Omega Products DP40-9SC2 Installation Manual

®
User ’s Guide
http://www.omega.com
e-mail: info@omega.com
DPF400 Series
High Accuracy Frequency or
Analog Input Flow Indicator
®
®
OMEGAnet® On-Line Service Internet e-mail
http://www.omega.com info@omega.com
Servicing North America:
USA: One Omega Drive, Box 4047
ISO 9001 Certified Stamford, CT 06907-0047
Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
Customer Service: 1-800-622-2378 / 1-800-622-BEST Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
Mexico and Latin America:
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Tel: (95) 800-TC-OMEGA
SM
En Espan˜ol: (203) 359-7803 e-mail: espanol@omega.com
SM
SM
SM
FAX: (95) 203-359-7807
Servicing Europe:
Benelux: Postbus 8034, 1180 LAAmstelveen, The Netherlands
Czech Republic: ul.Rude armady 1868, 733 01 Karvina - Hranice
France: 9, rue Denis Papin, 78190 Trappes
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
United Kingdom: One Omega Drive,
ISO 9002 Certified River Bend Technology Centre
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or commissioning this device as it contains important information relating to safety and EMC.
Tel: (31) 20 6418405 FAX: (31) 20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: nl@omega.com
Tel: 420 (69) 6311899 FAX: 420 (69) 6311114 e-mail: czech@omega.com
Tel: (33) 130-621-400 FAX: (33) 130-699-120 Toll Free in France: 0800-4-06342 e-mail: france@omega.com
Tel: 49 (07056) 3017 FAX: 49 (07056) 8540 Toll Free in Germany: 0130 11 21 66 e-mail: germany@omega.com
Northbank, Irlam, Manchester, M44 5EX, England Tel: 44 (161) 777-6611 FAX: 44 (161) 777-6622
Toll Free in England: 0800-488-488
e-mail: info@omega.co.uk
Table of Contents -
HOW TO USE THIS MANUAL
This manual is organized to follow a sequence of setting up the meter, configuring it, and operating it. The table of contents reflects this sequence. The contents also show that the manual contains a lot of information; we have taken care to answer likely questions and provide all the information you may need. We have purposely included all sections in the contents, to allow you to use it as an index–to easily pinpoint specific information and go
directly to it.
Page
SAFETY CONSIDERATIONS.......................................................................vi
SECTION 1 INTRODUCTION
1.1 Description ...................................................................................1
1.2 Features.........................................................................................1
1.3 Meter Modes.................................................................................2
1.3.1 Rate Meter/Totalizer.....................................................................2
1.3.2 Rate Meter/Totalizer/Square Root Extractor ...............................2
1.3.3 Batch Controller...........................................................................2
1.3.4 Clock.............................................................................................2
1.4 Optional Boards Overview...........................................................3
1.4.1 Isolated Pulse Input Board............................................................3
1.4.2 Isolated Analog Input Board........................................................3
1.4.3 Isolated Analog Output Board.....................................................4
1.4.4 Isolated Parallel BCD Output Board............................................4
1.4.5 Dual Relay Output and 4 Relay Output Board............................4
1.4.6 Isolated RS-232 Serial Communications Board ..........................4
1.4.7 Isolated RS-485 Serial Communications Board .........................5
1.5 Available Models and Options.....................................................5
Table of Contents
SECTION 2 SETUP
2.1 Unpacking ....................................................................................8
2.2 Safety Precautions ........................................................................10
2.2.1 Power Voltage...............................................................................10
2.2.2 Power Wiring................................................................................11
2.3 Assembly/Disassembly ..................................................................11
i
Table of
Contents
Page
2.3.1 Opening the Meter.......................................................................11
2.3.2 Checking and Installing Jumpers.................................................15
2.3.2.1 Main Board Jumpers .....................................................................16
2.3.2.2 Optional Input and Output Board Jumper Information .............18
2.3.3 Installing Optional Boards ...........................................................19
2.3.3.1 Isolated Pulse Input Board............................................................21
2.3.3.2 Isolated Analog Input Board........................................................22
2.3.3.3 Isolated Analog Output Board .....................................................23
2.3.3.4 Isolated Parallel BCD (Binary-Coded Decimal) Output Board ..24
2.3.3.5 Dual Relay and 4 Relay Output Board ........................................25
2.3.3.6 Isolated RS-232 Serial Communications Board ..........................25
2.3.3.7 Isolated RS-485 Serial Communications Board ..........................26
2.3.4 Reinserting the Main Board Assembly into the Case .................26
2.3.5 Wiring...........................................................................................26
2.3.5.1 P1 - AC Power Wiring.................................................................27
2.3.5.2 Battery Backup Connections........................................................27
2.3.5.3 P2 - Control Input/Output Wiring ..............................................28
2.3.5.4 Basic Meter Input Wiring.............................................................29
2.3.5.5 Isolated Pulse Input Board Wiring...............................................29
2.3.5.6 Isolated Analog Input Board Wiring ...........................................29
2.3.5.7 Isolated Analog Output Board Wiring.........................................29
2.3.5.8 Isolated Parallel BCD Output Board Wiring...............................29
2.3.5.9 Dual Relay and 4 Relay Output Board Wiring............................30
2.3.5.10 Isolated RS-232 or RS-485 Serial Comm. Board Wiring............30
2.4 Panel-Mount Assembly ................................................................30
SECTION 3 FRONT AND REAR FEATURES
3.1 Front-Panel Displays and Buttons................................................33
3.2 Rear Connectors...........................................................................35
SECTION 4 BASIC CONCEPTS AND APPROACHES
TO SETUP AND CONFIGURATION
4.1 Choosing an Operating Mode......................................................39
4.2 Mode-Associated Measurements..................................................39
4.3 Using Scale Factors ......................................................................40
4.4 Using Offsets.................................................................................40
4.5 Automatic Scale and Offset.........................................................40
4.6 Overflow Values and Exponential Format...................................41
4.7 Negative-True Logic.....................................................................41
4.8 Different Meter Modes.................................................................41
4.9 Setpoints.......................................................................................42
4.9.1 Setpoint Assignments...................................................................42
4.9.2 Configuring Setpoints ..................................................................44
ii
Page
4.10 Resets and Stops...........................................................................44
4.10.1 Power-On (Hard) Reset ...............................................................44
4.10.2 Configuration (Cold) Reset .........................................................44
4.10.3 Setpoint-Only Reset.....................................................................45
4.10.4 Rate and Sq Rt Mode Resets........................................................45
4.10.4.1 RESET-A ......................................................................................45
4.10.4.2 RESET-B.......................................................................................46
4.10.4.3 RESET-C.......................................................................................46
4.10.4.4 Front-Panel RESET......................................................................46
4.10.5 Batch Mode Resets.......................................................................47
4.10.5.1 RESET-A ......................................................................................47
4.10.5.2 RESET-B.......................................................................................47
4.10.5.3 RESET-C.......................................................................................47
4.10.5.4 Front-Panel RESET......................................................................48
4.10.5.5 Autoreset ......................................................................................48
4.10.6 STOP ............................................................................................48
4.11 Counting Up or Down .................................................................48
4.11.1 Counting Up.................................................................................49
4.11.2 Counting Down............................................................................50
4.12 An Example Showing Setpoint and Reset Control ....................52
Table of Contents
SECTION 5 CONFIGURING THE METER
5.1 Parameters, Choices, and Settings ...............................................54
5.2 The Basics of Configuration.........................................................55
5.3 Lockouts........................................................................................56
5.3.1 Lockouts and Corresponding Actions..........................................57
5.3.2 Setting Lockouts...........................................................................60
5.3.3 Removing the S1-A Jumper for Added Security.........................61
5.4 Other Meter Parameters...............................................................61
5.4.1 MODE (Operating Mode) ...........................................................61
5.4.2 CNFG 1 through CNFG 4 (Configuration Parameters)
Rate, Sq Rt, Batch........................................................................61
5.4.3 AVG.CNF (Running Average Configuration) Rate, Sq Rt........69
5.4.4 IN.SC.OF (Input Scale and Offset) Rate, Sq Rt, Batch .............70
5.4.5 OT.SC.OF (Output Scale and Offset) Rate, Sq Rt, Batch .........71
5.4.6 RTE DP (Rate Decimal Point) Rate, Sq Rt ................................73
5.4.7 RTE OF (Rate Offset) Rate, Sq Rt..............................................73
5.4.8 RTE SC (Rate Scale) Rate, Sq Rt...............................................74
5.4.9 TOT DP (Total Decimal Point) Rate, Sq Rt, Batch...................74
5.4.10 TOT OF (Total Offset) Rate, Sq Rt, Batch ................................75
5.4.11 TOT SC (Total Scale) Rate, Sq Rt, Batch..................................75
5.4.12 AL TI (Alarm Time) Rate, Sq Rt, Batch ....................................76
5.4.13 GATE T (Gate Time) Rate..........................................................76
iii
Table of
Contents
Page
5.4.14 BAUD, SER.CNF, DAT FT, BUS FT, ADDRES, SER TI (Baud Rate, Serial Configuration, Data Format, Bus Format,
Device Address, Serial Time) Rate, Sq Rt, Batch ......................77
5.4.15 SET TI (Set Time) Rate, Sq Rt, Batch ......................................80
5.4.16 BAT DP (Batch Decimal Point) Batch .......................................81
5.4.17 B LOAD (Batch Load) Batch......................................................81
5.4.18 BAT SC (Batch Scale) Batch ......................................................82
5.4.19 CAL VZ, CAL VS, CALmAZ, and CALmAS (Calibrate Voltage Zero, Calibrate Voltage Span,
Calibrate milliAmp Zero, Calibrate milliAmp Span).................82
5.5 Step-by-Step Programming Example 1:
Rate Mode; Shaft RPM from a Gear-Tooth Detector .................82
5.5.1 Determining the Rate Scale (RTE SC).......................................83
5.5.2 Configuring the Meter..................................................................83
5.6 Step-by-Step Programming Example 2: Linear Flow; Adding TOTAL to RATE
with Excitation from the Basic Meter..........................................85
5.6.1 Determining the Rate Decimal Point (RTE DP) and
Rate Scale (RTE SC)...................................................................85
5.6.2 Determining the Total Decimal Point (TOT DP) and
Total Scale (TOT SC) .................................................................86
5.6.3 Configuring the Meter..................................................................86
SECTION 6 OPERATING THE METER
6.1 Rate Mode ....................................................................................88
6.2 Sq Rt Mode...................................................................................89
6.3 Batch Mode ..................................................................................89
APPENDICES
A Basic Meter: General Information, Specifications,
Jumper Configuration, and Input Connections ..........................91
B Isolated Pulse Input Board: General Information,
Specifications, Jumper Configuration, and Wiring .....................97
C Isolated Analog Input Board: General Information,
Specifications, Jumper Configuration,
Wiring, and Calibration...............................................................107
D Isolated Analog Output Board: General Information,
Specifications, Wiring, and Calibration ......................................118
E Isolated Parallel BCD Output Board: General Information,
Specifications, Jumper Configuration, And Wiring....................122
F Dual Relay and 4 Relay Output Board: General Information,
Specifications, Jumper Configuration, and Wiring .....................128
iv
Page
G Isolated RS-232 and RS-485 Serial Communications Boards:
General Information and Specifications......................................131
H Error Messages ..............................................................................132
J Factory Preset Configuration Settings/ User Settings .................134
K Ramp Specifications.....................................................................147
L Specifications................................................................................149
Table of Contents
vv
Safety
Considerations
SAFETY CONSIDERATIONS
This device is marked with the international Caution symbol. It is important to read this
manual before installing or commissioning this device as it contains important information relating to Safety and EMC (Electromagnetic Compatibility).
Unpacking & Inspection
Unpack the instrument and inspect for obvious shipping damage. Do not attempt to operate the
Note
This instrument has no power-on switch. An external switch or circuit-breaker shall be included in the building installation as a disconnecting device. It shall be marked to indicate this function, and it shall be in close proximity to the equipment within easy reach of the operator. The switch or circuit-breaker shall not interrupt the Protective Conductor (Earth wire), and it shall meet the relevant requirements of IEC 947–1 and IEC 947-3 (International Electrotechnical Commission). The switch shall not be incorporated in the mains supply cord.
Furthermore, to provide protection against excessive energy being drawn from the mains supply in case of a fault in the equipment, an overcurrent protection device shall be installed.
unit if damage is found.
This instrument is a panel mount device protected in accordance with Class I of EN 61010 (115/230 AC power connections). Installation of this instrument should be done by Qualified personnel. In order to ensure safe operation, the following instructions should be followed.
Note
•The Protective Conductor must be connected for safety reasons. Check that the power cable has the proper Earth wire, and it is properly connected. It is not safe to operate this unit without the Protective Conductor Terminal connected.
• Do not exceed voltage rating on the label located on the top of the instrument housing.
• Always disconnect power before changing signal and power connections.
• Do not use this instrument on a work bench without its case for safety reasons.
• Do not operate this instrument in flammable or explosive atmospheres.
• Do not expose this instrument to rain or moisture.
• Unit mounting should allow for adequate ventilation to ensure instrument does not exceed operating temperature rating.
• Use electrical wires with adequate size to handle mechanical strain and power requirements. Install without exposing bare wire outside the connector to minimize electrical shock hazards.
EMC Considerations
•Whenever EMC is an issue, always use shielded cables.
•Never run signal and power wires in the same conduit.
•Use signal wire connections with twisted-pair cables.
•Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems persist.
vivi
l. Introduction
1.1 DESCRIPTION
The Counter Timer is a multifunctional panel meter that can measure counts, rates, flows, times, totals and batches from pulse and analog input sources over the range of 0.2 to 20,000 Hz.
The meter can function in four different modes and be applied to a wide range of applications, serving functions as simple as displaying incoming pulses–or as complex as detecting out-of-limits conditions, triggering alarms, and channeling communications for a process control system.
The front panel displays values and messages with six, 14-segment LEDs, indicates which mode-associated measurement is being displayed through three Measurement LEDs, and indicates alarm status through four Setpoint LEDs. Five buttons below the LED display allow easy front-panel configuration and access to the meter’s many features. These features can also be accessed through digital communications.
1
Introduction
The meter can be mounted in a panel or simply placed on a bench top.
1.2 FEATURES
The following list outlines the meter’s features.
• Six-digit display
• Microprocessor-based, with nonvolatile memory-no battery backup required
• Configurable via front-panel push buttons and/or through RS-232 or RS-485 ports
• High accuracy
• Large digital offset enabling easy scaling in engineering units
• Five open-collector outputs for RATE, TOTAL, BATCH count, BAT NO (Number of Batches), and TIME
• Programmable decimal point selection
• Read/Display/Output rates up to 25 per second
• Plug-in input and output optional boards
1
1
1.3 METER MODES
Introduction
The meter can be configured to operate in four different modes. These are Rate Meter/Totalizer (Rate), Rate Meter/Totalizer/ Square Root Extractor (Sq Rt), Batch Controller (Batch), and Clock (C).
1.3.1 RATE METER/TOTALIZER
As a rate meter/totalizer, the meter measures the rate, totalizes and displays pulses in any engineering units required, and provides unregulated sensor excitation. The meter can also provide running averages and process time in HH:MM:SS format in this mode.
With the addition of the Isolated Analog Input Board, analog signals can be received and displayed in the required engineering units.
1.3.2 RATE METER/TOTALIZER/SQUARE ROOT EXTRACTOR
As a rate meter/totalizer/Square Root Extractor, the meter can accept dc signals from any typical differential flow transmitter and extract the square root to provide highly accurate flow values. Digital calculation ensures accurate readings–which can be dis­played in any engineering units desired.
This mode requires the Isolated Analog Input Board.
1.3.3 BATCH CONTROLLER
As a batch controller, the meter can count the batch, number of batches completed, and the grand total of pulses received. In addition, it has an internal timer for process time displayed in HH:MM:SS format. Any of the above functions can be displayed during the process without interrupting or stopping the process.
1.3.4 CLOCK
As a clock, the meter can function in real, 24-hour time mode using the HH:MM:SS display format. It can also function as a 99-hour process timer. Time is derived from the power line frequency, 50 or 60 Hz, which ensures accuracy. If power fails and the meter is equipped with a backup battery, the clock function continues working (although the display will not be lit). Refer to Section 2.3.5.2 for information on battery backup.
2
1.4 OPTIONAL BOARDS OVERVIEW
1
The meter is designed to accommodate numerous optional boards.
These boards transform the meter into a single instrument that “can do it all.” On the input end, they allow the meter to accept signals from a wider variety of sources, and on the output end, to communicate with, or control a wide variety of other devices.
1.4.1 ISOLATED PULSE INPUT BOARD
The Isolated Pulse Input Board is a signal conditioning board that allows the meter to accept low- and high-level input signals from sources whose pulses may be too weak for the Main Board of the meter to handle, or that need selectable hysteresis and/or regulated excitation. This board is used mainly for weak magnetic pick-ups, high-level line voltage sources, and NAMUR sources.
If this board is used, the Isolated Analog Input Board cannot be used.
(See Appendix B for specifications, jumper configuration, wiring, and applications for the Isolated Pulse Input Board.)
1.4.2 ISOLATED ANALOG INPUT BOARD
Introduction
The Isolated Analog Input Board is a signal conditioning board that converts analog signals to frequency and allows the meter to accept input signals from non-pulsed or direct current sources. This board is often used with differential pressure flow transmitters to provide linear flow values. (First-time installation requires calibra­tion using calibration data on the back of the board.)
If this board is used, the Isolated Pulse Input Board cannot be used.
(See Appendix C for specifications, jumper configuration, wiring, and applications for the Isolated Analog Input Board.)
3
1
Introduction
1.4.3 ISOLATED ANALOG OUTPUT BOARD
The Isolated Analog Output Board converts display readings into voltage or current output. This board is often used as a control board in process applications. (First-time installation requires calibration using calibration data on the back-the solder or out­board side of the board.) (See Appendix D for specifications, jumper configuration, wiring, calibration, and applications for the Isolated Analog Output Board.)
1.4.4 ISOLATED PARALLEL BCD (BINARY-CODED DECIMAL) OUTPUT BOARD
The Isolated Parallel BCD Output Board produces binary-coded decimal output for direct communication with a printer or with an intelligent device such as a PLC (Programmable Logic Controller).
If this board is used, the Dual Relay Output Board cannot be used.
(See Appendix E for specifications, jumper configuration, wiring, and applications for the Isolated Parallel BCD Output Board.)
1.4.5 DUAL RELAY OUTPUT AND 4 RELAY OUTPUT BOARD
The Dual Relay Output and 4 Relay Output Board enables setpoint-triggered switching to an external device.
If either option board is used, the Isolated Parallel BCD Output Board cannot be used.
(See Appendix F for specifications, jumper configuration, wiring, and applications for the Relay Output Board.)
1.4.6 ISOLATED RS-232 SERIAL COMMUNICATIONS BOARD
The Isolated RS-232 Serial Communications Board provides an isolated digital communications channel between a single meter and a computer, serial printer, or other device.
If this board is used, the RS-485 Serial Communications Board cannot be used.
(See Appendix G for specifications, jumper configuration, wiring, and applications for the Isolated RS-232 Serial Communications Board.)
4
1.4.7 ISOLATED RS-485 SERIAL COMMUNICATIONS BOARD
The Isolated RS-485 Serial Communications Board provides an isolated digital communications channel where multiple meters (addressed from 0 to 199) can communicate with a single computer.
If this board is used, the RS-232 Serial Communications Board cannot be used.
(See Appendix G for specifications, jumper configuration, wiring, and applications for the Isolated RS-485 Serial Communications Board.)
1.5 AVAILABLE MODELS AND OPTIONS
The following models and options are available. Optional boards are either installed at the time of purchase, or available as separate items and installed by the user after purchase.
BASIC MODELS AND INPUT OPTIONS
MODEL NUMBER DESCRIPTION
DPF401 No optional input board; Basic Meter accepts low
level and high level pulses from TTL/CMOS sources, open collector outputs and contact closures; provides 14-20 V unregulated sensor excitation
1
Introduction
DPF402 Isolated Pulse Input Board. Single-input for magnetic
pick-ups with low- and high-level signals; direct connection to NPN, PNP, or NAMUR sensors. Includes electrically floating, regulated sensor excita­tion rated 12.6 V @ 70 mA or 8.2 V or 5 V
DPF403 Isolated Analog Input Board. Input ranges of
0-1 mA, 4-20 mA, 0-5 V, 1-5 V, and 0-10 V are jumper selectable. Standard factory default calibration: 4-20 mA dc = 0-10000
FOR OPTIONS ADD SUFFIX TO MODEL NUMBER
-GR Green LED display
-230V 230 V ac, power
5
1
Introduction
CONTROL/BCD OUTPUT OPTIONS
MODEL NUMBER DESCRIPTION
DP40-B Isolated Parallel BCD (Binary-Coded Decimal)
Output Board
DP40-R Dual Relay Output Board; Two 7-amp,
Form-C Relays
DP40-R4 4 Relay Output Board; Two 7-amp & two 1 amp
Form-C Relays.
NOTE: Choose only one Control/BCD output option per meter. A 40-socket plug is included with the BCD option.
ANALOG OUTPUT OPTIONS
MODEL NUMBER DESCRIPTION
DP40-A Isolated Analog Output Board: Configurable output;
0-20 mA, 4-20 mA, 0-5 V, 0-10 V. Standard factory default calibration: 0-10000 = 4-20 mA dc
DATA COMMUNICATIONS OPTIONS
MODEL NUMBER DESCRIPTION
DPF400-S2 Isolated RS-232 Serial Communications Board* DPF400-S4 Isolated RS-485 Serial Communications Board**
NOTE: Choose only one option per meter. Both computer communications come with a 6-foot communications cable with phone plug termination.
* We recommend purchase of 9SC2 or 25SC2 (see OPTIONS below) ** We recommend purchase of 9SC4 or 25SC4 (see OPTIONS below)
6
OPTIONS
MODEL NUMBER DESCRIPTION
DP40-9SC2 9-pin Serial Connector for RS-232 port DP40-9SC4 9-pin Serial Connector for RS-485 port DP40-25SC2 25-pin Serial Connector for RS-232 port DP40-25SC4 25-pin Serial Connector for RS-485 port
1
Introduction
7
2
Setup
2. Setup
2.1 UNPACKING
Unpack all items and make sure that every item on the packing list is present. The items you should receive are listed below. If something is missing, use the phone number for the Customer Service Department nearest you.
Also, inspect the shipping container and enclosed equipment for any signs of damage. Take particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
NOTE: The shipping agent will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save all packing material and containers in the event that reshipment is required.
You should receive the following items :
QTY DESCRIPTION ILLUSTRATION
1 Basic meter with
main board (and optional boards if ordered by number); in plastic case with sleeve, panel gasket, two thumbnuts
1 Front-panel
button cover available with return of the postcard (see meter box)
8
QTY DESCRIPTION ILLUSTRATION
1 Orange, 3-socket
power connector (P1); for AC input
1 Gray, 3-socket
input connector (P3); for signal input
1 Rear cover with
holddown screw
1 20-Socket Ribbon
Connector (P2)
L
2
1 2
N
3
1 2 3
Setup
2 Panel-mount gasket
(1 spare)
1 Operator’s manual
9
2.2 SAFETY PRECAUTIONS
2
Setup
The meter is protected in accordance with Class I of EN61010. Refer to Safety Considerations page.
WARNING: If your meter is to be wired to sensors or to control
!
2.2.1 POWER VOLTAGE
inputs that could have hazardous potentials, these potentials will be carried to the 20-pin output connector (P2) at the rear. They will also be present on the meter’s circuit boards. Follow all instructions carefully, inserting the electronics into the case and installing connec­tors BEFORE connecting the meter to any source of power.
DO NOT contact any exposed metal parts, install optional board(s), change jumpers, or in any way disassemble or assemble the meter while it is connected to AC voltage.
Note the following information and guidelines for safe operation of your meter:
Your power source voltage rating should agree with the voltage under which the meter has been configured to operate. The first thing you should do is verify this.
The meter’s operating voltage is shown in the VOLTS: entry of the ID and Serial Number Label. Figure 2-1 shows a copy of this label. It is located on the meter packing box and is clearly visible. (Another label is also affixed to the meter case.)
MODEL: SERIAL NO.: PART NO.: VOLTS: WATTS:
Figure 2-1. ID and Serial Number Label showing the Meter's operating voltage.
10
2.2.2 POWER WIRING
CAUTION: The meter has no power-on switch; it will be ON
!
when power is applied.
Section 2.3.5 shows you how to wire not only the AC power connector, but all other connectors as well.
2.3 ASSEMBLY/DISASSEMBLY
2.3.1 OPENING THE METER
Your meter is fully assembled, but not wired. In most cases, if you have ordered optional boards with the meter, these boards will already be installed.
You will need to remove only the rear cover to complete wiring, but you will have to open the meter to do one or more of the following:
WARNING: You must disconnect and turn-off the power and
!
connector signals before accessing the unit and installing optional boards. Failure to do so may result in injury!
2
Setup
a. Check or reconfigure the Transformer Jumpers on the Main
Board so that they correspond to your line voltage (W1 and W2 for 115 V ac, or W3 for 230 V ac). See Section 2.3.2.1.
b. Access jumpers on the Main and optional boards.
See Section 2.3.2.
c. Install optional boards. See Section 2.3.3.
11
2
Setup
REAR PROTECTIVE COVER
COVER MOUNTING SCREW
THUMBNUTS
GASKET
SIGNAL CONNECTOR BOARD
SLEEVE
AC POWER BOARD
MAIN BOARD ASSY
CASE
BEZEL
Figure 2-2. Exploded View of the Meter
12
Using Figures 2-2 as a guide, follow these simple instructions to open the meter:
2
!
IMPORTANT: Turn-off the power and input signals from the unit before proceeding. Failure to do so may result in injury!
1. Remove the cover mounting screw that secures the rear protective cover to the meter, and remove the Rear Protective Cover.
If you are simply wiring the meter–but not checking jumpers or
installing or removing boards–this is as far as the meter needs to be disassembled. Go to Section 2.3.5.
2. Remove all wiring connectors from the rear of the meter.
3. Remove the two thumbnuts that secure the case to the sleeve.
4. Remove the sleeve completely by sliding it back from the front bezel.
Setup
13
l. Introduction
1.1 DESCRIPTION
The Counter Timer is a multifunctional panel meter that can measure counts, rates, flows, times, totals and batches from pulse and analog input sources over the range of 0.2 to 20,000 Hz.
The meter can function in four different modes and be applied to a wide range of applications, serving functions as simple as displaying incoming pulses–or as complex as detecting out-of-limits conditions, triggering alarms, and channeling communications for a process control system.
The front panel displays values and messages with six, 14-segment LEDs, indicates which mode-associated measurement is being displayed through three Measurement LEDs, and indicates alarm status through four Setpoint LEDs. Five buttons below the LED display allow easy front-panel configuration and access to the meter’s many features. These features can also be accessed through digital communications.
1
Introduction
The meter can be mounted in a panel or simply placed on a bench top.
1.2 FEATURES
The following list outlines the meter’s features.
• Six-digit display
• Microprocessor-based, with nonvolatile memory-no battery backup required
• Configurable via front-panel push buttons and/or through RS-232 or RS-485 ports
• High accuracy
• Large digital offset enabling easy scaling in engineering units
• Five open-collector outputs for RATE, TOTAL, BATCH count, BAT NO (Number of Batches), and TIME
• Programmable decimal point selection
• Read/Display/Output rates up to 25 per second
• Plug-in input and output optional boards
1
1
1.3 METER MODES
Introduction
The meter can be configured to operate in four different modes. These are Rate Meter/Totalizer (Rate), Rate Meter/Totalizer/ Square Root Extractor (Sq Rt), Batch Controller (Batch), and Clock (C).
1.3.1 RATE METER/TOTALIZER
As a rate meter/totalizer, the meter measures the rate, totalizes and displays pulses in any engineering units required, and provides unregulated sensor excitation. The meter can also provide running averages and process time in HH:MM:SS format in this mode.
With the addition of the Isolated Analog Input Board, analog signals can be received and displayed in the required engineering units.
1.3.2 RATE METER/TOTALIZER/SQUARE ROOT EXTRACTOR
As a rate meter/totalizer/Square Root Extractor, the meter can accept dc signals from any typical differential flow transmitter and extract the square root to provide highly accurate flow values. Digital calculation ensures accurate readings–which can be dis­played in any engineering units desired.
This mode requires the Isolated Analog Input Board.
1.3.3 BATCH CONTROLLER
As a batch controller, the meter can count the batch, number of batches completed, and the grand total of pulses received. In addition, it has an internal timer for process time displayed in HH:MM:SS format. Any of the above functions can be displayed during the process without interrupting or stopping the process.
1.3.4 CLOCK
As a clock, the meter can function in real, 24-hour time mode using the HH:MM:SS display format. It can also function as a 99-hour process timer. Time is derived from the power line frequency, 50 or 60 Hz, which ensures accuracy. If power fails and the meter is equipped with a backup battery, the clock function continues working (although the display will not be lit). Refer to Section 2.3.5.2 for information on battery backup.
2
1.4 OPTIONAL BOARDS OVERVIEW
1
The meter is designed to accommodate numerous optional boards.
These boards transform the meter into a single instrument that “can do it all.” On the input end, they allow the meter to accept signals from a wider variety of sources, and on the output end, to communicate with, or control a wide variety of other devices.
1.4.1 ISOLATED PULSE INPUT BOARD
The Isolated Pulse Input Board is a signal conditioning board that allows the meter to accept low- and high-level input signals from sources whose pulses may be too weak for the Main Board of the meter to handle, or that need selectable hysteresis and/or regulated excitation. This board is used mainly for weak magnetic pick-ups, high-level line voltage sources, and NAMUR sources.
If this board is used, the Isolated Analog Input Board cannot be used.
(See Appendix B for specifications, jumper configuration, wiring, and applications for the Isolated Pulse Input Board.)
1.4.2 ISOLATED ANALOG INPUT BOARD
Introduction
The Isolated Analog Input Board is a signal conditioning board that converts analog signals to frequency and allows the meter to accept input signals from non-pulsed or direct current sources. This board is often used with differential pressure flow transmitters to provide linear flow values. (First-time installation requires calibra­tion using calibration data on the back of the board.)
If this board is used, the Isolated Pulse Input Board cannot be used.
(See Appendix C for specifications, jumper configuration, wiring, and applications for the Isolated Analog Input Board.)
3
1
Introduction
1.4.3 ISOLATED ANALOG OUTPUT BOARD
The Isolated Analog Output Board converts display readings into voltage or current output. This board is often used as a control board in process applications. (First-time installation requires calibration using calibration data on the back-the solder or out­board side of the board.) (See Appendix D for specifications, jumper configuration, wiring, calibration, and applications for the Isolated Analog Output Board.)
1.4.4 ISOLATED PARALLEL BCD (BINARY-CODED DECIMAL) OUTPUT BOARD
The Isolated Parallel BCD Output Board produces binary-coded decimal output for direct communication with a printer or with an intelligent device such as a PLC (Programmable Logic Controller).
If this board is used, the Dual Relay Output Board cannot be used.
(See Appendix E for specifications, jumper configuration, wiring, and applications for the Isolated Parallel BCD Output Board.)
1.4.5 DUAL RELAY OUTPUT AND 4 RELAY OUTPUT BOARD
The Dual Relay Output and 4 Relay Output Board enables setpoint-triggered switching to an external device.
If either option board is used, the Isolated Parallel BCD Output Board cannot be used.
(See Appendix F for specifications, jumper configuration, wiring, and applications for the Relay Output Board.)
1.4.6 ISOLATED RS-232 SERIAL COMMUNICATIONS BOARD
The Isolated RS-232 Serial Communications Board provides an isolated digital communications channel between a single meter and a computer, serial printer, or other device.
If this board is used, the RS-485 Serial Communications Board cannot be used.
(See Appendix G for specifications, jumper configuration, wiring, and applications for the Isolated RS-232 Serial Communications Board.)
4
1.4.7 ISOLATED RS-485 SERIAL COMMUNICATIONS BOARD
The Isolated RS-485 Serial Communications Board provides an isolated digital communications channel where multiple meters (addressed from 0 to 199) can communicate with a single computer.
If this board is used, the RS-232 Serial Communications Board cannot be used.
(See Appendix G for specifications, jumper configuration, wiring, and applications for the Isolated RS-485 Serial Communications Board.)
1.5 AVAILABLE MODELS AND OPTIONS
The following models and options are available. Optional boards are either installed at the time of purchase, or available as separate items and installed by the user after purchase.
BASIC MODELS AND INPUT OPTIONS
MODEL NUMBER DESCRIPTION
DPF401 No optional input board; Basic Meter accepts low
level and high level pulses from TTL/CMOS sources, open collector outputs and contact closures; provides 14-20 V unregulated sensor excitation
1
Introduction
DPF402 Isolated Pulse Input Board. Single-input for magnetic
pick-ups with low- and high-level signals; direct connection to NPN, PNP, or NAMUR sensors. Includes electrically floating, regulated sensor excita­tion rated 12.6 V @ 70 mA or 8.2 V or 5 V
DPF403 Isolated Analog Input Board. Input ranges of
0-1 mA, 4-20 mA, 0-5 V, 1-5 V, and 0-10 V are jumper selectable. Standard factory default calibration: 4-20 mA dc = 0-10000
FOR OPTIONS ADD SUFFIX TO MODEL NUMBER
-GR Green LED display
-230V 230 V ac, power
5
1
Introduction
CONTROL/BCD OUTPUT OPTIONS
MODEL NUMBER DESCRIPTION
DP40-B Isolated Parallel BCD (Binary-Coded Decimal)
Output Board
DP40-R Dual Relay Output Board; Two 7-amp,
Form-C Relays
DP40-R4 4 Relay Output Board; Two 7-amp & two 1 amp
Form-C Relays.
NOTE: Choose only one Control/BCD output option per meter. A 40-socket plug is included with the BCD option.
ANALOG OUTPUT OPTIONS
MODEL NUMBER DESCRIPTION
DP40-A Isolated Analog Output Board: Configurable output;
0-20 mA, 4-20 mA, 0-5 V, 0-10 V. Standard factory default calibration: 0-10000 = 4-20 mA dc
DATA COMMUNICATIONS OPTIONS
MODEL NUMBER DESCRIPTION
DPF400-S2 Isolated RS-232 Serial Communications Board* DPF400-S4 Isolated RS-485 Serial Communications Board**
NOTE: Choose only one option per meter. Both computer communications come with a 6-foot communications cable with phone plug termination.
* We recommend purchase of 9SC2 or 25SC2 (see OPTIONS below) ** We recommend purchase of 9SC4 or 25SC4 (see OPTIONS below)
6
OPTIONS
MODEL NUMBER DESCRIPTION
DP40-9SC2 9-pin Serial Connector for RS-232 port DP40-9SC4 9-pin Serial Connector for RS-485 port DP40-25SC2 25-pin Serial Connector for RS-232 port DP40-25SC4 25-pin Serial Connector for RS-485 port
1
Introduction
7
2
6. Using Figure 2-4 as a guide, bend the side-panel detents on the case outward and pull the board assembly out of the case by the mounting screw stem.
Setup
These six steps are known as “accessing the Main Board assembly.”
ID AND SERIAL NUMBER LABEL
SIGNAL INPUT JUMPER LABEL
Figure 2-4. Removing the Main Board Assembly from the Case
TO INSTALL
TO REMOVE
BEND DETENTS OUTWARD TO INSTALL MAIN BOARD
CASE
MOUNTING SCREW STEM
MAIN BOARD ASSY
The meter is now disassembled to the point where you can check and configure jumpers and install boards.
NOTE: When the sleeve is removed from the case in Step 5 above, the Connector Label on the case will be exposed. See Figure 2-5. Use this label for reference when wiring or con­necting plugs.
14
P2/CABLE CONNECTOR
ALARM 1 OUTPUT
ALARM 3 OUTPUT
ALARM, BATTERY RETURN
P4/RS-232
6 5 4 3 2 1
P1/AC PWR
1 2 3
TWICE LINE FREQ
1 1
N/C COMM RTN RX TX RTS N/C
{
LINE NEUTRAL AC GND
RESET-A
RESET-B
TEST TX
PULSE OUTPUT
RESET-C
STOP
P4/RS-485
6 5 4 3 2 1
P6/RELAY
1 2 3
1 3 5 7 9
10
11
12
13
14
15
16
17
18
19
20
N/C B RX A RX COMM RTN B TX/RX A TX/RX
NO1 CM1 NC1
ALARM V + INPUT
2
ALARM 2 OUTPUT
4
ALARM 4 OUTPUT
6
ALARM 5 OUTPUT
8
DEBOUNCE TEST RX DIGITAL RETURN HOLD / PRINT REQ GATE INPUT BACKUP BATTERY +
P5 ANALOG OUT
P7/RELAY
1 2 3
1 2 3
1
2
Setup
1
RETURN 4-20 mA 0-10 V
NO2 CM2 NC2
BCD
P5
1
.
.
.
P18 for 4 RELAY
1
P8
...
...
1
P6
P2
1
.
..........
.
..........
P1
.
1
1
CONNECT TO LOW VOLTAGE LIMITED ENERGY CIRCUITRY ONLY.
P8
1
P7
P4
. . . .
1
.
19
Figure 2-5. Connector Label
2.3.2 CHECKING AND INSTALLING JUMPERS
This section contains figures and instructions for checking and installing jumpers, but it gives full information on the Main Board only. If you have any optional boards, refer to the appropriate appendix for specific jumper information.
NOTE: When referring to jumpers and the corresponding boards, the view is from the REAR of the meter.
P9
P3
.
1
.
.
.
1
.
.
PULSE INPUT
P9 & P3
FILTER SIG IN HI LEVEL PULL UP/DOWN
EXCITATION SIG IN LO LEVEL SIG RETURN
ANALOG INPUT
+24 V EXC SIGNAL IN RETURN
TEST TEST TEST
1
15
Loading...
+ 140 hidden pages