OM-PQR1010 Power Line and Environmental Monitor
Extended Warranty
Program
SM
⻬ Measures All Types of
Disturbances Including
Spikes, Sags, Surges,
Common Mode Noise,
High Frequency Noise,
Dropouts, Power Failures
⻬ Datalogging - Stores
Average Reading Every
Minute for Up to 11 Days
on Each Channel
⻬ 1500 Event Storage Per
Input, 9000 Events Total
for 6 Channels
⻬ Records Magnitude,
Time and Date of Each
Disturbance
Model OM-PQR1010, shown with
included temperature probe, AC
voltage test leads, DC voltage
test leads, RS-232 cable, power
cord and Windows software CD.
Shown
smaller than
actual size.
Electronic equipment is damaged
more often by power disturbances
than by fire, theft and vandalism
combined. Intermittent power
problems are the most expensive
hidden expense to users of
microprocessor-based equipment.
The OM-PQR1010 power quality
recorder is a state-of-the-art,
fully integrated instrument that
measures, records and reports
power disturbances, aiding in
the analysis of power quality
in medical, commercial and
industrial applications.
The OM-PQR1010 monitor unit
will detect a wide range of power
disturbances including spikes,
sags, surges, common mode noise,
dropouts and high frequency noise
on two phases of AC voltage and
one DC voltage channel; it will also
test the temperature, humidity and
ac current. For ac current
measurement, the OM-PQR1010
is equipped with a pair of banana
jacks designed to accept a 10 mV/A
output current transducer (CT).
Disturbances detected on multiple
channels are recorded by their
time, date, magnitude and duration
in a non-volatile RAM memory.
Stored data can then downloaded
and analyzed by connecting the
OM-PQR1010 to your PC’s serial port
and running the included
OM-PQR1010 software
Windows data analysis software.
Input connections to the
OM-PQR1010 are made between
the safety connectors on the back
of the unit and the circuit panel
to be tested. The OM-PQR1010
operates on standard 110 Vac/220
Vac power. Once plugged in, the
OM-PQR1010 immediately begins
testing the signals on the input
connectors.
Specifications
GENERAL
Summary Event Count:
56,000 of each event type
Detail Storage:
1500 events per channel
OM-PQR1010, shown
smaller than actual
size
Data Retention:
5 years without power
Datalogging Sample Rate:
Programmable from 1 reading per
second to 1 reading per 4 minutes
Impulse Detection: 500 nanoseconds
Communications: RS-232, 1200 to
19,200 baud (programmable),
8 data bits, 1 stop bit, no parity
Input Connections: ac voltage,
safety banana jacks; dc voltage,
BNC; current (CT), banana jacks;
temperature, phone jack; humidity,
mini-DIN
Operating Temperature: 0 to 50
(32 to 122°F), 10 to 80% RH
non-condensing
o
C
E-71
Operating Voltage: 12 to 30 Vdc,
80 to 260 Vac
Operating Current: 0.1 A
Operating Line Frequency:
DC to 400 Hz
Weight: 1.8 kg (4 lb)
Dimensions:
203 mm H x 254 mm W x 51 mm D
(8 x 10 x 2")
Power Cord: 1.8 m (6')
AC/DC VOLTAGE CHANNELS
DATA LOGGING
Range: ac voltage, 80 to 520 Vac
RMS; dc voltage, +1 to +52 Vdc
Accuracy: +1.5%
SAGS/SURGES
Threshold: 5% and 10%
of average, or user
programmed levels
Duration Limits: 1 cycle or 20 ms
Accuracy: +1.5%
DROPOUTS
Threshold: AC voltage, less than
10 V RMS; DC voltage, less than
10% of avg
Duration Limits: longer than 8 ms,
shorter than 80 ms
POWER FAILURES
Threshold: ac voltage, less than
10 V RMS; dc voltage, less than
10% of avg
Duration Limits: Longer than
80 ms
IMPULSES
Resolution: 2 channels:
20 V to > 2500 V peak
Threshold: 20, 70, 140, 300, 850,
2.5 kV
Duration Limits: 500 nanoseconds
above threshold
Accuracy: +10%
HIGH FREQUENCY NOISE
Range: 2 volts peak, 10 kHz
to 10 MHz
Accuracy: +10%
Response Time: 1 ms
AC VOLTAGE LINE FREQUENCY
Range: 40 to 400 Hz
Accuracy: +1%
Response Time: 1 AC cycle
Threshold: +2% deviation
from average
HUMIDITY CHANNEL
DATA LOGGING
Range: 10 to 90%
RH non-condensing
Accuracy: +5%
SAGS/SURGES
Threshold: 5% and 10%
of average, or user
programmed levels
Duration Limits: 2 seconds
Accuracy: +5%
CURRENT CHANNEL
DATA LOGGING
Input: Current transducer
(CT), 10 mV/A output
Range: 0.2 A to 1000 A RMS
Accuracy: +1.5%
SAGS/SURGES
Threshold: 5% and 10%
of average, or user
programmed levels
Duration Limits: 20 ms
Accuracy: +1.5%
DROPOUTS
Threshold: Less than 10%
of average
Duration Limits: Longer than
8 ms, shorter than 80 ms
POWER FAILURES
Threshold: Less than 10%
of average
Duration Limits:
Longer than 80 ms
TEMPERATURE CHANNEL
DATALOGGING
Sensor: Semiconductor transistor
Range: -7 to 121°C (20 to 250°F)
(0.02 to 2.5 Vdc)
Accuracy: +1%
Sample Rate: Programmable from
OM-PQR1010,
shown smaller
than actual size
1 reading per second to 1 reading
per 4 minutes
SAGS/SURGES
Threshold: 5 and 10% of average,
or user programmed levels
Duration Limits: 2 seconds
Accuracy:+1.5%
OM-PQR1010
software
AVAILABLE FOR FAST DELIVERY!
To Order
(Specify Model Number)
Model No. Description
OM-PQR1010 Power line and environmental monitor
OM-PQR-HUMIDITY Humidity probe for OM-PQR1010
Each OM-PQR1010 is supplied with 1.8 m (6') power cord, RS-232 communications cable with
DB9F termination, Windows 2000/XP data analysis software, temperature probe, one set of ac
voltage test leads and one set of dc voltage test leads. Humidity probe must be ordered
separately.
Ordering Example: OM-PQR1010 power line and environmental monitor
E-72