Omega Products CT6101 User Manual

Microprocessor-Based Circular Chart Recorder
CT6100 Series
Starts at
$
1630
Accepts Thermocouple,
RTD, Vdc, and mA DC Inputs
High Accuracy
and Stability
User Configurable via
Front-Panel Keypad
Menu-Driven Interface for Rapid Configuration and Calibration
Available in 1-, 2-, or
3-Pen Versions
Large-Character
Alphanumeric LCD with Backlight
Simultaneous Digital
Display of Process Variable for Each Channel
Programmable
Locks for Security
CT6103, $2312, shown smaller than actual size.
The CT6100 Series microprocessor­based circular chart recorders are available in 1-, 2-, or 3-channel variants. They combine the simplicity and clarity of pen drawing with the versatility of microprocessor control. Each channel is compatible with all industry-standard sensors and signals, including thermocouple Types J, K, T, E, N, R, S, and B; Pt100 platinum RTDs; and 4 or 20 mA current loops.
Low and high measurement ranges are provided for each input type; separate range cards are not required. Multi-input versions feature optoelectronic isolation of the input stages to eliminate troublesome installation ground loops. Multislope integrating 16-bit A/D converters ensure precise measurement by sampling the input every 0.5 seconds.
Thermocouple and RTD characteristics are fully linearized. The recorders use automatic cold-junction compensation for thermocouple measurement.
All CT6100 Series models feature alarm relays. Single-pen recorders are equipped with 3 fail-safe single­pole changeover relays; 6 relays are standard on the 2- and 3-pen versions. Color-coded LEDs indicate the status of each relay. All relay functions are user selectable; setpoint values and hysteresis levels are entered directly via the keypad, while relay action and channel assignment are selected from user-friendly menus. Circuit precision is matched by the backlash-free pen-drive mechanism, which has a positioning resolution greater than 0.1%. An integral
S-45
feedback potentiometer enables closed-loop monitoring of each pen position.
Each recorder uses a low-maintenance stepper motor. The rotation speed of the 244 mm diameter chart is microprocessor controlled and user programmable.
All CT6100 recorders are housed in a strong molded case that can be panel or surface mounted. A gasket-lockable door protects internal components from harsh industrial environments and offers protection rated to NEMA 12 (IP55). A tough acrylic window lets the user view the chart trace, digital channel readings, and alarm status with the recorder's door closed.
D
Extended Warranty
Program
SM
o
n
’t F
o
rg
e
to
O
rd
e
r A
R
e
c
o
rd
e
t
d
d
itio
n
a
r
P
l
a
p
e
r!
Specifications
Inputs
Input Types: Thermocouple K, T,
J, N, E, B, R, S; Platinum RTD (Pt100) 3-wire; ±2V, +20 Vdc; ±2 mA, ±20 mA
Input Ranges: Thermocouples to BS4937 (type B minimum temp 200°C); RTD to BS1904:1984
Cold Junction Compensation:
Automatic; ±0.02°C/°C stability
Input Resistance:
Thermocouple: 10 M ±2 mA: 200 ±20 mA: 20 ±2V, ±20V: >1 M
Min Span: 5°C (9°F), thermocouples K,
J, T, E, N and RTD; 50°C (90°F) other thermocouple types, 50 mV, 200 mA
T/C BURN-OUT: Pull-up or pull-down, selectable
RTD Lead Resistance:
3-wire, compensated up to 10 max per lead
Input Protection: ±50 Vdc on signal inputs
Input Isolation: Optoelectronic on 2 and 3 input models 500 Vrms channel­to-channel, 500 Vrms channel-to-ground
Sensor/Transmitter Power: 12 or 24 Vdc, 25 mA, selectable; for current loop inputs
Performance Accuracy:
RTD, Low Range (-200 to 200°C):
±0.2°C
RTD, High Range (200 to 850°C):
±0.8°C
Thermocouple: ±0.25% FS Voltage/Current: ±0.2% FS
Temperature Stability:
±0.02% FS/°C
Linearization Accuracy:
T/C Types J, K, T, N, E: ±0.1°C,
-50 to 200°C, ±1°C max T/C Types R and S: ±0.2°C,
-50 to 200°C, ±1°C max
T/C Type B: ±1°C max RTD: ±0.1°C, -200 to 850°C
Calibration Shift: ±10°C user
programmable to eliminate sensor errors (T/C and RTD)
hart and Display
C
Chart Size: 244 mm (9.6") dia; 40, 50, 60, 70 or 80 linear divisions
Chart Drive: DC stepper motor Chart Speeds: 1 to 24 hours in steps of
one hour, 2 to 31 days in steps of 1 day Writing Method: Disposable ink
cartridges; pen 1-red; pen 2-green; pen 3-blue
Pen Positioner: DC stepper motor Positioning Resolution: >0.1% FS
esponse Time:
R
Zero to full scale in 4.5 sec Pen Lift: Powered, activated from front
panel; chart fast time advance possible with pens raised
Display: 2 line x 20 character dot matrix liquid crystal with backlight and automatic temperature compensation;
9.6 mm (0.37") character height
Display Resolution:
0.1°C for temperature inputs, software programmable for voltage/ current inputs
Alarm Display: Relay status shown by red and green front panel LEDs
Relays
Type: SPDT; 30 Vdc or 250 Vac @ 6 A max; 150 W DC, 1660 Vac non-inductive
Action: Software selectable from: high alarm/low alarm/deviation alarm/control relay (high)/control relay (low); relays de-energize in alarm state; assignable to any channel
Hysteresis: User programmable 0.0% to 10% span
Snubber Network: 22 nF and 100 across each contact
Analog Output (Option)
ype: 0 to 20/4 to 20 mA; assignable to
T
any channel; range selectable within any segment of chart range
Resolution: 12-bit Compliance: 20V approx Isolation: Optoelectronic Action: direct or reverse
eneral
G
Security: 3-level software lock including password protection, internal hardware jumper lock, lockable
oor
d Power: 115 or 230 Vac ±10%,
switch selectable, 50/60Hz: terminal block connection
Power Requirements: <25 W Operating Ambient: 0 to 55°C
(32 to 131°F), 0 to 90% RH (non condensing)
Case: Steel, with glass-filled polyester resin door with acrylic window
Protection: NEMA 12 (IP55) Mounting: Panel or surface Dimensions: 336 W x 396 H x 171 mm D
(13.2 x 15.6 x 6.7") Weight: From single pen: 7 kg
(15.4 lb) to three pen: 7.7 kg (17.0 lb) Panel Cutout: 288 W x 356 mm H
(11.3 x 14.0")
OMEGACARESMextended warranty program
is available for models shown on this page.
Ask your sales representative for full details
when placing an order. OMEGACARE
covers parts, labor and equivalent loaners.
AVAILABLE FOR FAST DELIVERY!
To Order (Specify Model Number)
Model No. Price Description CT6101 $1630 1-pen recorder with 3 relays and transmitter power supply CT6102 1850 2-pen recorder with 6 relays and transmitter power supply CT6103 2312 3-pen recorder with 6 relays and transmitter power supply
Accessories
Model No. Price Description CT6100-RED $17 Red pens, pack of 3 CT6100-GREEN 17 Green pens, pack of 3 CT6100-BLUE 17 Blue pens, pack of 3 CT6100-0-100/24H 25 100 chart papers, 24 hours CT6100-0-100-31D 25 100 chart papers, 31 days POWERCORD-SE 7 Power cord with one end stripped
Option*
Order Suffix Add’l Price Description
-PV $140
Option is not field installable. It must be ordered at the time of purchase.
*
Comes complete with 1 package of 24 hour chart paper, pen(s) and operator’s manual.
Ordering Example: CT6101, 1-pen recorder with 3 relays and transmitter power supply and POWERCORD-SE, power cord, $1630 + 7 = $1637.
OCW-3, OMEGACARESMextends standard 1-year warranty to a total of 4 years ($350),
$1630 + 350 = $1980.
Recommended Reference Book:
See Section Y for Additional Books
S-46
Analog output of 0 to 20 mA or 4 to 20 mA, assignable to any channel
Statisical Quality Control Methods,
ST-0609, $75
SM
S
One Omega Drive | Stamford, CT 06907 | 1-888-TC-OMEGA (1-888-826-6342) | info@omega.com
EPG05
www.omega.com
UNITED KINGDOM
www. omega.co.uk
Manchester, England
0800-488-488
UNITED STATES
www.omega.com
1-800-TC-OMEGA
Stamford, CT.
CANADA
www.omega.ca
Laval(Quebec)
1-800-TC-OMEGA
GERMANY
www.omega.de
Deckenpfronn, Germany
0800-8266342
Karviná, Czech Republic
FRANCE
www.omega.fr
Guyancourt, France
088-466-342
CZECH REPUBLIC
www.omegaeng.cz
596-311-899
BENELUX
www.omega.nl
Amstelveen, NL 0800-099-33-44
More than 100,000 Products Available!
Temperature
Calibrators, Connectors, General Test and Measurement Instruments, Glass Bulb Thermometers, Handheld Instruments for Temperature Measurement, Ice Point References, Indicating Labels, Crayons, Cements and Lacquers, Infrared Temperature Measurement Instruments, Recorders Relative Humidity Measurement Instruments, RTD Probes, Elements and Assemblies, Temperature & Process Meters, Timers and Counters, Temperature and Process Controllers and Power Switching Devices, Thermistor Elements, Probes and Assemblies,Thermocouples Thermowells and Head and Well Assemblies, Transmitters, Wire
Flow and Level
Air Velocity Indicators, Doppler Flowmeters, Level Measurement, Magnetic Flowmeters, Mass Flowmeters, Pitot Tubes, Pumps, Rotameters, Turbine and Paddle Wheel Flowmeters, Ultrasonic Flowmeters, Valves, Variable Area Flowmeters, Vortex Shedding Flowmeters
pH and Conductivity
Conductivity Instrumentation, Dissolved Oxygen Instrumentation, Environmental Instrumentation, pH Electrodes and Instruments, Water and Soil Analysis Instrumentation
Data Acquisition
Auto-Dialers and Alarm Monitoring Systems, Communication Products and Converters, Data Acquisition and Analysis Software, Data Loggers Plug-in Cards, Signal Conditioners, USB, RS232, RS485 and Parallel Port Data Acquisition Systems, Wireless Transmitters and Receivers
Pressure, Strain and Force
Displacement Transducers, Dynamic Measurement Force Sensors, Instrumentation for Pressure and Strain Measurements, Load Cells, Pressure Gauges, Pressure Reference Section, Pressure Switches, Pressure Transducers, Proximity Transducers, Regulators, Strain Gages, Torque Transducers, Valves
Heaters
Band Heaters, Cartridge Heaters, Circulation Heaters, Comfort Heaters, Controllers, Meters and Switching Devices, Flexible Heaters, General Test and Measurement Instruments, Heater Hook-up Wire, Heating Cable Systems, Immersion Heaters, Process Air and Duct, Heaters, Radiant Heaters, Strip Heaters, Tubular Heaters
click here to go to the omega.com home page