Omega Products DP9800-RTD User Manual

8-Channel Benchtop Digital Thermometer/ Data Logger
For Thermocouple or RTD Sensors with USB Port
DP9800 Series
U High Precision U USB PC Interface U Thermocouple and
RTD Models
U Built-In Display for
Selected Channel or Scanned Channels
U Resolution 0.1°C on
LCD Display, 0.01°C in Software
U Self-Calibration Feature U Select °C/°F U PC Software Included
for Remote Control and Measure, Logging and Calibration
U Simple Operation
The DP9800 Series thermometer can be used in conjunction with a PC to provide accurate, versatile 8 channel thermocouple and Pt100 RTD temperature measure­ments, scanning and logging of measured values. It can also be used as a “stand alone” indicator/ logger and incorporates a digital backlit LCD display of measured temperature.
The built-in self-calibration facility for the DP9800-TC thermocouple model is a rapid and convenient method for on-site calibration and does not require any additional equipment other than a special, external link. Self-calibration of Pt100 ranges is equally simple and uses plug-in precision resistors.
The DP9800 is designed to provide exceptional stability with high measurement resolution and represents an ideal crossover between plant practicality and laboratory performance at a very competitive price.
The PC software supplied with the instrument allows control, measure and calibration functions.
Specifications
At an ambient temperature of 20°C Input/Ranges: Thermocouple to
IEC 584 Type J: -200 to 750°C (-328 to 1382°F) Type K: -200 to 1200°C (-328 to 2192°F) Type T: -200 to 350°C (-328 to 662°F) Type E: -200 to 900°C (-328 to 1652°F) Type N: 0 to 1300°C (32 to 2372°F) Type R: 0 to 1760°C (32 to 3200°F) Type S: 0 to 1760°C (32 to 3200°F) Type B: 300 to 1800°C (572 to 3272°F) Pt100 to IEC751, 3-Wire:
-200 to 850°C (-328 to 1562°F)
Note: all inputs are non-isolated and thermocouples must be ungrounded style.
Accuracy: Thermocouples J, K, T, E, and N, better than ±0.1°C ±0.1% of range zero to span ±0.15% of range
-100 to - 200°C (J, K, T, and E) Thermocouples: R, S, and B, better
than ±0.1°C ±0.15% of range
Linearization: ±0.05°C Pt100 Range: Better than ±0.05°C
±0.1% of range
Zero Drift: ±0.01% of span per °C
M-1
DP9800 shown smaller than actual size.
Span Drift: ±0.01% of span per °C Display: LCD, backlight Display Resolution: Thermocouple
ranges 0.1°C, Pt100 range 0.01°C User Interface: Front panel key
for channel number or auto-scan selection; PC software for all other functions
Indication: Channel number, measured temperature (°C or °F)
Reference Junction: Automatic, accurate reference
Self Calibration: User facility incorporated; the instrument auto-calibrates on every A/D cycle
Sensor: Open circuit upscale indication, detection and indication (independent alarms should be used for process safety if required)
Ambient Operating Temperature: 0 to 50°C (32 to 122°F)
Storage Temperature: -20 to 70°C (-4 to 158°F)
Input Terminations: 8 x thermocouple, mini connectors
Terminal Blocks: 8 x Pt100 PC Interface: USB Power Supply: 6 Vdc (5.5 to 9.0V)
via universal mains adaptor (supplied), 120 to 250V, 50/60 Hz
Logging Interval: 5 second to 1 hour
M
BB
AA
DP9800, rear view, shown smaller than actual size.
Dimensions: mm (inch)
DP9800 shown smaller than actual size.
61.50 (2.42)
75.89 (3)
77.83 (3.06)
143
(5.62)
214
(8.42)
To Order Visit omega.com/dp9800 for Pricing and Details
Model No. Description
DP9800-TC
DP9800-RTD
Comes complete with power adaptor, USB cable, software and operator’s manual (on CD).
Ordering Examples: DP9800-TC, 8-channel thermocouple digital thermometer/data logger. OCW-3, OMEGACARESM extends standard 2-year warranty to a total of 5 years.
8-channel thermocouple digital thermometer/data logger with USB
8-channel RTD digital thermometer/data logger with USB
M-2