DIGITAL PROCESS OR
STRAIN GAGE PANEL METERS
WITH BUILT IN EXCITATION
DP63800 Process Meters
DP63900 Strain Meters
Starts at
$
245
DP63800-E, $245, shown actual size.
⻬ Models Available for
Process Voltage and
Current or Strain
Gage Bridge Inputs
⻬ 5-Digit 0.56" Red
Sunlight Readable
Display
⻬ Variable Intensity
Display
⻬ 16 Point Scaling for
Non-linear Processes
⻬ Programmable
Function Keys/User
Inputs
⻬ Four Setpoint
Alarm Outputs
(W/Option Card)
⻬ Communication and
BUS Capabilities
(W/Option Card)
⻬ Retransmitted
Analog Output
(W/Option Card)
⻬ NEMA 4X IP65 Sealed
Front Bezel
See Section Y for a
Selection of Scientific,
Technical, and Reference
Books Available from
omega.com
The DP63000 Series panel meters
for process voltage and current or
strain gage bridge inputs offer many
features and performance
capabilities to suit a wide range of
industrial applications. The meters
employ a bright 14 mm (0.56"),
sunlight readable, red LED display.
The intensity of display can be
adjusted from dark room
applications up to sunlight
readable, making it ideal for
viewing in lighting conditions.
A standard green LED is available
as an option. They have been
specifically designed for harsh
industrial environments with NEMA
4X (IP65) sealed front bezel and
extensive testing of noise effects
to CE requirements.
The meters provide a MAX and MIN
reading memory with
programmable capture time. The
capture time is used to prevent
detection of false max or min
readings which may occur during
start-up or unusual process events.
A variety of field installable plug-in
option cards are also available to
enhance the unit’s capabilities.
Option boards are available for
setpoints/alarms, analog output
or serial communications.
SPECIFICATIONS
Display: 14.2 mm (0.56") red sunlight
readable LED, (standard green LED
optional)
Display Range: 5-Digit, -19999 to
99999 counts
D-55
Power Requirements:
Standard: 85 to 250 Vac, 50/60 Hz,
15 Va
“-LV” Option: 11 to 36 Vdc or
24 Vac ±10%, 50/60 Hz, 11 W)
Excitation:
Process: 24 Vdc @ 50 mA, ± 5%
regulated
Strain: selectable 5 Vdc @ 65 mA or
10 Vdc @ 125 mA (± 2%)
Keypad: 3 programmable function keys,
5 keys total
Operating Temperature Range:
0 to 50°C (32 to 122°F) derate to 45°C
(113°F) with 3 option boards installed
Storage Temperature Range:
-40 to 60°C (-40 to 140°F)
Operating and Storage Humidity:
0 to 85% max. RH (non-condensing)
Altitude: Up to 2000 meters
Enclosure Rating: IP65 Face only
Connections: High compression cage-
clamp terminal blocks
Dimensions: 96.5 x 49.5 x 106.6 mm
(3.80 x 1.95 x 4.20")
Panel Cutout: 92 x 45 mm (3.62 x 1.77")
Depth Behind Panel: 104.1 mm (4.10")
Weight: 295 g (10.4 oz)
OPTION BOARD
COMMUNICATIONS CARDS
RS232/RS485
Isolation to Sensor and User Input
Commons: 500 Vrms for 1 min
Working Voltage: 50V Not Isolated
from all other commons
Baud Rate: 300 to 19.2k
Data Format: 7/8 bits; odd, even, or
no parity working (Voltage: 50V)
RS232 Communication Card
Type: RS232 half duplex
v
49.5
(1.95)
96.5
(3.80)
2.5
(0.10)
104.1
(4.10)
44.5
(1.75)
Dimensions: mm (in)
Note: Recommended minimum clearance (behind the panel)
for mounting clip installation is 53.4 (2.1) H x 127 (5) W.
RS485 Communication Card
Type: RS485 multi-point balanced
interface
Bus Address: 0 to 99, max
32 meters per line
Transmit Delay: Selectable;
2 to 50 msec or 50 to 100 msec
MODBUS
Type: RS485; RTU and ASCII
MODBUS modes
Isolation to Sensor and User Input
Commons: 500 Vrms for 1 minute
Working Voltage: 50V, not isolated
from all other commons
Baud Rates: 300 to 38400
Data: 7/8 bits
Parity: No, Odd, or Even
Addresses: 1 to 247
Transmit Delay: Programmable
®
ANALOG OUTPUT CARD
Types (User Selectable): 0 to 20 mA,
4 to 20 mA and 0 to 10 Vdc
Isolation to Sensor and User Input
Commons: 500 Vrms for 1 min
Working Voltage: 50V Not isolated
from all other commons
Accuracy: 0.17% of FS (18 to
28°C); 0.4% of FS (0 to 50°C)
Resolution: 1/3500
Compliance:
10 Vdc: 10 kΩ load min
20 mA: 500 Ω load max
Update Time: 200 msec max to
within 99% of final readout value
(digital filter and internal zero
correction disabled) 700 msec max
(digital filter disabled, internal zero
correction enabled)
D-56
SETPOINT OUTPUT CARDS
(Four types of field installable cards)
Response Time: 200 msec max to
within 99% of final readout value
(digital filter and internal zero
correction disabled) 700 msec max
(digital filter disabled, internal zero
correction enabled)
Dual Relay Card (LDP6-CDS10):
Type: Two FORM-C relays
Isolation to Sensor and User Input
Commons: 2000 Vrms for 1 min
Working Voltage: 250V
Contact Rating: One Relay
Energized: 5 amps @ 120/240 Vac
or 28 Vdc (resistiveload),
@120 Vac, inductive load. Total
Current With Both Relays Energized
not to exceed 5 amps
Life Expectancy: 100 k cycles min
at full load rating. External RC
snubber extends relay life for
operation with inductive loads
Quad Relay Card (LDP6-CDS20):
Type: Four FORM-A relays
Isolation to Sensor and User Input
Commons: 2300 Vrms for 1 min
Working Voltage: 250V
Contact Rating: One Relay
Energized: 3 amps @ 250 Vac or
30 Vdc (resistive load),
120 Vac, inductive load. Total
Current With All Four Relays
Energized not to exceed 4 amps
Life Expectancy: 100 k cycles min
at full load rating. External RC
snubber extends relay life for
operation with inductive loads
Quad Sinking Open Collector
(LDP6-CDS30):
Type: Four isolated sinking NPN
transistors
Isolation to Sensor and User Input
Commons: 500 Vrms for 1 min
Working Voltage: 50V Not Isolated
from all other commons.
Rating: 100 mA max @ VSAT = 0.7V
max Vax = 30V
Quad Sourcing Open Collector
(LDP6-CDS40):
Type: Four isolated sourcing PNP
transistors
Isolation to Sensor and User Input
Commons: 500 Vrms for 1 min
Working Voltage: 50V
(Not Isolated from all other
commons
Rating: Internal supply: 24 Vdc
± 10%, 30 mA max total all four
External Supply: 30 Vdc max,
100 mA max each output
)
1
⁄
1
10
⁄8 HP
HP @
INSTRUMENTATION
D