Omega Products FMI-100 User Manual

PROGRAMMABLE INTERFACE DISPLAY FOR RATE, TOTAL AND CONTROL COMMANDS
FMI-100
U Displays Instantaneous
and Total Flow
U Digital RS232 or RS485
Interface (Multi-Drop Up to 64 Devices)
U Programmable, Optically-
U For Use with
FMA1700/1800 or FMA5400/5500 Series
U 16 Point Programmable
Controls, With Ramping Up/Down
U Free Configuration
and Monitoring Utility Software
FMI-102D with FMA1702
Both models
shown smaller
than actual size.
FMI-102D with FMA5404
D
The FMI-100 Series is for flowmeters and controllers with analog 0 to 5, 5 to 10, 0 to 10 Vdc or 4 to 20 mA input output interface, where flow indications/control, totalizers or alarm functions are required. Also when re-transmission of the flow rate and/or totalizer functions via optically-isolated pulse output or serial communication is desired. Local or programmable set point control for flow controllers (no host PC presence required).
The graphical LCD has large 13 mm (0.51") digits for flow rate and 5.5 mm (0.21") for total and can be set by user to simultaneously show different combination of the flow parameters: flow rate, totalizers, flow alarms, and diagnostic events. All configuration parameter settings are easily accessed via a simple user-interface menu driven by a 6 button key-pad which can be password-protected.
For flowmeters and/or flow controllers, FMI-100 provides jumpers selectable for 0 to 5 Vdc or 4 to 20 mA analog set point control signals. The flow rate set point can
be adjusted locally via key-pad, remotely via host PC using digital communication interface, or programmed in advance using built-in 16 steps batch table with ramping up/down support.
The programmable flow pulse output operates independently from the totalizers and based on configuration settings can provide pulse frequency proportional to instantaneous fluid flow rate. The FMI-100 provides two independent programmable flow totalizers. Both totalizers are updated every 100 ms and can be set to activate different events. Main totalizer accumulated total is backed-up
D-1
in EEPROM memory every second. And it provides the user with a flexible alarm/warning system that monitors the fluid flow for conditions that fall outside configurable limits as well as visual feedback for the user via the LCD or via an optically-isolated output.
All process data and settings can be read and modified manually via local LCD key-pad or through the digital RS232 or RS485 communication interface. Proprietary ASCII software interface command set and free Communication Utility software are provided.
SPECIFICATIONS
ADC/DAC Resolution: 12-bit Analog Inputs: 0 to 5 Vdc, 4 to 20 mA, 5 to 10 Vdc
(jumper-selectable), 0 to 10 Vdc (contact Flow Engineers at Omega for details)
Analog Outputs: 0 to 5 Vdc, 4 to 20 mA (jumper-selectable) LCD: 128 x 64 graphic LCD with instantaneous flow reading
and total volume indication, adjustable LCD contrast and back light
Key-Pad: Local 6 tactical push buttons Pulse Output: User-programmable, optically-isolated,
with preset active low time interval (10 to 6550 mS). Digital Output: Two programmable, optically-isolated, UCE @ 40 Vdc, ICE @ 150 mA (voltage isolation; 250 Vrms) Digital Interface: RS232 or RS485 (multidrop capability up to 64 devices)
Protocol: Proprietary ASCII software interface command set Speed: 1200/2400/4800/9600/19200/38400/57600/115200
baud (user selectable)
Configuration: Stop Bit: 1 Data Bits: 8 Parity: None Flow Control: None Addressing: Maximum 255 addresses (for RS485 option only)
Type: RS232 or RS485 2-wire Power Requirements: 12 to 26 Vdc (up to 60 mA maximum) Interface Connectors: Process I/O signals and digital
RS232/RS485 interface, miniature 9-pin female D-SUB connector; digital optically-isolated outputs; TERM BLOCK HEADER 4POS 3.5 mm male pins, shrouded Environment: Installation Level II, Pollution Degree II, (per IEC 664) Electromagnetic Compatibility: Compliant ref. 89/336/EEC as amended; emission standard; EN 55011:1991, Group 1, Class A immunity standard: EN 55082- 1:1992
Operating Temperature: -10 to 70°C (14 to 158°F) Dimensions: 86.4 W x 76.2 H x 19.1 mm D
(3.4 x 3.0 x 0.75") Weight: Approximately 125 g (0.3 lb)
To Order
Model No. Description FMI-102 RS232 monitor/control interface, no display FMI-105 RS485 monitor/control interface, no display FMI-102-10V RS232 monitor/control interface, no display for 0 to 10V input/output FMI-105-10V RS485 monitor/control interface, no display for 0 to 10V input/output FMI-102D Rate/total, display, RS232 monitor/control interface FMI-105D Rate/total, display, RS485 monitor/control interface FMI-102D-10V Rate/total, display, 0 to 10V input/output monitor/control interface FMI-105D-10V Rate/total, display, 0 to 10V input/output monitor/control interface
Comes complete with operator’s manual.
Mounting Kits (Sold Separately) Model No. Description FMI-100-MKM-DD Meter mounting kit two 9-pin D connectors FMI-100-MKM-RD Meter mounting kit RJ11 to 9-pin D FMI-100-MKM-FD Meter mounting kit flat wire to 9-pin D FMI-100-MKC-2C Controller mounting kit 12 Vdc North American with communication FMI-100-MKC-2N Controller mounting kit 12 Vdc North American no communication FMI-100-MKC-4C Controller mounting kit 24 Vdc North American with communication FMI-100-MKC-4N Controller mounting kit 24 Vdc North American no communication
For units with 230 Vac European plug add “-EU” to the model number, for additional cost. For units with 240 Vac Austrialian plug add “-AU” to the model number, for additional cost. For units with 240 Vac UK plug add “-UK” to the model number, for additional cost. Ordering Example: FMI-105, RS485 interface (no display) with FMI-100-MKM-RD meter mounting kit with cable.
Visit omega.com/fmi-100 for Pricing and Details
D-2