1
⁄8 DIN Digital Panel
Thermocouple Meters
M-49
The DP63300-TC thermocouple
meter accepts inputs from standard
thermocouples and precisely
linearizes them. A full 4-digit display
accommodates a wide range of
temperature inputs. The unit
automatically compensates for cold
junction, NBS linearity and the
meter’s zero and span.
The meter features a readout choice
of either Fahrenheit or Celsius with
0.1 or 1° resolution. Display prompts
and front panel buttons aid the
operator through set-up and
operation. With a few simple steps
the unit can be used as a millivolt
meter. This mode is useful in
monitoring and displaying the actual
voltage produced at the
thermocouple probe junction and as
an aid in troubleshooting for a faulty
thermocouple probe.
DP63300-TC
$
186
⻬ Thermocouple Meter
⻬ 4-Digit, 14.2 mm
(0.56") LED
⻬ Programmable
Thermocouple Type
(T, E, J, K, R, S, B, N or
mV Scale)
⻬ Conforms to ITS-90
Standards
⻬ Selectable °F or °C with
0.1 or 1° Display
Resolution
⻬ Programmable
Temperature Offset
⻬ Programmable Digital
Filtering Enhances
Stability
The meter provides a peak “HI” and
valley “LO” reading memory with
selectable capture delay time. The
capture delay is used to prevent
detection of false peak or valley
readings that may occur during
start-up or unusual process events.
The peak and valley readings are
stored at power-down so that
process limits can be monitored
over any length of time.
Programmable digital filtering
enhances the stability of the
reading.
All setup data is stored in EEPROM,
which will hold data for a minimum
of 10 years without power. The
meter has several built-in diagnostic
functions to alert operators of any
malfunction. Extensive testing of
noise interference mechanisms and
full burn-in makes the indicator
extremely reliable in industrial
environments. The front bezel meets
NEMA 4X (IP65) requirements for
wash-down applications.
⻬ Peak/Valley (High/Low
Reading) Memory
⻬ NEMA 4X (IP65) Sealed
Front Bezel
⻬ Custom Units Overlay
with Backlight
DP63300-TC, $186, shown larger than actual size.
Panel Punches
Available, Visit
omega.com/panelpunches
M-50
M
Specifications
Display: 4-digit, 14.2 mm H (0.56")
LED, minus sign displayed for negative
temperatures
Power: 85 to 250 Vac, 50/60 Hz, 6 VA
Isolation: 2300 Vrms for 1 min between
input and supply (300V working voltage)
Thermocouple Types: T, E, J, K, R, S,
B, N or mV scale
Resolution: 1° for all types, or 0.1° for
T, E, J, K and N only
Thermocouple Range and Accuracy:
All errors include NBS conformity, cold
junction effect and A/D conversion
errors at 23°C after 60 minutes
warm-up; relative humidity <85%
Input Impedence: 20 MΩ, all types
Lead Resistance Effect: 20 µV/350 Ω
Max Input Voltage Protection:
70 Vdc continuous
Cold Junction Compensation:
Automatic, 0.02 degree/degree
(disabled for linear mV scale)
Reading Rate: 2.5 readings/second
Response Time: 2 seconds to settle for
step input (increases with
programmable digital filtering)
Low-Frequency Noise Rejection:
Normal Mode Rejection: 45 dB
@ 50/60 Hz (may be improved by
programmable digital filtering)
Common Mode Rejection:
120 dB, DC to 50/60 Hz
Environmental Conditions:
Operating Temperature Range:
0 to 50°C (32 to 122°F)
Storage Temperature Range:
-40 to 80°C (-40 to 176°F)
Operating and Storage Humidity:
85% max (non-condensing) from
0 to 50°C (32 to 122°F)
Span Drift: 40 ppm/°C
To Order
(Specify Model Number)
Model No. Price Description
DP63300-TC $186 Thermocouple input, 85 to 250 Vac, 50/60 Hz
AVAILABLE FOR FAST DELIVERY!
2.5
(0.10)
104.1
(4.10)
44.5
(1.75)
Thermocouple
Type Range Accuracy
-200 to 400°C 0.8°C
(-328 to 752°F) (1.4°F)
-200 to 1000°C 0.8°C
(-328 to 1832°F) (1.4°F)
-200 to 760°C 0.8°C
(-328 to 1400°F) (1.4°F)
-200 to 1250°C 0.8°C
(-328 to 2282°F) (1.4°F)
0 to 1768°C 2.1°C
(32 to 3214°F) (3.8°F)
0 to 1768°C 2.1°C
(32 to 3214°F) (3.8°F)
150 to 1820°C 2.3°C
(302 to 3308°F) (4.1°F)
-200 to 1300°C 0.8°C
(-328 to 2372°F) (1.4°F)
mV -10.00 to 80.00 mV 0.01%
Thermocouple Types
Note: Recommended minimum clearance
(behind the panel) for mounting clip
installation is 53.4 (2.1) H x 127 (5) W.
Zero Drift: 1 µV/°C
Altitude: Up to 2000 m (6562')
Construction: This unit is rated for
NEMA 4X (IP65) indoor use, 1-piece
bezel/case, flame resistant—panel
gasket and mounting clip included
Connections: High-compression, cageclamp terminal block
Weight: 0.24 kg (0.65 lb)
Model No. Price Description
DPP-5 $525
1
⁄
8 DIN panel punch
EE-1319 85 Reference Book: Grounding and Shielding Techniques
Accessories
Comes complete with operator’s manual.