Omega Products OM-VLT User Manual

E-1
$
255
Basic Unit
OM-VLT
Temperature Data Logger with Built-In Printer
Model OM-VLT is a programmable temperature data logger, strip recorder and alarm device with integral digital display. The easy to follow menu driven programming is done via the front panel keys.
The unit records, saves and prints up to 290 temperature readings at predefined intervals. The OM-VLT produces an alarm when the temperature goes out of the defined range or after the temperature remains out of range for a certain period of time. The OM-VLT prints temperature data directly via the integral printer and can also send temperature data to an external PC. Communications with a PC is done by using Windows Hyperterminal software so no additional software is required. The OM-VLT can also send alarms to an external device via the built-in relay contact.
Specifications
Temperature Sensor: Solid state
IC, 10 m (33') long sensor probe wire, 0.2” diameter
Temperature Measurement Range: -40 to 130°C
(-40 to 266 °F)
Temperature Measurement Accuracy: ±1°C (±1.8°F) Temperature Writing and Storage Resolution:
0.1°C (0.2 °F) Sample Rate: Programmable between 1 and 90 minutes in 1 minute intervals Memory: Data logger saves the last 290 measurements; logging starts as soon as the unit is turned on Computer Interface: RS-232: 9600 baud, 8 data bits, parity none, 1 stop bit, flow control none LEDs: Power, low battery, out of paper, sensor, min, max
Monitors,
Logs, Prints and Displays Temperature Data vs Time
Menu Driven
Programming via Front Panel Keys
290 Historical
Temperature Readings – Data Can be Printed or Sent to PC on Request
Real Time
Temperature Printing – Three Printing Modes
Integral Temperature
Display – Max/Min on Request
Adjustable Hi/Lo
Alarm Thresholds and Alarm Delay
Local Audible/Visual
Alarm Indication
Alarm Relay
for Remote Alarming Applications
RS-232 Port for
Exporting Real Time Data to PC – No Software Needed
Low Battery, Chart
Ending and No Sensor Indicators
Extendable Remote
Temperature Sensor
OM-VLT, $255, shown smaller than actual size
E-2
E
Alarms: High and low temperature, end of paper, no sensor; audible (internal buzzer)/visual (alarm LED)/normally open relay contact
Audible Alarm (Internal Buzzer):
Some alarms trigger the internal alarm buzzer. The buzzer can be silenced by pressing EXIT on the front of the data logger or by connecting the optional remote buzzer silencer switch (OM-VLT­SK) which momentarily shorts the buzzer silencer contacts on the data logger. The alarm buzzer can also be enabled or disabled via the data logger front panel menu programming.
Temperature Alarm Operation:
If the measured temperature goes outside the maximum or minimum temperature limits for an amount of time equal to the defined alarm delay period (if any), an alarm is sent – the internal buzzer sounds (if enabled), an alarm is sent to an external device via the alarm relay contact (if enabled) and the alarm LED lights.
Disconnected Temperature Sensor Alarm Operation: If the
temperature sensor is disconnected the display shows the message nOS, the alarm LED lights and the internal buzzer sounds (if enabled).
To Order
(Specify Model Number)
Model No. Price Description
OM-VLT $255 Temperature data logger with printer OM-VLT-SENSOR 16 Spare temperature sensor, 10 m (33') OM-VLT-PAPER 4 Replacement paper roll, 40 m (131') OM-VLT-115-ADAPTOR 19 115 Vac/12 Vdc power adaptor OM-VLT-230-ADAPTOR 18 230 Vac/12 Vdc power adaptor OM-VLT-BAT 2 3.6 V lithium battery OM-VLT-BRKT 6 Wall mounting bracket OM-VLT-PACO 6 Paper compartment cover OM-VLT-BACO 6 Battery compartment cover OM-VLT-SK 55 Alarm buzzer remote silencer kit OM-VLT-CABLE10H 23 Temperature sensor extension cable, 3 m (10') OM-VLT-CABLE33H 28 Temperature sensor extension cable, 10 m (33') OM-VLT-PR 143 24 Vdc/ 12 Vdc power reducer OM-61-RS232-DB9 14 1.8 m (6') RS-232 cable DB9F termination
ALL MODELS AVAILABLE FOR FAST DELIVERY!
OM-VLT data loggers include 2 thermal chart paper rolls, wall mounting bracket, paper compartment cover, battery compartment cover,
3.6 V lithium battery, 115Vac/12 Vdc power adaptor, remote temperature sensor 10 m (33' long) and complete operator’s manual.
Ordering Example: OM-VLT temperature data logger with printer and OMEGACARE
SM
1 year extended warranty for OM-VLT
(adds 1 year to standard 1 year warranty), $255 + 25 = $280.
Download data using Windows Hyperterminal - no extra software is required.
End of Paper Alarm Operation:
If the printer paper runs out, the internal buzzer sounds (if enabled), an alarm is sent to an external device via the alarm relay contact (if enabled) and the out of paper LED lights.
Alarm Relay: 1 A, 12 Vac/Vdc Alarm Delay Range: User
selectable; No delay, 10 min, 30 min, 1 hr, 2 hrs Printing Modes: There are three printing modes; full text report (small font), time and temperature (large font), graph format
Chart Speed Accuracy: ±1% Time Accuracy: ±80 ppm
@ 75 °F (24°C)
Thermal Chart Paper Dimensions: 2" wide x 131' long Display: 4 character numeric LED Power: 12 Vac/Vdc via included
115 Vac 50/60 Hz adaptor Battery Backup: 9 V alkaline for up to 60 hours operation Clock Battery: CR2032 coin cell, used for saving logged data and configuration settings and for running the real-time clock
Operating Environment:
-18 to 65°C (0 to 150 °F), 0 to 95% RH non-condensing
Storage Temperature:
-40 to 65°C (-40 to 150 °F) Mounting: Vertical or horizontal, free standing, wall or panel mount (wall mount bracket included)
Dimensions:
209 W x 135 H x 69 mm D (8.2 x 5.3 x 2.7") Weight: 1.3 Kg (2.9 lbs) including chart paper