Economical High Accuracy
Temperature Indicators
Single and Six-Channel Models
DP460 Series
DP460 base meter.
Temperature Input Models
U Accurate to 0.5°C (0.9°F)
U Selectable Thermocouple Input:
J, K, T, E, R, S, and B
U RTD Input: 0.00385 or 0.00392 alpha
U 0.1°/1° Auto-Resolution
U °F/°C Switchable
Current and Voltage Input Models
U Current and Voltage Selectable
U 0.02% Accuracy, 1-Digit Repeatability
U Pushbutton Digital Scaling and Offset
U -999 to 9999 Display Range
U DP460-E Has a Transmitter/Transducer
Excitation Power Supply
Strain Gage and Load Cell Input Models
U 0 to 5 mV/V and 0 to 50 mV Selectable
U Pushbutton Digital Display Range
U -999 to 9999 Display Range
U Built-in 10 Vdc Excitation Power Supply
U Auto Re-Zero (Tare) Button
Optional
U Multi-Channel Inputs (for Units Without
Excitation Power Supply)
U Single or Dual 5A Alarm Relays
KTSS-14G-12 probe,
sold separately.
Visit omega.com/ktss_jtss_chb
DP461 meter with integral alarm.
Select Input Type Quickly
with internal Rotary Switch
DP462 meter with multi-input.
M-47

DP460
Base Meter
Meter with integral alarm
DP461
DP462
Meter with multi-input
KTSS-14G-12 Probe, sold separately.
Visit omega.com/ktss_jtss_chb
The DP460 Series is designed for
a variety of temperature, current,
voltage, strain gage and load cell
measuring applications. Excellent
accuracy and small size combine
to make it the ideal choice for most
industrial indications.
The DP460-T and DP460-RTD
temperature models feature a unique
0.1° or 1° automatic resolution
(unless the instrument is programmed
to display 1° only). Select among nine
temperature sensors or thermocouple
types, J, K, T, E, R, S, and B or 100
ohm; 3-wire, Platinum RTD’s for
either 0.00385 or 0.00392 alpha.
Behind the snap-off lens, each of
these nine thermocouples can be
selected via a miniature rotary switch.
No calibration is required to switch
between thermocouple types. Another
miniature DIP switch selects between
°F or °C display.
The DP460-V voltage models allow
selection between mA or voltage input
with a switch easily accessible behind
the snap-off front lens. Full scales are
0 to 20 mA or 0 to 10 V. These or any
subset (e.g., 4 to 20 mA, 1 to 5 V) can
be scaled. Two simple pushbuttons
conveniently located behind the snapoff lens are used to set zero (Lo), full
scale (Hi) and a decimal point. The
indicator can be scaled within its wide
-999 to 9999 display count.
The DP460-S strain gage models
are compatible with both bondedfoil type (2 mV/V, 3 mV/V, 4 mV/V)
transducers and higher output,
semiconductor (10 mV/V, 20 mV/V,
30 mV/V) type strain gages.
Its microprocessor design and
ratiometric measuring circuit
assure high accuracy with superior
repeatability. Scaling and calibrating
the DP460-S to display engineering
units from the strain gage input is
similar to the DP460-V models. There
are no pots to drift nor complicated
formulas to compute to set DIP
switch gain ranges. It can achieve
sensitivities as low as 1 µV per count
and can be scaled anywhere within
its wide -999 to 9999 display count.
Meters shown smaller than actual size
The alarm option provides single
(ALM) or dual (ALM2) limit setpoints
that can be set to any point within
the DP460 series’ measuring range.
The user digitally adjusts the setpoint
by pressing two buttons located on
the front panel. Selection of Hi or Lo
(single) alarm or Hi and Lo (dual)
alarms is set by moving a jumper
located on the option card. You can
view the setpoint by pressing the
View button on the front panel. A
light on the front panel illuminates
red for the single alarm or red/yellow
(dual alarm) when the alarm trips.
The DP462 series has the ability
to switch between up to six sensor
inputs of the same type with its
multi-input option. A miniature rotary
switch, accessible from the front
panel, allows convenient switching
between inputs. Sensor hook-up is a
snap-on/off terminal block located at
the rear of the instrument. It is easy
to modify a basic DP460 model and
add the multi-input board (DP460206 or DP460-403) in the field.
The Rack Adaptor Plate permits easy
mounting of DP460 Series indicators
in standard 483 mm (19") instrument
racks. Four versions are available for
mounting 1, 2, 3 or 4 indicators.
Common Specifications
Display: Four 7-segment red/
orange LED’s, 14.2 mm (0.56") H
Repeatability: ±1 count
Stability with Temperature
Zero: 1 µV/°C
Span: 0.01% reading/°C
Power: 115/230 Vac selectable
(±10%), 50 to 60 Hz, 400 Hz
Noise Rejection
NMRR: 60 dB @ 50/60 Hz
(±0.1 Hz)
CMRR: 120 dB @ 50/60 Hz
(±0.1 Hz with 250 unbalance)
Overload Protection (Power Lead
to Ground): Across inputs, for 1
minute T/C up to 250 Vac/dc, V+ to
V- RTD up to 62 Vac/dc, V+ or I-
Environmental Ranges
Operating: 5 to 45°C
(41 to 113°F)
Storage: -40 to 65°C
(-40 to 149°F)
Humidity: < 80% RH
non-condensing
Alarm Relay Contact Rating
5 A @ 120 Vac (noninductive load),
Form C (SPDT)
Input Types and Ranges
1° Resolution 0.1° Resolution
Input Type Range Accuracy* Range Accuracy*
Iron- -337 to 1403°F 0.03% -99.9 to 999.9°F 0.9°F
Constantan -205 to 762°C 0.03% -99.9 to 761.8°C 0.5°C
CHROMEGA
ALOMEGA
Copper- -346 to 775°F 0.03% -99.9 to 754.6°F 0.9°F
Constantan -210 to 401°C 0.03% - 99.9 to 401.4°C 0.5°C
CHROMEGA
Constantan -205 to 1002°C 0.03% -99.9 to 999.9°C 0.5°C
Pt-Pt13% Rh 32 to 3216°F 0.05%
Pt-Pt10% Rh 0 to 1769°C 0.05%
Pt30% Rh- 316 to 1829°C 0.05%
Pt6% Rh- 601 to 3325°F 0.05%
Pt100 -329 to 1563°F 0.03% -99.9 to 999.9°F 0.9°F
385 or 392 -200 to 851°C 0.03% -99.9 to 850.8°C 0.5°C
Process
Strain
* Accuracy is % reading + 1 digit. † For strain, refer to accuracy specifications for DP460-S
®
- -331 to 2510°F 0.03% -99.9 to 999.9°F 0.9°F
®
-202 to 1377°C 0.03% -99.9 to 999.9°C 0.5°C
®
-338 to 1835°F 0.03% -99.9 to 999.9°F 0.9°F
Not Available
Not Available
0 to 20 mA or 0 to 10 Vdc, Scalable -999 to 9999 0.02%
0 to 50 mV or 0 to 500 mV, Scalable -999 to 9999 0.02%†
-
-
M
†
M-48

DP460-E face shown smaller than actual size
Convenient pushbutton scaling allows the
DP460-E to display from -999 to 9999
Case: One-piece, black plastic,
screw terminal terminations
Case: 40 H x 90 W x 13 mm D
(1.57 x 3.54 x 5.35")
Bezel: 48 x 96 mm (1.89 x 3.78")
1
Panel Cutout:
⁄8 DIN, 45 x 92 mm
(1.77 x 3.62")
Weight: 454 g (1 lb)
DP460-T and DP460-R
Input Impedance:
Thermocouple, 20 MΩ
RTD, 16.9 kΩ
RTD: 16.9 kΩ + V + input - I input
RTD Lead Wire Error: At 150 µA
excitation current 40 mΩ/Ω of
equal resistance in V+ and Vleads,1 Ω/Ω of imbalance in
V+ or V- leads
Thermocouple Reference
Junction: Internal, automatic,
0.03%°C/°C, 5 to 45°C
(41 to 113°F)
Break Detection:
Upscale = 50 nA
Point Update Rate:
2 per second nominal (1° readings);
1 per second nominal
(0.01° readings)
Multi-Input Option
Accuracy: Add ±0.5°C/±1°F
to instrument accuracy
DP460-E and DP460-V
Read Rate: 2 per second nominal
Input Impedance:
Voltage: 1 MΩ, Current: 5 Ω
Excitation: DP460-E:
built-in 10 Vdc (±5%), 30 mA
DP460-E shown actual size
DP460-S
Display Scaling Range:
-999 to 9999
Accuracy:
0 to 5 mV/V: Sensitivity ±0.02% rdg,
±1 count for sensitivities ≤2 µV/ct;
±0.02% rdg, ±4 cts for sensitivities
≥ 2 µV/count
0 to 50 mV/V: Sensitivity ±0.02% rdg,
±1 ct for sensitivities ≥ 20 V/ct;
±0.02% rdg, ±4 cts for sensitivities
≥ 20 µV/count.
Sensitivity
(minimum 1 display count):
0 to 5 mV/V: maximum 1 µV/count,
minimum 1 display count
0 to 50 mV/V: maximum 10 µV/count
Stability with Temperature Zero:
1 µV/°C
Span: 0.01%rdg/°C
Read Rate: 2 per second nominal
Input Impedance: Inputs ≥-1 MΩ
Excitation Sense: 1 MΩ
Measuring Technique:
Dual slope, ratiometric,
differential input max
Excitation: Built-in 10 Vdc (±5%),
30 mA
Input and Output Options*
Model No. Description
DP460-ALM 5A Alarm board
DP460-ALM2 Dual 5A alarm board
DP460-206** 6-channel input board
DP460-403 3-channel input board
*Only one option board can be installed into
a DP460 base meter.
**DP460-206 cannot be used with models
with “-E” or “-S” inputs.
To Order Visit omega.com/dp460-t for Pricing and Details
Model No. Description
DP460-(*) Base meter
DP461-(*) Meter with single alarm relay output
DP461A-(*) Meter with dual alarm relay output
DP462-(*) Meter with six-channel inputs
Comes complete with user’s manual and self-adhesive engineering unit front panel labels.
*Insert “T” for thermocouple input, “RTD” for RTD input, “V” for Voltage, “E” for Voltage with
10 Vdc excitation, “S” for Strain input with 10 Vdc excitation
Note: E and S inputs are not available for the DP462 models.
Rack adaptor plates are available for mounting 1 to 4 indicators in a standard 483 mm (19") rack.
4 models: DP460-RKS1, DP460-RKS2, DP460-RKS3, DP460-RKS4.
OMEGACARE
OMEGACARE
full details when placing an order.
Ordering Example: DP462-T, base meter with multi-channel thermocouple input.
OCW-3 OMEGACARE
SM
extended warranty program is available for models shown on this page.
SM
covers parts, labor, and equivalent loaners. Ask your sales representative for
SM
extends standard 2-year warranty to a total of 5 years.
M-49