Omega Products DP63400-T-LV User Manual

Page 1
1
8 DIN Digital Panel
Temperature Meters
For Thermocouple and RTD Inputs
M-47
Specifications
Display: 5-digit, 14.2 mm (0.56") red
Power:
AC Versions:
AC Power: 85 to 250 Vac,
50/60 Hz, 10 VA Isolation: 2300 Vrms for 1 minute
to all inputs
Low Voltage Versions:
DC Power: 11 to 36 Vdc, 11 W AC Power: 24 Vac, ± 10%,
50/60 Hz, 10 VA
Isolation: 500 Vrms for 1 minute to all inputs (50V working)
The DP63400 Series panel meters offer many features and performance capabilities to suit a wide range of industrial applications. These meters accept thermocouple or RTD inputs. The meters provide a maximum and minimum reading memory with programmable capture time. The capture time is used to prevent detection of false max or min readings which may occur during start-up or unusual process events. These meters have been specifically designed for harsh industrial environments. With NEMA 4X (IP65) sealed bezel and extensive testing of noise effects to CE requirements, the DP63400 Series meters provide a tough, yet reliable, application solution.
DP63400-T
Starts at
$
201
Thermocouple and RTD
Panel Meter
5-Digit,
14 mm (0.56") LED
Conforms to ITS-90
Standards
°F or °C Unit Overlay
with Backlight
NEMX 4X (IP65)
Keypad: 3 programmable function keys,
5 keys total
A/D Converter: 16-bit resolution Update Rates:
A/D Conversion Rate: 10 readings/s Step Response: 200 ms max to
within 99% of final readout value (digital filter and internal zero correction disabled), 700 ms max (digital filter disabled, internal zero correction enabled)
Display Update Rate: 1 to 10 updates/s Max/Min Capture Delay Time: 0 to 3275 s
Low-Frequency Noise Rejection:
Normal Mode: >60 dB @ 50 or 60 Hz
±1%, digital filter off
Common Mode: >100 dB, DC to 120 Hz
Memory: Non-volatile E
2
PROM retains all programmable parameters and display values
Environmental Conditions:
Operating Temperature Range:
0 to 50°C (32 to 122°F)
Storage Temperature Range:
-40 to 60°C (-40 to 140°F)
Operating and Storage Humidity:
0 to 85% max RH non-condensing
Altitude: Up to 2000 m (6562')
Connections: High-compression,
cage-clamp terminal block
Wire Strip Length: 7.5 mm (0.3") Wire Gage: 30 to 14 AWG copper wire Torque: 0.51 N/m (4.5 lb/in) max Construction: This unit is rated for
NEMA 4X (IP65) outdoor use, IP20 touch-safe, Installation Category II, Pollution Degree 2, 1-piece bezel/case, flame-resistant, synthetic rubber keypad—panel gasket and mounting clip included
Weight: 200 g (7 oz)
49.5
(1.95)
96.5
(3.80)
v
Note: Recommended minimum clearance (behind the panel) for mounting clip installation is
53.4 H (2.1) x 127 W (5).
Dimensions: mm (in)
DP63400-T, $201, shown smaller than actual size.
91.4
(3.60)
44.5
(1.75)
2.5
(0.10)
104.1 (4.10)
44.5
(1.75)
Panel Punches
Available Visit
omega.com/
panelpunches
Page 2
M-48
M
Thermocouple Inputs
Resolution: Variable: 0.1, 0.2, 0.5, or
1, 2, or 5 degrees
Scale: °F or °C
Input Type Range Accuracy* (18 to 28°C) Accuracy* (0 to 60°C)
-200 to 400°C (-328 to 752°F)
-270 to -200°C (-454 to -328°F)
1.2°C** 2.1°C
-200 to 871°C (-328 to 1600°F)
-270 to -200°C (-454 to -328°F)
1.0°C** 2.4°C
-200 to 760°C (-328 to 1400°F) 1.1°C 2.3°C
-200 to 1372°C (-328 to 2502°F)
-270 to -200°C (-454 to -328°F)
1.3°C** 3.4°C
-50 to 1768°C (-58 to 3214°F) 1.9°C 4.0°C
-50 to 1768°C (-58 to 3214°F) 1.9°C 4.0°C 100 to 300°C (100 to 572°F) 3.9°C 5.7°C
300 to 1820°C (572 to 3308°F) 2.8°C 4.4°C
-200 to 1300°C (-328 to 2372°F)
-270 to -200°C (-454 to -328°F)
1.3°C** 3.1°C
0 to 2315°C (32 to 4199°F) 1.9°C 6.1°C
*
After 20 minute warm-up. Accuracy is specified in 2 ways—accuracy over an 18 to 28°C (64 to 82°F) in a 15 to 75% RH environment and accuracy over a 0 to 50°C (32 to 122°F) in a 0 to 85% RH (non-condensing) environment. Accuracy specified over the 0 to 50°C (32 to 122°F) operating range includes meter tempco and ice point tracking effects. The specification includes the A/D conversion errors, linearization conformity, and thermocouple ice point compensation. Total system accuracy is the sum of meter and probe errors. Accuracy may be improved by field calibrating the meter readout at the temperature of interest.
**
The accuracy over the interval -270 to -200°C (-454 to -328°F) is a function of temperature, ranging from 1°C at -200°C and degrading to
7°C at -270°C. Accuracy may be improved by field calibrating the meter readout at the temperature of interest.
Offset Range: -19,999 to 99,999 display units Input Impedance: 20 MΩ Lead Resistance Effect: 0.03µV/Ω Maximum Continuous Overvoltage: 30V
Input Accuracy* Accuracy* Type Range (18 to 28°C) (0 to 50°C)
100 Ω Pt -200 to 850°C alpha = .00385
(-328 to 1562°F)
0.4°C 1.6°C
100 Ω Pt -200 to 850°C alpha = .003919
(-328 to 1562°F)
0.4°C 1.6°C
120 Ω Nickel -80 to 260°C alpha = .00672
(-112 to 1562°F)
0.2°C 0.5°C
10 Ω Copper -100 to 260°C alpha = .00427
(-148 to 500°F)
0.4°C 0.9°C
RTD Inputs
Type: 3- or 4-wire, 2-wire can be
compensated for lead wire resistance
Excitation Current:
100 Ω Range: 165 µA 10 Ω Range: 2.6 mA
Accuracy* Accuracy*
Input Type Range (18 to 28°C) (0 to 50°C) Direct mV -10 to 65 mV 0.02% of rdg 0.12% of rdg
Range (1 µV res.) + 4µV + 5µV Direct 100 Ω 0 to 400 W 0.02% of rdg 0.12% of rdg
Range (10 MW res.) + 0.04 Ω + 0.05 Ω Direct 10 Ω 0 to 25 W 0.04% of rdg 0.20% of rdg
Range (1 MW res.) + 0.005 Ω + 0.007 Ω
Direct Readout
Input Range: -10 to 65 mV
0 to 400 Ω: High range 0 to 25 Ω: Low range
Display Range: -19999 to 99999
To Order
(Specify Model Number)
Model No. Price Description
DP63400-T $201 Temperature input, 85 to 250 Vac, 50/60 Hz
DP63400-T-LV 228 Temperature input, 11 to 36 Vdc, 24 Vac
Comes complete with operator’s manual. Ordering Example: DP63400-T-LV, temperature input, 11 to 36 Vdc, $228.
AVAILABLE FOR FAST DELIVERY!
Lead Resistance:
100 Ω Range: 10 Ω/lead max 10 Ω Range: 3 Ω/lead max
Model No. Price Description DPP-5 $525
1
8 DIN Panel Punch
Accessories
Recommended Reference Book: Sensor Technology Handbook, CS-3766, $140. See Section Y for Additional Books.
Page 3
One Omega Drive | Stamford, CT 06907 | 1-888-TC-OMEGA (1-888-826-6342) | info@omega.com
EPG05
www.omega.com
UNITED KINGDOM
www. omega.co.uk
Manchester, England
0800-488-488
UNITED STATES
www.omega.com
1-800-TC-OMEGA
Stamford, CT.
CANADA
www.omega.ca
Laval(Quebec)
1-800-TC-OMEGA
GERMANY
www.omega.de
Deckenpfronn, Germany
0800-8266342
Karviná, Czech Republic
FRANCE
www.omega.fr
Guyancourt, France
088-466-342
CZECH REPUBLIC
www.omegaeng.cz
596-311-899
BENELUX
www.omega.nl
Amstelveen, NL 0800-099-33-44
More than 100,000 Products Available!
Temperature
Calibrators, Connectors, General Test and Measurement Instruments, Glass Bulb Thermometers, Handheld Instruments for Temperature Measurement, Ice Point References, Indicating Labels, Crayons, Cements and Lacquers, Infrared Temperature Measurement Instruments, Recorders Relative Humidity Measurement Instruments, RTD Probes, Elements and Assemblies, Temperature & Process Meters, Timers and Counters, Temperature and Process Controllers and Power Switching Devices, Thermistor Elements, Probes and Assemblies,Thermocouples Thermowells and Head and Well Assemblies, Transmitters, Wire
Flow and Level
Air Velocity Indicators, Doppler Flowmeters, Level Measurement, Magnetic Flowmeters, Mass Flowmeters, Pitot Tubes, Pumps, Rotameters, Turbine and Paddle Wheel Flowmeters, Ultrasonic Flowmeters, Valves, Variable Area Flowmeters, Vortex Shedding Flowmeters
pH and Conductivity
Conductivity Instrumentation, Dissolved Oxygen Instrumentation, Environmental Instrumentation, pH Electrodes and Instruments, Water and Soil Analysis Instrumentation
Data Acquisition
Auto-Dialers and Alarm Monitoring Systems, Communication Products and Converters, Data Acquisition and Analysis Software, Data Loggers Plug-in Cards, Signal Conditioners, USB, RS232, RS485 and Parallel Port Data Acquisition Systems, Wireless Transmitters and Receivers
Pressure, Strain and Force
Displacement Transducers, Dynamic Measurement Force Sensors, Instrumentation for Pressure and Strain Measurements, Load Cells, Pressure Gauges, Pressure Reference Section, Pressure Switches, Pressure Transducers, Proximity Transducers, Regulators, Strain Gages, Torque Transducers, Valves
Heaters
Band Heaters, Cartridge Heaters, Circulation Heaters, Comfort Heaters, Controllers, Meters and Switching Devices, Flexible Heaters, General Test and Measurement Instruments, Heater Hook-up Wire, Heating Cable Systems, Immersion Heaters, Process Air and Duct, Heaters, Radiant Heaters, Strip Heaters, Tubular Heaters
click here to go to the omega.com home page
Loading...