Omega Products FPD1000-HP Installation Manual

FPD1000-HP SERIES
HIGH PRESSURE
Positive Displacement Flowmeters
To the owner:
Thank you for purchasing an OMEGA FPD Series Flowmeter. Please take a few min­utes to read through the manual before in­stalling and operating your meter. If you have any problems with the meter, refer to the Maintenance and Troubleshooting sec­tions of the manual.
This manual contains connection and op­erating instructions for the OMEGA FPD Series meters with pulse outputs. This in­cludes the following models:
FPD1202-HP FPD1203-HP
The OMEGA FPD Series flowmeter has in­corporated the oval rotor principal into its design. This has proven to be a reliable and highly accurate method of measuring flow. Exceptional repeatability and high accuracy over a wide range of fluid viscosities and flowrates are features of the OMEGA FPD Series flowmeter design. With low pressure drop and high pressure rating, means the OMEGA FPD Series flowmeters are suit­able for both gravity and pump (in-line) ap­plications.
INSTALLATION
1. Use thread sealant on all pipe threads.
2. Ensure the meter is installed so that rotor shafts are always in a horizontal plane (See Figure 1). Flow is bi-direc­tional.
Figure 1
DO NOT install meter this way.
3. OMEGA recommends use of flexible connections.
4. Extreme care must be taken when in­stalling the meter. Pipe strain or over­tightening meter connections can cause meter damage.
PULSER DETAILS
IMPORTANT INFORMATION
Please read this information
!
carefully before use!
Before use, confirm the fluid to be used is compatible with the meter or consult with OMEGA for advice.
To prevent damage from dirt or foreign matter, OMEGA recommends a Y or bas­ket type 200 mesh strainer be installed as close as possible to the inlet side of the meter. (If required, contact OMEGA for fur­ther information.)
To prevent damage to the meter, slowly fill the system with fluid. This will prevent dam­age caused by air purge.
NOTE: Failure to do this could damage the
meter.
For pump applications, turn off the pump at the end of each day.
Hall Effect Sensor Specifications
•4.5 V to 24 V (4.6 - 9 mA) operation needs only an unregulated supply.
•Open collector 25 mA output NPN (Current Sink) compatible with digital logic.
•Reverse battery precaution.
Figure 2
Supply
Signal
Ground
12 3
Hall Effect Sensor Wiring Details
V
cc
(HP Models Only)
X
1
Reed Relay Specifications
•Two wire SPST N/O.
•Switching voltage 150 VDC; maximum current 0.25 Amps.
•Rating 3 watts.
•Duty cycle 20% ON, 80% OFF.
Figure 3
12
Reed Switch Wiring Details
(HP Models Only)
MAINTENANCE
Disassembly:
1. Ensure the fluid supply to the meter has been disconnected, and the line pres­sure has been released before disas­sembly.
2. Remove the four screws and remove the pulser cap.
3. Remove the gasket.
4. Remove eight screws and remove the meter cap.
5. Remove O-ring and inspect. Replace O­ring if damaged.
6. Remove rotors, clean and inspect. Re­place rotors if damaged.
7. To remove the PCB, remove the two screws.
NOTE: Reed Switch PCB’s cannot be re-
moved.
Reassembly:
1. Replace rotors into the meter body. The rotors should be at 90 degrees to each other.
2. Lightly rotate the rotors by hand. They must rotate freely.
3. Install O-ring.
4. Replace the meter cap and tighten the eight screws uniformly to 35Nm (25 Ft. Lbs.).
5. Replace the pulser cap and tighten the four screws.
2
Loading...
+ 8 hidden pages