Omega Products FP-5070 Installation Manual

Page 1
User’s Guide
Shop online at
www.omega.com
e-mail: info@omega.com
FP-5070 Series Flow Sensors
Page 2
OMEGAnet® Online Service Internet e-mail
www.omega.com info@omega.com
Servicing North America:
USA: One Omega Drive, P.O. Box 4047
ISO 9001 Certified Stamford CT 06907-0047
TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
Customer Service: 1-800-622-2378 / 1-800-622-BEST Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
®
®
®
Mexico: En Español: (001) 203-359-7803 e-mail: espanol@omega.com
FAX: (001) 203-359-7807 info@omega.com.mx
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
TEL: +31 (0)20 3472121 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: nl@omega.com
Czech Republic: Rudé armády 1868, 733 01 Karviná 8
TEL: +420 (0)69 6311899 FAX: +420 (0)69 6311114 Toll Free: 0800-1-66342 e-mail: czech@omega.com
France: 9, rue Denis Papin, 78190 Trappes
TEL: +33 (0)130 621 400 FAX: +33 (0)130 699 120 Toll Free in France: 0800-4-06342 e-mail: france@omega.com
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
TEL: +49 (0)7059 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: germany@omega.com
United Kingdom: One Omega Drive, River Bend Technology Centre
ISO 9002 Certified Northbank, Irlam, Manchester
M44 5EX United Kingdom TEL: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-488 e-mail: sales@omega.co.uk
Page 2
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient-connected applications.
OMEGA FP-5070 Series Flow Sensors
Page 3
OMEGA FP-5070 Series Flow Sensors
SAFETY INSTRUCTIONS
1. Do not remove from pressurized lines.
2. Confirm chemical compatibility before use.
3. Do not exceed maximum temperature/pressure specifications.
4. Wear safety goggles or faceshield during installation/service.
5. Do not alter product construction.
6. Failure to follow safety instructions could result in severe personal injury.
1. Description
The FP-5070 series Mini Flow Sensors contain a free-running rotor which is driven by the fluid flow. Within the given measurement range, the rotational speed of the rotor is proportional to the fluid flow rate. Permanent magnets built into the rotor actuate an electronic switch in the top of the sensor generating a square-wave output signal proportional to flow rate. Both opaque and transparent fluids can be measured from 0.2 to
20.0 centistokes.
Wetted sensor parts are constructed of PVDF and FPM, making the sensor suitable for use with most process fluids, including most acids, bases, light oils, and solvents.
2. Specifications
General
Flow Range:
• -2V sensor: 400 to 2800 mL/m (0.105 to 0.740 U.S. gpm)
• -3V sensor: 700 to 4200 mL/m (0.185 to 1.123 U.S. gpm)
• -4V sensor: 1300 to 6000 mL/m (0.343 to 1.585 U.S. gpm)
• -6V sensor: 3200 to 12000 mL/m (0.845 to 3.170 U.S. gpm) Linearity: ±0.25% of full range Repeatability: ±0.25% of full range Viscosity range: 0.2 to 20.0 centistokes Pipe connections: G 1/4 in. ports, 1/4 in. NPT (male) pipe adapters
(2 included) Cable length: Std: 7.6m (25 ft.), max.: 300 m (1000 ft.) Cable type: 2-conductor shielded, twisted-pair, 22 AWG Shipping Weight: 0.4 kg (0.8 lb.)
Pressure Drop:
Pressure Drop Across Sensor vs. Flow Rate
PSI
BAR
1.50
21.76
18.13
14.50
10.88
7.25
3.63
1.25
1.00
0.75
0.50
0.25
02040607080
Flow rate (%) of full scale
-2V
-3V
-4V
-6V
Cable plug
Upper body with electronic switch
25 ft. cable
FLOW
1
/4 in.
G port threads
Lower body with
built-in rotor
Pipe fitting adapter
1
/4 x 1/4 in. NPT
G (2 included)
WARNING!
Polar organic solvents (i.e., ketones and chlorinated hydrocarbons) and aromatic hydrocarbons are not compatible with this sensor.
Wetted Materials
• Housing: PVDF
• Flow insert: PTFE
• Quad ring seal: FPM
• Rotor: PVDF
• Pipe thread adapters: PVDF
• Suitable for clean fluids only
Electrical
Power: 5 to 24 VDC @ 10 mA max. Output type: Open-collector transistor, 10 mA max.
sink
Max. pressure/temperature:
• 5.5 bar @ -30°C (80 psi @ -22°F)
• 5.5 bar @ 24°C (80 psi @ 75°F)
•3 bar @ 120°C (45 psi @ 248°F)
psibar 100
6.9
80
5.5
60
4.
1
40
2.8
20
1.4
0
-40 0 40 80 120 160 200 240 280
-40 -18 4 27 49 71 93 116 138
°F
°C
OMEGA FP-5070 Series Flow Sensors
Page 3
Page 4
Dimensions:
95 mm (3.74 in.)
FLOW
Mounting tabs (3), #8 or M4 self-tapping screws required (customer supplied)
32.5 mm (1.28 in.) bolt circle
68 mm (2.7 in.)
65 mm (2.6 in.)
3. Installation
The sensor may be installed in any position, although horizontal flow is recommended (sensor mounted upright). If the sensor is not installed upright, the linearity error may be greater in the lower part of the sensor's measurement range.
Mounting tabs are provided using #8 or M4 self-tapping screws (customer supplied). See Dimensions illustration for mounting tab hole pattern specifications.
Install sensor with the arrow pointing in the direction of flow.
Always maximize distance between the sensor and pump source. Never install immediately downstream of valves,
Pipe Fitting
Adapters (included)
FLOW
FLOW
100-150 mm (4-6 in.)
Mounting
tabs (3)
Compatible pipe/tubing connections (customer supplied):
Female SxT coupling
Female TxT coupling
Female TxT
coupling
Installation Hints
Avoid vibrations and shocks
Avoid solids in the fluid
Install a filter or line strainer upstream to protect sensor
Page 4
Hand
tighten
only!
Coupling
Pipe or tubing
100-150 mm (4-6 in.)
Female TxT union
Hose adapter
fittings, etc. For optimum performance, a straight flow run of at least 100 to 150 mm (4 to 6 in.) should be provided before and after the sensor.
Two pipe fitting adapters (included) convert the G 1/4 in. straight threads to 1/4 in. NPT pipe threads. Hand tighten only! Apply 1-2 turns of sealing tape to all threaded connections to prevent leaks.
CAUTION!
Use an adjustable wrench to prevent the fitting adapters from overtightning while installing mating pipe connectors. Sensor damage will occur if the ports are overtightened.
4. Wiring Details
4.1 Cable Extensions
The standard 25 foot sensor cable can be extended to 300 m (1000 ft.) using 2-conductor shielded twisted-pair cable.
Always maintain cable shield through cable splice.
For splice-free cable replacement up to 300 m (1000 ft.), refer to the sensor plug connection diagram (below) for connection details.
Sensor Signal Output (Red wire)
Sensor Power:
5 to 24 VDC
(Black wire)
1
Sensor ground
2
(Silver wire)
Disregard ground symbol on connector
Max cable length:
300 m (1000 ft.)
OMEGA FP-5070 Series Flow Sensors
Sensor Plug (side)
Page 5
4.2 Instrument Connections
A
C
FP-5072
FLOW
FP-5072
FLOW
OMEGA
FPM-5500/FPM-9020A
Shield
Black Red
Freq. IN
Freq. IN
Std. Sensor
Iso. Gnd
Sen. Pwr.
Open Collector Sensor
OMEGA
FPM-9010A
SENSOR
BLK RED SHLD
Black Red
Shield
12 13 14
B
OPEN
COLLECTOR
SENSOR
STANDARD
SENSOR
D
OMEGA
FP90 Series
Frequency In
Sensor Power
Gnd
Frequency In
FP-5072
FLOW
Red
Black
Shield
Black
Shield
Red
Other Brands
10 kΩ
FP-5072
FLOW
External Power supply
5 to 24
+
Gnd.
Input
-
VDC
Other
instrument
• Pull-up resistor required (10 kΩ recommended). Consult your instrument manual for
• Configure instrument input card for FP-5072 sensor input
 additional information.
5. Calibration
The K-Factors listed below represent the number of pulses the sensor will generate for each measured engineering unit. They are listed in U.S. gallons, liters, and mL by sensor model.
-------------------------------- K-FACTORS ----------------------------------
Sensor Flow PULSES PER PULSES PER PULSES PER Model Insert U.S. GAL LITER mL
FP-5072-PV 2 mm 5685 1502 1.502 FP-5073-PV 3 mm 3308 874 0.874 FP-5074-PV 4 mm 2316 612 0.612 FP-5076-PV None 1249 330 0.330
6. Replacement Parts
Upper body
Cable plug, #3-2507.080-5
OMEGA FP-5070 Series Flow Sensors
Rotor, #3-2507.080-2
Flow inserts: 2 mm: #3-2507.081-2 3 mm: #3-2507.081-3 4 mm: #3-2507.081-4
Quad ring seal, 3-2507.080-3
FLOW
Lower body, FP-5074-PV
Page 5
Page 6
7. Replacing The Flow Insert
Sensor range can be modified by changing the flow insert. The sensor must be removed from service and disassembled prior to installing the new flow insert. See section 2 specifications for flow range data.
Flow Insert Replacement Procedure:
1. Depressurize system and remove sensor.
2. Rotate the upper sensor body clockwise until it releases from the lower half, then lift off.
3. Remove rotor and quad ring seal from lower body.
4. Push the flow insert outward using a small screwdriver.
5. Install the new flow insert (small diameter inward) with the eraser end of a pencil. Apply light pressure until insert seats against the step in the lower body. Do not force!
6. Install rotor into lower body. Spin rotor with finger and check for free rotation. If rotor hits flow insert, remove rotor and push insert back until free rotor rotation is established. Use a rounded object like a pen or pencil body to adjust flow insert depth.
7. Install rotor, quad ring, and upper body. Hand tighten only! Do not overtighten upper body or the lower body assembly tabs will break.
8. Troubleshooting
8. Reprogram instrument with new K-Factor, see calibration section 5.
WARNING!
Do not use tools of any kind on the sensor body or port connections. Hand tighten only! Excessive force will damage sensor.
Upper body
Assemble
(counterclockwise)
flow
insert
Inlet port
Rotor
FLOW
Lower Body
Disassemble
(clockwise)
Quad ring seal
Assembly Tabs (3)
Outlet port
Condition Recommendation
Erratic or missing sensor signal
A) Verify ALL cable and instrument connections (section 4). B) Verify proper sensor installation (section 3). C) Remove power from instrument and disconnect sensor inputs. Power up instrument and check across
Black and Shield terminals with a digital voltage meter for 5 VDC. If 5 VDC is not present, the instrument requires service or may be misconfigured. OMEGA FPM 9010A
controllers require input card configuration for FP-5072 sensor inputs, see instrument manual. Perform steps A-E If sensor problems persist, contact OMEGA.
D) Verify the FP-5072 paddlewheel is spinning freely by blowing into the flow chamber. If the paddlewheel
dosen’t spin freely, the following conditions may exist:
• The sensor may be dirty or clogged. Disassemble and clean with hot tap water and soft brush (see section 7).
• The rotor may be hitting the flow insert. Disassemble and adjust flow insert depth (see section 7).
E) Test sensor with flow system active and sensor powered. Use an oscilloscope to check the sensor input
signal across the Red (Signal IN) and Shield terminals. A square wave signal should appear at these terminals. If no signal is present replace sensor.
9. Ordering Information
Mfr. Part No. Description
FP-5072-PV Mini-Flow Sensor, 2mm insert FP-5073-PV Mini-Flow Sensor, 3mm insert FP-5074-PV Mini-Flow Sensor, 4mm insert FP-5076-PV Mini-Flow Sensor, 6mm inlet, no insert
Page 6
OMEGA FP-5070 Series Flow Sensors
Page 7
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR
WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the
product was PURCHASED,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2000 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of
the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA FP-5070 Series Flow Sensors
Page 7
Page 8
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at www.omega.com
TEMPERATURE

Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies

Wire: Thermocouple, RTD & Thermistor

Calibrators & Ice Point References

Recorders, Controllers & Process Monitors

Infrared Pyrometers
PRESSURE, STRAIN AND FORCE

Transducers & Strain Gages

Load Cells & Pressure Gages

Displacement Transducers

Instrumentation & Accessories
FLOW/LEVEL

Rotameters, Gas Mass Flowmeters & Flow Computers

Air Velocity Indicators

Turbine/Paddlewheel Systems

Totalizers & Batch Controllers
pH/CONDUCTIVITY

pH Electrodes, Testers & Accessories

Benchtop/Laboratory Meters

Controllers, Calibrators, Simulators & Pumps

Industrial pH & Conductivity Equipment
DAT A ACQUISITION

Data Acquisition & Engineering Software

Communications-Based Acquisition Systems

Plug-in Cards for Apple, IBM & Compatibles

Datalogging Systems

Recorders, Printers & Plotters
HEATERS

Heating Cable

Cartridge & Strip Heaters

Immersion & Band Heaters

Flexible Heaters

Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL

Metering & Control Instrumentation

Refractometers

Pumps & Tubing

Air, Soil & Water Monitors

Industrial Water & Wastewater Treatment

pH, Conductivity & Dissolved Oxygen Instruments
6-2507.090-OM/(J-10/02) M-2972/1002
Loading...