Omega Products DP7700 User Manual

HIGH-SPEED LOAD/STRAIN METERS AND PROCESS/VOLT METERS
DUAL DIFFERENTIAL INPUTS AVAILABLE
DP7600 Series
All Models
$
750
1000 Readings
per Second Maximum
5-Digit DisplayDual Inputs
Optional (Serial Interface Required)
Calibrating via 5 Front-Panel Switches
Programmable Digital
Filter, 1 to 500 Hz Cutoff
16-Bit A/D Converter
(65,000 Counts)
Tare (Zero) Up to 50%
of Span with Single Pushbutton
Peak and Runout Modes115/230 Vac Power
The DP7600 is a high-speed load/strain meter that performs 1000 conversions per second. It does limit checking and peak/ runout tracking at a user-selectable rate (from 1 to 500 Hz), with data displayed at the push of a button. The tare function, up to 50% of span, has absolute and relative reading modes. LEDs clearly indicate limit status and display mode settings. Five front-panel switches allow easy programming, and a security code prevents unwanted corruption of calibration data. Excitation voltage of 5 or 10 Vdc is standard.
The text at the top of the switch (NORMAL, PEAK, RUNOUT, TARE, ABS/REL) indicates its normal function. The text at the lower half of the switch (ENTER, UP, DOWN, NEXT, EXIT) indicates the switch’s function during setup and calibration.
The DP7700 is a high-speed process/volt meter with many of the same features. It accepts most process inputs from 0-200 mV to 0-10 Vdc and 4 to 20 mA current loops. Transducer/loop power of ±15 Vdc @ 50 mA and 4 open­collector outputs are standard.
SPECIFICATIONS
DP7600 Load/Strain Gage Meter
Excitation: 10 Vdc @ 50 mA,
5 Vdc @ 25 mA Coarse Gain Setting: Selectable by
switches on rear of unit
DP7600, $750, shown actual size, with LC105, load cell, $295, see page F-49.
Full Scale mV Ranges: 5.1, 10.2,
15.3, 20.0, 26.3, 30.6, 35.1, 39.2 Bridge Resistance: 100 to 5000 ;
4-wire or 6-wire configuration Calibration: Shunt calibration capability
with span programmable from front panel; no potentiometer adjustments; access code prevents unauthorized calibration switches
DP7700 Process/Volt Meter
Excitation: ±15 Vdc, ±5 Vdc @ ±50 mA Range/Input Impedance:
0 to 200 mVdc 10 x 10 0 to 5 Vdc 1 M 0 to 10 Vdc 1 M 4 to 20 mA 10
Range Selection: Selectable by DIP switch on rear of unit
Calibration: Digital calibration via front-panel button switches; no potentiometer adjustments; access code prevents unauthorized calibration switches
12
D-77
HIGH-SPEED PROCESS/ STRAIN METERS
GENERAL SPECIFICATIONS
Accuracy: ±0.01% of reading
±1 count of A/D @ 25°C (77°F)
Temperature Coefficient: 50 ppm/°C Operating Temperature: -20 to 60°C
(-4 to 140°F) Display: -32768 to 32767 with
programmable decimal points; 14 mm (0.56") red LED
Resolution of A/D: 16-bit Input Bias Current: ±50 nA max;
±100 pA/°C (per input channel) Front-Panel Controls: 5 pushbutton
switches for programming digital filter, offset, span, limits and recall of peak reading or runout from storage
Normal-Mode Rejection: 70 dB Rollover Error: ±2 counts of
A/D @ 25°C (77°F) Channels: One channel standard;
second channel optional Digital Filter: Programmable cutoff
frequency, 1 to 500 Hz; display update, 3 times per second
Reading Rate: Limited to 1000/s divided by number of channels
Input Analog Filter: 4-pole active anti-alias; 500 Hz cutoff on single-channel unit and 250 Hz cutoff on dual-channel unit
Peak Detect: Updated at rate selected by digital filter. Peak is normally stored in memory and recalled via front-panel switch, over the serial interface, or by selecting peak display mode on rear connector. If peak display mode is selected, an internal timer can be programmed to give limit #1 output up to 100 seconds after a programmed limit is reached. This output can be used to reset the peak reading by tying limit #1 output to remote input.
Runout Mode: Updated at rate selected by digital filter. Runout is normally stored in memory and recalled via front-panel switch, over the serial interface, or by selecting runout display mode on rear connector. If runout display mode is selected, an internal timer can be programmed to give limit #1 output up to 100 seconds after programmed limit is reached. This output can be used to reset the runout reading by tying limit #1 output to remote input.
Convert/Hold Input: Normally high; if pulled low, will hold display; requires 1 ms positive pulse for single conversion
Analog Output (Optional): 0 to 10 Vdc, 0 to 5 Vdc or 4 to 20 mA; tracks display reading; voltage outputs will sink or source 1 mA; 300V isolation between analog outputs and instrument ground
Serial Interface (Optional): ASCII RS232 or RS485 compatible, 300V isolation to system ground
Limit Outputs: 4 isolated open-collector transistor outputs rated 30 Vdc maximum; will sink up to 50 mA; 300 V isolation between limit outputs and system ground; limit #1 can be programmed to be delayed from 1 to 100 seconds
Relays (Optional): Dual 115 Vac,
2.5 A, form "A" relay contacts (isolated transistor outputs not available when relays are installed); limit #1 and limit #2 activate the relay contacts
Remote Inputs: Tare, display peak, display runout, and reset peak reading or runout reading; active low TTL compatible, 50 ms negative pulse required
I/O Method: Quick-disconnect screw terminal block; RJ11 jack for serial interface
* Requires “-S2” or ”-S4” option.
DP7600, $750, shown smaller than actual size.
See Section Y for a
Selection of Scientific,
Technical, and Reference
Books Available from
omega.com
Power: 115/230 Vac selectable by switch on rear of unit
DC Power (Optional): 10 to 30 Vdc @ 600 mA
Power Consumption: 8 W Construction: Aluminum case Dimensions:
69 H x 132 W x 175 mm D (2.7 x 5.18 x 6.9") including I/O connector
Panel Cutout: 63 H x 129 mm W (2.45 x 5.06")
Weight: 370 g (13 oz)
AVAILABLE FOR FAST DELIVERY!
To Order (Specify Model Number)
MODEL NO. PRICE DESCRIPTION
DP7600 $750 High-speed load/strain meter DP7700 750 High-speed process/voltmeter
Comes complete with operator’s manual. Ordering Example: DP7600-DUAL-S2, strain meter with dual-inputs and RS232,
$750 + 125 + 65 = $940.
OPTIONS
ORDER SUFFIX ADD’L PRICE DESCRIPTION
-DUAL* $125 Second input channel
-S2 65 ASCII RS232 serial interface
-S4 65 ASCII RS485 serial interface
-AMA 125 Analog output, 4 to 20 mA
-A10 125 Analog output, 0 to 10 Vdc
-A05 125 Analog output, 0 to 5 Vdc
-R 45 Dual 2.5 A
-DC9/18 125 9 to 18 Vdc power @ 600 mA
-DC18/36 125
18 to 36 Vdc power @ 500
D-78
@115 Vac form “A”
mA
relays