1
⁄8 DIN Temperature/Process
Meter/Scanner
With Data Logging and RS232 Communications
DP470-T-C2, $365,
shown larger than
actual size.
DP470 Series
Starts at
$
365
Single- and Multiple-
Input Models
Thermocouple, RTD, and
Process Input Models
Selectable 0.1/1° Resolution
Selectable Digital °C/°F
Temperature Units
4-Digit, 14-Segment
Alphanumeric Display
RS232 Communications
Standard
Captures Min
and Max Readings
Free Windows-Based
Software
View all Channel Data,
Alarm Status, and
Min/Max Values
Simultaneously
Log Real-Time Data
Create Exportable
Data Files
The DP470 Series of single- and
multiple-input digital temperature
and process indicators sets a
standard for performance, features,
value, and reliability. With built-in
digital programmability,
communications, and powerful
Windows-based software, the
DP470 is ideal for monitoring,
1
⁄8 DIN
testing, and process-control
applications. Complete configuration
of the instrument can be quickly
performed from the front panel,
including designating sensor input
type, sampling rate, limit settings,
and resolution. Display of process
and min/max values is also user
selectable from the front panel. The
standard single-input model features
easy-to-use front-panel configuration
and display of temperature or
process measurement with either
0.1 or 1° resolution, min/max values,
and limit conditions.
Windows-Based RS232
Communications Software
The DP470 is far more than a
stand-alone, high-performance
digital temperature or process
indicator. Thanks to the exclusive
DP470 software, the DP470 is a
powerful real-time data monitoring,
control, and data logging system.
Clear, menu-driven displays guide
the user through instrument setup
and allow the monitoring, collection,
and logging of data remotely through
a PC serial port. Software alarms,
channel ID, and min/max values are
presented, along with a simulated
LED display of the selected channel,
in an on-screen control panel format.
In addition, data can be exported as
text to create graphs and print
hard-copy reports. (No Protocol is
available).
M-81
Multiple Inputs
With the multiple-input model
installed, the DP470 Series can
monitor up to 6 thermocouple or
process inputs or 3 RTD channels.
On single-input models, an optional
board can be installed. This board
provides enhanced alarm functions
as well as a scalable 4 to 20 mA/0 to
10V analog output.
All active inputs can be manually
selected or automatically scanned.
A front-panel channel indicator
accompanies the displayed data.
Individual limits can be assigned to
each channel. A built-in 5 A rated
form “C” relay is energized
whenever the limit on the selected
channel is exceeded.
Dual Alarms
With the dual-alarm model, the DP470
can be configured to perform on/off
control functions as well as process
monitoring. Setpoints can be set for
high, low, or deviation. In addition,
hysteresis (deadband) and delay can
be programmed into each setpoint.
Alarm conditions can be programmed
for manual acknowledgment or for
automatic reset. The dual-alarm option
comes equipped with two 5 A rated
form “C” (SPDT) relays, one for each
alarm setpoint. A flashing display
alerts the user to an alarm condition.
Panel punches
available, visit
omega.com/
panelpunches
Analog Output Models
Extended Warranty
Program
SM
Analog output capability is also
available in combination with the dual
alarm option. A linearized output
(4 to 20 mA or 0 to 10V) with fully
adjustable zero and span allows the
instrument to serve as an indicating
transmitter, or as a high-level input to
a PC-based converter or other
recording device.
Specifications
Power: 100 to 240 Vac (±10%),
50 to 400 Hz
Display: 4-digit, 14-segment red
14.2 mm H (0.56") LED,
10 mm (0.4") green °C/°F LED, option
10 mm (0.4") green
Accuracy:
1° Resolution:
J, K, T, E: ±1°C/F + 0.03% rdg
R, S: ±1°C/F + 0.05% rdg
2-, 3-, 4-Wire RTD:
±0.8°C/F + 0.03% rdg
0.1° Resolution:
J, K, T, E: ±0.5°C (0.9°F) ±1 digit
2-, 3-, 4-Wire RTD:
±0.5°C/0.9°F ±1 digit
Process: 0.02% rdg ±1 count
Multi-Input Option Accuracy:
Add ±0.5°C/0.9°F to instrument
accuracy specification
Repeatability: ±1 count
Temperature Stability:
Zero: 1 µV/°C
Span: 0.01%
Thermocouple Reference Junction:
Internal, automatic, 0.03°C/°C,
5 to 45°C (41 to 113°F)
Break Detection: Up-scale 50 nA,
unit displays OL
Environmental Ranges:
Operating Temperature:
0 to 50°C (32 to 122°F)
Storage Temperature:
-40 to 65°C (-40 to 149°F)
Humidity:
<80% RH, non-condensing
Stability with Time: 0.8° per year
Noise Rejection: NMRR 60 dB @
50/60 Hz, CMRR 120 dB @ 50/60 Hz
(±0.1 Hz with 250 Ω unbalanced)
Overload Protection:
Power Leads to Ground:
1500 Vdc or Vac rms
Across Inputs, for 1 Minute:
Inputs up to 250 Vac/dc
Input Impedance:
Thermocouple: 20 MΩ, exclusive of
break detect current effects
RTD: 16.9 kΩ
Process:
Voltage, 1 MΩ; current, 5 Ω
Point Update Rate: 2/s nominal
(1° readings); 1/s nominal (0.1° readings)
Alarm Relay Contact Rating:
5 A @ 120 Vac SPDT
(non-inductive load) form “C”
Multiple-input viewer monitors channels, alarm conditions
and data logging functions.
RS232 Serial Communications:
Full-duplex voltage, isolated from ground to
500 Vac; 9600 baud, 8 bits, 1 stop bit, no
parity, ASCII, (No Protocol is available).
Options/Accessories
Dual Alarm Model: Two 5 A/120V,
form “C” relay outputs
Analog Output Model: 0 to 10 Vdc,
fully scalable (load current 2 mA max);
4 to 20 mA DC, fully scalable
(load resistance 300 Ω max)
Analog Output Accuracy:
0.25% FS of display value
Analog Output Resolution:
Approximately 0.0125 FS
Analog Output Isolation:
Isolated between input and
internal circuit to 500 Vac
Multiple-Input Option: Up to
6 thermocouple/process inputs or three
3- or 4-wire RTDs; auto/manual scanning;
LED channel indicator, 10 mm (0.4")
green; individually programmable
setpoints on each input; 5 A/120V, form
“C” relay (SPDT) output energized
whenever limit exceeded on any channel
Size: 48 x 96 mm (1.89 x 3.78")
Panel Cutout:
1
⁄8 DIN,
45 x 92 mm (1.8 x 3.6")
Weight: 454 g (1 lb)
OMEGACARESMextended
warranty program is
available for models
shown on this page.
OMEGACARESMcovers
parts, labor, and
equivalent loaners.
Ask your sales
representative for full
details when placing
an order.