Omega Products FP1402-RTD User Manual

DIGITAL PLASTIC PADDLEWHEEL FLOWMETERS
FP1400 Series
U Polypropylene or
PVDF Models
U 29 Engineering Units
U Two Programmable Totalizers U Programmable Alarms U Isolated Analog 0 to 5 Vdc or
4 to 20 mA
U RS232 Communications
Standard
U Local Key Pad and 2 x 16
Characters LCD Display
U Free Communication Software
with Data Logging Capability
F
FP1400 Series flowmeters support various functions including; two independently programmable flow totalizers, user programmable low, high or range flow and temperature alarm, two sets of user programmable optically isolated outputs, self diagnostic alarm, and flow pulse output. The flow rate can be displayed in 29 different volumetric or mass flow engineering units. Flowmeter parameters and functions can be programmed locally via optional key pad and LCD or remotely via the RS232/RS485 interface. Local 2 x 16 LCD readout with adjustable backlight provides flow rate, temperature, total volume reading in currently selected engineering units, diagnostic events indication and feature a password protected access to the process parameters to ensure against tampering or resetting.
Liquid flowing through the unit causes the paddlewheel to spin. As the magnets embedded in the paddle spin past the sensor, electrical pulses are produced in which frequency is proportional to the flow rate. The number of pulses per desired time interval and the K-factor (number of pulses/gal) make it is possible to calculate the flow rate and volume passing through the unit. On board CPU and signal conditioner circuitry perform accurate flow and total computation, digital communication and analog 0 to 5 Vdc or 4 to 20 mA output signals. Non-volatile memory stores all hardware specific and user programmable variables, including flow linearization table.
FP1408 shown larger
than actual size.
SPECIFICATIONS
Viscosity: 1 cSt (water) can be used for liquids up to 50 cSt
with field calibration (max flow range may be affected)
Flow Accuracy (Including Linearity): ±1% of FS Repeatability: ±0.25% of FS Temperature Measurement Range: 60°C (140°F) Temperature Accuracy (Including Linearity): ±0.5°C Flow Response Time: Approx 1 sec (above 10% of full scale
flow), approx 2 seconds (below 10% of full scale flow)
Maximum Pressure: 10 bar (150 psig) Maximum Pressure Drop: @ 15 psi for sized units @ 20 psi
all other sizes
Ambient Temperature: -10 to 60°C (14 to 140°F) Output Signals:
Linear 0 to 5 Vdc (3000 Ω minimum load impedance); linear 4 to 20 mA (500 Ω maximum loop resistance). Maximum noise 20 mV peak to peak (for 0 to 5 Vdc output)
Flow Pulse Output: 3.3 Vdc amplitude (3000 Ω min load impedance)
Optically Isolated Outputs: UCE 40 Vdc, ICE 150 mA Power: 11 to 26 Vdc, 100 mV maximum peak to peak
output noise Consumption: +12 Vdc (150 mA maximum); +24 Vdc (100
mA maximum); circuit board have built-in polarity reversal protection, 300 mA resettable fuse provide power input protection
Communications: RS232 standard, RS485 (optional)
F-1
Also available on remote display. FPB1400 Series with DPF701 meter. Visit omega.com/fpb1400
FPB1406
Both models shown
smaller than actual size.
PVDF Models
Body: PVDF LID: PVDF Paddle Wheel: PVDF Shaft: Zirconia ceramic Bearings: Sapphire jewels O-RINGS: PTFE RTD: 316 SS casing
Electrical Connections: Built-in
female 12-pin M16, NEMA 4X (IP67) connector. To be mated with 12-pin M16 male EMI shielded NEMA 4X (IP67) connector (sold separately)
Display: Optional local 2 x 16 characters LCD with adjustable backlight (2 lines of text)
Keypad: Optional 4 pushbutton key pad CE Compliant: EMC Compliance with
89/336/EEC as amended. Emission Standard: EN 55011:1991, Group 1, Class A Immunity Standard: EN 55082-1:1992
To Order
Visit omega.com/fp1400 for Pricing and Details
DPF701 Visit omega.com/dpf701
Wetted Materials Polypropylene Models
Body: Polypropylene LID: Acrylic Paddlewheel: PVDF Shaft: Nickel tungsten carbide Bearings: Sapphire jewels
O-Rings: EPDM RTD: 316 SS casing
Maximum Model No. Ports Pressure Drop Range Model No. Description with RTD NPT bar (psi) GPM
FP1402 Polypropylene flowmeter, with display FP1402-RTD ³⁄₈ 1 (15) 0.5 to 5 FP1404 Polypropylene flowmeter, with display FP1404-RTD ¹⁄₂ 1.4 (20) 1 to 10 FP1406 Polypropylene flowmeter, with display FP1406-RTD ³⁄₄ 1.4 (20) 2 to 20 FP1408 Polypropylene flowmeter, with display FP1408-RTD 1 1.4 (20) 3.5 to 35 FP1412 PVDF flowmeter, with display FP1412-RTD ³⁄₈ 1 (15) 0.5 to 5 FP1414 PVDF flowmeter, with display FP1414-RTD ¹⁄₂ 1.4 (20) 1 to 10 FP1416 PVDF flowmeter, with display FP1416-RTD ³⁄₄ 1.4 (20) 2 to 20 FP1418 PVDF flowmeter, with display FP1418-RTD 1 1.4 (20) 3.5 to 35
Accessories
DPF701 ¹⁄₈ DIN digital display, 115 Vac 7.5 to 13 Vdc powered FP1400-CABLE 12-pin cable
Comes complete with operator’s manual. For units with RS485 communications in place of RS232 add suffix “-485” to the model number, no additional charge.
Ordering Example: FP1404, polypropylene flowmeter with display 1 to 10 GPM range,
1
NPT connections.
2
F-2