
Mfr. Part No. Description
PHE-3271 Flat pH Electrode PHE-3271-HF Flat pH Electrode, HF Resistsant ORE-2715 Flat ORP Electrode PHE-2716 Bulb pH Electrode PHE-2716-DI Bulb pH Electrode, DI Resistant ORE-2717 Bulb ORP Electrode PH-270-PA Preamplifier, 3/4 in. NPT P31515-0V200 PVDF Pipe Adapter FPP-1220-0021 O-ring, FPM (standard) FPP-1224-0021 O-ring, EPR FPP-1228-0021 O-ring, Kalrez
CAUTION!
Apply O-ring lubricant to sensor O-rings prior to assembly. Unlubricated O-rings may score the sealing surface in the 2720.
Technical Features
a) Male BNC connector b) Keyed for a simple sure fit c) Viton® O-rings d) Silicone-bushing seal e) Double junction f) Ag/AgCl reference element g) Platinum sensing surface (ORP) or flat pH glass h) Porous UHMW polyethylene reference junction
i) 3K Balco thermistor j) Solidified acrylamide reference
electrolyte k) Ag/AgCl measuring element l) Large reference volume m) Epoxy seal n) Shielding o) 10KΩ I.D. resistor for ORP sensor
a
o n
m
b
c
d
c
l
k
j
i
e
f
h
g
page 6 OMEGA PHE/ORE Series pH/ORP Electrode Instructions
Page 7
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR
WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the
product was PURCHASED,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2000 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of
the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA PHE/ORE Series pH/ORP Electrode Instructions
page 7
Page 8
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at www.omega.com
TEMPERATURE

Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies

Wire: Thermocouple, RTD & Thermistor

Calibrators & Ice Point References

Recorders, Controllers & Process Monitors

Infrared Pyrometers
PRESSURE, STRAIN AND FORCE

Transducers & Strain Gages

Load Cells & Pressure Gages

Displacement Transducers

Instrumentation & Accessories
FLOW/LEVEL

Rotameters, Gas Mass Flowmeters & Flow Computers

Air Velocity Indicators

Turbine/Paddlewheel Systems

Totalizers & Batch Controllers
pH/CONDUCTIVITY

pH Electrodes, Testers & Accessories

Benchtop/Laboratory Meters

Controllers, Calibrators, Simulators & Pumps

Industrial pH & Conductivity Equipment
DATA ACQUISITION

Data Acquisition & Engineering Software

Communications-Based Acquisition Systems

Plug-in Cards for Apple, IBM & Compatibles

Datalogging Systems

Recorders, Printers & Plotters
HEATERS

Heating Cable

Cartridge & Strip Heaters

Immersion & Band Heaters

Flexible Heaters

Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL

Metering & Control Instrumentation

Refractometers

Pumps & Tubing

Air, Soil & Water Monitors

Industrial Water & Wastewater Treatment

pH, Conductivity & Dissolved Oxygen Instruments
6-2714.090-OM/(H-4/03) M-2922/0403
Loading...